Thermo Scientific HyPerforma S.U.B. TK (Jacketed)Quick Start Guide
Thermo Scientific HyPerforma S.U.B. TK (Jacketed)Quick Start GuideRev. 1.0
Thermo Scientific 3 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
Section 1 TCU Setup and Installation
NOTES:
• TheseinstructionsonlyapplytojacketedsystemsusingtheLaudaT1200TCUfortemperaturecontrol.
• Materials/Suppliesneeded:o TCUhosekitshippedwiththecontroltowero TCUcommunicationcableshippedwiththetower(● ● ●)o Coolantliquido FunnelusedforaddingtheCoolantLiquido Trayorbucketforcatchingfluidduringthepurgeprocess
• PleaserefertoLAUDATCUmanualprovidedifmoredetailsarerequired.
1. PlacetheTCUnexttotheS.U.B.andcontroltowerunit.2. Ensurethevalveforthedrainisclosed(seeFigure1).
Figure 1. TCU connection
Thermo Scientific 4 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
3. Connections:(CablesandHosesprovided)a. ConnectoneoftheprovidedhosestotheoutletoftheTCU
andtotheinletoftheS.U.B.b. ConnecttheremaininghosetotheinletoftheTCUandthe
outletoftheS.U.B.c. Connectthecommunicationcable(● ● ●)fromthefrontof
theTCUtotherearofthetower.NotethattheconnectionontheTCUisbelowthedisplayandcanonlybeaccessedwhenthefaceistiltedup.
d. ConnectmainpowertoTCU.
4. Reservoir:a. Reservoirneedstobefilledtomid-levelwithcoolant/water
mix–seeLaudamanualfor(centermarkonthefluidlevelgage).AccesstofillthereservoirislocatedonthefrontoftheTCU.Fillslowlyandmonitorfillgaugeperiodically.
Figure 2. TCU fill level gauge at half full
Thermo Scientific 5 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
5. PurgeTCUandJacket:a. Openvalves.Ensureinletandoutletvalvesonthejacketare
open.b. Turnonthemainpowerusingthetoggleswitchatthefrontof
theTCU.c. Onstartup,theTCUwillself-testandbegintopurgetheair
outoflinesandjacket.d. Duringthepurgeprocess,thereservoirlevelwilldropinorder
tofillthejacketoftheS.U.B.NOTE:VerifybypassvalveisnotrecirculatingfluidinTCU
e. Whenthelevelgetstoolow,theTCUwillsetoffanalarm.Repeatstep4andfillthereservoirtomid-levelandthealarmwillturnoffandthemachinewillcontinuetopurgeandfillthejacket.
f. Thisprocessmayneedtoberepeatedseveraltimesuntilthejacketisfullandmostoftheairisremovedfromthesystem.
g. Placeatrayorbucketunderneaththejacketventvalve.h. Keeptheventvalveopenuntilcoolingfluidcomingoutof
thevalveisfreeofair(TCUneedstoberunningduringthisprocess).Thismaytake1-2minutes.Thenclosethevalve.
i. DoafinalcheckonthefluidleveloftheTCUandensurethefluidlevelisatmid-level.
j. Ifnecessary,adjustdefaulttemperature.
CAUTION: Donotfillreservoirtotopoffilllevel,thecoolantmayneedroomforexpansion.
6. WhentheTCUisinitsfinallocation,rollthecontroltowerintoplaceovertheTCUanduptothestoplocatedinsidethetower.Notethatanaccessdoorislocatedonthefrontofthetowertoviewormanipulatethecontrols.
Figure 3. Jacket Purge Valve
Thermo Scientific 1 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
2.1 Orientation and Location
Section 2 Hardware Installation
TheSingle-UseBioreactorhardwareisshippeddirectlyfromthemanufacturerandwillarrivecrated.Forunpackinginstructionsanddetailedcontentsofthecrate,pleaserefertotheuncratinginstructionsthatcamewiththeshipment.Besuretofollowtheunpackinginstructionsprovidedandretainallpackagingmaterials.
Tools Needed:• Allenwrenchprovidedinshippingdocumentationpacket• Adjustablewrenchneededforlevelingfeet(notsupplied)
a. Positionreactorandtowerindesiredlocationandlockwheelsensuringavailabilityofneededutilities(power,gas,etc.).Rightorlefthandorientationwasdeterminedattimeoforder.Ensurethattowerislocatedonthesamesideasthecableandtubingorganizer.
b. Beforerunningcables,notethatthecableandtubingorganizer
hasanadjustmentfeaturetokeepthecableslinedupwiththerearofthecontroltower.Thearmhasaslidethatcanextendandretractmanuallyusingtheclampingknob.
WARNING:Anyproceduresthatrequirethecontrolboxtobeopenshouldbeperformedwiththemainelectricaldisconnectinthelockedoutpositionandallpowersourcesremovedfromthecontrolbox.Foroperatorsafety,securethelocationoftheS.U.B.outersupportcontainerbydisablingtheswivelcastersbeforeservicing.
Figure 4. Cable and Tubing Organizer Figure 5. Cable and Tubing Organizer Adjustment Feature
Thermo Scientific 2 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
2.2 Attaching Cables and Gas Lines
Figure 6. Back panel labeling
• ConnectMotorPowercableintodesignatedTowerconnector(● ● ●)• ConnectMotorFeedbackcableintodesignatedTowerconnector(● ● ●)• ConnectLoadCellcableintodesignatedTowerconnector(● ● ●)
• Usethethree(3)color-coded,labeledgaslinesandconnecteachlinetotherespectivecoloreddotsourceontheTower.
i. OverlaySparge–Black(●)ii. MicroSparge–Red(●)iii. MacroSparge-White(●)
• Servicegaspressureshouldbeaminimumof30psi,andamaximumof35psiforproperoperation.Verifythatgasconnectionsaresecureandtherearenoleaks(e.g.Snooporsimilarproducttodetectleaks).
• ConnecttheRTD(temperatureprobe)cable(● ● ●).Thiscableshouldbeinaboxthatwasshippedwiththetower.
ConnectthecorrespondingfacilitygassourcestothedesignatedconnectorsontheTower.Theconnectorsarefractional(inches),soifthefacilityusesmetriclines,simplyusetheadaptersprovidedintheplasticbaginsidetheTower(6mm.to1/4”adapters).
Thermo Scientific 3 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
• Attachprobeclipstoprobewindow(frontoftheS.U.B.).TheprobeclipswereshippedwiththeS.U.B.unit.
Figure 7. 250 L S.U.B. TK System
Positionandattachvarioushardwareitems:• Securethetoolbrackettothetank.Thispositionisnotcritical,
butwouldbeconvenientnearthemotor.Placetheratchetandthecustomopenendwrenchinthetoolholder.Attachdriveshaft,asshown.
• Attachtheadjustableventfilterheaterbracketusingtheclampleverhandles.Positionthishardwaredirectlyoppositeofthemotor,asshown.
Thermo Scientific 4 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
NOTE:Beforeconnectingcables,routethemthroughthetubingandcableorganizerfirst• IfyouintendtousetheFinesseTruFluorsingle-usesensors,route
theappropriatecablesfromtheTowerintothecablemanagementarmanddowntothesensorwindowareaontheS.U.B.ThesecableswillalreadybeattachedtotherearoftheControlTowerviaaliquidtightconnectionandhavethefollowingdesignation,DO-2andpH-2(Nocolorcode).
• Ifyouintendtousetraditionalre-usablesensingprobes,routetheappropriatecablesinthesamemannerasdescribedaboveforthesingle-usesensors.Ensuretheyarepluggedintothebackofthecontroltower(● DO-1●,● pH-1●).
• LocatethebagpressuresensorcableontheTower(● ● ●)androuteitviathecablemanagementarmanduptowardthetoparmofthetank.
Ifusingaventheatermakethefollowingconnections.• ConnectcabletolabeledVH-1POWER(● ●)powersourceon
Tower.• ConnectYellowThermocoupletolabeledconnectiononTower
VH-1FDBK(● ●)• Routecablethroughcablemanagementarmanduptoventheater
bracketsontopofS.U.B.unit.• Ifusingasecondventheater,useVH-2connections
VH-2POWER(● ● ●)andVH-2FDBK(● ● ●)androuteparallelwithVH-1cable.
Vent Filter Heater
Ifusingaresistiveheatermakethefollowingconnections.• ConnectcablelabeledRESHEATER(● ● ●)tothepower
sourceontheTower• ConnectcablelabeledHEATERPOWERIN(● ● ●)into
designatedtowerconnector.
Resistive Heater Power Connection
Thermo Scientific 5 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
Beforeconnectingpower,verify,again,allgaslineswerecheckedandnotleaking.
• Connectmainpowercord(nocolorcode)tocorrectutilitysource.• ItisstronglyrecommendedthataUPS(UninterruptedPower
Supply)(● ● ●)beused.Thepurposeofthisconnectionistomaintainpowertothecontrolsduringapoweroutage.
• IfUPSpowerisnotavailableornotused,thentheUPSJumperconnectormustbeputinplaceoftheUPScable.
2.3 Power Connections
UPS Jumper
Figure 8. UPS Jumper
EnsuretheS.U.B.unitisinitsintendedposition.• Lockthevesselinplacebyloweringthelevelingfeet.Ensurethat
unitislevelbeforeoperation.• UsingtheAllenwrenchprovidedwithS.U.B.unitshipping
documents,removetheloadcellshippingplatesandsecuretheminthebracketattherearofthecontroltower.Therearetwo¼-20socketheadcapscrewsthatneedtoberemovedfromeachloadcellprotectionplate.Movethetowertodesiredpositionandlockinplaceandlevelusingthelevelingfeet.
2.4 Securing Position of Control Tower and
S.U.B.
Thermo Scientific 6 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
• RefertoUser’sGuidefordetailedinstructionsoncalibrationsteps(loadcells,sensors,etc).
• Avirtualkeyboardiconislocatedatthebottomofthewindowdisplayfornotes,values,etc.
• StartbyturningontheMAINDISCONNECTswitchlocatedatthebackofthecontroltower.
• EnsuretheEmergencyStop(locatedatthefrontofthetower)isnotactivated(shouldbeintheoutposition).Ifitisengaged,thentwisttodisengage.NotethattheE-stoplightwillremainred,eventhoughitisdisengaged,untilthenextstephasbeencompleted.
• TurnpoweronbypressingthegreenStart/Re-setbuttonlocatedatthefrontofthetowerbehindtheoperatorscreen.E-stoplightwillthenturnoff.
• Notethatagreenoperationlight,locatedatthebackofthetower,willalsoturnon.
Themonitorcanbetiltedupanddownandrotatelefttoright.2.5 HMI (Monitor)
2.6 Power Up the System
2.7 System Will Begin Booting Up
Thermo Scientific 7 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
Section 3 BPC Installation
1. CarefullyunpacktheS.U.B.BioProcessContainer(BPC)fromitstwoprotectivepolybags.
2. Uncoilandremoveprotectivepackagingfromthesparger,drain,andauxiliarylinesonthebottomoftheBPC.Don’tremovetheprotectivepackagingontopoftheBPCatthistime.
3. CarefullylowerthebottomendoftheBPCintotheS.U.B.withanytubingsetsonthetopsidelayingoverthesideuntilthebearingassemblyislevelwiththedriveshaftmotor.
4. Removetheprotectiveblackcapfromthebearingassemblyandslideintotheretentionmechanismontheagitationmotorandlockitintoplace.
Figure 9. Bearing port inserted into motor mount slot and clamp
Thermo Scientific 8 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
5. Guidebottomtubinglinesetthroughtheirproperchannels:• Insertfritandopenpipespargerlinesthroughthespargeplate
(round,whiteplasticwithslots)first,beforelocatingitinthebaseofunit.
• Insertdrainlinethroughbottomfrontfacedirectlybelowagitationmotor.Hangthedrainlinecoilonthearmprovided.Alsoroutethesamplelinethroughthesnapinslotsonthedraincoilarm.
• AttachBPCpositioningtabs(bottomofchamber)ontothealignmentpinslocatedatthebaseoftheS.U.B.tank.Notethatthe50LBPCdoesnotrequirethisstepandisnotrecommended,buthastabholdersavailable,ifdesired.
• Insertsensors,KleenPaks,andauxiliarylinesetsthroughthe“probebeltopeningonthetank”.
6. Uncoverventfilters,wrapheatingblankets(ifused)aroundthefilters,attach/adjustventbracket,andmountventfiltersinbracket.
Figure 10. BPC tab holders used for proper BPC positioning
Figure 11. Filter being placed in bracket
Thermo Scientific 9 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
7. Routetheoverlayspargeset(locatedatthetopoftheBPCmarked“OverlaySparge”)throughthecableandtubingorganizerarmtothetopofthemainbeamofthecableandtubingorganizerandconnecttothefittingmarkedOVERLAYSPARGE.Thiswillbethelinesetthatwillbeusedtopre-filltheBPCwithair,soensurethattheclamponthislinesetisreleased.
8. RoutetheMicroSpargeLine(labeled),locatedatthebottomoftheBPC,upthroughthemainbeamofthecableandtubingorganizerusingtheblacksnapinfeatures.ConnecttothecorrectgasfittinglabeledMICROSPARGE.Unclampthislinesetwhenreadytospargethroughfrit.
9. RouteMacroSpargeline(labeled),locatedatthebottomoftheBPC,runningparalleltotheMicroSpargeLine,usingthesamesnapinfeatures.ConnecttothecorrectgasfittinglabeledMACROSPARGE.Ensuretheclampisreleasedonthisline,aswellbeforeoperation.NOTE:Toreducetheriskoffiltermembranewetting,positioneachspargefilterabovethelowpointofthespargeline.
10.ConnectthepressuresensorcableontothesensorlocatedatthetopoftheBPC.
Figure 13. Pressure sensor
Figure 12. BPC linesets
Thermo Scientific 10 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
11.EnsurethecablehasbeenroutedthroughthecableandtubingorganizerandconnecttothefittinglocatedattherearofthecontrolcabinetlabeledBAGPRESS(● ● ●).
12. IMPORTANTDisengageanyclampsonthelinesetfortheexhaustfilter(s).Iftheclampsareleftengaged,theBPCcouldburstwhenfillingitwithairorliquid.
13.Positionandcloseabarclamp(PartNo.SV20664.01)onthebottomdrainlineascloseaspossibletotheBPCport(Figure14).Alsoplacebarclampsonprobeportsintendedforconventionalprobes(Figure15).
14.InsertRTDorselectedtemperaturesensorintothermo-well(Figure16,Figure17).• Placesmallamountofglycerol(0.5mL)inwelltoaidinheat
transfer.• Sensorshouldbeinserteduntilbaseofprobemeetsthemouth
ofthethermo-well.• Securebytwistingtheluerlockcollar.Thethermo-wellwill
stretchslightlywhenRTDisseated.
Figure 16
Figure 14
Figure 17
Figure 3.15
Thermo Scientific 11 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
Withthesoftwarebootedup,clickonthe“HOME”buttonontheupperleftsideofthescreen.Inthetoprighthandcornerofthescreen,clickontheicon“Guest.”
ThenclickontheUserpulldownmenuandselect“Administrator.”
EnterthecorrectPassword,whichis“admin.”
TotypeinformationwithoutpluggingaUSBkeyboardintothecontroltower,clickontheminiaturekeyboardiconlocatedatthebottommiddleofthescreen.Thiswillbringupatouchscreenkeyboard.Whenfinishedusingthetouchscreenkeyboardjustclickontheupperrighthandcornertocloseit.Note,youmustbringupthekeyboardFIRSTbeforeopeningadropdownmenu.Alsonotethatthekeyboardcanbemovedandrepositioned.
Figure 18 Initial TruBio View With Log On Breakout
Section 4 Software Start Up
Thermo Scientific 12 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
Nowtheheadingshouldread“Administrator.”
ThenextstepistoloadthecorrectconfigurationfilefortheSUBunit.Clickonthe“Configure”buttonontheleftsideofthescreen.
NOTE:youcanalsoclickontheyellow“LoadConfigurationFile”buttonwhenfirstturningonthesystem.
Clickthetabonthetopthatreads“File.”
SelecttheappropriateS.U.B.TKsystemtobeusedbyclickingtohighlightit.
• ThermoTurnkey250LElectricSUB• ThermoTurnkey250LJacketedSUB• ThermoTurnkey50LElectricSUB• ThermoTurnkey50LJacketedSUB
Loadthenewconfigurationbyclickingthe“Load”buttonontheright.
Select“Yes”whenthepopupwindowappears.Thescreenautomaticallyjumpstothe“Home”displaywindow.
Figure 19 Configure Files View
Thermo Scientific 13 S.U.B. TK Jacketed Quick Start Guide
Quick Start Guide
Figure 20 “Home” Monitor View
Thermo Scientific 14 S.U.B. TK Jacketed Quick Start Guide