Hobgen test1940-1960

Page 3: Hobgen test 1940-1960 - UZ Leuven

GC-MS (steroïden)1930-today

- Laborious (derivatisation)- Time consuming (long run times)- But still considered reference method

Page 4: Hobgen test 1940-1960 - UZ Leuven

RosalynYalow

RIA

1960-today

Page 5: Hobgen test 1940-1960 - UZ Leuven

Concentration range of analytes in blood

Page 6: Hobgen test 1940-1960 - UZ Leuven

The workhorse of the endocrine laboratory

Immunoassay and its pitfalls

Page 7: Hobgen test 1940-1960 - UZ Leuven
Page 8: Hobgen test 1940-1960 - UZ Leuven

Competitive immunoassay

• Only one epitope• One antibody• Competion for antibody

• Less sensitive• Less specific• Small dynamic range

Page 9: Hobgen test 1940-1960 - UZ Leuven
Page 10: Hobgen test 1940-1960 - UZ Leuven

Sandwich assay

• Two epitopes• Excess capture antibody• Excess detection antibody

• Most sensitive immunoassays• Large dynamic range• High specificity

Page 11: Hobgen test 1940-1960 - UZ Leuven

Immuno-assaypro contra• Relatively sensitive• Easy• Fully automated• Direct measurement

• Lack of accuracy• Lack of sensitivity• Lack of specificity• Limited synamic range• Expensive (kit) reagents• Black box

Page 12: Hobgen test 1940-1960 - UZ Leuven

Serum tubePlastic

Vacuum

Rubber

Separator gel

Surfactants

Clot activator

Page 13: Hobgen test 1940-1960 - UZ Leuven

Plasma tubeLi heparin (green)

fast centrifugation

no clotting

EDTA (purple)

ACTH

renine

aldosteron

Page 14: Hobgen test 1940-1960 - UZ Leuven

Stability and storage

STAT analysis• PTH• Insulin• …

Frozen till analysis

• ACTH• Plasma Renin Activity• Aldosterone• C peptide• Gastrin• Glucagon• …

Page 15: Hobgen test 1940-1960 - UZ Leuven

hemolysisSmall effect on immunoassaysbut:

proteolysisinsulingastringlucagonPTH

cellular releaseNSE

lipemiaSmall effect on immunoassaysBut:

turbidimetricnephelometric

Page 16: Hobgen test 1940-1960 - UZ Leuven

Carry-over

• Most of today’s diagnostic analyzers take every possible precaution to avoid sample carry-over.

• In spite of these efforts, a sample having a VERY high result may affect one or more samples that follow it. e.g. hCG

Page 17: Hobgen test 1940-1960 - UZ Leuven

Cross-reactivity: hCG and LH

Page 18: Hobgen test 1940-1960 - UZ Leuven

Cross-reactivity: steroids

Page 19: Hobgen test 1940-1960 - UZ Leuven

Hormone binding proteins

Albumin, SHBG, TBG, CBG

For total hormone measurement, it is essential to displace all bound hormone from endogenous binding sites

solvent extraction

anilino naphtalene sulfonic acid

For free hormone measurement, displacement can alter the equilibrium

(always happens to a certain degree in free hormone assays + low concentration!)

NEFA in free thyroxine assays

Page 20: Hobgen test 1940-1960 - UZ Leuven

Autoanalyte antibodiesThyroid hormones

Thyroglobulin

Insulin

Prolactin

Page 21: Hobgen test 1940-1960 - UZ Leuven

MacroprolactinBiologically inactive

Cleared slowly

Causes false high prolactin

Page 22: Hobgen test 1940-1960 - UZ Leuven

Heterophile antibodiesHAMA

Page 23: Hobgen test 1940-1960 - UZ Leuven

Bridging of capture anddetector antibodies=> Falsely elevated result

Exclusive binding of capture or detector antibody only=> Falsely lowered result

Page 24: Hobgen test 1940-1960 - UZ Leuven
Page 25: Hobgen test 1940-1960 - UZ Leuven

Anti-Ru antibodies (Roche)

Mainly in areas with textile industry Use of Ru in dying process of clothing

Ru in environment, clothing or food chain

Estimated frequency of interference in first generation Elecsys FT3 assay (Roche Diagnostics): 0.2% (Sapin, Clin Chem Lab Med 2007)

Page 26: Hobgen test 1940-1960 - UZ Leuven

Case report: anti-rhutenium antibodies in 7 y old girl

TSH 0 .66 mIU/L 0.27 - 4.20 Vrij T4 76 pmol/L 12 - 22 Vrij T3 27.4 pmol/L 3.1 - 6.8

Oestradiol (LC-MSMS) < 1 .3 ng/L Oestradiol (Roche) 399 ng/LTestosteron 82 .1 ng/dL 15.0 - 45.0 SHBG 66 nmol/L 41 - 103 Androstenedion 14 .1 ng/dL 30.0 - 200.0

Page 27: Hobgen test 1940-1960 - UZ Leuven

High dose hook effectTumor markers

hCG

Ferritin

Thyroglobulin

High analyte concentration

Page 28: Hobgen test 1940-1960 - UZ Leuven

Poor agreement with reference method

• Poor standardization• Cross reactivity (poor quality of antibody)

e.g. androstenedione, cortisol, oestrone

Page 29: Hobgen test 1940-1960 - UZ Leuven

0 100 200 300 400 5000

50

100

150

200

250

300

350

400

450

Andr RIA

Andr

LC

MS

Page 30: Hobgen test 1940-1960 - UZ Leuven

Lack of sensitivity and bad assay design

Page 31: Hobgen test 1940-1960 - UZ Leuven
Page 32: Hobgen test 1940-1960 - UZ Leuven

Free hormone assaysFree Thyroid Index (FTI)

Page 33: Hobgen test 1940-1960 - UZ Leuven
Page 34: Hobgen test 1940-1960 - UZ Leuven
Page 35: Hobgen test 1940-1960 - UZ Leuven

Total T4Immuno A

Free T4Immuno A

Total T3Immuno A

Free T3Immuno A

LC-MS/MS LC-MS/MS

Page 36: Hobgen test 1940-1960 - UZ Leuven

Mass SpectrometryA reference method for hormones

Page 37: Hobgen test 1940-1960 - UZ Leuven

1) Sample preparation

• Protein precipitation (methanol, acetonitrile, ZnSO4)

• Liquid liquid extraction• Solid phase extraction

Page 38: Hobgen test 1940-1960 - UZ Leuven

2) HPLC

Page 39: Hobgen test 1940-1960 - UZ Leuven

HPLC in practice- Pumps- Fluids

- Tubings- Sample loader- Complex valves

- Column

Page 40: Hobgen test 1940-1960 - UZ Leuven

Massaspectrometer (Q)

CortisolPrednisolon

362,460 Da360,444 Da

++

+MS1

+

Page 41: Hobgen test 1940-1960 - UZ Leuven

Tandem Massaspectrometer (QQQ)

++

++

++

MS1 MS2

+++

Page 42: Hobgen test 1940-1960 - UZ Leuven

Isobaric interference

Page 43: Hobgen test 1940-1960 - UZ Leuven

Mass Spectrometryin practice

- Ion source- Switching valves- Tubings- Vacuum pump- Nitrogen generator

Page 44: Hobgen test 1940-1960 - UZ Leuven

Mass Spectrometryin practice

- Spray probe- Heating probes- Curtain plate- Curtain plate aperture- Orifice

Page 45: Hobgen test 1940-1960 - UZ Leuven
Page 46: Hobgen test 1940-1960 - UZ Leuven

LC-MS/MS in EndocrinologyAnalytical Superiority

Page 47: Hobgen test 1940-1960 - UZ Leuven

Added value LC-MS/MS: Accuracy

Page 48: Hobgen test 1940-1960 - UZ Leuven

van den Ouweland JMW, Kema IP. J Chromatography B 2012

Added LC-MS/MS: Sensitivity

Page 49: Hobgen test 1940-1960 - UZ Leuven

Cortisol Prednisolon

Cross reactivity!

Added value LC-MS/MS: Specificity

Page 50: Hobgen test 1940-1960 - UZ Leuven

???

???

Page 51: Hobgen test 1940-1960 - UZ Leuven
Page 52: Hobgen test 1940-1960 - UZ Leuven

Added value LC-MSMS: steroid profiling

Page 53: Hobgen test 1940-1960 - UZ Leuven

Added value LC-MS/MS: pro and contra

Advantages• Analytical superiority• “Cheap” reagents

• Organic solvents• Deuterated internal standards• Robust colums• Calibrator sets• Pipetting tips and plates

• High troughput possible• Develop new assays

Disadvantages• Expensive instruments• Batch mode• Skilled labor required• LIS interface• Regulartory uncertainty• Ruggedness