20
2008 [HALL PROBE GRADIOMETRY ] A novel Scanning Hall probe gradiometer has been developed and a new method to image x, y & z components of the magnetic field on the sample surface has been demonstrated for the first time with 1μm spatial resolution. 3D field distribution of a Hard Disk sample is successfully measured at 77K using this novel approach to prove the concept.

HALL PROBE GRADIOMETRY · 2019. 10. 14. · [HALL PROBE GRADIOMETRY ] A novel Scanning Hall probe gradiometer has been developed and a new method to image x, y & z components of the

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

  • 2008   

    [HALL PROBE GRADIOMETRY ]

    A novel Scanning Hall probe gradiometer has been developed and a new method to image x, y & z components of the magnetic  field on the sample surface has been demonstrated  for the first  time  with  1μm  spatial  resolution.    3D  field  distribution  of  a  Hard  Disk  sample  is successfully measured at 77K using this novel approach to prove the concept. 

  • www.nanomagnetics‐inst.com  2

    1.1 Novel Hall Probe Gradiometry

    The gradiometry in the general sense describes the method used to measure the

    variation of a physical quantity in space. In our case it is the magnetic field variation

    over the sample. From Figure 1 it can be seen that if a different voltage value is

    measured through terminals V1-2 and V3-4 due to the presence of an external magnetic

    field up on the Hall array, this indicates a field variation along the current axis. The

    magnitude of the voltage difference can be used to quantify this variation. This is a one

    dimensional first order gradiometer configuration. Number of terminals can be

    increased but the separation between them should be as much as the width of the Hall

    crosses.

    Figure 1: Hall gradiometer.

    The method used in this study, however, is completely different than the methods

    described above. There is no need for the complex design principles or the growth

    requirements. Any semiconducting material suitable for a Hall effect device is

    applicable with a simple fabrication process. The spatial resolution is just defined by the

    width of the Hall cross, thus the limitation only comes from the fabrication capability or

    from the material properties, such as the surface charge depletion after certain size. In

    fact, probes with ~50nm spatial resolution has already been shown [1], which can

    readily be employed to measure the three dimensional magnetic fields using that

    method.

    2w 

    w  IH 

    V1 

    V2 

    V3 

    V4 

  • www.nanomagnetics‐inst.com  3

    The low noise amplifier is modified such that, the current and voltage signals can be

    passed through any four leads of the Hall cross. The restriction of having the current and

    the voltage leads to be mutually perpendicular to each other is eliminated. If the current

    flow does not follow a straight path as shown in Fig. 2, we still observe the Hall effect,

    which we may call the bending resistance [2-3]. As the current flows at any particular

    point along the path, the magnetic force FM is perpendicular to the direction of motion.

    In uniform magnetic field however, the Hall voltage would be equal to zero. But, if the

    magnetic field distribution across the device is spatially non-uniform and

    nonsymmetrical along the gradient axis, there will be a finite voltage developed at the

    voltage terminals. The lateral electric field generated to balance the magnetic force is

    not uniform across the device; in other words, it has a gradient along the device. As a

    result, the measured potential is the derivative of the perpendicular magnetic field in the

    active region of the Hall cross. Depending on the current flow path the derivative about

    to different diagonals of the Hall cross is measured as shown in Fig. 3 with the non-

    perpendicular current and voltage leads with different configurations and the field

    gradient axis. This has been accidentally observed by Bath University group [4] during

    an SHPM experiment, where the voltage and current leads were mixed up. The

    orientation of the sensor over the magnetic field distribution of the measured sample is

    important. In our setup the probe has a fixed position with respect to the quadrants of

    the scanner piezo. Thus, the orientation of the probe is not changing during the scan

    which is placed by 45º with respect to the X&Y scan directions (Fig.4). On the other

    hand the sample can be aligned in any desired rotational angle with respect to our Hall

    sensor. In the experiments a hard disk sample is imaged with a 1µm size PHEMT Hall

    probe in LN2 environment to minimize the thermal drift. First, an image is obtained

    using the normal Hall sensor configuration, where current and voltage leads were

    perpendicular to each other.

  • www.nanomagnetics‐inst.com  4

    Figure 2: Hall effect with non-perpendicular current and voltage leads with different

    configurations under uniform magnetic field.

    1

    4

    3

    FL

    B

    IH

    VG=0 

    .

    ‐‐

    ‐‐

    1

    4

    3

    FL

    IH

    VG=0 

    .

    ‐‐

    ‐‐

  • www.nanomagnetics‐inst.com  5

    Figure 3: Hall effect with non-perpendicular current and voltage leads with different

    configurations under non-uniform magnetic field.

    1

    3

    FL

    BIH 

    VG=∂Bz(x,y)/∂y

    .+

    ‐‐

    x B

    +

    4

    3

    FLB

    IH 

    VG=∂Bz(x,y)/∂x. ‐

    ++

    x B

    1

    +

    ‐ 

    Gradient Axis  

    Gradient Axis  

  • www.nanomagnetics‐inst.com  6

    The scan performed with a speed of 5µm/s at 256 pixel resolution, applying a 500µA

    DC drive current to the probe in the AFM tracking mode. Figure 5 shows the SHPM

    image of the Hard Disk sample resolving individual bits. The Hall sensor alignment and

    the lead configurations are given at the right hand side of the figure.

    Figure 4: Orientation of the Hall sensor, over the sample, with respect to the scan

    direction of the microscope: Hall cross has 45° alignment.

    N

    S

    W  E

    Y scan direction 

    X scan direction

  • www.nanomagnetics‐inst.com  7

    Figure 5: SHPM scan of a Hard disk sample at 77K with normal Hall sensor

    configuration of mutually perpendicular current and voltage leads. Hall cross diagram

    shows the relative alignment of the probe over the sample and the leads’ positions.

    The magnitude of the Hall voltage signal is stored in a 256×256 matrix. Hence, matrix

    arithmetic operations can be applied on the acquired image. Equipped with these, the

    rows are differentiated along the y direction to obtain ∂Bz/∂x, and columns are

    differentiated to get ∂Bz/∂y along the x direction using forward differences method. The

    results are shown in Fig 6 and Fig 7 which are calculated ∂Bz/∂x and ∂Bz/∂y

    respectively. Immediately after obtaining the standard SHPM image, drive current

    direction is changed to the proper leads so that the ∂Bz/∂x and ∂Bz/∂y images can be

    obtained. To have a healthy comparison, the probe was not pulled back off the surface

    while the switch positions on the amplifier are changed. The Hall probe was driven by

    smaller, 100µA current, as it was not possible to properly offset null the probe. The

    other scan parameters are held the same as in the case of normal scan. Images obtained

    as a result of these experiments and the current/ voltage lead configurations are shown

    in Fig. 8 and 9. The results are in perfect agreement with the calculated ones, except the

    field values are not absolute since the probe is offset nulled at the presence of the field.

    I+V+ 

    V‐I‐ 

  • www.nanomagnetics‐inst.com  8

    Figure 6: Calculated ∂Bz/∂x from the measured Bz(x,y) data matrix by forward

    differences.

    Figure 7: Calculated ∂Bz/∂y from the measured Bz(x,y) data matrix by forward

    differences.

  • www.nanomagnetics‐inst.com  9

    Figure 8: SHPM scan of a HDD at 77K with Hall sensor configuration shown in the

    diagram. The image represents ∂Bz/∂x due to the relative positions of current and

    voltage leads

    Figure 9: SHPM scan of a HDD at 77K with Hall sensor configuration shown in the

    diagram. The image represents ∂Bz/∂y due to the relative positions of current and

    voltage leads.

    V+ı+ 

    V‐I‐ 

    I+I‐ 

    V+V‐ 

  • www.nanomagnetics‐inst.com  10

    1.2 Three Dimensional Magnetic Field Imaging

    As different sample orientations are scanned, different images are obtained in lateral

    derivative, which was actually expected as the gradient of the field changes in the space.

    One may argue that since we can calculate the local gradient of the Bz(x,y) along x and

    y directions, why should we bother imaging in this obscure way? On the other hand this

    measurement mode clearly gives us the local gradient of the Bz(x,y), which can be used

    in a number of applications like gradiometry etc.

    A more important ramification is the possibility of calculating in-plane components,

    Bx(x,y) and By(x,y) of the magnetic field if we measure Bz(x,y), ∂Bz(x,y)/∂x and

    ∂Bz(x,y)/∂y across the space. If we start with the Maxwell equation derived from

    Ampere’s law [5],

    B ε µ J ε µ E (1)

    In a source free region the equation simplifies to,

    B 0 (2) which can be written, in open form, as

    ̂ ̂ B ̂ B ̂ B 0 (3)

    from which we can get,

    B B ̂ B B ̂ B B 0 (4)

    To satisfy the equation each vector component must individually be equal to zero.

  • www.nanomagnetics‐inst.com  11

    B B 0 (5) B B 0 (6) B B 0 (7)

    The first two equations can be solved in terms of the parameters we measure at the

    beginning of the problem. Hence we can write,

    B B 0 (8)

      B B (9) B B dz (10)

    Hence if we measure or obtain ∂Bz/∂y as a function of z, then we can calculate the By at

    this specific z. The SHPM data should be obtained at increasing sample-sensor

    distances, until the signal decays to zero or below the noise levels. A similar equation

    can be written for Bx as well,

    B x, y, z B , , dz (11)

    As, it is not possible to integrate the equations analytically, we have to compute them

    numerically from the measured data.

    z dz ∑ (12)

  • www.nanomagnetics‐inst.com  12

    The situation can visualized with the aid of Fig. 10 given below. While we can measure

    the quantitative value of the perpendicular magnetic field, Bz, at the closest sample-

    sensor position, this is not exactly the case for the derivatives. The change in magnitude

    at a particular point for, ∂Bz(x,y,z)/∂x and ∂Bz(x,y,z)/∂y is not necessarily a decrease

    with increasing sample-sensor separation. However, the overall dynamic field will

    decrease with increasing probe-sample separation.

    Note that we do not need to have an infinite sum as the field value f(jh) will decay after

    a certain value of jh that can be set as the finite upper limit n. Thus, ∂Bz(x,y,z)/∂x and

    ∂Bz(x,y,z)/∂y are acquired at different heights (zi) until the field decays to zero,

    carefully recording the values. A fixed incremental separation h is used between the

    scans for ease of calculation.

    Figure 10: Visualization of incremental scan for Bz(x,y,z), ∂Bz(x,y,z)/∂x and

    ∂Bz(x,y,z)/∂y

    While the sample is scanned the surface is never perfectly smooth. In addition to this,

    some inclination angle can unintentionally be given, while mounting the sample. For

    this reason, while the sample is scanned with a liftoff, this distance must be applied

    throughout the sample evenly. The magnetic force microscopy (MFM) techniques,

    where the magnetic and interaction forces have to be separated, can be applied to this

    h1 

    h0 

    h2 

    h3 

    hn z

    x

    y

  • www.nanomagnetics‐inst.com  13

    case. Hence, while the forward scan is conducted using the feedback following the

    surface texture; the backward scan is performed following the same texture, however

    this time, by adding the required lift-off value at each pixels. By this procedure, the

    value of h is held fixed throughout the whole scan area. Fig 11 shows the image of Hard

    disk with the forward scan, (a) while the probe is in closest proximity of the sample, and

    backward scan (b) while the probe is lifted off by 3.5µm. The decay of the magnetic

    signal when the probe is lifted off by 3.5µm away from the surface can clearly be seen

    and obviously this distance can be accepted as the upper limit of the integration while

    calculating the Bx and By. This value is within the range of retraction capability of the

    used scanner piezo at room temperature. But, unfortunately, the effect of the thermal

    drift is also high at room temperature and it is virtually impossible to scan the same

    portion of the HDD by giving successive offsets, which is required to integrate ∂Bz/∂x

    or ∂Bz/∂y. Fig 12 shows this effect. At each level of 0.5µm offset, three images, Bz,

    ∂Bz/∂x and ∂Bz/∂y, had to be recorded. Each image requires ~40 minutes to be scanned.

    Hence, approximately two hours spent at each level. As a result, the experiments had to

    be done at low temperature to eliminate the thermal drift, which unfortunately reduces

    the scan area of the piezo tube in X, Y and Z directions. To satisfy both requirements,

    the stick-slip coarse approach mechanism is used to successively pull the Hall sensor off

    the sample. The motion of the puck is first calibrated for a given voltage pulse. Then

    appropriate numbers of pulses were applied in order to lift-off the sensor from the

    sample. We adjusted the voltage pulses applied to the piezo to provide 250nm backward

    steps. Overall 26 scans are performed, starting from the closest proximity of the sample,

    with 250nm steps, until the probe is 6.5µm away from the sample in height. At each

    height level, Bz(x,y), ∂Bz(x,y)/∂x and ∂Bz(x,y)/∂y were imaged. The thermal drift is

    insignificant at 77K as evident in Fig 13. ∂Bz/∂x and ∂Bz/∂y images are used to integrate

    Bx and By respectively using the method described above. Figures 14-16 show the Bz,

    By and Bx fields of an Hard disk sample.

  • Figure 11

    and (b) ba

    STM feedb

    voltage ap

    scan.

    : SHPM im

    ackward sca

    back. Scan

    pplied to th

    (

    www.na

    mage of Hard

    an (3.5µm a

    speed was

    he sample a

    a) 

    anomagneti

    d disk samp

    away from t

    5µm/s, reso

    and the tunn

    cs‐inst.com

    ple, (a) forw

    the surface)

    olution set t

    neling curre

    ward scan (i

    . Both imag

    to 256 × 25

    ent of 1nA m

    in the tunne

    ges were ob

    56 pixels, -1

    maintained

    (b) 

    14

    eling range)

    btained with

    100mV bias

    d during the

    )

    h

    s

    e

  • Figure 12

    effect of t

    with a hei

    not seen a

    time passe

    images we

    5µm/s, res

    and the tu

    (

    (

    2: SHPM im

    the thermal

    ight of (a) 0

    as the offset

    ed between

    ere obtained

    solution set

    nneling cur

    www.na

    a) 

    c) 

    mages of Ha

    drift. Pictu

    0.5µm, (b) 1

    t is only give

    the shown

    d with STM

    t to 256 ×

    rrent of 1nA

    anomagneti

    ard disk sam

    ures are the

    1.5µm, (c) 2

    en during th

    images. Th

    M feedback u

    256 pixels,

    A maintained

    cs‐inst.com

    mple obtain

    e forward s

    2.5µm, (d) 3

    he backwar

    he shift is to

    using a 1µm

    -100mV bi

    d during the

    (

    (

    ned at room

    scans of pr

    3.5µm respe

    rd scan. App

    owards the b

    m PHEMT s

    ias voltage

    e scan.

    (b) 

    (d) 

    m temperatu

    robe above

    ectively. Fie

    proximately

    bottom left

    ensor. Scan

    applied to

    15

    ure showing

    the sample

    eld decay is

    y 4 hours of

    corner. All

    n speed was

    the sample

    g

    e

    s

    f

    l

    s

    e

  • Figure 13

    thermal dr

    (a) 0.25 µ

    passed be

    insignifica

    sensor. Sc

    magnetic f

    (

    (

    3: SHPM im

    drift. Picture

    µm, (b) 1.25

    tween the f

    ant after the

    can speed w

    field can als

    www.na

    a) 

    c) 

    mages of Ha

    es are the s

    5µm, (c) 2.5

    first and the

    ermal stabil

    was 5µm/s,

    so be seen a

    anomagneti

    Hard disk sa

    scans of pro

    5µm, (d) 3.

    e last image

    lization. All

    resolution

    as the senso

    cs‐inst.com

    ample obtain

    obe levels a

    75µm resp

    es. There is

    l images we

    set to 256

    ors moves aw

    (

    (

    ned at 77K

    above the sa

    ectively. Ap

    a very littl

    ere obtained

    6 × 256 pix

    way from th

    (b) 

    (d) 

    K showing ef

    ample with

    pproximatel

    le drift whic

    d using a 1µ

    xels. The de

    he sample.

    16

    effect of the

    a height of

    ly 10 hours

    ch becomes

    µm PHEMT

    ecay of the

    e

    f

    s

    s

    T

    e

  • www.nanomagnetics‐inst.com  17

    (a) 

    (b)  (c) 

    Figure 14: SHPM image of Hard disk sample, shows the Bz of the field (a), obtained at

    77K in the feedback tracking zone. Image was obtained using a 1µm PHEMT sensor.

    Scan speed was 5µm/s, resolution set to 256×256 pixels. ∂Bz(x,y)/∂x (b) and ∂Bz(x,y)/∂y

    (c) are calculated from Bz image by differentiating rows and columns of the image

    matrix respectively.

  • www.nanomagnetics‐inst.com  18

    Figure 15: By field calculated by integrating ∂Bz/∂y over a finite range. Each image file used in calculation was obtained using the same 1µm PHEMT sensor. Scan speed

    was 5µm/s, resolution set to 256×256 pixels. The increments along the z-direction, h, is

    set to 0.25μm in the range of [0, 6.5μm].

  • www.nanomagnetics‐inst.com  19

    Figure 16: Bx field calculated by integrating ∂Bz/∂x over a finite range. Each image file

    used in calculation was obtained using the same 1µm PHEMT sensor. Scan speed was

    5µm/s, resolution set to 256×256 pixels. The increments along the z-direction, h, is set

    to 0.25μm in the range of [0, 6.5μm].

  • www.nanomagnetics‐inst.com  20

    1. Sandhu, A., et al., 50 nm Hall sensors for room temperature scanning Hall probe microscopy. Japanese Journal of Applied Physics Part 1-Regular Papers Short Notes & Review Papers, 2004. 43(2): p. 777-778.

    2. Grundler, D., et al., Bend-resistance nanomagnetometry: spatially resolved magnetization studies in a ferromagnet/semiconductor hybrid structure. Physica E-Low-Dimensional Systems & Nanostructures, 2002. 12(1-4): p. 248-251.

    3. Peeters, F.M. and X.Q. Li, Hall magnetometer in the ballistic regime. Applied Physics Letters, 1998. 72(5): p. 572-574.

    4. Oral, A., unpublished. 5. Jackson, J.D., Classical Electrodynamics. 3rd ed. 1999, New York: Wiley. xxi,

    808 p.