Upload
others
View
8
Download
1
Embed Size (px)
Citation preview
Handling high matrix & challenging samples with Agilent ICP-MS
Uwe Noetzel
Agilent Technologies
Waldbronn, Germany
Agilent’s History in ICP-MS30 Years of ICP-MS Innovation
1987 2000 20091994
First computer-controlled ICP-MS
PMS series
4500
7500Enabling high
sensitivity
metal analysis
Enabling
routine robust
ICP-MS analysis
Enabling control
of common
interferences
First benchtop ICP-MSCool plasma
Enabling ease
of use and
productivity
9 orders detectorORS cell
HMIISIS-DS MassHunter SW
Enabling controlled
reaction chemistry
8800 ICP-QQQ
2012
World’s first ICP-QQQ
2014
#1 selling ICP-MS !
77007900
UHMIODS detectorISIS 3
A new era in ICP-
MS performance
2015 2016
7800
Solution readyMethod automation
Enabling simplified
ICP-MS workflows
Flexible, high
performance MS/MS
New 8900 ICP-QQQ
Second generation ICP-QQQ
November 2016
Spectroscopy Seminar Greece
Page 2
Performances
Flame AA
MP-AES
ICP-OES
Working rangeppt %
ICP-MS
Furnace AA
Mono-element
technique
ppb
11/15/2016
Agilent ICP-MS Systems
3
Processes in ICP-MS Plasma
Mostly the same as ICP-OES
November 2016
Spectroscopy Seminar Greece
4
General ICP-MS Components/Technology
1. The liquid
sample is mixed
with argon gas by
the nebulizer to
form an aerosol..
2. The smallest
droplets pass
through the spray
chamber and into
the ion source -
the plasma
3. The sample
is desolvated
and ionized in
the plasma
4. Ions are
extracted from the
plasma by
extraction lenses in
the interface region
5. Ion lenses focus and
collimate the ions. The Omega
lens bends the ion beam off-
axis to prevent photons
from striking the detector
7. The quadrupole mass
spectrometer separates ions based
on their mass to charge ratio. The
selected ions continue on to the
detector
8. Ions are measured
using a discrete
dynode detector
providing 9 orders of
linear dynamic
range6. Interference Ions
are removed from the
ion path by
reaction/collision/KED
effects in the cell
November 2016
Spectroscopy Seminar Greece
5
3 Key Benefits of Agilent ICP-MS
1) Matrix Tolerance
Ultra High Matrix
Introduction System (UHMI)
Handles tough sample
matrices better than any
other ICP-MS. Highest total
dissolved solids(TDS)
capability up to 3% with
7800 and 25% with 7900
• Reduces sample prep time
and error
• Better long-term stability
Page 6November 15, 2016
November 2016
Spectroscopy Seminar Greece
6
Processes in ICP-MS Plasma
November 2016
Spectroscopy Seminar Greece
7
The key technology – HMI and UHMI- Aerosol Control
High matrix sample
Carrier Gas
Sampling Cone
Torch
Spray chamber
Nebulizer
Dilution Gas
Skimmer
Peristaltic pump Aerosol DilutionDilution gas reduces sample load on plasma.
Dilution factor is selectable from 1 to 1/100
Agilent’s patented HMI dilutes the sample in the gas
state, using aerosol dilution. This removes the main
problems of liquid sample dilution:
• Time / Reagents / Errors / Contamination
Agilent UHMI
November 2016Spectroscopy
Seminar GreecePage 8
Ultra HMI (High Matrix Introduction) system
November 2016
Spectroscopy Seminar Greece
Carrier gas
Dilution gas
Carrier gas flow (L/min) Total gas
flow
7700 HMI 7900 UHMI
HMI-4 (HMI-L) 0.6 0.8 0.95
HMI-8 (HMI-M) 0.35 0.68 0.95
HMI-25 (HMI-H) 0.23 0.5 0.95
HMI-50 N.A. 0.4 0.95
HMI-100 N.A. 0.33 0.95
• Increased dilution range to x100 – even
higher matrix capability
• Less matrix loading to interface, so
better long-term stability
• UHMI maintains high carrier gas flow
through spray chamber, so faster gas
replacement and washout
9
HMI – a Patent by Agilent Technologies
November 2016
Ken Sakata
Spectroscopy Seminar Greece
10
Lowest Oxide AND Double-Charged Ratios
Figure-of-Merit AgilentAgilent
HMIBrandX
Oxide ratio % CeO+/Ce+ < 1.2* 0.2 - 1.0 < 2.5
Double-Charged ratio
%Ce2+/Ce+ < 2* < 2 < 3
No other ICP-MS can achieve these
low Oxide AND double-charged ratios
HMI
* Demonstrated at Installation
November 2016
Spectroscopy Seminar Greece
11
45Sc 52Cr 56Fe 63Cu 66Zn 111Cd 115In 136Ba
100
50
0
% Recovery of a 10ppb mixed standard spiked into undiluted NASS-5 Seawater
0.2% CeO+/Ce+ 1 % CeO+/Ce+ 2 % CeO+/Ce+
Greater Matrix Tolerance = Lower Oxides
November 2016
Spectroscopy Seminar Greece
12
Expanding the ICP-MS application to High matrix sample types
• Marine organisms
• Sediments
• Groundwater
• Open ocean
• Coastal seawater
• Enclosed lakes and lagoons
- with salt levels far in excess of oceanic seawater. Also known as
hypersaline lakes, where the level of salinity may be many times
that of seawater
- See Agilent application note 5991-4257EN
November 2016
Spectroscopy Seminar Greece
13
3 Key Benefits of Agilent ICP-MS
1) Matrix Tolerance 2) Interference Removal
Ultra High Matrix
Introduction System (UHMI)
Handles tough sample
matrices better than any
other ICP-MS. Highest total
dissolved solids(TDS)
capability up to 25%
• Reduces sample prep time
and error
• Better long-term stability
Octopole Reaction System
(ORS4) He Mode
He Mode effectively removes
common polyatomic
interferences in the samples
• Easy to use - He collision
mode only
• Effective for a wide range of
sample types
• Accurate measurements
Page 14 November 2016
Spectroscopy Seminar Greece
14
He Mode – Quite Brilliant
Page 15
Spectroscopy Seminar Greece
General ICP-MS Components/Technology
1. The liquid
sample is mixed
with argon gas by
the nebulizer to
form an aerosol..
2. The smallest
droplets pass
through the spray
chamber and into
the ion source -
the plasma
3. The sample
is desolvated
and ionized in
the plasma
4. Ions are
extracted from the
plasma by
extraction lenses in
the interface region
5. Ion lenses focus and
collimate the ions. The Omega
lens bends the ion beam off-
axis to prevent photons
from striking the detector
7. The quadrupole mass
spectrometer separates ions based
on their mass to charge ratio. The
selected ions continue on to the
detector
8. Ions are measured
using a discrete
dynode detector
providing 9 orders of
linear dynamic
range6. Interference Ions
are removed from the
ion path by
reaction/collision/KED
effects in the cell
November 2016
Spectroscopy Seminar Greece
16
Polyatomic Interference Formation
40Ar 35Cl75As
75
November 2016
Spectroscopy Seminar Greece
17
Helium interactions in an Octopole Reaction CellCollision: Energy discrimination
He
Collision
Ar
Cl
Ar
Cl
As
As
Electrical potential (Octopole)
Reaction cell
Electrical potential (Q-pole)
Molecular interference(ArCl) has larger cross
section than the analyte (As).
More frequent interactions with He.
A significant reduction in kinetic energy relative to
the analyte (As). Energy filtering can be used to
ensure only the analyte
enters the quadrupole analyzer.
Page 18 Spectroscopy Seminar Greece
Page 18
2E5
cps
45 50 55 60 65 70 75 80Mass
Blank Acid Matrices and IPA in No Gas Mode
No Gas Mode
Unspiked 5% HNO3 + 5% HCl + 1% H2SO4 + 1% IPA Matrix
Unspiked Matrix – ALL peaks are due to polyatomic interferences
Multiple polyatomic interferences affect almost every
mass – Interferences are matrix-dependent
Color of spectrum indicates which matrix gave each interfering peak
November 2016
Spectroscopy Seminar Greece
19
45 50 55 60 65 70 75 80Mass
2E5
cps
He Mode
ALL polyatomic interferences are removed
in He Mode
Unspiked 5% HNO3 + 5% HCl + 1% H2SO4 + 1% IPA Matrix
ALL polyatomic interferences are removed in He Mode (same cell conditions)
Blank Acid Matrices and IPA in He Mode
Color of spectrum indicates which matrix gave each interfering peak
November 2016
Spectroscopy Seminar Greece
20
45 50 55 60 65 70 75 80Mass
2E5
cps
He Mode
Now we will add a 10 ppb spike
Unspiked 5% HNO3 + 5% HCl + 1% H2SO4 + 1% IPA Matrix
ALL polyatomic interferences are removed in He Mode (same cell conditions)
Blank Acid Matrices and IPA in He Mode
Color of spectrum indicates which matrix gave each interfering peak
November 2016
Spectroscopy Seminar Greece
21
2E5
cps
45 50 55 60 65 70 75 80Mass
10ppb Spike in 5% HNO3 + 5% HCl + 1% H2SO4 + 1% IPA Matrix
Consistent high sensitivity for all isotopes of all elements in He Mode
Matrix Mix with Spike (10ppb) in He Mode
He Mode
Good signal for all spike
elements at 10ppb Spike.
Perfect template fit for all
elements – no residual
interferences and no loss of
analyte signal by reaction
Consistent sensitivity and perfect template match for all elements
November 2016
Spectroscopy Seminar Greece
22
2E5
cps
45 50 55 60 65 70 75 80Mass
Matrix Mix with Spike (10ppb) in He Mode
He Mode
Let’s zoom for Fe and check the perfect template 56Fe/57Fe
Perfect agreement 56Fe/57Fe
ArO interference removed
40Ar
16O
November 2016
Spectroscopy Seminar Greece
23
2E5
cps
45 50 55 60 65 70 75 80Mass
Matrix Mix with Spike (10ppb) in He Mode
He Mode
Let’s zoom for Cr and check the perfect template 52Cr/53Cr
Perfect agreement 52Cr/53Cr
ArC interference removed
40Ar
12C
November 2016
Spectroscopy Seminar Greece
24
3 Key Benefits of Agilent ICP-MS
1) Matrix Tolerance 2) Interference Removal 3) Dynamic Range
Ultra High Matrix
Introduction System (UHMI)
Handles tough sample
matrices better than any
other ICP-MS. Highest total
dissolved solids(TDS)
capability up to 25%
• Reduces sample prep time
and error
• Better long-term stability
Octopole Reaction System
(ORS4) He Mode
He Mode effectively removes
common polyatomic
interferences in the samples
• Easy to use - He collision
mode only
• Effective for a wide range of
sample types
• Accurate measurements
Wider Dynamic Range
10 orders dynamic range with
7800 and 11 orders with 7900:
0.1ppt(DL) to 10,000ppm
• Simplifies method
development
• Easy sample prep.
• Improves productivity
Page 25 November 2016
Spectroscopy Seminar Greece
25
Far Wider Measurement Range Than Any Other ICP-MS11 orders - low and high level calibrations in a single run
Cd (1ppt - 1ppb) and Na (100ppb - 10,000ppm (1%)) in the same run
Concentration range (11 orders)
and upper measurement limit
(>1%) are at least 10x better
than any other ICP-MS
Both calibrations are linear.
Total concentration range
covered from Cd blank (BEC
of <0.1ppt) to Na top
standard (1%) is 11 orders
November 2016
Spectroscopy Seminar Greece
26
Putting All Together the advantages of the Agilent ICP-MSsimple, one cell gas, interference-free ICP-MS for real world sample analysis
One Configuration (standard –
robust plasma conditions)
One gas mode* (helium
collision)
One method for waters,
soils, wastes, food, urine,
blood, serum…
28 elements in less than 3.5 minutes per
run
* Plus no gas mode for non-interfered elements
Page 28 November 2016
Spectroscopy Seminar Greece
28
Agilent ICP-MS for Drinking Water Analysis
Graphite Furnace Atomic Absorption (GFAAS)(for trace metals – Cd, Cr, Ni, Ag, Pb)
Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES)(Al, Mn, Fe, Cu, Zn, Ba, B, Ca, Mg)
Hydride Generation/Atomic Fluorescence Detection(Hg, As, Se, Sb)
each with different sample preparation requirements
Agilent ICP-MS: The perfect replacement!
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS)(Cd, Cr, Ni, Ag, Pb, Al, Mn, Fe, Cu, Zn, Ba, B, Ca, Mg, Hg, As, Se, Sb, U)
November 2016
Spectroscopy Seminar Greece
29
High-Throughput Routine Analysis
November 2016
Spectroscopy Seminar Greece
30
Elemental coverage, dynamic range
and multielement capability has
meant ICP-MS is widely accepted
in routine labs
Method Detection Limits - 3 (n = 10)in Parts Per Billion for Drinking Waters(in 1% HNO3/0.5% HCl)
•All elements except low mass non-interfered elements were run in helium mode
•Low mass, non-interfered elements run in no gas mode
•No correction equations used
Element MassMDL
(ppb)Cell mode Element Mass
MDL
(ppb)Cell mode
Be 9 0.005 No gas Zn 66 0.014 He
B 11 0.005 No gas As 75 0.011 He
Na 23 0.058 No gas Se 78 0.017 He
Mg 24 0.002 No gas Sr 88 0.002 He
Al 27 0.007 No gas Mo 95 0.006 He
K 39 0.076 He Ag 107 0.002 He
Ca 44 0.090 No gas Cd 111 0.002 He
V 51 0.014 He Sb 121 0.006 He
Cr 52 0.004 He Ba 137 0.005 He
Mn 55 0.008 He Hg 202 0.001 He
Fe 56 0.014 He Tl 205 0.002 He
Co 59 0.004 He Pb 208 0.001 He
Ni 60 0.014 He Th 232 0.001 He
Cu 63 0.002 He U 238 0.001 He
Page 31 November 2016
Spectroscopy Seminar Greece
31
Analysis of Food CRM Digests Using Agilent ICP-MS with He Mode Only (single mode)Single sample preparation is required
•Food CRMs microwave digested with 6ml HNO3 + 2ml H2O2
•Final volume 100 mL
Single analysis by Agilent ICP-MS for all the analytes!
Full mass range scan of food
digest (e.g. microwave digestion)
Page 32 November 15, 2016
Agilent Spectroscopy Seminar Iceland
Few calibrations curves (prepared in 1% HNO3/0.5% HCl)
Na,
LD= 160 ppt
Mg,
LD= 31 ppt
As,
LD= 8 ppt
Cd,
LD= 0.7 ppt
Hg,
LD= 3 ppt
Pb,
LD= 1 ppt
Excellent measurement capabilities for all the range of masses
November 15, 2016November 15, 2016
Agilent Spectroscopy Seminar Iceland
Biological fluidsWhole blood, serum, urine, plasma
Traces elements measurements (toxic and essential ones)
in biological fluids
Samples simply diluted (10 to 20 times) with specific basic solution
containing:
• Butanol
• EDTA
• Triton X100
• Ammonia
More than 100 samples can be prepared per hour !
November 2016
Spectroscopy Seminar Greece
34
Biological fluidsWhole blood, serum, urine, plasma
Diluent Stock Solution
November 2016
Spectroscopy Seminar Greece
35
10 20 30 40 50 60 70 80 90 100110120130140150160170180190200210220230240250260
5.0E4
1.0E5
[1] Spectrum No.1 [ 181.525 sec]:002SMPL.D# / Tune #1 [CPS] [Linear]
m/z-> 10 20 30 40 50 60 70 80 90 100110120130140150160170180190200210220230240250260
5.0E4
1.0E5
[1] Spectrum No.1 [ 181.531 sec]:004SMPL.D# / Tune #1 [CPS] [Linear]
m/z->
Analysis of Toxic and Harmful ElementsElemental Screening of 1:10 diluted Urine (with interference removal in He mode)
• Unique capability of ICP-MS to acquire a scan across the entire mass range in
about 2 minutes, screening elements from 1000’s ppm to sub-ppb levels
Pb
Rb
Fe
Br Ba
Ca
Li MoI
CsSr
Zn
Cu
Sb
Sn
C Na
Mg
As
10 20 30 40 50 60 70 80 90 100110120130140150160170180190200210220230240250260
5.0E4
1.0E5
[1] Spectrum No.1 [ 181.531 sec]:004SMPL.D# / Tune #1 [CPS] [Linear]
m/z-> 10 20 30 40 50 60 70 80 90 100110120130140150160170180190200210220230240250260
2.5E5
5.0E5
[1] Spectrum No.1 [ 181.531 sec]:004SMPL.D# / Tune #1 [CPS] [Linear]
m/z->
PbFe
Br Ba
Li MoI
CsSr
Zn
Cu
Sb
SnAs
RbC Na
Mg
C
a
Identifying poisons
• “Unknown” element spiked into urine sample
Unknown element
Presence of element can be confirmed by spectral
template
November 2016
Spectroscopy Seminar Greece
36
10 20 30 40 50 60 70 80 90 100110120130140150160170180190200210220230240250260
5.0E4
1.0E5
[1] Spectrum No.1 [ 181.525 sec]:002SMPL.D# / Tune #1 [CPS] [Linear]
m/z-> 10 20 30 40 50 60 70 80 90 100110120130140150160170180190200210220230240250260
5.0E4
1.0E5
[1] Spectrum No.1 [ 181.531 sec]:004SMPL.D# / Tune #1 [CPS] [Linear]
m/z->
Analysis of Toxic and Harmful ElementsElemental Screening of 1:10 diluted Urine (with interference removal in He mode)
• Unique capability of ICP-MS to acquire a scan across the entire mass range in
about 2 minutes, screening elements from 1000’s ppm to sub-ppb levels
Pb
Rb
Fe
Br Ba
Ca
Li MoI
CsSr
Zn
Cu
Sb
Sn
C Na
Mg
As
10 20 30 40 50 60 70 80 90 100110120130140150160170180190200210220230240250260
5.0E4
1.0E5
[1] Spectrum No.1 [ 181.531 sec]:004SMPL.D# / Tune #1 [CPS] [Linear]
m/z-> 10 20 30 40 50 60 70 80 90 100110120130140150160170180190200210220230240250260
2.5E5
5.0E5
[1] Spectrum No.1 [ 181.531 sec]:004SMPL.D# / Tune #1 [CPS] [Linear]
m/z->
PbFe
Br Ba
Li MoI
CsSr
Zn
Cu
Sb
SnAs
RbC Na
Mg
C
a
Identifying poisons
• “Unknown” element spiked into urine sample
Unknown element
194 196 198 200 202 204 206 208 210 212 214 216
5.0E4
1.0E5
[1] Spectrum No.1 [ 175.118 sec]:003SMPL.D / Tune #1 [CPS] [Linear]
m/z->
Pb
194 196 198 200 202 204 206 208 210 212 214 216
2.5E5
5.0E5
[1] Spectrum No.1 [ 175.118 sec]:005SMPL.D# / Tune #1 [CPS] [Linear]
m/z->
Tl
210Po
208Pb
205Tl
203Tl
Presence of element can be confirmed by spectral
template
November 2016
Spectroscopy Seminar Greece
37
Speciation: HPLC-ICP-MS with Agilent 1260
November 2016
Spectroscopy Seminar Greece
38
Arsenic concentration in food varies over 4 orders of magnitude(slide courtesy J. Feldmann, U. Aberdeen)
0.01 mg/kg 0.1 mg/kg 1 mg/kg
20-200
0.5-1,5
>500
0.001
0.05-0.09
0.005-
0.01 0.22-1.1
3.0
[Arsenic]
Schoof et al. 1999, Williams et al. EST 2005, Raab et al. 2004
11/15/2016
Agilent ICP-MS Systems
39
As Speciation - Toxicity
Toxic!
Non-Toxic
?
Courtesy Ute Kohlmeyer GALAB, Germany
Many As species
exist – the
inorganic As
species are known
to be toxic and
most organic
species are
relatively harmless
to humans.
The potential
toxicity of some
species, such as
the huge variety of
arsenosugars, has
not yet been
established.
Less-Toxic
11/15/2016
Agilent ICP-MS Systems
40
Certified urine: NIES CRM No.18
Certified urine : NIES CRM No.18, diluted 10 times
Retention Time / min0.0 2.0 4.0 6.0 8.0 10.0 12.0
2000
6000
10000
14000
18000
22000
26000
30000
34000
1/10 NIES No.18
1/10 NIES No.18 + 1ug/L
1/10 NIES No.18 + 5ug/L
Sig
na
l / co
un
ts
Found mg/L Found2 mg/L Recovery1 % Found3 mg/L Recovery2 %
Total As 0.137 ± 0.011
AB 0.069 ± 0.012 0.0658 0.0758 100.0 0.1164 100.5
DMAA 0.036 ± 0.009 0.0313 0.0407 98.5 0.0817 100.5
As(III) 0.0025 0.0123 98.4 0.0486 92.6
MMAA 0.0026 0.0131 104.0 0.0518 98.5
As(V) 0.0024 0.0121 97.6 0.0526 100.4
Found2: 1.0 µg/L each STD spiked 10 time dilution urine CRM
Found3: 5.0 µg/L each STD spiked 10 time dilution urine CRM
Certified Value
November 2016
Spectroscopy Seminar Greece
Page 41
Agilent ICP-MS Software Development
ChemStation
MassHunter 1
MassHunter 2
MassHunter 3
MassHunter 4.x
November 2016
Spectroscopy Seminar Greece
Page 42
Method Wizard – Automatic ModeCreate a new method with just a few clicks
November 2016
Spectroscopy Seminar Greece
43
1. Click Automatic mode
2. Click on required elements
from the periodic table
3. Select Reference Stock
Solution from a list of
available standard mixes.
Method Wizard – Automatic Mode (2)
November 2016
Spectroscopy Seminar Greece
44
1. Specify locations of Blank,
typical sample and Tune
solution
2. Choose fastest speed or
lowest possible DLs
3. Click and
MassHunter builds, tests
and runs your method.
Agilent ICP-MS Applications
November 2016
Spectroscopy Seminar Greece
45
Controlling InterferencesComparing ICP-QMS and ICP-QQQ
November 2016
Spectroscopy Seminar Greece
46
Works at least as well as Agilent ICP-QMS in He mode
Also works superbly with reactive cell gases:
• Double mass filter (MS/MS)
• Q1 allows only single m/z to enter cell
• Ions in the cell are independent of matrix
• Reactions are controlled and consistent even with unusual or variable samples
Market-leading He mode reduces polyatomic interferences effectively
Can use reactive gas, but reaction mode has limitations in ICP-QMS:
• Single mass filter, positioned after cell
• No mass selection before cell
• Ions in cell vary with sample composition
• Reaction chemistry varies, so ions passed to detector vary
Agilent
ICP-QMS
Agilent
ICP-QQQ
November 2016
Spectroscopy Seminar Greece
47
8900 ICP-QQQ Key Product Hardware
Low flow
sample
introduction
system
High matrix
introduction
(UHMI)
technology
Fast, frequency-
matching 27MHz
RF generator Efficient twin-turbo
vacuum system
Dual conical Extraction and
Omega lens focuses ions
across the mass range
Up to 11 orders
dynamic range
electron
multiplier (EM)
detector
Analyzer quad Q2:
High frequency
hyperbolic quadrupole
– selects ions that
pass to detector
High-transmission,
matrix tolerant interface
First quad Q1: High frequency
hyperbolic quadrupole mass filter –
selects ions that enter the cell
4th generation collision/
reaction cell (ORS4)
with 4 cell gas lines
Peltier-
cooled
spray
chamber
Robust, high-temperature
plasma ion source
Agilent ICP-QQQ Features
November 2016
Spectroscopy Seminar Greece
48
MS/MS reaction cell having high selectivity• MS/MS reaction cell removes the most challenging interferences.
• S, Si and P can be quantified below – 50ppt
High sensitivity and Low noise• Sensitivity of Agilent 8900 is twice the Agilent7900, 700Mhz/ppm for
Yttrium is spec. !!
• Random noise is extremely low, < 0.2cps.
Excellent Abundance Sensitivity• MS/MS scan achieves < 1x10-10 abundance sensitivity.
High matrix tolerance and Reliability • never compromises in matrix tolerance and reliability established with
Agilent 7900 series ICP-MS.
Summary
8900 ICP-QQQ – High-end ICP-MS
• Unique power of MS/MS
• For difficult interference problems
• Maximum sensitivity and flexibility
7800 ICP-MS – ICP-MS New Standard
• HMI – Proven robust performance
• Proven He mode performance
• 10 orders wide dynamic range
7900 ICP-MS – Class by itself
• Highest performance in a Single Quad ICP-MS
• UHMI – up to 25% TDS
• Fast TRA mode for Single nanoparticle research
November 2016
Spectroscopy Seminar Greece
49