INSTRUCTIONS MANUAL

Free and total SO

automatic analyser2Free and total SO

automatic analyser2

Page 2: INSTRUCTIONS MANUAL Free and total SO automatic …
Page 3: INSTRUCTIONS MANUAL Free and total SO automatic …

·

·

·

·

·

·

·

4 glass tubes for Toning

4 rubber stoppers nº5

4 stirrer magnets 6x20 mm

1 toning burette + dispenser stopper

1 nut clip for burette

1 inox rack for 12 tubes

1 power voltage feeder

ref. 3026056

ref. 47185050

ref. 137115

ref. 3019065

ref. 63204115

ref. R292

ref. 0002065

·

·

·

·

·

1 titration liquid 500 mL

1 starch indicator 250 mL

1 acid 500 mL

1 alkali 500 mL

1 instructions manual

ref. 1006009

ref. 1005001

ref. 1006008

ref. 1006010

Included material

ref. 1010015

·

·

·

1 pipette 25 mL double line A class

2 pipettes 5 mL graduated

1 pipette 10 mL graduated

ref. 4018034

ref. 4018034

ref. 4018034

Required material

·

·

·

·

·

·

·

·

·

·

·

Ripper method

automatic dosification

fast

objetive

highly repetitive (precision s=+/- 2 mg/L)

simple and easy to use

very practical

without maintenance nor calibrations

use of conomic reagents

power voltage: 12V

compact design

Characteristics

GAB Sistemática Analítica S.L.

Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436

www.gabsystem.com [email protected]

175m

m

150

mm

165m

m

500

mm

Peso2,20 Kg

Page 4: INSTRUCTIONS MANUAL Free and total SO automatic …

1011

06070809

Operative Technique

Free sulphurous:

05

Instructions

Assembly

01

02

03

04

Set the rod in the corresponding hole from the back and turn it

clockwise, later set the nut clip.

Place a 25 mL burette in the burette support. with

Connect the power voltage feeder and press . The leds will

shine on indicating current in the instrument. will remain

.

Place the glass tubes in the rack and put one stirrer magnet inside

Fill it Titration

Liquid (Potassium Iodate N/128).

power

Led #1

illuminated

With the help of pipettes add in one of the tubes

to analyse, and

Place the tube inside the Toning+.

till the limit of the upper side of the instrument.

Place the dispenser in the center of the tube like a stopper.

Press the key.A led will shine counterclockwise, the

sample will stir and will autotune. Once finished the autotune a

double “pip” will sound indicating the begining of the analysis.

, white ones, will remain illuminated indicating

. Otherwise and indicating in

adjustment. In this case .

The instrument will add reagent automatically

The end of analysis will show with

In this case 3 long beeps will sound indicating end of analysis.

25 mL of the

wine sample 1-2 mL of Starch Indicator 5 mL of

Acid.

Add distilled water

run/stop

Led

#1 and #2 adjustment

OK led #1 led #2 will blink error

go to observations

all leds illuminated during 10 sec.

GAB Sistemática Analítica S.L.

Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436

www.gabsystem.com [email protected]

Page 5: INSTRUCTIONS MANUAL Free and total SO automatic …

14

15

16

17

18

With the help of pipettes add in one of the tubes

to analyse and .

Cover it with the rubber stopper, mix it and let it

in the rack.

Once time elapsed uncover it and add

and .

Follow

For further analysis follow .

25 mL of the

sample of wine 10 mL of Alkali

rest during 15

minutes

1-2 mL of Starch Indicator

5 mL of Acid

steps from 06 to 12

steps from 14 to 17

without adding distilled water and will

obtain mg Total SO2 /L.

GAB Sistemática Analítica S.L.

Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436

www.gabsystem.com [email protected]

Take note of the spent volume in the burette.

factor and will get the content of free sulphurous

in milligrams per liter [ ].

For further analysis follow .

The read volume is

multiplied by 10

mg Free So

steps from 05 to 122 /L

13

12

Observations

It is recommended to add the sample and reagents with the glass

tube inside the rack to avoid spill liquids inside the instrument.

The tube must be before each analysis.

after having added the Acid reagent.

It is important not to touch the tube once it is inserted inside the

instrument as glass walls from the tube are not completely regular

so a turn or movement will alter the detection of light beam.

clean and dry

Analyse immediately

Total sulphurous:

Page 6: INSTRUCTIONS MANUAL Free and total SO automatic …

Observations

If led #1 and led #2 blink means that the sample of wine have not

been able to pass the autotune because it may be a very intensive in

colour wine (bigger than ) or very cloudy

(bigger than ). In such case

for the Free SO or for Total

SO with and where the

according to attached table from next page.The graphics show

an illustrative summary:

For the totals analysis it is not necessary to add water.

To the instrument press button during

.

Try to maintain always the same working conditions.

After 15 minutes of not working (no analysis done) the instrument

is auto shutdown.

button is used to set the level of the reagent where you need

it. Purge everyday you use it.Titration liquid must be in the tip of

the doser.

thus it can damage the silicon tube

inserted in the electrovalve and then damage the electovalve itself

inducing malfunction of the instrument.

2

2

15 units of intensity

3000 units of turbidity NTU repeat steps

05 to 12 repeat steps from 14 to 17

10 mL of sample multiplier factor will be

25

disconnect power 3

seconds

Purge

Attention: Do not use Iodine

I

0 15

25 mL 10 mL

NTU

0 3000

25 mL 10 mL

GAB Sistemática Analítica S.L.

Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436

www.gabsystem.com [email protected]

Page 7: INSTRUCTIONS MANUAL Free and total SO automatic …

0301

10

03 02

08

GAB Sistemática Analítica S.L.

Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436

www.gabsystem.com [email protected]

02

06

09 01

12

10 20 25

IODINE N/50 64 32 25.6

IODINE N/128 25 12.5 10

TABLE OF MULTIPLIER COEFFICIENTS

SAMPLE VOLUME [mL]

Page 8: INSTRUCTIONS MANUAL Free and total SO automatic …

Codes guide

GAB Sistemática Analítica S.L.

Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436

www.gabsystem.com [email protected]

Led #1

Current

run/stop

power

, white, lighted:

in the instrument.

Connect: press once.

Shut down: press during 3 seconds.

Led #1 Led #2 led

#3

OK

and , white, lighted and

, blue, blinking:

Adjustment . Beggining of analysis.

Led #1 Led #2

not OK

10 mL of sample multiplier factor 25

and , white, blinking:

Adjustment . Repeat analysis with

and .

Page 9: INSTRUCTIONS MANUAL Free and total SO automatic …

Codes guide

Warranty

GAB Sistemática Analítica S.L.

Sant Jordi, 30. Moja-Olèrdola 08734, BCN [Spain]Tel. +34 938 171 842 Fax. +34 938 171 436

www.gabsystem.com [email protected]

All blue leds lighted:

.Will keep like this till

or is pressed.

Read the burette spent volume and

multiply by the corresponding factor

according to table.

End of analysis

run/stop power

It covers all the manufacture defects, provided that no manipulation

has been done by not authorised personnel, during 2 years, except

defects induced for bad use.

All leds lighted:

: If it keeps like this

during 10 seconds the analysis will be

finished, otherwise 1 or 2 more drops of

Titration Liquid will be needed.

Testing end of analysis