Page 2: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

GENERAL OVERVIEW Fume hoods are one of the most important components used to protect laboratory personnel from exposure to hazardous chemicals and agents used in the laboratory. When used properly, they can provide an effective backup safety device for the containment and exhaust of toxic, offensive or flammable materials when the containment of an experiment or procedure fails and vapors or dusts escape from the apparatus used.

Average face velocity is the main criterion that is used to determine if a hood is functioning satisfactorily as a safety device for the containment and elimination of airborne contaminants. Factors that can affect face velocity and thus hood performance include baffle position, cross drafts, air volume, hood features, hood location, foot traffic and user work practices. STANDARD CONFORMANCE AND COMPLIANCE The design and performance of fume hoods are generally conform to the following international standards:-

SEFA 1-2002 ASHRAE 110-95 AS2243-8 EN 14175 BS 7258

FUME HOOD CONSTRUCTION Fume Hood Superstructure In general, the fume hood shall come with worktop, sash, base storage cabinet, control and display panel, electrical distribution board, lighting encased with a vapor proof casing and utility facilities. Where the construction material used is concern, it must be selected with a predominant view to safety in a hazardous environment. The fume hood shall have dual wall construction with an air gap within for safety. Materials used must be resistant to chemicals, heat abrasion, impact vandalism and fungi. The fume hood design needs to have access for easy decontamination, cleaning and maintenance. All electrical fittings shall be properly insulated and exposed surface shall be electrically earthed to the earth terminal of the isolator. Typically construction acceptable for fume hood shall have an external wall of powder coated epoxy steel material or fire resistant glass-fiber reinforce plastic and an internal chamber molded, one piece, fire resistant fiberglass material. It shall be furnished with clear lamination, safety glass of 6mm thickness, tampered proof sash assembly with counter balanced weight. The fume hood shall be constructed with a well designed baffle and internal lining with coved corners. Materials used in the construction of the fume cupboard working space shall be able to minimize resistant to air flow and to avoid turbulence, backflow and air eddied. Fluctuation in the sash velocity as the sash is raised or lowered should be minimal.

FUME CUPBOARD GENERAL OVERVIEW

Page 3: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

Work Top The epoxy table top shall be dished at certain angle to prevent spill over. The worktop shall be 900mm form final floor level and have resistant to solvents, acids, heat and impact. A suitable sink shall be recessed in the left corner beneath the gooseneck faucets. General Purpose Base Cabinet The storage cabinet shall be of metallic type and come with removable shelf and rear panels to provide access to service plumbing. The shelf and bottom shall be lined ribbed rubber matting. The toe space shall be protected by coved rubber molding. Doors shall be close fitting type and no parts of the cabinet shall support combustion. Air vent piping shall be provided to extract any escaped fume from within the base cabinet storage area. Air vent pipe shall be permanently connected to the main exhaust PVC pipe from the fume hood. Polypropylene material sheet or drip tray shall be provided in base cabinet shelf to protect shelving from coming into contact with any split chemical solution. Fume Hood Utility All service valves or devices shall be fitted to the sides of the fume hood fascia. All handles shall be fitted with international coding system.

FUME CUPBOARD GENERAL OVERVIEW

Page 4: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

FUME HOOD PERFORMANCE The fume hood shall be of a demonstrable aerodynamic design capable of containing effluent generated within the hood at a face velocity of at least0.5 m/sec. The interior lighting shall have a minimum of 700 lux at worktop level. GENERAL SPECIFICATION Material Hood Exterior is epoxy coated steel

Front Panel Easily removed without tools for easy accessibility

to electrical wiring facilities. Front End Fixture Panel Quickly and easily removed to provide convenient

to service fixture from the front of the hood. Side panel Convenient access to sash weight, service fixture

and plumbing lines. Inner Liner Fiberglass/stainless steel Sash 6mm double layered glass with safety lamination Exhaust Point The exhaust outlet reduces static pressure loss to

ensure smooth, efficient airflow. The integral condensation collection rings prevents condensed vapors in the ductwork from entering the work area and channels them to drain or collection container via the rubber tubing supplied.

Vapor Proof Light Casing Shielded with glass plate properly sealed. Lights Uses energy saving fluorescent light. Switches Two rocker switches located with easy reach of

the operator control both the internal lights and blower.

Service Fixtures Remote valve with side outlet Fixture Lines Water fixtures are plumbed with copper tubing

while gas is plumbed with tin-lined tubing Electrical Outlets 220V, 13Amp single-phase electrical outlets are

mounted on the front-end panel Table Top Epoxy table top specially designed to provide

chemical resistant inside the fume hood Table Top Material & Construction One-piece corrosion resistant from a special

formulated of epoxy resins to withstand chemicals. Dished and contoured

CERTIFICATION In providing a fume hood that serve the purpose as a protection shield to laboratory personnel from a hazardous chemicals and agents in laboratory, ASL fume hoods have been tested by independent laboratory and certified conformed to the international standards of ASHRAE 110-95 PROTOCOL.

SL

FUME CUPBOARD Performance/ Specification/ Certification

Page 5: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

By-Pass

Light switch

Blower switch

6mm thick double coated sash

Emergency push button

Spare

Service control

13Amp socket outlets

Fiberglass Inner Liner

Fume Outlet

2 x 36WFluorescent

Lights

Services

Duratop EpoxyDrip Cup / Sink

Anti-siphon PP Bottle trap

FUME CUPBOARD Constant Volume/ Bypass General Purpose

Page 6: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

Fume HoodController

Airflow SensorFumeOutlet

Airflow Sensor CMS electronic device Individual circuit testing INERIS approval Conform to CEI 1000-4-2 Conform to EN 55022

ADVANTAGES• Easy maintenance

• Energy Saving

• Safety

• User Friendly

Variable Speed Drive Controller

Spare

Service control

6mm thick double coated sashEmergency push button

Fiberglass Inner Liner

13Amp socket outlets

2 x 36WFluorescent

Lights

Services

Duratop EpoxyDrip Cup / Sink

Anti-siphon PP Bottle trap

FUME CUPBOARD – Variable Speed Drive General Purpose

Page 7: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

Light switch

Blower switch

6mm thick double coated sash

Emergency push button

Spare

Service control

13Amp socket outlets

Fiberglass Inner Liner

Venturi Valve

2 x 36WFluorescent

Lights

Services

Duratop Epoxy Drip Cup / Sink

Anti-siphon PP Bottle trap

Fume Hood Monitor

Sash Sensor

Fume Outlet

Fume Hood Monitor For VAV System

VAV Output Bar Graph or Alarm Time Line

Velocity Display m/sec or fpm

Control Pushbuttons

LED Indicators

Function and up/down buttons for Menu Configuration and Calibration.ENTER – also used as Mute button for audible alarm

FUME CUPBOARD – Variable Air Volume General Purpose

Page 8: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

EmergencyPushdown Button

DURATOP Epoxy resin worktop 19 or 25 mm thick with integral 6 mm marine edge

RecirculationPump

NeutralizingTank

Demister

Wash-down Unit

Packing

Air FlowController

Electrical Socket

Scrubber Unit

Scrubbed liquor to Neutralizing tank

Stainless SteelInner Liner

Ball Valve

IncomingServices

Overflow Pipe

Drain

Perchloric Acid Fume Hood is used when work with perchloric acid. The inner line should be seamless to prevent potential trap of the fume / particles that will crystallized. Fume from the hood will go through a scrubber unit that will wash down the fume before it goes to the ducting system. These type of fume hood need dedicated ducting and exhaust system. VSD or VAV system can be used together with the fume hood to increase protection of the operator.

FUME CUPBOARD – Perchloric Acid Hoods

Page 9: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

Electrical Socket

6mm thick DoubleCoated laminated sash

Air Flow Controller

High EfficiencyParticles Filter

Pressure Differential MeterPre-Filter

Fan

Bag-in / Bag-outFilter

Non-returnDamper

FlexibleDuct

Services

EmergencyPushdown Button

Services Remote Control

Seamless covedInner Liner

DURATOP Epoxy resin worktop 19 or 25 mm thick with integral 6 mm marine edge

Radioisotope fume hood should be designed too efficiently ventilate procedures using radio-chemicals. The fume hood must be able to provide a flow of air sufficient to extract the fumes from within the work area and keep away from the operator to avoid ingestion, inhalation or skin contact. Hood interior is constructed of seamless, ground and polished 304 or 316L stainless steel with covered interior corner. Free of joints, cracks or gaskets to protect against spills and radioactive build-up. The hood comes with integral cup-sink which can be used to dispose of liquids safely and may act as a drain during cleaning procedure.

FUME CUPBOARD – Radio Isotope Hood

Page 10: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

All electrical component, semi-conductor or component that can produce spark should be isolated or at the external of the fume hood. Lights should be housed in an explosion proof casing. Sash material should be an explosion proof material.

FUME CUPBOARD – Explosion Proof Hood

Page 11: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

FUME CUPBOARD – Acid Digestion Hood

Page 12: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

WALK-IN FUME HOOD

STUDENTWORKSTATION FUME HOOD

FUME CUPBOARD – Walk-in and Student Hood

Page 13: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

DURATOP EpoxyWORK SURFACES

FUME CUPBOARD – Worktop Surface & Base Cabinet

For use with fume hood superstructures and base cabinets. DURATOP modified epoxy resin surfaces are 19 or 25 mm Thick with integral 6 mm molded or applied Marine Edge.

Special cupsink cutout must be specified if alternative cup sinks are selected. DURATOP Epoxy Cupsinks 3” x 6” or 3” x 9” (76 mm x 152 or 76 mm x 229)

Model No. Depth Width_ EPT1-48 762mm (30”) 1219 mm (48”) EPT1-60 762mm (30”) 1524 mm (60”) EPT1-72 762mm (30”) 1829 mm (72”)

Other designs and molded shapes available upon request.

Flush front panel above cupboardOne adjustable shelfFiller end panel for closing service at rear of base cabinet

Product No. Width_ RFB1-23 584 mm (23”) RFB1-30 750 mm (30”) RFB1-36 914 mm (36”) RFB1-48 1194 mm (48”)

BASE CABINETS

Page 14: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

Project ReferenceLABORATORY FUME CAPBOARD

Project: ERECTION OF 1 BLOCK SCHOOL OF BIOLOGICAL SCIENCES COMPRISING PART 2/4/5 – STOREY WITH 4 BASEMENTS AND 1 BLOCK OF PART 1 AND 2 STOREY ANIMAL HOUSE FOR NANYANG TECHNOLOGICAL UNIVERSITY (NTU).

Side to side & Back to back 6 FT Fume Cupboard

Combination of 4 & 5 FT Fume Cupboard

Project: DELFI COCOA - MALAYSIA

5 FT Fume Cupboard

FUME CUPBOARD MOTOR & BLOWER AT ROOFTOP

6 FT Fume Cupboard

GP 4 FT Fume Cupboard

Page 15: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

Project ReferenceLABORATORY FUME CAPBOARD

5FT EXPLOSION PROOF FUME HOOD WITH FLAMMABLE CABINET

5FT GENERAL PURPOSE FUME

HOOD

4FT GENERAL PURPOSE FUME

HOOD

Project: PROPOSED ERECTION OF A 5-STOREY ANIMAL & PLANT HEALTH CENTRE AT AGRI-BIO PARK, MK 12, LOT 10-3

PT, PLOT 9 & 8 PT, LIM CHU KANG ROAD, SINGAPORE

5FT PERCHLORIC ACID HOOD WITH SCRUBBER

ERECTION OF A 5-STOREY ANIMAL & PLANT HEALTH CENTRE AT AGRI-BIO PARK, MK 12, LOT 10 – 3PT. PLOT 9 & PT, LIM CHU

KANG ROAD, SINGAPORE (AVA)

Supply and Install of the following:02 Units – 4ft General Purpose Fume hood01 Unit – 5ft General Purpose Fume hood01 Unit – 5ft Explosion Proof Fume hood01 Unit – 6ft Explosion Proof Fume hood01 Unit - 5ft Perchloric Acid Fume hood 06 Units of Fume hoods

Page 16: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

Disclaimer This catalogue is just a basic guide to Free Standing laboratory casework system. The types, sizes, materials and components are not limited to what mentioned in this catalogue. The main purpose is to understand how the system was made-up and material selection of most consistently demanded in the industrial, medical, research and teaching facilities. Physical environment functionally and aesthetically conducive must be considered in proposing the laboratory casework. Designing laboratory casework can be considered as a trendy work, it needs to suit the rapid changes research, technology, equipments development and evolution, shape, size, laboratory application and end the user taste. Refer to the latest product catalogue and latest manufacturer catalogue for any changes in materials size, limitation and features.

Page 17: FUME CUPBOARDS CUPBOARDS ASHRAE 110-95 CERTIFIED TOTAL LABORATORY SOLUTIONS

Total Laboratory Solutions

Science-Lab Fume Hood products are used in:

• Electronics industry

• Chemical industry

• Pharmaceutical industry

• Schools

• Hospitals