Upload
anonymous-ahvji5kwaz
View
1.316
Download
318
Embed Size (px)
Citation preview
8/18/2019 Instructions for Use Cobas b121 en[1]
1/192
cobas b 121 systemInstructions for Use
8/18/2019 Instructions for Use Cobas b121 en[1]
2/192
COBAS, COBAS B and LIFE NEEDS ANSWERS
are trademarks of Roche.
©2008 Roche Diagnostics
Roche Diagnostics GmbH
D-68298 Mannheim
Germany
www.roche-diagnostics.com
8/18/2019 Instructions for Use Cobas b121 en[1]
3/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 1
cobas b 121 system
Revision History
Edition notice
cobas b 121 system
In the course of 2005 the Roche OMNI C system was rebranded under the RocheDiagnostics professional IVD user brand cobas®.
Systems with a serial number of 5000 or above are cobas b 121 system.
Systems with a serial number up to 4999 are Roche OMNI C systems.
Copyright
© 2008, Roche Diagnostics GmbH. All rights reserved
The contents of this document may not be reproduced in any form or communicatedto any third party without the prior written consent of Roche Diagnostics.While every effort is made to ensure its correctness. Roche Diagnostics assumes noresponsibility for errors or omissions which may appear in this document.Subject to change without notice.
Brands
COBAS, COBAS B, COBAS BGE LINK, LIFE NEEDS ANSWERS, ROCHE OMNI,AUTOQC, ROCHE MICROSAMPLER, COMBITROL and AUTO-TROL aretrademarks of Roche.
Manual Version Software Version Revision date Changes
2.0 October 2001
3.0 1.2 January 2002
4.0 1.31 December 2002
5.0 1.60 December 2004
6.0 1.70 May 2005 English only!
7.0 1.71 Juni 2006 cobas branding
8.0 1.71 Juni 2006
9.0 1.72 May 2007
10.0 1.74 June 2008 Change of the template
8/18/2019 Instructions for Use Cobas b121 en[1]
4/192
cobas b 121 system
Roche Diagnostics June 2008
2 Instructions for Use · Version 10.0
Contact addresses
Manufacturer
Edition
Revision 10.0, June 2008
First edition: October 2001
REF/No. 03260992001
Roche Diagnostics GmbH
D-68298 Mannheim / Germany
www.roche.com
8/18/2019 Instructions for Use Cobas b121 en[1]
5/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 3
cobas b 121 system
Table of contents
Revision History 1
Edition notice 1Copyright 1Brands 1Contact addresses 2Edition 2Table of contents 3Preface 5How to use this manual 5Where to find information 5Conventions used in this manual 5
Introduction and specifications Part A
1 Safety information
Important information A-5Operating safety information A-6
2 General description
Introduction A-9General notes A-10Measurement and calibration procedure A-12Safety instructions for specific dangers A-13System description A-15
3 Installation and shutdown
Installation A-21Shutdown A-36
4 Specifications
Performance data A-47Default and input values A-49Sample throughput A-49Sample volume A-49Sample types A-50Calibrations A-50Environmental parameters A-51
Product data A-53Printer A-53Screen A-54Barcode scanner A-54
Operation Part B
5 Measurement
Preanalytics B-5Measuring procedure B-8
6 Quality control
Quality control - general B-17General QC concept B-17
Important information concerning the analysis
of QC measurement results B-19Material setup B-20QC measurement B-26Multirules B-28QC consequences B-30QC unlock B-31QC troubleshooting B-32
7 Calibration
Calibration - general B-37
8 Operating modes
Operating modes - general B-43User interface B-43Analyzer mode B-49Software overview B-61
Maintenance Part C
9 Maintenance
Maintenance - general C-5Decontamination C-5Daily C-7Weekly C-8
Semi annual C-9Sample-dependent maintenance procedures C-10Unscheduled C-17Additional maintenance procedures C-30Maintenance overview C-31Maintenance scheduler C-32
Troubleshooting Part D
10 Troubleshooting
Troubleshooting - general D-5
System stops D-5List of system stops D-7Electrode status D-14
Appendix Part E
11 List of consumables
Order information E-5
Index Part F
Index F-3
8/18/2019 Instructions for Use Cobas b121 en[1]
6/192
Roche Diagnostics June 2008
4 Instructions for Use · Version 10.0
cobas b 121 system
8/18/2019 Instructions for Use Cobas b121 en[1]
7/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 5
cobas b 121 system
Preface
This manual has detailed descriptions of cobas b 121 system features and general
operational concepts, specification functions and use of controls, operatingtechniques, emergency procedures, product labeling and maintenance procedures.
How to use this manual
To help you find information quickly, there is a table of contents at the beginning ofthe book and each chapter. In addition, a complete index can be found at the end.
Where to find information
In addition to the Instructions for Use, the following documents are also provided toassist in finding desired information quickly:
o cobas b 121 system Reference Manual
o cobas b 121 system Short Instruction
Conventions used in this manual
Visual cues are used to help locate and interpret information in this manual quickly.This section explains formatting conventions used in this manual.
Symbols Helping to locate and interpret information in this manual the following symbols areused:
Keep this manual in a safe place to ensure that it is not damaged and remains available for
use.
This Instructions for Use should be easily accessible at all times.
Symbol Description
a Procedural step
o List item
e Cross-reference
h Call up of screen
Note
8/18/2019 Instructions for Use Cobas b121 en[1]
8/192
cobas b 121 system
Roche Diagnostics June 2008
6 Instructions for Use · Version 10.0
IVD symbols The IVD symbols are used in accordance with DIN EN 980:2003, ISO 15223:2000 (asper the In Vitro Diagnostics Directive 98/79/EC), and DIN EN ISO 780:1997.
Attention, Warning, Danger
All sections / passages that are marked with this symbol describe
procedures and/or indicate conditions or dangers that could damage
or lead to a malfunction in the cobas b 121 system, and which
therefore should never be attempted and/or must be observed to
avoid potential injuries (to patients, users and third parties).
Risk of infection!
All sections and parts of texts that are marked with this symbol
describe procedures that may involve risk of infection.
Symbol Description
Symbol Description
This product complies with the requirements in the guideline for In
Vitro Diagnostic 98/79/EC.
Lot designation
Consumables: use by,.. (expiry date)
The consumables must be completely consumed by the indicated
date.
If a day is not indicated, apply the last day of the respective month.
Store at ,..The conditions necessary to preserve the product's shelf life before
opening.
For in vitro diagnostic use
"Grüner Punkt" (in Germany)
Manufacturer — according to In Vitro Diagnostic guidelines 98/79/
EG
Store upright
Risk of infection!
(according to the standard DIN EN 61010-2-101:2002) (Instrument)
Risk of infection!
(according to the standard DIN ISO 15223-1:2005)(Consumables)
Catalogue number
8/18/2019 Instructions for Use Cobas b121 en[1]
9/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 7
cobas b 121 system
Other symbols The following symbols are listed as additional information:
Caution (refer to accompanying documents). Please refer to safety-
related notes in the manual, accompanying this instrument.
Please consult instructions for use
Serial number (model plate)
Symbol Description
Symbol Description
Electrodes:This date indicates the limit of the maximum storage time of
an electrode. The electrode must be installed in the instrument
no later than the imprinted date.
If the installation takes place on the imprinted date, it still falls
within the specifications. The calculation of the “Install
before” date is based on the production date of the elctrode.
Danger symbol: "Irritant" (on the label and the packaging of
C3 Fluid Pack)
Rating: Although not corrosive, momentary, longer-lasting, or
repeated contact with skin or mucous membrane may result in
inflammation. Danger of sensitization during contact with
skin (when classified with R 43).
Caution: Avoid contact with eyes and skin, do not inhale
vapors.
Do not use content if the packaging is damaged
Protective gloves, protective goggles and suitable protective
clothing must be worn.
8/18/2019 Instructions for Use Cobas b121 en[1]
10/192
cobas b 121 system
Roche Diagnostics June 2008
8 Instructions for Use · Version 10.0
Abbreviations The following abbreviations are used:
Abbreviation Definition
A
ANSI American National Standards Institute
AQC Automatic Quality Control
B
BG Blood gas
C
CLIA Clinical Laboratory Improvement Amendments
CLSI Clinical and Laboratory Standards Institute
cond Conductivity
CSA Canadian Standards Association
C1 Calibration Solution 1
C2 Calibration Solution 2
C3 Fluid Pack
D
dBA Decibel weighted against the A-frequency response curve. This curve
approximates the audible range of the human ear.
DIL Diluent
DNS Domain Name Server
E
EC European community
e.g. exempli gratia – for example
EN European standard
F
FMS Fluid mixing system
H
Hct Hematrocrit
HIV Human immunodeficiency virus
HW Hardware
I
i.e. id est – that is to say
ISE Ion selective electrode
IVD In vitro Diagnostic Directive
L
LCD Liquid cristal display
LIS Laboratory Information System
LJ Levey Jennings
M
MAC Media Access Control
MC Measuring chamber
MCon Measuring contact
MSDS Material safety data sheet
MSS Metabolite sensitive sensor
MV Mean value
8/18/2019 Instructions for Use Cobas b121 en[1]
11/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 9
cobas b 121 system
e For writing the measuring, calculated and input values see Chapter 8 Operating modes >
Parameter on page B-73!
P
PP Peristaltic pump
Q
QC Quality control
R
REF Reference solution
S
SIP Sample inlet path
SD Standard deviation
SDC Sample distributor cartridge
SO2 Oxygen saturation
T
tHb Total hemoglobinU
UL Underwriters Laboratories Inc.
V
VDE Association of German Electrical Engineers (Verband Deutscher
Elektrotechniker)
Abbreviation Definition
http://../cobas%20b%20221%20System/Operator%20Manual/Book%20DE/Softwaremodi.pdfhttp://../cobas%20b%20221%20System/Operator%20Manual/Book%20DE/Softwaremodi.pdfhttp://../cobas%20b%20221%20System/Operator%20Manual/Book%20DE/Softwaremodi.pdf
8/18/2019 Instructions for Use Cobas b121 en[1]
12/192
cobas b 121 system
Roche Diagnostics June 2008
10 Instructions for Use · Version 10.0
8/18/2019 Instructions for Use Cobas b121 en[1]
13/192
June 2008
1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
2 General description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7
3 Installation and shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-19
4 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-45
Introduction and specifications A
8/18/2019 Instructions for Use Cobas b121 en[1]
14/192
8/18/2019 Instructions for Use Cobas b121 en[1]
15/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-3
cobas b 121 system 1 Safety information
Table of contents
Safety information
The information provided in this chapter is essential for the safe, trouble-free
operation of the instrument and must be read and understood by the user.
Important information ...............................................................................................A-5
Operating safety information .....................................................................................A-6
In this chapter Chapter 1
8/18/2019 Instructions for Use Cobas b121 en[1]
16/192
Roche Diagnostics June 2008
A-4 Instructions for Use · Version 10.0
1 Safety information cobas b 121 system
Table of contents
8/18/2019 Instructions for Use Cobas b121 en[1]
17/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-5
cobas b 121 system 1 Safety information
Important information
Important information
These Instructions for Use contain vital warnings and safety information.
This instrument is intended to be used only for the specialized purpose described in
the instructions. The most important prerequisites for use, operation, and safety areexplained to ensure smooth operation. No warranty or liability claims will be coveredif the machine is used in ways other than those described or if the necessaryprerequisites and safety measures are not observed.
The instrument may be operated only by persons whose qualifications enable them tocomply with the safety measures that are necessary during operation of theinstrument.
Suitable protective equipment, like laboratory clothing, protective gloves, protectivegoggles and if necessary mouth protectors, must be worn to prevent direct contactwith biological working materials. In addition, a face mask is required if there is a
risk.
Adjustments and maintenance performed with covers removed and power connectedmay be attempted only by a qualified technician who is aware of the associateddangers.
Instrument repairs are to be performed only by the manufacturer or qualified servicepersonnel.
Only accessories and supplies either delivered by or approved by Roche are to be usedwith the instrument. These items are manufactured especially for use with thisinstrument and meet the highest quality requirements.
Operation of the instrument with solutions whose composition is not consistent with
that of the original solutions can negatively affect the long-term measurementaccuracy. Deviations in the composition of the solutions can also decrease the servicelife of the electrodes.
For safety reasons, quality control measurements must be performed daily. Since themeasurements of the instrument depend not only on the correct characteristicfunction, but also on a series of marginal conditions (e.g. pre-analysis), resultsobtained from the instrument should be submitted for an expert opinion beforetaking additional measures based on the supplied measurements.
WARNING
Caution (refer to accompanying documents)!
Please refer to safety-related notes in the manual accompanying this instrument.
8/18/2019 Instructions for Use Cobas b121 en[1]
18/192
Roche Diagnostics June 2008
A-6 Instructions for Use · Version 10.0
1 Safety information cobas b 121 system
Operating safety information
Operating safety information
The instrument has been constructed and tested according to the following European
Standards:
o IEC/EN 61010-1:2001
o IEC/EN 61010-2-101:2002
o IEC/EN 61010-2-081:2002 + A1:2003
It was delivered from the factory in flawless condition with regards to safety features.In order to preserve this condition and ensure safe operation, the user must respectthe notices and warnings that are contained in these Instructions for Use.
o This instrument is classified under the protection class I according toIEC/EN 61010-1.
o The instrument meets the conditions for overvoltage category II.
o The instrument meets the conditions for contamination level 2.o Do not operate the instrument in an explosive environment or in the vicinity of
explosive anesthetic mixtures containing oxygen or nitrous oxide.
o If objects or liquids enter the internal areas of the instrument, remove theinstrument from its power supply and allow an expert to check it thoroughlybefore using it again.
o The instrument is suitable for long-term operation indoors.
WARNING
o The power cord must be plugged into a grounded power receptacle. When using an
extension cord, make sure it is properly grounded.
o Any rupture of the ground lead inside or outside the instrument or a loose ground
connection may result in hazardous operating conditions for the operating personnel.Intentional disconnection of the grounding is not permitted.
o The instrument is not suitable for operation with a direct current power supply.
o Use only the original power plug delivered with the cobas b 121 system.
8/18/2019 Instructions for Use Cobas b121 en[1]
19/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-7
cobas b 121 system 2 General description
Table of contents
General description
This chapter contains a general description of the instrument, as well as
precautionary measures against special dangers and the proper handling of sensorsand solutions.
Introduction ................................................................................................................A-9
General notes .............................................................................................................A-10
Application area ...................................................................................................A-10
Operating instructions ........................................................................................A-10Important buttons on the screen ........................................................................A-10
Measurement and calibration procedure .................................................................A-12
Measurement procedure .....................................................................................A-12
Calibration procedure .........................................................................................A-12
Measurement evaluation .....................................................................................A-12
Safety instructions for specific dangers ....................................................................A-13
Handling samples ................................................................................................A-13
Disposal of waste water, bottles, electrodes, and the instrument ..................... A-13
Decontamination ................................................................................................A-13
Handling solutions ..............................................................................................A-13
Handling electrodes ............................................................................................A-14System description ....................................................................................................A-15
Visual identification ............................................................................................A-15
Screen ...................................................................................................................A-15
Printer ..................................................................................................................A-15
Measurement chamber ........................................................................................A-16
Pump ....................................................................................................................A-16
Flap .......................................................................................................................A-16
Bottle compartment ............................................................................................A-16
Power supply ........................................................................................................A-16
Position of the power switch .........................................................................A-16
Reverse side ..........................................................................................................A-17Interface .........................................................................................................A-17
In this chapter Chapter 2
8/18/2019 Instructions for Use Cobas b121 en[1]
20/192
Roche Diagnostics June 2008
A-8 Instructions for Use · Version 10.0
2 General description cobas b 121 system
Table of contents
Barcode scanner .............................................................................................A-18
Warning and identification labels .................................................................A-18
8/18/2019 Instructions for Use Cobas b121 en[1]
21/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-9
cobas b 121 system 2 General description
Introduction
Introduction
The cobas b 121 system is a modular analyzer for measuring blood gases, electrolytes,total hemoglobin, oxygen saturation and hematocrit in whole blood, serum, plasma,acetate and bicarbonate containing dialysis solutions, and QC materials.
It is possible to complete database procedures or to make simultaneous adjustmentsduring measurement or calibration.
The individual, mutually independent operating modes are defined as follows:
o Analyzer: measuring, QC, system, calibration, commonly used functions
o Database: data about patients, measurement, calibration, QC, and the instrument
o Setup: instrument settings
o Info: Roche info, version number, fill levels, help, sensor report
The instrument is currently available in the following configuration:
o BG (pH, P O2, P CO2) / ISE (Na+, K+, Cl–, Ca2+) and tHb/SO2
A new fluid calibration system eliminates the need for expensive calibration gases.This change results in easier handling, a smaller footprint, and reduced costs.
An easily understood "Touch Screen" interface facilitates easy operation and savescostly and time-consuming user training.
The patented electrodes are completely maintenance-free, and only require a verysmall sample volume.
Figure A-1
8/18/2019 Instructions for Use Cobas b121 en[1]
22/192
Roche Diagnostics June 2008
A-10 Instructions for Use · Version 10.0
2 General description cobas b 121 system
General notes
General notes
Application area
The instrument is designed to measure BG / ISE / tHb/SO2 in whole blood. Theaccuracy of measurement values is checked accordingly.
In order to achieve accurate measurements of recommended aqueous controlsolutions (with regards to deviations from biological samples), choose the propercomponents and make the corresponding corrections in the QC measurement mode.
The accuracy of measurement values of undefined aqueous solutions cannot beguaranteed (e.g. due to the possibility of interfering components and/or missing orinsufficient buffer systems, and/or differences in ionic strength and diffusionpotential when compared to biological samples).
Operating instructions
The instrument should be powered on at all times!
If the instrument will remain turned off for longer than 24 hours, it is necessary tocarry out shutdown procedures.
e For more information, see Chapter 3 Installation and shutdown!
Prevent fluids leaking inside the analyzer.
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each exchangeof solutions and packs and after startup of the instrument.
Additionally complete a quality control test on one level between two automatic2P calibrations. The levels have to be alternated (low, normal, high).
e For more information, see Chapter 6 Quality control !
Important buttons on the screen
Buttons Description
Operating mode "Analyzer"
Operating mode "Database"
Operating mode "Setup"
Button "Info"
More functions
8/18/2019 Instructions for Use Cobas b121 en[1]
23/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-11
cobas b 121 system 2 General description
General notes
e For more information, see Chapter 8 Operating modes, section Buttons on page B-47!
Confirm
Cancel
Activate / deactivate
Start action
Buttons Description
8/18/2019 Instructions for Use Cobas b121 en[1]
24/192
Roche Diagnostics June 2008
A-12 Instructions for Use · Version 10.0
2 General description cobas b 121 system
Measurement and calibration procedure
Measurement and calibration procedure
Measurement procedure
Use the following methods are used to determine the various measurement variables:
P O2 : Use of the Clark measurement principle: measurement of current generated bythe reduction of oxygen
P CO2 : Use of the Severinghouse principle: potentiometric measurement of the pHchange in the electrode caused by CO2.
pH, Na+, K+, Ca2+ and Cl– are potentiometric electrodes. Special glasses are used asthe sensitive element for pH and Na+. The potassium and calcium membranescontain special neutral carriers. A special ion exchanger is used for chloridemembranes. Calculation of these variables also requires the use of a reference
electrode - a permanently contacted chloride electrode in the instrument.
tHb /SO2 : Light absorption in whole blood is measured at four different wavelengths,whereby on one hand the sample is subjected to light radiation and on the other handthe dispersed light is also evaluated.
Hematocrit : Measurement of the sample's conductivity
Calibration procedure
tHb and SO2 are factory calibrated. Ambient air and a zero point solution are used tocalibrate oxygen. The remaining parameters are calibrated using two solutions, which
are mixed in various ratios. Gas containers are not used. The use of at least twocalibration points for each measurement variable and constant internal monitoring ofthe calibrations ensures the accuracy and precision of the measurement values.
Measurement evaluation
The results from the instrument must be carefully examined by a clinical-medicalspecialist who will take the patient's clinical condition into consideration before anyclinical decisions are reached based on the test results.
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each exchangeof solutions and packs and after startup of the instrument.
Additionally complete a quality control test on one level between two automatic2P calibrations. The levels have to be alternated (low, normal, high).
e For more information, see Chapter 6 Quality control !
8/18/2019 Instructions for Use Cobas b121 en[1]
25/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-13
cobas b 121 system 2 General description
Safety instructions for specific dangers
Safety instructions for specific dangers
Handling samples
Respect the necessary hygienic regulations when handling samples because samplesmay contain dangerous viruses.
e For more detailed information, see Chapter 5 Measurement !
Disposal of waste water, bottles, electrodes, and the instrument
Decontamination
The purpose of this procedure is to minimize risk when handling items that were incontact with biological samples.
Roche recommends following a decontamination procedure in addition toregulations specific to the laboratory.
These decontamination procedures should be performed periodically to minimize therisk of infections (incl. hepatitis virus and HIV).
e For more detailed information about decontamination, see Chapter 9 Maintenance!
Handling solutions
Store the instrument solutions according to the specified packaging requirements.The solutions should be adapted to the ambient temperature before use.
The shelf life of the solutions is limited.
Please read the bottle label and the packaging for the correct storage temperature andthe maximum shelf life.
Dispose waste water, bottles, electrodes and the instrument according to local and/or
labour regulations (biologically contaminated—hazardous waste!)
Always wear gloves!
ATTENTION
CAUTION! DO NOT FREEZE!
If frozen, the solution's concentration may change and cause calibration errors!
Do not use damaged C3 Fluid Packs! Do not mix the individual components!
8/18/2019 Instructions for Use Cobas b121 en[1]
26/192
Roche Diagnostics June 2008
A-14 Instructions for Use · Version 10.0
2 General description cobas b 121 system
Safety instructions for specific dangers
Handling electrodes
Store the electrodes according to the packaging specifications.
The shelf life of the electrodes is limited.
See the electrode label and the packaging for the correct storage temperature and themaximum shelf life.
NOTE IMPORTANT! Installation note for theP CO2 electrode
Insert the electrode into the measuring chamber within 5 minutes of opening the ALU-PE
packaging.
A special protective gas atmosphere designed to condition the P CO2 electrode during
storage is found inside the ALU-PE packaging.
This gas atmosphere ensures immediate potential stability during insertion of the electrode
into the measurement chamber and/or immediate readiness for measuring after the
2P calibration is completed.
If more than 5 minutes elapse after opening the ALU-PE packaging, the level of gas
conditioning could be lost and the time required for the first-time calibration could be
increased.
8/18/2019 Instructions for Use Cobas b121 en[1]
27/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-15
cobas b 121 system 2 General description
System description
System description
Visual identification
Screen
All information (results, error messages, warnings, etc.) is displayed on the screen.The screen consists of a 5.7" colour LCD that is covered with a touch-sensitive film("touch screen").
Printer
The low-noise 2" thermal printer with integrated paper cutter for roll paper is locatedunderneath the printer cover.
A Printer cover
B Pump
C Power pack / main switch
D Unlocking knob for the AutoQC module
E Bottle compartment
F Measurement chamber with the electrodes
and tHb/S O2 module
G Flap
H Contrast setting
I Screen
Figure A-2
A
C
B
D
E
I
F
H
G
8/18/2019 Instructions for Use Cobas b121 en[1]
28/192
Roche Diagnostics June 2008
A-16 Instructions for Use · Version 10.0
2 General description cobas b 121 system
System description
Measurement chamber
The measurement chamber with the electrodes and the tHb/SO2 module are locatedbeneath the instrument cover.
The electrodes are flow-through electrodes with a visible sample channel.
The tHb/SO2-module is an optical sensor module for determining the levels of totalhemoglobin (tHb) and oxygen saturation (SO2) in whole blood.
Pump
A peristaltic pump transports the sample and the operating fluids inside theinstrument.
Flap
When opening the flap, notice two definitive locking positions:
o Flap position 1 (half opened) – Syringe mode—for syringes and ampoules
o Flap position 2 (completely opened) – Capillary mode—for capillaries andRoche MICROSAMPLERs
Bottle compartment
The calibration solutions and the waste container are located behind the bottle
compartment cover.
Power supply
This unit contains the power switch and the power connector.
Position of the power switch
Power switch "OFF" Power switch "ON"
Figure A-3
8/18/2019 Instructions for Use Cobas b121 en[1]
29/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-17
cobas b 121 system 2 General description
System description
Reverse side
Interface
Only data processing units manufactured according to the standards IEC 950(UL1950) may be attached to the interface connections!
COM 1 This interface can be assigned to a ticket printer and Host FMT.
e For an exact description of the assignment, see Reference Manual chapter Setup, section
Interfaces > COM 1.
COM 2 This interface can be assigned as serial interface to ASTM.
e For an exact description of the assignment, see Reference Manual chapter Setup, section
Interfaces > COM 2.
A COM 1
B COM 2
C Barcode Scanner
D Network connection
Figure A-4
COM 1* RS 232 Schnittstelle
COM 2* RS 232 Schnittstelle
Barcode PS/2 DIN - 6p Buchse
Network 10BaseT Ethernet (RJ45)
A B C D
*TIP:
Always use a filter adapter when using the serial interface. Order this part from your
customer service representative!
8/18/2019 Instructions for Use Cobas b121 en[1]
30/192
Roche Diagnostics June 2008
A-18 Instructions for Use · Version 10.0
2 General description cobas b 121 system
System description
Barcode scanner
o Scanning of solution data (type, lot number, expiration date current composition,etc.)
o Scanning of electrode data (type, lot number, expiration date)
o Scanning of patient or user identity
o Scanning of QC data (QC material, lot number, basis, expiration date,target values, etc.)
o Scanning of desired alphanumeric code
e For further information, see the enclosed manual for the PS2 hand-held scanner (included
in scope of delivery).
Warning and identification labels
Figure A-5
ATTENTION
Press the button on the underside to activate the scanner! A beeping sound and a brief
illumination of the LED on the upper side indicate the successful scanning of the barcode.
Figure A-6 Identification label
A Identification label
Figure A-7
A
8/18/2019 Instructions for Use Cobas b121 en[1]
31/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-19
cobas b 121 system 3 Installation and shutdown
Table of contents
Installation and shutdown
In this chapter, the software-guided installation and shutdown of the instrument are
described step by step. The sequence of the steps described must be strictly followed.
Installation .................................................................................................................A-21
Location ...............................................................................................................A-21
Accessories ...........................................................................................................A-22
Installation ...........................................................................................................A-22
1. Attach power cord and barcode scanner ..................................................A-22
2. Installing the AutoQC module (Automatic Quality Control Module) ..A-23
3. Switch on ....................................................................................................A-25
4. Select language ...........................................................................................A-25
5. Setting the date and time ............................ ............................. ................. A-25
6. Check/Activate the parameters tHb and SO2 ...........................................A-25
7. Checking the barometer value ..................................................................A-26
8. Checking the AutoQC module .............................. ................................. .. A-26
9. Installation .................................................................................................A-26
10. Check tubes at V2 and V9 .......................................................................A-26
11. Insert FMS tubes (Fluid Mixing System) ........................... .................... A-27
12. AutoQC module (option) .......................................................................A-27
13. Attach pump tube ....................................................................................A-28
14. Insert needle and fill port holder ............................................................A-28
15. Insertion of printer paper ............................. .............................. ............ A-29
16. Insertion of bottles .............................. ............................. ....................... A-30
17. Insert electrodes .......................................................................................A-32
18. Insertion of the reference electrode ........................................................A-33
19. Begin installation routines ......................................................................A-35
20. Quality control ............ .............................. ............................ .................. A-35
Shutdown ...................................................................................................................A-36
Less than 24 hours ...............................................................................................A-36
Longer than 24 hours ..........................................................................................A-36
1. Remove bottles ...........................................................................................A-36
In this chapter Chapter 3
8/18/2019 Instructions for Use Cobas b121 en[1]
32/192
Roche Diagnostics June 2008
A-20 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Table of contents
2. Remove the V2 and V9 tube ends from the guides ..................................A-37
3. Insert the shutdown kit and connect it ............................... ..................... A-37
4. Remove the shutdown kit ..........................................................................A-38
5. Remove the waste container ......................................................................A-38
6. Return the V2 and V9 tube ends to the guides ............................. ........... A-387. Remove the pump tube .............................................................................A-39
8. Remove the electrodes ...............................................................................A-39
9. Remove needle and fill port ......................................................................A-40
10. Remove the printer paper ............................... .............................. .......... A-40
11. Remove the tube under V1 and V2 ........................... ............................. A-41
12. Shutting down the AutoQC module ............................. ......................... A-41
13. Uninstall the AutoQC module ................................................................A-42
8/18/2019 Instructions for Use Cobas b121 en[1]
33/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-21
cobas b 121 system 3 Installation and shutdown
Installation
Installation
Location
For best results, choose a suitable, level location that is not subject to direct sunlightfor the device.
When installing an instrument that was stored in a cool room or was transported atlow temperatures, be aware that condensation may have formed and could causedisturbances to the instrument. Be sure that the instrument is climatized for at leastone hour at room temperature before beginning operation.
The following conditions must be fulfilled:
o Ambient temperature: + 15 °C to + 33 °C
o Ambient air pressure: 462 - 800 mmHg (61.63 - 106.60 kPa)
o Avoid direct sunlight, vibration and strong electromagnetic fields (electricmotors, transformers, X-ray equipment, cellular phones, ...).
o Use a stable and level work surface (max. 1° incline with bottles installed).
o Relative humidity:
O 20 to 95 % (at ≥ +15 °C to ≤ 31 °C)
O 20 to 90 % (at > +31 °C to ≤ +33 °C)
o Allow sufficient free space around the instrument for air circulation and the
electrical connections.o Check for the correct voltage: 100 to 240 VAC (±10%)
After setting up the instrument at a location that meets the necessary conditions,execute the following steps to ensure the instrument is ready for operation:
o Check that the instrument and all accessories are present by comparing contentswith the delivery packing slip. If anything is missing, inform your Rocherepresentative immediately.
o Inspect the instrument and the accessories for damage. If damage is noted, informthe transportation company immediately. Retain the packaging material andproducts as evidence for the damage claim.
An atmospheric pressure of 462 mmHg (61.63 kPa) corresponds to approx. 4.000 m above
sea level.
8/18/2019 Instructions for Use Cobas b121 en[1]
34/192
Roche Diagnostics June 2008
A-22 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Installation
Accessories
The following parts are delivered as standard equipment with the instrument:
Installation
1. Attach power cord and barcode scanner
1 Plug the power cord into the power supply (Figure A-8/1).
2 Connect the barcode scanner and. if necessary, the network cable (Figure A-8/2)to the respective interface(s) (Figure A-8/2) on the reverse side of the instrument.
3 Open the bottle compartment cover and the docking mechanism.
1 barcode scanner 1 shutdown kit1 power cord 1 micro electrode dummy
2 pump tubes 2 fill port holders
1 drip tray 2 needles
1 paper roll 2 MCon
(1) Power supply (2) Interfaces
Figure A-8
A Docking mechanism
Figure A-9
A
8/18/2019 Instructions for Use Cobas b121 en[1]
35/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-23
cobas b 121 system 3 Installation and shutdown
Installation
a If available (optional equipment), install the AutoQC module.
2. Installing the AutoQC module (Automatic Quality Control Module)
1 Open the AutoQC cover and remove the transport safety band and both safetyscrews (see Figure A-11/1).
2 Remove the red plastic relief clamp by lifting from the AutoQC valve V17, which islocated toward the back of the AutoQC unit on the inside (see Figure A-11/2).
ATTENTION
Attempt to install and operate an AutoQC module only if the instrument is prepared for usewith the AutoQC module. The valves V12 and V13 must be present in the bottle compart-
ment (see Figure A-10)!
Not prepared Prepared
Figure A-10
ATTENTION
Check to be sure that the power cord is connected to the power supply. After engaging the
AutoQC module, it will no longer be possible to connect the cord to the power supply!
The safety screws are intended to fix the longitudinal and transversal slides and should
only be re-inserted whenever the AutoQC module will be transported.
(1) (2)
A Transport safety band B Red plastic relief clamp
Figure A-11
A
B
8/18/2019 Instructions for Use Cobas b121 en[1]
36/192
Roche Diagnostics June 2008
A-24 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Installation
3 On the right side of the instrument. remove the gray plastic cover from theAutoQC docking port. Close the AutoQC cover and push the AutoQC moduleinto the tracks on the instrument until it locks into place. The module's dockingport will slide into the opening on the side wall on the instrument and lock intoplace.
4 Connect both tube ends in the bottle compartment to the docking part accordingto Figure A-13.
5 Connect the cable of the AutoQC module to the jack on the reverse side of theinstrument.
A AutoQC docking port
Figure A-12
A
A
Figure A-13
8/18/2019 Instructions for Use Cobas b121 en[1]
37/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-25
cobas b 121 system 3 Installation and shutdown
Installation
3. Switch on
Switch the instrument on and wait until the program has loaded completely and hasstarted. The instrument is in the "System stop" mode. Before beginning the start upprocedures, you must select the language with which the instrument will be operated,set the date and time, verify/activate the parameters tHb and SO2, check the baro-meter value and if the AutoQC module is activated.
4. Select language
1 Press:
h [Setup] > [Instrument] > [Language]
2 Select the language and confirm your selection.
e For a detailed description, see Reference Manual chapter Setup!
5. Setting the date and time
Press:
h [Setup] > [Times & intervals] > [Date/Time]
e For a detailed description, see Reference Manual chapter Setup!
6. Check/Activate the parameters tHb and S O2
Press:
h [Setup] > [Parameters] > [Misc. settings] > [Act. / deact. f. measurement]
e For a detailed description, see Reference Manual chapter Setup!
Figure A-14
Should a hardware problem occur during the booting process, the system is stopped and
an error message is given.
The Customer Service should be informed if the error occurs again.
8/18/2019 Instructions for Use Cobas b121 en[1]
38/192
Roche Diagnostics June 2008
A-26 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Installation
7. Checking the barometer value
Press:
h [More functions] > [System] > [Test] > [Control Sensors] > [Barosensor]
If the barometer value deviates by more than +/- 2 mmHg from the value indicated by
a precision barometer, it will be necessary for technical support to calibrate yourbarometer!
8. Checking the AutoQC module
Press:
h [Setup] > [Instrument] > [AutoQC]
If the AutoQC module is not activated, press this button.
Changes will be accepted after reboot.Press [Yes].
9. Installation
Press:
h [More functions] > [System] > [Tools] > [Installation]
o Follow the instructions on the screen!
Confirm every step by pressing the [Confirm] button.
10. Check tubes at V2 and V9
Check the bottle compartment's back wall to make sure that the air mixture tube(valve V2) and the ventilation tube (valve V9) are in the guides (see Figure A-15).
A Air mixture tube B Ventilation tube
Figure A-15
A
A
B
8/18/2019 Instructions for Use Cobas b121 en[1]
39/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-27
cobas b 121 system 3 Installation and shutdown
Installation
11. Insert FMS tubes (Fluid Mixing System)
1 Slide the feed tube of the C1 solution under the tube clip (valve V1).
2 Slide the feed tube of the C2 solution under the tube clip (valve V2)(see Figure A-16).
3 Both of these valves are located on the back wall of the bottle compartment.
12. AutoQC module (option)
1 Open the cover of the AutoQC module.
2 Insert the AutoQC ampoule block.
3 Close the cover of the AutoQC module.
4 Perform the mat assignments and program the AutoQC times.
e See Chapter 6 Quality control !
A Tube of the C1 solution B Tube of the C2 solution
Figure A-16
B
A
8/18/2019 Instructions for Use Cobas b121 en[1]
40/192
Roche Diagnostics June 2008
A-28 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Installation
13. Attach pump tube
1 Open the instrument cover.
2 Open the peristaltic pump's clear plastic cover (tension lever). Push the linearbracket (white plastic part) upwards (Figure A-17).
3 Place the tube around the rolling wheel.
4 Close the clear plastic cover (tension lever). The tubing holder is then pressed intothe sealer.
14. Insert needle and fill port holder
1 Open the flap to the labelled capillary position (completely open position).
2 Insert the needle to the left into the sealing piece and then push the needlecarefully down until it clicks into place.
3 Rotate the fill port holder so that the underside shows on the top (seeFigure A-19/1) and then push this over the needle (see Figure A-19/2).
4 Snap the fill port holder evenly onto the axis (see Figure A-19/3).
Pump closed
A Tension lever
B Pump head
C Linear clamp
Figure A-17
A
B
C
Figure A-18
Do not bend the needle when pushing it up!
8/18/2019 Instructions for Use Cobas b121 en[1]
41/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-29
cobas b 121 system 3 Installation and shutdown
Installation
5 Close the flap.
15. Insertion of printer paper
1 Open the printer cover.
2 Place the new paper roll into the holder.
3 Make sure that the printer lever is in the "down" position (see Figure A-20).
4 Cut off, at a right angle, the start of the paper.
5 Feed in the start of the paper according to the sticker on the inside of the printercover. The paper is automatically pulled into the printer.
(1) (2) (3)
Figure A-19
The printer paper is heat sensitive on one side only. Please make sure that you insert the
paper roll correctly.
A Printer lever
Figure A-20
A
8/18/2019 Instructions for Use Cobas b121 en[1]
42/192
Roche Diagnostics June 2008
A-30 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Installation
6 Close the printer cover and feed the paper outward through the slit in the printercover.
16. Insertion of bottles
1 To do this. place the bottle tool on the screw cap of the C1 or C2 (Figure A-22/1).
2 Press the grips together and press the transparent disk downward Figure A-22/2).
3 Rotate the transparent disk in the clockwise direction. Stop when you notice (aftera short distance) resistance (Figure A-22/3).
4 Scan the barcodes on the bottles of C1 calibration solution 1. C2 calibrationsolution 2 and the C3 fluid pack (the waste water bottle does not have a barcode)(see Figure A-23).
A Slit in the printer cover.
Figure A-21
A
If your facility is 3000 meters above sea level or higher deaerate the bottle before inserting
to avoid splashing the C1 and C2 solutions.
(1) (2) (3)
Figure A-22
8/18/2019 Instructions for Use Cobas b121 en[1]
43/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-31
cobas b 121 system 3 Installation and shutdown
Installation
5 The instrument recognizes the correct solution and checks the expiration date.On the screen, the respective bottle starts blinking. If the bottle has passed the
expiration date, the screen displays a warning.
6 Confirm with this button.
7 Insert the bottle completely into the appropriate position, following theinstructions on the docking mechanism (see Figure A-24/2). The bottles areopened automatically.
8
Close the docking mechanism.Check the positions of the bottles by matching the labels on the dockingmechanisms to the labels on the bottles (see Figure A-24/3).
9 Close the bottle compartment cover.
Figure A-23
Remove rubber sealings from C3 Fluid Pack before inserting it!
(1) (2)
Figure A-24
8/18/2019 Instructions for Use Cobas b121 en[1]
44/192
Roche Diagnostics June 2008
A-32 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Installation
17. Insert electrodes
1 Open the measuring chamber cover.
2 Open locking lever.
3 Follow the instructions on the screen.
4 If necessary, remove air bubbles by holding the electrode vertically and by tappinglightly with your fingernail against the electrode body (see Figure A-26).
5 Insert the electrodes, beginning at the right and proceeding left according to thecolour code.
6 Push all electrodes slightly to the right so that they are lined up together withoutgaps.
Figure A-25
Make sure that no air bubbles have formed in the inner electrolytes of the electrodes (see
Figure A-26). If there are air bubbles between the contact pin and the membrane, there
will not be effective electrical conduction. Result: calibration and measurement errors!
A Fee of air bubbles
Figure A-26
A
8/18/2019 Instructions for Use Cobas b121 en[1]
45/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-33
cobas b 121 system 3 Installation and shutdown
Installation
18. Insertion of the reference electrode
1 Insert the reference electrode.
2 Place the tube into the tube guide slot (seeFigure A-28/2).
3 Affix the white plug at the end of the reference electrode tube (see Figure A-29).
Figure A-27
Be sure that the tube lies precisely in the guide slot. Otherwise it may be pinched, thereby
preventing aspiration of the reference solution.
Result: calibration and measurement errors!
(1) (2)
A Locking lever
B Reference electrode
Figure A-28
A B
Push the plug in completely.
8/18/2019 Instructions for Use Cobas b121 en[1]
46/192
Roche Diagnostics June 2008
A-34 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Installation
4 Close the locking lever.
5 Inspect the electrical contact of the electrodes by checking if they fit tightly. Thecorrect position of the various electrodes is easy to recognize by looking at thecolours of the contact strips or at their labels.
6 Scan the barcodes located on the inner packaging of each electrode or enter thebarcodes manually with the help of the keyboard.
7 The corresponding electrode starts blinking on the screen.Confirm with this button.
A White plug at the end of the reference
electrode
Figure A-29
Figure A-30
A A
Always save the inside packaging materials! Whenever the instrument is taken out of
service and then brought back into service, the barcodes of the remaining functional
electrodes must be scanned again!
Figure A-31
8/18/2019 Instructions for Use Cobas b121 en[1]
47/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-35
cobas b 121 system 3 Installation and shutdown
Installation
8 Close the measuring chamber and then the instrument cover.
19. Begin installation routines
1 Press [Start].
2 All solutions are aspirated and the system is calibrated; this may take longer thanthe respective actions during operation.
3 Installation is complete.
20. Quality control
1 You must define the material before executing a quality control measurement.
e For instructions, see Chapter 6 Quality control !
2 Perform a quality control on all 3 levels (low, normal, high).Make sure that the results agree with the target values.
e See Chapter 6 Quality control !
If an error occurs during one of the steps, a system stop is displayed, but the instrument
has been brought "into operation".
See Chapter 10 Troubleshooting , for instructions on clearing the system stop.
8/18/2019 Instructions for Use Cobas b121 en[1]
48/192
Roche Diagnostics June 2008
A-36 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Shutdown
Shutdown
Less than 24 hours
If the instrument will be needed within the next 24 hours, press:
h [More functions] > [System] > [Tools] > [Software shutdown]
Switch off the instrument.
Longer than 24 hours
If the instrument will be shut down for longer than 24 hours, perform the following
procedure.
e See Chapter 8 Operating modes, section Database > Data export – PCMCIA card (for
example) on page B-55
e See Chapter 9 Maintenance, section Decontamination on page C-5 for a detailed
description!
While in the "Analyzer" operating mode, press:
h [More functions] > [System] > [Tools] > [Shutdown]
Follow the instructions on the screen.
Confirm every step by pressing this button!
1. Remove bottles
Open the bottle compartment cover and remove bottles C1, C2 and C3.
Before shutting down the instrument, backup the data to a PCMCIA card or an interface.
Roche Diagnostics recommends decontaminating all surfaces and tubing before shutting
down the instrument.
8/18/2019 Instructions for Use Cobas b121 en[1]
49/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-37
cobas b 121 system 3 Installation and shutdown
Shutdown
2. Remove the V2 and V9 tube ends from the guides
1 Remove the air mixture valve tube (valve V2) from the guide.
2 Remove the ventilation tube (valve V9) from the guide.
3 Place both ends of the tube onto an absorbent pad, such as a paper towel or gauze(see Figure A-33).
3. Insert the shutdown kit and connect it
1 Fill the shutdown kit about halfway with distilled water.
2 Insert the shutdown kit into space C3.
3 Connect the tubes of the set with the connectors from C1 and C2.
A Ventilation tube V9 B air mixture valve tube V2
Figure A-32
A End of the tubes
Figure A-33
A
A
B
A A
8/18/2019 Instructions for Use Cobas b121 en[1]
50/192
Roche Diagnostics June 2008
A-38 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Shutdown
4 Begin the tube washing procedure by pressing the [Start ] button.
4. Remove the shutdown kit
1 Remove the shutdown kit.
2 Start the procedure for emptying the tubes by pressing the [Start] button.
5. Remove the waste container
Remove the waste container (W waste container).
6. Return the V2 and V9 tube ends to the guides
1 Return the air mixture valve tube (valve V2) to the guides.
2 Return the ventilation tube (valve V9) to the guides (see Figure A-35).
3 Close the docking mechanisms.
Figure A-34
Remove the remaining fluid from the tubes of the shutdown kit by briefly holding the tubes
vertically, allowing the fluid to run back into the container.
A Ventilation tube B Air mixture valve tube
Figure A-35
A
A
B
8/18/2019 Instructions for Use Cobas b121 en[1]
51/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-39
cobas b 121 system 3 Installation and shutdown
Shutdown
7. Remove the pump tube
1 Open the instrument cover.
2 Open the peristaltic pump's clear plastic cover (tension lever). Push the linearclamp (white plastic piece) upwards (see Figure A-36).
3 You can now remove the entire tube set (tube holder with tubes)(see Figure A-36).
4 Close the tension lever (clear plastic cover).
8. Remove the electrodes1 Open the measuring chamber cover.
2 Open the locking lever.
3 Remove the electrodes.
4 Close the locking lever and the measurement chamber cover.
A Tension lever
B Pump head
C Linear clamp
Figure A-36
B
A
C
A Locking lever B
Figure A-37
A
8/18/2019 Instructions for Use Cobas b121 en[1]
52/192
Roche Diagnostics June 2008
A-40 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Shutdown
9. Remove needle and fill port
1 Open the flap to the labelled capillary position (completely open position).
2 Rotate the fill port holder in the direction of the arrow (see the marking on the fillport holder). It will come out of the axis and snap out of place (seeFigure A-38/1).
3 Carefully remove the fill port holder from the needle (see Figure A-38/2).
4 Remove the needle.
5 Push the needle to the left and into the sealer. Then carefully pull it upward(see Figure A-38/3).
6 Close the flap.
10. Remove the printer paper
1 Open the printer cover.
2 Pull out the printer paper to the rear and remove it.
3 Close the printer cover.
Do not bend the needle!
(1) (2) (3)
Figure A-38
8/18/2019 Instructions for Use Cobas b121 en[1]
53/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-41
cobas b 121 system 3 Installation and shutdown
Shutdown
11. Remove the tube under V1 and V2
1 Slide the feed tube of the C1 solution under the tube clip (valve V1) and out.
2 Slide the feed tube of the C2 solution under the tube clip (valve V2) and out.
3 Pressure is removed from the tubes.
If available (option):
12. Shutting down the AutoQC module
1 Open the cover and remove the AutoQC ampoule block.2 The carriage in the AutoQC module moves to the service position.
3 Close all covers.
4 Switch the instrument OFF.
5 Shut down is complete.
A Feed tube C1 solution B feed tube: C2 solution
Figure A-39
B
A
After successfully shutting down the instrument, it will be in the system stop mode (shut
down). Only a renewed installation procedure can alter this status.
8/18/2019 Instructions for Use Cobas b121 en[1]
54/192
Roche Diagnostics June 2008
A-42 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Shutdown
13. Uninstall the AutoQC module
1 Pull the cable of the AutoQC module from the jack on the reverse side of theinstrument (see Figure A-40).
2 Remove both tube ends in the bottle compartment from the docking port.
3 Unlock the AutoQC module by pulling the unlocking knob (see Figure A-40) onthe reverse side of the instrument and remove the AutoQC module from thetracks on the instrument.
A Unlocking knob
Figure A-40
A Tube ends B
Figure A-41
A
A
A
8/18/2019 Instructions for Use Cobas b121 en[1]
55/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-43
cobas b 121 system 3 Installation and shutdown
Shutdown
4 Insert the red plastic relief clamp for the AutoQC valve V17.
5 If you plan to transport the instrument, be sure to unplug the power cord, thescanner, and the network cable. Then install the transport safety device for theAutoQC module.
6 Use the original packaging when transporting the instrument!
A Relief clamp for the AutoQC valve V17
Figure A-42
A Transport safety band
Figure A-43
A
A
8/18/2019 Instructions for Use Cobas b121 en[1]
56/192
Roche Diagnostics June 2008
A-44 Instructions for Use · Version 10.0
3 Installation and shutdown cobas b 121 system
Shutdown
8/18/2019 Instructions for Use Cobas b121 en[1]
57/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-45
cobas b 121 system 4 Specifications
Table of contents
Specifications
In this chapter, the performance data, as well as product and environmental data are
described.
Performance data ......................................................................................................A-47
Measured parameters ..........................................................................................A-47
Default and input values ...........................................................................................A-49
Sample throughput ...................................................................................................A-49
Sample volume ..........................................................................................................A-49
Sample types ..............................................................................................................A-50
Calibrations ...............................................................................................................A-50
Environmental parameters .......................................................................................A-51
Temperature / humidity / stability .............................. ............................. .......... A-51
Product data ..............................................................................................................A-53
Electrical data ......................................................................................................A-53
Classification (according IEC/ISO) ....................................................................A-53
Dimensions ..........................................................................................................A-53
Weight ..................................................................................................................A-53
Printer ........................................................................................................................A-53
Screen .........................................................................................................................A-54
Barcode scanner ........................................................................................................A-54
In this chapter Chapter 4
8/18/2019 Instructions for Use Cobas b121 en[1]
58/192
Roche Diagnostics June 2008
A-46 Instructions for Use · Version 10.0
4 Specifications cobas b 121 system
Table of contents
8/18/2019 Instructions for Use Cobas b121 en[1]
59/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-47
cobas b 121 system 4 Specifications
Performance data
Performance data
Measured parameters
specified
for
Specified range Precision(a) (b)
(within-run)
Precision (a) (b)
(day-day)
Accuracy(a)
P O2 B 0 - < 60 mmHg SD < 4.0 mmHg SD < 6.0 mmHg < ± 8.0 mmHg
60 - 140 mmHg SD < 2.0 mmHg SD < 3.0 mmHg ± 4.0 mmHg
> 140 - 500 mmHg SD < (2% - 0.8 mmHg) SD < (4% - 2.6 mmHg) < ± (6% - 4.4 mmHg)
> 500 - 800 mmHg SD < (4% - 11.8 mmHg) SD < (8% - 22.6 mmHg) < ± (12% - 34.4 mmHg)
0 - > 7.998 kPa SD < 0.533 kPa SD > 0.800 kPa < ± 1.066 kPa
7.998 - 18.662 kPa SD < 0.267 kPa SD < 0.400 kPa ± 0.533 kPa
> 18.662 - 66.650 kPa SD < (2% - 0.107 kPa) SD < (4% - 0.347 kPa) < ± (6% - 0.587 kPa)> 66.650 - 106.640 kPa SD < (4% - 1.573 kPa) SD < (8% - 3.013 kPa) < ± (12% - 4.586 kPa)
Q 0 - < 60 mmHg SD < (10.7 - 12%) mmHg SD < (12.2 - 12%) mmHg < ± (16.8 - 13%) mmHg
60 - 200 mmHg SD < 3.5 mmHg SD < 5.0 mmHg ± 9.0 mmHg
> 200 - 800 mmHg SD < (7% - 10.5) mmHg SD < (7% - 9.0) mmHg < ± (6..5% - 4) mmHg
0 -< 7.998 kPa SD < (1.426 - 12%) kPa SD < (1.626 - 12%) kPa < ± (2.239 - 13%) kPa
7.998 - 26.660 kPa SD < 0.467 kPa SD < 0.667 kPa ± 1.200 kPa
26.660 - 106.640 kPa SD < (7% - 1.400) kPa SD < (7% - 1.200) kPa < ± (6.5% - 0.533) kPa
P CO2 B/Q 4 - < 15 mmHg SD < 2 mmHg SD < 3 mmHg ± 4 mmHg
15 - 80 mmHg SD < 1.5 mmHg SD < 2.0 mmHg ± 2.5 mmHg
> 80 - 200 mmHg SD < 4.5 mmHg SD < 6 mmHg ± 8 mmHg0.533 - < 2.00 kPa SD < 0.267 kPa SD < 0.400 kPa ± 0.533 kPa
2.00 - 10.664 kPa SD < 0.200 kPa SD < 0.267 kPa ± 0.333 kPa
10.664 - 26.660 kPa SD < 0.600 kPa SD < 0.800 kPa ± 1.066 kPa
pH B/Q 6.0 - < 6.8 SD < 0.020 SD < 0.035 ± 0.06
6.8 - 7.6 SD < 0.008 SD < 0.015 ± 0.02
> 7.6 - 8.0 SD < 0.015 SD < 0.030 ± 0.04
Na+ B/S/A/D/Q 20 - < 120 mmol/L SD < 4.5 mmol/L SD < 6 mmol/L ± 8 mmol/L
120 - 170 mmol/L SD < 1.5 mmol/L SD < 2.0 mmol/L ± 2.5 mmol/L
> 170 - 250 mmol/L SD < 6 mmol/L SD < 8 mmol/L ± 10 mmol/L
K +
B/S/A/D/Q 0.2 - < 3.0 mmol/L SD < 0.15 mmol/L SD < 0.35 mmol/L ± 0.5 mmol/L3.0 - 6.0 mmol/L SD < 0.06 mmol/L SD < 0.15 mmol/L ± 0.2 mmol/L
> 6.0 - 20 mmol/L SD < 0.4 mmol/L SD < 1 mmol/L ± 1.4 mmol/L
Cl- B/S/A/D/Q 20 - < 70 mmol/L SD < 3 mmol/L SD < 4 mmol/L ± 8.0 mmol/L
70 - 130 mmol/L SD < 1.5 mmol/L SD < 2.0 mmol/L ± 4.0 mmol/L
> 130 - 250 mmol/L SD < 4.5 mmol/L SD < 6 mmol/L ± 12.0 mmol/L
Table A-1
8/18/2019 Instructions for Use Cobas b121 en[1]
60/192
Roche Diagnostics June 2008
A-48 Instructions for Use · Version 10.0
4 Specifications cobas b 121 system
Performance data
Ca2+ B/S/A/D/Q 0.1 - < 0.6 mmol/L SD < 0.06 mmol/L SD < 0.1 mmol/L ± 0.2 mmol/L
0.6 - 1.5 mmol/L SD < 0.03 mmol/L SD < 0.05 mmol/L ± 0.1 mmol/L
> 1.5 - 4.0 mmol/L SD < 0.15 mmol/L SD < 0.25 mmol/L ± 0.5 mmol/L
0.4008 - < 2.4048 mg/dL SD < 0.2405 mg/dL SD < 0.4008 mg/dL ± 0.8016 mg/dL
2.4048 - 6.0120 mg/dL SD < 0.1202 mg/dL SD < 0.2004 mg/dL ± 0.4008 mg/dL
> 6.0120 - 16.0320 mg/dL SD < 0.6012 mg/dL SD < 1.0020 mg/dL ± 2.0040 mg/dL
Hct B/Q 10 - < 20% SD < 3% SD < 5% ± 6.0%
20 - 60% SD < 1.5% SD < 2.5% ± 3.0%
< 60 - 80% SD < 3% SD < 5 % ± 6.0%
0.100 - < 0.200 SD < 0.030 SD < 0.050 ± 0.060
0.200 - 0.600 SD < 0.015 SD < 0.025 ± 0.030
< 0.600 - 0.800 SD < 0.030 SD < 0.050 ± 0.060
tHb B 3 - < 6 g/dL SD< (3.0% + 0.27 g/dL) SD < (3.3% + 0.30 g/dL) < ± (-3.3% + 0.70 g/dL)
6 - 18 g/dL SD < 0.45 g/dL SD < 0.5 g/dL < ± 0.5 g/dL
> 18 - 25 g/dL SD < (2.1% + 0.06 g/dL) SD < (2.1% + 0.11 g/dL) 180.00 - 250.00 g/L SD < (2.1% + 0.60 g/L) SD < (2.1% + 1.10 g/L)
8/18/2019 Instructions for Use Cobas b121 en[1]
61/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-49
cobas b 121 system 4 Specifications
Default and input values
Default and input values
Sample throughput
Sample volume
Parameter Default values Range (default values) Range (input values)
tHb 15.0 g/dL 11,0 - 16.0 g/dL 1.0 - 26.0 g/dL
150.0 g/L 110.0 - 160.0 g/L 10.0 - 260.0 g/L
9.3 mmol/L 6.8 - 9.9 mmol/L 6.2 - 161.3 mmol/L
P 50 Adult 26.7 mmHg not editable not editable
3.56 kPa
P 50 Fetal 21.5 mmHg not editable not editable
2.87 kPa
F IO2 0.21 0.1 - 1.0 0.1 - 1.0
R 0.84 0.7 - 2.0 0.7 - 2.0
Patient temperature 37.0 °C 2.0 - 44.0 °C 2.0 - 44.0 °C
98.6 °F 35.6 - 111.0 °F 35.6 - 111.0 °F
Table A-2
Activated / installed
electrodes
Typical sample throughput
[samples / hours]
Syringe Capillary
BG - tHb/SO2 30 30
BG - ISE - tHb/SO2 30 30
Table A-3
If the sample contains a higher Hct concentration, a sample volume of up to 70 µL is
required!
Activated / installed electrodes Typical sample volume
[µL](a) Typical specimen volume
(capillary position)
[µL](b)
Typical specimen volume
(syringe position)
[µL](c)
BG - ISE - tHb/SO2 60 68 90
(a) typical for Hct≤ 45%
(b) Volume limitation through sample sensor: If the specified volume is reached, an aspiration stop occurs (approx. 1 second). If the capillary is notremoved, the aspiration process continues.
(c) No aspiration stop in the syringe position.
8/18/2019 Instructions for Use Cobas b121 en[1]
62/192
Roche Diagnostics June 2008
A-50 Instructions for Use · Version 10.0
4 Specifications cobas b 121 system
Sample types
Sample types
o Whole blood
o Serum
o Plasma
o Dialysis solutions containing acetate
o Dialysis solutions containing bicarbonate
o QC material
Calibrations
Calibrations Time intervals Length (typical)
[min]
Sys cal every 24 hours
(alternatively 8, 12 or 24 hours)
< 15
1P cal every 30 minute (alternatively 1 hour) < 2
2P cal every 12 hours
(alternatively 4, 8, 12 hours)
< 7
Warm-up phase when turning (a)
(a) incl. calibration
< 25
Warm-up phase power fail < 1 minute < 2
Electrode exchange as needed < 27
Table A-4
8/18/2019 Instructions for Use Cobas b121 en[1]
63/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-51
cobas b 121 system 4 Specifications
Environmental parameters
Environmental parameters
Temperature / humidity / stability
Instrument
Electrodes
Operating conditions
o Ambient temperature +15 °C to +33 °C
o Ambient air pressure 462 - 800 mmHg (61.63 - 106.60 kPa)(a)
(a) TIP: An atmospheric pressure of 462 mmHg (61.63 kPa) corresponds to approx. 4,000 m above sea level.
o Sea level -400 m to +4000 m
o Relative humidity 20 - 95%, if T > +15 °C to ≤ +31 °C
20 - 90%, if T > +31°C to +15 °C to ≤ +31 °C
20 - 90%, if T > +31 °C to
8/18/2019 Instructions for Use Cobas b121 en[1]
64/192
Roche Diagnostics June 2008
A-52 Instructions for Use · Version 10.0
4 Specifications cobas b 121 system
Environmental parameters
Solutions
QC material
Operating conditions
o Ambient temperature +15°C to +33 °C
o Relative humidity 20 - 95%, if T > +15 °C to ≤ +31 °C
20 - 90%, if T > +31 °C to
8/18/2019 Instructions for Use Cobas b121 en[1]
65/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 A-53
cobas b 121 system 4 Specifications
Product data
Product data
Electrical data
Classification (according IEC/ISO)
Dimensions
Instrument
AutoQC-Modul
Weight
Printer
Mains voltage range: 100 to 240 VAC (±10% permissible tolerance)
Frequency: 50/60 Hz
Required power: 150 W
Protection class: I
Overvoltage category: IIContamination level: 2
Width: 35.4 cm
Heigth: 46.7 cm
Depth: 41.0 cm
Width: 19.5 cmHeigth closed: 19.8 cm (open: 38.7 cm)
Depth: 39.5 cm
cobas b 121 system (instrument): about 17 kg (without calibration solutions)
cobas b 121 system ready for shipment: about 23 kg
AutoQC module: 5.9 kg (without ampoule mats)
Type: thermal printer with integrated paper cutter
Resolution: 12 dots/ mm
Full graphics: 576 dots / line
Printing speed: 8 mm / sec
Paper width: 58 mm
Paper length: ca. 30 m
8/18/2019 Instructions for Use Cobas b121 en[1]
66/192
Roche Diagnostics June 2008
A-54 Instructions for Use · Version 10.0
4 Specifications cobas b 121 system
Screen
Screen
Barcode scanner
Type: integrated flat LCD screen
Format: 5.7 inch
Resolution: 320 x 240 pixel
Type: PS2 hand-held scanner with integrated decoder
Reading speed: up to 45 scans/sec.
Resolution: 0.1 mmReading distance: up to 5 cm
Reading width: up to 8 cm
Preprogrammed code types(a):
(a) Further available barcode types can be programmed in accordance with the enclosed manual of thePS2 hand-held scanner (included in scope of delivery).
o China Postal Code
o Codabar
o Code 39
o Code 128
o EAN-8
o EAN-13
o EAN-128
o Interleaved 2 of 5
o UPC-A
o UPC-E
8/18/2019 Instructions for Use Cobas b121 en[1]
67/192
June 2008
5 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
6 Quality control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-15
7 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-35
8 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-41
Operation B
8/18/2019 Instructions for Use Cobas b121 en[1]
68/192
8/18/2019 Instructions for Use Cobas b121 en[1]
69/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 B-3
cobas b 121 system 5 Measurement
Table of contents
Measurement
In this chapter, all information necessary for carrying out measurements is described.
Preanalytics .................................. ................................. ............................... ................ B-5
Sample collection ............................. ................................ ................................ ..... B-5
Sample acquisition ............................ ............................... ............................... B-5
Acceptable anticoagulants ........................... ............................. ....................... B-5
Sample collection especially for tHb, SO2 and Hct measurement ................ B-5
Sample collection containers ....................................... ............................... .......... B-6
Syringes ............................... ................................. ............................... ............. B-6
Capillary tubes ............................ .................................. ................................. .. B-6
Roche MICROSAMPLER® ........................... ............................ ...................... B-6
Clot catcher .......................... ................................ .............................. .............. B-6
Sample handling .............................. ................................ ................................ ...... B-7
Whole blood ......................................................... .............................. ............. B-7
Serum ............................. ................................ .............................. .................... B-7
Plasma ............................ ................................ ............................. ..................... B-7
Measuring procedure ........................... ................................ ................................ ....... B-8
Syringe mode ........................... ................................ .............................. ................ B-9
Capillary mode ............................ ................................ ............................. ........... B-10
Data input .............................. ................................ .............................. ................ B-12
Results display / measurement report ........................... ............................. ........ B-13
In this chapter Chapter 5
8/18/2019 Instructions for Use Cobas b121 en[1]
70/192
Roche Diagnostics June 2008
B-4 Instructions for Use · Version 10.0
5 Measurement cobas b 121 system
Table of contents
8/18/2019 Instructions for Use Cobas b121 en[1]
71/192
Roche Diagnostics June 2008
Instructions for Use · Version 10.0 B-5
cobas b 121 system 5 Measurement
Preanalytics
Preanalytics
Sample collection
e Please refer to CLSI document M29-A3, "Protection of Laboratory Workers from
occupationally acquired infections." Approved guidelines - Third Edition 2005.
Sample acquisition
Only qualified personnel may perform the collection of blood needed for analyticalpurposes.
e For detailed information about blood sampling, storage and handling, see for example
CLSI document H11-A4, "Procedures for the collection of arterial blood specimes;
Approved Standard (Fourth Edition 2004)".
Acceptable anticoagulants
The only clot inhibitors that may be used for analyses in the cobas b 121 system areheparin salts. Other clot inhibitors, such as EDTA, citrate, oxalate, fluoride, andammonium-based materials have significant influence on the blood's pH and otherparameters and may not be used for this reason.
Sample collection especially for tHb, S O2 and Hct measurement
Whole blood, especially for the analysis of tHb, SO2 and Hct, must be thoroughly
mixed immediately before analysis in order to achieve consistent distribution of redblood cells and plasma before insertion of the sample.
Carefully rotate the sample about two axes using your hand or a mechanical device orinsert a metal disk or ball in the syringe before collecting the sample. Shortly beforeusing the sample, carefully shake the syringe. The up and down motion of the disk orball inside the syringe cylinder ensures consistent mixing.
e Refer to CLSI document C46-A "Blood gas an pH analysis related measurements;
Approved Guideline 2001".
Follow the usual applicable safety precautions when drawing blood samples. When
handling blood samples, there always exists the danger of transmission of HIV, hepatitis B
and C viruses or other pathogens transmissible by blood. Employ suitable blood sampling
techniques in order to reduce risk to personnel.
Suitable protective equipment, like laboratory clothing, protective gloves, protective
goggles and if necessary mouth protectors, must be worn to prevent direct contact with
biological working materials. In addition, a face mask is required if there is a risk.
The puncture site may never be squeezed! Mixing the blood sample with tissue fluid may
lead to the premature onset of clotting despite sufficient heparinization of the sample
collection containers! Incorrect sample collection or the use of an unsuitable sample
collection container may lead to errors and discrepancies in the measurement values.
8/18/2019 Instructions for Use Cobas b121 en[1]
72/192
Roche Diagnostics June 2008
B-6 Instructions for Use · Version 10.0
5 Measurement cobas b 121 system
Preanalytics
Sample collection containers
Syringes
If using another manufacturer's product with liquid heparin as a clot inhibitor, thecollection container should not be larger than required for the blood volume. Thiswill minimize the effects of the clot inhibitor on the thinning of the blood. The use ofplastic syringes is common, but there are cases when the use of plastic syringes is notappropriate, for example, when P O2 values are expected to be outside the normalrange. If very high P O2 values are expected, the sample should be analyzed as quicklyas possible after the specimen collection.
Capillary tubes
The capillary tubes must have a minimum volume of 100 μ L.
Capillary tubes with ceramic sealing caps should not be used because the fracture thatforms when opening the capillary can damage the fill port of the instrument. Use onlycapillary tubes with heat-treated ends to avoid damage to the instrument. When usingstirring rods like those offered by a few manufacturers, remove these rods beforeinserting the sample in order to avoid clogging the instrument.
Roche MICROSAMPLER®
The Roche MICROSAMPLER, which consists of two capillary tubes (115 μL each) ina plastic container, is ideally suited to atraumatic arterial blood collection.
Each laboratory should document the permissibility of sample containers that areused. These products vary from manufacturer to manufacturer and sometimes fromlot to lot.
Clot catcher
The use of a clot catcher is recommended to prevent clogging of the sample path dur-ing measurement of critical