Instrument Disassembly and Assembly Manual
Table of Contents
Knees
Modular Case Bracket .......................................................................................... 4
Posterior Referencing Sizer .................................................................................. 6
Spacer Block Assemblies ...................................................................................... 7
Tensor Assemblies ................................................................................................ 8
Tibial Articular Surface Provisional (TASP) .......................................................... 9
Hips
Liner Inserter ...................................................................................................... 12
Modular Tap Handle ........................................................................................... 14
Straight Universal Handle .................................................................................. 16
Acetabular Drill ................................................................................................... 18
Trauma
Tensioner and Bit ................................................................................................ 22
AO Collet Handle with Spin Cap ......................................................................... 24
Drill Guide Block ................................................................................................. 26
Impactor ............................................................................................................. 28
Locking Entry Cannula ....................................................................................... 30
3.5/2.7 mm Locking Screw Standard Depth Gauge .......................................... 32
Retrograde Femur Depth Gauge ........................................................................ 34
Small Depth Gauge 60 mm ................................................................................. 36
Extremities
T-Handle ............................................................................................................. 40
Suction Device .................................................................................................... 42
Appendix
Parts Index .......................................................................................................... 44
2 | Instrument Disassembly and Assembly Manual
3 | Instrument Disassembly and Assembly Manual
KNEES
4 | Instrument Disassembly and Assembly Manual
Disassembly Instructions1. Remove instruments and provisionals from tray.
2. From the back side of the tray, locate bracket to be removed.
3. Grasp bracket removal tool.
4 Fully place removal tool over desired fastener.
5. Fully depress plunger to disengage each fastener on the desired bracket.
6. After all fasteners have been disengaged, the bracket will drop from the tray. If bracket remains engaged repeat steps 3, 4, and 5.
7. End of disassembly.
Note: The Modular Case Bracket must be fully disassembled for cleaning and steam sterilization.
Modular Case Bracket
KNEE
S
Step 1
Step 2
Step 6
Step 5
Step 3 & 4
Bracket
Tray
Back Side of Tray
Bracket Removal
Tool
Fastener
5 | Instrument Disassembly and Assembly Manual
5 | Disassembly and Assembly Manual
KNEES
Assembly Instructions1. Choose a desired Case Bracket.
2. Ensure wave springs are present on each fastener.
3. Pick a desired location to place bracket ensuring thatit will not overlap neighboring brackets.
4. Press down on bracket until there is an audible clickfrom each fastener.
5. Invert tray and inspect each fastener to ensurewedges are fully engaged. If fasteners are not fullyengaged,repeat step 4.
6. End of assembly.
Wave Spring
Fastener
Modular Case Bracket (cont.)
Step 2
Step 1
Case Bracket
Step 3Step 4
Step 5
KNEES
Assembly Instructions1. Choose a desired case bracket.
2. Ensure wave springs are present on each fastener.
3. Pick a desired location to place bracket ensuring no overlap of neighboring brackets.
4. Press down on bracket until there is an audible click from each fastener.
5. Invert tray and inspect each fastener to ensure wedges are fully engaged. If fasteners are not fully engaged,repeat step 4.
6. End of assembly.
Modular Case Bracket
Wave Spring
Fastener
Step 2
Step 3
Step 5
Step 4
5 | Disassembly and Assembly Manual
KNEES
Assembly Instructions1. Choose a desired Case Bracket.
2. Ensure wave springs are present on each fastener.
3. Pick a desired location to place bracket ensuring thatit will not overlap neighboring brackets.
4. Press down on bracket until there is an audible clickfrom each fastener.
5. Invert tray and inspect each fastener to ensurewedges are fully engaged. If fasteners are not fullyengaged,repeat step 4.
6. End of assembly.
Wave Spring
Fastener
Modular Case Bracket (cont.)
Step 2
Step 1
Case Bracket
Step 3Step 4
Step 5
Case Bracket
Step 1
Tray
Fastener
6 | Instrument Disassembly and Assembly Manual
Disassembly Instructions1. Separate the magnetic sizer drill guide from the
top surface of the spacer block or tensor.
2. Pull the posts of the sizer stylus up and out of the holes in the sizer drill guide until fully released.
3. End of disassembly.
Note: The Posterior Referencing Sizer must be fully disassembled for cleaning and steam sterilization.
Posterior Referencing Sizer
Step 3
KNEE
S
Sizer Stylus
Magnetic Sizer Drill Guide
Step 2
Magnetic Sizer Drill Guide
Spacer Block or Tensor
(Spacer Block Shown)
Step 1
7 | Instrument Disassembly and Assembly Manual
KNEES
Step 2
or
Step 1
Disassembly Instructions1. Pull the tab of the shim away from the magnetic
handle of the spacer block until fully released.
2. Separate the magnetic +9 mm femoral paddle from the +0 mm femoral paddle.
3. Depress the release button of the spacer block and slide the +0 mm femoral paddle or 4-in-1 adapter until fully released.
4. End of disassembly.
Note: The 4-in-1 Adapter is an optional attachment to the spacer block.
Note: The Spacer Block Assemblies must be fully disassembled for cleaning and steam sterilization.
Magnetic +9 mm Femoral Paddle
Spacer Block
Shim
Spacer Block Assemblies
Step 3
4-in-1 Adapter(optional part)+0 mm Femoral Paddle
Step 3
Step 4
Magnetic +9 mm Femoral Paddle
+0 mm Femoral Paddle
Spacer Block
Shim
8 | Instrument Disassembly and Assembly Manual
+0 mm Femoral Paddle
Tibial Tensor
Magnetic +9 mm Femoral Paddle
Step 4
Femoral Tensor
Tensor Assemblies
Disassembly Instructions1. Depress the release button of the tibial tensor and
pull the femoral tensor up until fully released.
2. Separate the magnetic +9 mm femoral paddle from the +0 mm femoral paddle.
3. Depress the release button of the femoral tensor and slide the +0 mm femoral paddle or 4-in-1 adapter until fully released.
4. End of disassembly.
Note: The 4-in-1 Adapter is an optional attachment to the femoral tensor.
Note: The Tensor Assemblies must be fully disassembled for cleaning and steam sterilization.
Step 1
Femoral Tensor
Tibial Tensor
Release Button
Step 3
Femoral Tensor
+0 mm Femoral Paddle
+0 mm Femoral Paddle
Step 2
Magnetic +9 mm Femoral Paddle
KNEE
S
Femoral Tensor4-in-1 Adapter(optional part)
9 | Instrument Disassembly and Assembly Manual
KNEES
Disassembly Instructions1. Insert tibial sizing plate handle into assembled
TASP construct by depressing button and inserting into shim.
2. Remove the shim from the top and bottom by pulling the tibial sizing plate handle away from the assembled TASP Construct without depressing button. To avoid damage to the top and bottom, do not impact or apply twisting forces to the tibial sizing plate handle during shim removal.
3. Separate top from bottom.
4. Depress button on handle to separate handle from shim.
5. End of disassembly.
Note: The TASP must be fully disassembled for cleaning and steam sterilization.
Step 3
TASP Shim
TASP Top
TASP Bottom
Tibial SizingPlate Handle
Step 1
Assembled TASP Construct
Step 2
Step 4
TASP Shim
Tibial SizingPlate Handle
Tibial Articular Surface Provisional (TASP)
10 | Instrument Disassembly and Assembly Manual
KNEE
S Signs of Potential Mechanical FailureDue to multiple reuse cycles, damage of the TASP components resulting in mechanical failure is possible. Upon disassembly, evaluate components for damage and preliminary signs of failure. If damage is noticed, do not use the device and contact your Zimmer Biomet representative for a replacement.
Below are preliminary signs of failure associated with the TASPs as seen in internal testing.
TASP TOPS Deformation along edges of dovetail TASP TOPS Cracks along length of dovetail
TASP BOTTOM Cracks/crazing along central post
Tibial Articular Surface Provisional (TASP) (cont.)
11 | Instrument Disassembly and Assembly Manual
HIPS
12 | Instrument Disassembly and Assembly Manual
Disassembly Instructions1. Unthread the inserter sleeve by turning it
clockwise until the threads are fully disengaged. Sleeve and washer should be loose on the shaft of the liner inserter.
2. Slide the rubber o-ring off of the head towards the black handle. O-ring should be loose on the shaft of the liner inserter.
3. End of disassembly.
Note: The Liner Inserter must be fully disassembled for cleaning and steam sterilization.
Liner Inserter
HIP
S
Inserter Sleeve
Inserter Washer
Step 1
Head
O-Ring
Step 2
Step 3
13 | Instrument Disassembly and Assembly Manual
Assembly Instructions1. Pull o-ring onto the head until it is fully seated
underneath the head.
2. Push inserter washer firmly up against o-ring and the inserter sleeve up against inserter washer.
3. Secure the assembly by turning the sleeve clockwise until the threads are fully engaged.
4. End of assembly.
Inserter Sleeve
Step 3
Head
O-Ring
Step 1
Step 2
Inserter Sleeve
Inserter Washer
Liner Inserter
HIPS
14 | Instrument Disassembly and Assembly Manual
Disassembly Instructions 1. Turn the outer collar counter clockwise to
completely unthread the outer collar from the inner collet. There will be increased resistance as you continue to unthread the outer collar.
2. End of disassembly.
Note: The Modular Tap Handle must be fully disassembled for cleaning and steam sterilization.
Modular Tap Handle
Outer Collar
Inner Collet
Step 2
Step 1
HIP
S
15 | Instrument Disassembly and Assembly Manual
HIPS
Assembly Instructions1. Put outer collar onto inner collet.
2. Turn the outer collar clockwise to thread the outer collar onto the inner collet. There will be increased resistance as you continue to thread the outer collar.
3. End of assembly.
Modular Tap Handle
Step 2
Outer Collar
Inner Collet
Step 1
16 | Instrument Disassembly and Assembly Manual
Disassembly Instructions1. Pull the outer sleeve towards the blue handle
and twist clockwise to unlock the sleeve.
2. Slide sleeve away from blue handle.
3. End of disassembly.
Note: The Straight Universal Handle must be fully disassembled for cleaning and steam sterilization.
Outer Sleeve
Blue Handle
Straight Universal Handle
HIP
S
Outer Sleeve
Shaft
Step 2
Step 1
17 | Instrument Disassembly and Assembly Manual
Assembly Instructions1. Align pin in tip of main shaft with slot in outer
sleeve and slide outer sleeve over the shaft.
2. Pull sleeve towards blue handle until it stops and Rotate outer sleeve counter clockwise until it locks.
3. End of assembly.
Straight Universal Handle
Outer Sleeve
Blue Handle
HIPS
Outer Sleeve
Shaft
Step 1
Step 2
18 | Instrument Disassembly and Assembly Manual
Acetabular Drill
Step 2
Disassembly Instructions1. Remove polymer C-shaped collar from drill shaft.
2. Remove polymer sleeve from drill shaft.
3. End of disassembly.
Note: The Acetabular Drill must be fully disassembled for cleaning and steam sterilization.
Step 1
Shaft
C-shaped Collar
Polymer Sleeve
Polymer Sleeve
HIP
S
Shaft
C-shaped Collar
19 | Instrument Disassembly and Assembly Manual
Assembly Instructions1. Slide polymer sleeve over drill shaft.
2. Attach the C-shaped collar to the groove in the drill shaft.
3. End of disassembly.
Note: The Acetabular Drill must be fully disassembled for cleaning and steam sterilization.
Acetabular Drill
Step 2
Shaft
C-shaped Collar
Polymer Sleeve
Step 1
HIPS
20 | Instrument Disassembly and Assembly Manual
21 | Instrument Disassembly and Assembly Manual
TRAUMA
22 | Instrument Disassembly and Assembly Manual
Disassembly Instructions 1. Pull the tensioning bit out from the tensioner.
2. Pull the lever arm out from tensioning bit.
3. End of disassembly.
Note: The Tensioner and Bit must be fully disassembled for cleaning and steam sterilization.
Tensioner and Bit
Step 2
Lever Arm
Tensioning Bit
Step 1
Tensioner Tensioning Bit
Step 3
Tensioner
TRAU
MA
23 | Instrument Disassembly and Assembly Manual
Assembly Instructions1. Insert the lever arm into tensioning bit.
2. Insert the tensioning bit into tensioner.
3. End of assembly.
Tensioner and Bit
Step 1
Lever Arm
Tensioning Bit
Step 2
Tensioner Tensioning Bit
Step 3
Tensioner
TRAU
MA
24 | Instrument Disassembly and Assembly Manual
Disassembly Instructions1. Pull the spin cap off the handle.
2. End of disassembly.
Note: The AO Collet Handle with Spin Cap must be fully disassembled for cleaning and steam sterilization.
AO Collet Handle with Spin Cap
TRAU
MA
Step 1
Step 2
Handle
Connector
Spin Cap
25 | Instrument Disassembly and Assembly Manual
TRAU
MA
Assembly Instructions1. Place the spin cap on the connector on the end of
the handle and press firmly. The cap should spin freely on the handle.
2. End of assembly.
AO Collet Handle with Spin Cap
Step 2
Step 1
Handle
Connector
Spin Cap
26 | Instrument Disassembly and Assembly Manual
Disassembly Instructions
1. Pull the thumb screw out from the guide block.
2. End of disassembly.
Note: The Drill Guide Block must be fully disassembled for cleaning and steam sterilization.
Thumb Screw
Guide Block
Drill Guide Block
TRAU
MA
27 | Instrument Disassembly and Assembly Manual
Thumb Screw
Guide Block
Assembly Instructions
1. Firmly press the thumb screw into the guide block.
2. End of assembly.
Drill Guide Block
TRAU
MA
28 | Instrument Disassembly and Assembly Manual
Disassembly Instructions 1. Turn the impactor nose counter clockwise to
unscrew from the impactor.
2. End of disassembly.
Note: The Impactor must be fully disassembled for cleaning and steam sterilization.
Impactor
Impactor
Impactor Nose
Step 1
Step 2
TRAU
MA
Impactor
29 | Instrument Disassembly and Assembly Manual
Assembly Instructions1. Thread the impactor nose into the impactor until it
is fastened completely.
2. End of assembly.
Impactor
Impactor
Impactor Nose
Step 1
Impactor
TRAU
MA
30 | Instrument Disassembly and Assembly Manual
Disassembly Instructions1. Push and hold down button.
2. Remove the sleeve from the blue handle.
3. End of disassembly.
Note: The Locking Entry Cannula must be fully disassembled for cleaning and steam sterilization.
Locking Entry Cannula
Button
Step 1
Step 2
TRAU
MA
31 | Instrument Disassembly and Assembly Manual
Sleeve
Step 2
Blue Handle
Locking Entry Cannula
TRAU
MA
Assembly Instructions1. Push and hold down button.
2. Insert the sleeve into blue handle until it is fastened completely.
3. End of assembly.
Locking Entry Cannula
Step 3
Button
Step 1
TRAU
MA
32 | Instrument Disassembly and Assembly Manual
3.5/2.7 mm Locking Screw Standard Depth Gauge
Shaft
Calibrated Gauge
Shaft
Calibrated Gauge
Disassembly Instructions1. Hold the shaft and pull the calibrated gauge
away from the shaft.
2. End of disassembly.
Note: The 3.5/2.7 mm Locking Screw Standard Depth Gauge must be fully disassembled for cleaning and steam sterilization.
TRAU
MA
Step 1
33 | Instrument Disassembly and Assembly Manual
Shaft
Calibrated Gauge
Calibrated Gauge
Shaft
Assembly Instructions1. Insert the calibrated gauge into the large
opening on the shaft.
2. End of assembly.
3.5/2.7 mm Locking Screw Standard Depth Gauge
TRAU
MA
Step 1
34 | Instrument Disassembly and Assembly Manual
Disassembly Instructions 1. Remove the calibrated gauge from the shaft.
2. End of disassembly.
Note: The Retrograde Femur Depth Gauge must be fully disassembled for cleaning and steam sterilization.
Retrograde Femur Depth Gauge
Calibrated Gauge
Shaft
TRAU
MA
Step 1
35 | Instrument Disassembly and Assembly Manual
TRAU
MA
Assembly Instructions1. Insert the Calibrated Gauge into the Shaft.
2. End of assembly.
Retrograde Femur Depth Gauge
Shaft
Calibrated Gauge
TRAU
MA
Step 1
36 | Instrument Disassembly and Assembly Manual
Disassembly Instructions1. Turn the cap counter clockwise to unscrew cap
from shaft and remove.
2. Remove the calibrated gauge by holding the shaft and push on the flat end of the calibrated gauge untill it is all the way out.
3. End of disassembly.
Note: The Depth Gauge must be fully disassembled for cleaning and steam sterilization.
Small Depth Gauge 60 mm
Calibrated Gauge
Shaft
Cap
Shaft
Cap
Step 1
Flat end of Calibrated Gauge
Step 2
ShaftTRAU
MA
37 | Instrument Disassembly and Assembly Manual
Assembly Instructions1. Insert the calibrated gauge into the shaft.
2. Place the cap over the probe of the calibrated gauge.
3. Turn the cap clockwise to thread onto the shaft.Tighten firmly.
4. End of assembly
Small Depth Gauge 60 mm
Cap
Step 3Step 2
Shaft
Calibrated Gauge
Shaft
Cap
Cap
Step 1
TRAU
MA
38 | Instrument Disassembly and Assembly Manual
39 | Instrument Disassembly and Assembly Manual
EXTREMITIES
40 | Instrument Disassembly and Assembly Manual
Disassembly Instructions1. Push t-handle lock towards t-handle slide.
2. While pressed, turn the t-handle lock clockwise and release.
3. Slide the spring and the t-handle slide up the t-handle shaft to disengage the four pegs from the four holes.
4. End of disassembly.
Note: The T-Handle must be fully disassembled for cleaning and steam sterilization.
T-Handle
EXTR
EMIT
IES
T-Handle Slide
Spring
T-Handle Shaft
T-Handle Lock
Step 2
Step 3
T-Handle Lock
T-Handle Slide
Step 1
41 | Instrument Disassembly and Assembly Manual
Assembly Instructions1. Move the t-handle slide, spring, and t-handle lock
down the shaft and push the four pins of the into the four holes of the t-handle shaft.
2. Push the t-handle lock towards the t-handle slide, compressing the spring.
3. While pressed, turn the t-handle lock counter clockwise and release.
4. End of assembly.
T-Handle
EXTR
EMITIES
Step 3
Step 4
T-Handle Lock
Step 2
Spring
T-Handle Shaft
T-Handle Lock
T-Handle Slide
Step 1
42 | Instrument Disassembly and Assembly Manual
EXTR
EMIT
IES
Disassembly Instructions1. Unscrew the screw cap.
2. Slide screw cap off of the shaft.
3. Pull the suction cup from the shaft.
4. End of disassembly.
Note: The Suction Device must be fully disassembled for cleaning and steam sterilization.
Suction Device
Screw Cap
Step 1
Suction Cup
Step 2
Suction Cup
Shaft
Step 3
Screw Cap
43 | Instrument Disassembly and Assembly Manual
EXTR
EMITIES
Assembly Instructions1. Push the suction cup into the shaft until the shaft’s
fingers are in the suction cup’s groove.
2. Slide the screw cap all the way down the shaft with the large hole facing the suction cup.
3. Screw the screw cap completely to the shaft.
4. End of assembly.
Suction Device
Screw Cap
Step 3
Suction Cup
Step 2
Suction Cup
Shaft
Step 1
Screw Cap
44 | Instrument Disassembly and Assembly Manual
Appendix A – Parts Index
Posterior Referencing Sizer
42-5097-060-10 FUZION P-REF SIZER STYLUS
42-5097-061-00 FUZION P-REF SIZER DR GDE L
42-5097-062-00 FUZION P-REF SIZER DR GDE R
42-5097-011-00 FUZION SPACER LEFT
42-5097-012-00 FUZION SPACER RIGHT
42-5097-021-00 FUZION TENSOR L FEM ASSY
42-5097-022-00 FUZION TENSOR R FEM ASSY
Spacer Block Assemblies
42-5097-011-11 FUZION SPACER SHIM L 11 MM
42-5097-011-20 FUZION SPACER SHIM L 20 MM
42-5097-012-11 FUZION SPACER SHIM R 11 MM
42-5097-031-00 FUZION +0 MM FEM PADDLE L
42-5097-032-00 FUZION +0 MM FEM PADDLE R
42-5097-012-20 FUZION SPACER SHIM R 20 MM
Tensor Assemblies
42-5097-021-00 FUZION TENSOR L FEM ASSY
42-5097-021-01 FUZION TENSOR L TIB ASSY
42-5097-022-00 FUZION TENSOR R FEM ASSY
42-5097-022-01 FUZION TENSOR R TIB ASSY
42-5097-030-09 FUZION +9 MM FEM PADDLE
42-5097-031-00 FUZION +0 MM FEM PADDLE L
42-5097-032-00 FUZION +0 MM FEM PADDLE R
42-5097-041-00 FUZION 4-IN-1 ADPT L
42-5097-042-00 FUZION 4-IN-1 ADPT R
Tibial Articular Surface Provisional (TASP)
42-5279-002-00 PSN CONST TASP SHIM CD 10 MM
42-5279-002-02 PSN CONST TASP SHIM CD 12 MM
42-5279-002-04 PSN CONST TASP SHIM CD 14 MM
42-5279-003-00 PSN A/S PROV SHIM, 10 MM CD
42-5279-003-01 PSN A/S PROV SHIM, 11 MM CD
42-5279-003-02 PSN A/S PROV SHIM, 12 MM CD
42-5279-003-03 PSN A/S PROV SHIM, 13 MM CD
42-5279-003-04 PSN A/S PROV SHIM, 14 MM CD
42-5279-004-00 PSN CONST TASP SHIM EF 10 MM
42-5279-004-02 PSN CONST TASP SHIM EF 12 MM
42-5279-004-04 PSN CONST TASP SHIM EF 14 MM
42-5279-005-00 PSN A/S PROV SHIM, 10 MM EF
42-5279-005-01 PSN A/S PROV SHIM, 11 MM EF
42-5279-005-02 PSN A/S PROV SHIM, 12 MM EF
42-5279-005-03 PSN A/S PROV SHIM, 13 MM EF
42-5279-005-04 PSN A/S PROV SHIM, 14 MM EF
42-5279-006-00 PSN CONST TASP SHIM GH 10 MM
42-5279-006-02 PSN CONST TASP SHIM GH 12 MM
42-5279-006-04 PSN CONST TASP SHIM GH 14 MM
42-5279-007-00 PSN A/S PROV SHIM, 10 MM GH
42-5279-007-01 PSN A/S PROV SHIM, 11 MM GH
42-5279-007-02 PSN A/S PROV SHIM, 12 MM GH
42-5279-007-03 PSN A/S PROV SHIM, 13 MM GH
42-5279-007-04 PSN A/S PROV SHIM, 14 MM GH
42-5279-008-00 PSN CONST TASP SHIM J 10 MM
42-5279-008-01 PSN CONST TASP SHIM J 11 MM
42-5279-008-02 PSN CONST TASP SHIM J 12 MM
42-5276-008-03 PSN CONST TASP SHIM J 10 MM
42-5279-008-04 PSN CONST TASP SHIM J 14 MM
42-5279-009-00 PSN A/S PROV SHIM, 10 MM J
42-5279-009-00 PSN A/S PROV SHIM, 10 MM J
42-5279-009-01 PSN A/S PROV SHIM, 11 MM J
42-5279-009-01 PSN A/S PROV SHIM, 11 MM J
42-5279-009-02 PSN A/S PROV SHIM, 12 MM J
42-5279-009-02 PSN A/S PROV SHIM, 12 MM J
42-5279-009-03 PSN A/S PROV SHIM, 13 MM J
42-5279-009-03 PSN A/S PROV SHIM, 13 MM J
42-5279-009-04 PSN A/S PROV SHIM, 14 MM J
42-5279-009-04 PSN A/S PROV SHIM, 14 MM J
Liner Inserter
00-6260-017-22 LINER INSERTER MECHANICAL 22 MM
00-6260-017-26 LINER INSERTER MECHANICAL 26 MM
00-6260-017-28 LINER INSERTER MECHANICAL 28 MM
00-6260-017-32 LINER INSERTER MECHANICAL 32 MM
00-6260-017-36 MECH LINER INSERTER 36 MM
00-6260-017-40 MECH LINER INSERTER 40 MM
AO Collet Handle with Spin Cap
00-2366-015-00 AO COLLET HANDLE
Modular Tap Handle
00-8790-005-30 MODULAR TAP HANDLE
Straight Universal Handle
00-8790-010-00 STRAIGHT UNIVERSAL HANDLE
APP
END
IX A
45 | Instrument Disassembly and Assembly Manual
Acetabular Drill
00-8000-214-00 ZCA ACETABULAR DRILL, 6 MM
00-8000-215-00 ZCA ACETABULAR DRILL,13 MM
00-8000-216-00 ZCA ACETABULAR DRILL,10 MM
Tensioner and Tensioner Bit
00-2232-060-01 CABLE TENSIONER
00-2232-060-18 CABLE TENSIONER BIT STD LNG
Drill Guide Block
00-2366-099-01 DRILL GUIDE BLOCK RTN SCREW
00-2366-099-50 NARROW DRILL GUIDE BLOCK LEFT
00-2366-099-51 NARROW DRILL GUIDE BLOCK RIGHT
00-2366-099-52 STD. DRILL GUIDE BLOCK LEFT
00-2366-099-53 STD. DRILL GUIDE BLOCK RIGHT
00-2366-099-54 WIDE DRILL GUIDE BLOCK LEFT
00-2366-099-55 WIDE DRILL GUIDE BLOCK RIGHT
Impactor
00-1193-006-00 IMPACTOR
00-1193-055-00 IMPACTOR
Locking Entry Cannula
00-2490-013-04 LOCKING ENTRY CANNULA
3.5/2.7 MM Screw Depth Gauge
00-2360-040-35 3.5/2.7 MM LOCKING SCREW STANDARD
DEPTH GAUGE
Retrograde Femur Depth Gauge
00-2490-065-85 RETROGRADE FEMUR DEPTH GAUGE
Small Depth Gauge 60 mm
00-4810-002-01 SMALL DEPTH GAUGE 60 MM
T-Handle
00-8401-002-00 T-HANDLE
Suction Device
01.04239.850 SUCTION DEVICE
Modular Case Bracket
00-5907-083-72 MOD BRKT PSN MC TASP L 4-9 CD
00-5907-083-73 MOD BRKT PSN MC TASP L 4-11 EF
00-5907-083-74 MOD BRKT PSN MC TASP L 8-12 GH
00-5907-083-75 MOD BRKT PSN MC TASP L 12J
00-5907-071-04 PSN ¼ SIZE SHELL
00-5907-071-02 PSN 1/2 SIZE SHELL
00-5907-071-00 PSN FULL SIZE SHELL
00-5907-083-77 MOD BRKT PSN MC TASP R 4-9 CD
00-5907-083-78 MOD BRKT PSN MC TASP R 4-11 EF
00-5907-083-79 MOD BRKT PSN MC TASP R 8-12 GH
00-5907-083-80 MOD BRKT PSN MC TASP R 12J
Appendix A – Parts Index (cont.)
APPEN
DIX A
All content herein is protected by copyright, trademarks and other intellectual property rights, as applicable, owned by or licensed to Zimmer Biomet or its affiliates unless otherwise indicated, and must not be redistributed, duplicated or disclosed, in whole or in part, without the express written consent of Zimmer Biomet.
This material is intended for health care professionals, the Zimmer Biomet sales force and Zimmer Biomet employees. Distribution to any other recipient is prohibited.
Zimmer Biomet does not practice medicine. The information being presented is of a general nature and does not represent or constitute medical advice or recommendations and is for medical education purposes only.
©2017 Zimmer Biomet
1258.2-GLBL-en-REV0617
Authorized RepresentativeZimmer U.K. Ltd.9 Lancaster PlaceSouth Marston ParkSwindon, SN3 4FPUK
Legal ManufacturerZimmer, Inc.1800 West Center StreetWarsaw, Indiana 46580 USA
www.zimmerbiomet.comCE mark is not valid unless there is a CE mark on the product label.
0086