1
Bottom Loading Hydrogen Purging Furnace (BLF - H - 1700) Maximum Temperature : 1700°C Working Chamber* : 130 x 130 mm Controller : PID Temperature Controller Controlling sensor : Thermocouple “B” type Heating element : “U” MoSi2 Heating element Insulation : Multiple layers of high temperature fiber insulation boards ensure minimum heat loss. Flages : Water Cooling Flanges with viton “O” ring Time to reach max : 1.5 Hr temperature Temperature uniformity : approx +/- 5°C Bottom lifting arrangement : Bottom lifting platform with D.C motor ensures smooth lifting and lowering of material Total lift movement : Approx 200mm Outer Construction : CRCA sheets duly powder coated. Power supply : 230V AC, 50 Hz * Customized chamber size available on request. General Information Bottom loading furnace BLF - H - 1700 is a hydrogen purging furnace with inner tube dimension 100 mm (ID), 115 mm (OD) & 280 mm length. Hydrogen gas is purged in presence of vacuum at 1700°C and sample is being tested. It is a special furnace, which is made to use in reducing atmosphere. Bottom loading furnaces are designed for uniform thermal distribution inside the chamber and for easy loading and unloading of sample with help of lifting arrangement. Advantages 1. Easy loading and unloading of sample 2. Uniform Temperature distribution inside chamber 3. Fast Cooling 4. Low heat loss 5. smooth lifting arrangement Applications It is a special type of furnace which is made to use for reducing atmosphere. Specifications Tempsens Instruments (l) Pvt. Ltd. U#ll A-190, Road No.5, M.I.A., Udaipur-313003 (Rajsthan) INDIA Ph.:+91-294-3052962, 3052959, Fax.:+91-294-3052950 Email: [email protected] www.tempsens.com

Hydrogen Purging Furnace (H 1700) · Bottom loading furnace BLF - H - 1700 is a hydrogen purging furnace with inner tube dimension 100 mm (ID), 115 mm (OD) & 280 mm length. Hydrogen

  • Upload
    others

  • View
    8

  • Download
    0

Embed Size (px)

Citation preview

Bottom Loading Hydrogen Purging Furnace (BLF - H - 1700)

Maximum Temperature : 1700°C

Working Chamber* : 130 x 130 mm

Controller : PID Temperature Controller

Controlling sensor : Thermocouple “B” type

Heating element : “U” MoSi2 Heating element

Insulation : Multiple layers of high temperature fiber

insulation boards ensure minimum heat

loss.

Flages : Water Cooling Flanges with viton “O” ring

Time to reach max : 1.5 Hr temperature

Temperature uniformity : approx +/- 5°C

Bottom lifting arrangement : Bottom lifting platform with D.C motor

ensures smooth lifting and lowering of

material

Total lift movement : Approx 200mm

Outer Construction : CRCA sheets duly powder coated.

Power supply : 230V AC, 50 Hz

* Customized chamber size available on request.

General Information

Bottom loading furnace BLF - H - 1700 is a hydrogen purging furnace with inner tube dimension 100 mm (ID),

115 mm (OD) & 280 mm length. Hydrogen gas is purged in presence of vacuum at 1700°C and sample is being

tested. It is a special furnace, which is made to use in reducing atmosphere. Bottom loading furnaces are

designed for uniform thermal distribution inside the chamber and for easy loading and unloading of sample with

help of lifting arrangement.

Advantages

1. Easy loading and unloading of sample

2. Uniform Temperature distribution inside chamber

3. Fast Cooling

4. Low heat loss

5. smooth lifting arrangement

Applications

It is a special type of furnace which is made to

use for reducing atmosphere.

Specifications

Tempsens Instruments (l) Pvt. Ltd. U#llA-190, Road No.5, M.I.A., Udaipur-313003 (Rajsthan) INDIA

Ph.:+91-294-3052962, 3052959, Fax.:+91-294-3052950

Email: [email protected]