Filtration + Separation Sept Okt 2012

Embed Size (px)

Citation preview

  • 7/30/2019 Filtration + Separation Sept Okt 2012

    1/1

    Feature44Filtration+SeparationSeptember/October 2012

    Sealed (Dead End)capillary membranes

    Epoxy

    Epoxy

    Untreated Water

    Sterile

    Water Exits

    capillary membranes

    Bacteria control:

    Testing membranefilter cartridges

    The standard test method for determiningbacterial retention in membrane filtersrequires a one-time challenge of 107 bacteriaper cm2 of the effective membrane surface[ASTM F838-05 Standard Test Method forDetermining Bacterial Retention of MembraneFilters Utilized for Liquid Filtration]. In factthis method replicates what normally occursgradually in a sterilising filtration process. Themethod used originated from the point of viewthat the bacterial load of a membrane filteroccurs over time and over the service life of the

    filter. The goal was to determine if the filterretains a total bacterial load of 107 per cm2 (1010bacteria/filter cartridge) over its anticipatedservice life of approximately 5000 L.

    ASTM F838-05 prescribes Brevundimonasdiminuta as the test organism. However, asthere is no specific selective culture medium forthis organism, it is not possible to differentiateBrevundimonas diminuta from other bacteria.Therefore the tests were carried out with asubstitute, the Pseudomonas aeruginosa bacteriumfor which specific culture media exists. Anotherreason that Pseudomonas aeruginosa was used

    is that it has a higher passage percentagethan Legionella pneumophila [Influence of size,shape and flexibility on bacterial passagethrough micro pore membrane filters. T.Egli Environmental Science & Technology, 008,4, 6749-6754.], which is the target bacteriastrain to be retained. The choice to perform the

    esearchers at the Flemish Institute for Technological Research VITO

    NV have undertaken tests on membrane filter cartridges to determine

    their suitability for use in creating Legionella Safe shower units.

    validation with Pseudomonas aeruginosa is wellfounded: the smallest width ofPseudomonasaeruginosa bacteria is 0.50 m, whereas thesmallest width ofLegionella pneumophila is 0.55m (versus the smallest width ofBrevundimonasdiminuta of 0.40 m).

    For the dynamic test method two cubitainerswere filled with tap water filtered overcarbon and micro-filtration cartridges inorder to ensure sterility of the test water. Thecontainers in which bacterial suspensions were

    stored were placed in parallel. The bacterialsuspensions were consecutively pumpedthrough Legionella Safe shower filters (-3bar). Filtration was achieved by means of asuper-efficient hollow fibre micro-filtrationmembrane, with a maximum pore size of0.15 m, which has a high affinity for waterby virtue of a unique blend of polymericcomponents. It only needs low pressure tostart the filtration process and delivers highflow rates at a low pressure gradient.

    The membrane filters (manufactured by PrimeWater bvba) consist of a bundle of asymmetric

    hollow fibre membranes in a tube, where thespace between the membranes and the tube ishermetically sealed with an epoxy resin (seeFigure 1). In the so called Dead end filtrationtechnique (see Figure ), the water flowsfrom the outside to the inside and the filteredparticles are caught in the porous membrane

    wall, which gradually clogs. A decrease in flowrate to an unacceptable level is a signal forreplacement of the filter. This micro filtrationtechnique is primarily applied in LegionellaSafe shower filters.

    From both containers the inlet bacterialsuspension was continuously sampled todetermine the exact inlet concentrationofPseudomonas aeruginosa. With intervalsof approximately 1000 L permeate sampleswere taken from each of the filters. The

    flow rate at the end of the test was reducedto approximately 5 LpM, from 10 LpM atthe start. After a simulated service life of5000 and 4500 L respectively with watergradually contaminated with up to E+10 CFUPseudomonas aerigunosa bacteria, the bacterialretention of both filters was calculated by theanalysis of permeate samples at the beginningand the end of two days of filtration. Itwas concluded that both filter cartridgeseffectively stop bacteria in the inflowing waterand demonstrate a log reduction of at least6 (99.9999%). The ASTM F838-05 criteriawere satisfied in the determination of the

    bacterial retention of membrane filters. Contact:

    Sabine Kreps

    VITO NV Gebouw Prodem

    Tel. + 32 14 33 69 42

    E-mail:[email protected]

    www.vito.be

    R

    Figure 1: Cross section of hollow fibre micro-filtration membrane filter. Figure 2: Dead end filtration principle.

    mailto:[email protected]:[email protected]://www.vito.be/mailto:[email protected]://www.vito.be/