Upload
kenneth-jay-ednaco
View
216
Download
1
Embed Size (px)
Citation preview
7/31/2019 Indications for Mechanical Ventilation (2)
1/25
1
Indications for mechanical
ventilation
Dr Aidah Abu Elsoud Alkaissi
An-Najah National University
Faculty of Nursing
7/31/2019 Indications for Mechanical Ventilation (2)
2/25
2
Indications for mechanical
ventilation If a patient has a continous decrease in oxygination (PaO2) , an
increase in arterial carbon dipxide level (PaCO2), and a persistent
acidosis (Deceased PH) , mechanical ventilation may be necessary
Conditions such as thoracic or abdominal surgery, drug overdose,
neuromuscular disorders, inhalation injury, COPD, multiple trauma,
shock multisysem failure and coma all may lead to respiratory failure
and the need for mechanical ventilation (chart 25-11) page 616 guide
the decision to plce a patient on a ventilator
A patient with apnea that is not readily reversible also is a candidate formechanical ventilation
7/31/2019 Indications for Mechanical Ventilation (2)
3/25
3
Classification of ventilators
Classified according to the manner in which theysupport ventilation
The rwo general categories are:
Negative pressure and positive pressure ventilators The most common category in use today is the
positive pressure ventilator
7/31/2019 Indications for Mechanical Ventilation (2)
4/25
4
Negative-pressure ventilators
Exert a negative pressure on the external chest
Decreasing the intrathoracic pressure during inspiration allows air toflow into the lung, filling its volume
Physiologically, this type of assissted ventilation is similar tospontaneous ventilation
It is used mainly in chronic respiratory failure associated withneuromascular conditions such as poliomylitis, muscular dystrophy, amyotrophic lateral sclerosis, and mysthenia gravis
It is inappropriate for th unstable or complex patient or the patientwhose condition requires frequent ventilatory changes
Negative pressure ventilators are simple to use and do not requireintubation of the airway consequently they are especially adaptablefor home use
7/31/2019 Indications for Mechanical Ventilation (2)
5/25
7/31/2019 Indications for Mechanical Ventilation (2)
6/25
6
Pressure-Cycled Ventilators
Ends inspiration when a preset pressure hs been reached
The ventilator cycles on, delivers a flow of air until it reachesa predetermined pressure, then cycles off
Its major limitation is that the volume of air or oxygen canvary as the patients airway resistance or complicationschanges
As a result the tidal volume delivered may be inconsistent,possibly compromising ventilation
Consequently in adults, pressure cycled ventilators areintended ony for short term use
The most common type is the IPPB machine
7/31/2019 Indications for Mechanical Ventilation (2)
7/25
7
Intermittent Positive Pressure Breathing
Is a form of assisted or controlled respiration produced by a ventilatory
apparatus in which compressed gas is delivered under positive
pressureinto a persons airway until a preset pressure is reached
Passive exhalation is allowed through a valve
The specific pressure and volume amounts, a long with the use of any
nebulizing medications, are prescribed individually for patients
The nurse should encourage patients to relax and reassure them that the
machine will automatically shut off airflow at the end of inspiration
The IPP machine may be powered by elecricity or gas and may be
connected with a mouthpiece, mask, or tracheostomy adapter
7/31/2019 Indications for Mechanical Ventilation (2)
8/25
7/31/2019 Indications for Mechanical Ventilation (2)
9/25
7/31/2019 Indications for Mechanical Ventilation (2)
10/25
10
Time-Cycled Ventilation
Terminate or control inspiration after a preset time
The volume of air the patient receives is regulated by the length of
inspiration and the flow rate of the air
Most ventilators have a rate control that determines the respiratory rate,But pure time cycling is rarely used for adults
These ventilators are used in newborns and infants
7/31/2019 Indications for Mechanical Ventilation (2)
11/25
11
Volume ctcled ventilators Volume-cycled ventilators are by far the most commonly
used possitive pressre ventilators today
With this type of ventilator, the volume of air to be
delivered with each inspirationis preset
Once this preset volume is deliveredto the patient, theventilator cycles off and exhalation occurs passively
From breath to breath, the volume of air delivered by the
ventilator is relatively constant, ensuring consistent,adequate breaths despite varying airway pressures
7/31/2019 Indications for Mechanical Ventilation (2)
12/25
7/31/2019 Indications for Mechanical Ventilation (2)
13/25
13
Noninvasive positive pressure ventilation
Patients are considered candidates for noninvasive ventilation if theyhave acute or chronic respiratory failure, acute pulmonary edema,COPD or chronic heart failurewith a sleep related breathing disorder
The device also may be used at home to improve tissue oxygenationand to rest the respiratory muscles while the patient sleeps at night
It is contraindicated for thise who have experienced respiratory arrest,serious dysrhythmias, cognitive impairment, or head or facial trauma
Noninvasive ventilation may also be used for patients at the end of lifeand those who do not want endotracheal intubation but may need short
or long term ventilatory support Bilevel positive airway pressure (biPAP) ventilation offers independenr
7/31/2019 Indications for Mechanical Ventilation (2)
14/25
14
Noninvasive positive pressure ventilation
It delivers two level of positive airway pressure provided viaa nasal or oral mask, nasal pillow or mouthpiece with a tightseal and a portable ventilator
Each inspiration can be intiated either by the patient or by the
machine if it is programmed with a backup rate The back up rate ensure the patient will receive a set number
of breaths per minute
Bi-PAP is most often used for patients who require
ventilatory assisstance at night such as those with severeCOPD or sleep apnea
Tolerance is variable; BiPAP is most successful with highlymotivated patient
7/31/2019 Indications for Mechanical Ventilation (2)
15/25
15
Adjusting the ventilator
The ventilator is adjusted so that the patient is comfortable and
breaths in sync with the machine
If the volume ventilator is adjusted appropriately, the patients
arterial blood gas values will be satisfactory and there will be
little or no cardiovascular compromise
Please read chart 25-12 page 618
Table 25-2 page 619
7/31/2019 Indications for Mechanical Ventilation (2)
16/25
16
Anesthesia machines Oxygen and nitrous oxide are usually supplied from the hospital
pipelines to the anesthesia machine at pressures of 50-55 psi The gas hoses (A narrow channel through which data flows under
pressure) going to the machine are color-coded, and the connecter arespecific for each gas so that nitrous oxide cannot be inadvertently(Done unintentionally, accidentally, often with no one accepting blame)
connected to the oxygen hose or vise versa
If central gas supply is not available or the hospital piping system fails,the machines are equipped with E-size cylinders of both gases, one ortwo cylinders of each gas are connected to yokes (To become joined
securely) on the machine In cylinder, oxygen is stored as a compressed gas
A full E-size cylinder contains about 660 L of oxygen at 2000 psi
As the oxygen is used the pressure falls in direct proportion to theremaining volume, because the E-size cylinder is used to provideoxygen while transporting patient
7/31/2019 Indications for Mechanical Ventilation (2)
17/25
7/31/2019 Indications for Mechanical Ventilation (2)
18/25
18
Anesthesia machines Knowing how much oxygen is left in a partially used tank is
important Thus 1000 psi would indicate 330 L remaining and 500 psi would
indicate 165 L remaining or sufficient oxygen for 5 L/min flow
for more than 25 min
When the pressure has dropped to about 250 psi, the cylindershould not be used because it no longer has an adequate reserve
Nitrous oxide is stored as a liquid in cylinder and the pressure
above the liquid is 750 psi
A full, E-size cylinder contains about 1600 L of nitrous oxide As the nitrous oxide is used the pressure above the liquid remains
constant
7/31/2019 Indications for Mechanical Ventilation (2)
19/25
19
Anesthesia machines Only when the liquid has been completely vaporized does the
pressure begin to fall
Therefore the nitrous oxide can be almost gone but still show thesame pressure
In contrast to oxygen, the amount remaining in the tank can not bereadily determined
The gases in the cylinders flow through regulators thatreduce thepressure to about 50 psi
The hoses from the hospital gas sources are connected to themachine at the outlet or these regulators
In most machines of recent vintage (A period of origin) a pressureinterlock device shuts off the nitrous oxide flow if oxygen pressureis not present
The gases then flow through individual flowmeters (orrotameters)on the front of the machine so that the gas flow and the
ratio of oxygen to nitrous oxide can be selected by theanesthesiolo ist
7/31/2019 Indications for Mechanical Ventilation (2)
20/25
20
Anesthesia machines From the top of the flowmeters, the gases are mixed and then flow
through a vaporizer in which the inhalational anesthetic of choice is
vaporized and added to the oxygen nitrous oxide mixture
The total gas flow is then delivered from the machine to the patient
With a flow through vaporizer, by definition, all of the fresh gas goingto the patient from the anesthesia machine flows through the vaporizer
The control dials are usually located on top of these vaporizers and are
calibrated in percentages
Most recently manufactured vaporizers are flow and temperaturecompensated, meaning that they are reasonably accurate at all flows
and temperatures used clinically
The filling ports on the vaporizers are usually key indexed so that only
the appropriate vilatile agent can be used
7/31/2019 Indications for Mechanical Ventilation (2)
21/25
21
Anesthesia machines
The copper kettle is one type of bypass vaporizer that may still be
found on many older machines A low flow of oxygen (usually less than 1 L/min) goes through a
separate flowmeter and then through the bypass vaporizer, where thisoxygen is totally saturated with the anesthetic vapor
The saturated oxygen is then combined with the oxygen-nitrous oxide
mixture from the other flowmeter and the total mixture flows fromthe machine to the patient
To calculate the concentration of anesthetic going to the patient, theanesthesiologist must know the barometric pressure, the vapor
pressure of the anethetic agent being used, the oxygen flow throughthe bypass vaporizer, the temperature of the anesthetic liquid in thevaporizer and the combined flow of the oxygen nitrous oxidemixture
The major advantages of bypass vaporizers is that any volatile
anesthetic agent can be used in them
7/31/2019 Indications for Mechanical Ventilation (2)
22/25
22
Anesthesia machines Another important feature of the anesthesia machine is the oxygen flush button
With all new machines, pushing the oxygen flush button allows 100% oxygenfrom the 50 psi line to flow directly to the fresh gas outlet on the machine and thusto the patient
This oxygen flow completely bypasses the flow meters and vaporizers
In most hospitals a semiclosed circle system is used to deliver the fresh gas flow
(including anesthetic gases) to patients The circle system is composed of a container filled with a CO2-absorbing material
(such as soda lime or baralime), two unindirectional valves, an adjustable pressurelimiting valve (APL), a reservoir bag, an inlet connection for fresh gas flow andtwo connections to the patient through corrgated (To shape into folds) breathingor anesthesia hoses
As the patient inspires, gases are drawn through the CO2 absorber and from thefresh gas supply through the inspiratory limb of the corrugated hoses
As the patient exhales, the one-way valve on the inspiratory limb preventsbackflow, and the exhale gases flow through the expiratory limb and through theexpiratory one-way valve
7/31/2019 Indications for Mechanical Ventilation (2)
23/25
23
Anesthesia machines The expiratory limb and valve are easily identified by the condensation
of water vapor along this portion of the circuit The 3L reservoir bag absorbs the peak flow of expired gases and allows
the anesthesiologist to force gas through the CO2 absorber along theinspiratory limb of the circuit and thereby ventilate the patient
The expired gases flow through the CO2 absorber, where the carbon
dioxide is removed from them
Any excess gas is vented through the APL valve, which is yasulltmounted just ahead of the CO2 absorber
The FiO2 sensor is usually mounted in the inspiratory limb just after the
one-way valve It measures the fraction of inspired O2 and can set to alarm if a low
concentration is detected
A low pressure sensor is usually mounted in the expiratory limb near theother one way valveto detect a ventilator malfunction or a disconnectionin the circuit
7/31/2019 Indications for Mechanical Ventilation (2)
24/25
7/31/2019 Indications for Mechanical Ventilation (2)
25/25