© 2010–2013 Onset Computer Corporation. All rights reserved. Onset, HOBO, HOBOlink, HOBOware, BoxCar, and FlexSmart are trademarks or registered trademarks of Onset Computer Corporation for its data logger products and configuration/interface software. Macintosh is a registered trademark of Apple, Inc. Microsoft and Windows are registered trademarks of Microsoft Corporation. All other trademarks are the property of their respective companies. 12730L MANBHWUG HOBOware® User’s Guide Onset Computer Corporation 470 MacArthur Blvd. Bourne, MA 02532 www.onsetcomp.com Mailing Address: P.O. Box 3450 Pocasset, MA 025593450 Phone: 1800LOGGERS (18005644377) or 5087599500 Fax: 5087599100 Email: [email protected] Technical Support Hours: 8AM to 8PM ET, Monday through Friday Customer Service Hours: 8AM to 5PM ET, Monday through Friday

HOBOware User's Guide - 12730

Embed Size (px)

DESCRIPTION

hoboware user guide complete

Citation preview

  • 20102013OnsetComputerCorporation.All rights reserved.Onset,HOBO,HOBOlink,HOBOware,BoxCar,andFlexSmartare trademarksor registered trademarksofOnsetComputerCorporation for its data logger products and configuration/interface software.Macintosh is a registered trademark ofApple, Inc.Microsoft andWindows are registered trademarks ofMicrosoftCorporation.Allothertrademarksarethepropertyoftheirrespectivecompanies.12730LMANBHWUG

    HOBOwareUsersGuide

    OnsetComputerCorporation

    470MacArthurBlvd.Bourne,MA02532

    www.onsetcomp.comMailingAddress:

    P.O.Box3450Pocasset,MA025593450

    Phone:1800LOGGERS(18005644377)or5087599500Fax:5087599100

    Email:[email protected]:8AMto8PMET,MondaythroughFridayCustomerServiceHours:8AMto5PMET,MondaythroughFriday

  • HOBOwareUsersGuide 2

    ContentsChapter 1: An Overview of HOBOware .............................................. 6IntroductiontoHOBOwarePro....................................................................................................7IntroductiontoHOBOwareLite....................................................................................................7AdvancedHOBOwareProFeatures..............................................................................................8NewHOBOwareFeatures.............................................................................................................9InstallingHOBOware..................................................................................................................11ActivatingHOBOware.................................................................................................................12HOBOwareSetupAssistant(HOBOwareProOnly)....................................................................13UsingaTrialVersionofHOBOware............................................................................................14ATouroftheHOBOwareInterface............................................................................................15OverviewofaPlot......................................................................................................................16TheDetailsPane.........................................................................................................................17ThePointsTable.........................................................................................................................18ThePlotSetupWindow..............................................................................................................19HOBOwareSoftwareUpdates....................................................................................................19

    Chapter 2: Working with HOBO Data Loggers .................................. 20ConnectingDevices....................................................................................................................20ConnectingaUSBDevice............................................................................................................21ConnectingaBaseStationorCoupler........................................................................................21ConnectingaSerialDevice(HOBOwareProOnly).....................................................................22ConnectingMultipleDevices......................................................................................................22LaunchingDevices......................................................................................................................23LaunchLoggerWindow..............................................................................................................24LaunchOptionsforHOBOUX90SeriesLoggers.........................................................................25LaunchOptionsforHOBOUX100SeriesLoggers.......................................................................28LaunchOptionsfortheHOBO4ChannelPulseInputDataLogger(UX120017x).....................31LaunchOptionsforOtherUSeriesLoggers...............................................................................34LaunchOptionsforStationLoggers............................................................................................37MultipleLoggingIntervals..........................................................................................................39FastLoggingIntervals.................................................................................................................40RepeatinganActiononMultipleUSeriesDevices....................................................................40"ReadHeaderFailed"Error........................................................................................................41LaunchUtilities...........................................................................................................................41FilterSeriesatLaunch.................................................................................................................41DataAssistantsWindow(Scaling)..............................................................................................43ConfigureAlarms........................................................................................................................44

  • HOBOwareUsersGuide 3

    AdvancedSensorProperties:PulseFrequencyandLockoutTime.............................................46AdvancedSensorProperties:Calibration...................................................................................47AdvancedSensorProperties:Occupancy...................................................................................48Statistics:Maximum,Minimum,Average,andStandardDeviation...........................................49FilteredSeriesvs.StatisticsLogging...........................................................................................50BurstLogging..............................................................................................................................51CheckingDeviceStatus...............................................................................................................52TheStatusWindow.....................................................................................................................53DeviceDetails.............................................................................................................................53CurrentReadings........................................................................................................................55StoppingaDevice.......................................................................................................................55LabCalibration............................................................................................................................56UpdatingUXSeriesLoggerFirmware.........................................................................................59

    Chapter 3: Reading Out, Plotting, and Analyzing Data...................... 61ReadingOutData........................................................................................................................61SettingupaPlot.........................................................................................................................62DataTypesinPlots.....................................................................................................................63OpeningFiles..............................................................................................................................64OpeningFilesfromUnsupportedLoggers..................................................................................65ModifyingaPlot.........................................................................................................................67SettingSeriesandPlotProperties..............................................................................................69SettingPropertiesforallPlots....................................................................................................70SettingAxisProperties................................................................................................................72SettingSeriesProperties............................................................................................................73AddingaGraphLabel.................................................................................................................75Zooming,Panning,andSmartScalingPlots................................................................................76CopyingaSeriestoanotherPlot................................................................................................77UndoingandRedoingPlotChanges...........................................................................................77SavingProjectFiles.....................................................................................................................78PrintingPlotElements................................................................................................................78ClosingaPlot..............................................................................................................................79ViewingaPlot.............................................................................................................................79ViewingaPieChart.....................................................................................................................79ConvertingUnits.........................................................................................................................81ViewingSubsetStatistics............................................................................................................81OverviewofaGraphSubset.......................................................................................................82SettingSubsetStatisticsProperties............................................................................................83

  • HOBOwareUsersGuide 4

    FilteringaSeries.........................................................................................................................84CroppingaSeries........................................................................................................................85OverlayingSeries........................................................................................................................86MergingDatafiles.......................................................................................................................87RemovingaSeriesfromaPlot....................................................................................................88DataAssistants...........................................................................................................................88BarometricCompensationAssistant..........................................................................................89ConductivityAssistant................................................................................................................91DissolvedOxygenAssistant........................................................................................................94GrainsPerPoundAssistant.........................................................................................................99GrowingDegreeDaysAssistant................................................................................................100kWhAssistant...........................................................................................................................103LinearScalingAssistant.............................................................................................................105PulseScalingAssistant..............................................................................................................107ImportingTextFiles..................................................................................................................108ImportTextFilesRequirements...............................................................................................110ImportingFilesfromHOBOlink.................................................................................................112ExportingData..........................................................................................................................112ExportingTableData................................................................................................................113ExportOptions(MultiFileExport)............................................................................................114ExportingDatainClassic(BoxCarPro)Format.........................................................................114ExportingDetails.......................................................................................................................114CopyingPointsDataintoanotherApplication.........................................................................114FormattingtheDate/TimeColumninExcel.............................................................................115TheBulkExportTool.................................................................................................................115WorkinginSecureMode:21CFRPart11Compliance.............................................................118SecurityModeIssues................................................................................................................119

    Chapter 4: Other Hardware ............................................................ 120WorkingwiththeHOBOU30Station(HOBOwareProOnly)...................................................120ControllingtheRelayontheHOBOU30Station(HOBOwareProOnly)..................................121ChangingtheStateoftheU30RelayContact..........................................................................122TheAlarmsWindowfortheU30/NRC......................................................................................123TheAlarmConfigurationWindowfortheU30/NRC................................................................124ConfiguringFlexSmartModules/AnalogSensorPorts..............................................................125TheConfigureChannelWindow...............................................................................................126SettingVoltageRangesandScalingParameters......................................................................127ConfiguringExcitationPower...................................................................................................129

  • HOBOwareUsersGuide 5

    LoadingaSavedConfiguration.................................................................................................130ResettingaModuletotheDefaultConfiguration....................................................................131WorkingwithaShuttle(HOBOwareProOnly).........................................................................132TheShuttleManagementWindow..........................................................................................132OffloadingandSavingShuttleFiles..........................................................................................134SavingFilesfromaShuttle.......................................................................................................135DeletingFilesfromaShuttle....................................................................................................136ConfiguringLoggerswiththeE50B2Power&EnergyMeter...................................................136WorkingwiththeHOBOU26DissolvedOxygenLogger...........................................................139

    Chapter 5: HOBOnode Manager ...................................................... 140ATourofHOBOnodeManager.................................................................................................141ViewingRealTimePlots...........................................................................................................142DeterminingDataNodeType...................................................................................................143ChangingLoggingandConnectionIntervals.............................................................................144SettingupExternalSensors......................................................................................................145ConfiguringScalingonaSensor................................................................................................146SettingHOBOnodeManagerPreferences................................................................................159WorkingwithHOBOnodeManagerData.................................................................................161EnablingHOBOnodeViewer.....................................................................................................167UsingHOBOnodeViewer..........................................................................................................169SettingupDataDelivery...........................................................................................................170DataStorageLocation..............................................................................................................175Alarms.......................................................................................................................................175TheNetworkMap.....................................................................................................................185Maintenance&Troubleshooting..............................................................................................189

    Chapter 6: Reference ..................................................................... 201Windows...................................................................................................................................201Menus.......................................................................................................................................204SettingPreferences..................................................................................................................214TipsforWorkingwithMultipleLoggers...................................................................................231SettingtheLanguage/FormatonYourComputer....................................................................234Alarm&ReadoutTool..............................................................................................................235

  • HOBOwareUsersGuide 6

    Chapter1:AnOverviewofHOBOwareHOBOwaresoftwareisusedforlaunching,readingout,andplottingdatafromHOBOdataloggersandwirelessHOBOdatanodes.HOBOwareProcombinesfastloggerlaunchandreadoutfunctionswithpowerfuldataplottingcapabilities,makingiteasytoanalyzedatarecordedwithHOBOdataloggersandHOBOwirelessdatanodes.Anintuitiveinterfaceallowsyoutoselectdataparameterstodisplay,formatgraphs,performanalysis,andsaveprojectsforfutureuse.YoucanalsoexportdatatoMicrosoftExcelorotherASCIIcompatibleprogramsforadditionalanalysis.HOBOwareLiteisentrylevelsoftwareforHOBOUSBloggers,offeringeasyloggerlaunch,readout,anddataplotting.ForacomparisonbetweenHOBOwareProandHOBOwareLite,seehttp://www.onsetcomp.com/sites/all/themes/foliage/hoboware_comparison.html.

    GettingStartedIfyouareusingHOBOdataloggers,followtheinstructionsintheGettingStartedGuidetoinstallthesoftware,connectyourlogger,launch,andreadoutdata.YoucanexpecttobecomfortablewithHOBOwarefundamentalswithinafewminutes.Refertothesectionsoutlinedbelow,dependingonwhathardwareyouareusing.

    AllUsers Installation Reference

    DataLoggersIfyouareusingHOBOdataloggers(suchastheU14orH12series),refertothefollowingsections:

    WorkingwithHOBODataLoggers PlottingandAnalyzingData

    WirelessHOBOdatanodesIfyouareusingHOBOwirelessdatanodes,refertotheHOBOnodeManagersection.

    OtherHardwareRefertoOtherHardwareifyouareusingaHOBOU30Station,aHOBOShuttle,orFlexSmartModules/AnalogSensorPorts.

    Notes AnydifferencesforHOBOwareProandHOBOwareLite,forWindowsorMacintosharenoted. ScreencapturesofwindowsareEnglishversions.IfyouhaveaversionofHOBOwareinanother

    language,seethesoftwareforthetranslatedwindows.

  • HOBOwareUsersGuide 7

    IntroductiontoHOBOwareProOverviewWithHOBOwarePro,youcanquicklylaunch,readout,andplotdatarecordedbyHOBOdataloggers.HOBOwareProallowsyoutoanalyzeandviewdatanumerousways,including:

    Filteringdatatoextractkeyseriesinformationsuchasminimum,maximum,oraveragevalues. Scalingdatatosensorunits, Savinggraphsforfutureuse, Singlefileexportingorbulkexporting,and WorkinginSecuremodefor21CFRPart11compliance.

    Inaddition,HOBOwareProincludesHOBOnodeManagerforconfiguringandmaintainingawirelessHOBOdatanodenetwork.SeeNewHOBOwareProFeaturesandAdvancedHOBOwareProFeaturesforadditionaldetails.ToupgradefromHOBOwareLitetoHOBOwarePro,call18005644377andaskforOnsetCustomerService.

    GettingStartedTogetupandrunningquickly,followtheinstructionsintheGettingStartedGuidetoinstallthesoftware,connectyourlogger,launch,andreadoutdata.YoucanexpecttobecomfortablewithHOBOwarefundamentalswithinafewminutes.

    DevicesHOBOwareProiscompatiblewiththefollowingdevices:

    HOBOUSeriesloggers HOBOWeatherStations,includingHOBOU30Station HOBOMicroStations HOBOEnergyLoggerSystems HOBOUShuttleandWaterproofShuttle HOBOFlexSmartModules HOBODataNodes(WirelessSensors) HOBO4ChannelPulseInput(UX120017) EnergyandPowerMeter(TVERE50B2)

    IntroductiontoHOBOwareLiteHOBOwareLiteisentrylevelsoftwareforHOBOUSeriesloggers,offeringthefollowingfeatures:

    Easyloggersetup/launchandreadout Quickcheckofloggerstatusandcurrentreadings Graphingandanalysiswiththeabilitytoviewmultipledatafilesononegraph Filterstoextractkeyseriesinformationsuchasminimum,maximum,oraveragevalues

  • HOBOwareUsersGuide 8

    Theabilitytosavechangestographsinaprojectfile ExportofASCIItextdatatoMicrosoftExcelorotherprogramsforfurtheranalysis

    YoucanopenthefollowingfilesinHOBOwareLite: .hoboincludingfilescreatedwithOnset'sGreenLinesoftwareandbothMacintoshandWindows

    compatibleversionsofHOBOwareLiteandHOBOwarePro. .hsec .hproj

    ToupgradefromHOBOwareLitetoHOBOwarePro,call18005644377andaskforOnsetCustomerService.AnupgradetoHOBOwareProisrequiredforcertaindevices.

    AdvancedHOBOwareProFeaturesSetupAssistantWhenyouopenHOBOwareProforthefirsttime,youaregreetedbytheHOBOwareSetupAssistant,aseriesofwindowstohelpyouquicklychangesomekeypreferencesettingsfordevicetypesandserialports,unitsofmeasurement,anddataassistants.SeeHOBOwareProSetupAssistantformoreinformation.

    DataAnalysisAnalysisfeaturescanbeusedtoextractkeyinformationfromloggeddata.Youcancreatenewdataseries,whichcanbegraphedorexported.Youcanalsofilterdataoveruserspecifiedintervalsthatcanbeinseconds,minutes,hoursordays.TheSubsetStatisticstool(includedwithHOBOwarePro)calculatesstatisticsoverauserdefinedinterval.

    TimeSavingOptionsLaunchingBydefault,HOBOwareProdisplaystheLaunchwindowwhenyouinitiatealaunch.Ifyouwanttoskipthelaunchwindowandlaunchtheloggerwiththedefaultsettings,chooseAutomaticallylaunch.Thisisparticularlyusefulifyouwanttolaunchmanyloggersquicklywithoutchanginganysettings.Formoreinformation,seeLaunchTimeSavingOptions.

    ReadoutThisfeatureallowsyoutoskipanumberofdefaultactionswhenreadingoutalogger(suchasbeingpromptedtosavingorthePlotSetupwindow).Formoreinformation,seeReadoutTimeSavingOptions.

    ViewingSubsetStatisticsTheSubsetStatisticsfeatureallowsyoutoselectarangeofdatainagraph,anddisplaythemaximum,minimum,averageandstandarddeviationforthemeasurementsinthatrange.SeeViewingSubsetStatisticsformoreinformation.

    DataAssistantsHOBOwareProDataAssistantsletyoucreatenewdataseriesbycombiningdatarecordedbytheloggerwithadditionaldatathatyouenterbeforeyoudisplaytheplot.Forexample,thekWhAssistantconvertsloggedpulsedatafromanenergytransducertokWh,averagekW,andenergycost.AllDataAssistantscanberunfromthePlotSetupwindowafteryoureadoutaloggeroropenadatafile.WiththismethodyouneedtoruntheDataAssistanteachtimeyoureadouttheloggeroropenadatafile.

  • HOBOwareUsersGuide 9

    LinearScaling,PulseScaling,andkWhDataAssistantscanalsobeusedatlaunchtimeforsomeloggerstoautomaticallycreateserieseachyoureadouttheloggeroropenadatafile.ThisfeatureisavailablefortheHOBOU30,HOBOH21(WeatherStationandMicroStation),HOBOH22(EnergyLogger),HOBOU12,andHOBOUX120017xloggerswithsupportedsensorsattached.Formoreinformation,seeHOBOwareProDataAssistants.

    ExportDatatoOtherPrograms OneclickexporttoMicrosoftExcelorotherASCIIcompatibleprograms BulkFileExportToolprovideseasywayofexportingmanydatafilesatonce ExportinOnsetBoxCarProformat Internationaldateandtimeformats Copyandpastegraphimagesintootherprograms

    NewHOBOwareFeaturesThefollowingnewfeatureisavailableinHOBOware3.4.1:

    DevicemenuoptionsspecifictoHOBOU30StationsarenowavailablethroughanewManageU30menuchoice.

    ThefollowingnewfeaturesareavailableinHOBOware3.4: SupportforthenewHOBOUX100seriesdataloggers,includinglaunchcapability,prelaunchfilters,alarms,

    burstlogging,statisticslogging(maximum,minimum,average,andstandarddeviation),andtheabilitytostopandrestartloggingduringasingledeployment.

    Piechartsforloggerswithstateseriesthatyoucanview,print,andsaveas.pngfilesforadditionalanalysisbeyondlinegraphs.ThisisespeciallyhelpfulforlightandoccupancydatafromUX90005xand006xloggers.

    AfirmwareupgradetoolthatautomaticallydetectswhenanewfirmwarefileisavailableforUX90andUX100seriesdataloggersandwalksyouthroughtheupdateprocess.

    Improvementstotheprocessofexportingdatapoints,includingtheabilitytoselectindividualseriesoreventtypesandtosortcolumns.

    Anewpreferenceforsettingtheloggerlaunchdescriptionastheserialnumbersothatyoucaneasilydifferentiatefileswhenlaunchingandreadingoutseveralloggersofthesametype.

    TheabilitytoshowsgapsinserieswhenmergingdatafilesorwhenplottingaUX100loggerthathadstoppedandresumedlogging.

    ImprovedtheNetworkMapfeatureinHOBOnodeManager.(ThisrequiresaZWReceiverfirmwareupgrade.)

    SupportforthenewWindows8operatingsystem.ThecurrentlistofsupportedoperatingsystemsisWindows8,Windows7(Pro,Ultimate,andHomePremium),andWindowsXP(ProandHome)onthePCandOSXVersions10.6.x,10.7.x,and10.8.xonMac.

    Note:TheAlarm&ReadoutToolisnolongerinstalledwithHOBOwareasofversion3.4.However,youwillstillhaveaccesstothesoftwareifitwasinstalledwithanolderversionofHOBOware.

  • HOBOwareUsersGuide 10

    NewFeaturesinPreviousReleasesHOBOware3.3.2:

    Supportforthefollowinglanguagesandtheirassociatedformatsinthesoftwareuserinterface:Spanish(Spain),Portuguese(Portugal),andGerman(Germany);HelpanduserguidesareavailableinEnglishonly.

    ImprovedfieldcalibrationaccuracyintheConductivityAssistant.HOBOware3.3.1:

    SupportforthenewHOBOU26DissolvedOxygenlogger,includinglaunchandreadoutcapability,aLabCalibrationtooltocalibratetheloggerto100%and/or0%saturation,andaDissolvedOxygenDataAssistantthatcorrectsformeasurementdriftfromfoulingandgeneratessalinityadjustedDOconcentrationaswellaspercentsaturationdata.

    TheabilitytoupdatefirmwareforreceiversanddatanodesintheHOBOZWWirelessSystem,withtheoptiontoupdateasingledeviceatatimeormultipledatanodesinagroup.

    AdataencodingpreferenceaddedtotheGeneralpreferencesthatcontrolswhetherdatainHOBOwareisimportedandexportedbasedonUTF8oroperatingsystemstandards.

    Frenchlanguagesupportforthesoftwareuserinterface(HelpanduserguidesareavailableinEnglishonly).HOBOware3.3:

    SupportforthenewHOBOUX90seriesdataloggers,includinglaunchcapability,advancedsensorconfiguration,andprelaunchfilters.ThemodelssupportedwiththisreleasearetheState/Pulse/Event/Runtimelogger(UX90001x),LightOn/Offlogger(UX90002x),MotorOn/Offlogger(UX90004x),andOccupancy/Lightlogger(UX90005x/006x).

    HOBOware3.2.2: EnhanceddisplaypreferencesforsortingdataseriesintheStatusandPlotSetupwindows. AutilityforupgradingtheHOBOnodeManagerdatabasetoimproveperformance. PerformanceimprovementsfortheHOBOdatanodesdatadeliveryfeature. AchangetotheFTPoptionfortheHOBOdatanodesdatadeliveryfeaturesothatitusespassivemode,

    whichallowsforbetterconnectionsthroughfirewalls. AnupdatetotheConductivityAssistant,whichincludesrefinementsinthecalculationoftemperature

    compensationandanewoptionfornonlinear,seawatercompensationbasedonPSS78. ArevisiontothecalculationforTCDI520010SandTCDI540020SsensorswhenusedwithUSeries

    loggersensuringdataisdisplayingproperly. SupportforJava7.

    HOBOware3.2.1: TheintegrationoftheAlarm&ReadoutTool,whichisnowautomaticallyinstalledandavailableforusefrom

    theToolsmenu. TheabilityforallnonadministratoruserstorunHOBOwareonWindows(administratorprivilegesare

    requiredtoinstallHOBOware,mapandunmapfileassocations,andloadnewDataAssistants). CompatibilitywithiMacandMacBookProandIntelCorei5andi7processors. ArevisiontothecalculationfortheSSMDSoilMoistureSensorensuringdataisdisplayingaccurately. TheoptiontocreatenewfilesoroverwriteexistingonesviaFTPwhenusingtheDataDeliveryfeaturefor

    HOBODataNodes.ThisallowsyoutoautomaticallyimportwirelessnodeCSVdataintootherapplications,suchasMicrosoftExcel.

  • HOBOwareUsersGuide 11

    HOBOware3.2: Aredesignedlaunchwindowforquickloggerconfigurationandeasysensorsetup. Fasterprocessingtimesforopeninglargedatafiles(512KBandup). Theabilitytoconfigurefilteredserieswhenlaunchingthelogger,whichautomaticallygeneratescustom

    series,suchasaveragetemperatureperday,whenyoureadouttheloggerandplotdata. TheabilitytoconfigurePulseScaling,LinearScaling,andkWhDataAssistantserieswhenlaunchingany

    loggerwithapplicableexternalsensors SupportforthenewHOBO4ChannelPulseInputDataLogger(UX120017x),includingadvancedsensor

    configurationforsettingmaximumpulsefrequencyandlockouttimesasneededinrawpulseandeventchannels.

    SupportforthenewE50B2Power&EnergyMeter(TVERE50B2),includingsinglestepconfigurationwiththeHOBO4ChannelPulseInputDataLogger(UX120017x)andautomaticcalculationofnumerousadditionaldataseriesforanalysis.

    SupportforthenewHOBOConductivityLogger(U24002). Theoptiontodisableloggingthebatterychannelbydefaultonsomeloggers,whichcanextendbatterylife

    andmemoryspace.Note:LoggerslaunchedinHOBOware3.2orlaterwithseriescreatedbythePulseScaling,LinearScaling,andkWhDataAssistantsorwithfilteredseriescannotbereadoutinearlierversionsofHOBOware.

    InstallingHOBOwareYoucaninstallHOBOwarefromtheOnsetWebsiteorfromtheInstallationCD.Note:YoumustbeanadministratortoinstallHOBOware.ThefirsttimeyouinstallHOBOwareitwillbeinTrialMode.Youwillneedtoactivateitenteringyourlicensekey,orcontinuetorunitinTrialModeuntilthetrialexpires.IfyouareupdatinganexistingversionofHOBOwareyouwillnotneedtoactivateHOBOware.SeeActivatingHOBOware(page12).

    BeforeYouBegin READMEfile

    RefertotheHOBOwarewebsiteathttp://www.onsetcomp.com/HOBOwareortheREADMEfileontheinstallationCDfordetailedinformationaboutthehardwareandsoftwareneededtorunHOBOware.Ifyouhaveadditionalquestions,pleasecontactOnsetTechnicalSupport.

    LicenseKeyThelicensekeyislocatedintheemailconfirmationoronlinereceiptyoureceivedwhenpurchasingthesoftware.IfyouhaveaHOBOwareCD,thelicensekeyisprintedonthebackoftheCDcase.AfteryouinstallHOBOwareandrunitforthefirsttimeyouwillbepromptedtoenterthelicensekey.

    InstallingHOBOwarefromtheOnsetWebSiteTodownloadHOBOwarefromtheOnsetWebsite,gotothefollowingURLandfollowtheinstructionsfordownloadingandinstallingHOBOware:http://www.onsetcomp.com/products/softwareWhenyouinstallHOBOwareformthewebitwillbeinTrialmodeuntilyouenteryourLicenseKey.

  • HOBOwareUsersGuide 12

    InstallingHOBOwarefromtheInstallationCDToinstallthesoftwarefromtheCD:

    Windows1. InserttheCDinyourcomputersCDROMdrive.Theinstallationprogramshouldstartautomatically.Ifitdoesnot,navigatetotheCDdriveinMyComputerorWindowsExploreranddoubleclickHOBOware_Setup.exetolaunchtheHOBOwareinstaller.NOTE:HOBOwareforWindowsrequirestheJavaRuntimeEnvironment.IfthecorrectJavaRuntimeEnvironmentversionisnotalreadyinstalledonyourcomputer,theHOBOwareSetupprogramwillaskifyouwanttoinstallitnow.(YoumustbeloggedinasanadministratoronyourcomputertoinstalltheJavaRuntimeEnvironment.)ClickYesandfollowtheprompts.AfterJavaRuntimeEnvironmentisinstalled,theHOBOwareinstallationwillcontinue.2. FollowthepromptstoinstallHOBOware.TostartHOBOware,doubleclicktheHOBOwareicononyourdesktoporselectPrograms>OnsetApplications>HOBOwarefromtheStartmenu.

    Mac1. InserttheCDinyourcomputersCDROMdrive.2. DoubleclicktheCDiconintheFinderanddoubleclicktheHOBOware.pkgtolaunchtheHOBOware

    installer.3. Whenyoureachthefinalscreenoftheinstallationprogram,clickRestart.4. Afterrestarting,openHOBOwarebydoubleclickingtheHOBOwareiconinyourcomputers

    Applicationsfolder.5. ProceedtoActivatingHOBOware.

    SoftwareUpdatesIfyouhaveanInternetconnection,HOBOwarewillperiodicallyaskifyouwouldlikeittochecktheOnsetwebsiteforsoftwareupdates.Thedefaultistocheckonceperweek,butyoucanconfigureHOBOwaretocheckdailyormonthly.GotoPreferences>General>Startup>CheckforHOBOwareupdates.IfyoudonothaveanInternetconnection,chooseNever.YoumaycheckforupdatesmanuallyatanytimebyclickingCheckforUpdatesfromtheHelpmenu.NOTE:IfyouareusingHOBOnodeManagerwithHOBODataNodeWirelessNetwork,youmuststopdevicecommunicationtoupgradeHOBOware.

    UpgradingtoHOBOwareProToupgradefromHOBOwareLitetoHOBOwarePro,call18005644377andaskforOnsetCustomerService.

    ActivatingHOBOwareThefirsttimeyouinstallHOBOwareitwillbeinTrialMode.Youwillneedtoactivateitenteringyourlicensekey,orcontinuetorunitinTrialModeuntilthetrialexpires.

  • HOBOwareUsersGuide 13

    NOTE:IfyouareupdatinganexistingversionofHOBOwareyouwillnotneedtoactivateHOBOware.

    LocatingyourLicenseKeyThelicensekeyislocatedintheemailconfirmationoronlinereceiptyoureceivedwhenpurchasingthesoftware.IfyouhaveaHOBOwareCD,thelicensekeyisprintedonthebackoftheCDcase.

    LicenseKeyManagerWindowThefirsttimeyouopenHOBOwaretheLicenseKeyManagerwindowwillappear.EnteryourlicensekeyintheLicenseKeytextboxtoactivatetheFullVersionofHOBOware.ThisistheonlytimeyouwillhavetoenteryourLicenseKey,evenifyouupgradetoanotherversion.

    HOBOwareSetupAssistant(HOBOwareProOnly)WhenyouopenHOBOwareProforthefirsttime,youaregreetedbytheHOBOwareSetupAssistant,aseriesofwindowsintendedtohelpyouquicklychangesomekeysettingsfordevicetypesandserialports,unitsofmeasurement,anddataassistants.ToruntheSetupAssistant,clickStartandfollowthepromptstochangesomekeypreferences.NOTE:TheSetupAssistantisoptional.IfyouclickCancel,youwillbeabletouseHOBOwareProwithitsdefaultsettings.

    SetupPreferencesDeviceTypesSerialPort(s)touse:

    SupportserialdevicesonCOM1(Windows) Supportserialdevicesonthedefaultserialportonly(Mac)

    ThedefaultisUSB.IfyouenableSerialDevicesyoucanselecttheporttouse.

    UnitSystem US(default) SI

    DataAssistantsSelecttheinstalledDataAssistantsyouwanttoshowwhenrelevantdataispresent.AllloadedDataAssistantsarecheckedbydefault.YoucanalsoloadanewDataAssistant.

  • HOBOwareUsersGuide 14

    RunningtheSetupAssistantataLaterTimeToruntheSetupAssistantatanytime,goto:Windows:File>Preferences>GeneralMac:HOBOware>Preferences>GeneralandclicktheSetupAssistantbuttoninthebottomrightofthepane.

    UsingaTrialVersionofHOBOwareThefirsttimeyouopenHOBOwaretheLicenseKeyManagerwindowwillappear.TorunaTrialVersion,checkIwanttouseHOBOwareProTrialandclickOK.

    ThedateyourtrialwillexpireappearsatthetopoftheHOBOwarewindow.

    EachtimeyouopenaTrialVersiontheLicenseKeyManagerwindowappears.CheckIwanttouseHOBOwareProTrialtocontinueusingtheTrialVersion,or,ifyouhavepurchasedHOBOware,enterthelicensekeyintheLicenseKeytextbox.

    TrialExpirationWhenthetrialexpiresyouwillbepromptedtoenteralicensekeybeforeyoucancontinueusingHOBOware.

  • HOBOwareUsersGuide 15

    ATouroftheHOBOwareInterfaceTheillustrationbelowshowstheHOBOwaremainwindow,includingaplotgeneratedfromopeninganexistingdatafile.

    1. MainMenuFromtheMainMenuyoucanperformoperationsonadeviceoraproject.2. ToolbarThetoolbarincludesiconsforperformingoperationsonadeviceaswellastoolto

    manipulateaplotorgraph.FormoreinformationseeTheToolbaronpage210.3. PointsTableLoggerdatapresentedinatabularformat.SeeThePointsTableonpage18.4. DetailsPaneInformationandstatisticsaboutthelaunch.SeeTheDetailsPaneonpage17.5. PlotAgraphicalrepresentationofthesensorreadingsandevents.SeeOverviewofaPlotonpage

    16.6. ViewsUsetheseiconstotogglebetweenviewswhenyouhavemultipleplotsopen.

  • HOBOwareUsersGuide 16

    7. MessageBarIndicatesdevicesconnectedaswellasvariousactionsinprogress(suchasreadout,launch).

    OverviewofaPlotAplotisavisualrepresentationoftheseriesyouselectedtoviewwhenyouopenedtheplot.Tochangedefaultplotsettings,seePlottingPreferencesonpage222.

    1. TitleTitleofthegraph(defaultisthedescriptionenteredatlaunch).ShoworhidetheTitlefromtheViewmenu.

    Tochangethetitleproperties,doubleclickthetitleorselectthetitleandclickthePropertiesicon .ThefollowingoptionsareavailableontheTitlePropertieswindow:Name:Enteruptoa40charactertitleforthegraph.Bydefault,thenameisthedescriptionenteredwhenlaunchingthelogger.Location:Movethetitletoeitherthetoporbottomofthegraph.Thedefaultlocationisthetopofthegraph.Font:Selectafonttype,size,andappearance.2. AxesThegraphcancontainseveralvalueaxesdependingonthenumberofseriesdisplayed.YaxisAlsocalledthevalueaxes;displaysthevaluesforallsensorseries,suchastemperatureindegreesFahrenheitorCelsius,orbatterylevelinVolts.XaxisAlsocalledthetimeaxes;displaysthetimerangesforthedata.3. Gridlines

  • HOBOwareUsersGuide 17

    ShoworhideverticalandhorizontalgridlinesfromtheViewmenu.4. LegendThelegendistheareathatliststhekeysforeachseries(linecolorsandmarkertypes).Ifthecrosshairtoolisinuse,italsoliststhedate,timeandpointvaluescurrentlyselectedbythecrosshair.ShoworhidetheLegendfromtheViewmenu.Tochangethelegendproperties,doubleclickthelegendorselectthelegendandclickthePropertiesicon

    .ThefollowingoptionsareavailableontheLegendPropertieswindow:Name:Enteruptoa40charactertitleforthelegend.Thereisnodefaultname.Location:Movethelegendtoeithertheleftorrightsideofthegraph.Bydefault,thelegendappearsontheright.ShowBorder:Toggleswhetheraboxisdrawnaroundthelegend.Bydefault,aborderisshown.Forsecurefiles(HOBOwarePro:.hsecand.dsec;HOBOwareLite:.hsec)apadlockicon appearsinthelegendnexttothenamesoftheseriesthatwereverifiedasoriginaldata.

    TheDetailsPaneForeachseries,event,andstate,theDetailsPanecontainsinformationandstatistics,including:

    DevicesDeviceInfosuchasthemodelandserialnumber. DeploymentInfoSuchasseriesname,launchdescription,andlaunchtime.(DoesnotapplytoHOBO

    datanodes) SeriesStatisticsSuchasnumberofsamples,averagereading,andsampletime. Sensorinformation(forstationtypeloggers)Includingthesensorlocation,serialnumber,and

    resolutionInHOBOwarePro,ifyoudonotwantplotstoopenwiththesensorlocationshowninthePointsandDetailspanes,gotoPreferences>Plotting>PointsTableandDetailspaneanddisablethe"IncludesensorLocation(ifavailable)inpointstableanddetailspane"option.TodisplaytheDetailspaneinHOBOwareLite,chooseDetailsPanefromtheViewmenu.ChooseDetailsPaneagaintohidethepane.

    DeploymentinformationSuchasthelaunchdescription,deploymentnumber,starttimeand

    timezone,logginginterval,andbatteryvoltageatlaunch. SeriesstatisticsIncludingthetotalsensorsamplesandevents,timeofthefirstandlastsample,

    andthemaximum,minimum,average,andstandarddeviationforeachsensorseriesintheplot

  • HOBOwareUsersGuide 18

    AudittrailinformationforsecurefilesTofullyexpandthedetailstree,clicktheShowAllbutton.Tofullycollapsethetree,clickHideAll.ClickaseriesnodeinthedetailstreetoselectthecorrespondingseriesonthegraphandcorrespondingcolumninthePointsTable.Toprintthedetailstree,gototheFilemenuandchoosePrintDetails.Toexportthedetailstreetoatextfile,gototheFilemenuandchooseExportDetails.Toresizethepane,dragthedividerbetweentheDetailspaneandthegraph.YoucanhidetheDetailspane.FromtheViewmenu,chooseDetailsPane.ChooseDetailsPaneagaintoredisplaytheDetailspane.NOTE:YoucanconfigureiftheDetailspaneisshownbydefaultwiththeShowthedetailspanewhenplottingdataoptioninPreferences>Plotting>Layout.

    ThePointsTableThePointstablelistsloggerdatainatable.ThePointstableislinkedtothegraph:onlythedataforthesensormeasurements,states,andeventsdisplayedonthegrapharelistedinthePointstable.

    Thedatapointsarelistedchronologically.Allthemeasurements,statechanges,andeventsrecordedbytheloggeratagiventimearelistedinasinglerow.Eachseriesisdisplayedinacolumn.Youcanresizethecolumnsbyclickinganddraggingthedividersbetweenthecolumnheaders.Usethearrowkeysorscrollbarstomoveup,down,left,andrightinthePointstable.Additionally,asyouclickthecrosshairtool inthegraph,thecorrespondingpointisselectedinthePointstable;ifyouclickavalueoreventcellinthePointstable,thecrosshairwillappearonthegraphatthetimecorrespondingtothatcell.Toprintthepoints,gototheFilemenuandchoosePrintPoints(availableonlywithJava1.5orhigher).Tocopypoints,usethemousetoselectthecellsyouwanttocopy,thenpressCtrl+CommandCtocopythemtotheclipboard.Tip:YoucanexportthedatapointstoatextfileforuseinMicrosoftExcelorotherprograms.Toresizethepane,dragthedividerbetweenthePointstableandthegraph.TohidethePointstable,fromtheViewmenu,choosePointsTable.IfyoudonotwantthePointstabletobedisplayedbydefaultwhenyouviewdata,gotoPreferences>Plotting>Layoutanddisable"Showthepointstablewhenplottingdata"option.HOBOwareProIfyoudonotwantplotstoopenwiththesensorlocationshowninthePointsandDetailspanes,disablethe"IncludesensorLocation(ifavailable)inpointstableanddetailspane"settingunderPointsTable&DetailsPaneonthePlottingPreferencepane.

  • HOBOwareUsersGuide 19

    HOBOwareLiteTodisplaythePointstable,choosePointsTablefromtheViewmenu.ChoosePointsTableagaintohidethePointstable.

    ThePlotSetupWindowAfteryoureadoutaloggeroropenasaveddatafile,thePlotSetupwindowappears.

    1. DescriptionEntertexttodescribetheplot.2. SelectSeriestoPlotSelecttheseriesyouwanttoplotandtheunits,ifapplicable.3. SelectInternalLoggerEventstoPlotSelectinternalloggereventstoplot(onlyavailableforsome

    loggers).4. OffsetfromGMTSettheTimeZoneOffsetfromGMTifrequired.5. DataAssistants(HOBOwareProOnly)SeeDataAssistantsonpage88.HOBOwareProDataAssistantsletyoucreatenewdataseriesbycombiningdatarecordedbytheloggerwithadditionaldata.NOTE:ThefollowingDataAssistantscanalsoberunfromtheLaunchwindow:PulseScaling,LinearScaling,kWh.6. PlotButtonClickthisbuttontoplottheselecteddataonagraph.

    HOBOwareSoftwareUpdatesIfyouhaveanInternetconnection,HOBOwarewillperiodicallyaskifyouwouldlikeittochecktheOnsetwebsiteforsoftwareupdates.Thedefaultistocheckonceperweek,butyoucanconfigureHOBOwaretocheckdailyormonthly.GotoPreferences>General>Startup>CheckforHOBOwareupdates.IfyoudonothaveanInternetconnection,chooseNever.YoumaycheckforupdatesmanuallyatanytimebyclickingCheckforUpdatesfromtheHelpmenu.

  • HOBOwareUsersGuide 20

    Chapter2:WorkingwithHOBODataLoggersWorkingwithHOBOdataloggersinvolvesthefollowingbasicsteps:

    1. Connecttheloggertothecomputerfortheinitialsetupandlaunch.2. Launchthelogger.Enterorselecttheappropriateparametersfortheloggerdeploymentandthen

    launch,orstart,thelogger.3. Checkthestatusofthelogger.Youhavetheoptionofcheckingthecurrentstatusoftheloggerand

    anycurrentreadingswhileitisstillconnectedtothecomputer.Thiscanbehelpfultoverifythatthelaunchconfigurationisasexpected.

    4. Readoutthelogger.Afterdeployingthelogger,readoutallrecordeddataandsaveittoafile.Manydevicesalsoworkwithshuttlesthatallowyoutokeeptheloggerinthefield,butstillreadoutanyrecordeddata.Youcanthenbringtheshuttlebacktotheoffice,connectittothecomputer,offloadandsavethefiles.

    5. Plotthedata.Afterreadingouttheloggerandsavingthedata,youcanselectanddefinethedataseriesyouwishtoplotinagraph.HOBOwareProalsoprovidesDataAssistantsforautomaticscalingofcertaindatawhenplotting.

    6. Analyzethedataandcustomizetheplot.Therearenumeroustoolsavailableforworkingwiththedataandchangingtheplot,suchasexporting,filtering,mergingfiles,cropping,andmore.

    7. Savechangestotheplotasaprojectfile.Changestotheaxis,series,plot,andlegendpropertiesaswellasanyfilteringormergingcanbesavedasa.hprojfile.

    NOTE:IfyouareconfiguringHOBOdatanodes,seethesummaryoftasksonpage1forsettingupdatanodes.

    ConnectingDevicesMostHOBOdataloggersandshuttlesconnecttothecomputerwithanOnsetsuppliedUSBcable.TheHOBOWeatherStation,HOBOMicroStation,andHOBOEnergyLoggerconnecttothecomputerwithaserialcable(ortoaUSBportwithaKeyspanserialadapter).SomeloggersrequireanopticUSBbasestation/couplertoconnecttothecomputer.Consultthemanualthatcamewithyourdeviceforspecificinformationaboutrequiredbasestations/couplersorcables.Refertooneofthesetopicsfordetailsaboutconnectingyourdevicetothecomputer:

    ConnectingaUSBDevice(page21) ConnectingaBaseStation/Coupler(page21) ConnectingaSerialDevice(HOBOwareProOnly)(page22) ConnectingMultipleDevices(page22)

    NOTES: ForinstructionsonconnectingHOBOwarecompatibleshuttlesorusingtheHOBOWaterproofshuttle

    asabasestation,refertotheshuttle'suserguide. HOBOwareLitecanbeusedwithmostHOBOUSeriesDataLoggers,whichcommunicatewiththe

    computerbyconnectingtoaUSBport. FordetailsonworkingwithHOBODataNodes,seeSystemOverview.

  • HOBOwareUsersGuide 21

    ConnectingaUSBDeviceYoucanconnectmostdevicestotheUSBportbyusinganOnsetsuppliedUSBinterfacecable.SomeloggersrequireanopticUSBbasestation/couplerorserialcable(HOBOwareProonly)toconnecttothecomputer.Consultthemanualthatcamewithyourdeviceforspecificinformationaboutrequiredbasestations/couplersorcables.NOTE(HOBOwarePro):Forinstructionsonconnectingshuttles,orusingtheHOBOWaterproofShuttleasabasestation,refertotheshuttlesuserguide.NOTE:IfyouareusingaBaseStationorCoupler,seeConnectingaBaseStationorCoupler(page21).

    StepsToconnectaloggerorshuttletoHOBOwareusingaUSBcable:

    1. OpenHOBOware.2. PlugthelargeendoftheUSBinterfacecableintoaUSBportonthecomputer.3. PlugthesmallendoftheUSBinterfacecableintotheportonthedevice.(Refertothediagramand

    instructionsthatcamewiththedeviceifyouneedhelpfindingtheport.)WhenthedeviceisrecognizedbyHOBOware,therightsideofthestatusbaratthebottomoftheHOBOwarewindowwillupdatetoreflectthenumberofdevicesconnected.Atthispoint,youcanbeginusingthedevice.NOTES(WindowsOnly):

    Ifthedevicehasneverbeenconnectedtothiscomputerbefore,itmaytakesometimeforthecomputertodetectthenewhardwareandreportthatithasconnectedsuccessfully.Oneormoremessageswillappear,indicatingthatnewhardwarehasbeenfound.Youmayalsohearachime.

    Yourcomputermaytellyoutorebootbeforeyoucanusethedevice.Itisnotnecessarytoreboot.DisconnectingaDeviceTodisconnectthedevice,simplyunplugitfromtheUSBcable.Waitforthestatusbartoupdatethenumberofdevicesbeforecontinuing.

    ConnectingaBaseStationorCouplerSomeloggersrequireanopticUSBbasestation/couplerorserialcabletoconnecttothecomputer.Consultthemanualthatcamewithyourdeviceforspecificinformationaboutrequiredbasestations/couplersorcables.NOTES(WindowsOnly):

    Ifyouareconnectingabasestation/couplerorloggerthathasneverbeenconnectedtothiscomputerbefore,itmaytakesometimeforthecomputertodetectthenewhardwareandreportthatithasconnectedsuccessfully.Oneormoremessageswillappear,indicatingthatnewhardwarehasbeenfound.Youmayalsohearachime.

    Youmaybepromptedtorebootyoucomputerbeforeyoucanusethelogger.Itisnotnecessarytoreboot.

    StepsToconnectaloggertoabasestation/coupler:

    1. OpenHOBOware.

  • HOBOwareUsersGuide 22

    2. Plugthebasestation/couplercableintoaUSBportonthecomputer.Important:Makesurethebase/stationcoupleristhecorrectmodelfortheloggeryouwanttouse.Consultthemanualthatcamewiththeloggerifyouarenotsure.3. Attachtheloggertothebasestation/couplerasdescribedinthedocumentationthatcamewithyour

    logger.WhentheloggerisrecognizedbyHOBOware,therightsideofthestatusbaratthebottomoftheHOBOwarewindowwillupdatetoreflectthenumberofloggersconnected.Atthispoint,youcanbeginusingthelogger.Todisconnectalogger,simplyunplugitfromthebasestation/coupler.Waitforthestatusbartoupdatethenumberofdevicesbeforecontinuing.Toattachanotherloggerviaabasestation/coupler,removethelogger,leavingthebasestation/couplerconnected.Thenconnectthenextlogger.

    ConnectingaSerialDevice(HOBOwareProOnly)Toconnectadevicetoacomputerusingaserialcable:

    1. OpenHOBOwarePro.2. Plugthe9pinendoftheserialinterfacecableintoaserialportonthecomputerorKeyspanadapter.3. Plugtheotherendoftheserialinterfacecableintothecommunicationsportonthedevice.(Referto

    thediagramandinstructionsthatcamewiththedeviceifyouneedhelpfindingtheport.)

    4. ClicktheSelectDeviceicon onthetoolbartoensurethatthedeviceislistedintheSelectDevicewindow.Clickthebuttonnexttothedeviceinthewindow,thenclickOK.

    5. ConfigureSerialPortSettings(Optional)IfyouareusingaserialportotherthanCOM1(PC)orDefault(Macintosh),youwillneedtosetupHOBOwaretolookatanotherport.TochangetheSerialPortssettinggoto:PC:File>Preferences>CommunicationsMac:HOBOwarePro>Preferences>CommunicationsNotethatcheckingmultipleserialportscantakesometime,evenwhennodevicesareattached.

    ThestatusbaratthebottomoftheHOBOwareProwindowwillupdatetoreflectthatthedeviceisconnectedandselected.Atthispoint,youcanbeginusingthedevice.Todisconnectthedevice,simplyunplugitfromtheserialcable.

    ConnectingMultipleDevicesOverviewIfyourcomputerhasmultipleUSBandserialports,youcanconnectmultipledevicesandworkwithoneatatime(HOBOwareLitecanonlybeusedforUSBPorts).Whenmultipledevicesareconnected,theSelectDevicewindowopenseverytimeyouselectLaunch,Readout,Stop,orStatusfromtheMainMenu.Clickthebuttonnexttoadeviceonthelisttoselectit,thenclickOK.Theselectedactionwillproceed.ToaccesstheSelectDevicewindowatanytime,clicktheSelectDeviceicononthetoolbar,orchooseSelectDevicefromtheDevicemenu.

  • HOBOwareUsersGuide 23

    WithHOBOwarePro,ifyouworkwithmultipleUSeriesdevicesfrequently,youcansettheDefaultActiontoLaunch(Devicemenu)andenableLaunchTimeSavingOptionstoworkmoreefficiently.NOTE(HOBOwarePro):IfserialdevicesareselectedinPreferences,itmaytakesometimeforHOBOwareProtoscanalloftheserialports.Youcanselectthedeviceyouwantassoonasitisdisplayed.Youdonothavetowaitforalloftheportstobescanned.

    VerifyingSelectedDeviceIftheBlinkDeviceLightbuttonisavailable,youcanclickittoverifythatyouhaveselectedtheappropriatedevice.Thisbrieflyilluminatesthelightonthedevicesthatyouhaveselected.Ifthedevicedoesnothavealight,orifthelightisnotvisiblewhenthedevicesisinthebasestation/coupler,simplychecktheserialnumber(S/N)onyourdevicestomakesurethatitmatchestheonethatisselectedinthiswindow.WhentheSelectDevicewindowisnotopen,youcancheckthestatusbaratthebottomofthescreentoquicklydeterminewhichdeviceisselected.Theleftsideofthestatusbarindicatesthemodel,description,andserialnumberofthedevice;therightsideindicatesthetotalnumberofdevicesconnected.NOTE(HOBOwarePro):SerialdevicesthatyouhaveaddedorremovedarenotreflectedinthedevicecountuntilyouclicktheSelectDeviceicononthetoolbartoupdatethedevicelist.

    LaunchingDevicesTosetupaloggertorecorddata,youmustspecifyseveralparametersandlaunch,orstart,thelogger.Tip:Checkthatthetimeonthecomputeriscorrectbeforeconfiguringalaunch.Ifitisnot,closeHOBOware,updateyourcomputer'sclockmanuallyorsynchronizeitwithanonlinetimeserver(ifavailable)andthenreopenHOBOware.Tolaunchalogger:

    1. Connecttheloggertothecomputer.2. ClicktheLaunch icononthetoolbar,orchooseLaunchfromtheDevicemenu.3. Differentmessagesmayappeardependingonthestateofthelogger.Answereachoftheprompts

    accordingly:Awarningappearsiftheloggerhasalreadybeenlaunched.ClickYestocontinue,orclickNotocancel.ForUSeriesloggersonly(otherthantheHOBOU30Station),awarningappearsiftheloggerhasrecordeddatasinceitslastreadout.ClickYestoproceedtothelaunchwindow,orclickNotocancelthelaunchandreadoutthedatafirst.Becauseloggersaretestedbeforeshipping,theremaybedatainyourloggerthefirsttimeyoulaunchit.Inthiscase,itisnotnecessarytoreadoutthelogger.ClickYestoproceed.4. ConfigurethelaunchparametersintheLaunchLoggerwindow.5. Toverifythatthesensorsareworkingcorrectlyortoreviewsettingsfromthepreviouslaunch,click

    theStatusbutton.6. ClicktheStartbuttonwhenyouarefinishedsettingthelaunchparameters(notethatthistextonthis

    buttonchangesdependingonwhenyouchoseloggingtobegin).HOBOwaredisplaystheprogressofthelaunchandwarnsyounottounplugtheloggerwhileitisbeingconfigured.

    Oncetheloggerbeginslogging,itwillcontinuelogginguntilyoustopit,thememoryisfull,orthebatteryrunsout.

  • HOBOwareUsersGuide 24

    LaunchLoggerWindowUsetheLaunchLoggerwindowtosetupyourloggertorecorddata.TheLaunchLoggerwindowisdividedintothefollowingthreepanes:

    LoggerInformation:Thenameoftheloggercurrentlyselectedappearsatthetopofthispane,whichalsoincludestheserialnumber,deploymentnumber,andcurrentbatterylevel.UsetheDescriptionfieldtotypeuptoa40characternamefortheloggerdeployment.Thedescriptionyouenterherewillbeusedasthedefaultfilenamewhenyoureadouttheloggerandsavethedata.Itwillalsobethedefaulttitleontheplot.ClicktheStatusbuttoninthispanetoseethecurrentstatusoftheloggerandthesettingsusedonthepreviouslaunch.AUserNotesbuttonmayalsoappearforsomeloggers.Clickittoentermoreextensivenotesaboutthedeployment.

    Sensors:Thispanedisplaysalistofthesensorsavailableforthelogger.Choosethesensors,orchannels,thatyouwishtologinthisdeployment,andselectanyexternalsensorsyoumaybeusing.Thispanealsodisplaysanyutilitiesavailableforyourlogger,suchasAlarms,Scaling,andFilters.

    Deployment:Usethispanetoselectthelogginginterval(therateatwhichyouwanttheloggertorecorddata)andtochoosewhentheloggershouldbeginrecordingdata.Thispanealsodisplaystheloggingduration,whichistheapproximatetimeitwilltaketofilltheloggermemorybasedonthelogginginterval,sensors,andothersettingscurrentlyselected.

    Forsomeloggers,youcanalsosetasamplinginterval,choosewhentostoplogging,andselectotheroptionsinthispane.

    Whenyouaredonechoosingthelaunchsettings,clickthebuttoninthelowerrighthandcornerofthewindowtosendthesettingstothelogger.Notethatthebuttontextchangesbasedonwhenyouchosetostartlogging.HOBOwareProTip:Ifyouwillbeusingthesamelaunchsettingsformultipledeploymentsofthesameloggertype,selecttheSkipLaunchWindowNextTimecheckboxtobypassthiswindowthenexttimeyoulaunchadevice.ThiswillcausethenextloggertobelaunchedwitheitherthepreviouslaunchsettingsorthecurrentloggersettingsassetinLaunchTimeSavingOptionspreferences.TheoptionsavailableintheLoggerLaunchwindowvarydependingonthetypeofloggeryouareusing.Formoredetailsaboutlaunchingyourparticulardevice,refertothefollowingtopics:

    LaunchOptionsfortheHOBOUX90SeriesLoggers LaunchOptionsfortheHOBOUX100SeriesLoggers LaunchOptionsfortheHOBO4ChannelPulseInputDataLogger(UX120017x) LaunchOptionsforUSeriesLoggers LaunchOptionsforStationLoggers

  • HOBOwareUsersGuide 25

    LaunchOptionsforHOBOUX90SeriesLoggersThefollowinglaunchoptionsareavailableforHOBOUX90seriesdataloggers.NotethattheLaunchLoggerwindowmayvaryfromtheexampleshownbelow.

    LoggerInformation

    Description:Enteradescriptionforthelaunch,upto40characters,inthisfield.Thisdescriptionisusedasthedefaultfilenamewhenyoureadoutandsavethedatarecordedbythelogger.Itisalsothedefaulttitleoftheplot.

    SerialNumber:Thisistheserialnumberforthelogger.Notethatsomeexternalsensorsordevicesalsohaveserialnumbers,whicharenotlistedhere.

    DeploymentNumber:Thisisthenumberoftimes(includingthistime)theloggerhasbeenlaunched.Eachtimeyoustartanewlaunch,thedeploymentnumberincreasesbyone.

    BatteryLevel:Thisshowsthecurrentbatterylevelinthelogger. Status:ClicktheStatusbuttontochecktheloggersettingsfromthepreviouslaunch.

    SensorsTheSensorsListdisplaysallinternalandexternalsensorsavailableforrecordingdata.Choosethesensors,orchannels,thatyouwishtologinthisdeployment.Toconfigureasensor:

    1. Selectthesensor(s)youwishtolog.2. Choosethemeasurementtolog(e.g.State,Runtime,etc.)fromthedropdownlist.3. Completeadditionaldetailsforthesensor.Thiswillvarydependingonwhattheloggerwillbemeasuring.

    Formostsensors,typeanameandselectadescriptionpairfromthedropdownlist.ForexternalPulsechannels,clickthebuttonforthecorrespondingchannelinthesensorlisttoselectthesensortypeasshownbelow.IfyouselectRawPulseforthesensortype,youcanuseDataAssistantstosetupkWhandpulsescalinginformationasnecessary(HOBOwareProonly),whichwillalsodisplayonthelogger

  • HOBOwareUsersGuide 26

    LCDscreen.NotethatsensorsthatsupportDataAssistantsatlaunchtimedisplaythekWhand/orscalingiconnexttotheirname.

    4. Repeatsteps1through3ifyourloggersupportsmultiplesensors.5. Settheunitsthatdisplayonthelogger'sLCDscreen.WhenthesensorisconfiguredtologStateor

    Runtime,chooseeither: Timetoshowthetotalamountoftimetheswitchhasbeenclosedoronsinceloggingbegan,

    rangingfromsecondstodays;or %toshowthepercentageoftimetheswitchhasbeenclosedoronsinceloggingbegan.WhenthesensorisconfiguredtologPulseorEvent,theLCDwilldisplaytheunitsinthreecharacters.Accepteitherthedefaultunitsortypeyourownthreecharacterunits.

    Notesonconfiguringsensors: ClicktheFiltersbuttontocreateadditionalfilteredseriesforanyofthechannelsyouconfigured.The

    filteredserieswillthenbeautomaticallycalculatedwhenyoureadouttheloggerandplotthedata. ClicktheAdvancedbuttontoaccessthefollowingsettingsspecifictothelogger:PulseFrequencyand

    LockoutTime,Occupancy,andCalibration. Whensettingupapulsesensor,considersettingupthelaunchforadelayedorpushbuttonstart,and

    thenconnectthesensors/deviceslater,beforeloggingbegins.Important:Sensorsthatareconnectedafterloggingbeginsmaynotlogaccuratedata.

    TheOccupancy/Lightlogger(UX90005x/006x)alsoautomaticallygeneratesadditionalseriesuponreadingouttheloggertogiveyoucombinedstatistics,suchashowlongthelightswereleftonwhiletheroomwasunoccupied.TheseserieswillbeavailableinthePlotSetupdialogwhenopeningthedatafile.

    Youhavetheoptiontorecordthelogger'sbatteryvoltageateachsampletime;thisisthelastchannelinthesensorslist.Liketheotherdatachannels,loggingtheinternalbatterychannelconsumessomeofthelogger'smemory.Unlessyoususpectabnormalbatteryperformance,youdonotneedtologthebatteryvoltage.Thisloggerwillrecordabadbatteryeventevenifitisnotloggingthebatterychannel.YoucanalsohidethebatterychannelwithSeriespreferences(inPreferences,selectDisplay,thenSeries).

    Deployment LoggingInterval.Selecthowoftentheloggerwillrecorddata(onlyavailableforsensorsconfiguredtolog

    PulseorRuntime).Youcanchooseeitheroneofthepresetloggingintervalsorspecifyacustomlogginginterval.Theminimumloggingintervalisonesecondandthemaximumformostloggersis18hours,12minutes,and15seconds.Theshorterthelogginginterval,themorequicklymemoryfillsandbatterypowerisconsumed.

    LoggingDuration.Thisliststheapproximatetimeitwilltaketofilltheloggermemorybasedontheloggingintervalandsensorscurrentlyselected.Thisisatheoreticalestimateonly;batterylifeandfrequentstateandeventloggingwillaffectyourdeployment.ThisestimateisonlyavailablewhenthesensorsareconfiguredtologPulseorRuntime.ForestimateswithStateandEventlogging,refertotheloggerdocumentation,alsoavailableathttp://www.onsetcomp.com/support/manuals.

  • HOBOwareUsersGuide 27

    StartLogging.Selectwhentolaunchthelogger.Thisdefaultstothesettingforthelogger'spreviouslaunch.SeeLaunchTimeSavingOptions(HOBOwarePro).Youcanchoosetolaunchthislogger:

    Now.LoggingbeginsassoonasyouclicktheStartbuttonintheLaunchLoggerwindow. AtInterval.Loggingwillbeginatanexactinterval(forexample9:00:00ratherthan8:47:00when

    youchooseaonehourlogginginterval).Theexactstarttimedependsontheloggingintervalyouchoose.ThisoptionisonlyavailablewhenthesensorsareconfiguredtologPulseorRuntime.

    PushButton.LoggingwillnotstartuntilyoupresstheStart/Stopbuttonontheloggerandholditdownforatleastthreeseconds.TheLCDscreenontheloggerwilldisplay"Start"untilyoupressthebutton.

    OnDate/Time.Loggingwillbeginatadateandtimeyouspecify,uptoapproximatelysixmonthsfromthepresent.TheLCDscreenwillcountdowntothatstartdate/timeandthenloggingwillbegin.

    Stoplogging.Selectwhenyouwanttheloggertostoplogging.Youcanchoose: WhenmemoryfillsorNever(wrapping).Ifyouselect"Whenmemoryfills,"thentheloggerwill

    stoprecordingdataoncethememoryisfull.Ifyouselect"Never(wrapping),"theloggerwillrecorddatacontinuouslyuntileithertheloggerbatteryrunsoutoryoustopit.Oncetheloggerisfull,thenewestdatawilloverwritetheoldestdata.Whenwrapping,thememorysegmentontheloggerLCDscreenwillblink.

    Pushbutton.Whenthisoptionisselected,theloggerwillstoprecordingdatawhenyoupresstheStart/Stopbuttonontheloggeritselffor3seconds.TheLCDscreenontheloggerwilldisplay"Stop"whenthisoptionhasbeenselected.

    Specificstopdate.Selectthedateyouwanttheloggertostoprecordingdata.Chooseeitherapresettimeorsetyourowncustomdateandtime.

    Options.Select"TurnLCDoff"ifyouwanttheloggertooperatein"stealthmode"withtheLCDscreenturnedoff.YoucanoverridethistemporarilybypressingtheStart/Stopbuttononthelogger.TheLCDwillthenremainilluminatedfor10minutes.

    Skiplaunchwindownexttime(HOBOwarePro)CheckthisboxifyouwouldliketobypasstheLaunchWindowthenexttimeyouchooseLaunchfromtheDevicemenuorclicktheLaunchicon.ThiswillcausethenextloggertobelaunchedwitheitherthepreviouslaunchsettingsorthecurrentloggersettingsassetinLaunchTimeSavingOptionsintheGeneraltabofpreferences.

  • HOBOwareUsersGuide 28

    LaunchOptionsforHOBOUX100SeriesLoggersThefollowinglaunchoptionsareavailableforHOBOUX100seriesdataloggers.NotethattheLaunchLoggerwindowmayvaryfromtheexampleshownbelow.

    LoggerInformation

    Description.Enteradescriptionforthelaunch,upto40characters,inthisfield.Thisdescriptionisusedasthedefaultfilenamewhenyoureadoutandsavethedatarecordedbythelogger.Itisalsothedefaulttitleoftheplot.

    SerialNumber.Thisistheserialnumberforthelogger.Notethatsomeexternalsensorsordevicesalsohaveserialnumbers,whicharenotlistedhere.

    DeploymentNumber.Thisisthenumberoftimes(includingthistime)theloggerhasbeenlaunched.Eachtimeyoustartanewlaunch,thedeploymentnumberincreasesbyone.

    BatteryLevel.Thisshowsthecurrentbatterylevelinthelogger. Status.ClicktheStatusbuttontochecktheloggersettingsfromthepreviouslaunch.

    SensorsTheSensorsListdisplaysallsensorsavailableforrecordingdata.Choosethesensors,orchannels,thatyouwishtologinthisdeployment.Ifconnectingathermocouple,selecttheappropriatetypefromthedropdownlist.Notes:

    ClicktheAlarmsbuttontosetanalarmforthislogger.NotethattheLoggingModemustbesettoNormalorStatisticstoconfigureanalarm.

    ClicktheFiltersbuttontocreateadditionalfilteredseriesforanyofthechannelsyouconfigured.Thefilteredserieswillthenbeautomaticallycalculatedwhenyoureadouttheloggerandplotthedata.

  • HOBOwareUsersGuide 29

    Youhavetheoptiontorecordthelogger'sbatteryvoltageateachsampletime;thisisthelastchannelinthesensorslist.Liketheotherdatachannels,loggingtheinternalbatterychannelconsumessomeofthelogger'smemory.Unlessyoususpectabnormalbatteryperformance,youdonotneedtologthebatteryvoltage.Thisloggerwillrecordabadbatteryeventevenifitisnotloggingthebatterychannel.YoucanalsohidethebatterychannelwithSeriespreferences(inPreferences,selectDisplay,thenSeries).

    Deployment LoggingInterval.Selecthowoftentheloggerwillrecorddata.Youcanchooseeitheroneofthe

    presetloggingintervalsorspecifyacustomlogginginterval.Theminimumloggingintervalisonesecondandthemaximumformostloggersis18hours,12minutes,and15seconds.Theshorterthelogginginterval,themorequicklymemoryfillsandbatterypowerisconsumed.

    LoggingMode.Selectthetypeofloggingmodeyouwishtouse:Normal,BurstLogging,orStatistics.WithBurstLoggingmode,youcanconfiguretheloggertouseadifferentloggingintervalwhenspecificconditionsaremet.WithStatisticsmode,youcanconfiguretheloggertocalculatemaximum,minimum,average,andstandarddeviationforallenabledsensors(exceptbatteryvoltage)duringloggingatasamplingintervalyouspecify.Keepinmindthatthemorestatisticsyourecord,theshortertheloggerdurationandthemorememoryisrequired.Onceyoulaunchthelogger,theselectedstatisticswillbedisplayedontheloggerLCD.AfteryousetupBurstorStatisticslogging,therewillbeanEditbuttonnexttoLoggingModeintheLaunchLoggerwindowtomakeadditionalchangesasnecessary.NotethatNormalorStatisticsmodeisrequiredifyouwanttosetupalarmsforthelogger.

    LoggingDuration.Thisliststheapproximatetimeitwilltaketofilltheloggermemorybasedonthelogginginterval,loggingmode,andsensorscurrentlyselected.Thisisatheoreticalestimateonly;batterylifeandotherfactorswillalsoaffectthedeployment.

    StartLogging.Selectwhentolaunchthelogger.Thisdefaultstothesettingforthelogger'spreviouslaunch(youcanchangethisinHOBOwareProwiththeLaunchTimeSavingOptionssubcategoryoftheGeneralpreferences).Youcanchoosetolaunchthislogger: Now.LoggingbeginsassoonasyouclicktheStartbuttonintheLaunchLoggerwindow. AtInterval.Loggingwillbeginatanexactinterval(forexample9:00:00ratherthan8:47:00when

    youchooseaonehourlogginginterval).Theexactstarttimedependsontheloggingintervalyouchoose.

    OnDate/Time.Loggingwillbeginatadateandtimeyouspecify,uptoapproximatelysixmonthsfromthepresent.TheLCDscreenwillcountdowntothatstartdate/timeandthenloggingwillbegin.

    PushButton.LoggingwillnotstartuntilyoupresstheStart/Stopbuttonontheloggerandholditdownforatleastthreeseconds.TheLCDscreenontheloggerwilldisplay"Start"untilyoupressthebutton.

    StopLogging.Selectwhenyouwanttheloggertostoplogging.Youcanchoose: WhenmemoryfillsorNever(wrapping).Ifyouselect"Whenmemoryfills,"thentheloggerwill

    stoprecordingdataoncethememoryisfull.Ifyouselect"Never(wrapping),"theloggerwillrecorddatacontinuouslyuntileithertheloggerbatteryrunsoutoryoustopit.Oncetheloggerisfull,thenewestdatawilloverwritetheoldestdata.When"Never(wrapping)"isselected,awrapindicatoriconwilldisplayontheloggerLCDscreen.Note:YoucannotselectNever(wrapping)ifBurstisselectedfortheLoggingMode.

    Pushbutton.Whenthisoptionisselected,theloggerwillstoprecordingdatawhenyoupresstheStart/Stopbuttonontheloggerforthreeseconds.TheLCDscreenontheloggerwilldisplay"Stop"whenthisoptionhasbeenselected.

  • HOBOwareUsersGuide 30

    Resumeloggingonnextbuttonpush.Thissettingisavailablewhen"Pushbutton"isselectedasa"StopLogging"option.Ifyouselectthisoption,youcanresumeloggingonastoppedloggerbypressingtheStart/Stopbuttonforthreeseconds.Forexample,presstheStart/Stoploggingbuttonontheloggeroncetostoplogging.Then,afewminutes,hours,orevendayslaterwhenyouarereadytocontinuelogging,presstheStart/Stopbuttonontheloggeragainandloggingwillresumeusingthesameloggingintervalfromthelastlaunch.Youcanthencontinuetostartandstoploggingasoftenasyou'dlikeduringthissamedeployment.Anygapsbetweenwhenyoustoppedandrestartedloggingduringthisdeploymentwillbereflectedontheplotteddatawhenyoureadoutthelogger.OnceyourelaunchtheloggerwithHOBOwareortheUShuttle,anewdeploymentandnewdatafilewillbegin.Thedatafromthepreviousdeploymentwillnotbecarriedovertothenewone.

    Specificstopdate.Selectthedateyouwanttheloggertostoprecordingdata.Chooseeitherapresettimeorsetyourowncustomdateandtime.Note:Ifyouselectaspecificstopdateandalsohavetheloggerconfiguredtoresumeloggingonthenextbuttonpush,thentheloggerwillstoploggingatthedateyouselectregardlessofhowmanytimesyoustopandrestarttheloggerwiththeStart/Stopbutton.

    Options.Select"TurnLCDoff"ifyouwanttheloggertooperatein"stealthmode"withtheLCDscreenturnedoff.YoucanoverridethistemporarilybypressingtheStart/Stopbuttononthelogger.TheLCDwillthenremainilluminatedfor10minutes.

    Skiplaunchwindownexttime(HOBOwarePro)CheckthisboxtobypasstheLaunchWindowthenexttimeyouchooseLaunchfromtheDevicemenuorclicktheLaunchicon.ThiswillcausethenextloggertobelaunchedwitheitherthepreviouslaunchsettingsorthecurrentloggersettingsassetintheLaunchTimeSavingOptionssubcategoryoftheGeneralpreferences.Important:AfteryouchoosethelaunchsettingsinthiswindowandclickStart,thesettingsarethenloadedintothelogger.TheLCDscreenontheloggerwilldisplayLOADduringthisprocess.IfyoudisconnecttheloggerfromtheUSBcablebeforethisprocessisfinished,ErrwillappearontheLCDscreeninstead.IfyouseeErratanypointduringlaunchconfiguration,checktheUSBconnectionbetweenthecomputerandthelogger,reopentheLaunchLoggerwindowandclickStartagain.

  • HOBOwareUsersGuide 31

    LaunchOptionsfortheHOBO4ChannelPulseInputDataLogger(UX120017x)ThefollowinglaunchoptionsareavailablefortheHOBO4ChannelPulseInputDataLogger(UX120017x).NotethattheLaunchLoggerwindowmayvaryfromtheexampleshownbelow.

    LoggerInformation

    Description.Enteradescriptionforthelaunch,upto40characters,inthisfield.Thisdescriptionisusedasthedefaultfilenamewhenyoureadoutandsavethedatarecordedbythelogger.Itisalsothedefaulttitleoftheplot.

    SerialNumber.Thisistheserialnumberforthelogger.Notethatsomeexternalsensorsordevicesalsohaveserialnumbers,whicharenotlistedhere.

    DeploymentNumber.Thisisthenumberoftimes(includingthistime)theloggerhasbeenlaunched.Eachtimeyoustartanewlaunch,thedeploymentnumberincreasesbyone.

    BatteryLevel.Thisshowsthecurrentbatterylevelinthelogger. Status.ClicktheStatusbuttontochecktheloggersettingsfromthepreviouslaunch.

    SensorsTheSensorsListdisplaysallinternalsensorsandexternalchannelsavailableforrecordingdata.Choosethesensors,orchannels,thatyouwishtologinthisdeployment.Toconfigureachannel:

    1. IfyouareusingtheloggerwithanE50B2Power&EnergyMeter,thenselectforE50B2Power&EnergyMeterintheConfigureSensorsdropdownlistatthetopoftheSensorspane.Thiswillautomaticallypreselecttheappropriatechannelsandpulsescalingfactors,whichyoucanthencustomizefurtherinthefollowingsteps.

  • HOBOwareUsersGuide 32

    Forallothersensors/devices,selectManuallyintheConfigureSensorsdropdownlist.2. Makesurethecheckboxisenabledfortheexternalchannelyouwishtolog.3. ChooseeitherPulse,State,Event,orRuntimefromthedropdownlist.4. ForPulsechannels,clickthebuttonforthecorrespondingchannelinthesensorlisttoselectthe

    sensortypeasshownbelow.IfyouselectRawPulseforthesensortype,youcanuseDataAssistantstosetupkWhandpulsescalinginformationasnecessary(HOBOwareProonly.YoucanalsoadjustmaximumpulsefrequencyandlockouttimebyclickingtheAdvancedbutton.NotethatsensorsthatsupportDataAssistantsatlaunchtimedisplaytheScalingand/orkWhiconnexttotheirname.

    ForStatechannels,typeanameandselecttheOpenandClosedstatedescriptions.ForEventchannels,typeaname,adjusttheincrement,andtypeaunitdescription(oracceptthedefault"units"forthedescription).ForRuntimechannels,typeanameandselectthe%ortimeOff/Ondescriptions.

    5. Repeatsteps1through4foreachchannelyouwishtoconfigure.Notesonconfiguringsensors:

    Althoughitishelpfultoseetheconnectedsensors/deviceswhensettingupthelaunch,itisnotrequiredthatyouphysicallyconnectthemwhileselectingsensortypeoptions.Youmaysetupthelaunchforadelayedorpushbuttonstart,thenconnectthesensors/deviceslater,beforeloggingbegins.Important:Sensorsthatareconnectedafterloggingbeginsmaynotlogaccuratedata.

    ClicktheFiltersbuttontocreateadditionalfilteredseriesforanyofthechannelsyouconfigured.Thefilteredserieswillthenbeautomaticallycalculatedwhenyoureadouttheloggerandplotthedata.

    Youhavetheoptiontorecordthelogger'sbatteryvoltageateachsampletime;thisisthelastchannelinthesensorslist.Liketheotherdatachannels,loggingtheinternalbatterychannelconsumessomeofthelogger'smemory.Unlessyoususpectabnormalbatteryperformance,youdonotneedtologthebatteryvoltage.Thisloggerwillrecordabadbatteryeventevenifitisnotloggingthebatterychannel.YoucanalsodisablethebatterychannelwithSeriespreferences(inPreferences,selectDisplay,thenSeries).

    Deployment LoggingInterval.Selecthowoftentheloggerwillrecorddata.Youcanchooseeitheroneofthe

    presetloggingintervalsorspecifyacustomlogginginterval.Theminimumloggingintervalisonesecondandthemaximumformostloggersis18hours,12minutes,and15seconds.Theshorterthelogginginterval,themorequicklymemoryfillsandbatterypowerisconsumed.Ifyoudidnotselectanychannelstolog,thisfieldwillbedisabled.

  • HOBOwareUsersGuide 33

    LoggingDuration.Thisliststheapproximatetimeitwilltaketofilltheloggermemorybasedontheloggingintervalandsensorscurrentlyselected.Thisisatheoreticalestimateonly;batterylifeandfrequentstateandeventloggingwillaffectyourdeployment.Ifyouareloggingonlystatechangesandevents,noestimationispossible.

    StartLogging.Selectwhentolaunchthelogger.Thisdefaultstothesettingforthelogger'spreviouslaunch.SeeLaunchTimeSavingOptions(HOBOwarePro).Youcanchoosetolaunchthislogger: Now.LoggingbeginsassoonasyouclicktheStartbutton. AtInterval.Loggingwillbeginatanexactinterval(forexample9:00:00ratherthan8:47:00when

    youchooseaonehourlogginginterval).Theexactstarttimedependsontheloggingintervalyouchoose.

    PushButton.Loggingwillnotstartuntilyoupressthebuttonontheloggerandholditdownforatleastthreeseconds.TheWaitingLEDwillblinkontheloggeruntilyoupresstheStart/Stopbutton.

    OnDate/Time.Loggingwillbeginatadateandtimeyouspecify,uptoapproximatelysixmonthsfromthepresent.TheWaitingLEDwillblinkontheloggeruntilloggingbeginsontheselecteddate/time.

    Stoplogging.Selectwhenyouwanttheloggertostoplogging.Youcanchoose: WhenmemoryfillsorNever(wrapping).Ifyouselect"Whenmemoryfills,"thentheloggerwill

    stoprecordingdataoncethememoryisfull.Ifyouselect"Never(wrapping),"theloggerwillrecorddatacontinuouslyuntileithertheloggerbatteryrunsoutoryoustopit.Oncetheloggerisfull,thenewestdatawilloverwritetheoldestdata.

    Pushbutton.Whenthisoptionisselected,theloggerwillstoprecordingdatawhenyoupresstheStart/Stopbuttonontheloggeritselffor3seconds.IfyoualsoconfiguredaPushButtonstart,thenyoumustwait5minutesafterloggingbeginsbeforeyoucanusethebuttontostoplogging.

    Specificstopdate.Selectthedateyouwanttheloggertostoprecordingdata.Chooseeitherapresettimeorsetyourowncustomdateandtime.

    Options.Select"TurnLEDsoff"ifyouwanttheloggertooperatein"stealthmode,"whichdisablestheLoggingLEDonthelogger.TheWaitingLEDandtheTestbutton/Activitylightswillstillremainoperationalwhenthisoptionisselected.

    Skiplaunchwindownexttime(HOBOwarePro)CheckthisboxifyouwouldliketobypasstheLaunchWindowthenexttimeyouchooseLaunchfromtheDevicemenuorclicktheLaunchicon.ThiswillcausethenextloggertobelaunchedwitheitherthepreviouslaunchsettingsorthecurrentloggersettingsassetinLaunchTimeSavingOptionspreferences.

  • HOBOwareUsersGuide 34

    LaunchOptionsforOtherUSeriesLoggersThefollowinglaunchoptionsareavailableforUSBUSeriesloggersotherthantheUX90,UX100,andUX120017x.NotethattheLaunchLoggerwindowmayvaryfromtheexampleshownbelow.

    LoggerInformation

    Description.Enteradescriptionforthelaunch,upto40characters,inthisfield.Thisdescriptionisusedasthedefaultfilenamewhenyoureadoutandsavethedatarecordedbythelogger.Itisalsothedefaulttitleoftheplot.

    SerialNumber.Thisistheserialnumberforthelogger.Notethatsomeexternalsensorsalsohaveserialnumbers,whicharenotlistedhere.

    DeploymentNumber.Thisisthenumberoftimes(includingthistime)theloggerhasbeenlaunched.Eachtimeyoustartanewlaunch,thedeploymentnumberincreasesbyone.

    BatteryLevel.Thisshowsthecurrentbatterylevelinthelogger. Status.ClicktheStatusbuttontochecktheloggersettingsusedinthepreviouslaunch.

    SensorsTheSensorsListdisplaysallinternalsensorsandexternalchannelsavailableforrecordingdata.Choosethesensors,orchannels,youwishtologinthisdeployment.Toconfigureanexternalsensor:

    1. Makesurethecheckboxisenabledfortheexternalchannelyouwishtoconfigure.2. Clickthebuttonforthecorrespondingchannelinthesensorlisttoselectthesensortypeasshown

    below.Iftheexternalsensorhasnotbeenconfigured,thebuttonreads.Otherwise,theprevioussensortypeselectedwillbedisplayedonthebutton.

    3. Selectthemeasurementcategoryandthenthespecificsensorpartnumberandrange(ifapplicable).Notethatthesensorpartnumbermustmatchthepartnumberlistedontheshrinktubeonthesensorcabletorecordaccuratedata.

  • HOBOwareUsersGuide 35

    4. Repeatsteps1through4foreachadditionalexternalsensoryouwishtoconfigure.

    Notesonconfiguringsensors:

    Althoughitishelpfultoseetheconnectedsensorswhensettingupthelaunch,itisnotrequiredthatyouconnectthesensorsduringlaunchsetup.Youmaysetupthelaunchforadelayedorpushbuttonstart,thenconnectthesensorslater,beforeloggingbegins.Important:Sensorsthatareconnectedafterloggingbeginswillnotloganydata.Alternatively,ifyouselectanexternalchannel,butdonotplugthesensorin,falsedatawillberecordedforthatchannel.

    SomeloggerswithRHorpressuresensorsrequirethatyoualsoselecttheinternaltemperaturesensortologrelativehumidityorpressure.Thedependencymaybenotedinthesensorslist(seeRelativeHumiditychannelintheexampleabove).Refertothemanualthatcamewithyourloggerforadditionalinformation.

    WhenboththeinternaltemperatureandRHsensorsareselected,dewpointwillbecalculatedautomatically.

    Ifexternaleventsorstatesareavailableforyourlogger,youcanenternamestoidentifytheeventandstatechannelsthatareinuse.YoucanalsorenametheOpenandCloseddescriptionsforeachstate,andenterincrementsandunitsforeacheventchannel.

    HOBOwareProDataAssistantsletyoucreatenewdataseriesbycombiningdatarecordedbytheloggerwithadditionaldata.WhenyouuseaDataAssistantfromtheLaunchLoggerwindow,theparametersetissavedandappliedtodataeverytimeyoureadoutaloggeroropentheresultantdatafile.ThederiveddataisalsoshownwhenyoucheckthelatestreadingsintheStatuswindow.ClicktheScalingbuttontoaccesstheDataAssistantswindowandchoosetheLinearScalingorPulseScalingAssistant.ClicktheCreatebuttontoruntheassistant.Afteryouentervalues,clickSave.SensorsthatsupportDataAssistantsatlaunchtimedisplaytheScalingiconnexttotheirname.

    Someloggershavealarmfunctionality.ClicktheAlarmsbutton,ifavailable,tosetalarmsasdesired.SensorsthathavealarmfunctionalitydisplaytheAlarmsiconnexttotheirname.

    ClicktheFiltersbuttontocreateanadditionalfilteredseries,suchasaveragetemperatureperday,automaticallywhenyoureadouttheloggerandplotthedata.

    Youhavetheoptiontorecordthelogger'sbatteryvoltageateachsampletime;thisisthelastchannelinthesensorslist.Liketheotherdatachannels,loggingtheinternalbatterychannelconsumessomeofthelogger'smemory.Unlessyoususpectabnormalbatteryperformance,youdonotneedtolog

  • HOBOwareUsersGuide 36

    thebatteryvoltage.AUSerieslogger(otherthanaHOBOU30Station)willrecordabadbatteryeventevenifitisnotrecordingthebatterychannel.YoucanalsodisablethebatterychannelwithSeriespreferences(inPreferences,selectDisplay,thenSeries).

    Deployment LoggingInterval.Selecthowoftentheloggerwillrecorddata.Youcanchooseeitheroneofthepreset

    loggingintervalsorspecifyacustomlogginginterval.Theminimumloggingintervalisonesecondandthemaximumformostloggersis18hours,12minutes,and15seconds.Theshorterthelogginginterval,themorequicklymemoryfillsandbatterypowerisconsumed.Ifyoudidnotselectanychannelstolog,thisfieldwillbedisabled.SeealsoMultipleLoggingIntervalsandFastLoggingIntervals.

    LoggingDuration.Thisliststheapproximatetimeitwilltaketofilltheloggermemorybasedontheloggingintervalandsensorscurrentlyselected.Thisisatheoreticalestimateonly;batterylifeandfrequentstateandeventloggingwillaffectyourdeployment.Ifyouareloggingonlystatechangesandevents,noestimationispossible.

    DOSensorCapExpires(HOBOU26DissolvedOxygenloggeronly).IfaDOsensorcapisinstalledontheU26logger,thedatethecapexpiresislisted.Theloggerwillnotcollectanydataafterthecapexpires.FormoredetailsontheDOsensorcap,seeWorkingwiththeHOBOU26DissolvedOxygenLogger.

    StartLogging.Selectwhentolaunchthelogger.Thisdefaultstothesettingforthelogger'spreviouslaunch.SeeLaunchTimeSavingOptions(HOBOwarePro).ForUSeriesloggers,youcanchoosetolaunchthelogger: Now.LoggingbeginsassoonasyouclicktheStartbutton. AtInterval.Loggingwillbeginatanexactinterval(forexample9:00:00ratherthan8:47:00when

    youchooseaonehourlogginginterval).Theexactstarttimedependsontheloggingintervalyouchoose.

    PushButton/UsingCoupler.(Availableonlyoncertainloggers.)Loggingwillnotstartuntilyoupressthebuttonontheloggerandholditdownforatleastthreeseconds,orreturntheloggertothecouplerwithoutthebasestationforatleastthreeseconds(refertoyourloggermanualformoreinformation).Iftheloggerhasalight,itwillflashquicklywhenloggingbegins.

    OnDate/Time.Loggingwillbeginatadateandtimeyouspecify,uptoapproximatelysixmonthsfromthepresent.

    Skiplaunchwindownexttime(HOBOwarePro)CheckthisboxifyouwouldliketobypasstheLaunchWindowthenexttimeyouchooseLaunchfromtheDevicemenuorclicktheLaunchicon.ThiswillcausethenextloggertobelaunchedwitheitherthepreviouslaunchsettingsorthecurrentloggersettingsassetinLaunchTimeSavingOptionspreferences.

  • HOBOwareUsersGuide 37

    LaunchOptionsforStationLoggersThefollowinglaunchoptionsareavailableforStationloggers,whichconnecttothecomputerviaserialcableinsteadofUSBcable.NotethattheLaunchLoggerwindowmayvaryfromtheexampleshownbelow.

    LoggerInformation

    Description.Enteradescriptionforthelaunch,upto40characters,inthisfield.Thisdescriptionisusedasthedefaultfilenamewhenyoureadoutandsavethedatarecordedbythelogger.Itisalsothedefaulttitleoftheplot.

    SerialNumber.Thisistheserialnumberforthelogger.Notethatsomeexternalsensorsalsohaveserialnumbers,whicharelistedintheSensorspane.

    DeploymentNumber.Thisisthenumberoftimes(includingthistime)theloggerhasbeenlaunched.Eachtimeyoustartanewlaunch,thedeploymentnumberincreasesbyone.

    BatteryLevel.Thisshowsthecurrentbatterylevelinthelogger. UserNotes.Clickthisbuttontotypeupto2000charactersofinformationaboutthedeployment.

    ThistextwillbedisplayedintheDetailspaneafteryoureadouttheloggerandplotthedata. Status.ClicktheStatusbuttontochecktheloggersettingsfromthepreviouslaunch.

    SensorsOnlythesensorsthatarecurrentlypluggedinorbuiltintotheloggerarelisted.Sensorsarelistedinascendingorderbyserialnumber,regardlessoftheirphysicalpositioninthelogger.Ifyouaddorremovesensors,clicktheRefreshbuttontomakesureyourchangesareseenbytheloggeranddisplayedinthislist.Youcanassignanametoeachsensor(upto30characters).Thisishelpfulifyouwanttospecifyalocationwherethesensorwillbeplacedorifyouneedtodifferentiatemultiplesensorsofthesametype.

  • HOBOwareUsersGuide 38

    SensorsmustbeconnectedpriortolaunchingifyouwillbeenteringanameforthesensororconfiguringFlexSmartmodulesorAnalogSensorPorts(usetheConfigurebuttontotherightoftheLabelfield).Otherwise,sensorsdonothavetobeconnectedwhensettingupthelaunch.Youmaysetupthelaunchforadelayedorpushbuttonstart,thenconnectthesensorslaterbeforeloggingbegins.Notesaboutconfiguringsensorchannels:

    Donotattachmorethan15channelsofsensors.Theloggercannotaccommodatemorethan15channels.

    Inadditiontologgingupto15channels,theHOBOEnergyLoggerandHOBOU30Stationcanrecordtheirinternalbatterychannel(othersensorsmustbeattached;theloggercannotlogbatteryalone).Thelogger'sbatteryvoltageisrecordedateachinterval.However,loggingtheinternalbatterychannelconsumessomeofthelogger'smemory.Unlessyouareusingexcitationpowerorsuspectabnormalbatteryperformance,youprobablydonotneedtologthebatteryvoltageinloggersotherthantheHOBOU30Station.

    Alwaysenablethelogger'sinternalbatterychannelwhenloggingwithexcitationpower.Thiswillhelpyoudeterminewhetherandwhenexcitationpowerwasturnedoffduringadeployment.

    HOBOwareProDataAssistantsletyoucreatenewdataseriesbycombiningdatarecordedbytheloggerwithadditionaldata.WhenyouuseaDataAssistantfromtheLaunchLoggerwindow,theparametersetissavedandappliedtodataeverytimeyoureadoutaloggeroropentheresultantdatafile.ThederiveddataisalsoshownwhenyoucheckthelatestreadingsintheStatuswindow.YoucanusethefollowingDataAssistantsfromtheLaunchwindow(otherscanonlybeusedfromthePlotSetupwindow):PulseScaling,LinearScaling,andkWh.ThisfeatureisavailablefortheHOBOU30,theHOBOH21(WeatherStationandMicroStation)andtheHOBOH22(EnergyLogger)loggers.ClickthekWhorScalingbuttontoaccesstheDataAssistantswindow.SelectthedesiredassistantandclicktheCreatebuttontoruntheselectedassistant.Afteryouentervalues,clickSave.SensorsthatsupportDataAssistantsatlaunchtimedisplaytheScalingand/orkWhiconnexttotheirname.

    ClicktheFiltersbuttontocreateanadditionalfilteredseries,suchasaveragetemperatureperday,automaticallywhenyoureadouttheloggerandplotthedata.

    Deployment LoggingInterval.Selecthowoftentheloggerwillrecorddata.Youcanchooseeitheroneofthe

    presetloggingintervalsorspecifyacustomlogginginterval.Theminimumloggingintervalisonesecondandthemaximumformostloggersis18hours,12minutes,and15seconds.Theshorterthelogginginterval,themorequicklymemoryfillsandbatterypowerisconsumed.Somesensors(specifically,theFlexSmartTRMSmodules)requirealoggingintervaloftwosecondsorgreater.Ifyouchooseafasterlogginginterval,erroneousdatawillbeloggedonthesechannels.

    SamplingInterval.Thesamplingintervalallowsyoutotakemultiplemeasurementswithinthelogginginterval,thenaveragethemtogethertocreateasingleloggedmeasurement.Thesamplingintervalisoptionalandisvalidonlyforsensorsthatsupportmeasurementaveraging.Refertothesensor'susermanualtodeterminewhethermeasurementaveragingisavailableonthesensor.Ifyouhaveatleastonesensorthatsupportsmeasurementaveraging,clicktheEnablebutton,thensetthesamplingintervalatlessthanorequaltothelogginginterval(uptofourminutes).Rapidsampling(fasterthanoneminute)willreducethelogger'sbatterylife.Ifyoudonothaveanysensorswithmeasurementaveragingorwishtoturnoffthesamplinginterval,clicktheDisablebutton.

    LoggingDuration.Thisliststheapproximatetimeitwilltaketofilltheloggermemorybasedontheloggingintervalandsensorscurrentlyselected.Thisisatheoreticalestimateonly;batterylifeandfrequentstateandeventloggingwillaffectyourdeployment.Ifyouareloggingonlystatechangesandevents,noestimationispossible.IfyouaddorremovesensorswhileviewingtheLaunchwindow,clicktheRefreshbuttontogetanupdatedLoggingDuration.

    StartLogging.Selectwhentolaunchthelogger.Thisdefaultstothesettingforthelogger'spreviouslaunch.SeeLaunchTimeSavingOptions(HOBOwarePro).Youcanchoosetolaunchthelogger:

  • HOBOwareUsersGuide 39

    Now.LoggingbeginsassoonasyouclicktheStartbutton.Ifyoudonothaveatleastonesensorattached,theloggerwillnotlaunch.

    AtInterval.Loggingwillbeginatanexactinterval(forexample9:00:00ratherthan8:47:00whenyouchooseaonehourlogginginterval).Theexactstarttimedependsontheloggingintervalyouchoose.

    PushButton.Loggingwillnotstartuntilyoupressthebuttonontheloggerandholditdownforatleastthreeseconds.Iftheloggerhasalight,itwillflashquicklywhenloggingbegins.ThisisnotavailablefortheU30Station.

    OnDate/Time.Loggingwillbeginatadateandtimeyouspecify,uptoapproximatelysixmonthsfromthepresent.

    SaveSettingsinLogger.Loggingwillnotstart,butthelaunchsettingswillbesavedsothatyoudonothavetoreenterthemwhenyouarereadytolaunchatalatertime.ThenexttimeyouviewtheLaunchwindowforthislogger,thesettingsyouenteredwillstillbeinplace.

    StopLogging.Therearetwooptionsforstoppingthelogger:either"whenmemoryfills"or"never(wrapping)."Ifyouselect"whenmemoryfills,"thentheloggerwillstoprecordingdataoncethememoryisfull.Ifyouselect"never(wrapping),"theloggerwillrecorddatacontinuouslyuntileithertheloggerbatteryrunsoutoryoustopit.Oncetheloggerisfull,thenewestdatawilloverwritetheoldestdata.

    Skiplaunchwindownexttime(HOBOwarePro)CheckthisboxifyouwouldliketobypasstheLaunchWindowthenexttimeyouchooseLaunchfromtheDevicemenuorclicktheLaunchicon.ThiswillcausethenextloggertobelaunchedwitheitherthepreviouslaunchsettingsorthecurrentloggersettingsassetinLaunchTimeSavingOptionspreferences.Note:ForadditionaldetailsonlaunchingtheHOBOU30Station,seeWorkingwiththeHOBOU30Station.

    MultipleLoggingIntervalsOverviewSomeloggerscanbeconfiguredwithmultipleloggingintervals,whichallowyoutodefineseparateintervalsfordifferentphasesofthedeployment.Ifmultipleloggingintervalsareavailableforyourlogger,youwillseeanAddNewIntervalbuttonontheLaunchwindowforyourlogger.Becauseinternalloggerevents,suchascouplerevents,dotakeupsomeofthelogger'smemory,thenumberofsamplesandendtimesareonlyestimates.Frequenteventswillcausetheloggertofillupsooner.

    StepsToconfiguremultipleloggingintervals:

    1. ClicktheAddNewIntervalbuttonuntilyouhaveasmanyintervalsasyouwant(uptothenumberofintervalsavailableforthelogger).

    2. Setthehours,minutes,andsecondsforeachinterval,andthenumberofsamplesyouwanttheloggertorecordateachinterval.

    3. UsetheStartandEndtimesintheDurationcolumntohelpyoudecidehowmanysamplestorecordateachinterval.

    Todeleteaninterval,clicktheRemovebuttonnexttotheinterval.

  • HOBOwareUsersGuide 40

    FastLoggingIntervalsSomeloggerscanbeconfiguredwithfastloggingintervals,whichallowyoutologmorethanoncepersecond,upto100measurementspersecond(hertz).Iffastloggingintervalsareavailableforyourlogger,theLaunchwindowforyourloggerwillofferLoggingIntervalchoicesofNormalandFast.Tousefastloggingintervals,clicktheFastbutton.ThenchoosealoggingfrequencyfromtheHertzdropdownlist,orentertheloggingintervalindecimalformintheSecfield.

    NOTES Aloggercannotcommunicatewithacomputerorshuttlewhileitisloggingatafastinterval.Onceit

    hasstartedlogging,ifyouconnectittoacomputer,basestation,orshuttletocheckstatusorreaditout,theloggerwillstop.

    BeawarethatafastloggingintervaltypicallymeansaveryshortLoggingDuration.Makesuretheloggerwillrunlongenoughtocollectthedatayouneed.Disablingunnecessarychannelswillgiveyoumoretime.

    Becausefastloggingdeploymentsaresobrief,theloggermustbelaunchedwithatriggerstart.Followtheinstructionsthatcamewithyourloggerforatrigger(couplerorbutton)start.Iftheloggerhasalight,thelightwillflashquicklywhenloggingbegins.