Upload
others
View
23
Download
5
Embed Size (px)
Citation preview
AP 5301/8301Instrumental Methods of Analysis
and Laboratory
Lecture 8
Secondary ion mass spectrometry (SIMS)
Prof YU Kin Man
E-mail: [email protected]
Tel: 3442-7813
Office: P6422
1
Lecture 9: outline Introduction: general features of secondary ion mass spectrometry
SIMS theory:─ Ion-solid interactions
─ Sputtering process
─ Ion yield
─ Quantification: relative sensitivity factor
Instrumentation:─ Ion sources
─ Mass spectrometer
─ Ion detector
─ Time-of-flight SIMS
Common modes of SIMS:─ Static SIMS
─ Dynamic SIMS
Depth profiling:─ Crater effects
─ Depth resolution
Strengths and weaknesses
2
The technique involves bombarding the surface of a sample with a beam of
ions, thus emitting secondary ions. These ions are later measured with a
mass spectrometer to determine either the elemental or isotopic composition of
the surface of the sample.
Secondary ion mass spectrometry (SIMS)
A well established analytical technique that was first pioneered in 1949
SIMS is generally used for
surface, bulk, microanalysis,
depth profiling, and
impurity analysis.
Primary ion beam
(O-, O2+, Ar+, Cs+, Ga+ are often
used with energies between 1
and 30 keV)
Primary ions are implanted and
mix with sample atoms to depths
of 1 to 10 nm.
The bombarding primary ion beam produces monatomic and polyatomic particles
of sample material and re-sputtered primary ions, along with electrons and
photons. The secondary particles carry negative, positive, and neutral charges
and they have kinetic energies that range from zero to several hundred eV.
http://atomika.com/
3
SIMS analysis4
Secondary ion mass spectrometry (SIMS) is a technique used to analyze the
composition of solid surfaces and thin films by sputtering the surface of the
specimen with a focused primary ion beam and collecting and analyzing
ejected secondary ions with a mass spectrometer to determine the elemental,
isotopic, or molecular composition of the surface to a depth of 1 to 2 nm.
SIMS analysis5
Cameca IMS 6f secondary ion mass spectrometer6
7
SIMS: comparison with other techniques
8
Ion-solid interaction
Cs+, O2+, Ar+ and Ga+
at energies ~ 1-30 keV
negative, positive, and neutral
charges with kinetic energies ranging
from zero to a few hundred eV.
Energy is transferred from the energetic primary ions to atoms in the sample.
Some of these atoms receive enough energy to escape the sample
Sputtered species:
Monatomic and
polyatomic particles of
sample material (+ve,
-ve or neutral)
Re-sputtered primary
species (+ve, -ve or
neutral)
Electrons
photons
Sputtering9
Sputtering is a process whereby particles are
ejected from a solid target material due to
bombardment of the target by energetic particles.
The kinetic energy of the incoming particles is
typically hundreds eV to keV, leading erosion of the
target materials.
Sputtering is commonly used as a
tool for thin film deposition
Eroding material from a target
source onto a substrate using a
gaseous plasma (Ar)
targets
substrate
For thin film analysis:
Mass analyze the sputtered ejected
ions─SIMS
To expose atoms underneath the
surface for analysis─depth profiling
The Sputtering Process10
Sputter rates in typical SIMS experiments vary between 0.5 and 5 nm/s.
Sputter rates depend on sputter yield, which in turn depends on the primary
beam species, energy, intensity, sample material, and crystal orientation.
Sputter yield: ratio of number of atoms sputtered
to number of impinging ions, typically 5-15
─ Commonly in SIMS, oxygen or cesium is used as a
primary ion source, which chemically changes the
surface and the sputter rate.
Sputter yields of silicon as a function of ion energy for
noble gas ions at normal incidence.
The variation of the sputter yield with angle
for the three metals. Below approximately 60
degrees, the sputter rate increases with
angle before passing through a maximum
Secondary ion yield11
Secondary ion current of species 𝑚
𝐼𝑠𝑚 = 𝐼𝑝𝑦𝑚𝛼+𝜃𝑚𝜂
𝐼𝑝 = primary particle flux
𝑦𝑚 = sputter yield
𝛼+ = ionization probability to positive ions
𝜃𝑚 = fractional concentration of m in the layer
𝜂 = transmission of the analysis systemIon yield is influenced by
─ Matrix effects
─ Surface coverage of reactive elements
─ Background pressure
─ Orientation of crystallographic axes with respect to the sample surface
─ Angle of emission of detected secondary ions
The number of secondary particles (atoms/ions) emitted by the surface for each
impinging primary ion is defined as sputtering yield and can range between 5
and 15. The fraction of ionized emitted particles is called secondary ion yield
and ranges typically between 10-4 to 10-6.
In SIMS, it is the secondary ions that are eventually detected by the mass
spectrometer
Secondary ion yields: primary ion beams
Oxygen works as a medium which strips off electrons from the speeding
sputtered atoms when they leave surface, while Cesium prefers to load an
electron on the sputtered atoms.
𝑶𝟐+ ions beam:
─ Oxygen tends to bind with metal (Me)
atoms, if present in the sample.
─ During secondary emission the Me-O
bonds break thus generating 𝑀𝑒𝑛+
𝑪𝒔+ ions beam:
─ The implanted Cs ions lower the
sample work function
─ More secondary electrons are excited
over the surface potential barrier
─ Increased availability of electrons
leads to increased negative ion
formation especially for elements with
high electron affinity.
12
Secondary ion yield depends critically on the primary ion beam species. Typically
𝐶𝑠+and 𝑂2+ ion beams are used in SIMS measurements.
Selection of primary ions:
Inert gas (Ar, Xe, etc.)
─ Minimize chemical modification
Oxygen
─ Enhance positive ions
Cesium
─ Enhance negative ions
Liquid metal (Ga)
─ Small spot for enhanced
lateral resolution
Secondary ion yields: primary ion beam Oxygen bombardment increases the yield of positive ions
Cesium bombardment increases the yield of negative ions.
The increases can range up to four orders of magnitude.
13
Relative secondary ion yield14
10 20 30 40 50 60 70 80 90 100
Atomic number
106
105
104
103
102
108
107
106
105
104
103
102
Rela
tive s
econdary
positiv
e ion y
ield
Rela
tive s
econdary
negative ion y
ield
0 10 20 30 40 50 60 70 80 90
Atomic number
16.5 keV Cs+ 13.5 keV O-
One of the main obstacles preventing the derivation of a universal theory of the
secondary ion emission is a fact that the secondary ion yield of any chemical
element strongly depends on its chemical environment─matrix effect.
This may cause variations in the ion yield over several orders of the
magnitudes, from one matrix to another.
For example, yields of Al+ ions from Al2O3 and Al metal differ by a factor of 100;
Si+ ion emission from SiO2 is 2500x higher than that from Si.
Quantification in SIMS
The SIMS signal intensity for a particular element M (𝐼𝑀) is related to its
concentration in the analyzable layer (𝐶𝑀) by several parameters:
𝐼𝑀 = 𝐽𝑝𝐴𝑆𝛽𝑀𝑇𝐶𝑀
𝐽𝑝 = primary ion current
𝐴 = analyzed surface area
𝑆 = sputtering yield
𝛽𝑀 = secondary ion yield for element M
𝑇 = transmission of SIMS spectrometer
Since many of these parameters are not known, an approach based on
relative sensitivity factors is adopted in SIMS to evaluate atomic
concentrations of minor constituents when that of the major constituent is
known.
15
Relative sensitivity factor (RSF)16
For absolute quantification using SIMS, standards as similar as possible to
the real sample are needed. It is typical to use implanted samples as
standards.
For example for the concentration profile of an impurity 𝑖 (𝐶𝑖) in a matrix
(𝑚𝑎𝑡), a standard with a known dose (𝐷 𝑖𝑛 𝑎𝑡𝑜𝑚𝑠/𝑐𝑚2) of 𝑖 in the same
matrix is created. So that the relative sensitivity factor
𝑅𝑆𝐹 = 𝐷𝐶𝐼𝑚𝑎𝑡𝑡
𝑧𝐼𝑖 𝑠𝑡𝑑
,
where 𝐶 is the number of data cycles, 𝐼𝑚 is the matrix element secondary ion
intensity (counts/sec), 𝑡 is the count time/cycle, 𝑧 is the depth of the crater, 𝐼𝑖 is the
summation of secondary ion intensity of 𝑖 in counts.
𝐶𝑖 = 𝐼𝑖
𝐼𝑚𝑎𝑡 𝑠𝑎𝑚𝑝𝑙𝑒
∙ 𝑅𝑆𝐹
Implanted standards have the advantages of:
─ Any element (isotope) can be implanted into any matrix
─ Depth and peak concentration can be tuned by the energy and the dose
─ Multiple element can be implanted
─ A detection limit can be established
SIMS: ion implanted standards17
The procedure is based on the exposure, for a controlled time, of the
matrix to a beam of primary ions of the element of interest.
The primary ion energy usually ranges between 50 and 300 keV, whereas
the dose is about 1013-1016 ions/cm2.
After implantation the sample is analyzed under Dynamic SIMS conditions
and the element signal is monitored as a function of time (e.g. of depth
reached due to erosion).
After the SIMS measurement
the crater is measured to reveal
the real depth
the implantation dose/crater
depth ratio provides an estimate
for the average atomic
concentration (atoms/cm3) of
the element in the matrix
A RSF can be established
𝑅𝑆𝐹 = 𝐷𝐶𝐼𝑚𝑎𝑡𝑡
𝑧𝐼𝑖 𝑠𝑡𝑑
SIMS: instrumentation
SIMS CAMECA 6F
Ion Sources
Ion sources with electron impact
ionization - Duoplasmatron: Ar+,
O2+, O-
Ion sources with surface ionization -
Cs+ ion sources
Ion sources with field emission -
Ga+ liquid metal ion sources
Mass Analyzers
Magnetic sector analyzer
Quadrupole mass analyzer
Time of flight analyzer
Ion Detectors
Faraday cup
Dynode electron multiplier
Vacuum < 10−6 torr
18
Schematic Diagram of a SIMS instrument19
20
Ion source: DuoplasmatronA duoplasmatron is an ion source with electron impact ionization
A cathode filament emits electrons into a vacuum chamber
Small quantity of gas (Ar, O2, Ne, etc.) leaks into the chamber and interacts
with the electrons forming a plasma
The plasma is accelerated through a series of highly charged grids to the
desired energy and extracted through the aperture.
It can operate with almost
any gas
When O2 is used, O-, O2- or
O2+ can be extracted
depending on the electrical
polarity selected
Probe diameter typically
between 5 mm to 1 mm
Ion current densities >10
mA/cm2
Ion source: Cs+ source21
The Cs atoms are ionized
during evaporation because
the work function of W (4.52
eV) is substantially greater
then the ionization potential of
Cs (3.88 eV)
The Cs+ ions are extracted and
accelerated to an energy up to
10 keV.
Depending on the gun design,
fine focus or high current can
be obtained.
Cs gun is typically more
expensive to operate
Cs metal (or compound) is heated in the reservoir (~400oC) forming a vapor
The Cs vapor flows through a feed tube to a porous tungsten plug
The Cs vapor diffuses through the pores in the plug to the front of surface
which is maintained at >1100oC by the ionizer heater
Operates with low melting point metals or metallic alloys, which are liquid at
room temperature or slightly above (Ga, Cs).
The liquid metal covers a W tip and emits ions under influence of an intense
electric field.
Ion current densities > 1A/cm2 with sub mm probe diameter.
Beam can be focused to <50 nm with moderate intensity and rastered to provide
secondary electron image or elemental mapping over the specimen surface.
Ion source: Liquid Metal Ion Source (LMIS)
W
Capillary
500 mm
22
Dual source SIMS23
Many SIMS spectrometers are
equipped with two sources,
usually a Cesium gun and an
Oxygen Duoplasmatron
source.
A mass filter (typically a
quadrupole), enables the
selection of the ion of interest.
Selected ions are then focused
and accelerated towards the
sample by electrostatic
lenses.
In the final stage of the dual
source electrostatic deflectors
drive primary ions towards
specific regions of the sample
surface.
Magnet Sector
Electrostatic Analyzer and Mass Spectrometer24
ESA is to minimize fluctuation
of kinetic energy of ions so as
to reduce the interference of
ions, providing a higher
mass resolution of mass
spectrometers
All paraxial ions of particular
energy will follow the central
lines to be focused in a plane
of the ESA slit
Fluctuation in kinetic energy
of ions is substantially
suppressed
The sputtering process produces ions with a range of ion energies. An energy
slit can be set to intercept the high energy ions. Sweeping the magnetic field in
MA provides the separation of ions according to mass-to-charge ratios in
time sequence.
The mass analyzer select the particular
species according to the mass-to-charge ratio
𝑚𝑞 =
𝐵2
2𝑉𝑟2
where B is the magnetic field, V is the ion
accelerating voltage, r is the radius of curvature
of the ion
For an analogy, think of how a prism
refracts and scatters white light separating it
into a spectrum of rainbow colors.
Mass spectrometer
In a mass spectrometer, ions travel different paths
through the magnet to the detector due to their
mass/charge ratios. A mass analyzer sorts the
ions according to mass/charge ratios and the
detector records the abundance of each ratio.
25
Ion Detectors
Faraday cupSecondary electron multiplier
20 dynodes Current gain 107
A Faraday cup measures the ion current
hitting a metal cup, and is sometimes used
for high current secondary ion signals.
With an electron multiplier an impact of a
single ion starts off an electron cascade,
resulting in a pulse of 108 electrons which
is recorded directly.
Usually it is combined with a fluorescent
screen, and signals are recorded either
with a CCD-camera or with a fluorescence
detector.
26
Mass resolution27
Typically a higher mass resolution will accompany a loss of ion intensity
Mass resolution is usually specified in terms of 𝑚/∆𝑚 where 𝑚 is the mass of
the ion and ∆𝑚 is the FWHM of the detected signal.
─ For example, 56Fe+ and 28Si2+ (𝑚/𝑞=55.9349 and 55.9539) require 𝑚/∆𝑚 of
2,950 for separation while Au and 133Cs32S2 (𝑚/𝑞=196.9666 and 196.8496)
require 𝑚/∆𝑚 of 1700.
𝑚/∆𝑚 for the two species is 21160FWHM
∆𝒎
Time of flight SIMS28
Time-of-Flight SIMS (ToF-SIMS) uses a pulsed ion beam to remove
molecules from the very outermost surface of the sample. These particles are
then accelerated into a "flight tube" and their mass is determined by
measuring the exact time at which they reach the detector (i.e. time-of-flight).
ToF-SIMS is based on the fact that ions with the same energy but different
masses travel with different velocities.
mass resolutions >18,000 can be achieved
It also has extremely high transmission with the parallel detection of all
masses and the unlimited mass range.
Time-of-flight mass analyzer
In order to provide higher resolution the pulse should be as narrow as 1-10 ns.
The pulse repetition frequency is usually in a kHz range. Typical flight times
10 ns to 800 µs
During a short pulse of ion beam,
sputtered ions are accelerated
and acquire a constant kinetic
energy:
𝐾𝐸 = 𝑚𝑣2/2
with different 𝑚/𝑞 and velocity 𝑣.
The ions arrive to the detector in
time sequence (𝑡) after traveling
a distance 𝑙.
𝑡 =𝑙
𝑣=
𝑙
2𝑞𝑉𝑚
2
𝑚
𝑞=
2𝑉𝑡2
𝑙2
𝑙
29
Reflectron ToF spectrometer
The kinetic energy distribution in the direction of ion flight can be corrected by
using a reflectron. The reflectron uses a constant electrostatic field to reflect
the ion beam toward the detector.
The more energetic ions penetrate deeper into the reflectron, and take a
slightly longer path to the detector.
Less energetic ions of the same mass-to-charge ratio penetrate a shorter
distance into the reflectron and, correspondingly, take a shorter path to the
detector.
Twice the flight path is achieved in a given length of instrument.
30
SIMS: modes of operation31
Static SIMS: 0.1-10 keV ions are
employed, with current surface
densities in the nA/cm2 range,
Under these conditions the total
erosion of the sample first
monolayer (1 nm) may take even
an hour.
Dynamic-SIMS: 10-30 keV ions,
with current surface densities in
the mA-mA/cm2 range, are used.
Under these conditions the sample
is eroded continuously and the
acquired mass spectra enable the
monitoring of constituting elements
along the sample depth (depth
profiling).
According to the primary ion energy and current, the SIMS technique can be
divided into two variants:
Profiling
Material removal
Elemental analysis
Ultra surface
analysis
Elemental or
molecular analysis
Analysis completed
before significant
fraction of molecules
destroyed
Static SIMS32
Under the ion bombardment, fragment ions or
even intact molecular ions are emitted from
the top monolayer.
If the primary ion dose is limited to a level at
which every primary ion should (statistically)
always hit a fresh area, the (static) SIMS
spectrum reveals molecular information.
Progressively, as the ion dose increases, the molecular signal decreases then
vanishes when the whole area has been damaged.
To stay in static SIMS mode, the primary ion dose must be < 𝟏𝟎𝟏𝟐 𝒊𝒐𝒏𝒔/𝒄𝒎𝟐
Static SIMS gives rise to a fingerprint mass spectrum that contains "low mass"
(< 500 amu) ion fragments, identifying organic surface composition.
Due to the complexity of the static SIMS mass spectrum, it is mostly used as a
qualitative characterization of the molecular composition of the top surface.
By focusing and scanning the primary ion beam, molecular information can be
obtained with sub-micron lateral resolution, and molecular surface distribution
can be imaged.
Static SIMS33
Positive ion TOF mass spectrum of polydimethylsiloxane
contaminated polyethylene terephthalateSilicon wafer contaminated with copper, iron and chromium
Range of elements H to U: all isotopes
Destructive Yes, if sputtered long enough
Chemical bonding Yes
Depth probed Outer 1 to 2 monolayers
Lateral resolution Down to below 100 nm
Imaging/mapping Yes
Quantification Possible with suitable standard
Mass range Typically up to 1000 amu, 10000 amu (ToF)
Main application Surface chemical analysis, organics, polymers
Dynamic SIMS34
Ion dosage and sputter rates are high resulting more
fragmentation.
Must be equipped with Oxygen and Cesium primary
ion beams in order to enhance, respectively, positive
and negative secondary ion intensity by 2 to 3 orders
of magnitude compared to the use of noble gas ions.
As the primary ion dose implanted in the target increases, the primary
species concentration (oxygen or cesium) will reach an equilibrium and this
corresponds to a sputtering steady state when reliable quantification is
possible with reference standard samples, using RSF.
One of the main application of dynamic SIMS is the in-depth distribution
analysis of trace elements (for example, dopant in semiconductors).
Impact ion energy is adjusted depending on the applications.
─ Low energy (down to 150eV) is used to reduce atomic mixing and
improve depth resolution down to the sub-nanometer level.
─ High energy (up to 20 keV) is chosen to investigate deeper (10-20
microns), faster (sputter rate of µm per min range), and improve
detection limits and image resolution.
Dynamic SIMS35
Range of elements H to U: all isotopes
Destructive Yes, material removed during sputtering
Chemical bonding In rare cases only
Depth probed Depth resolution 2-30 nm, probe into mm below surface
Quantification Standard needed
Accuracy 2%
Detection limits 1012-1016 atoms/cm3 (ppb-ppm)
Imaging/mapping Yes
Sample requirements Solid; vacuum compatible
Near surface B depth profiles
from a 2.2 keV BF implant in
Si using different energies O2+
primary beam
SIMS depth profiling: example36
Sputter time: 700 sec
Depth: 9310 Å
Erosion rate:13.3 Å/sec
Using an ion implanted
sample: P dose 1015 P/cm2
RSF: Relate the intensity to
atomic concentration:
𝑅𝑆𝐹 =𝑑𝑜𝑠𝑒
𝐼 𝑥 𝑑𝑥
RSF: 1counts/s=3.4x1015 P/cm3
Phosphorus doped Silicon
Dynamic SIMS – Depth Profiling
Factors affecting depth resolution:
Crater edge rejection:─ Raster beam for flat bottomed crater
─ Accept ions only from the center of crater
Ion beam mixing─ Primary ion mass
─ Impact energy
─ Impact angle
Surface roughness─ Metal worse than single crystal materials
The depth profile can be affected by:
Redeposition by sputtering from
the crater wall onto the analysis
area
Direct sputtering from the crater
wall
37
Crater Effect
(a)
(b)
The analyzed area is usually
required to be much smaller
than the scanned area.
a. Ions sputtered from a selected central
area (using a physical aperture or
electronic gating) of the crater are
passed into the mass spectrometer.
b. The beam is usually swept over a large
area of the sample and signal detected
from the central portion of the sweep.
This avoids crater edge effects.
38
Depth resolution39
Reducing ion penetration depth can reduce
effects of ion mixing, this can be achieved by
─ Larger angle of incidence from normal
─ Lower bombarding energy
─ Increased mass of primary beam
Simulation of the effect of 1% and 10%
unevenness on crater bottom for a sinusoidal
dopant distribution, according to the uneven
etching model
D. S. McPhail, et al., Scanning Microscopy 2, 639 (1989)
Luftman et al., J. Vac. Sci. Technol. B10, 323 (1992)
1.5 keV O2+ beam incident at 60° from
normal on a delta doped Be in GaAs sample
FWHM depth
resolution <3 nm
Sample Rotation Effect
SEM micrographs of a)
aluminum surface, b) bottom of
crater sputtered through 1 μm
aluminum layer into underlying
silicon without rotation and c)
with rotation
F. A. Stevie and J. L. Moore, Surf. Interf.
Anal. 18, 147 (1992)
B B
SiSi
Al Al
No rotation With rotation
SIMS profiles of 11B ion implantation into 1 μm Al/Si. With sample rotation, B at interface is clearly
defined and silicon from Al-Si-Cu layer shows movement to Al/Si interface
Reduction of preferential sputtering of different grains of polycrystalline materials
40
The SIMS detection limits for most trace elements are between 1012 and
1016 atoms/cm3.
The primary limiting factor is ionization efficiencies.
The dark current (or dark counts) arises from stray ions, electrons in
vacuum systems, and from cosmic rays
Count rate limited sensitivity:
─ When sputtering produces less secondary ion signal than the detector dark current.
─ If the SIMS instrument introduces the sample element, then the introduced level
constitutes background limited sensitivity, e.g. Oxygen, present as residual gas in
vacuum systems
Atoms sputtered from mass spectrometer parts by secondary ions constitute
another source of background.
Typically, sensitivity and depth resolution cannot be optimized simultaneously
Best sensitivity is achieved with high sputtering rate and large detected
area
Best depth resolution is achieved with low impact energy, reduced ion
penetration into sample, low sputtering rate and small detected area
Sensitivity and resolution41
42
Element M+ (O2+) M- (Cs+)
Li 3E13 1E16
Be 3E14 1E20
B 1E15 3E15
Na 3E14 2E17
Mg 1E14 1E20
Al 2E15 1E17
K 2E14 2E18
Ca 3E14 1E20
Ti 2E14 1E18
V 1E14 1E17
Cr 1E15 2E17
Mn 3E14 1E18
Fe 1E15 3E17
Ni 1E16 5E17
Cu 3E16 1E16
Zn 1E16 1E20
Sr 5E15 1E20
Y 1E17 1E20
Zr 1E15 4E17
Nb 1E16 1E18
Mo 1E16 1E18
Cd 5E16 1E21
In 3E15 3E17
Detection Limits: in InP, GaAs, GaN (atoms/cm3)
For electropositive elements:
Element M- (Cs+) M+ (O2+)
H 2E17 2E18
C 1E16 2E18
N 5E15 (NGa-) 5E18
O 1E16 1E20
F 2E14 5E16
P 2E15 1E16
Si 2E15 1E16
S 1E15 1E19
Cl 3E15 2E17
Ge 5E15 2E16
Se 5E14 2E17
Br 5E13 1E17
Te 1E15 2E17
Ag 2E16 2E16
Au 1E15 1E17
For electronegative elements:
Comparison: static and dynamic SIMS43
Imaging SIMS
The mass spectrometer is set to
only detect one mass.
The particle beam traces a raster
pattern over the sample with a
low ion flux beam, much like
Static SIMS.
Typical beam particles consists of
Ga+ or In+ and the beam diameter
is approximately 100 nm.
The analysis takes usually less
than 15 min.
The intensity of the signal detected for the particular mass is plotted
against the location that generated this signal.
Absolute quantity is difficult to measure, but for a relatively
homogeneous sample, the relative concentration differences are
measurable and evident on an image.
Images or maps of both elements and organics can be generated.
44
Imaging SIMS
Scanning ion image of granite from the Isle of Skye.
-University of Arizona SIMS 75 x 100 micrometers.
45
Imaging SIMS46
Detection of micrometric spots due to an organic contaminant (pentaerythritol-
tetraoctanoate, C37 H68 O8 , a lubricant) on an hard disk surface.
Charging of insulating samples in SIMS47
A positive charge is accumulated on the sample surface during a
SIMS analysis, due to ionic bombardment.
In insulating samples this charge cannot be neutralized by electrons
drawn from the ground through the sample stage.
Sample charging diffuses the primary beam and diverts it from the
analytical area, changes the energy distribution and direction of
secondary ions.
Several techniques are available to manage sample charging:
─ Flooding the sample surface with a low energy (a few eV) electron
beam, like in the case of XPS
─ Placing a conducting grids over the sample, similarly samples are
often coated with conducting materials such as gold or carbon.
─ Bombarding the sample with negative ions (for example O-)
─ Applying a continuously variable voltage offset to the accelerating
voltage for samples that are only slightly charging.
Example: Gate oxide breakdown48
Example: GaAs quantum well structure49
Negative secondary ions with 5keV
Cs primary ion bombardment
Example: ion beam mixing in isotope superlattice50
SIMS concentration profiles of the stable isotopes 74Ge (upper
solid line) and 70Ge (lower solid line) in crystalline (natGe/70Ge)10
and amorphous (natGe/73Ge)10 as-grown multilayers
The structures were implanted with 310 keV
Ga ions with a dose of 1.0 × 1015 and 3.3 ×1014/cm2.
Self-atom mixing in crystalline Ge is mainly
controlled by radiation enhanced diffusion
during the early stage of mixing before the
crystalline structure turns amorphousBracht et al., J. Appl. Phys. 110, 093502 (2011).
310 keV Ga+ 1.0 × 1015 /cm2
310 keV Ga+ 3.3 × 1014/cm2
Example: self-diffusion un GaSb51
Ga and Sb profiles of the
as-grown 69Ga121Sb/71Ga123Sb heterostructure
After annealing the
isotope structure under
Sb-rich conditions at
700oC for 105 min
After annealing at 700oC
for 18 days
Near the melting
temperature, Ga diffuses
more rapidly than Sb by
over three orders of
magnitude. This surprisingly
large difference in atomic
mobility is a consequence of
reactions between defects
on the Ga and Sb
sublattices, which suppress
the defects that are required
for Sb diffusion.
Bracht et al., Nature 408, 69 (2000).
Advantages and weaknesses of SIMS
Advantages Weaknesses
Excellent sensitivity, especially for
light elements
Destructive method
High surface sensitivity Element specific selectivity
Depth profiling with excellent depth
resolution (nm) (dynamic)
Standards needed for quantification
Good spatial resolution (<1-25 mm) Sample must be vacuum compatible
Small analyzed volume (down to
0.3mm3) so little sample is needed
Sample mist have a flat surface
Information about the chemical
surface composition due to ion
molecules (static)
High equipment cost (>1M-3M USD)
Elements from H to U can be
detected with excellent mass
resolution
52