Carrier and makeup gas settings selectable for He, H2, N2 and Ar Turnkey Laboratories Solutions Gas Chromatography Optima GC - 3007 Optional advanced autosampler Sophisticated sample inlets EPC/AFC gas system Multi-valve and multi- column technology Column oven temperature- ramp programmable Psi, KPa, Bar units selectable Pressure control range: 1~150psi, Pressure accuracy: 0.001Psi Programmable pressure ramp RSD < 0.5% Pressure/flow rate : -400 to 400 kPa/min and -400 to 400 ml/min Flow rate control range: 0~600ml/min, flow rate accuracy: 0.1ml/min flow rate RSD < 0.1% Programmable pressure/flow rate ramping: 20 steps Excellent FID wide range at <1.5* 10-12 pg/s

GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

  • Upload
    others

  • View
    19

  • Download
    0

Embed Size (px)

Citation preview

Page 1: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

Carrier and makeup gas settings selectable for He, H2, N2 and Ar

Turnkey Laboratories Solutions

Gas ChromatographyOptima GC - 3007

Optionaladvanced

autosampler

Sophisticatedsample inlets

EPC/AFC gassystem

Multi-valveand multi-

columntechnology

Column oventemperature-

rampprogrammable

Psi, KPa, Bar units selectable

Pressure control range: 1~150psi, Pressure accuracy: 0.001Psi Programmable pressure ramp RSD < 0.5%

Pressure/flow rate : -400 to 400 kPa/min and -400 to 400 ml/min

Flow rate control range: 0~600ml/min, flow rate accuracy: 0.1ml/minflow rate RSD < 0.1%

Programmable pressure/flow rate ramping: 20 steps

Excellent FIDwide range at<1.5* 10-12 pg/s

Page 2: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

" ~450" "

Detector Limit of detection Baselinenoise

Maxoperatingtemp.

Baseline drift(after 2hrsstabilization)

Lineardynamicrange

FID 1.5PgC/s(N-C16)

4.4fg /MI

S: <2.0 x 10-11 g/sP: <5.0 x 10-13 g/sor 2.0 x 1013 g/s

<4 x 10-13 A

<30uV

<2 x 10-14 A/min zH 004 5 x 10-14 A/30min

400 Hz

100 Hz

100 Hz

100 Hz

> 107

> 105

5*104

S > 102

P: > 103

N: > 103

P: > 103

< 100V /30min

< 50uV /30min

S: <2 x 10-13 AP: <8 x 10-13 A

S: <1 x 10-12 A/30minP: <2 x 10-12 A/30min

2 x 10-12 A/30min

TCD

ECD

FPD

NPD

400oC

400oC

400oC

400oC

450oC

Programming temp. - ramp speed : 0-120oC/min

Fast cool down: 450~50oC< 1.5minutes

Temp. Contrl range: RT+5oC ~ 450oC (0.1oC increment)

N: <0.3 x 10-12 g/s(Azobenzene)P: <0.1 x 10-13 g/s(Malathion)

Dataacquisitionrate:

<20uV

Column oven dimension: 278x 310x 165mm=15.5L

Programming ramp : 20 ramp, programing time of 9999.99 mins.

Temp. accuracy: + 0.01℃C (Rapid column oven heating with rates -250oC to +250oC)

< 400pg tidecane/ ml

Page 3: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

Clarity workstation feature

Multi-channel and multi-user universal workstation to control all parameter Can collect signal from 4 detectors via RS232 or USB

Basic parameter: max sample-collecting frequency @100Hz/unlimited peak process quantity/integral sensitivity @1µV*s

Batch processing makes machine control, auto sampling sequence collecting, auto integral correction much easierauto integral correction much easier

Fully support FDA-21CFR PART 11, SST AND IQ/OQ

Powerful post-treatment facilitate chromatography comparing, re-correction and data input&output.

Feature like easy to use report publisher, online help and answer, fully compatible with windows 7/10.

Standard setup

Optionalsamplingdevice

Capillary inlet Split/splitless capillary

Packed inlet Packed inlet

Optima-3007 Clarity counter-controlworkstationWorkstation

Injection Value

Headspacesampler

Thermaldesorption

6 - port valve or 10-port valve

Available

Available

16 or 150Auto sampler

Offers minimum sampling time snapshot function, single analysis capability. Automatic and manual peak integration, manipulation, identification, calibration points and levels and manual calibration curve creation, column performance calibration, data comparison function, Capability of connecting micro-bore column of 0.1 mm ID etc.

Page 4: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

AFC: split/splitless mode, carrier gas can fulfill constant-pressure, constant-flow, constant-speed, programmable-temperature-rise, programmable-flow-rise andprogrammable-speed-rise to reduce sample decomposition and discrimination whileincrease separation and shorten analysis time.

EPCAFC fulfill digitalization and automatization, only need input column parameters, EPC/AFC can set best flow of column and show digitally.EPC/AFC can fulfill gas leaking self-diagnosis and cut off flow&gas source and alarm at the same time.

Traditional GC(Column temp @140oC) Analysis time 43min

One button access to routine maintenance information.

-rise and programmable-speed-rise; unique constant linear velocity mode for carrier gas.Unlimited valve events to fulfill accurate switching of multi-valve;Outside power & voltage checking system, over-heating protection system and flow monitoring system to make it intellectualized.Vacuum fluorescent display with english/chinese changeable;Multi-function keyboard can set complicated parameters and store 16 chromatography method;FID and FPD can provide reminder for auto ignite and turn-off, TCD with overflow/cut-off protection, flameout detection.Excellent FID wide-range to enhance linear range;Multi-valve and multi-column switch system to make sure complicated analysis at one times injection;PCB system shielding function to reduce interference;Carrier gas saving mode available to reduce cost;Autosampler, headspace and thermal desorption can be incorporated;Clarity workstation can fulfill 3Q certification to meet GMP/GLPstandard.

Instrument feature

AFC characteristic

GC9720(Column temp @140oC) with programmable pressure-rise Analysis time 17min

PCM control module much precise with independent-development AFC system;8-channel high-accuracy temperature control system and 8-channel outside events to fulfill counter-control ;Every gas circuit can achieve constant-pressure, constant-flow, constant-speed, programmable-temperature-rise, programmable-flow

GC should be capable of upgradation to any 3rd detector or mass Detector.Built in display and computer interface, display chromatograms, method parameters like temp., pressure and flow rates etc.ECD complete with 63NI source and low voltage heatersAdvance self Diagnostic function (inbuilt)

Page 5: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

Carrier gas saving mode:After injeciton, can proceed low split flow modeautomatically to reduce carrier gas significantly.

13 min.4 11.6min

Traditional GC constant-flow mode Analysis time 13.4min Optima-3007 with programmable pressure-rise Analysis time 17min

Multi-valve and multi-column switch

Adopting AFP pneumatic valve,can setup 3-valve- 4-column and 4-valve- 5- column switch system to fully analyze only atone time sinpetro chemical, coalgas, trace C2H2 inethylene and trace CO and CO2 in ethylene.

Multi-valve and multi-column flow diagram

Page 6: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

Clarity workstation is applicable

Multi-channel and multi-user universal workstation to control all parameter

Can collect signal from 4 detectors via RS232 or USB

Basic parameter: max sample-collecting frequency @100Hz/unlimited peak process

quantity/integral sensitivity @1*sμV

Batch processing makes machine control, auto sampling sequence collecting, auto integral

correction much easier

Fully support FDA-21CFR PART11, SSTAND IQ/OQ

Powerful post-treatment facilitate chromatography comparing, re-correction and data

input&output.

1 2 3

4

5

67

8

9

10

1

2

3

4

5

6 1

2

3

4

5

6

1

2

3

4

5

6

PCM

Hayesep Q(Separate )2mCO2, CO and SO2

m2)(A5 block H2, O2 and N2

2ml Hayesep Q pre-column 1

Release

Air block

Port 3 o�

Capillar y Pre-column 2S/SL

5UL Port 2 o�

Out

Port 1 o�

Cap. Al O2 3

TCD

FIDCAP

FL9720 workstation feature

Fully counter-control to set all parameter in workstationEasy operationalMulti-channel signal sampling, multi outside events control

Page 7: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

Autosampler for optima-3007 Big displayer with double -tower automatic injection;Tray vial quantity: 16 or 150;Injection volume: 0.1-250uL;Sampling accuracy:+0.01uL;Precision: 0.4%Injection Port: split/splitless capillarycapillary columns: 50, 100, 250, 350 µInjection loop: multiport(0.25ml, 0.5ml and 1ml) Injection repeatability: <0.5%Maximum Temperature: 450oCMaximum pressure: 0-150psi(with EPC)Reproducibility: 0.3% RSD

Temperature set point Resolution: 0.1℃

Column oven dimension: 278x310x165mm=15.5L;

Temp. control range: RT+5℃~450℃(0.1℃ increment);Temperature Ramp: multi-ramp(>14)with plateaus

Wide split ratio setting range: 1 to 4500.Packed injection, capillary injection, flash-evaporationinjection, PTV injection and liquid injection are available;Easy consumables changeover .

Temp. accuracy: ±0.01℃

Programming temp.-ramp speed: 0-120℃/minProgramming temp.-ramp : 20Fast cool down: 450~50℃≤ 5min

Ignition recognition, Hprotection, anti-ponding;Solvent no tailing peak.Dual Temperature control.

Excellent wide range FID design, no ceiling limit for solvent peak;Dual flow rate differential system.Limit of detection can be 1.5*10 -12g/s;

Page 8: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

Performance introductionINSTRUMENTS PERFORMANCE IS

INTRODUCED

Excellent qualitative repeateability

[pA]

00.5200.0200.5100.0100.500.0

00.0500.5400.0400.5300.0300.52

[pA]

00.5700.0700.5600.0600.5500.05

200.000

179.500

159.00

158.500

118.00

97.500

77.00

55.500

35.500

15.500

-5.00000.00100.5900.0900.5800.0800.57

[pA]

200.000

179.500

159.00

158.500

118.00

97.500

77.00

55.500

35.500

15.500

-5.000

200.000

179.500

159.00

158.500

118.00

97.500

77.00

55.500

35.500

15.500

-5.000

200.000

179.500

159.00

158.500

118.00

97.500

77.00

55.500

35.500

15.500

-5.000

[pA]

Retention time RSD ≤0.05%

Can incorporate flash-evaporation and high pressure liquid injection

Flash-evaporation injection for gas-liquid mixture, high pressure valve injection for liquid.

PTV sample injectionWith multiple accumulated injection and solvent release, increase temperature program-rise to achieve trace analysis.

Page 9: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

300ppm FIDstandard sample

Peak at , no ceiling limit130000

150.000

130.000

110.000

90.000

70.000

50.000

30.000

10.000

-10.000

150.000

130.000

110.000

90.000

70.000

50.000

30.000

10.000

-10.000

150.000

130.000

110.000

90.000

70.000

50.000

30.000

10.000

-10.000

150.000

130.000

110.000

90.000

70.000

50.000

30.000

10.000

-10.000

0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 20.00 22.00 24.00 26.00 28.00 30.00 32.00 34.00 36.00 38.00 40.00

40.00 42.00 44.00 46.00 48.00 50.00 52.00 54.00 56.00 58.00 60.00 60.00 62.00 64.00 66.00 68.00 70.00 72.00 74.00 76.00 78.00 80.00

[pA] [pA]

[pA][pA]

[pA]

40.000

35.000

30.000

25.000

20.000

15.000

10.000

5.000

0.000

-5.00000.2100.0100.800.600.400.200.0

130000.0

116500.0

103000.0

89500.0

76000.0

62500.0

49000.0

35500.0

22000.0

8500.0

-5000.00.00 2.00 4.00 6.00 8.00 10.00 12.00

[pA]

Excellent quantitative repeatability

Excellent wide-range design makes no ceiling limit of solvent peak

GC-3007 is suitable but no limited to below application: food safety, environmental protection, energy(), medicine Petroleum refining industry and etc.

Typical application

Peak area RSD ≤0.8%

Page 10: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

【 】Food safety

[pA]70.000

55.000

40.000

25.000

10.000

-5.00015.00 20.00 25.00 30.00 35.00 40.00 45.00

Setup

Setup

DEHP(Di-(2-ethylhexyl)phthalate)

[pA]500.000

450.000

400.000

350.000

300.000

250.000

200.000

150.000

100.000

50.000

0.000

-50.0000.00 5.00 10.00 15.00 20.00 25.00 30.00 3 5.00 40.00 45.00

Detector: FIDInlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: LZP950 for wineWorkstation: Optima-3007

Setup

Detector: FIDInlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: HP-5 cap.Workstation: Optima-3007

Chinse white wine

200ppb organochlorine in pesticide residue

Setup

Peak sequence:α-BHC,β-BHC,γ-BHC, δ-BHC, op-DDE, pp-DDD, op-DDT, pp-DDT

[mV]

10.00 13.00 16.00 19.00 22.00 25.00 28.00 31.00 34.00

200.000

180.000

160.000

140.000

120.000

100.000

80.000

60.000

40.000

20.000

0.000

-20.000

Detector: ECDInlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: HP-5Workstation: Optima-3007

Page 11: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

Setup

Complicated Organochlorine

Setup

100ppb organophosphoruspesticideresidue

60.000

55.000

50.000

45.000

40.000

35.000

30.000

25.000

20.000

15.000

10.000

5.000

0.000

-5.000

30.000

26.600

23.200

19.800

16.400

13.500

9.600

6.200

2.800

-0.600

-4.00030.00 35.00 40.00 45.00 50.00 55.00

】【Environmental protection

Peak sequence: COS, CS2, SO2, Methyl Mercaptan, Ethyl Mercaptan

450.000

400.000

350.000

300.000

250.000

200.000

150.000

100.000

0.00 3.00 6.00 9.00 12.00 15.00 18.00 21.00 24.00

[mV]Setup

[mV]

00.600.200.0 00.8100.6100.4100.2100.0100.800.4

[pA]

Detector: FPDInlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: DB-35Workstation: Optima-3007

Peak sequence: DDVP, methamidophos, acephate, omethoate, dimethoate, parathionmethyl, fenitrothion, parathion, quinalphos, tiguron triazophos

Detector:TCDInlet: PackedGas control module: 2 setsInjection mode: liquid autosamplerColumn: Hayesep QWorkstation: Optima-3007

Detector: ECDInlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: HP-5Workstation: Optima-3007

Page 12: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

Peak sequence: CHCl3, CCl4

Setup

Detector: ECDInlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: DM-1Workstation: Optima-3007

【 Energy 】

Petroleum refining

Distillationsimulation

inorganic gas

[mV]

0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00

Hydrocarbon

0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00

47.500

42.500

37.500

32.500

27.500

22.500

17.500

12.500

7.500

2.500

-2.500

[pA]

[pA]

0.00 2.00 4.00 6.00 8.00 10.00 12.00

700.000

629.000

558.000

487.000

416.000

345.000

274.000

203.000

132.000

61.000

-10.000

Setup

Detector:FID+TOCInjector: gas/capillary

Inorganic gas peak sequence: H2, CO2, O2, N2, COHydrocarbon peak sequence: CH4, C2H6, C2H4, C3H6, C3H8, cyclopropane, C2H2, iso-butane,Propadiene, n-butane, trans-2-Butene,, n-butylene, isobutene, cis-2-Butene, isopentane, n-pentane,allylene, 1,3-butadiene, 2-methyl-2-butene, trans-2-Pentene, 1-pentene, cis-2-Pentene, Hexane

4-valve-5-columnGas control module: 4setsValve injection

Column: Hayesep Q packed, 5A packed,AL2O3 capillary, DB-1 capillaryWorkstation: GC-3007

Setup

600.000

500.000

400.000

300.000

200.000

100.000

0.000

-100.000

Detector: FIDInlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: HP-1 cap.Workstation: GC-3007

Page 13: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

Aromatic compounds

[pA]

0.00 6.00 12.00 18.00 24.00 30.00 36.00 42.00 48.00 54.00

1750.000

1550.000

1350.000

1150.000

950.000

750.000

550.000

350.000

150.000

-50.000

-250.000

Setup

Detector: FIDInlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: Innowax cap.Workstation: Optima-3007

Peak sequence: Benzene, Toluene, Ethane, P-xylene, M-xylene, P-Ethyltoluene, M-Ethyltoluene,S-Butybenzene, Diethylbenzene, O-diethylbenzene

,

Dimethyl etherinLNG

15.000

13.300

11.600

9.900

8.200

6.500

4.800

3.100

1.400

-0.300

-2.000

[pA]Setup

0.00 2.00 4.00 6.00 8.00 10.00 12.0 0 14.00 16.00 18.00

Setup

Setup

Organic acid

Cresolisomer

100.000

80.000

60.000

40.000

20.000

0.000

-20.0000.00 4.00 8.00 12.00 16.00 20.00 24.00 28.00

700.000

625.000

550.000

475.000

400.000

325.000

250.000

175.000

100.800

25.000

-50.0000.00 2.30 4.60 6.90 9.20 11.50 13.80 1 6.10 18.40 20.70 23.60

【 Medicine】

[pA][pA]

[pA]

Peak sequence:o-cresol, p-cresolm-cresol,

Detector: FID

Peak sequence: acetic acid, propionic acid, butyrate, valeric acid, sovaleric acid, caproic acid, heptylic acid, octanoic acid, n-nonanoic acid,lactic acid, 2-Hydroxy-2-Methylbutyric Acid.

Inlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: DB-FFAPWorkstation: Optima-3007

Detector: FIDInlet: CapillaryGas control module: 2 setsInjection mode: Gas valve injectionColumn: PLOTQ cap.Workstation: Optima-3007

Peak sequence: CH4, C2H2, prepene, propane,methylal, Dimethyl ether, n-butene, cis-bitene,isoamylene, methyl alcohol, n-pentane

Detector: FIDInlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: specialized for cresolWorkstation: Optima-3007

Page 14: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

Organic solvent:

VOCS2nd-level solution

50ppb thiopheneinBenzene

0.00 1.90 3.80 5.70 7.80 9.50 11.40 13.00 15.20 17.10 19.00

60.002

53.177

40.352

39.527

32.702

25.877

19.052

12.227

5.402

-1.423

-8.248

[pA]

Peaksequence:methyl alcohol, ethanol,acetone+ isopropanol, acetonitrile,dichloromethane,chloroform, isobutanolCCI4 Benzene, n-heptane, isoamylol,pyridine, toluene, n-amyl alcohol,ethylbenzene, p-xylene, n-Hexanol, o-xylene

Setup

0.00 3.00 6.00 9.00 12.00 15.00 18.00 21.00 24.00 27.00 30.00 33.00

50.000

45.000

40.000

35.000

30.000

25.000

20.000

15.000

10.000

5.000

0.000

-50.000

[pA]

Setup

Detector: FPDInlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: HP-InnowaxWorkstation: Optima-3007

Setup[pA]

30.000

25.000

20.000

15.000

10.000

5.000

0.000

-5.0000.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00

Detector: FPD

Peak sequence: methanol, ethanol,isopropanol, acetone, methyl acetate, n-butyl alcohol, butanone, ethyl acetate,acetic acid isopropyl ester benzene, 1-Methoxy-2-propanol, propyl acetat, 4-methyl-2-pentanone, 1-Ethoxy-2-propanol, toluene, n-butyl acetate,ethylbenzene, o-xylene, styrene

Inlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: PC-VOCOLWorkstation: Optima-3007

Detector: FPDInlet: CapillaryGas control module: 2 setsInjection mode: liquid autosamplerColumn: DB-624 cap.Workstation: Optima-3007

Page 15: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

HPLC Servicing :We have team of service engineers who can attend to any make of HPLC promptly @the most affordable cost. :We also take up preventive Maintenace to reduce downtime of HPLC’s Trainings

AMC’s/CMC :We offer user training both in-House and at customer sites on HPLC principles, operations, troubleshooting.Validations :We have protocols for carrying out periodic Validations as per GLP/GMP/U SFDA norms

Analytical Technologies is synonymous for offering technologies for doing analysis and is the Fastest Growing Global Brand having presence in at least 96 countries across the globe. Analytical Technologies Limited is an ISO :9001 Certified Company engaged in Designing, Manufaturing, Marketing & providing Services for the Analytical, Chromatography, Spectroscopy, Bio Technology, Bio Medical, Clinical Diagnostics, Material Science & General Laboratory Instrumentation. Analytical Technologies, India has across the Country operations with at least 4 Regional Offices, 6 Branch Offices & Service Centers. Distributors & Channel partners worldwide.

Instruments :We offer instruments / Renting Services Modules like pumps, detector etc. on Rent.

Support & Services

HPLC Servicing, Validation, Trainings and Preventive Maintenance :

About Analytical Technologies

About Analytical Technologies

Optical EmissionSpectrophotometer

DSC/TGA NOVA 2020 plusAutomated Bio Chemistry

Analyzer

HEMA 2020 Hematology

Analyzer

Micro PlateReader/Washer

Liquid ParticalCounter

Fully AutomatedCLIA

NOVA Basic Semi -AutoChemistry Analyzer

PCR/Gradient PCR/RTPCR

Blood GasAnalyzer

Random access Analyzerfor immunoassay

Proteins & clinical chemistry

Semen Analyzer Water puri�cationsystem

Total Organic Carbon

UV/VISSpectro 2060+Double Beam

FTIR Gas Chromatograph3000

Gas Chromatograph2979 Plus

FlashChromatograph

Atomic AbsorptionSpectrophotometer

!

URINOVA 2800Urine Analyzer

Page 16: GC optima 3007 (8 April) - AIS-India Optima 3007.pdf · Optima-3007 Clarity counter-control workstation Workstation Injection Value Headspace sampler Thermal desorption 6 - port valve

Regulatory compliances

Corporate Social Responsibility

Reach us @

Analytical Foundation is a nonprofit organizat (NGO) found for the purpose of:1.Research & Innovation Scientist’s awards / QC Professional Award : Quality life is possible by innovation only and the innovation is possible by research only, hence ANALYTICAL FOUNDATION is committed to identify such personallities for their contributions across various field of Science and Technoligy and awarding them yearly. To participate for award, send us your details of research / testing / publication at [email protected]

2. Improving quality of life by offering YOGA Training courses, Work shops / Seminars etc.

3. ANALYTICAL FOUNDATION aims to DETOXIFY human minds, souls and boby by means of yoga, Meditation, Ayurveda, Health Care, Awards, Media, Events, Comps etc.

Not

e : C

ompa

ny re

serv

es ri

ghts

to a

dd/d

elet

e/m

odify

the

cont

ents

/ te

chni

cal s

peci

ficat

ions

of th

e ca

talo

gue

with

out p

rior n

otic

e.

HPLC Solutions MultipleLabs Analytical Bio-Med Analytical Distributors Analytical Foundation (Trust)

Corporate & Regd. Office:Analytical House, # E67 & E68,Ravi Park, Vasna Road, Baroda,Gujarat 390 015. INDIA

T: +91 265 2253620 +91 265 2252839 +91 265 2252370F: +91 265 2254395

E: [email protected] [email protected] [email protected] [email protected]

W. www.ais-india.com www.analycalgroup.net www.hplctechnologies.com www.multiplelabs.com

Sales & Support Offices:across the country :Distributors & Channelpartners World Wide

VER.17/09-1