Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
1 | P a g e
Goal: Develop a system that enables 19 or fewer vials to be utilized for protocol development. Theory: A small batch will freeze dry faster than a large batch in a lab freeze dryer. First Experimental Set: A full tray in a lab freeze dryer and then run 19 vials in the same dryer using the same cycle (using CM-Pirani convergence to indicate end of primary drying).
Run a REVO with a full tray of vials and time Experimental Design 3ml fill in a 10 ml vial Water Freezing step ramp 0.5C/min to-40C, hold 60 mins Primary drying shelf set point +20C Vacuum set point 100 mT
Run a REVO with a 19 vials and time Experimental Design 3ml fill in a 10 ml vial Water Freezing step ramp 0.5C/min to -40C, hold 60 mins Primary drying shelf set point +20C Vacuum set point 100 mT
REVO Full Tray 636 minutes in primary
drying
REVO 19 Vials 512 minutes in primary
drying
0
50
100
150
200
250
-50
-40
-30
-20
-10
0
10
20
30
0 100 200 300 400 500 600Time (Minutes)
REVO 19VAC-PREVO 19VAC-CREVO FULLTRAY VAC-PREVO FullTray VAC-C
Full tray CM and Pirani convergence
Results REVO full tray vs 19 vials
Experimental Group # 1
Water
Full tray in REVO
Vs 19 Vials in
REVO
19 vial CM and Pirani convergence
636 minutes is the target for all other water runs
Results
19 vials dry quicker than
a full tray based on
end of primary
drying being detected by CM vs Pirani convergence
Results: 19 Vials Take less time. Why? Assumption: Radiant energy has a significant effect on the freeze drying process via edge vial effect. Smaller batch runs are more susceptible to this effect because a larger percentage of the vials are “edge effect vials”.
2 | P a g e
Experimental Group # 1
5% Sucrose
Full tray in
REVO
Run a REVO with a full tray of vials and time Experimental Design 2ml fill in a 10 ml vial 5% sucrose by weight Freezing step -10 hold 30 mins, -40C hold 120 mins Primary drying shelf set point -20C Vacuum set point 60 mT
REVO Full Tray 2026 minutes in primary
drying
-200
-150
-100
-50
0
50
100
150
200
-50
-40
-30
-20
-10
0
10
20
30
40
0
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
1900
2000
2100
2200
REVO2ml fill in 10ml vial
5% sucrose1 tray filled
TP1 TP2 TP3 TP4 TP5 TP6 TP7
TP8 TP9 TP10 TSHELF_S Vacuum-P Vacuum-c
Results
Full tray took 2026
mins based on end of primary
drying being detected by CM vs Pirani convergence
2026 minutes is the target for all other sucrose runs
3 | P a g e
Theory: Eliminate radiant energy, and thus edge effect vials, and your small batch runs will be more like the larger batch runs.
Second Experimental Set: Eliminate radiant energy from the freeze dryer when using 19 vials. Build a small freeze dryer with the ability to control wall temperature. Note primary drying time.
Made controlled temperature walls. Also tried adding commercial insulation and aluminum shield to minimize edge effect.
ALPHA design
Run ALPHA with a 19 vials and wall at +20C Experimental Design 3ml fill in a 10 ml vial Water Freezing step ramp at 0.5C/min to -40C, hold 30 mins Primary drying shelf set point 20C Vacuum set point 100 mT
Run ALPHA with a 19 vials and wall at -20C Experimental Design 3ml fill in a 10 ml vial Water Freezing step ramp at 0.5c/min to -40C, hold 30 mins Primary drying shelf set point 20C Vacuum set point 100 mT
ALPHA with +20C Walls 532 minutes in primary
drying. (reminder: target is 636 mins)
ALPHA with -20C Walls 557 minutes in primary
drying (reminder: target is 636 mins)
Experimental Group #2
Water
Eliminate
radiant energy to minimize edge effect
vials by cooling and/or
shielding the chamber walls of the freeze
dryer.
Results
The minimization
of radiant energy does
not eliminate edge effect.
4 | P a g e
Run ALPHA with a 19 vials and wall at +20C Experimental Design 2ml fill in a 10 ml vial 5% sucrose by weight Freezing step -10 hold 30 mins, -40C hold 120 mins Primary drying shelf set point -20C Vacuum set point 60 mT
ALPHA, wall at +20C 789 minutes in primary
drying
-200
-150
-100
-50
0
50
100
150
200
-50
-40
-30
-20
-10
0
10
20
30
40
0 30 60 90 120
150
180
210
240
270
300
330
360
390
420
450
480
510
540
570
600
630
660
690
720
750
780
810
840
870
900
930
960
990
1020
1050
1080
Alpha UnitWalls at +20C
2 ml in 10 ml vial5% sucrose
TP1 TP2 TP3 TP4 TP5 TP6 TSHELF_S Vacuum-P Vacuum-c
2
2 3 5 4
6
1
Array and thermocouple placement in
Alpha
Experimental Group #2
5% Sucrose
Eliminate
radiant energy to minimize edge effect
vials by cooling and/or
shielding the chamber walls of the freeze
dryer.
5 | P a g e
ALPHA, wall at -20C 1096 minutes in primary
drying
-200
-150
-100
-50
0
50
100
150
200
-50
-40
-30
-20
-10
0
10
20
30
40
0 30 60 90 120
150
180
210
240
270
300
330
360
390
420
450
480
510
540
570
600
630
660
690
720
750
780
810
840
870
900
930
960
990
1,02
01,
050
1,08
01,
110
1,14
01,
170
1,20
01,
230
1,26
01,
290
1,32
01,
350
1,38
0
Alpha Wall at -20C2 ml fill in 10 ml vial
5% sucrose
TP1 TP2 TP3 TP4 TP5 TP6 TSHELF_S Vacuum-P Vacuum-c
Pres
sure
(mT)
Tem
pera
ture
(C)
Results: Cooling the chamber wall to -20C appeared to make some difference but the elimination of radiant energy does not appear to be the answer to simulating a larger freeze dryer. Why?
Run ALPHA with a 19 vials and wall at -20C Experimental Design 2ml fill in a 10 ml vial 5% sucrose by weight Freezing step -10 hold 30 mins, -40C hold 120 mins Primary drying shelf set point -20C Vacuum set point 60 mT
Experimental Group #2
5% Sucrose Continued
6 | P a g e
Theory: Eliminating radiant energy did not appear to eliminate edge effect in the 19 vial array. Taking a closer look at how the individual vials are responding may help us understand this further. The common perception is that when we eliminate radiant energy the edge effect should be eliminated.
Third Experimental Set: Utilizing the gravemetric measurement method determine the rate of freeze drying by various vials
within the array. Do perimeter vials freeze dry differently than center vials even minimal radiant energy is present?
Run ALPHA with a 19 vials and wall at -40C Experimental Design 3ml fill in a 10 ml vial Water Freezing step at 0.5 to -40C, hold 30 mins Primary drying shelf set point +20C Vacuum set point 100 mT
Run ALPHA with a 19 vials and wall at -40C with aluminum shield and additional insulation. (10+ runs conducted) Experimental Design 3ml fill in a 10 ml vial Water Freezing step at 0.5 to -40C, hold 30 mins Primary drying shelf set point +20C Vacuum set point 100 mT
% dry by weight 25% of the way
into primary drying.
33 37 39
34 36
29 34 35
33 35
44 34 38
32 29
27
28 33
25
Experimental Group #3
Water
Run systems with wall at -40C. Study the rate of
sublimation in individual vials
in a freeze dryer that has
eliminated radiant energy to gain a better perspective on
what is happening in
the dryer.
31 37 36
34 32
27 44 36
38 37
33 31 34
32 29
26
27 30
25
Results: Edge effect NOT eliminated by lowering the temperature of the chamber wall. If the presumption that radiant energy caused edge effect and radiant energy is eliminated, then edge effect should be eliminated. This experiment does not verify this
assumption. Gravemetric studies indicate that edge vials are drying faster than the other vials, even when there is minimal radiant energy
Results
Low temperature
chamber walls does not eliminate
“edge effect”
Std Dev 4.9 Std Dev 4.4
7 | P a g e
Run ALPHA with a 19 vials and wall at -40C
Experimental Design 2ml fill in a 10 ml vial 5% sucrose by weight Freezing step -10 hold 30 mins, -40C hold 120 mins Primary drying shelf set point -20C Vacuum set point 60 mT
ALPHA, wall at -40C 1042 minutes in primary
drying
-200
-150
-100
-50
0
50
100
150
200
-50
-40
-30
-20
-10
0
10
20
30
0 30 60 90 120
150
180
210
240
270
300
330
360
390
420
450
480
510
540
570
600
630
660
690
720
750
780
810
840
870
900
930
960
990
1020
1050
1080
1110
1140
1170
1200
1230
1260
1290
Alpha Wall at -40C2 ml fill in 10 ml vial
5% sucrose
TP1 TP2 TP3 TP4 TP5 TP6 TSHELF_S Vacuum-P Vacuum-c
Experimental Group #3
5% Sucrose
Run systems with wall at -40C. Study the rate of
sublimation in individual vials
in a freeze dryer that has
eliminated radiant energy to gain a better perspective on
what is happening in
the dryer.
Results
Low temperature
chamber walls does not eliminate
“edge effect” as evidence by run time not approaching
REVO run time of 2026 mins
Gravemetric tests not yet conducted.
Warning: At -40C in these conditions the wall starts behaving
like an internal condenser
8 | P a g e
Theory: The product vials have a thermal effect on one another (essentially acting as cold sinks and hot sinks). Simulate adjacent vials by directly contacting the edge vials and controlling the temperature.
Fourth Experimental Set: Find a way to contact the cold temperature walls to the outside edge vials. If “edge effect” is eliminated this indicates that vial to vial contact has a significant effect in “edge effect”.
No longer using the Alpha unit. Designed and build the MicroFD with
“thermal simulator” temperature controlled blocks.
Run MicroFD with a 19 vials and wall at -40C Experimental Design 3ml fill in a 10 ml vial Water Freezing step ramp 0.5 to -40C, hold 30 mins Primary drying shelf set point +20C Vacuum set point 100 mT
MicroFD with simulator blocks (aka LyoSim™) 633 minutes in primary drying
ONLY 3 MINUTES DIFFERENT THAN THE FULL TRAY RUN DONE IN THE REVO
EDGE EFFECT ELIMINATED ALL VIALS NOW BEHAVE LIKE CENTER VIALS
The use of thermal simulator blocks in the MicroFD created an environment where the sublimation rate was very similar (essentially the same) in every vial. The elimination of edge effect indicates the MicroFD
can be utilized for protocol development for larger freeze dryers based on its ability to simulate center vials.
Experimental Group # 4
Water
Adjacent vials
affect the thermal
characteristics of one another. Since edge vials
do not have the same
number of adjacent vials
they sublimate differently.
Have the controlled
temperature wall come in contact with the vials to
eliminate edge effect.
2 22
22
2 2 22
2 2
2 2 2
2
2
23 23
25
2
2
25 2
26 24 26
25
26
24 25 24
23
22
25
24 25
25
26
Std Dev 4.4
9 | P a g e
(see specific experiments for additional details)
Experiment #
Product Conditions (see experiments for details)
Time Vial variation in sublimation rate
1 Water REVO full tray 636 mins Not studied
1 Water REVO 19 vials 512 mins High disparity, data not reported here
2 Water Alpha 19 vials +20C walls 532 min High disparity, data not reported here
2 Water Alpha 19 vials -20C walls 557 mins High disparity, data not reported here
3 Water Alpha 19 with -40C wall Not timed High disparity, edge effect seen (see pg 3)
3 Water
Alpha 19 vials with -40C wall and additional radiant energy
shielding
Not timed High disparity, edge effect seen (see pg 3)
4 Water MicroFD with LyoSim (thermal sink emulators
touching vials)
633 mins (successfully
simulates REVO run with full tray
Variation in vials minimized, edge effect eliminated
(see pg 4)
Summary
Water
Patent # 9121637 Accuflux Heat flux sensor in a freeze dryer LyoSim Patent Pending
10 | P a g e
Additional & Subsequent Studies. Big Pharma Collaboration: As part of our on-going work we conducted a blind study with a large pharma group. They created a cycle in a SP Scientific LyoStar. They supplied us with vials and product that had a critical temperature of -24C. Within 3 runs we were able to fine tune the Micro FD to simulate their LyoStar freeze dryer run to their satisfaction. They have subsequently purchased a unit for further study.
Experiment #
Product Conditions (see experiments for details)
Time Vial variation in sublimation rate
1 5% Sucrose
REVO full tray 2026 mins Not yet studied
2 5% Sucrose
Alpha 19 vials +20C walls 789 mins Not yet studied
2 5% Sucrose
Alpha 19 vials -20C walls 1096 mins Not yet studied
3 5% Sucrose
Alpha 19 with -40C wall 1042 mins Not yet studied
4 5% Sucrose
MicroFD with LyoSim (thermal sink emulators
touching vials)
Not yet studied Not yet studied
Product Experiments
with customer
Summary
5% Sucrose