Finite Element Analysis of Composite Hip Prosthesis M. Sivasankar, D.Chakraborty, S.K.Dwivedy M. Sivasankar, D.Chakraborty, S.K.Dwivedy Department of Mechanical Engineering Department of Mechanical Engineering Indian Institute of Technology, Guwahati Indian Institute of Technology, Guwahati

Finite Element Analysis of Composite Hip Prosthesis

  • Upload
    acacia

  • View
    38

  • Download
    0

Embed Size (px)

DESCRIPTION

Finite Element Analysis of Composite Hip Prosthesis. M. Sivasankar, D.Chakraborty, S.K.Dwivedy Department of Mechanical Engineering Indian Institute of Technology, Guwahati. An overview of hip replacement References on total hip replacement (THR) Objective Material selection - PowerPoint PPT Presentation

Citation preview

Page 1: Finite Element Analysis of Composite Hip Prosthesis

Finite Element Analysis of Composite Hip Prosthesis

M. Sivasankar, D.Chakraborty, S.K.Dwivedy M. Sivasankar, D.Chakraborty, S.K.Dwivedy

Department of Mechanical EngineeringDepartment of Mechanical EngineeringIndian Institute of Technology, GuwahatiIndian Institute of Technology, Guwahati

Page 2: Finite Element Analysis of Composite Hip Prosthesis

2

Scope of the Present Work

An overview of hip replacementReferences on total hip replacement (THR)ObjectiveMaterial selection Finite element modelComparison of results

Page 3: Finite Element Analysis of Composite Hip Prosthesis

3

An Overview of Hip Replacement

Anatomy of hip jointHip prosthesisTypes of prosthesis fixationReasons for hip failure

A Typical Hip Prosthesis

A Typical Hip Prosthesis

Page 4: Finite Element Analysis of Composite Hip Prosthesis

4

Anatomy of Hip Joint

Largest weight bearing joint Composed of rounded head of the femur joining the acetabulum of pelvis in a ball and socket arrangement

Page 5: Finite Element Analysis of Composite Hip Prosthesis

5

Reasons for Hip Failure

Long-term aseptic loosening.Primary hip arthoplasties are subjected to failure due to bone resorption i.e. bone loss.Failure due to fatigue loading of hip joint. Relative micro motions resulting from improper implant fitting in the bone cavity.

Page 6: Finite Element Analysis of Composite Hip Prosthesis

6

In cementless implants load transfer between a stiff implant and relatively flexible bone results in extremely unnatural stress distribution in bone, i.e. excessive stress concentrations near to the implant ends.Stress shielding followed by bone resorption in the other areas of bone-implant interface.

Reasons for Hip Failure Cont..

Page 7: Finite Element Analysis of Composite Hip Prosthesis

7

Reasons for Hip Failure Cont..

Hip failure due to bone loss is caused by the production of wear particles associated with the deterioration of the prosthesisFor an average hip patient, the prosthesis have to resist thirty-four million blows

Page 8: Finite Element Analysis of Composite Hip Prosthesis

8

Damaged Femoral Head

Femoral head cartilage The neck is cut-off as in figure Marrow cavity is made inside the femur Hip prosthesis is fitted either by PMMA cement or press fitted

Page 9: Finite Element Analysis of Composite Hip Prosthesis

9

References on Total Hip Replacement

Many researchers carried out advanced researches in the field of THR using Finite Element Method and other methods

Page 10: Finite Element Analysis of Composite Hip Prosthesis

10

Literature Review

Researches in this area has been carried out in:

Cemented Joint

Cement less Joint

Finite Element Analysis

Experimental with design models

Page 11: Finite Element Analysis of Composite Hip Prosthesis

11

Researches Carried Out

S.No. Experiment Based FEM based Cemented Joint Cementless Joint

1 G.Bergmann(2001) C.F.Scifert (1999) S.K.Senapathi (2002) P.Kowalczyk (2001)2 G.Selvaduray(2002) A.Philips (2001) P.Colombi (2002)  3 H.Katoozian (2001) P.Kovalczyk(2001) A.B.Lennon (2003)  4 J.A.Simoes (2000) P.B.Chang (2001) Sheryl Zimmarman (2002)  5 M.Baleani(2000) R.Huiskes (1992) V.Waide (2004)  6 C.Kaddick (1997) C.Li (2002) P.J.Prendergast (1997)  7 N.H.Tai (1995) W.Van Papegem (2001) A.Philips (2001)  8 G.Dillon (1995) S.Srinivasan (2000) C.Li (2002)  9 X.Diao (1997) H.F.El.Sheikh (2003)    10 B.W.Stansfield (2003) S.Gross (2001)    11 M.Pawlikowwski S.H.Teoh (2002)    12 S.L.Evans (1998) Bernard Weisse (2003)    13 R.P.Morris H.Katoozian (2001)    14 M.T.Raimondi (1999)      15 C.M.Styles (1998)      16 I.Hilal (1999)      17 Darryl.D      18 S.Srinivasan (2000)      19 L.J.Lee (1996)      20 W.Vanpaepegem (2002)      21 S.Ramakrishna (2001)      22 K.L.Reifs nider (1991)      23 S.K.Roy Choudhury (2004)      24 A.Rajadurai (2002)      25 R.De.Santis (2004)      26 Vesa Saikko (2002)      27 Debera.E Hurwitz (2003)      28 B.Mavcic (2002)      

Page 12: Finite Element Analysis of Composite Hip Prosthesis

12

Recent Work

Few researchers like A.Phillips[1], P.J.Prendergast [2] ,H.Katoozian [6], C.Li [9] etc., work in the area of cemented prosthesis.

Page 13: Finite Element Analysis of Composite Hip Prosthesis

13

Biomaterials

Stainless Steel AlloysCobalt-Chrome alloys (Vitallium)Titanium alloysComposites

Page 14: Finite Element Analysis of Composite Hip Prosthesis

14

Comparison of Characteristics

Characteristic S-Steel Co-Cr alloy Titanium Alloy

Stiffness High Medium Low

Strength Medium Medium High

Corrosion -resistance Low Medium High

Biocompatibility Low Medium High

Page 15: Finite Element Analysis of Composite Hip Prosthesis

15

Need of Composites

The isotropic alloys used for stem have much higher stiffness than that of the boneAlmost all monoclinic implants have 5 to 20 times more stiffness than the boneA stiff shaft of a total hip prosthesis stress shields the upper part of the thigh bone

The shielded bone does not thrive, loses its substance and becomes weakThe total hip joint has weak anchorage in a weak skeleton and may failThe remedy is a prosthetic shaft manufactured from metal alloys with stiffness similar to bone

Page 16: Finite Element Analysis of Composite Hip Prosthesis

16

Advantages of Composites

Low stiffness of composite stems can enhance proximal bone ingrowthsTailorability property in strength and stiffness.Excellent biocompatibility

A controlled stiffness prosthesis can reduce stress shielding and bone resorptionLess weight of the prosthesis

Page 17: Finite Element Analysis of Composite Hip Prosthesis

17

Composite Prosthesis

Clinical studies reported early fatigue fracture of a femoral component made from laminated fiber reinforced composites.The new designs are Constructed of short glass fibers/epoxy resin and CF/PEEK composites.

Page 18: Finite Element Analysis of Composite Hip Prosthesis

18

SOLID 92 Element

Page 19: Finite Element Analysis of Composite Hip Prosthesis

19

Composite Model

Basic Composite Model With Elements

Page 20: Finite Element Analysis of Composite Hip Prosthesis

20

Conical Stem

Cemented prosthesis model contains three main parts:

Conical Stem with headCement layerCortical bone

Basic Model

Page 21: Finite Element Analysis of Composite Hip Prosthesis

21

Chopped Fiber Core

Model With Chopped Fiber Core

Page 22: Finite Element Analysis of Composite Hip Prosthesis

22

Material Properties

Material Properties Used for Analysis of Total Hip Prosthesis

Parts Material Young’s Modulus (MPa)

Poisson's Ratio

Geometrical Parameter (All dimensions are in mm) 

Head and Stem

 Ti6Al4V 110x103 0.33 Sphere radius 25Stem radius 10Stem outer radius 10Stem inner radius 7.5 

Cement Layer

UHMWPE-AL2O3

1x103 0.39 Inner radius 10.5Outer radius 12.2182Length 100 

Cortical Bone

AS4/PEEK 3x103 0.30 Inner radius 20.5Outer radius 30

Page 23: Finite Element Analysis of Composite Hip Prosthesis

23

Maximum Shear Stress Region

Enlarged View of the Deformed Stem and Cortical Bone Showing the Maximum Shear Stress Region (Path Aa)

Page 24: Finite Element Analysis of Composite Hip Prosthesis

24

Variation of Shear Stresses

Variation of Maximum Shear Stress With System Parameters

Stem Length (in mm)

Maximum Shear Stress(in MPa)

 145 17.314

145.5 15.522

147.5 21.033

150 20.919

152.5 17.144

155 20.262

Neck Inclination (in degree)

Maximum Shear Stress(in MPa)

 45 17.314

47.5 20.383

50 22.964

Neck Length (in mm)

Maximum Shear Stress(in MPa)

 45 13.337

47.5 17.376

50 17.314

52.5 25.363

Stem Inner Radius (in mm)

Maximum Shear Stress(in MPa)

 

7.5 17.314

8 20.655

8.5 19.443

Page 25: Finite Element Analysis of Composite Hip Prosthesis

25

The variation in the above parameters do not show a particular trend

Hence the design optimization has been carried out to minimize the magnitude of maximum shear stress

Continued...

Page 26: Finite Element Analysis of Composite Hip Prosthesis

26

Hip Prosthesis 

Parts State Variables Design Variables

Femur

Sphere Radius 25 mm

Stem Outer Radius

10 mm

Stem Inner Radius

7.5 mm

Neck Inclination 450

Stem Length 145.5 mm

Neck length 50 mm

Dimensions of Hip Prosthesis Before Optimization

Page 27: Finite Element Analysis of Composite Hip Prosthesis

27

Femoral Components

Design Variables of Femoral ComponentsAfter Optimisation

Design Variables Dimension (mm)

Stem outer radius 9.9301

Stem inner radius 8.0405

Stem length 153.22

Neck length 50.975

Page 28: Finite Element Analysis of Composite Hip Prosthesis

28

Shear Stresses

SXY x-y component

SYZ y-z component

SXZ z-x component

Shear Stresses in the Interface of Stem and Cortical Bone

Page 29: Finite Element Analysis of Composite Hip Prosthesis

29

Shear Stresses - Continued…

SXY x-y component

SYZ y-z component

SXZ z-x component

Shear Stresses in the Interface of Stem and Cortical Bone

Page 30: Finite Element Analysis of Composite Hip Prosthesis

30

Conclusion

A 3D finite element analysis has been done for analysis of composite hip prosthesis which consists of a conical stem with a cement layer.Location and magnitude of shear stresses show the region of failure which is in agreement with the earlier published results.

Page 31: Finite Element Analysis of Composite Hip Prosthesis

31

Continued…

As the variation of the parameters do not show a particular trend, design optimization has been carried out to minimize the magnitude of maximum shear stressThe optimum dimensions obtained from the present analysis show considerable reduction in shear stress

Page 32: Finite Element Analysis of Composite Hip Prosthesis

32

References

[1] A. Phillips, 2001, Finite element analysis of the acetabulum after impaction grafting, The University of Edinburgh.[2] P.J.Prendergast, 1997, Review paper – Finite element models in tissue mechanics and orthopaedic implant design, Clinical Biomechanics, Vol. 12, No. 6, 343-366.[3] C. F. Scifert, T. D.Brown, J. D.Lipman, 1999, Finite element analysis of a novel design approach to resisting total hip dislocation, Clinical Biomechanics, 14, 697-703.[4] M.Baleani, M.Viceconti, R. Muccini, M. Ansaloni, 2000, Endurance verification of custom-made hip prostheses, International journal of fatigue 22, 865-871. [5] P. B. Chang, B. J. Williams, K.S. B.Bhalla, T. W. Belknap, T. J. Santner, W. I. Notz, D. L. Bartel, 2001, Design and analysis of robust total joints replacements: Finite element model experiments with environmental variables, ASME, Journal of Biomechanical .Engineering, 123, 239-246.[6] H. Katoozian, D. T. Davy, A. Arshi, U. Saadati, 2001, Material Optimization of femoral component of total hip prosthesis using fiber reinforced polymeric composites, Medical Engineering and Physics, 23, 503-509.[7] S. K. Senapati, S. Pal, 2002, UHMWPE-ALUMINA ceramic composite, an improved prosthesis material for an artificial cemented hip joint, Trends in Biomaterials Artificial. Organs, 16(1), 5-7.[8] J.Stolk, N. Verdonschot, L. Cristofolini, A. Toni, R. Huiskes, 2002, Finite element and experimental models of cemented hip joint reconstructions can produce similar bone and cement strains in pre-clinical tests, ASME, Journal of Biomechanics 35, 499-510.[9] C. Li, C. Granger, H. D. Schutte Jr, S. B. Biggers Jr, J. M. Kennedy, R. A. Latour Jr, 2003, Failure analysis of composite femoral components for hip arthroplasty, Journal of Rehabilitation Research and Development, 40(2), 131–146.

Page 33: Finite Element Analysis of Composite Hip Prosthesis

33

Questions?

Page 34: Finite Element Analysis of Composite Hip Prosthesis

34

Thank You for Attending…

A copy of my slides will be available on my website www.biosankar.4t.com

My email address is: [email protected]

IITG Biomechanics center: www.iitg.ernet.in

Work supported by Department of Mechanical Engineering,

contact +91 (361) 2582697