9
Implantation of transvenous nonthoracotomy cardioverter-defibrillator systems in patients with permanent endocardial pacemakers Among 177 patients in whom a nonthoracotomy approach was initially used to implant a cardioverter-defibrillator system, 11 (6%) patients also received a separately implanted permanent pacemaker. The main problem encountered in these patients were previously implanted unipolar pacemakers (n = 3) and ventricular pacing leads positioned at the right ventricular apex, the latter interfering with optimal placement of the tripolar implantable cardioverter-defibrillator (ICD) lead (n = 9). The approaches used to solve these problems were individualized and included placement of the ICD sensing lead at the right ventricular outflow tract (n = 3), initial placement (n = 1) or subsequent repositioning (n = 2) of the right ventricular pacing lead at the outflow tract, upgrade from unipolar to bipolar systems (n = 2), reprogramming from the DDD to AAI mode (n = 2), inactivation of the pacemaker (n = 1), and simultaneous placement of a single-chamber atrial pacemaker with the ICD lead (n = 2). These revisions fulfilled the pacing needs in each patient and prevented unfavorable sensing interaction between the two systems. {AM HEART J 1995;129:45-53.) Ross Brooks, MD, Hasan Garan, MD, Brian A. McGovern, MD, and Jeremy N. Ruskin, MD Boston, Mass. An estimated 15 % to 20 % of patients with implant- able cardioverter-defibrillator (ICD) systems require concomitant single- or dual-chamber pacing. 1-3 Be- cause these systems are separately implanted, po- tential interaction of sensing function may result. Most important, the rate-sensing lead of the ICD system may detect pacemaker stimuli leading to multiple counting and spurious discharges or detec- tion inhibition may occur, potentially resulting in suppression of life-saving ICD therapy during ven- tricular fibrillation. 4 Newer ICD systems with inte- grated sensing and VVI pacing functions have elim- inated interaction between separately implanted sys- tems. However, these systems do not permit rate responsive pacing, which may be desirable in some patients and may drain the battery prematurely in others who are pacemaker dependent. Moreover, most patients with ICD systems requiring pacemak- ers need atrial or dual chamber pacing and are not well served by VVI pacing alone. 1-~ With conventional thoracotomy-implanted ICD From the Cardiac Arrhythmia Service, Massachusetts General Hospital, and Harvard Medical School. Received for publication Feb. 3, 1994; accepted March 21, 1994. Reprmt requests: Ross Brooks, MD, Cardiac Unit, Massachusetts General Hospital, Boston, MA 02114. Copyright ® 1995 by Mosby-Year Book, Inc. 0002-8703/95/$3.00 + 0 4/1/59119 systems, interaction between ICD and pacemaker functions can be avoided by implanting epicardial ICD rate-sensing leads remote from endocardial- placed pacemaker leads. Patients with preexisting unipolar pacemaker leads require revision to bipolar pacing systems to prevent sensing of the large unipo- lar pacer stimuli. 4 With newer nonthoracotomy ICD systems, wide separation of the leads can no longer be achieved because the transvenous ICD and pace- maker leads are both placed in an endocardial posi- tion in the right ventricle. In addition to potential sensing interaction, other technical considerations related to lead insertions and placement may also limit transvenous ICD implantation in patients with preexisting permanent pacemakers. No published data have been reported on patients with combined transvenous ICD and permanent endocardial pace- maker systems. Data on the management of patients with combined systems is relevant because >80 % of all ICD systems currently implanted are nonthorac- otomy and require placement of one or more trans- venous lead. 5 We describe our clinical experience managing 11 patients with combined transvenous ICD and endocardial-placed pacemaker systems. METHODS Patients. Among 177 consecutive patients undergoing transvenous ICD placement at our institution by means of 45

Implantation of transvenous nonthoracotomy cardioverter-defibrillator systems in patients with permanent endocardial pacemakers

Embed Size (px)

Citation preview

Implantation of transvenous nonthoracotomy cardioverter-defibrillator systems in patients with permanent endocardial pacemakers

Among 177 patients in whom a nonthoracotomy approach was initially used to implant a cardioverter-defibrillator system, 11 (6%) patients also received a separately implanted permanent pacemaker. The main problem encountered in these patients were previously implanted unipolar pacemakers (n = 3) and ventricular pacing leads positioned at the right ventricular apex, the latter interfering with optimal placement of the tripolar implantable cardioverter-defibrillator (ICD) lead (n = 9). The approaches used to solve these problems were individualized and included placement of the ICD sensing lead at the right ventricular outflow tract (n = 3), initial placement (n = 1) or subsequent repositioning (n = 2) of the right ventricular pacing lead at the outflow tract, upgrade from unipolar to bipolar systems (n = 2), reprogramming from the DDD to AAI mode (n = 2), inactivation of the pacemaker (n = 1), and simultaneous placement of a single-chamber atrial pacemaker with the ICD lead (n = 2). These revisions fulfilled the pacing needs in each patient and prevented unfavorable sensing interaction between the two systems. {AM HEART J 1995;129:45-53.)

Ross Brooks, MD, Hasan Garan, MD, Brian A. McGovern, MD,

and Jeremy N. Ruskin, MD Boston, Mass.

An estimated 15 % to 20 % of patients with implant- able cardioverter-defibrillator (ICD) systems require concomitant single- or dual-chamber pacing. 1-3 Be- cause these systems are separately implanted, po- tential interaction of sensing function may result. Most important, the rate-sensing lead of the ICD system may detect pacemaker stimuli leading to multiple counting and spurious discharges or detec- tion inhibition may occur, potentially resulting in suppression of life-saving ICD therapy during ven- tricular fibrillation. 4 Newer ICD systems with inte- grated sensing and VVI pacing functions have elim- inated interaction between separately implanted sys- tems. However, these systems do not permit rate responsive pacing, which may be desirable in some patients and may drain the battery prematurely in others who are pacemaker dependent. Moreover, most patients with ICD systems requiring pacemak- ers need atrial or dual chamber pacing and are not well served by VVI pacing alone. 1-~

With conventional thoracotomy-implanted ICD

From the Cardiac Arrhythmia Service, Massachusetts General Hospital, and Harvard Medical School.

Received for publication Feb. 3, 1994; accepted March 21, 1994.

Reprmt requests: Ross Brooks, MD, Cardiac Unit, Massachusetts General Hospital, Boston, MA 02114.

Copyright ® 1995 by Mosby-Year Book, Inc. 0002-8703/95/$3.00 + 0 4 / 1 / 5 9 1 1 9

systems, interaction between ICD and pacemaker functions can be avoided by implanting epicardial ICD rate-sensing leads remote from endocardial- placed pacemaker leads. Patients with preexisting unipolar pacemaker leads require revision to bipolar pacing systems to prevent sensing of the large unipo- lar pacer stimuli. 4 With newer nonthoracotomy ICD systems, wide separation of the leads can no longer be achieved because the transvenous ICD and pace- maker leads are both placed in an endocardial posi- tion in the right ventricle. In addition to potential sensing interaction, other technical considerations related to lead insertions and placement may also limit transvenous ICD implantation in patients with preexisting permanent pacemakers. No published data have been reported on patients with combined transvenous ICD and permanent endocardial pace- maker systems. Data on the management of patients with combined systems is relevant because >80 % of all ICD systems currently implanted are nonthorac- otomy and require placement of one or more trans- venous lead. 5 We describe our clinical experience managing 11 patients with combined transvenous ICD and endocardial-placed pacemaker systems.

METHODS Patients. Among 177 consecutive patients undergoing

transvenous ICD placement at our institution by means of

45

January 1995 46 Brooks et al. American Heart Journal

Table I. Clinical features of implantable cardioverter-defibri l lator and pacemaker systems

Patient

1 2 3 4 5

Age/Gender 74/M 59/M 58/M 78/F 75/M Heart disease CAD CAD CAD DCM CAD LVEF 0.55 0.30 0.48 0.43 0.30 Arrhythmia

Clinical VT VT VF VT VF Induced VT VT VF VT NI

Indication for SSS CHB SSS SSS SSS pacemaker

Pacemaker CPI Vista 940 CPI Vista 940 CPI Triumph CPI Vista T PS II Mode DDD DDD AAIR DDD -~ AAI DDDR Venous access LSCV LSCV LSCV RSCV LSCV Lead B B B U -~ B U Lead position Apex RVOT Atrium Apex Apex Pacemaker Yes Yes Yes Yes Yes

dependent Transvenous lead Transvene Endotak Endotak Endotak Endotak Venous access RSCV RSCV LSCV LSCV RSCV Lead position RVOT Apex Apex Apex Apex ICD model Medtronic PCD CPI 1600 CPI 1600 CPI PRX CPI P2 Order of implant P --* ICD P --~ ICD P -~ ICD P -~ ICD P --* ICD Pacemaker revision None P implanted before P implanted with New atrial lead P reprogrammed

ICD ICD and P DDD --~ AAI implanted; old P inactivated

CAD, Coronary artery disease; DCM, dilated cardiomyopathy; LVEF, left ventricular ejection fraction; VT, ventricular tachycardia; VF, ventricular fibril- lation; NI, noninducible: SSS, sick sinus syndrome; CHB, complete heart block; CPI, Cardiac Pacemakers, Inc.; PS, Pacesetter Synchrony; PP, Pacesetter Paragon; LSCV, left subclavian vein; RSCV, right subclavian vein; RCV, right cephalic vein; B, bipolar; U, unipolar; RVOT, right ventricular outflow tract; RVIT, right ventricu]ar inflow tract; P, pacemaker.

an initial nonthoracotomy approach, the 11 pat ients (10 men, 1 woman) who required concomitant permanent pacemakers are the subject of this report. The clinical fea- tures of these pat ients are shown in Table I. Ten pat ients had endocardial pacemaker systems implanted a mean of 3.2 years (0.05 to 12 years) before ICD insertion. Seven pa- t ients received dual-chamber pacemakers (5 bipolar and 2 unipolar systems), one pat ient received a VVI pacemaker, aud 1 pa t ien t each received a unipolar and bipolar VVI pacemaker. In 8 pat ients with previously implanted pace- makers, the ventricular pacing lead was posit ioned at or near the right ventricular apex, interfering with the ideal posit ion of the ICD transvenous lead.

ICD lead insertion. Two investigational nonthoracot- omy ICD systems were used. Three pat ients received a Transvene system (Medtronic, Inc., Minneapolis, Minn.) and 8 pat ients received a CPI Endotak system (Cardiac Pacemakers, Inc., St. Paul, Minn.). The 110 cm, 10.5F tri- polar Transvene lead (model 6966 [n = 3] and model 6936 [n = 1]) consists of a screw-in tip, an adjacent sensing ring, and an adjacent 50 mm-long coil electrode. A separate 8F unipolar coil electrode is also inserted in conjunction with the t r ipolar lead to serve as the second electrode. The 100 cm, 12F tr ipolar Endotak lead (model 0062 [n = 1], model 0064 In = 3], and model 0074 [n = 3]) consists of a t ined tip,

an adjacent distal coil electrode, and a proximal coil elec- trode. In 3 pat ients with chronically implanted ventricular pacing leads at the apex of the right ventricle, the active fixation t r ipolar Medtronic lead was posit ioned in the re- gion of the right ventricular outflow tract (Fig. 1). In all 7 pat ients receiving an Endotak system, the t ined t r ipolar Endotak lead was placed at the apex of the right ventricle (Figs. 2 to 4). The transvenous ICD leads were implanted in the pacemaker laboratory before intraoperat ive ICD implanta t ion in 6 pat ients and at the t ime of ICD implan- ta t ion in the other 5 patients .

ICD testing. Defibril lation threshold testing and ICD implanta t ion were performed intraoperat ively under gen- eral anesthesia. Defibril lation threshold test ing was per- formed according to the manufacturer ' s protocol as out- l ined previously. 5 During defibril lation threshold testing all pacemakers were reprogrammed to the asynchronous (DOO or VOO) pacing mode at 80 beats /min the maximal programmable output and pulse width. Ventricular fibril- lat ion was induced by brief delivery of 60 Hz al ternat ing current through the t r ipolar lead. In pat ients receiving an Endotak system (n = 7), the external cardioverting device (model 2806) was set in the automatic mode during defibril lation threshold testing to simulate the actual detect ion algori thm of the ICD. In one pat ient receiving a

Volume 129, Number 1

American Heart Journal ~roo~,3 et al. 47

6 7 8 9 10 I1

66/M 55/M 82/M 43/M 81/M 77/M DCM CAD CAD DCM CAD CAD 0.27 0.29 0.26 0.41 0.29 0.22

VF VT VT VT VT VT VT VT VT VT VT VF CHB CHB SSS SSS SSS SSS

PS PS II PS II Medtronic 8420 PP II PS VVI -~ DDD DDD -~ ODO DDD --~ AAI VVI DDD DDD -- AAI LSCV LSCV RSCV RSCV RSCV RSCV U - ~ B U B B B B Apex Apex Apex RVIT RVIT Apex Yes No Yes No Yes No

Transvene Transvene Endotak Transvene Endotak Endotak RSCV RCV LSCV LSCV LSCV LSCV RVOT RVOT Apex Apex Apex Apex Medtronic PCD Ventritex Cadence CPI PRX Medtronic Jewel CPI PRX CPI P2 P -~ ICD P -* ICD P ~ ICD P - - ICD P ~ ICD P ~ ICD

P revised after P inactivated, Ventricular lead P rate set at <ICD P lead repositioned P reprogrammed ICD: back-up VVI repositioned to rate to RVOT to AAI mode VVI --~ DDD pacing from ICD RVOT

Medtronic jewel ICD, the external cardioverting device (model 9760) also allowed ar rhythmia detect ion during de- fibril lation testing in an automatic mode. Two pat ients re- ceiving Medtronic PCD systems underwent test ing with an external cardioverting device (model 9715B) in the manual mode. One final pat ient underwent testing in the manual mode with a Ventr i tex (Sunnyvale, Calif.) external cardio- verting device (model HVS-02). In these three instances, defibril lat ion was a t t empted after 10 seconds of elapsed arrhythmia. The rate-sensing channel of the ICD and two surface electrocardiograms were recorded on paper (25 mm/sec) before, during, and after all a r rhythmia induc- tions. The implanted ICD device was tested with the pace- maker programmed in the asyncbronous mode at the aforementioned settings.

Pacemaker system revisions. Revision of the pace- maker system was performed before ICD implanta t ion in 2 patients , in t raoperat ively in 7 patients, and after ICD implanta t ion in 1 pa t ien t (Table I). Pa t ien t 1 did not require revision of his pacemaker system. In pat ient 2 it was determined tha t both a pacemaker and ICD were neces- sary, and the pacemaker was implanted 10 days before the ICD. In this pat ient the ventr icular pacing lead was posi- t ioned at the right ventr icular outflow tract to allow sub- sequent placement of an Endotak lead at the right ventric- ular apex (Fig. 2). In pa t ien t 3, a bipolar atr ial lead and single-chamber pacemaker were placed at the t ime of insertion of the Endotak lead (Fig. 3). In pat ient 4, who had a preexist ing unipolar dual-chamber pacing system, the

pacemaker was init ially reprogrammed to the AAI mode during intraoperat ive testing. However, despite repro- gramming to this mode, large pacemaker st imulus art ifacts were present on the ICD rate channel, resulting in multi- ple counting, part icular ly in the immediate postshock pe- riod. At intraoperat ive ICD implantat ion, a new bipolar lead and a single-chamber atr ial pacemaker were im- planted with an ICD containing VVI pacing functions (Fig. 4). In pat ients 5 and 11, who had bipolar dual-chamber pacemakers, the Endotak lead was placed in close proxim- ity to the pacemaker lead, which was also posit ioned at the right ventricular apex. These patients , who received ICDs capable of VVt pacing, bad their DDD pacemakers reprogrammed to the AAI mode. Pa t ien t 6 underwent re- vision of his pacemaker system after ICD implantat ion. The original system consisted of a unipotar VVI pacemaker tha t was upgraded to a dual-chamber system by placing new bipolar atrial and ventricular leads. Pa t ien t 7, who had a unipolar dual-chamber pacemaker placed for t ransient peri infarction heart block, had his pacemaker inactivated in lieu of replacing the pacemaker leads. Pa t ien t 7 also re- ceived an ICD with VVI pacing functions. Pa t ien t 8, who had a bipolar dual chamber pacemaker placed 1 month earlier, underwent repositioning of the ventricular pacing lead at the t ime of ICD implantat ion to facilitate place- ment of an Endotak lead at the right ventricular apex. In pat ient 9 with a bipolar ventricular pacemaker lead im- planted at the mid right ventricle, a Medtronic t r ipolar lead was posit ioned at the right ventricular apex (Fig. 5). The

January 1995

48 Brooks et al. American Heart Journal

Fig. 1. Posterior-anterior chest x-ray film shows previ- ously implanted unipolar pacemaker in patient 8, who had transient postinfarction heart block. Pacemaker leads were positioned in atrium and at right ventricular apex. Tripo- lar Medtronic lead was inserted with tip placed at right ventricular outflow tract. Second coil electrode is posi- tioned in right atrium.

rate of the VVI pacemaker was set below ICD VVI pacing rate. Patient 10, who also had a ventricular pacemaker lead positioned at the right ventricular inflow tract, had an En- dotak lead placed at the apex of the right ventricle. How- ever, because of detection of pacemaker stimuli on the ICD rate-sensing lead, this patient's pacemaker lead was repo- sitioned to the right ventricular outflow tract.

RESULTS ICD implantation. Satisfactory R-wave amplitudes

(12 + 3 mV), pacing thresholds (0.64 _+ 0.3V), and lead resistances (445 _+ 41 ohms) were obtained in all patients. In all cases, local pacemaker electrograms were absent or only minimally detectable on the ICD rate-sensing lead, and in no case did the pacemaker stimulus exceed the peak-to-peak R-wave amplitude of the ICD by >20 %. With the final pacemaker revi- sion, no clinically significant sensing interaction was observed during defibrillation threshold or ICD test- ing in any patient. All patients received an ICD sys- tem during the same intraoperative session. Eight (73 % ) of 11 patients received a nonthoracotomy sys- tem with a mean defibrillation threshold established

at 15 + 5 J. Three patients required thoracotomy because of inability to achieve satisfactory defibril- lation thresholds <18 J by means of a nonthoracot- omy approach despite testing multiple configura- tions. In these patients a hybrid configuration incor- porating a single left ventricular extrapericardial patch and two transvenous coil electrodes were used. The ICD pulse generators implanted are shown in Table I. None of the pacemakers used (Table I) were reset to start-up modes by ICD shock delivery. In two patients there was transient loss of pacemaker cap- ture after ICD shock therapy, which reaolved after one or two noncaptured beats.

Follow-up. All patients underwent repeat ICD test- ing before discharge. All ICD devices showed normal sensing, antitachycardia pacing, and cardioversion/ defibrillation functions. All patients were discharged alive from hospital and have been monitored for a mean of 8.8 _+ 9 (range 3 to 26) months. The pace- maker and ICD systems have been observed to func- tion normally during follow-up visits, including rou- tine 8-week ventricular fibrillation inductions, as part of the investigational protocols. Four (36 %) pa- tients received appropriate ICD shock or antitachy- cardia pacing therapy during the follow-up period. No pacemaker resets were noted at the time of ther- apy delivery in these patients. No additional modifi- cation was necessary to any patient's system after discharge.

DISCUSSION

The 11 patients who required combined pace- maker and ICD systems were successfully managed. The main problems were previously implanted uni- polar pacemakers (n = 3) and ventricular pacing leads positioned at the right ventricular apex (n = 8), the latter problem interfering with optimal place- ment of the tripolar ICD lead. The approaches used to solve these problems were individualized to each patient's set of circumstances and included place- ment of the ICD transvenous lead at the right ventricular outflow tract (n = 3), initial placement (n = 1) or subsequent repositioning (n = 2)' of the right ventricular pacing lead at the outflow ~ract, up- grade from unipolar to bipolar pacemaker systems (n = 2), reprogramming from DDD to AAI pacing modes (n = 2), inactivation of the pacemaker (n = 1), reduction of the VVI pacemaker rate to below the ICD pacing rate (n = 1); and simultaneous placement of a single-chamber atrial pacemaker with the ICD lead (n = 2). These modifications fulfilled the pacing requirements in all patients and prevented adverse sensing interaction between the two systems. Con- siderations in managing patients who required com-

Volume 129, Number 1 American Heart Journal Brooks et al. 49

- ~ ' ' : ' ~ ' I , , .-,',~," " ~ = ,~ ~ .

Fig. 2. Patient 2 had complete heart block and under- went implantation of bipolar dual-chamber pacing sys- tem 14 days before ICD implantation. In anticipation of ICD implantation, ventricular pacemaker lead was placed at right ventricular outflow tract to allow position- ing of tripolar Endotak lead at apex of right ventricle. Re- lation between ventricular pacing and ICD leads in this patient is considered optimal when combined systems are required.

bined pacemakers and nonthoracotomy transvenous ICD systems are summarized in Table II.

Determinants of cross-talk between pacemaker and ICD s y s t e m s . The potential incompatability of sepa- rately implanted pacemaker and transvenous ICD systems results from the detection of pacemaker stimuli on the ICD rate-sensing channel. Sensing of pacemaker stimuli on the ICD rate lead depends to a large extent on the far-field effect of the pacemaker stimuli, that is, the size of the pacemaker spike rel- ative to the local electrogram on the ICD sensing channel. 6 In general, pacemaker leads that are in close proximity to the ICD lead have a large pace- sense dipole, and are in parallel to them will result in a greater far-field effect and larger signals on the ICD lead. Unipolar pacemaker artifacts may lead to sub- stantial far-field effect and in most cases to very large signals, even if the two leads are placed far apart. The far-field effect of pacemaker stimuli may be more of an issue with the Endotak relative to the Transvene system because of inherent sensing differences be-

' .~ ; :;. =.=, J

• - ~ " i ' ' , , , ' " ~ i " ,

: ~,~-~; : ~ : :;" -.. . ' . . ' .

Fig. 3. Patient 3 had symptomatic sinus bradycardia and underwent implantation of single-chamber atrial pace- maker at time of insertion of tripolar Endotak lead.

tween these two systems. Bipolar sensing in the Medtronic system occurs between the tip of the elec- trode and an adjacent sensing ring whereas, with the Endotak lead, bipolar sensing occurs between the tip and the distal coil electrode. With the Medtronic system, true bipolar sensing takes place between two closely spaced electrodes, resulting in a smaller pace-sense dipole. In contrast, with the CPI system a relatively greater interelectrode distance is present and results in a larger pace-sense dipole. Conse- quently, the potential for detecting far-field signals may be somewhat greater with the Endotak lead and relatively less with the Transvene lead. Owing to these differences, a pacemaker lead positioned at the right ventricular inflow tract may be compat- ible with a Transvene lead placed at the right ven- tricular apex despite their parallel positions, as dem- onstrated in one of our patients (Figs. 5 and 6). In contrast, Endotak and pacemaker leads should be placed farther apart, ideally at right angles to one another, to minimize far-field effect. Because the tined Endotak lead must be positioned at the right ventricular apex, the pacemaker lead in effect should optimally be placed at the right ventricular outflow tract (Fig. 2).

January 1995

50 Brooks et al. American Heart Journal

' ' . , :

, ' e l

irish: ,.r

, i.~,~,

1 . . . .

Fig. 4. Patient 4 had sinus node dysfunction and previously implanted unipolar dual-chamber pacemaker underwent placement of tripolar Endotak lead at right ventricular apex. Despite reprogramming pacemaker to AAI mode, large pacemaker stimuli were present on ICD rate-sensing channel. At time of ICD implan- tation new bipolar atrial lead placed in conjunction with single-chamber pacemaker resolved sensing in- teraction. Old pacemaker system was not removed.

Fig. 5. Patient 9 with previously implanted bipolar VVI pacemaker lead in mid right ventricle underwent place- ment of Medtronic tripolar lead at right ventricular apex. Despite close proximity of two ventricular leads, ventricu- lar pacemaker spikes were not detected on ICD rate-sens- ing channel (see Fig. 6).

Table II, Considerations in patients receiving combined transvenous ICD and endocardial pacemaker systems

1. In a patient with a previously implanted pacemaker, does the patient still need the pacemaker?

2. Is the current mode of pacing necessary (e.g., DDD) or adequate (e.g., VVI)?

3. Is the lead system unipolar or bipolar? 4. With the present pacemaker generator, is there potential for

reset to unipolar pacing during ICD shock therapy? 5. Could the patient be managed with VVI pacing from the ICD

system alone? 6. In a patient requiring separately implanted DDD or VVI

pacemaker systems, the ventricular pacing lead (rather than the ICD transvenous lead) should be preferentially implanted in the region of the RVOT.

7. In a patient with a DDD pacemaker, could the patient be managed with AAI pacing along with back-up VVI pacing from the ICD?

8. Be sure to test the pacemaker in the asynchronous mode dur- ing defibrillation and ICD testing to ensure appropriate ICD sensing.

Reprogramming of pacemakers to eliminate interac- tion. Consistent with other reported series, a major- ity of our patients (n = 9) required atrial or dual- chamber pacing, most commonly for sinus node dys- funct ion (Table I). Because most pat ients with sinus node dysfunct ion can be satisfactorily managed by single-chamber atrial pacing, 7, s one option tha t can be used to prevent sensing interaction in pat ients

Volume 129, Number 1 American Heart Journal Brooks et al. 51

Oct. 6, 1993 . . . 25mm/sec

,~ SINUS RHYTHM WITH VO0 PACING

ECG 2 --.~ I j - ~ r . ~ , . . - -

M A R K E R CHANNEL

[ I I 0.44 0:48 0 :50

Arrhythmia Induction

VENTRICULAR - - FIBRILLATION

I • ' ' 1 ' " I 0:46 0:52 0:54

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

ICD Shock I

- - ~ - VENTRICULAR FIBRILLATION ~ f - - SINUS RHYTHM WITH VO0 PACING

I

I

ECG 1

ECG 2

M A R K E R CHANNEL

0:56 ~ O: 58 t :00 |

Detection Satisfied

v - - 7 I ̧ I

1'02 t : 04 1 0 6

Fig. 6. Two surface ECG leads and ICD marker channel during testing in patient 9, who had VVI pace- maker placed previously in conjunction with Medtronic transvene ICD system. Pacemaker is reprogrammed to VOO mode during testing. Despite presence of bipolar pacemaker spike artifact on ECG leads, no pace- maker spike artifact is detected on ICD marker channel and sensing before, during, and after induced yen- tricular fibrillation is normal.

with previously implanted bipolar dual-chamber pacemaker systems is to reprogram the pacemaker to the AAI mode. The atrial bipolar stimulus is remote from the ventricular lead and generally so small on the ICD rate-sensing channel that it does not inter- fete with arrhythmia detection. Four patients in the present series were managed by AAI pacing alone. Two patients with preexisting bipolar DDD pace- makers were reprogrammed to the AAI mode. A third patient with a unipolar DDD system had her pace- maker inactivated. In this patient a new bipolar atrial lead was placed in conjunction with a single-chamber atrial pacemaker. In each patient reprogramming from DDD to AAI pacing proved satisfactory and, of practical importance, obviated the need to revise the

chronically implanted ventricular pacing lead, which was left in its original position. Moreover, it permit- ted placement of an Endotak lead at the right ventricular apex in all three patients. A fourth patient with no previously implanted pacemaker re- ceived a single-chamber atrial pacemaker at the time of implantation of an Endotak lead. All three pa- tients with previously implanted DDD pacemakers who received AAI pacing also received an ICD with VVI pacing functions (Table I). Although AAI pacing satisfied the needs of these patients, we preferred to have back-up VVI pacing available because of the potential risk of heart block.

Frequency of combined ICD and pacemaker systems. The frequency of combined pacemaker and ICD sys-

January 1995

52 Brooks et al. American Heart Journal

tems in the present series is 6 %, well below the 15 % to 20 % reported previously in series of patients with thoracotomy-implanted systems. 1-3 The lower num- ber in the present series may in part be accounted for by the use of ICD pulse generators that incorporated VVI pacing functions in 9 of 11 patients. The pace- maker function of ICDs subserves a back-up role and is generally satisfactory in patients with postshock bradyarrhythmias or long pauses during atrial fibril- lation or in patients at potential risk for heart block. In the past, patients who required back~up VVI pac- ing for these indications required separate pacemak- ers in conjunction with their ICD systems; this prac- tice may account for the higher percentage of pa- tients with pacemakers in earlier series of ICD recipients.

Frequency of ICD-pacemaker interaction. In earlier series the reported frequency of significant pacemak- er-ICD interaction was <1% .1 The low rate of inter- action is the result of widespread recognition of po- tential problems and meticulous intraoperative test- ing to ensure that sensing interaction is absent. The most significant problems encountered were double rate counting and arrhythmia nondetection. These problems were first recognized in patients with uni- polar pacing systems because of the large pacemaker stimulus artifact, but have also been described in pa- tients with bipolar systems in which the pacemaker stimulus is also large. 9 Another potential problem in patients with pacemakers is transient loss of capture and "reset" of the pacemaker to the start-up mode (usually unipolar) after ICD shock delivery. The po- tential for the reset problem exists in patients with older pacemakers, particularly those capable of uni- polar and bipolar pacing, but is not a problem in those with a dedicated bipolar system. This problem did not occur in the patients in the present series with the pacemakers used (Table I), and therefore the po- tential risk of inadvertent unipolar pacing was con- sidered extremely small.

ICDs with pacemaker functions. The latest genera- tion of ICD pulse generators also contain VVI pacing functions. However, in VVI pacemaker-dependent patients continuous pacing will result in substantial battery drain, even if the pacemaker function of the ICD is programmed at nominal settings. With all of the ICDs implanted in the present series, continuous pacing would potentially result in premature battery depletion and, in some cases, shorten the life of the battery by _>50% or more. Because of the high cost of the ICD, it is not cost-effective to use it as the pri- mary pacing modality in pacemaker-dependent pa- tients. In these patients separate pacemaker and ICD systems are necessary. Although 9 patients received

an ICD with VVI pacing functions, 8 of the 11 patients in the present series were pacemaker de- pendent and required separately implanted pace- makers. Technologic advances in future generations of ICD pulse generators may require less energy to perform pacing functions and may not substantially reduce battery life in pacemaker-dependent patients. More advanced ICD systems may contain dual- chamber pacing capabilities, rendering separately implanted pacemaker-ICD systems unnecessary in most patients. Until these systems become available, separately implanted pacemaker systems will be re- quired in appropriately selected patients with trans- venous ICDs. The present study suggests that im- plantation of separate endocardial pacing and ICD systems is feasible, that potential technical difficul- ties can be overcome, and that potentially adverse interactions can be avoided with careful planning and meticulous testing in these patients. Physicians implanting pacemakers into potential ICD recipients are encouraged to use dedicated bipolar systems and to ideally implant these leads at the right ventricular outflow tract.

Limitations. One option we did not use to facilitate optimal placement of the transvenous ICD leads was extraction of chronically implanted pacing leads. We did not believe that this procedure was generally worth the risk or effort because in most cases the lead had been implanted for many years. Of potential concern to defibrillation threshold testing with nonthoracotomy configurations is the position of the distal coil electrode at the right ventricular outflow tract rather than its customary location at the apex. In the present series two patients with leads posi- tioned at the right ventricular outflow tract required thoracotomy, although a third patient did not. Al- though the eccentric position of the lead may have contributed to a failure to achieve adequate defibril- lation thresholds, it was not believed to be the primary factor responsible for unsuccessful nontho- racotomy implantation in these patients because other variables were identified. 5 Nevertheless, for reasons of optimal defibrillating efficacy with nontho- racotomy configurations and to ensure long-term lead stability, the tripolar lead should ideally be po- sitioned at the apex of the right ventricle. Finally, in some patients a more practical solution was sought, for example, reprogramming from DDD to AAI modes. It may be argued that this avoided rather than solved potential sensing interaction between the two ventricular leads, although in these instances the modification was sufficient and the patient was thereby exposed to a minimal number of interven- tions.

Volume 129, Number 1 American Heart Journal ~roo~8 et aL 53

REFERENCES

1. Winkle RA, Mead RH, Ruder MA, Gaudiani VA, Smith NA, Buch WS, Schmidt P, Shipman T. Long-term outcome with the automatic implantable cardioverter-defibrillator. J Am Cell Cardio11989;13:1353- 61.

2. Epstein AE, Kay GN, Plumb VJ, Shepard RB, Kirkhn JK. Combined automatic implantable cardioverter-defibrillator and pacemaker sys- tems: implantation techniques and follow-up. J Am Coil Cardiol 1989;13:121-31.

3. Calkins H, Brinker J, Velm E, Guarnieri T, Levine JH. Clinical rater- actions between pacemakers and automatic implantable cardioverter- defibrillators. J Am Cell Cardiol 1990;16:666-73.

4. Kim SG, Gurman S, Waspe LE, Brodman R, Fisher JD. Unipolar pacer artifacts induced failure of automatic implantable cardioverter-de- fibri]lator to detect ventricular fibrillation. Am J Cardio11986;57:880-1.

5. Brooks R, Garan H, Torchiana D, Vlahakes GJ, Newell J, McGovern

BA, Garan H, Ruskin JN. Determinants of successful nonthoracotomy cardieverter-defibrillator implantation: experience in 101 patients us- ing two different lead systems. J Am Cell Cardiel 1993;22:1835-42.

6. Furman S. Sensing and timmg the cardiac electrogram. In: Furman S, Hayes DL, and Holmes DR, eds. A practice of cardiac pacing. Mount Kisco, N.Y.: Futura, 1989:79-114.

7. Rosenqvist M, Obel I. Atrial pacing and the risk of AV block: is there a time for change in attitude? PACE 1989;12:97-10L

8. Barold SS. The fourth decade of cardiac pacing: optimal modes of stimulation. Primary Cardiol 1993;19:16-32.

9. Kelly PA, Cannom DS, Garan H, Mirabal GS, Harthorne JW, Hurvitz RT. Vlahakes GJ, Jacobs MA, Ilvento JP, BuckIey M J, Ruskin JN. The automatic implantable cardioverter-defibrillator; efficacy, complica- tions, and survival in patients with malignant ventricular arrhythmias. J Am Cell Cardiol. 1988;11: 1278-86.

AVAILABILITY OF JOURNAL BACK ISSUES

As a service to our subscribers, copies of back issues of the AMERICAN HEART JOURNAL for the preceding 5 years are main- tained and are available for purchase from the publisher, Mosby, at a cost of $10.00 per issue. The following quantity discounts are available: 25 % off on quantities of 12 to 23. and one-third off on quantities of 24 or more. Please write to Mosby-Year Book, Inc., Subscription Services, 11830 Westline Industrial Drive, St. Louis, Me 63146-3318. or call (800)453-4351 or (314)453-4351 for in- formation on availability of particular issues. If unavailable from the publisher, photocopies of complete issues are available from University Microfilms International, 300 N. Zeeb Rd., Ann Arbor, MI 48106, (313)761-4700.