J. Alozy, S. George , F.Murtas , M. Silari 1 CERN 17-06-2014 GEMPIX detectors as hadron therapy beam monitors and radioactive waste detectors GEMPIX construction Beam monitor for Hadrotherapy GEMPIX for Radioactive Waste Conclusions F.Murtas INFN-CERN-Ardent J. Alozy, E.Frojdh, S. George , M. Silari CERN

GEMPIX detectors as hadron therapy beam monitors and radioactive waste detectors

  • Upload
    allayna

  • View
    94

  • Download
    0

Embed Size (px)

DESCRIPTION

F.Murtas INFN-CERN-Ardent J. Alozy, E.Frojdh, S . George , M. Silari CERN. GEMPIX detectors as hadron therapy beam monitors and radioactive waste detectors. GEMPIX construction Beam monitor for H adrotherapy GEMPIX for Radioactive Waste Conclusions. Gempix Detector . - PowerPoint PPT Presentation

Citation preview

Page 1: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 1CERN 17-06-2014

GEMPIX detectors as hadron therapy beam monitors and radioactive

waste detectors

• GEMPIX construction• Beam monitor for Hadrotherapy• GEMPIX for Radioactive Waste• Conclusions

F.Murtas INFN-CERN-ArdentJ. Alozy, E.Frojdh, S. George , M. Silari CERN

Page 2: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 2CERN 17-06-2014

Gempix Detector

Side view

Quad medipix board

The detector has two main parts :- The quad medipix without silicon sensors- The triple gem detector with HV filters and connector

New HV GEM board

Page 3: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 3CERN 17-06-2014

Gem foils and frames A new GEM layout has been designedActive area of 28x28 mm2 The electordes path have been designed to avoid the medipix wire bonding.Produced by Rui De Oliveira.

New frames were designed 10x10 cm2

to fit the holes of the Quadmedipix board

5 different thickness (from 1 to 5 mm)

Wire bonding position

Page 4: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 4CERN 17-06-2014

Assembling the GEM foils

The three GEM foils are assembled on top of HV GEM board

Page 5: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 5CERN 17-06-2014

Final detector

Page 6: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 6CERN 17-06-2014

Head on detector

Gas flux AR CO2 or AR CO2 CF4

The anode is a quad naked without silicon sensor :The active area is 9 cm2

Triple GEMMylar window

Particles to be analysed

HVGEM Console

FITPIX Console

HV

FITPIX

Page 7: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 7CERN 17-06-2014

A small TPC

Gas fluxThe anode is a quad medipix without silicon sensor:The active area is 9 cm2

The particle track is analysed with 512 pixel in length of 3 cm

This detector could be used also as Tissue Equivalent Proportional Chamber The track path is equivalent to 30 microns of tissue … with 17 samples/microns

Triple GEM

Kapton windowParticles to be analysed

Length analysed

Page 8: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 8CERN 17-06-2014

GEMPIX Detectors

Head-on version(4 detectors)

Side-on version(2 detectors )

HV Connector

MedipixDAQ box

Page 9: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 9CERN 17-06-2014

Measurements atCNAO (Pavia - Italy)Carbon ion beam

at 480 Mev/nucleon

Page 10: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 10CERN 17-06-2014

MeasurementsBeam : 480 MeV/A Carbon Ion

Measurements in air

Measurements in water phantom :23 different depths throughout water phantomEach position given spot 5.108 carbon ion treatment (clinical treatment intensities)

Frame length = 1ms, gas = ArCO2, gain = 750 V (~0.43 keV/TOT - Proper calibration awaits the return of the detector from CNAO RP).

Page 11: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 11CERN 17-06-2014

Mylar windowGas in/out

Fitpix Readout

External HV (HVGEM)

GEMPIX at CNAO (Pavia)

Page 12: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 12CERN 17-06-2014

GEMPIX as a Beam Monitor

480 MeV/A carbon ions, 10 ms frame, ASIC in Medipix (particle counting) Mode, IKrum = 5

Counts

Single Carbon Ion

Other particle

Page 13: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 13CERN 17-06-2014

The overall beam shape is clearly visible

Beam profile

As number of hits As energy released

Page 14: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 14CERN 17-06-2014

Gempix Linearity

Number of counts per spills

Time (s)

Page 15: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 15CERN 17-06-2014

Phantom used at CNAO

Bragg Peak finder at CNAO

Thin WindowGempix

Stepper Motor

Several measurement of energy deposition along the beam (5x108 Carbon ions per 10 spills)

Page 16: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 16CERN 17-06-2014

Energy deposition is ~0.43 keV/TOT Spot width here is sigma of gaussian (FWHM = 2.3 sigma)

Depth profile and beam dimension

Page 17: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 17CERN 17-06-2014

Linear Log 3D reconstruction

Energy deposition: 3D reconstruction

This part (neutrons and gammas) could be analyzed with really high resolution

Page 18: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 18CERN 17-06-2014

Single Event Upset

Depth = 124 mm (in bragg peak), ~0.43 keV/Count (9.6 MHz, TOT mode), IKrum = 1, 1ms frame

SEU Events

Page 19: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 19CERN 17-06-2014

No Cfg Resets

Reset at start

Reset/10 frames

Reset every frame

Reconfiguring

Page 20: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 20CERN 17-06-2014

Radioactive Waste :a detector for 55Fe

Page 21: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 21CERN 17-06-2014

Sample Preparation

Fraiseuse DECKEL FP4M for powder productionreducing the 60CO background

Boxes supplied by SMIPA s.r.lDimension: 38x38x4 mm3

(now 38x38x8 mm3)Weight: 3.951 ± 0.15%

Measurements for Fe-55 concentration with GEMPIX ,two different configuration :Drift gap 11 mmDrift gap 3 mm

Support conceived for measurements with samples and calibration sources

Page 22: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 22CERN 17-06-2014

Radioactive waste: 55Fe

1 sec radioactive source

100s counts

Page 23: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 23CERN 17-06-2014

X-ray candidates

Bck

Signal vs Background

Gas AR CO2CF4Gate: 1 sHV 1200 VGas flux: 3.5 l/h

Clk: 24Ikrum: 5Polarization: -

Page 24: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 24CERN 17-06-2014

Cluster parameters measurements

Cluster type Cluster volume Inner size

Border size Size Size y vs size x

55Fe

60Co

55Fe

60Co

55Fe

60Co

55Fe

60Co

55Fe

60Co

Page 25: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 25CERN 17-06-2014

Online analysis

Once that the calibration values are set in the Python script, an online analysis can be performed

Number of X-Rays measured in 1000 sec

Number of big blob

Cut on sizeCut on charge

Page 26: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 26CERN 17-06-2014

Drift 3 mm vs drift 11 mm

Attenuation length 2.91 cm

Theory: 11 mm/ 3 mm = 2.6

Measurements:11 mm/ 3 mm = 2.7

Waiting the measurements from certificated company for the final calibration

Page 27: GEMPIX detectors  as  hadron therapy beam monitors and radioactive waste detectors

J. Alozy, S. George , F.Murtas , M. Silari 27CERN 17-06-2014

Summary Six GEMpix detectors have been built and they show good performances

from single particle detections to high intensity beam monitor

Really good cluster analysis is possible thanks to the Fitpix software pakage

Software tools and detector tuning are still needed for the cluster analysis in gas (dE/dX, particle id, 3D track reconstruction....)

The detector could be easily improved with Timepix3 chips

Other applications and measurements are scheduled for next months Monitor for burning plasma (KSTAR - PETAL) Measurements for micro and nano dosimetry with tissue equivalent gasMonitor for gamma radiotherapy