Thermo Scientific HyPerforma S.U.B. TK (Jacketed)Quick Start Guide

Page 2: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific HyPerforma S.U.B. TK (Jacketed)Quick Start GuideRev. 1.0

Page 3: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 3 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

Section 1 TCU Setup and Installation

NOTES:

• TheseinstructionsonlyapplytojacketedsystemsusingtheLaudaT1200TCUfortemperaturecontrol.

• Materials/Suppliesneeded:o TCUhosekitshippedwiththecontroltowero TCUcommunicationcableshippedwiththetower(● ● ●)o Coolantliquido FunnelusedforaddingtheCoolantLiquido Trayorbucketforcatchingfluidduringthepurgeprocess

• PleaserefertoLAUDATCUmanualprovidedifmoredetailsarerequired.

1. PlacetheTCUnexttotheS.U.B.andcontroltowerunit.2. Ensurethevalveforthedrainisclosed(seeFigure1).

Figure 1. TCU connection

Page 4: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 4 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

3. Connections:(CablesandHosesprovided)a. ConnectoneoftheprovidedhosestotheoutletoftheTCU

andtotheinletoftheS.U.B.b. ConnecttheremaininghosetotheinletoftheTCUandthe

outletoftheS.U.B.c. Connectthecommunicationcable(● ● ●)fromthefrontof

theTCUtotherearofthetower.NotethattheconnectionontheTCUisbelowthedisplayandcanonlybeaccessedwhenthefaceistiltedup.

d. ConnectmainpowertoTCU.

4. Reservoir:a. Reservoirneedstobefilledtomid-levelwithcoolant/water

mix–seeLaudamanualfor(centermarkonthefluidlevelgage).AccesstofillthereservoirislocatedonthefrontoftheTCU.Fillslowlyandmonitorfillgaugeperiodically.

Figure 2. TCU fill level gauge at half full

Page 5: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 5 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

5. PurgeTCUandJacket:a. Openvalves.Ensureinletandoutletvalvesonthejacketare

open.b. Turnonthemainpowerusingthetoggleswitchatthefrontof

theTCU.c. Onstartup,theTCUwillself-testandbegintopurgetheair

outoflinesandjacket.d. Duringthepurgeprocess,thereservoirlevelwilldropinorder

tofillthejacketoftheS.U.B.NOTE:VerifybypassvalveisnotrecirculatingfluidinTCU

e. Whenthelevelgetstoolow,theTCUwillsetoffanalarm.Repeatstep4andfillthereservoirtomid-levelandthealarmwillturnoffandthemachinewillcontinuetopurgeandfillthejacket.

f. Thisprocessmayneedtoberepeatedseveraltimesuntilthejacketisfullandmostoftheairisremovedfromthesystem.

g. Placeatrayorbucketunderneaththejacketventvalve.h. Keeptheventvalveopenuntilcoolingfluidcomingoutof

thevalveisfreeofair(TCUneedstoberunningduringthisprocess).Thismaytake1-2minutes.Thenclosethevalve.

i. DoafinalcheckonthefluidleveloftheTCUandensurethefluidlevelisatmid-level.

j. Ifnecessary,adjustdefaulttemperature.

CAUTION: Donotfillreservoirtotopoffilllevel,thecoolantmayneedroomforexpansion.

6. WhentheTCUisinitsfinallocation,rollthecontroltowerintoplaceovertheTCUanduptothestoplocatedinsidethetower.Notethatanaccessdoorislocatedonthefrontofthetowertoviewormanipulatethecontrols.

Figure 3. Jacket Purge Valve

Page 6: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 1 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

2.1 Orientation and Location

Section 2 Hardware Installation

TheSingle-UseBioreactorhardwareisshippeddirectlyfromthemanufacturerandwillarrivecrated.Forunpackinginstructionsanddetailedcontentsofthecrate,pleaserefertotheuncratinginstructionsthatcamewiththeshipment.Besuretofollowtheunpackinginstructionsprovidedandretainallpackagingmaterials.

Tools Needed:• Allenwrenchprovidedinshippingdocumentationpacket• Adjustablewrenchneededforlevelingfeet(notsupplied)

a. Positionreactorandtowerindesiredlocationandlockwheelsensuringavailabilityofneededutilities(power,gas,etc.).Rightorlefthandorientationwasdeterminedattimeoforder.Ensurethattowerislocatedonthesamesideasthecableandtubingorganizer.

b. Beforerunningcables,notethatthecableandtubingorganizer

hasanadjustmentfeaturetokeepthecableslinedupwiththerearofthecontroltower.Thearmhasaslidethatcanextendandretractmanuallyusingtheclampingknob.

WARNING:Anyproceduresthatrequirethecontrolboxtobeopenshouldbeperformedwiththemainelectricaldisconnectinthelockedoutpositionandallpowersourcesremovedfromthecontrolbox.Foroperatorsafety,securethelocationoftheS.U.B.outersupportcontainerbydisablingtheswivelcastersbeforeservicing.

Figure 4. Cable and Tubing Organizer Figure 5. Cable and Tubing Organizer Adjustment Feature

Page 7: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 2 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

2.2 Attaching Cables and Gas Lines

Figure 6. Back panel labeling

• ConnectMotorPowercableintodesignatedTowerconnector(● ● ●)• ConnectMotorFeedbackcableintodesignatedTowerconnector(● ● ●)• ConnectLoadCellcableintodesignatedTowerconnector(● ● ●)

• Usethethree(3)color-coded,labeledgaslinesandconnecteachlinetotherespectivecoloreddotsourceontheTower.

i. OverlaySparge–Black(●)ii. MicroSparge–Red(●)iii. MacroSparge-White(●)

• Servicegaspressureshouldbeaminimumof30psi,andamaximumof35psiforproperoperation.Verifythatgasconnectionsaresecureandtherearenoleaks(e.g.Snooporsimilarproducttodetectleaks).

• ConnecttheRTD(temperatureprobe)cable(● ● ●).Thiscableshouldbeinaboxthatwasshippedwiththetower.

ConnectthecorrespondingfacilitygassourcestothedesignatedconnectorsontheTower.Theconnectorsarefractional(inches),soifthefacilityusesmetriclines,simplyusetheadaptersprovidedintheplasticbaginsidetheTower(6mm.to1/4”adapters).

Page 8: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 3 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

• Attachprobeclipstoprobewindow(frontoftheS.U.B.).TheprobeclipswereshippedwiththeS.U.B.unit.

Figure 7. 250 L S.U.B. TK System

Positionandattachvarioushardwareitems:• Securethetoolbrackettothetank.Thispositionisnotcritical,

butwouldbeconvenientnearthemotor.Placetheratchetandthecustomopenendwrenchinthetoolholder.Attachdriveshaft,asshown.

• Attachtheadjustableventfilterheaterbracketusingtheclampleverhandles.Positionthishardwaredirectlyoppositeofthemotor,asshown.

Page 9: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 4 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

NOTE:Beforeconnectingcables,routethemthroughthetubingandcableorganizerfirst• IfyouintendtousetheFinesseTruFluorsingle-usesensors,route

theappropriatecablesfromtheTowerintothecablemanagementarmanddowntothesensorwindowareaontheS.U.B.ThesecableswillalreadybeattachedtotherearoftheControlTowerviaaliquidtightconnectionandhavethefollowingdesignation,DO-2andpH-2(Nocolorcode).

• Ifyouintendtousetraditionalre-usablesensingprobes,routetheappropriatecablesinthesamemannerasdescribedaboveforthesingle-usesensors.Ensuretheyarepluggedintothebackofthecontroltower(● DO-1●,● pH-1●).

• LocatethebagpressuresensorcableontheTower(● ● ●)androuteitviathecablemanagementarmanduptowardthetoparmofthetank.

Ifusingaventheatermakethefollowingconnections.• ConnectcabletolabeledVH-1POWER(● ●)powersourceon

Tower.• ConnectYellowThermocoupletolabeledconnectiononTower

VH-1FDBK(● ●)• Routecablethroughcablemanagementarmanduptoventheater

bracketsontopofS.U.B.unit.• Ifusingasecondventheater,useVH-2connections

VH-2POWER(● ● ●)andVH-2FDBK(● ● ●)androuteparallelwithVH-1cable.

Vent Filter Heater

Ifusingaresistiveheatermakethefollowingconnections.• ConnectcablelabeledRESHEATER(● ● ●)tothepower

sourceontheTower• ConnectcablelabeledHEATERPOWERIN(● ● ●)into

designatedtowerconnector.

Resistive Heater Power Connection

Page 10: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 5 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

Beforeconnectingpower,verify,again,allgaslineswerecheckedandnotleaking.

• Connectmainpowercord(nocolorcode)tocorrectutilitysource.• ItisstronglyrecommendedthataUPS(UninterruptedPower

Supply)(● ● ●)beused.Thepurposeofthisconnectionistomaintainpowertothecontrolsduringapoweroutage.

• IfUPSpowerisnotavailableornotused,thentheUPSJumperconnectormustbeputinplaceoftheUPScable.

2.3 Power Connections

UPS Jumper

Figure 8. UPS Jumper

EnsuretheS.U.B.unitisinitsintendedposition.• Lockthevesselinplacebyloweringthelevelingfeet.Ensurethat

unitislevelbeforeoperation.• UsingtheAllenwrenchprovidedwithS.U.B.unitshipping

documents,removetheloadcellshippingplatesandsecuretheminthebracketattherearofthecontroltower.Therearetwo¼-20socketheadcapscrewsthatneedtoberemovedfromeachloadcellprotectionplate.Movethetowertodesiredpositionandlockinplaceandlevelusingthelevelingfeet.

2.4 Securing Position of Control Tower and

S.U.B.

Page 11: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 6 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

• RefertoUser’sGuidefordetailedinstructionsoncalibrationsteps(loadcells,sensors,etc).

• Avirtualkeyboardiconislocatedatthebottomofthewindowdisplayfornotes,values,etc.

• StartbyturningontheMAINDISCONNECTswitchlocatedatthebackofthecontroltower.

• EnsuretheEmergencyStop(locatedatthefrontofthetower)isnotactivated(shouldbeintheoutposition).Ifitisengaged,thentwisttodisengage.NotethattheE-stoplightwillremainred,eventhoughitisdisengaged,untilthenextstephasbeencompleted.

• TurnpoweronbypressingthegreenStart/Re-setbuttonlocatedatthefrontofthetowerbehindtheoperatorscreen.E-stoplightwillthenturnoff.

• Notethatagreenoperationlight,locatedatthebackofthetower,willalsoturnon.

Themonitorcanbetiltedupanddownandrotatelefttoright.2.5 HMI (Monitor)

2.6 Power Up the System

2.7 System Will Begin Booting Up

Page 12: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 7 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

Section 3 BPC Installation

1. CarefullyunpacktheS.U.B.BioProcessContainer(BPC)fromitstwoprotectivepolybags.

2. Uncoilandremoveprotectivepackagingfromthesparger,drain,andauxiliarylinesonthebottomoftheBPC.Don’tremovetheprotectivepackagingontopoftheBPCatthistime.

3. CarefullylowerthebottomendoftheBPCintotheS.U.B.withanytubingsetsonthetopsidelayingoverthesideuntilthebearingassemblyislevelwiththedriveshaftmotor.

4. Removetheprotectiveblackcapfromthebearingassemblyandslideintotheretentionmechanismontheagitationmotorandlockitintoplace.

Figure 9. Bearing port inserted into motor mount slot and clamp

Page 13: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 8 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

5. Guidebottomtubinglinesetthroughtheirproperchannels:• Insertfritandopenpipespargerlinesthroughthespargeplate

(round,whiteplasticwithslots)first,beforelocatingitinthebaseofunit.

• Insertdrainlinethroughbottomfrontfacedirectlybelowagitationmotor.Hangthedrainlinecoilonthearmprovided.Alsoroutethesamplelinethroughthesnapinslotsonthedraincoilarm.

• AttachBPCpositioningtabs(bottomofchamber)ontothealignmentpinslocatedatthebaseoftheS.U.B.tank.Notethatthe50LBPCdoesnotrequirethisstepandisnotrecommended,buthastabholdersavailable,ifdesired.

• Insertsensors,KleenPaks,andauxiliarylinesetsthroughthe“probebeltopeningonthetank”.

6. Uncoverventfilters,wrapheatingblankets(ifused)aroundthefilters,attach/adjustventbracket,andmountventfiltersinbracket.

Figure 10. BPC tab holders used for proper BPC positioning

Figure 11. Filter being placed in bracket

Page 14: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 9 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

7. Routetheoverlayspargeset(locatedatthetopoftheBPCmarked“OverlaySparge”)throughthecableandtubingorganizerarmtothetopofthemainbeamofthecableandtubingorganizerandconnecttothefittingmarkedOVERLAYSPARGE.Thiswillbethelinesetthatwillbeusedtopre-filltheBPCwithair,soensurethattheclamponthislinesetisreleased.

8. RoutetheMicroSpargeLine(labeled),locatedatthebottomoftheBPC,upthroughthemainbeamofthecableandtubingorganizerusingtheblacksnapinfeatures.ConnecttothecorrectgasfittinglabeledMICROSPARGE.Unclampthislinesetwhenreadytospargethroughfrit.

9. RouteMacroSpargeline(labeled),locatedatthebottomoftheBPC,runningparalleltotheMicroSpargeLine,usingthesamesnapinfeatures.ConnecttothecorrectgasfittinglabeledMACROSPARGE.Ensuretheclampisreleasedonthisline,aswellbeforeoperation.NOTE:Toreducetheriskoffiltermembranewetting,positioneachspargefilterabovethelowpointofthespargeline.

10.ConnectthepressuresensorcableontothesensorlocatedatthetopoftheBPC.

Figure 13. Pressure sensor

Figure 12. BPC linesets

Page 15: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 10 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

11.EnsurethecablehasbeenroutedthroughthecableandtubingorganizerandconnecttothefittinglocatedattherearofthecontrolcabinetlabeledBAGPRESS(● ● ●).

12. IMPORTANTDisengageanyclampsonthelinesetfortheexhaustfilter(s).Iftheclampsareleftengaged,theBPCcouldburstwhenfillingitwithairorliquid.

13.Positionandcloseabarclamp(PartNo.SV20664.01)onthebottomdrainlineascloseaspossibletotheBPCport(Figure14).Alsoplacebarclampsonprobeportsintendedforconventionalprobes(Figure15).

14.InsertRTDorselectedtemperaturesensorintothermo-well(Figure16,Figure17).• Placesmallamountofglycerol(0.5mL)inwelltoaidinheat

transfer.• Sensorshouldbeinserteduntilbaseofprobemeetsthemouth

ofthethermo-well.• Securebytwistingtheluerlockcollar.Thethermo-wellwill

stretchslightlywhenRTDisseated.

Figure 16

Figure 14

Figure 17

Figure 3.15

Page 16: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 11 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

Withthesoftwarebootedup,clickonthe“HOME”buttonontheupperleftsideofthescreen.Inthetoprighthandcornerofthescreen,clickontheicon“Guest.”

ThenclickontheUserpulldownmenuandselect“Administrator.”

EnterthecorrectPassword,whichis“admin.”

TotypeinformationwithoutpluggingaUSBkeyboardintothecontroltower,clickontheminiaturekeyboardiconlocatedatthebottommiddleofthescreen.Thiswillbringupatouchscreenkeyboard.Whenfinishedusingthetouchscreenkeyboardjustclickontheupperrighthandcornertocloseit.Note,youmustbringupthekeyboardFIRSTbeforeopeningadropdownmenu.Alsonotethatthekeyboardcanbemovedandrepositioned.

Figure 18 Initial TruBio View With Log On Breakout

Section 4 Software Start Up

Page 17: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 12 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

Nowtheheadingshouldread“Administrator.”

ThenextstepistoloadthecorrectconfigurationfilefortheSUBunit.Clickonthe“Configure”buttonontheleftsideofthescreen.

NOTE:youcanalsoclickontheyellow“LoadConfigurationFile”buttonwhenfirstturningonthesystem.

Clickthetabonthetopthatreads“File.”

SelecttheappropriateS.U.B.TKsystemtobeusedbyclickingtohighlightit.

• ThermoTurnkey250LElectricSUB• ThermoTurnkey250LJacketedSUB• ThermoTurnkey50LElectricSUB• ThermoTurnkey50LJacketedSUB

Loadthenewconfigurationbyclickingthe“Load”buttonontheright.

Select“Yes”whenthepopupwindowappears.Thescreenautomaticallyjumpstothe“Home”displaywindow.

Figure 19 Configure Files View

Page 18: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 13 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide

Figure 20 “Home” Monitor View

Page 19: HyPerforma Single-use Bioreactor TK (SUB TK): Quick Start ...tools.thermofisher.com/content/sfs/manuals/SUB TK Jacketed Quick Start Guide.pdf · the jacket. f. This process may need

Thermo Scientific 14 S.U.B. TK Jacketed Quick Start Guide

Quick Start Guide