INF380 - Proteomics-6 1 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable for protein and peptide analyzes. They are easy to handle, and rapid results are obtained. The sample for MALDI-TOF instruments preferably should contain the peptides from a single protein or a very low number of proteins (two or three at maximum). Thus, extensive sample preparation must be performed before the analysis in the instrument. Analysis by MALDI instruments are therefore often combined with 2-D gel electrophoresis of the sample to separate the proteins, and single spots are picked for proteolytic treatment.

INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

Embed Size (px)

Citation preview

Page 1: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 1

INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF

• The MALDI-TOF instruments are the simplest MS instruments suitable for protein and peptide analyzes.

• They are easy to handle, and rapid results are obtained. • The sample for MALDI-TOF instruments preferably should contain

the peptides from a single protein or a very low number of proteins (two or three at maximum).

• Thus, extensive sample preparation must be performed before the analysis in the instrument.

• Analysis by MALDI instruments are therefore often combined with 2-D gel electrophoresis of the sample to separate the proteins, and single spots are picked for proteolytic treatment.

Page 2: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 2

MALDI TOF - properties• MALDI instruments are robust, fast and easy to use.• Not all of the peptides become ionized. There is a ``competition'' between the peptides for

being ionized, and which peptides that are ionized depend partly on the matrix used. • The ionization process by MALDI is not fully understood, but several studies have

investigated how the ionization depends on peptide properties. • MALDI is able to ionize molecules of masses from a few hundred Da up to several

hundred kDa, although the mass range studied in each single analysis is much more restricted.

• It is usually known whether we study peptides or large proteins, and the instrument settings can be optimized for each purpose. The lower mass range is limited by noise generated from matrix ions and other unwanted ions.

• Almost all of the ionized peptides have a single positive charge due to the acquisition of a proton. As a general rule, only very large peptides or proteins may pick up two or three protons. The spectra are therefore relatively easy to interpret.

• The accuracy in the peptide mass area (500-6,000 Da) is good. The exact accuracy obtainable is dependent on the instrument and the calibration procedures. When internal calibration is used, the accepted accuracy during database searches is often set to 50 ppm, meaning that the difference between the experimental $m/z$ value and the theoretical m/z value must be less than 50 ppm.

• The resolution is good, meaning that all newer instruments can distinguish the isotopic peaks in the peptide mass range and correctly assign the monoisotopic peak.

• The peptides ionized by MALDI are generally quite stable, and there is little fragmentation. • There can be contaminants: the matrix (which is unavoidable), contaminants introduced

by sample handling, and contaminants due to insufficient sample separation.• It is possible to connect to other ``intelligent systems'' (robots) for automatic proteomics. • MALDI instruments are generally quite sensitive.

Page 3: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 3

TOF analyzers• Time-of-flight analyzers are the dominating analyzers for single MS. • When connected to a MALDI ionization source, the instrument is

called a MALDI-TOF mass spectrometer. • The simplest form of TOF analyzers is the linear TOF• Ions are sent to the analyzer in short pulses due to the pulses of the

laser and the synchronized changes in the electric field.• The ions are accelerated in the electric field. Let the potential of the

acceleration field be P, the velocity at the end of the acceleration v, the distance to the detector d, and e the charge of an electron.

Page 4: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 4

TOF analyzers• Time-of-flight analyzers are the dominating analyzers for single MS.

• When connected to a MALDI ionization source, the instrument is called a MALDI-TOF mass spectrometer.

• As described in the previous Chapter, resolution of an MS instrument is its ability to discriminate between different components with small differences in the m/z values. We will illustrate how this is related to time

and masses for TOF analyzers.

Page 5: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 5

TOF analyzers• Another important factor in relation to resolution is that not all ions of a peptide

(or several peptides of the same m/z) will reach the detector at exactly the same time.

• Therefore, there will be a spreading in the measured time (and therefore the calculated m/z) of a peptide.

• The reasons for different detected time are mainly because not all ions come from exactly the same point in the matrix/sample, and they are spread out in a three-dimensional space when they enter the electrical field.

• Thereby the velocities after acceleration may vary slightly. • Overlaps may therefore occur between different peptides

Page 6: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 6

TOF analyzers

Two ways are commonly used to reduce the spreading (and thus increasing the resolution): Delayed extraction and reflectron.

• Delayed (pulsed) extraction When the peptides are ejected from the plate due to the laser pulse, they exhibit a range of velocities, and would reach the detector at different times. To remedy for this, a delayed extraction voltage can be used. When this voltage is turned on, the ions with lowest velocities get greatest acceleration voltage, such that all ions (of the same $m/z$) reach the detector at (approximately) the same time.

Page 7: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 7

TOF analyzers• Reflectron is an electrostatic mirror placed at the end of the linear flight tube. When

the ions enters the reflectron, they are exposed to another electric field that forces them to turn around and fly back, slightly angled relative to the direction they came from.

• The reflectron has two effects. First, it increases the flying distance d by sending the ions back through another TOF analyzer that has a detector placed at the far end. Second, the reflectron is able to 'collect' the ions of the same m/z that had slightly different velocities in the first drift tube.

• All advanced MALDI-TOF instruments have the possibility to choose between linear TOF and reflectron TOF in addition to delayed extraction.

Page 8: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 8

Constructing the peak list• The raw data spectrum contains signal from peptides from the sample

proteins, and signal due to different forms of noise. • Revealing the signals (peaks) corresponding to the sample peptides

is a multistep task. • The detailed procedures for peak list construction varies with the

instruments and programs used, and the type of spectra. • The main challenges are to remove noise peaks without removing

any of the peptide peaks

Page 9: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 9

Constructing the peak list• The main challenges are to remove noise peaks without removing

any of the peptide peaks, and to determine the m/z and intensity values with the best accuracy possible.

• Noise may be divided into two main types. – Chemical noise may have several sources.

• It may come from chemical contaminants introduced during the sample handling,

• It may be proteins or peptides that unintentionally have been introduced during sample handling, for example human keratins from our skin or hair.

• In MALDI-TOF analyzes it may also come from the instability of peptides after receiving energy from the laser pulse, resulting in for example fragmentation

• In all cases unexpected peaks are produced. – Electronic noise is the result of electronic disturbances, and occurs with

random fluctuations between the chemical noise. • The noise level of a spectrum (or part of a spectrum) is typically

calculated as a signal-to-noise ratio (s/n ratio).• Noise and contaminations may cause so much disturbance that the

mass spectrometric analysis is unusable. Both sample handling and instrument settings may therefore be critical during an experiment.

Page 10: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 10

Constructing the peak list• Baseline correction

– The baseline is an offset of the intensities of masses, and should be subtracted from the determined intensities. It often depends on the m/z value such that it is highest at low m/z-values, and shows an exponential decay.

– The noise varies from spectrum to spectrum. Each spectrum must therefore be treated separately.

– Many algorithms for baseline correction have been developed. The simplest baseline correction finds the lowest point in the spectrum and drags this point down to zero. At the same time, the highest point in the spectrum (the base peak) is kept at 100 percent intensity. The result is that the intensities in the spectrum is "extended" along the vertical axis.

• Smoothing and noise reduction– A spectrum can be jagged, making it difficult to detect peaks from noise.– A smoothing is therefore performed on the spectrum. Again this can be done in

different ways. – The straightforward way is to use a sliding window, where a new value is calculated

for the point in the middle of the window, based on the values for the points in the window. The calculation is very often a weighted middle.

– The length of the window and the weights are determined from known forms of the noise, often taking the resolution into account. Often a Gaussian filter is used, based on the Gaussian function which is symmetric.

– Due to variation of the noise level along the mass range, they are calculated for small intervals (down to under one unit).

– More advanced smoothing functions are Fourier transformations and Wavelet transformations.

Page 11: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 11

Constructing the peak list• Peak detection is the process of distinguishing interesting peaks

from noise. • An aim is that every isotope in the spectrum should be represented

by exactly one data point, and several techniques are used. – One way is to first identify the apex of a peak. This is the point where the

intensity changes from increasing to decreasing.– Then the intensity at this apex is compared to the surrounding noise level.– Alternatively, the valleys on each side of the apex can be used to

determine start and end of the peak. – Then the area of the peak is calculated and examined if it is above the

limit for being a peak. – Other methods consider a peak as a continuous range of points with

intensity above the noise level. – Some also consider the shape of the peak to distinguish peptides from

contaminants and noise, as peptides tend to have different forms than noise have.

– The number of expected raw data points per peak (which can vary over the spectrum) is also used in some peak detection algorithms.

• Whatever peak detection algorithm used, high intensity peaks of single peptides are usually detected without failure.

• However, when the peaks have low intensities and there are complex patterns of overlapping peptides, errors are much more prone to occur.

Page 12: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 12

Constructing the peak list

Page 13: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 13

Constructing the peak list• Intensity normalization • Intensities are used to distinguish noise from real signals, in

scoring of comparisons, and in classifying spectra as good or bad.

• The intensity of peaks are highly variable from spectrum to spectrum, and for using them in a uniform way they should be normalized somehow.

• The normalization should be such that the normalized values reflect the probabilities that the ions are real ions from peptides (not noise).

• The traditional way has been to transform to relative intensities, relative to the total intensity (TIC) or to the maximum peak intensity

• This can then be used to remove noise, by considering peaks with a normalized intensity below a threshold as noise.

• To obtain higher robustness (across the spectra) rank based intensity normalization is proposed. This means that the most intense peak gets rank one, the next highest rank two, and so on.

Page 14: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 14

Constructing the peak list - calibration• The spectra obtained from MS instruments must be calibrated to

achieve the accuracy needed for database searches• Internal calibration is most commonly used, and is done by setting

the exact m/z values of known peaks in the experimental spectrum. Such peaks are obtained in one of two ways: – known standards may be added to our sample– the protease used for digesting the sample will produce known

autolytic peaks by self-digestion. – It is also possible to use external calibration by spotting known

standards close to our sample on the sample plate of the instrument, and the spectra from the sample spot and the standard spot are acquired consecutively.

Page 15: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 15

Constructing the peak list - calibration• Different forms of regression curves for calculating the calibrated values

are used, from a linear m*=am+b to higher order, for example m*=(am+b sqrt(m)+c)2 where m* is the new m/z-value.

Page 16: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 16

Constructing the peak list - calibration

• In the case of internal calibration, the new constants are applied to the same spectrum, correcting the m/z values of all peaks in the spectrum.

• In the case of external calibration, the constants from the standard spectrum are applied into the sample spectrum.

• External calibration will usually be less exact than internal calibration.

Page 17: INF380 - Proteomics-61 INF380 – Proteomics Chapter 6 – Mass Spectrometry – MALDI TOF The MALDI-TOF instruments are the simplest MS instruments suitable

INF380 - Proteomics-6 17

Peak list preprocessing• Monoisotoping and deisotoping• This process reduces a cluster of isotopic peaks to a single peak, with intensity

equal to the sum of the isotope intensities. • Monoisotoping reduces to the lowest detected peak in the cluster. This is

usually the procedure for MALDI experiments where delayed extraction and reflectron mode have been used.

• Deisotoping reduces to a centroided peak,with the m/z value determined from the intensities of the individual isotopes.

• The centroid m/z value corresponds to the value that would be obtained if average masses of the atoms in the peptide had been used for calculation of its mass.

• Removing spurious peaks• The fractional masses can be used to try to remove non-peptide masses, by

removing those whose fractional masses do not follow the pattern for peptides. • Removing other peaks• The background noise increases at low m/z ratios, and masses below a specified

limit (sometimes 500, but more typically 700 to 800) are therefore removed. • Also high masses (typically above 3,000 to 4,000) are removed, since the

sensitivity and the accuracy is usually lower at high massed. • Additionally, there are relatively few peptides in this high-mass range • It is also common to filter the masses against known contaminants, like keratin or

from the used protease.