FUNDAMENTALS OF ULTRASONIC FLOW METERS.pdf

Embed Size (px)

Citation preview

  • 8/16/2019 FUNDAMENTALS OF ULTRASONIC FLOW METERS.pdf

    1/10

     PAGE 52 2002 PROCEEDINGS 

     AMERICAN SCHOOL OF GAS MEASUREMENT TECHNOLOGY 

    FUNDAMENTALS OF ULTRASONIC FLOW METERSKeven Conrad and Larry Lynnworth

    Panametrics, Inc.7255 Langtry, Houston, TX 77040-6626 and 221 Crescent Street, Waltham, MA 02453-3497

     ABSTRACT

    Ultrasonic contrapropagation methods have been usedto measure the flow of natural gas since the 1970s, flaregases since the 1980s, and smokestack gases in cem(continuous emissions monitoring) since the 1990s. Sincethe early 2000s, ultrasonic clamp-on flow measurements,previously restricted mainly to liquids, were foundeffective in measuring in standard steel pipes, the flowof steam, natural gas and other gases and vapors,including air, as long as the flow velocity was not so highas to cause excessive beam drift or excessive turbulence(in other words, below about Mach 0.1), and provided

    the acoustic impedance of the gas was equivalent to airabove about six bar and no important molecularabsorption or scattering mechanisms were present.

     Although the flow of gases by ultrasonics has long beenthought to be more difficult to measure than liquids, infact the measurement is easier in two important respects.One is, for the contrapropagation method, the upstream- downstream time difference is generally much greaterfor gases, as a consequence of the much lower soundspeeds in gases compared to liquids. The othersignificant factor that becomes important in mass flowmetering (including scfm output) is the existence oftheoretical and/or empirical relationships betweenultrasonic propagation and density, where either of such

    relationships is easier to exploit for gases than for liquids.To provide an idea of the scope of applicationsaddressable with ultrasonic technology that iscommercially available now or likely to be available inthe near future, this paper starts with an analysis fromthe point of view of acoustic impedance; considersdesigns as a function of the number of nozzles, fromzero to a dozen; and lists factors conducive to highaccuracy versus factors detrimental to high accuracy,i.e., conducive to uncertainty.

    INTRODUCTION

    Ultrasonic contrapropagation methods have been used

    to measure the flow of natural gas since the 1970s, flaregases since the 1980s, and smokestack gases in cem(continuous emissions monitoring) since the 1990s.Contrapropagation means sound waves are timed in adirection with the flow and later or simultaneously, againstthe flow. At low Mach number,

  • 8/16/2019 FUNDAMENTALS OF ULTRASONIC FLOW METERS.pdf

    2/10

    2002 PROCEEDINGS PAGE 53

     AMERICAN SCHOOL OF GAS MEASUREMENT TECHNOLOGY

    molecular weight times pressure. At high molecularweight and/or high pressure, these simple (linear)approximations are inadequate. Virial equations andsupercompressibility provide a remedy.

    To provide an idea of the scope of applicationsaddressable with ultrasonic technology that iscommercially available now or likely to be available in thenear future, this paper starts with a short theoretical section

    in which the contrapropagation equations are derived. Thisis followed by profile considerations. Then we go toacoustic impedance. Lastly, we consider designs as afunction of the number of nozzles, from zero to a dozen;and list factors conducive to high accuracy versus factorsdetrimental to high accuracy, i.e., conducive to small orlarge uncertainty. There are other “fundamentals” besidesthese. However, to keep this paper to reasonable length,only these topics are treated. The references, particularlythe 2002 paper by Walters et al. [15] deal with aspectsbeyond the scope of this paper.

    DISCUSSION

    Theory 

    1. Equations for a Contrapropagation Flowmeter

    In the contrapropagation method, ultrasonic (orsometimes audible) waves are transmitted upstream anddownstream. From the transit times t 

    1  and t 

    2  in each

    direction, and knowledge of the path and flow profile,the average flow velocity V 

     A  is determined. A rather

    simple derivation of the basic flow-sensing equation ispossible if one imagines a fluid of sound speed c flowing

     at a uniform velocity V < c in a duct of cross-sectionalarea  A, interrogated by two point sensors on the axisand spaced a distance L apart. The transit times in theupstream and downstream directions, respectively, are

      t 1 = L/(c – V) and t 

    2 = L/(c + V). (1 & 1a)

    The reciprocals of these transit times, when multipliedby the axial projection of the axial interaction path L, are

      L/t 1 = c – V and L/t 

    2 = c + V. (2 & 2a)

     Accordingly,

    The upstream-downstream time difference can beobtained from Equation (1) as ∆t  = 2LV  /( c2-V 2 ). This canbe expressed in terms of the Mach number M

    N =V/c for

    MN  10 megarayl range. Gases are in the 0.1 to 1kilorayl range. [Plots for gases only are given in Figs.1(b-d). When c yields average MW  (molecular weight) ordensity, mass flowrate M

    F  is not far behind.]

    Entering a gas on such a chart requires that the gas’sound speed and density are known. Entries are usuallyassociated with STP (0˚ C, 760 mm Hg). The locus formethane, if examined carefully, reveals a slight increasein c  as P  increases, as well as the expected nearlyproportionate increase in density ρ as P increases.

    ]L 1 1 L t 2 t 2

    t 1

    2 t 1t 2

    ]   [c = =[

    L 1 1 L ∆t 2 t 

    2t 1

    2 t 1t 2

    ]   [V = =[   ] (3)and

    (4)

  • 8/16/2019 FUNDAMENTALS OF ULTRASONIC FLOW METERS.pdf

    3/10

     PAGE 54 2002 PROCEEDINGS 

     AMERICAN SCHOOL OF GAS MEASUREMENT TECHNOLOGY 

    The departure from linearity between density andpressure is often treated by a supercompressibility factor.See for example, pages 29-52 in the chapter by BillBuzzard in David W. Spitzer’s 2001 book cited in [2].

    One of the ways of overcoming some of the largemismatch in impedances for gases at or near atmosphericpressure is to build the transducer with a quarterwaveimpedance matcher between the solid piezoelectric (of

    high Z  ) and the gas. Figure 2(a) shows a T7 transducerbuilt this way. To preserve the transducer from unwantedeffects of the “gas,” a Ti, SS or other thin metal layersurrounds the piezo and matcher, in these designs [8; 8(b)].Sometimes, as in cem (continuous emissions monitoring),one transducer is not enough, so arrays are used [Fig. 2(b)].Fiberacoustic “BWT” bundled waveguides [6] have beenused in many gas applications (up to ~ 500˚C) without any matcher [Fig. 2 (c), (d)].

    This matching idea works if the transducer is “wetted,” i.e., in contact with the gas as in Figure 2(a). In clamp-on(Figure 3) the transducer wedge contacts the steel (orother material), not the gas, so another solution is

    necessary to overcome the mismatch. Clamp & installation details are given in Fig. 3(b-d).

    One might say the low Z  of a gas is mostly due to its lowdensity, compared to steel or other elastic solids. We canalso ask, are there special consequences of the low c of agas, say methane, compared to steel? One consequenceis, in clamp-on, the angle of the sound beam refractedinto the gas is small, say 5 to 8 deg, according to acalculation using Snell’s Law of Refraction. This implies,a V  measurement along a path only a few degrees offnormal, is going to be very sensitive to crossflow. Ifcrossflow is significant, crossed clamp-on paths as shownin Figure 3(f) may be necessary. A second consequenceof low c and small refracted angle is that paths off thediameter (as seen in the end view) are not reachable withtoday’s technology. This means a clamp-on quadratureis out of reach. Exception: hybrid [Figure 4(g)].

    The “good” news associated with low c  is the timedifference ∆t   between upstream and downstreaminterrogations is larger, typically in the many (tens of)microseconds range, compared to liquids, where∆t  maybe just a few microsences and must be resolved to 1 nsor less. On the other hand, the attenuation coefficient ingases is usually much higher than in liquids, and alongwith impedance mismatch, constrains the ultrasonicfrequency to be < 1 MHz for gases, whereas for liquids ƒ> 1 MHz is common.

    How long have we waited for a practical clamp-onultrasonic flowmeter for gases to become availablecommercially? One answer is thirty-nine years. Thisanswer is based on (a) the GC868 announcement in [9],and on (b) the last paragraph in [14]. That 1966 patent isone of the earliest (perhaps the earliest) clamp-on U.S.patents for  liquid  flow, and claims a Japanese prioritydate of 1962.

    4. Designs As a Function of the Number of Nozzles

    Zero Nozzles. This means clamp-on (Figure 3). Fordetails see [1 & 9] or www.panametrics.com, PCI or PCIR&D pages, Ultrasonic Report UR-264. Accuraciesaround 2% have been achieved with one path. Crosseddiameters can improve the accuracy, as in [13]. As ofmid-2002, applications were restricted in general to metalpipes of diameter >3 inches (>75 mm) but less than about

    2 ft (600 mm). These are guidelines. At high pressure, ifV  is not too high, gas flows in pipes as large as 30 incheshave been measured by clamp-on. Bad pipe, heavy wall,noisy environment, and/or access limited to one side,can frustrate the method. Success depends on a numberof factors, including the state of the equipment, e.g., itssoftware version, as well as pipe and flow conditions.

     As has been the practice for liquids since the 1960s and1970s [11, 13, 14], measurements in ⊥ planes (X in endview) yield better accuracy than measurements in oneplane only. Unlike liquids [Fig. 3(a), parts (1) & (7)], thevee path, with both transducers on the same side of thepipe, is hard to use when the fluid is a gas, because of

    too much acoustic crosstalk. Sometimes crosstalk iscalled acoustic short circuit noise.

    One Nozzle.  Insertion probes can go in oblique ornormal. Some ultrasonic solutions use an intrusivereflector but the transducers themselves are flush to theID or recessed by up to a foot (300 mm) or so if the gasis hot. The advantages of the one-port compared to >1port are economy and ease of installation. Drawbacksare accuracy, to the extent the profile is uncertain. Theone-port samples only a small chord segment. If thereflector is located about 60% of the radius in from thewall, and to the extent the profile obeys a Nikuradsepower law, then the flow averaged from the wall to thatreflector nearly equals V  AVG. However, the length of thechord segment, in the end view, is only 0.3D  whereD  = pipe inside diameter. This small sample may becontrasted with multipath sampling for a 12-nozzlespoolpiece referred to at the end of this section. One-ports are shown in [15].

    Two Nozzles. Examples from the “flare gas world” aregiven in [10 & 15]. The bias 90 arrangement samples ashort chord segment but seems to provide adequateaccuracy. It is probably the most common hot-tapgeometry for flare gas flow measurements.

    Twelve Nozzles. Custody transfer spoolpieces shownin Figure 4(d) use crossed paths (hence four transducersper plane) in each of three parallel planes (hence 12transducers total).

     Accuracies (by which we mean uncertainties) are 0.5%or better, 0.3% in some tests. By using transducersshown in Figure 4(e), tests can be run in air at atmosphericpressure. Air calibration allows relatively economicaltesting under various disturbed-flow conditions such asupstream elbows, diameter changes, partly open valve,

  • 8/16/2019 FUNDAMENTALS OF ULTRASONIC FLOW METERS.pdf

    4/10

    2002 PROCEEDINGS PAGE 55

     AMERICAN SCHOOL OF GAS MEASUREMENT TECHNOLOGY

    or their combinations. [3] Quadrature integration (Figure4) is intended to suppress the effects of flow profile. The

     measurement of secondary flow, especially circulation,is dealt with in [5].

    TABLE 1.Factors Conducive to High Accuracy Versus Factors Detrimental to

    High Accuracy, i.e., Conducive to Uncertainty 

     High accuracy if: Low accuracy or uncertain results if:

    Long straight run, steady flow, single phase Nearby upstream disturbances; intermittently two-

    phase or multiphase

    Quiet environment, no electrical interference Noisy environment; crosstalk unavoidable

    Good pipe (concentric ID & OD), roughness known Rough or scaled pipe, roughness unknown and subjectand not changing to changes over time; gas fouls transducers

    Gas is known and does not exhibit molecular Highly attenuative, e.g., carbon dioxide, halogens;absorption; T & P allow good transmission; P too low or T  too high, such that absorption due toequipment provides adequate reciprocity (signals classical viscosity and thermal conductivity effectsfrom A → B look like signals B → A) is high

    Flow velocity is in a range such that the ∆t  falls into V  is too low or too high to be accurately and reliablyan easily measured range; no cycle skip; no packet measured by the available sensors and electronicsskip; no ambiguity about arrival time

    FIGURE 1. Some basic acoustic ideas and facts for gases. (a) Impedance nomogram for gases, liquids and solids. (b) Inthis ρc plot for gases, the c’s range from ~100 to 1300 ms-1, the MW ’s from 2 to 240, but γ , the ratio of specific heats, liesbetween 1 and 1.67. The effect of gas absolute temperature T  is shown on c

     Ar  when argon is heated at constant density,

    drawn assuming c Ar 

     increases in proportion to T 1/2. (c) Flare gas: empirical relation between c at 38C˚ and average molecularweight (  MW  ) fo r 2≤MW ≤ 58. (d)  Ai r dens it y ρ  vs temperature T .   Sound speed c  vs T . ➂  ρ  vs c.(  P  = 760 mm for graphs -➂.) Note that even if RH  (relative humidity) is not known, between 0˚C and 60˚C, c yields withsmall uncertainty. For details see Matson, J., Mariano, C. F., Khrakovsky, O., and Lynnworth, L. C., Ultrasonic MassFlowmeters Using Clamp-On or Wetted Transducers Proc of the 5th International Symposium on Fluid Flow Measurement (April 7-10, 2002) or refer to the website of the authors’ firm, where Ultrasonic Report UR-240 may be posted.© 2002 Panametrics.

  • 8/16/2019 FUNDAMENTALS OF ULTRASONIC FLOW METERS.pdf

    5/10

     PAGE 56 2002 PROCEEDINGS 

     AMERICAN SCHOOL OF GAS MEASUREMENT TECHNOLOGY 

    TABLE 2.Diameter, midradii, their combination, tomographic and quadrature arrangements.Complex flow patterns, high-accuracy, motivate multipaths on or off the diameter.

  • 8/16/2019 FUNDAMENTALS OF ULTRASONIC FLOW METERS.pdf

    6/10

    2002 PROCEEDINGS PAGE 57

     AMERICAN SCHOOL OF GAS MEASUREMENT TECHNOLOGY

    FIGURE 2. Some transducer fundamentals. (a) T7 “air transducer” for 100-kHz operation has a flange OD = 19 mm and isinternally impedance-matched [8]. (b) Array of sources as used in cem (continuous emissions monitoring) applications[4]. (c) Left: BWT® bundle waveguide icon represents a fiberacoustic waveguide. Right: Close-up photo of a BWTtransducer. (d) Details of an early BWT transducer from the mid-1990s [6].

  • 8/16/2019 FUNDAMENTALS OF ULTRASONIC FLOW METERS.pdf

    7/10

     PAGE 58 2002 PROCEEDINGS 

     AMERICAN SCHOOL OF GAS MEASUREMENT TECHNOLOGY 

    FIGURE 3. (a) Refraction with clamp-on, calculated assuming a refracted angle θ2 of 60˚ for a shear wave in the steel pipe,

    depends on sound speed c3 in the fluid. Referring to items (1)-(3) commercial contrapropagation clamp-on flowmetersavailable since the early 1990s include the PT868 and 6068 for measuring the flow of liquids. In water , the vee path usually works, and θ

    3 is about 25˚ at room temperature. In air, θ

    3.AIR is only about 6˚ and the transducers usually need to be placed

    on opposite sides of the pipe. (5) For liquid clamp-on, the vee path [shown in (1)] tends to cancel crossflow as well asdouble the sensitivity to flow compared to a single traverse. For gases, odd numbers of traverses are preferred, to reducecrosstalk. This means, if crossflow is significant, crossed paths are recommended. The velocities measured along thelegs of the X should be averaged. Best solution: find a long straight run far from disturbances and joints. For  gases[diagrams (4) & (5)] the flowmeter instrument (6) introduced in 2001 is the GC868 (Ao, 1999; Ao, et al. (2002); Lynnworth2001). Diagram (7), drawn for LOX or LN2, shows the refracted angle θ

    3 ≈ 16˚, nearly midway between water (25˚) and air

    (6˚). (b) Clamp and gas paths, schematic. (c) Clamps and instrument for clamp-on gas flow measurement [1, 12]. (d) Application on steam.© 2001 Panametrics.

  • 8/16/2019 FUNDAMENTALS OF ULTRASONIC FLOW METERS.pdf

    8/10

    2002 PROCEEDINGS PAGE 59

     AMERICAN SCHOOL OF GAS MEASUREMENT TECHNOLOGY

    FIGURE 4. (a) Example of a solid PanAdapta® precision plug for liquids. This plug is not  to be removed when the line ispressurized as it is a permanent part of the pressure boundary. Its outer surface is prepared to receive a removabletransducer. Frequency range: 0.5 to 5 MHz. (b) Similar to preceding case except the plug for gases consists of a bundleof thin waveguides welded within a sleeve. The bundle plug can be as short as one inch or as long as several feet (25.4mm to ~1 m). This sleeved sealed waveguide construction allows the removable piezoelectric transducer assembly to beseparated from the buffer bundle. (c) Schematics show planes of measurement in end view, and the crossed paths in athree-dimensional SolidWorks rendition. (d) Photos of a spoolpiece manufactured by RMG and corresponding to theschematics in (c). (e) Example of a T11 transducer. It is Ti-housed, internally quarter-wave matched, and twelve of themare used in (d). (f) Plugged version of (c). (g) Liquid version of the concept shown in (f).

  • 8/16/2019 FUNDAMENTALS OF ULTRASONIC FLOW METERS.pdf

    9/10

     PAGE 60 2002 PROCEEDINGS 

     AMERICAN SCHOOL OF GAS MEASUREMENT TECHNOLOGY 

    Schematic

    FIGURE 5. Plastic pipe, air flow at atmospheric pressure. These measurements, made by our colleague Oleg Khrakovsky,used ordinary liquid  flowmeter clamp-on equipment, which was sufficient because the pipe was plastic, not steel.

    FIGURE 6.Short, thick-walled N-path spoolpiece whose OD matches

    flange raised face dimensions. The “flanged transducer”is the T7 air transducer [ 8(c)] shown in Figure 2(a).

    FIGURE 7.Hybrid example: “cow” = clamp-on + wetted

    transducers.

     ACKNOWLEDGMENTS

    The work reported here includes important contributionsfrom Shirley Ao, Jim Hill, and their and the authors’colleagues. The authors acknowledge Panametrics’permission to reproduce passages, tables andillustrations from its copyrighted reports including

    UR-273 and -274.www. Saul Jacobson, Toan Nguyen,David Hesketh, Jed Matson, Paul Ceglia and others atPanametrics and Hans J. Kastner, Andreas Weber andothers at RMG contributed to the equipment in Figure 4(d).The manuscript was prepared by Lin L. Leeming.

  • 8/16/2019 FUNDAMENTALS OF ULTRASONIC FLOW METERS.pdf

    10/10

    2002 PROCEEDINGS PAGE 61

     AMERICAN SCHOOL OF GAS MEASUREMENT TECHNOLOGY

    BIBLIOGRAPHY AND REFERENCES

      1. Ao, X., Matson, J., Kucmas, P., Khrakovsky, O., andLi, X. S., Ultrasonic Clamp-On Flow Measurementof Natural Gas, Steam and Compressed Air,Proceedings of the 5th ISFFM (5th International Symposium, Fluid Flow Measurement)  (April 7-10,2002).

      2. Brown, A. E., Ultrasonic Flowmeters, in Spitzer, D.W. (Editor): Flow Measurement,  pp. 415-442, ISA (1991); (b) Brown, A., and Lynnworth, L., UltrasonicFlowmeters, in Spitzer, D. W. (Editor): FlowMeasurement - Practical Guides for Measurement Control, 2nd Edition, Ch. 20, pp. 517-575, ISA (2001).

      3. Hill, J., Weber, A., and Koyama, T., Qualification ofUltrasonic Flowmeters for Custody Transfer ofNatural Gas Using Atmospheric Air CalibrationFacilities, submitted for Proc. 20th North Sea FlowMeasurement Workshop, St. Andrews, Scotland(Oct. 22-25, 2002).

      4. Jacobson, S. A., Flow Measurement SystemIncluding Ultrasonic Transducers, U. S. Patent5,460,047 (October 24, 1995).

      5. Johari, H., and Durgin, W. W., Direct Measurementof Circulation Using Ultrasound, Experiments inFluids 25 pp. 445-454 (September 1998). (Later andearlier related references are in [15].)

      6. Liu, Y., Lynnworth, L. C. and Zimmerman, M. A.,Buffer Waveguides for Flow Measurement in HotFluids, Ultrasonics, 36 (1-5) pp. 305-315 (February1998); (b) Liu, Y., and Lynnworth, L.C., U. S. Patent5,962,790 (October 5, 1999); (c) CIP (continuation-in-part) U. S. Patent 6,343,511 (February 5, 2002).

      7. Lynnworth, L. C., Ultrasonic Flowmeters, Chap. 5,in Physical Acoustics - Principles and Methods,Mason, W. P., and Thurston, R. N., (Eds.), 14 pp.407-525, Academic Press (1979); (b) UltrasonicMeasurements for Process Control Theory,Techniques, Applications, Academic Press (1989);(c) with Magori, V., Industrial Process Control Sensorsand Systems, Chapter 4, pp. 275-470 in: Papadakis,E. P., (Guest Editor), Ultrasonic Instruments and Devices: Reference for Modern Instrumentation,Techniques, and Technology , 23 in the series Physical 

     Acoustics, Academic Press (1999).

      8. Lynnworth, L. C., Patch, D. R. and Mellish, W. C.,Impedance-Matched Metallurgically SealedTransducers, IEEE Transactions on Sonics and Ultrasonics, SU-31  (2) pp. 101-104 (March 1984);(b) Lynnworth, L. C., Fowler, K. A. and Patch, D. R.,Sealed, Matched Piezoelectric Transducer, U. S.Patent 4,297,607 (Oct. 27, 1981); (c) LCL, UltrasonicTransducer System with Crosstalk Isolation, U. S.Patent 5,515,733 (May 14, 1996).

      9. Scelzo, M., A Clamp-on Ultrasonic Flowmeter forGases, Flow Control 7 (9) pp. 34-37 (Sept. 2001).

    10. Smalling, J. W., Braswell, L. D., Lynnworth, L. C. andWallace, D. R., Flare Gas Ultrasonic Flow Meter,Proceedings 39th Texas A&M Annual Symposium onInstrumentation for the Process Industries, pp. 27-38 (January 17-20, 1984); (b) Smalling, J. W.,Braswell, L. D. and Lynnworth, L. C., Apparatus and Methods for Measuring Fluid Flow Parameters, U.S. Patent 4,596,133 (June 24, 1986); (c) U. S. Patent4,754,650 (July 5, 1988); (d) U. S. Patent 4,856,321(August 15, 1989).

    11. Suzuki, H., Nakabori, H., and Yamamoto, M.,“Ultrasonic Method of Flow Measurement in LargeConduits and Open Channels,” p. 11538 in C. G.Clayton (Ed.), Modern Developments in FlowMeasurement, Peregrinus, London (1972).

    12. Ting, V. C. and Ao, X., Evaluation of Clamp-OnUltrasonic Gas Flowmeters for Natural Gas

     Appl ica tio ns , Proc. 20th North Sea FlowMeasurement Workshop, St. Andrews, Scotland(Oct. 22-25, 2002).

    13. Yamamoto, M. and Amano, A., Ultrasonic FlowQuantity Measuring System, U. S. Patent 3,555,899(Jan. 19, 1971).

    14. Yamamoto, M., and Ito, K., Ultrasonic Flowmeter System, U. S. Patent 3,237,453 (March 1, 1966).

    15. Walters, J., Smalling, J. W., Ao, S., Hill, J. andLynnworth, L. C., Transit-time Ultrasonic Flowmetersfor Gases, presented at and published in the Proc.

     41st Annual CGA (Canadian Gas Association) GasMeasurement School , Grand Okanagan, KelownaBC, Canada (June 4-6, 2002). See also,UR-274.www, a Panametrics website version of thispaper.

    ©2002 Panametrics. All rights reserved. Version May 2, 2002