Go to ASL
Last edited: 30-Sep-101
IngMar Medical, Ltd.Respiratory Simulation Specialist
TutorialActive Servo Lung 5000
(ASL 5000) High Fidelity
Adult/Pediatric/NeonatalBreathing Simulator
IngMar Medical, Ltd.P.O. Box 10106Pittsburgh, PA 15232T: 800.583.9910 or 412.441.8228www.ingmarmed.com
2
Go to ASL
Last edited: 30-Sep-102
Tutorial Overview
ASL 5000™
Active Servo Lung
This tutorial is being provided with gratitude to our valued customers to enhance your experience with the ASL 5000.
It is not intended to replace the User’s Manual, but rather to complement it. Most of the information contained in this tutorial can also be found in the User’s Manual. However, some additional detail has been added to the tutorial in order to provide a more complete step-by-step process.
Authors: Cathy Brennan, Jenn Bacior, Stefan Frembgen
3
Go to ASL
Last edited: 30-Sep-103
The tutorial is designed to offer additional guidance in preparing and running respiratory simulations with the IngMar Medical ASL 5000.
You may choose to follow along slide by slide. Or you may prefer to skip to those slides that appear to be most relevant to your use of the ASL 5000.
Hyperlinks have been provided to facilitate moving quickly to the slides of interest to you. (Please refer to the Table of Contents for a detailed listing of the tutorial contents.)
Tutorial Overview
4
Go to ASL
Last edited: 30-Sep-104
In addition to hyperlinks, the tutorial takes advantage of other PowerPoint features such as Action Buttons and Action Settings. Clicking on these buttons will take you to other slides in the tutorial.
Some examples are:
Tutorial Overview
Review slide #8: Synchronizing
with the ASL 5000
Returns you to the last slide viewed.
Takes you to the next slide.
Hyperlinked to the indicated slide.
5
Go to ASL
Last edited: 30-Sep-105
Tutorial Overview
Sample Cases
ASL 5000
Included with the tutorial are several sample case files (in addition to scripts in the scenarios folder). These files can be easily loaded into the ASL 5000 folder on your computer. By doing so, you will have a head start in setting up your simulation models.
Please see the ASL Sample Cases slides later in this tutorial for instructions on loading the sample case files into your ASL 5000 directory.)
6
Go to ASL
Last edited: 30-Sep-106
Slide Number:
2 – 5 Tutorial Overview
6 – 7 Table of Contents
8 Synchronizing the ASL
9 – 17 Opening the ASL Software
Creating a Simulation Model
19 – 34 The Script
35 – 39 Step 2: Choose a Lung Model
40 – 44 Step 3: Choose a Chest Wall Model
45 Step 4: Save Simulation Parameter Set
46 – 53 Running a Simulation
ASL Tutorial: Table of Contents
7
Go to ASL
Last edited: 30-Sep-107
Slide Number:
54 – 60 Adjusting the Run Time Display
61 – 73 The Interactive Control Panel
74 – 80 Time-Varying Parameters
81 – 84 Non-Linear Compliance
85 – 104 ASL Sample Cases:89 – 81 Apnea
92 – 95 Bagging a Patient
96 – 98 Compliance Step-Down
99 – 101 Overdistension
102 – 104 Trigger Challenge
105 – 107 Troubleshooting
108 – 109 Glossary of Terms
ASL Tutorial: Table of Contents
8
Go to ASL
Last edited: 30-Sep-108
Synchronizing with the ASL 5000
Turn on the ASL first.
Then, start up Host PC software application.
It is best if the ASL 5000 is started first, in order to allow proper synchronization with the computer.
Wait until the red light of the Motor Enable/Disable switch goes out (ASL has finished booting up) before starting the software.
1 2Depress the Enable Switch if red light is still on after 20 seconds.
Note:
9
Go to ASL
Last edited: 30-Sep-109
Opening the ASL 5000 software
Select the application from the Windows Start menuoropen the ASL 5000 executable, (the application, file with the IngMar Medical icon).from the installation folder (Windows Explorer).
10
Go to ASL
Last edited: 30-Sep-1010
The Welcome screen appears and offers you several choices for operation and utility.
Click on the desired mode of operation.
Note:
Stand-alone (demo) mode requires no access to the ASL device. Use the demo mode to create or modify scripts, review archived data or to practice with or train on the software.
Opening the ASL 5000 software
11
Go to ASL
Last edited: 30-Sep-1011
You may also customize graph colors (see Slide 58) or check the version of the software by clicking the About ASL 5000 key.
Opening the ASL 5000 software
12
Go to ASL
Last edited: 30-Sep-1012
The Welcome screen appears and offers you several choices for operation and utility.
Click on the desired mode of operation.
Note:
Stand-alone (demo) mode requires no access to the ASL device. Use the demo mode to create or modify scripts, review archived data or to practice with or train on the software.
Opening the ASL 5000 software
13
Go to ASL
Last edited: 30-Sep-1013
Choose “Use software with ASL Simulator” and several stacked windows will open at once.
Opening the ASL 5000 software
Note: As the ASL 5000 software opens, it begins to search for the ASL device immediately.
14
Go to ASL
Last edited: 30-Sep-1014
Notice the panel at the top of the Run Time window labeled Status.
This panel tracks the progress of the synchronization between the computer and the ASL device.
Opening the ASL 5000 software
Note: Depending on the installation, connection may only occur specifically with one serial number.
15
Go to ASL
Last edited: 30-Sep-1015
As the “Waiting for synchronization…” message appears, the iteration numbers will count upwards.
Opening the ASL 5000 software
16
Go to ASL
Last edited: 30-Sep-1016
At this point, if the ASL had not yet been turned on, you will see the dialog box at the right.
To correct this issue, switch the green On/Off power switch on the back of the ASL to the On position.
Wait until the red light of the Motor Enable Switch is off, then try again.
Opening the ASL 5000 software
The Emergency Stop switch needs to be in the depressed position for motor operation (red light OFF after ASL start-up).
DO NOT push this button unless you are trying to stop the ASL mid-operation for some reason.
17
Go to ASL
Last edited: 30-Sep-1017
Upon successful completion of the synchronization process, you will see this message in the Status panel.
Opening the ASL 5000 software
18
Go to ASL
Last edited: 30-Sep-1018
If your computer has successfully synchronized with the ASL device, you should hear a soft “click” from inside the ASL 5000 as the piston stops at the home position .
Please go on to the next slide.
If the synchronization process was not successful, consult the Troubleshooting section of this tutorial.
Opening the ASL 5000 software
19
Go to ASL
Last edited: 30-Sep-1019
Now, open the window that is labeled Script Editor.
Click on this window to activate the Script Editor function.
Creating a Simulation Model: The Script
Note: If you can’t see the Script Editor window, you can use the drop-down menu at the top of the screen to navigate from one screen to another. Click on “Windows” for a listing of all available ASL 5000 screens.
Alternatively, you can click on the ASL 5000 folder marked with the IngMar Medical icon that has been minimized and placed on the taskbar at the bottom of your screen.
20
Go to ASL
Last edited: 30-Sep-1020
In the ASL 5000: Simulation Script Editor window - Scenario Scripts tab, double-click on one of the listed scripts,
or
highlight an item and click “Choose Scenario Script File”.
Creating a Simulation Model: The Script
21
Go to ASL
Last edited: 30-Sep-1021
Alternatively:
In the ASL 5000: Simulation Script Editor window, click on the “ScriptFile” drop down menu to display its selections.
Click “Open Script.”
Creating a Simulation Model: The Script
22
Go to ASL
Last edited: 30-Sep-1022
This will bring up a file dialog box listing all of the files with the extension .sct. in the \vars - subdirectory These are the simulation scripts that have already been created.
Creating a Simulation Model: The Script
23
Go to ASL
Last edited: 30-Sep-1023
Note: It is important to distinguish between a patient parameter file (.vr3) and a script file (.sct)!
A script file is made up of one or more parameter files. The .vr3 files are the parameter sets for a simulation.
The script shown here consists of seven different .vr3 files. The file called “..\pause.vr3” is used several times for two repetitions each.
Creating a Simulation Model: The Script
24
Go to ASL
Last edited: 30-Sep-1024
Note:
Once you have created specific patient parameter sets (.vr3 files), you can call them up again and again to run in different combinations.
Different combinations of patient parameter sets run in sequence make up the simulation scripts.
For selecting and editing the patient parameter files, see slide 26
Creating a Simulation Model: The Script
25
Go to ASL
Last edited: 30-Sep-1025
If this is your first time opening the software, and you have not yet created your own scripts, click on the script labeled Bag_Patient.sct .in the ..\vars\examples subdirectory
Creating a Simulation Model: The Script
26
Go to ASL
Last edited: 30-Sep-1026
The filename and path for the script file now appear in the box in the upper part of the Script Editor window - Manual Scripting tab.
The color of this box indicates whether the opened script file has been saved or not.
The patient parameter files of the chosen script are listed in the window.
Each .vr3 file has a number of repetitions assigned to it.
Creating a Simulation Model: The Script
27
Go to ASL
Last edited: 30-Sep-10
The script has only been saved as a current.sct, for immediate use with the simulator, but not as a script with its ow name
The script has both been saved with a unique name after changes were made, and also as a current.sct, for immediate use with the simulator
27
The background color scheme of the Script File Name box can assume five colors.
A red background indicates that files listed in the script are not available under the defined name
Creating a Simulation Model: The Script
The script has not been saved at all after a change or when invoked for the first time
The script has only been saved with a unique name after changes were made, but not as a current.sct, to be used with the simulator right away
28
Go to ASL
Last edited: 30-Sep-1028
Directory Tokens are shortcuts used to label file paths. They allow easy change of path names (creating relative paths) when collections of files are moved on a PC or network.
Click “Configure Token/Path” to create or edit tokens.
Creating a Simulation Model: Directory Tokens
29
Go to ASL
Last edited: 30-Sep-1029
Creating a Simulation Model: Directory Tokens
To create a token, click “New”
An Editor window will open where you can browse for a path to create the shortcut for
30
Go to ASL
Last edited: 30-Sep-1030
Creating a Simulation Model: Directory Tokens
The actual path
can be given a meaningful token name(a configurablerelative path name)
31
Go to ASL
Last edited: 30-Sep-1031
Creating a Simulation Model: Directory Tokens
In the Relative Path Configuration window, the token is now available to replace long, absolute pathnames.
You can chooseto have a singleor all path names showing
or
to use tokennames (single file or for all parameterfiles)
32
Go to ASL
Last edited: 30-Sep-1032
You can also change the number of repetitions that will run with this parameter set during your simulation.
Double click on the second .var file in the default script (segment1.var).
The “Select Simulation Parameter Set” dialog box will appear.From this dialog box, you can also browse for a different .vr3 file to use in your script.
Creating a Simulation Model: Editing Parameter Sets
33
Go to ASL
Last edited: 30-Sep-1033
To continue with editing a vr3-file, click on the “Edit” button in the lower right corner of the Select a Parameter File dialog box.
Creating a Simulation Model: The Script
34
Go to ASL
Last edited: 30-Sep-1034
Click on the line drawing of a lung model or pump model to choose among four options:
The Lung Model
A. One compartment
B. Two compartments
C. Flow pump
D. Volume pump
AB
CD
35
Go to ASL
Last edited: 30-Sep-1035
From the Lung Model screen, set the desired Resistance and Compliance for your “patient.”
Do this either by using the up and down arrows; or by highlighting the numbers that are currently in the value boxes and by changing them to the desired values.
The Lung Model
36
Go to ASL
Last edited: 30-Sep-1036
(See slide 61 for instructions on setting time-varying parameters.)
Turn Compensations on or off by clicking on the On/Off button.
With Compensations on, the desired values may be entered to the right of the respective compensations.
Choose between Constant and Time-Varying Parameters by clicking on the button below the Time-Varying Parameters label.
The Lung Model
37
Go to ASL
Last edited: 30-Sep-1037
Choose between parabolic and linear resistance.
Click on the button under Independent Exp. Resistance if you wish to simulate an expiratory resistance that is different from inspiratory.
The Lung Model
38
Go to ASL
Last edited: 30-Sep-1038
If you are using a two-compartment model, you have the option of simulating a compliance that is non-linear.
Click on the button below the “Non-Linear Compliance” label to do so. (See slide 69 for more details on creating a non-linear compliance curve.)
The Lung Model
39
Go to ASL
Last edited: 30-Sep-1039
On the Lung Model page, you can also choose the update rate for waveform sampling.
Choices are from ultra-high resolution (512 Hz) to 64 Hz.
Click on the “Next >” button to move to Step 3, the Chest Wall Model.
The Lung Model
40
Go to ASL
Last edited: 30-Sep-1040
From the Chest Wall Model screen, the respiratory profile is created, including the following parameters:
The Chest Model
• Pressure Waveform
• Breath Rate (bpm)
• Muscle Pressure (Pmax) in cm H2O
• Expiratory, Hold, Release Time (%)
• Insp/Exp for real-time patient effort adaptations
41
Go to ASL
Last edited: 30-Sep-1041
Click on the waveform sketch in the upper right quadrant of the screen or on the black triangles to choose a spontaneous breathing profile.
You may alsomake the patient passive
The Patient Effort Model
42
Go to ASL
Last edited: 30-Sep-1042
When a spontaneous pattern is chosen, you will need to provide values for the parameters listed, i.e.:
The Chest Model
• Spontaneous Rate (in bpm)
• Muscle Pressure(in cm H2O)
• Insp. Effort Rise Time (as % of the breath cycle time)
• Insp. Hold Time(as %)
• Insp. Effort Release Time(as %)
43
Go to ASL
Last edited: 30-Sep-1043
In ”sinusoidal” and “trapezoidal” patterns, you can also define forced expiration (new in SW 3.1)
The Chest Model
• Muscle Pressure(in cm H2O)
• Exp. Effort Rise Time (as % of the breath cycle time)
• Exp. Hold Time(as % of the breath cycle time)
• Exp. Effort Release Time(as % of the breath cycle time)
44
Go to ASL
Last edited: 30-Sep-1044
Click on the “Next >” button to proceed to Step 4.
The Chest Model
45
Go to ASL
Last edited: 30-Sep-1045
Click in the filename box (Path to parameter file) to edit the name of the parameter set you have created. Click on “Browse” to navigate to another folder and/or to overwrite another file.
-or-
Click on “Save >” to save the parameter set with the name that is displayed in the box.
Save Simulation Parameter Set
46
Go to ASL
Last edited: 30-Sep-1046
Move the cursor onto the Central Run Time Display window and click to bring the window to the top.
Note: If the message in the status box at the top of the Variable Parameters Run Time display is anything other than “Ready. Use Simulation switch to begin…”, please return to slide number 8 “Synchronizing with the ASL.”
Running a Simulation
47
Go to ASL
Last edited: 30-Sep-1047
Click on the Simulation switch to move from “Off” to “On.”
The message in the status bar will change to “Host data path and file name selection” and a dialog box will appear.
Running a Simulation
48
Go to ASL
Last edited: 30-Sep-1048
The Select an Output File dialog box requires that you choose a destination file for the data to be collected during the simulation run.
Running a Simulation
Click “OK” if you wish to retain the filename displayed. Otherwise, click in the filename box to change the name. You may also click on the button with a folder on it to browse for an existing file that you want to overwrite.
49
Go to ASL
Last edited: 30-Sep-1049
Note: You have the option of saving the simulation data either to the default data file, as shown in the dialog box, or with a filename of your choice.
The default data file, as well as the filename you choose, are base file names, requiring no extensions.
Data is actually saved to a set of files that together facilitate analysis of the simulation data.
Running a Simulation
50
Go to ASL
Last edited: 30-Sep-1050
You may, at this point, elect to save waveform data for the simulation run by checking this box
Running a Simulation
With this option chosen, data will be saved at a rate of 512 Hz, or lower, if desired. The waveform data sample rate is determined by the first patient file in a script only.
51
Go to ASL
Last edited: 30-Sep-1051
If you wish to save some of the waveform data later, you may do so at any time via the Breath Detection/Real Time Analysis (RTA) window.
Access the RTA window via the “Windows” drop-down menu
or via the ASL 5000 icon on the dock at the bottom or top of your screen.
Running a Simulation
-or-
52
Go to ASL
Last edited: 30-Sep-1052
Click in the Save Waveform Data box at the top of the screen to start saving data to high resolution waveform files.
This option can be turned on and off at any time throughout the simulation run.
Running a Simulation
53
Go to ASL
Last edited: 30-Sep-1053
After initialization is completed, the status bar will now read “Running simulation” and the simulation will begin.
You will now see waveforms moving across the screen, displaying flows, pressures and volumes.
Running a Simulation
54
Go to ASL
Last edited: 30-Sep-1054
Adjusting the Run Time Display
To change from a waveform to loop display, click the “Display” switch in the upper right corner of the Central Run Time Display screen.
55
Go to ASL
Last edited: 30-Sep-1055
Clicking on the “Freeze” switch next to the “Display” switch will allow you to freeze the waveforms display without stopping the simulation.
Adjusting the Run Time Display
56
Go to ASL
Last edited: 30-Sep-1056
Adjusting the Run Time Display
Click on the Pause switch to suspend the simulation without losing data.
Click again to start the simulation back up where it stopped.
57
Go to ASL
Last edited: 30-Sep-1057
From the Help/Customize menu, you can also access he “About” window.
Specifically, you can check the software version and build no.
Adjusting the Run Time Display
58
Go to ASL
Last edited: 30-Sep-1058
The scheme for background and line colors can be easily modified by selecting Graph Colors from the Help/Customize menu.
Click in the box labeled “Bkgr. Color” or on one of the plot lines above it. Use the pop-up color palette to choose the colors you want.
Adjusting the Run Time Display
59
Go to ASL
Last edited: 30-Sep-1059
Click in the box labeled “Bkgr. Color” or on one of the plot lines above it.
Use the pop-up color palette to choose the colors you want.
Adjusting the Run Time Display
60
Go to ASL
Last edited: 30-Sep-1060
Clicking on Launch Script Progress Window allows for a detailed view of the simulation in progress.
The Overall Script Progress shows the current and total number of repetitions from all script elements.
The Current Progress shows shows the current and total number of repetitions from the running script element.
Running a Simulation
61
Go to ASL
Last edited: 30-Sep-1061
Use the Interactive Control Panel (ICP) for making rapid adjustments to a simulation “on the fly,” i.e. while the simulation is running.
Switch to the ICP by clicking on the button in the upper left-hand corner of the Run Time Display screen.
The ICP can also be accessed via the “Windows” drop-down menu at the top of any of the ASL screens.
The Interactive Control Panel
62
Go to ASL
Last edited: 30-Sep-1062
The breath rate (bpm) and inspiratory tidal volume, Insp Vt (ml), are displayed in real time in the labeled windows on the left side of the screen.
The next window below indicates the type of patient model running in the current simulation. In this example, a passive patient has been modeled.
The Interactive Control Panel
63
Go to ASL
Last edited: 30-Sep-1063
The two blue buttons below the patient model display are controls for setting up a patient model.
You can choose to start with the “Default Lung Parameters.” Or click on “SETUP Lung Model” to access the existing list of simulation parameter files. Use one of these files as it is or click on the “Edit” button to modify it.
(See slide 19 – “Creating a Simulation Model.”)
The Interactive Control Panel
64
Go to ASL
Last edited: 30-Sep-1064
The Interactive Control Panel
In the panel to the right of the breath rate and Insp Vt windows, the filename and path to the originating parameters, (.var file), for the current simulation are displayed.
65
Go to ASL
Last edited: 30-Sep-1065
The last three buttons along the left side of the ICP allow you to start the simulation, record data to a high-res. file or quit the ICP and return to the Run-Time screen.
The Interactive Control Panel
66
Go to ASL
Last edited: 30-Sep-1066
Click on the Lung Model Param. tab to access the controls for Resistance and Compliance.
Use your mouse to click and drag the applicable knob. Alternatively, you can enter the desired value for Resistance or Compliance directly into the window below the corresponding control knob.
The Interactive Control Panel
67
Go to ASL
Last edited: 30-Sep-1067
The Spontaneous Breath Parameters tab provides an overview of the chest wall model parameters for the current simulation.
Changes can be made to these, as well as the Inspiratory, Hold and Release times using the “click-and-drag method” to turn the control knobs. Alternatively, simply highlight and change the displayed values.
The Interactive Control Panel
68
Go to ASL
Last edited: 30-Sep-1068
Trends for tidal volume, Pmax and breath rate for the current simulation are displayed under the Trends tab.
The Interactive Control Panel
69
Go to ASL
Last edited: 30-Sep-1069
To maintain a constant volume during a simulation, click the Closed Loop Control slider to activate this feature.
The Interactive Control Panel
70
Go to ASL
Last edited: 30-Sep-1070
In the case of a Constant Vt setting, the patient effort limits can be titrated by clicking on the Closed Loop Vt tab and using the “click-and-drag method” to manipulate the dials.
The Interactive Control Panel
71
Go to ASL
Last edited: 30-Sep-1071
The Interactive Control Panel
In the case of a Minute Volume Closed Loop setting, both the Pressure and Breath Rate limits can be manipulated by clicking on the Closed Loop MV tab and using the “click-and-drag method” to manipulate the dials.
72
Go to ASL
Last edited: 30-Sep-1072
The Interactive Control Panel
In the Closed Loop for CO2 (CO2Y) setting, both patient effort pressure and breath rate are algorithmically controlled to achieve a desired end-tidal CO2 value. Clicking on the Closed Loop CO2Y tab you can enter a value for CO2 production for the patient model and the target end-tidal CO2 level.
73
Go to ASL
Last edited: 30-Sep-1073
The Interactive Control Panel
From the New File tab you can specify a different patient parameter set as your starting point for making further interactive changes to the model
74
Go to ASL
Last edited: 30-Sep-1074
In Step 2 of the model creation process, you have several options for more advanced modeling. One of these is the ability to vary certain respiratory model parameters over the course of the simulation.
To choose this option, click on the Time-Varying Parameters switch in the Lung Model screen. The switch will go from “Constant” to “Time Varying” and an “Edit” button will appear.
Time-Varying Parameters
75
Go to ASL
Last edited: 30-Sep-1075
Clicking on the “Edit” button brings up the Time Varying Parameter Menu.
Double-click on the parameter you wish to profile to call up the Variable Parameters Editor screen.
Time-Varying Parameters
76
Go to ASL
Last edited: 30-Sep-1076
The graph at the top of the screen shows a breath-by-breath plot of the parameter to be profiled. The equation for this plot, or line, is displayed in the large white field in the bottom half of the screen.
Double-click on the line equation.
Time-Varying Parameters
77
Go to ASL
Last edited: 30-Sep-1077
Another screen appears--the Edit Curve Segment dialog box
Choose the type of line equation by clicking the down arrow under “Curve Type” on the upper left side of the screen.
Insert the desired values either by replacing current values in the boxes or by using the up/down arrows.
Click “OK” to return to the Variable Parameters Editor screen.
Time-Varying Parameters
78
Go to ASL
Last edited: 30-Sep-1078
When you are finished creating the profiles for the time-varying parameters, click “OK” in the Variable Parameters Editor to go to the Time-Varying Parameter Menu.
Click “Done” in the Time- Varying Parameter Menu dialog box to return to the Lung Model screen.
Time-Varying Parameters
79
Go to ASL
Last edited: 30-Sep-1079
Note: If you have chosen the “Time-Varying Parameters” option in the Lung Model screen, you are required to use this type of parameters throughout the rest of the modeling process.
Click on the “Next >” button in the Lung Model screen to edit the pressure profile in the Chest Wall Model (Step 3).
Time-Varying Parameters
80
Go to ASL
Last edited: 30-Sep-1080
Click on “Edit” beneath the grayed-out Pressure Profile graph of the Chest Wall Model window to bring up the Time-Varying Parameter Menu.
From there, parameters can be edited individually as in the Lung Model stage.
Time-Varying Parameters
81
Go to ASL
Last edited: 30-Sep-1081
When working with two-compartment lung models, you have the option of creating a non-linear compliance model.
(Please see slide number 33, “The Lung Model.”)
Click on “Non-Linear Compliance” in the lower right corner of the Lung Model screen.
Non-Linear Compliance
82
Go to ASL
Last edited: 30-Sep-1082
To edit the compliance curves, click on the “Edit C1” or “Edit C2” button.
-or-
Click in the white box next to “C2 = C1” to use the same compliance curve in both compartments of your lung model. Then click the “Edit C1” button.
Non-Linear Compliance
83
Go to ASL
Last edited: 30-Sep-1083
The Non-Linear Compliance Editor presents an interactive graph of the Volume-Pressure relationship. Click and drag the red dots to modify the Compliance.
-or-
Enter the desired values for the x- and y- coordinates, i.e. Pressure and Volume.
Non-Linear Compliance
84
Go to ASL
Last edited: 30-Sep-1084
The section between the two red horizontal lines is an area of proportional p/v response.
The green line represents the alternative “Nominal Compliance” value used when switching back to linear C response.
Click “OK” when finished to return to the Lung Model screen.
Non-Linear Compliance
85
Go to ASL
Last edited: 30-Sep-1085
ASL Sample Cases
Scenario: Objectives:
1. Apnea Managing Backup Ventilation
2. Bagging a Patient Teaching Resuscitation Skills
3. Compliance Step-Down
•Managing Alarming Events• -Low minute ventilation• -High pressure
4. Overdistension Identifying Excessive Pressures
5. Trigger Challenge Finding Appropriate Trigger Thresholds
86
Go to ASL
Last edited: 30-Sep-1086
Review slide #8: Synchronizing the ASL 5000
Review slide #10:Demo Mode
ASL Sample Cases
Note: Before executing any of the Sample Cases included in this tutorial, please remember to…
• Make sure that the software is not in “Demo” mode
• Start-up and synchronize the ASL 5000
87
Go to ASL
Last edited: 30-Sep-1087
ASL Sample Cases
Each of the Sample Cases includes a link to an associated script file. Make sure you have the ASL program open and open the Script Editor. In the Sample Case slides, you will be asked to use particular Script files. Be sure to choose the correct file name.
To return to the Tutorial from the ASL program, click on the Microsoft PowerPoint icon on the dock at the bottom of your screen.
Sample Cases
ASL 5000
General Instructions:
88
Go to ASL
Last edited: 30-Sep-1088
Scenarios:
• Apnea
2. Bagging a Patient
3. Compliance Step-Down
4. Overdistension
5. Trigger Challenge
ASL Sample Cases
Sample Cases
ASL 5000
89
Go to ASL
Last edited: 30-Sep-1089
Ventilator settings:
1. Ventilation Mode: VCV - A/C2. PEEP: 3 cmH2O3. Breath Rate (back-up): 8 bpm4. Tidal Volume: 700 mL5. Pressure (or Flow) trigger: 1 cmH2O
(1 L/min)
1. High Pressure Alarm: 20 cmH2O2. Low Minute Vent. Alarm: 7 L/min3. Apnea Delay Time: 30 s
From the Script Editor screen in your ASL 5000 program, choose the script file below and save it as “current.sct.”
Sample Cases: Apnea
C:\program files\ASL Software 3.2\vars\Examples\Apnea.sct
Open this script from your ASL 5000 program.
90
Go to ASL
Last edited: 30-Sep-1090
Review slide #46:Running a Simulation
Using your choice of patient circuit, connect the ventilator to the ASL 5000 at the port on the front of the unit.
Begin the simulation by clicking the “Simulation On” button in the Run Time screen.
Sample Cases: Apnea
91
Go to ASL
Last edited: 30-Sep-1091
Observe the following:Let the simulation run for several minutes…
Sample Cases: Apnea
1. Apnea alarm sounds.
2. Low minute ventilation volume alarm sounds.
3. Back-up breath rate kicks in--8 bpm.
92
Go to ASL
Last edited: 30-Sep-1092
Scenarios:
• Apnea
2. Bagging a Patient3. Compliance Step-Down4. Overdistension
5. Trigger Challenge
ASL Sample Cases
Sample Cases
ASL 5000
Please select another scenario
or
Return to the Table of Contents
93
Go to ASL
Last edited: 30-Sep-1093
From the Script Editor screen in your ASL 5000 program, choose the script file below and save it as current.sct.
Connect resuscitation bag with circuit to the vent port at the front of the ASL.
(See next slide for an alternate set-up.)
Sample Cases: Bagging a Patient
C:\program files\ASL Software 3.2\vars\Examples\Bag_Patient.sct
Open this script from your ASL 5000 program.
94
Go to ASL
Last edited: 30-Sep-1094
Alternate Setup:
Connect a manikin head with a functioning airway to the vent port at the front of the ASL 5000.
Use your choice of resuscitation bag with mask.
Sample Cases: Bagging a Patient
95
Go to ASL
Last edited: 30-Sep-1095
Have students bag the patient and ask them to feel the difference in compliance as it changes throughout the simulation.
Review Analysis Menu - Multi-Parameter Trend Display and evaluate consistency of delivered volumes, the presence or absence of excessive pressure peaks.
Sample Cases: Bagging a Patient
96
Go to ASL
Last edited: 30-Sep-1096
Scenarios:
• Apnea
2. Bagging a Patient
3. Compliance Step-Down
4. Overdistension
5. Trigger Challenge
ASL Sample Cases
Sample Cases
ASL 5000
Please select another scenario
or
Return to the Table of Contents
97
Go to ASL
Last edited: 30-Sep-1097
From the Script Editor screen in your ASL 5000 program, choose the script file below and save it as “current.sct.”
Sample Cases: Compliance Step-Down
Ventilator settings:
1. Breath Rate (back-up): 8 bpm2. High Pressure Alarm: 20 cm
H2O3. Low Minute Vent. Alarm: 7 L/min
C:\program files\ASL Software 3.2\vars\Examples\ C_Step_Down.sct
Open this script from your ASL 5000 program.
98
Go to ASL
Last edited: 30-Sep-1098
Observe the following:Let the simulation run for several minutes… 1. High pressure alarm sounds.
2. Minute ventilation changes very little.
Sample Cases: Compliance Step-Down
??????
What could cause this to happen? ???
???Answer: A dislocated ET tube could leave
just one lung compartment being ventilated. The ventilator will attempt to
deliver the same volume into the remaining lung compartment, effectively
over-inflating it.
99
Go to ASL
Last edited: 30-Sep-1099
Scenarios:
• Apnea
2. Bagging a Patient
3. Compliance Step-Down
4. Overdistension
5. Trigger Challenge
ASL Sample Cases
Sample Cases
ASL 5000
Please select another scenario
or
Return to the Table of Contents
100
Go to ASL
Last edited: 30-Sep-10100
From the Script Editor screen in your ASL5000 program, choose the script file below and save it as “current.sct.”
Sample Cases: Overdistension
Ventilator settings:
1. Breath Rate (back-up): 8 bpm2. Tidal Volume: varying
800-6003. High Pressure Alarm: 50 cm H2O
C:\program files\ASL Software 3.2\vars\Examples\ Overdistension.sct
Open this script from your ASL 5000 program.
101
Go to ASL
Last edited: 30-Sep-10101
Observe excessive pressure peaks.
Sample Cases: Overdistension
Let the simulation run for several breaths…
102
Go to ASL
Last edited: 30-Sep-10102
Scenarios:
• Apnea
2. Bagging a Patient
3. Compliance Step-Down
4. Overdistension
5. Trigger Challenge
ASL Sample Cases
Sample Cases
ASL 5000
Please select another scenario
or
Return to the Table of Contents
103
Go to ASL
Last edited: 30-Sep-10103
From the Script Editor screen in your ASL5000 program, choose the script file below and save it as “current.sct.”
Sample Cases: Trigger Challenge
Ventilator settings:
1. Ventilation Mode: VCV - A/C2. PEEP: 3 cm H2O3. Breath Rate (back-up): 8 bpm4. Tidal Volume: 700 mL5. Pressure (or Flow) trigger: 1 cm H2O
(1 L/min)
1. High Pressure Alarm: 20 cm H2O2. Low Minute Vent. Alarm: 7 L/min3. Apnea Delay Time: 30 s
C:\program files\ASL Software 3.2\vars\Examples\ trigger_challenge.sct
Open this script from your ASL 5000 program.
104
Go to ASL
Last edited: 30-Sep-10104
Use the Analysis Menu’s Real Time Graph Display option to show the trigger points.
Observe:
1. As Pmus decreases, at what point does the ventilator fail to trigger?
ASL Sample Cases: Trigger Challenge
Let the simulation run for several minutes…
105
Go to ASL
Last edited: 30-Sep-10105
No communication with the simulator!
Ensure that the correct Ethernet cable is connected to the RJ45 and/or a secure connection is made with the serial DB9 to the port labeled “Host”. The respective ports must also be connected and active on your PC.
Is the ASL device turned on? If not, it will take approximately 20 seconds to initialize after switching on..
Troubleshooting
blue cable for connection through a hub/switch (DHCP)
106
Go to ASL
Last edited: 30-Sep-10106
Did you choose “Use software with ASL Simulator” in the Welcome screen?
If using the connection by serial no. method (aslident.txt in installation folder), make sure that the ASL serial no. is entered that matches your hardware.
Troubleshooting
107
Go to ASL
Last edited: 30-Sep-10107
If you have confirmed the above four items, then try switching off the ASL device again. Close down the software by clicking on the “Exit ASL” button on the Variable Parameters Run Time screen.
Troubleshooting
108
Go to ASL
Last edited: 30-Sep-10108
Start up the ASL 5000 software again, then turn on the ASL device. (See slide # 10.)
On slower computers, it might be helpful to turn off any virus scan programs running on the computer.
If you are still unable to establish communication with the ASL 5000 device, please call IngMar Medical at: 1-800-583-9910.
Troubleshooting
109
Go to ASL
Last edited: 30-Sep-10109
An array of additional information for the ASL is available online:
User’s Manual
Instructional videos
User Forum
FAQs
All of these can be accessed via our website
Additional Information
110
Go to ASL
Last edited: 30-Sep-10110
Glossary of Terms
Hyperlink Hyperlinks are groups of text that are typically a different color than the balance of text on a particular slide or document. They are often underlined. When the mouse pointer is rolled over the underlined hyperlink text, the pointer converts to a hand with the index finger pointed at the text. Clicking on the text while the pointing hand is visible will advance you to some designated slide or file.
111
Go to ASL
Last edited: 30-Sep-10111
Glossary of Terms
Script File A Script File is a file containing various parameter sets that together form a specific simulation that you wish to run. A script file has the extension .sct.
Variable File Individual parameter sets are stored in files with the extension .var (for variables up to SW 2.2.) or vr3 (for variable sets made with SW 3.n).