Upload
sri-iriani-batorowati
View
92
Download
11
Tags:
Embed Size (px)
Citation preview
Lecture 15
Fluorescence spectroscopy and imaging:
Basic principles and sources of contrast
Outline for Fluorescence
I. Princip-Prisip dasar FluoresensiII. Quantum Yield dan LifetimeIII. Spektroskopi Fluoresensi IV. Fluorophores BiologikV. Instrumentasi FluoresensiVI. Pengukuran Fluoresensi
I. Princip-Prinsip Dasar Fluoresensi
1. Luminesensi• Emisi fotons dari keadaan tereksitasi electronik
• Dua tipes luminesensi:
Relakasasi dari keadaan eksitasi singlet excited Relakaksasi dari keadaan eksitasi triplet
I. Princip Dasar Fluoresensi 2. Keadaan singlet dan triplet states
• Keadaan dasar – dua electron per orbital; electron punya spin berlawanan dan berpasangan
• Keadaan eksitasi singlet Electron pada orbital energi lebih tinggi memiliki arah spin berlawanan relative thd electron dalam orbital lebih rendah
• Keadaan eksitasi triplet Elektron valence tereksitasi secara spontan berbailk arah spinnya (spin flip). Proes ini disebut intersystem crossing. Electrons dlm kedua orbital sekarang memiliki arah spin sama
I. Princip Dasar Fluoresensi
3. Jenis emisi• Fluoresensi – kembali dari keadaan eksitasi singlet ke
keadaan dasar; tidak memerlukan perubahan arah spin (relaksasi yang lebih lazim)
• Fosforesensi – Kembali dari keadaan eksitasi triplet ke keadaan dasar; electron perlu perubahan arah spin
• Laji emisi fluoresensi beberapa tingkat lebih cepat daripada fosforesensi (karena perubahan arah spin perlu waktu)
I. Princip Dasar Fluoresensi4. Diagram tingkat energy (Jablonski diagram)
I. Princip Dasar Fluoresensi
5a. Proses Fluoresensi : Population level energi• Pd suhu kamar (300 K), dan level energi electronic dan
vibrational tertentu, dpt dihitung ratio molekule dalam tingkat lebih tinggi dan lebih rendah
kTE
n
n
lower
upper exp
k=1.38*10-23 JK-1 (Boltzmann’s constant)E = perbedaan level energy
I. Prinsip Dasar Fluoresensi5b. Proses Fluoresensi: Eksitasi• Pd suhu kamar, segala sesuatu berawal dari level energi
vibrational terendah pd keadaan dasar.
• Jika sebuah molecule terpapar cahaya pada frekuensi resonan, maka
• Cahaya diserap; utk larutan encer , Hkm Beer-Lambert berlaku koefisien absorpsi molar (extinction coeff) (M-1 cm-1); besarnya menggambarkan probabilitas serapan pada panjang gelombang sesuai dengan pita absorpsinya.
• Eksitasi: setelah penyerapan cahaya, suatu kromofor tereksitasi ke level energi vibrational lbh tinggi, pada level S1 atau S2
• Proses absorpti berlangsung pd skala waktu (10-15 s) jauh lbh cepat d/p vibrasi molekuler → transisi “vertical” (Franck-Condon principle).
So
S1
clA
En
ergy
nuclear configuration
I. Prinsip Dasar Fluoresensi5c. Proses Fluoresensi: relaksasi Non-radiative
• Dlm keadaan tereksitasi, elektron dipromosikan ke orbital anti-bonding → atom dlm ikatan kurang kuat terikat→ bergeser ke kanan kurva energi potential S1 →eletron terpromosikan ke level energi vibrational eksitasi S1 lebih tinggi d/p level vibrational dalam keadaan dasar.
• Deaktivasi vibrational berlangsung lewat tabrakan intermolekuler pd skala waktu10-12 s (lbh cepat d/p proses fluoresensi)
.
So
S1
I. Prinsip Dasar Fluoresensi5d. Fluorescence process: Emission
• Molekule relaksasi dari level vibrasional terendah keadaan tereksitasi menuju level energi vibrasional keadaan tereksitasi (10-9 s)
• Relaksasi ke keadaan dasar berlangsung lebih cepat d/p waktu vibrasional molekuler → transisi “vertical”
• Energy foton yg diemisikan lebih kecil
dibandingkan foton datang So
S1
I. Prinsip Dasar Fluoresensi
6a. Pergeseran Stokes • Cahaya fluorescensi adalah “red-shifted” (λ lbh panjang d/p
λ cahaya eksitasi) relatif thdp cahaya terabsorbsi. ("Stokes shift”).
• Konversi internal (see slide 13) dapat mempengaruhi pergeseran Stokes (Stoke’s shift)
• Solvent mempengaruhi reaksi keadaan tereksitasi dan juga mempengaruhi besarnya Stoke’s shift
I. Prinsip Dasar Fluoresensi6b. Invariansi panjang gelombang emission dgn
panjang gelombang eksitasi
• λ emission hanya bergantung pd waktu relaksasi ke level vibrasi
terendah pada level S1
• Utk molekul, λ emisi fluoresensi yg sama muncul berapapun λ eksitasinya
So
S1
I. Prinsip Dasar Fluoresensi6c. Aturan bayangan cermin
• Level vibrational dlm keadaan tereksitasi dan keadaan dasar adalah sama.
• Spektra absorpsi merefleksikan level vibrational keadaan eksitasi electronik
• Spektra emisi merefleksikan level vibrational electron keadaan dasar
• Spektra emisi fluorescensi merupakan bayangan cermin spektra absorpsi
S0
S1
v=0
v=1
v=2v=3v=4v=5
v’=0v’=1v’=2v’=3v’=4v’=5
I. Prinsip Dasar Fluoresensi6d. Konversion Internal vs. Emisi fluoresensi
• Jika energi electronic bertambah, level energi tumbuh more closely spaced
• Sangat boleh jadi akan overlap antara energi vibrasional tinggi level Sn-1 and energi vibrational rendah levels of Sn
• Overlap meyebabkan transisi antar states sangat mungkin
• Konversi nternal mrpk transisi yg terjadi antara states dg multiplisitas sama dan berlangsung pd skala waktu 10-12 dt (lebih cepat d/p proses fluoresensi)
• Perbedaan energi antara S1 and S0 cukup besar dibanding antar states yg berdekatan → lifetime S1 lebih lama → emisi radiative dapat efektif sempurna dengan emisi non-radiatif.
Mirror-image rule typically applies when only S0 → S1 excitation takes place
Deviations from the mirror-image rule are observed when S0 → S2 or transitions to even higher excited states also take place
I. Prinsip Dasar Fluoresensi6e. Intersystem crossing• Intersystem crossing adalah transisi non-radiative antar states (keadaan)yg
berbeda multiplisitasnya
• Ini terjadi via inversi spin electron tereksitasi menghasilkan dua elektron tak berpasangan dengan arah spin sama, memberikan keadaan dg spin = 1 dan multiplisitas 3 (triplet state)
• Transisi antar state berbeda multiplisitas pada dasarna adalah terlarang
• Spin-orbit and vibronic coupling mechanisms decrease the “pure” character of the initial and final states, memungkinkan intersystem crossing terjadi
• Transisi T1 → S0 juga forbidden → lifetime T1 sangat lebih besar d/p lifetime S1 (10-3-102 s)
S0
S1
T1absorption
fluorescencephosphorescence
Intersystemcrossing
I. Prinsip Dasar Fluoresensi
I. Prinsip Dasar FluoresensiIn
ten
sity
Wavelength
Absorbance
DONOR
Absorbance
Fluorescence Fluorescence
ACCEPTOR
Molecule 1 Molecule 2
• Fluorescence energy transfer (FRET)
Inte
nsi
ty
Wavelength
Absorbance
DONOR
Absorbance
Fluorescence Fluorescence
ACCEPTOR
Molecule 1 Molecule 2
Non radiative energy transfer – a quantum mechanical process of resonance between transition dipolesEffective between 10-100 Å onlyEmission and excitation spectrum must significantly overlapDonor transfers non-radiatively to the acceptor
II. Quantum yield and lifetime• Quantum yield of fluorescence, f, is defined as:
• In practice, is measured by comparative measurements with reference compound for which has been determined with high degree of accuracy.
• Ideally, reference compound should have– the same absorbance as the compound of interest at given excitation wavelength– similar excitation-emission characteristics to compound of interest (otherwise, instrument
wavelength response should be taken into account)– Same solvent, because intensity of emitted light is dependent on refractive index
(otherwise, apply correction
– Yields similar fluorescence intensity to ensure measurements are taken within the range of linear instrument response
absorbed photons ofnumber
emitted photons ofnumber f
)(
)(2
2
sn
un
I
Isf
uf
sf
uf
1a. Quantum yield of fluorescence
II. Quantum yield and life time• Another definition for f is
where kr is the radiative rate constant and k is the sum of the rate constants for all processes that depopulate the S1 state.
• In the absence of competing pathways f=1• Radiative lifetime, r, is related to kr
• The observed fluorescence lifetime, is the average time the molecule spends in the excited state, and it is
k
krf
rr k
1
kf
1
1b. Fluorescence lifetime
II. Quantum Yield and Lifetime
2a. Characteristics of quantum yield• Quantum yield (hasil Kuantum) fluoresensi bergantung
pd lingkungan biologis
• Contoh: spektrum eksitasi Fura 2 dan spektrum emisi Indo-1 dan hasil quantum berubah ketika terikat pada Ca2+
Fura-2 changes in response to varying [Ca2+]
Indo-1 changes in response to varying [Ca2+]
II. Quantum yield and lifetime
2b. Characteristics of life-time• Provide an additional dimension of information
missing in time-integrated steady-state spectral measurements
• Sensitive thd mikrolingkungan biochemis, termasuk pH lokal, oksigenasi dan ikatan
• Lifetimes tak terpengaruh oleh variasi in intensitas eksitasi, konsentrasi atau sumber hilangnya keoptikan
• Kompatible dg pengukuran klinik in vivo
Courtesy of M.-A. Mycek, U Michigan
II. Quantum Yield and Lifetime
3a. Fluorescence emission distribution• Utk λ eksitasi tertentu,
transisi emisi terdistribusi di antara level energi vibrational yg berbeda
• Utk λ eksitasi tunggal dpt mengukur suatu spectrumemisi fluoresensi
Inte
nsit
y
Emission Wavelength (nm)
ExcEmm
II. Quantum Yield and Lifetime
3b. Heisenberg’s uncertainty principle• Values of particular pairs of observables cannot be
determined simultaneously with high precision in quantum mechanics
• Example of pairs of observables that are restricted in this way are:
• Momentum and position
• Energy and time
II. Quantum Yield and Lifetime
3c. Heisenberg’s uncertainty principle
• Momentum and position :
• Energy and time:
2
hxpx
2
htE
II. Quantum Yield and Lifetime
3d. Effect on fluorescence emission• Suppose an excited molecule emits fluorescence in
relaxing back to the ground state
• If the excited state lifetime, is long, then emission will be monochromatic (single line)
• If the excited state lifetime, is short, then emission will have a wider range of frequencies (multiple lines)
Inte
nsit
yEmission Wavelength (nm)
Exc Emm
Inte
nsit
y
Emission Wavelength (nm)
Exc Emm
Large – small Small – large
III. Fluorescence Intensity
1. Fluorescence intensity expression
2. Fluorescence spectra
III. Fluorescence Intensities
1a. Fluorescence intensityThe fluorescence intensity (F) at a particular excitation (x) and emission wavelength (m) will depend on the absorption and the quantum yield:
where,
IA – light absorbed to promote electronic transition – quantum yield
mxAmx IF ,
III. Fluorescence Intensities
1b. From the Beer-Lambert law, the absorbed intensity for a dilute solution (very small absorbance)
where,
Io – Initial intensity – molar extinction coefficient
C – concentrationL – path length
1 <<CL for
CLI303.2I
x
xoxA
III. Fluorescence Intensities
1c. Fluorescence intensity expressionThe fluorescence intensity (F) at a particular excitation (x) and emission wavelength (m) for a dilute solution containing a fluorophore is:
where,
Io – incident light intensity – quantum yieldC – concentration – molar extinction L – path length coefficient
mxomx CLIF 303.2,
III. Fluorescence Intensities
1d. Measured fluorescence intensityIf we include instrument collection angle:
where,Z – instrumental factor
Io – incident light intensity – molar extinction coefficientC = concentrationL – path length
ZCLIF mxomx 303.2,
III. Fluorescence Intensities2a. Fluorescence spectra
• Emission spectrum– Hold excitation wavelength fixed, scan emission
– Reports on the fluorescence spectral profile
» reflects fluorescence quantum yield, k(m)
ZCLIF mxomx 303.2,
III. Fluorescence Intensities2b. Fluorescence spectra
• Excitation spectrum– Hold emission wavelength fixed, scan excitation
– Reports on absorption structure
» reflects molar extinction coefficient, (x)
ZCLIF mxomx 303.2,
Fluo
resc
ence
Int
ensi
ty
Emission Wavelength (nm)
Fixed Excitation Wavelength
(b)
Fluo
resc
ence
Int
ensi
ty
Excitation Wavelength (nm)
Fixed Emission Wavelength
(a)
III. Fluorescence Intensities
III. Fluorescence Intensities2c. Fluorescence spectra
• Composite: Excitation-Emission Matrix» Good representation of multi-fluorophore solution
Flu
ores
cen
ce I
nte
nsit
y
Emission Wavelength (nm)
Fixed Excitation Wavelength
Exc
itat
ion
Wav
elen
gth
(n
m)
Emission Wavelength (nm)
Emission spectrum Excitation-emission matrix
III. Fluorescence Intensities
IV. Biological Fluorophores
1. Table
2. EEMs of Epithelial cell suspension
3. EEMs of Collagen
IV. Biological Fluorophores–Endogenous Fluorophores
amino acids
structural proteins
enzymes and co-enzymes
vitamins
lipids
porphyrins
–Exogenous Fluorophores
Cyanine dyes
Photosensitizers
Molecular markers – GFP, etc.
IV. Biological Fluorophores
TCL-1Emission(nm)
Exc
itatio
n(nm
)
300 350 400 450 500 550 600 650 700250
270
290
310
330
350
370
390
410
430
450
470
490
510
530
550
9.767e-001
6.646e+004
1.306e+005
1.947e+005
4.785e+005
1.120e+006
1.761e+006
2.923e+006
9.335e+006
1.575e+007
2.216e+007
Epithelial Cell SuspensionFluorescence intensity excitation-emission
matrix
FAD
NADH
Tryp.
Courtesy of N. Ramanujam
CarbohydratesFatty Acids and Glycerol Amino Acids
Acetyl CoA
CITRIC ACID CYCLE
CoA
CO2
FADH2
NADH
NADH-QReductase
Cytochrome Reductase
Cytochrome Oxidase
Q
Cytochrome C
O2
FAD NAD+
Oxidation of NADH and FADH2 by O2 drives synthesis of ATP
ELECTRON TRANSPORT
Metabolic Indicators
Metabolism
Redox Ratio: FAD / (FAD+NADH)
Redox ratio ~ Metabolic Rate
Emission(nm)
Exc
itatio
n(nm
)
Highest value: 13223388.1 270/260Scaled @: 4916595.85 305/270
M
S
UT Austin Mar-2000 UU
300 350 400 450 500 550 600 650 700250
270
290
310
330
350
370
390
410
430
450
470
490
510
530
550
7.551e+003
2.210e+004
3.613e+004
5.016e+004
1.123e+005
2.526e+005
3.929e+005
6.473e+005
2.051e+006
3.454e+006
4.857e+006
0 days
Emission [nm]
Exc
itatio
n [n
m]
Emission(nm)
Exc
itatio
n(nm
)
Highest value: 9476234.1 270/260Scaled @: 3424911.85 305/270
M
S
UT Austin Mar-2000 UU
300 350 400 450 500 550 600 650 700250
270
290
310
330
350
370
390
410
430
450
470
490
510
530
550
5.021e+003
1.514e+004
2.490e+004
3.467e+004
7.788e+004
1.755e+005
2.731e+005
4.501e+005
1.426e+006
2.403e+006
3.379e+006
7 days
Collagen I (gel)
K. Sokolov
Emission(nm)
Exc
itatio
n(nm
)
Highest value: 11516168.9 270/260Scaled @: 4868431.748 305/270
M
S
UT Austin Mar-2000 UU
300 350 400 450 500 550 600 650 700250
270
290
310
330
350
370
390
410
430
450
470
490
510
530
550
1.097e+004
2.559e+004
3.969e+004
5.379e+004
1.162e+005
2.572e+005
3.982e+005
6.538e+005
2.064e+006
3.474e+006
4.883e+006
39 days
IV. Biological Fluorophores
Collagen• It is the major extracellular matrix component, which is
present to some extent in nearly all organs and serves to hold cells together in discrete units
• Collagen fluorescence in load-bearing tissues is associated with cross-links, hydroxylysyl pyridoline (HP) and lysyl pyridinoline (LP).
• Collagen crosslinks are altered with age and with invasion of cancer into the extracellular matrix
V. Fluorescence Instrumentation
1. Introduction
2. Components of a spectrofluorometer
3. Description of key components
V. Fluorescence Instrumentation
1. Introduction• Fluorescence is a highly sensitive method (can measure
analyte concentration of 10-8 M)
• Important to minimize interference from:Background fluorescence from solventsLight leaks in the instrumentStray light scattered by turbid solutions
• Instruments do not yield ideal spectra:Non-uniform spectral output of light sourceWavelength dependent efficiency of detector and
optical elemens
V. Fluorescence Instrumentation• Illumination source
– Broadband (Xe lamp)– Monochromatic (LED, laser)
• Light delivery to sample– Lenses/mirrors– Optical fibers
• Wavelength separation (potentially for both excitation and emission)– Monochromator– Spectrograph
• Detector– PMT– CCD camera
2. Major components for fluorescence instrument
V. Fluorescence Instrumentation
Components of the spectrofluorometer (standard fluorescence lab instrument for in vitro samples)• Xenon lamp (> 250 nm)
• Excitation and emission monochromatorEach contains two gratings to increase purity of the lightAutomatic scanning of wavelength through motorized gratings
• Sample compartment
• Photo multiplier tube
PMT
Xenon Source
ExcitationMonochromator Emission
Monochromator
Sample compartment
V. Fluorescence Instrumentation2. Spectrofluorometer schematic
V. Fluorescence Instrumentation
3a. Xenon light source• Continuous output from Xenon: 270-1100 nm
• Power – typically 200-450 W
• Lifetime of ~2000 hours
• Strong dependence on wavelength
V. Fluorescence Instrumentation
3a. Xenon light source: broad illumination in the near UV-visible range
PMT
Xenon Source
ExcitationMonochromator Emission
Monochromator
Sample compartment
V. Fluorescence Instrumentation
V. Fluorescence Instrumentation3b. Monochromator: only a small range of wavelengths are focused
at the exit slit determined by angle of light incident on the diffraction grating
constant
wavelength
mmper lines of #n
ordern diffractioK where
,10sinsin 6
D
Kn
Principle of diffraction grating operation
V. Fluorescence Instrumentation
3b. Monochromator – Spectral ResolutionInversely proportional to product of dispersion
(nm/mm) of grating and the slit width (mm)
~ 5 nm sufficient for fluorescence measurements of
biological media
Signal increases with the slit width
V. Fluorescence Instrumentation
3b. Monochromator – Stray light• Light which passes through monochromator besides
that of desired wavelength
• Double grating monochromator (stray light rejection is 10 -8 – 10-12) but signal is decreased
V. Fluorescence Instrumentation
3b. Monochromator – Signal efficiency• Grating has a wavelength dependent efficiency
• Can choose the wavelength at which grating is blazed (maximal efficiency)
• Excitation monochromator should have high efficiency in the UV; emission monochromator should have high efficiency in the visible
PMT
Xenon Source
ExcitationMonochromator Emission
Monochromator
Sample compartment
V. Fluorescence Instrumentation
V. Fluorescence Instrumentation
3c. Photomultiplier tube• Contains a photocathode:
light sensitive material, which yields electrons upon interaction with photons based on photoelectric effect.
• Electrons are multiplied by a series of dynodes
• Provides current output proportional to light intensity
V. Fluorescence Instrumentation
3c. PMT – Linearity response• Current from PMT is proportional to light intensity
• Under high intensity illumination, PMT will saturate (dynamic range); at low intensity, limited by dark noise
• Excessive light can damage photocathode, resulting in loss of gain and increased dark noise (thermal noise)
V. Fluorescence Instrumentation
3c. PMT – Quantum efficiency• Quantum efficiency gives the photon to electron
conversion efficiency
• Highly dependent on wavelength
V. Fluorescence Instrumentation
3c. Key components – Noise• Dark current – Noise due to thermal generation;
increases with temperature and high voltage
• Shot noise – proportional to the square root of the signal
VI. Fluorescence Measurements
VI. Fluorescence measurements1. Instrument non-uniformities2. Excitation wavelength calibration3. Emission wavelength calibration4. Setup parameters for emission spectrum5. Routine experimental procedure6. Collection geometry 7. Blank scans8. Typical fluorescence spectrum
VI. Fluorescence Measurements
1a. Ideal spectrofluorometer• Light source must yield constant photon output at all
wavelengths
• Monochromator must pass photons of all wavelengths with equal efficiency
• The PMT must detect photons of all wavelengths with equal efficiency
VI. Fluorescence MeasurementsL
ight
Int
ensi
ty
Wavelength
LIGHT SOURCE
Eff
icie
ncy
Wavelength
MONOCHROMATOR
Eff
icie
ncy
Wavelength
PMT
VI. Fluorescence Measurements
1b. Distortions in excitation and emission spectra• Light intensity from light source is a function of
wavelength
• Monochromator efficiency is a function of wavelength
• The PMT does not have equal efficiency at all wavelengths
VI. Fluorescence Measurements
1c. Calibration• Correction of variations in wavelength of Xenon lamp
and excitation monochromatorNeed to do when measuring excitation spectra or emission spectra at multiple excitation wavelengths
• Correction of emission monochromator and PMTNeed to do when measuring emission spectra
VI. Fluorescence Measurements2a. Excitation wavelength calibration
• Excitation spectra are distorted primarily by the wavelength dependent intensity of the light source
• Can use reference photodetector (calibrated) next to sample compartment to measure fraction of excitation light
• The measured intensity of the reference channel is proportional to the intensity of the exciting light
Io
F
SampleCompartment
To PMT
To Reference Photodiode
VI. Fluorescence Measurements2b. Effect of excitation wavelength calibration
VI. Fluorescence Measurements3a. Emission wavelength calibration
• Need correction factors• Measure wavelength dependent output from a
calibrated light source• Standard lamps of known and calibrated
spectral outputs are available from the National Institute of Standards and Testing (NIST)
• This measurement is typically done by factory; it is difficult to perform properly with commercial fluorimeter
To PMT
Sample Compartment
Calibrated Lamp
VI. Fluorescence Measurements
3b. Emission wavelength calibration procedure• Measure intensity versus wavelength (I()) of standard
lamp with spectrofluorometer
• Obtain the spectral output data (L()) provided for the lamp
• Correction factor: S() = L()/ I()
• Multiply emission spectrum with correction factor
VI. Fluorescence Measurements3c. Emission wavelength calibration curve
0.0E+00
2.0E+00
4.0E+00
6.0E+00
8.0E+00
1.0E+01
1.2E+01
300 350 400 450 500 550 600 650 700
Wavelength (nm)
Inte
nsi
ty (
c.u
.)
VI. Fluorescence Measurements
5. Routine experimental procedures• Check wavelength calibration of excitation
monochromator
• Check wavelength calibration of emission monochromator
• Check throughput of spectrofluorometer
0.0E+00
1.0E-02
2.0E-02
3.0E-02
4.0E-02
5.0E-02
6.0E-02
260 310 360 410 460 510 560
Wavelength (nm)
Inte
nsity
(cp
s)
467 nm
Xe lamp scan
Hg lamp spectrum scan
0.0E+00
2.0E+06
4.0E+06
6.0E+06
8.0E+06
1.0E+07
1.2E+07
475 525 575 625 675
Wavelength (nm)
Inte
nsity
(c.
u.)
575 nm
Rhodamine standardscan
VI. Fluorescence Measurements6a. Collection geometry in sample compartment
• Front face – collection is at a 22 degree angle relative to the incident beam; appropriate for an optically absorbing / scattering sample; more stray light
• Right angle – collection is at a right angle to the incident light; appropriate for optically transparent sample; less stray light
Io F Io
F
Front Face Right Angle
VI. Fluorescence Measurements6c. Features of right angle illumination
• Appropriate for optically transparent sample
• At high optical densities, signal reaching detector will be significantly diminished
VI. Fluorescence Measurements
6d. Features of front face illumination• Appropriate for an optically absorbing / scattering
sample
• At high optical densities, light is absorbed near the surface of the cuvette containing the absorber; therefore fluorescence is detectable
• Fluorescence independent of concentration at high optical densities
VI. Fluorescence Measurements
7. Blank scan• Blank is identical to sample except it does not contain
fluorophore
• Measuring the fluorescence of these samples allows the scattering (Rayleigh and Raman) to be assessed
• In addition, such samples can reveal the presence of fluorescence impurities, which can be subtracted
VI. Fluorescence Measurements8. Typical fluorescence emission spectrum at 340 nm
excitation (the different components)
0
500000
1000000
1500000
2000000
2500000
3000000
300 350 400 450 500 550 600
Wavelength (nm)
Flu
ore
sc
en
ce
In
ten
sit
y (
a.u
.)
Raman
Rayleigh (exc = emm)
Fluorescence