Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106Dr Paul Holland, EffecTech Group & Jon Carlsen, Norsk AnalyseNorth Sea Flow Measurement Workshop 2017

Page 2: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Overview of presentation

• Importance of the GC and its performance

• Importance of a good quality calibration gas

• Introduce Norsok I-106 test & ISO 10723 and highlight different evaluation approaches

• Highlight some of the additional benefits from the ISO approach

• Results of using an ISO 10723 to generate a Norsok test report

• Summary

2

Page 3: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

3

Fit for purpose?

Sometimes it’s obvious if equipment is fit for purpose…..

Page 4: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Fit for purpose?

Sometimes it is not obvious…..

4

Page 5: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Gas Quality instrumentation (which to choose?)

Answer….Best performing instrument

Page 6: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

6

Natural Gas Quality Measurement

• Composition (content) of natural gas

– Inert gases

• Nitrogen, carbon dioxide, helium, argon

– Hydrocarbons

• Methane, ethane, propane, iso-butane, n-butane, pentanes, hexanes + ......

• Properties (characteristics) of natural gas

– Calorific value, Wobbe number, standard density (ISO 6976)

– Compression factor, line density (ISO 12213, AGA 8)

– Hydrocarbon dew point (ISO 23874)

– Emission factors (BS 8609)

Page 7: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Financial risks in energy metering

0

500,000

1,000,000

1,500,000

2,000,000

2,500,000

3,000,000

3,500,000

4,000,000

4,500,000

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6

Annual Value / €

U(Energy) / %

Typical gas fired power station

Power output : 500 MW

Energy Price : €60 / MWh

Typical

Page 8: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

NORSOK I-106 BS EN ISO 10723

Page 9: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Norsok I-106 ISO 10723

• Published 2014

• Norwegian Standard

• Defines pass / fail criteria of 0.3% for GCV based on:

– Linearity

– Repeatability

– Uncertainty

• Published 2012

• International Standard

• Defines a method, OIML & GUM compliant

• No pass / fail criteria

• Establishes

– Bias

– Uncertainty

• Bias and uncertainty compared to established limits

Page 10: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Calibration gas

• The gas chromatograph is a comparator

– wrist watch analogy

• calibration gas

– principal driver for accuracy of measurements

– inaccuracies in the calibration gas will produce errors in stream gas results

• NORSOK specifies uncertainty limits for calibration and test gases (Table2)

• ISO 10723 does not specify but requires uncertainties which are used in

the calculations

Page 11: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Issues: Low temperature

Page 12: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Using a non-homogeneous calibration gas

• Cup of tea analogy

Page 13: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Repeatability ComparisonNorsok ISO 10723

Component range (mole %)

SD (mole %)URXi/2

0 to 25

25 to 100

0.01

0.025

ComponentRange (mol%)

SD (mole %)

Not set

0

0.01

0.02

0.03

0.04

0.01 0.10 1.00 10.00 100.00

U (

mo

l %

)

mol %

NORSOK ISO 6974

Page 14: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Why does ISO 10723 not set a repeatability criteria?

• Now common practice to accept errors in the measurement of component amount

fractions provided their effect does not give rise to unacceptable errors in physical

properties.

• Question: If the repeatability of 50 ppm nitrogen in natural gas was greater than the defined limit, what would the effect be on the calculated calorific value?

• Answer: …….A very small effect….should this instrument fail?

• The ISO 10723 provides data which can be compared to Maximum Permissible

Bias (MPB) & Maximum Permissible Error (MPE) limits.

Page 15: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Linearity ComparisonNorsok

• 3 point check

ISO 10723

• 7 point check

0

200

400

600

800

1000

1200

65 75 85 95

Are

a

amount fraction %mol/mol

0

200

400

600

800

1000

1200

65 75 85 95

Are

a

amount fraction %mol/mol

Can only describe a linear response Can describe a response up to 3rd

order polynomial

Page 16: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Dealing with non-linearity

NORSOK

• Determines the maximum difference between the measured and reference values

ISO 10723

• GLS regression analysis to determine best fit polynomial

Page 17: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

ISO 10723 linearity results

Page 18: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Non-linear response errors

0

20

40

60

80

100

120

0 2 4 6 8 10 12

assumed response

true response

amount fraction / (%mol/mol)

instrument

response

Page 19: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Combining uncertainties - NORSOK

19

Page 20: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Modelling performance using ISO 10723

• The ISO 10723 models the GC performance

– Linearity

– Repeatability

– Calibration gas composition & uncertainty

• Monte Carlo is run on 10,000 simulated gas compositions to establish

– Mean error – Bias

– Uncertainty on the error

Page 21: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

-0.15

-0.10

-0.05

0.00

0.05

0.10

0.15

78 80 82 84 86 88 90 92 94 96 98

E(CVSUP) / MJ.m-3

methane content / (% mol/mol)

Errors and uncertainties on errors

Page 22: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Error Distribution

-0.15

-0.10

-0.05

0.00

0.05

0.10

0.15

78 80 82 84 86 88 90 92 94 96 98

E(CVSUP) / MJ.m-3

methane content / (% mol/mol)

Page 23: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Mean Error (Bias)

-0.16

-0.14

-0.12

-0.10

-0.08

-0.06

-0.04

-0.02

0.00

0.02

0.04

0.06

0.08

0.10

0.12

0.14

0.16

78 83 88 93 98

E(CVSUP) / MJ.m-3

methane content / (%mol/mol)

Maximum Permissible Bias

(MPB)

mean error = bias - B(P)

MPBP ≤δ

Page 24: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Uncertainty on Mean Error

-0.16

-0.14

-0.12

-0.10

-0.08

-0.06

-0.04

-0.02

0.00

0.02

0.04

0.06

0.08

0.10

0.12

0.14

0.16

78 83 88 93 98

E(CVSUP) / MJ.m-3

methane content / (%mol/mol)

Maximum Permissible Error

(MPE)uncertainty on the mean error≈ uncertainty on bias - U(B(P))

( ) MPEPUPc

≤+ δδ

Page 25: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Errors and Uncertainties - Summary

Page 26: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Uncertainty Comparison• Norsok • ISO 10723 (UK specification)

Property Uncertainty

Gross CV < 0.3%

Property Uncertainty

Gross CV

Gross CV

MPB < 0.09 MJ.m3

MPE < 0.18 MJ.m3

MJ.m3(0.45%)

Page 27: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Additional advantages of ISO 10723

27

Page 28: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Optimising Calibration Gas Design

xC1,cal = 0.88mean E(CV) = 0.001 ± 0.061 MJ.m-3

xC1,cal = 0.81mean E(CV) = 0.000 ± 0.028 MJ.m-3

Page 29: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Operating the GC in a type 1 mode (ISO6974)

Page 30: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Type 1 mode vs Type 2 mode

Page 31: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Calibration gas – reduce dewpoint = increase fill pressure

0 1 2 3 4 5 6 7 8 9 10

response / (peak area)

content / (% mol/mol)

y=Fass(x)

y=Ftrue(x)

xcal

Page 32: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Simulating a NORSOK Test using ISO 10723 data

32

Page 33: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Simulating a NORSOK Test using ISO 10723 data

33

Danalyser

Simulated Real

Component Linearity Uncertainty Linearity Uncertainty

Uncertainty Uncertainty

K =2 K =2

ULXi ULXi

C1 0.111 0.118

C2 0.028 0.031

C3 0.010 0.016

iC4 0.001 0.001

nC4 0.006 0.002

neoC5 0.000 0.000

iC5 0.001 0.000

nC5 0.000 0.001

C6 0.054 0.060

N2 0.003 0.002

CO2 0.008 0.008

O2 0.000 0.000

UHS (%rel) 0.18% 0.20%

Ums (% rel) 0.18% 0.20%

UEF (%rel) 0.22% 0.24%

Page 34: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Simulating a NORSOK Test using ISO 10723 data

34

Encal

Simulated Real

Component Linearity Uncertainty Linearity Uncertainty

Uncertainty Uncertainty

K =2 K =2

ULXi ULXi

C1 0.001 0.192

C2 0.001 0.019

C3 0.0004 0.007

iC4 0.0001 0.001

nC4 0.002 0.002

neoC5 0.000 0.000

iC5 0.000 0.000

nC5 0.000 0.000

C6 0.0002 0.001

N2 0.0002 0.002

CO2 0.000 0.007

O2 0.000 0.000

UHS (%rel) 0.03% 0.04%

Ums (% rel) 0.04% 0.06%

UEF (%rel) 0.04% 0.06%

Page 35: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Simulating a NORSOK Test using ISO 10723 data

35

Maxum II

Simulated Real

Component Linearity Uncertainty Linearity Uncertainty

Uncertainty Uncertainty

K =2 K =2

ULXi ULXi

C1 0.065 0.092

C2 0.043 0.049

C3 0.012 0.023

iC4 0.001 0.003

nC4 0.001 0.002

neoC5 0.000 0.000

iC5 0.001 0.001

nC5 0.000 0.001

C6 0.003 0.003

N2 0.002 0.008

CO2 0.001 0.003

O2 0.000 0.000

UHS (%rel) 0.04% 0.05%

Ums (% rel) 0.04% 0.05%

UEF (%rel) 0.05% 0.06%

Page 36: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech

Summary• Norsok test & ISO 10723 both determine instrument performance �

• ISO 10723 has some additional advantages:– Accurately model instrument response characteristics (up to 3rd order polynomial)

– Determine uncertainties over range or determine the instrument range within a

given uncertainty

– Provides data for GC to be operated in Type 1 mode according to ISO 6974

(minimising errors)

– Determine optimum calibration gas composition

– Decide if a new stream can be added to an existing GC

– Identify which component is at fault if issues are identified with GC

– Determine Limits of Detection (LOD)

– ISO 17025 calibration certificate can be issued for Emissions Trading

Requirements (EU/ETS)

– Can be used to simulate NORSOK test results

Page 37: Instrument Performance Evaluation, comparing ISO 10723 and NORSOK … · 2019-03-20 · Instrument Performance Evaluation, comparing ISO 10723 and NORSOK I-106 Dr Paul Holland, EffecTech