Upload
others
View
82
Download
0
Embed Size (px)
Citation preview
Radiation DetectorsStrahlungsdetektoren
Geiger-Mueller and Proportional Counter Tubes
Geiger-Müller- und Proportional-Zählrohre
� �
�
�
�
�
� �
�
� �
1) with compensation filter / mit Energiekompensation: A = Ambient Dose Equivalent H*(10), E = Air Kerma/Exposure 2) A in µSv/h, E in µGy/h; for the uncompensated types the range is given in dose rate unit at 662 keV (137Cs)3) shielded with / gemessen hinter 5 cm Pb + 0.2 cm Al 4) without connectors / ohne Anschluss
IntroductionSince more than 55 years VacuTec hasbeen developing and manufacturingcounter tubes in different versions andin top-standard glass technology. In thehigh-dose range and for detection ofparticle and low-energy X-radiationVacuTec can supply newly developedcounter tube types with chromium ironcathodes as miniature and endwindowtubes. With this wide range of productswe can offer the right counter tube fora multitude of applications.
EinführungVacuTec entwickelt und produziert seitmehr als 55 Jahren die verschiedenstenZählrohre vorwiegend für Quanten-strahlung in spezieller Glastechnologie.Für Anwendungen im Hochdosisbereichund zum Nachweis von Teilchen- undniederenergetischer Röntgenstrahlungstehen neuentwickelte Zählerrohrtypenmit Chrom-Eisen-Katode als Miniatur-und Fensterzählrohre zur Verfügung.VacuTec bietet mit seinem umfangrei-chen Sortiment das passende Zählrohrfür unterschiedlichste Anwendungen.
Typische AnwendungenVacuTec produziert Zählrohre für dieMessung von Alpha-, Beta-, Gamma-,Neutronen- und Röntgenstrahlung. Siefinden breite Anwendungen bei derMessung des Strahlenpegels im Strah-lenschutz und zur Umweltüberwa-chung. Weitere Anwendungen liegenim Bereich der zerstörungsfreien Werk-stoffprüfung sowie der berührungslo-sen Füllstandsmessung. Für Dosimetrieund Umweltmessungen sind die Zähl-rohre energiekompensiert nach „Luft-Kerma/Exposure” (Typ E) oder nach„Ambient Dose Equivalent” (Typ A)erhältlich.
Type
Typ
SensitivityEmpfindlichkeit
Dose-rate-rangeDosisleistungs-
bereich
(µSv/h /µGy/h)2)
Couting rate [137Cs]at 1 µSv/h / µGy/h
Zählrate [137Cs]bei 1 µSv/h / µGy/h2)
(counts/s /Impulse/s)
Dead-time
Totzeit
(µs)
Backgroundshielded
Nulleffektgeschirmt3)
(counts/min /Impulse/min)
PlateauPlateau
FinishAus-
führungband /Bereich
max.lenght /Längemm4)
RangeBereich
(V)
SlopeSteigung
(%/100 V)� � �
7000370003 A/E1)
• ••
4445
1 ... 3 x 104
1 ... 1051.91.6
� 90� 120
� 8� 8
400 ... 600400 ... 600
88
glassglass
7001370013 A/E1)
••
170167
0.3 ... 104
0.3 ... 10410 7.5
� 100� 100
� 60� 60
400 ... 600400 ... 600
1010
glass / CrFeglass / CrFe
7001470014 A/E1)
• ••
2931
4 ... 106
4 ... 1060.280.24
� 25� 25
� 3� 3
400 ... 600400 ... 600
1515
CrFeCrFe
7001770017 A/E1)
• ••
5959
2 … 105
3 … 3 x 1050.730.6
� 25� 25
� 5� 5
400 ... 550400 ... 550
1010
CrFeCrFe
7001870018 A/E1)
••
4355
102 … 2 x 107
102 … 2 x 1070.020.02
� 20� 20
� 2� 2
520 ... 620520 ... 620
3030
glassglass
7001970019 A/E1)
• ••
5151
1 … 4 x 104
1 … 4 x 1042.11.6
� 60� 60
� 7� 7
400 ... 600400 ... 600
44
CrFeCrFe
7003170031 A/E1)
••
270270
0.2 … 3 x 103
0.2 … 3 x 1031614
� 150� 150
� 120� 100
400 ... 600400 ... 600
1010
glassglass
70035 • • 106 0.4 … 8 x 103 6 � 80 � 60 400 ... 600 8 CrFe
End-windowEnd-fenster
Energy rangeEnergiebereich
WindowFenster
Dead-time
Totzeit
(µs)
Backgroundshielded
Nulleffektgeschirmt3)(counts/min /Impulse/min)
PlateauPlateau
FinishAus-
führung� � � ø
(mm)
ThicknessDicke
(mg/cm2)
MaterialMaterial
RangeBereich
(V)
SlopeSteigung
(%/100 V)(MeV) (keV) (keV)
70071700727007470075
3.53.53.53.5
50505050
3333
18133645
1.5 ... 21.5 ... 21.5 ... 21.5 ... 2
MicaMicaMicaMica
� 120� 90� 120� 30
572424
420 ... 650400 ... 600420 ... 650850 ... 1050
54310
CrFeCrFeCrFeCrFe
Typical ApplicationsThe counter tubes manufactured byVacuTec can be used for measuringalpha, beta, gamma, neutron and X-radiation. They have been introducedin the measurement of radiation levelsfor radiation protection and environ-ment monitoring applications. Furtherapplications include non-destructivematerial testing and contactless levelmeasurement. For usage in the fields ofdosimetry and environmental protec-tion the counter tubes are available withenergy compensation according to theAir Kerma/Exposure (type E) or AmbientDose Equivalent (type A) concepts.
alike / nur ähnlich
1
n2
n1
cou
nt/
s
startingvoltage
V1 V2recommendedvoltage
plateau voltagerange
supply voltage
plateau slope
7
3 x 105
2 x 107
2 x 107
… 4 x 104
… 4 x 104
… 3 x 103
… 3 x 103
… 8 x 103
Cross Reference List Austauschliste
Measuring Range Messbereichsübersicht
Type
Typ
10-1 100 101 102 103 104 105 106 107
70 003
70 013
70 014
70 017
70 018
70 019
70 031
70 035
Dose rateDosisleistung(mGy/h)
VacuTec Centronic LND Mullard1) Philips1)
Amperey1)
70 00370 003 E70 003 A
ZP1200ZP1201ZP1202
712171210
MX 146 18503
70 01370 013 E70 013 A
ZP1210 78016 MX 12018520
70 01470 014 E70 014 A
~ ZP1300~ ZP1313~ ZP1314
~ 714~ 7149
70 01770 017 E70 017 A
~ ZP1320~ ZP1321~ ZP1324
~ 713~ 71322~ 7139
MX 164 18550
70 01970 019 E70 019 A
ZP1200ZP1201ZP1202
712171210
MX 146 18503
70 03170 031 E70 031 A
ZP1220ZP1221
ZP1221/02
780177807
MX 145 18545
70 035
70 071 7241
70 072 ZP1401 7124 MX 147 18504
70 074 ZP1430 72314 MX 169 18526
70 075 ZP1460 7312 MX 167 18546
Geiger-Mueller TubesGeiger-Müller-Zählrohre
~ alike / nur ähnlich
1) for info only / nur zur Info
For detailed technical information please request our documents. On request we develop counter tubes and probes to customer specifications.
Für weiterführende Informationen zu diesen Produkten fordern Sie bitte unsere Unterlagen an. Die Entwicklung kundenspezifischerZählrohre und Sonden übernehmen wir auf Anfrage gern.
701/
2
VacuTec Meßtechnik GmbHDornblüthstraße 14 | D -01277 Dresden | Germany | Phone: + 49 (0) 351 31724-0Fax: + 49 (0) 351 3105085 | [email protected] | www.vacutec-gmbh.de
Proportional Counter Tube 70 020with end window for separate detection of ��and ��radiation
Alpha energy range
Beta energy range
Recommended supply voltage for ��radiation
Recommended supply voltage for ��radiation
Filling gas
Entrance window
Window diameter
Alpha-Energiebereich
Beta-Energiebereich
Empfohlene Arbeitsspannung für �-Strahlung
Empfohlene Arbeitsspannung für �-Strahlung
Gasfüllung
Eintrittsfenster
Fensterdurchmesser
> 3 MeV
> 40 KeV
1100 V
1400 V
He /organic vapour
Mica 2 mg/cm2
ø 31 mm
Proportional-Zählrohr 70 020mit End-Eintrittsfenster für den separaten Nachweis von �- und ��Strahlung
Proportional Counter Tube 70 021/022/023with mica/aluminium window in pancake style
Proportional Counter Tube 70 048Large-area proportional counter tube used for contamination measuring devices with separate detection of �, ��and � radiation
Alpha energy range
Beta energy range
Recommended supply voltage for ��radiation
Recommended supply voltage for �, ��and ��radiation
Filling gas
Entrance window
Active area
Dimensions
Alpha-Energiebereich
Beta-Energiebereich
Empfohlene Arbeitsspannung für �-Strahlung
Empfohlene Arbeitsspannung für �-, �- und �-Strahlung
Gasfüllung
Eintrittsfenster
Aktive Fläche
Abmessungen
> 3 MeV
> 40 KeV
800 V
1450 V
Ar/CO2
Al 3 mg/cm2
135 cm2 / 230 cm2
211.6 mm x 119 mm
211.6 mm x 181.4 mm
Proportional-Zählrohr 70 048Großflächenzählrohr für Kontaminations-Messgeräte mit separatem Nachweis für �-, �- und �-Strahlung
Proportional Counter Tubes for the Measurement of ActivitiesProportional-Zählrohre für Aktivitätsmessungen
Proportional-Zählrohr 70 021/022/023mit Glimmer-/Al-Fenster im „pancake style“
Recommended supply voltage for ��radiation
Recommended supply voltage for ��and ��radiation
Filling gas
Entrance window
Window diameter
Empfohlene Arbeitsspannung für �-Strahlung
Empfohlene Arbeitsspannung für �- und �-Strahlung
Gasfüllung
Eintrittsfenster
Fensterdurchmesser
1000 V
1450 V
Ar/CO2
Mica 2 / Al 3 / Mica 1.7 mg/cm2
ø 44.5 / 57.2 / 28.6 mm