Preventive Maintenance Checklist Equipment : Electrocardiograph Date :__________________ Merk : Period : __________________ Serial No. : _________________________ Room :__________________ Visual Inspection :Disconnect the cardiograph form AC power and inspect for the Following : L Loss or missing hardware ____________________________ Frayed or damage wiring ____________________________ Mechanicaldamage ____________________________ Evidence of liquid spill ____________________________ Printer drive gear wear ____________________________ Printer roller wear ____________________________ Wear or damage topower cord and Associatedstrain relief ____________________________ Corroded or damage electrodes ____________________________ Damage lead wires or patient module cable____________________________ Dirt on thermal printer head ____________________________ Connect the cardiograph to AC power and turn on the AC switch. Verify the following : The AC indicator islit ____________________________ One or more green battery indicator are lit when On-Standby is pressed Turn on the cardiograph ____________________________ Extended Self-test : Run Extended self-test, select “ALL” menu choice and verify that each test passes with no errors. Patient module and cable ____________________________ CPU assembly ____________________________ Printer ____________________________ Preview display ____________________________ Keyboard display ________________________

Formppp Ecg

Embed Size (px)

DESCRIPTION

dok

Citation preview

Rumah Sakit Siloam Gleneagles

Preventive Maintenance Checklist

Equipment: Electrocardiograph

Date

: __________________

Merk

:

Period : __________________

Serial No.: _________________________Room

: __________________

Visual Inspection :Disconnect the cardiograph form AC power and inspect for the

Following :

LLoss or missing hardware

____________________________

Frayed or damage wiring

____________________________

Mechanical damage

____________________________

Evidence of liquid spill

____________________________

Printer drive gear wear

____________________________

Printer roller wear

____________________________

Wear or damage to power cord and

Associated strain relief

____________________________

Corroded or damage electrodes

____________________________

Damage lead wires or patient module cable____________________________

Dirt on thermal printer head

____________________________

Connect the cardiograph to AC power and turn on the AC switch.

Verify the following :

The AC indicator is lit

____________________________

One or more green battery indicator are lit when On-Standby is pressed

Turn on the cardiograph

____________________________

Extended Self-test : Run Extended self-test, select ALL menu choice and verify that

each test passes with no errors.

Patient module and cable

____________________________

CPU assembly

____________________________

Printer

____________________________

Preview display

____________________________

Keyboard display

________________________

Electrocardiograph Simulation : Record an ECG wave using an ECG simulator.

Verify the following :

Trace activity for all 12 leads

No gross distortion of complexes or calibration pulses

Calibration pulses are of proper duration (200 ms) and amplitude (1 mV)

The trace will vary depending on simulation setting used and simulation type.

calibration pulse measurements will vary depending on the cardiograph gain

and speed setting.

Comments : _______________________________________________________________________

___________________________________________________________

___________________________________________________________

___________________________________________________________

Pulse Recorder :

Test Equipment Used

: ECG Stimulator BIOTEK, Fluke multi meterRemarks

: ________________________________________________________

Performed by : _____________________

Verified by : _______________________

BM.0308.46/1