INSTRUCTIONS MANUAL
Free and total SO
automatic analyser2Free and total SO
automatic analyser2
·
·
·
·
·
·
·
4 glass tubes for Toning
4 rubber stoppers nº5
4 stirrer magnets 6x20 mm
1 toning burette + dispenser stopper
1 nut clip for burette
1 inox rack for 12 tubes
1 power voltage feeder
ref. 3026056
ref. 47185050
ref. 137115
ref. 3019065
ref. 63204115
ref. R292
ref. 0002065
·
·
·
·
·
1 titration liquid 500 mL
1 starch indicator 250 mL
1 acid 500 mL
1 alkali 500 mL
1 instructions manual
ref. 1006009
ref. 1005001
ref. 1006008
ref. 1006010
Included material
ref. 1010015
·
·
·
1 pipette 25 mL double line A class
2 pipettes 5 mL graduated
1 pipette 10 mL graduated
ref. 4018034
ref. 4018034
ref. 4018034
Required material
·
·
·
·
·
·
·
·
·
·
·
Ripper method
automatic dosification
fast
objetive
highly repetitive (precision s=+/- 2 mg/L)
simple and easy to use
very practical
without maintenance nor calibrations
use of conomic reagents
power voltage: 12V
compact design
Characteristics
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
175m
m
150
mm
165m
m
500
mm
Peso2,20 Kg
1011
06070809
Operative Technique
Free sulphurous:
05
Instructions
Assembly
01
02
03
04
Set the rod in the corresponding hole from the back and turn it
clockwise, later set the nut clip.
Place a 25 mL burette in the burette support. with
Connect the power voltage feeder and press . The leds will
shine on indicating current in the instrument. will remain
.
Place the glass tubes in the rack and put one stirrer magnet inside
Fill it Titration
Liquid (Potassium Iodate N/128).
power
Led #1
illuminated
With the help of pipettes add in one of the tubes
to analyse, and
Place the tube inside the Toning+.
till the limit of the upper side of the instrument.
Place the dispenser in the center of the tube like a stopper.
Press the key.A led will shine counterclockwise, the
sample will stir and will autotune. Once finished the autotune a
double “pip” will sound indicating the begining of the analysis.
, white ones, will remain illuminated indicating
. Otherwise and indicating in
adjustment. In this case .
The instrument will add reagent automatically
The end of analysis will show with
In this case 3 long beeps will sound indicating end of analysis.
25 mL of the
wine sample 1-2 mL of Starch Indicator 5 mL of
Acid.
Add distilled water
run/stop
Led
#1 and #2 adjustment
OK led #1 led #2 will blink error
go to observations
all leds illuminated during 10 sec.
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
14
15
16
17
18
With the help of pipettes add in one of the tubes
to analyse and .
Cover it with the rubber stopper, mix it and let it
in the rack.
Once time elapsed uncover it and add
and .
Follow
For further analysis follow .
25 mL of the
sample of wine 10 mL of Alkali
rest during 15
minutes
1-2 mL of Starch Indicator
5 mL of Acid
steps from 06 to 12
steps from 14 to 17
without adding distilled water and will
obtain mg Total SO2 /L.
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
Take note of the spent volume in the burette.
factor and will get the content of free sulphurous
in milligrams per liter [ ].
For further analysis follow .
The read volume is
multiplied by 10
mg Free So
steps from 05 to 122 /L
13
12
Observations
It is recommended to add the sample and reagents with the glass
tube inside the rack to avoid spill liquids inside the instrument.
The tube must be before each analysis.
after having added the Acid reagent.
It is important not to touch the tube once it is inserted inside the
instrument as glass walls from the tube are not completely regular
so a turn or movement will alter the detection of light beam.
clean and dry
Analyse immediately
Total sulphurous:
Observations
If led #1 and led #2 blink means that the sample of wine have not
been able to pass the autotune because it may be a very intensive in
colour wine (bigger than ) or very cloudy
(bigger than ). In such case
for the Free SO or for Total
SO with and where the
according to attached table from next page.The graphics show
an illustrative summary:
For the totals analysis it is not necessary to add water.
To the instrument press button during
.
Try to maintain always the same working conditions.
After 15 minutes of not working (no analysis done) the instrument
is auto shutdown.
button is used to set the level of the reagent where you need
it. Purge everyday you use it.Titration liquid must be in the tip of
the doser.
thus it can damage the silicon tube
inserted in the electrovalve and then damage the electovalve itself
inducing malfunction of the instrument.
2
2
15 units of intensity
3000 units of turbidity NTU repeat steps
05 to 12 repeat steps from 14 to 17
10 mL of sample multiplier factor will be
25
disconnect power 3
seconds
Purge
Attention: Do not use Iodine
I
0 15
25 mL 10 mL
NTU
0 3000
25 mL 10 mL
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
0301
10
03 02
08
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
02
06
09 01
12
10 20 25
IODINE N/50 64 32 25.6
IODINE N/128 25 12.5 10
TABLE OF MULTIPLIER COEFFICIENTS
SAMPLE VOLUME [mL]
Codes guide
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
Led #1
Current
run/stop
power
, white, lighted:
in the instrument.
Connect: press once.
Shut down: press during 3 seconds.
Led #1 Led #2 led
#3
OK
and , white, lighted and
, blue, blinking:
Adjustment . Beggining of analysis.
Led #1 Led #2
not OK
10 mL of sample multiplier factor 25
and , white, blinking:
Adjustment . Repeat analysis with
and .
Codes guide
Warranty
GAB Sistemática Analítica S.L.
Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436
www.gabsystem.com [email protected]
All blue leds lighted:
.Will keep like this till
or is pressed.
Read the burette spent volume and
multiply by the corresponding factor
according to table.
End of analysis
run/stop power
It covers all the manufacture defects, provided that no manipulation
has been done by not authorised personnel, during 2 years, except
defects induced for bad use.
All leds lighted:
: If it keeps like this
during 10 seconds the analysis will be
finished, otherwise 1 or 2 more drops of
Titration Liquid will be needed.
Testing end of analysis