Upload
others
View
5
Download
0
Embed Size (px)
Citation preview
FFDM -FCRm QC Requirements-What You REALLY Need to Know
Melissa C. Martin, M.S., FACR, FAAPM, FACMP
AAPM Annual Meeting July 28, 2008
FDA Approval for Mammography
• Fuji FCRm System approved for Mammography on July 10, 2006 by the FDA as a FFDM Digital Mammography System
CR TechnologyCR TechnologyProtective LayerProtective LayerPhoto-stimuable phosphor layerPhoto-stimuable phosphor layer
SupportSupport
J. Ed Barnes, Ph.D. - MTMI
Fuji Fuji MammoMammo CRCR
•Pixel size 50 micron
•Use dual sided reading to DQE
•Imaging Plate has thicker phosphor layer
•Results in 40-50% increase in NEQ (DQE)
•Post processing enhances image
(e.g. filtering to recognize and enhance calcifications)
J. Ed Barnes, Ph.D. - MTMI
FCMFCM
J. Ed Barnes, Ph.D. - MTMI
1mm
1mm
Enlarged Images
Standard System (100 micron pixel)
FCRm (50 micron pixel)Dual side scanJ. Ed Barnes, Ph.D. - MTMI
Fuji System Components
ClearView-CSm
Breast Imaging Review Station
ClearView-1m Flash Plus IIPm
J. Ed Barnes, Ph.D. - MTMI
Fuji Clearview CSm Reader
• Four cassette stacker• Processes 20 screening exams/hour
(or 100 CR plates/hour)• Multi-objective Frequency Processing (MFP)
(enhances dense/peripheral tissue)• Pattern Enhancement Processing for
Mammography (PEM) (enhances calcifications)
J. Ed Barnes, Ph.D., - MTMI
Other CR Mammography Venders - World Wide
•Agfa
•Kodak
•Konica
•Canon
J. Ed Barnes, Ph.D. - MTMI
Kodak Mammography CR•Single sided plate•Thinner phosphor layer (improves resolution, reduces DQE)•Phosphor grain smaller and homogenous•BaFBr:Eu phospor•48.5 micron pixel•Read out speed reduced
•Twice as slow•Increases SNR•Use conventional reader (modified)
J. Ed Barnes, Ph.D. - MTMI
Agfa CR Mammography
•Single sided plate, flexible
•BaSrFBrI:Eu phosphor
•Smaller grain powdered phosphor
•50 micron pixel
•Scan slower than conventional CR
•Wider light collection guide
•Thinner plate to increase resolution
J. Ed Barnes, Ph.D. - MTMI
Konica CR Mammography
•Single sided plate, rigid
•BaFI:Eu phosphor
•43.75 micron pixel
•Utilize “Phased Contrast CR”
•Expose breast on a mag platform to a 14x17 cassette
•Demagnify to a 24x30 cm image(air grid)
J. Ed Barnes, Ph.D. - MTMI
CR Mammography Systems
for AGFA, FUJI, KODAK, KONICA
“ All measurements were equivalent to or exceeded the performance of a typical modern film/screen system”
From Centre for Evidence-based Purchasing Report 06047, October 2006
J. Ed Barnes, Ph.D. - MTMI
PSP Radiography (CR)
• Currently the major technology available for large field-of-view digital imaging
• Based upon the principles of photostimulated luminescence; 20+ years of experience
• Operation emulates the screen-film paradigm in use and handling.. (flexible but labor intensive)
• Manufacturing trends: – Smaller, faster, less expensive
J. Seibert - UC Davis
PSP Detector• Photostimulable Storage Phosphor (PSP)
Phosphor PlatePhosphor PlatePhosphor Plate Cassette HolderCassette HolderCassette Holder
• Coating thickness:– Standard resolution: ~100 μm BaFBr– High resolution: ~50-70 μm BaFBr
• Dual-side read; structured phosphor – CsBr
Laser Laser stimulationstimulation2 eV2 eV
ee
Incident xIncident x--rays rays
8 eV8 eV
F center trapF center trapee--PSLPSL
3 eV3 eVeeeeee
J. Seibert, Ph.D. - UC Davis
Computed Radiography Computed Radiography ““readerreader””
Information panelInformation panel
Plate Plate stackerstacker
FujiFuji
J. Seibert - UC Davis
CRCRReaderReader
CR QC CR QC WorkstationWorkstation
Display / ArchiveDisplay / Archive
Laser film printerLaser film printer
DICOM / PACSDICOM / PACS
Image AcquisitionImage Acquisition
CR CR ““systemsystem””: more than the IP: more than the IP’’s and the reader!!s and the reader!!
J. Seibert -UC Davis
Digital Image Networking
• Modality Worklist Input (from RIS via HL-7)
• Technologist QC Workstation– Image manipulation processing reconciliation– Reconciliation and image QA
• PACS and DICOM– Digital Imaging Communications in Medicine– Provides standard for modality interfaces, storage/retrieval, and
• DICOM image output
J. Seibert - UC Davis
Photostimulated Luminescence
Incident Laser Incident Laser BeamBeam
PMTPMT
Protective LayerProtective Layer
Phosphor LayerPhosphor Layer
Base SupportBase Support
LightLightScatteringScattering
Laser Light SpreadLaser Light Spread
PhotostimulatedPhotostimulatedLuminescenceLuminescence
"Effective" readout diameter"Effective" readout diameter
ExposedExposedImagingImaging
PlatePlate
Light guideLight guidePSLPSLSignalSignal
J. Seibert - UC Davis
CR PointCR Point--scan readoutscan readout
PMTPMT
Polygonal Polygonal MirrorMirror
LaserLaserSourceSource
Light channeling guideLight channeling guide
Output SignalOutput Signal
ReferenceReferencedetectordetector
CylindricalCylindricalmirrormirrorff--θθ
lenslens
ADCADC
Laser beam: Laser beam: Scan directionScan direction
Plate translation: Plate translation: SubSub--scan directionscan direction
To imageTo imageprocessorprocessor
ADCADC
x= 1279x= 1279y= 1333y= 1333z= 500z= 500
ERASURE → ReuseJ. Seibert - UC Davis
Scan DirectionScan Direction
SubSub--scan Directionscan Direction
Laser beam deflectionLaser beam deflection
Plate translationPlate translation
Typical CR resolution:35 x 43 cm -- 2.5 lp/mm (200 μm)24 x 30 cm -- 3.3 lp/mm (150 μm)18 x 24 cm -- 5.0 lp/mm (100 μm)
Screen/film resolution:7-10 lp/mm (80 μm - 25 μm)
J. Seibert - UC Davis
Phosphor Plate CyclePhosphor Plate Cycle
PSP PSP
Base supportBase support
reusereuse
plate erasure:plate erasure:remove residual signalremove residual signal
light erasurelight erasure
plate exposure:plate exposure:create latent imagecreate latent image
xx--ray exposureray exposure
plate readout:plate readout:extract latent imageextract latent image
laser beam scanlaser beam scan
J. Seibert - UC Davis
CR Innovations
• Dual side readout
• Mammography applications
• High-speed line scan systems (<10 sec)
• Structured PSP
• Low cost table-top CR readers
J. Seibert - UC Davis
MirrorMirrorTo imageTo imageprocessorprocessor
LightErasure
ADC
Light channeling guideLight channeling guide
Cylindrical mirrorCylindrical mirror
LaserLaserSourceSource
ReferenceReferencedetectordetector ff--θθ
lenslens
CR CR cassettecassette
ErasureErasureStageStageTransportTransport
PMTPMTOutput SignalOutput Signal
To imageTo imageprocessorprocessor
ADC
J. Seibert - UC Davis
CR Mammography
• Dual-side readout capabilities (increased DQE)• Fine laser beam spot size, 50 μm (10 lp/mm)• Mammography-specific image processing
CR cassette
ConventionalMammo system CR reader
Two cassette sizes
QC workstation
18x2424x30
Scanning laser beam
50 μm spot
2 light guides
Transparent base
Low attenuation cassette, special IP
Fuji CRmSystem
CR Plate Reader
CR Plates
Section F = Medical Physicist’s QC Tests
Medical Physics Annual Tests• Mammographic Unit Assembly Evaluation• Collimation Assessment• S-Value Confirmation• System Resolution• kVp Accuracy and Reproducibility• HVL Measurement• Breast Entrance Exposure and Glandular Average
Dose• Radiation Output Rate• Phantom Image Quality Evaluation
- Printed Images- Softcopy
Medical Physics Annual Tests• AEC System Performance• System Artifact Evaluation• CR Reader Scanner Performance• Dynamic Range• Primary Erasure
- Additive and Multiplicative Lag Effects• Inter Plate Consistency• Viewing and Viewing Conditions• Review Workstation Tests• Printer Tests• Technologist’s QC
1. Mammographic Unit Assembly EvaluationPass/Fail/NA
Free-standing unit is mechanically stable PassAll moving parts move smoothly, without obstructions to motion PassAll locks and detents work properly PassImage receptor holder assembly is free from vibrations PassImage receptor slides smoothly into holder assembly PassImage receptor is held securely by assembly in any orientation PassCompressed breast thickness scale is accurate to +/-0.5 cm, reproducible to +/-2 mm PassPatient or operator is not exposed to sharp or rough edges, or other hazards PassOperator technique control charts are posted PassOperator protected during exposure by adequate radiation shielding PassAll indicator lights working properly PassAuto decompression can be overridden to maintain compression (status displayed) PassManual emergency compression release can be activated in the event of a power failure Pass
Comments: The following four items satisfy MQSA Final Regulations:1.) Upon power interruption, the gantry remains stationary and stable.2.) Under compression, the paddle deflection was: 18x24 cm: 3 mm; 24x30 cm: 4 mm.3.) The collimator light illuminance was 420 lux.4.) The compression paddle indicates the location of the AEC sensor. Compression is set at 31 pounds.The technique chart above reflects the results of our measurements and should be posted on the unit. The unit will be used clinically in the Auto-Filter mode.
FCRm Collimation
Test for Routinely Used:
Collimator, Compression Paddle, Bucky Combinations
Target Materials
Fuji CRm System
Measure Exposure Rate at Surface of Breast Support
S-Value Confirmation
FCRm S Value Confirmation
• Make exposure at 25 kVp with Mo/Moto give exposure > 20 mR
• Expose cassette at this technique- wait 10 minutes and process
• Record S value• Calculate the “corrected S value”Action Limit: Corrected S value should not exceed
the range of 120 +/- 20% (96 to 144)
1 - S Value Confirmation
Date:Exposure Unit ID: Lorad M-IVImage Reader ID: Fuji ClearView CSm
Is the 95%uji ClearView CS
kVp: 25mAs: 3.0
Exposure 1 (mR): 37.0Exposure 2 (mR): 37.1Exposure 3 (mR): 37.0
Average Exposure: 37.0 Exposure readings within 5%? Yes
IP Exposure Time 1: 2.0 IP Read Time 1: 2 minIP Exposure Time 2: 2.0 IP Read Time 2: 2 minIP Exposure Time 3: 2.0 IP Read Time 3: 2 min
S Value 1: 66.0 S Value 2: 68.0S Value 3: 65.0
Average S Value: 66.3
Corrected S Value: 122.8
Corrected S Value Pass/Fail? PASSCorrective Action Needed? NO
Description of Corrective Action:Corrective Action Taken? (Yes, No?)
Date of Corrective Action:Results of New Test: (Pass, Fail?)
Comments:
November 21, 2007
FCRm System Resolution
• Up to 10 lp/mm Test Pattern• 4 cm Acrylic Phantom• Parallel and Perpendicular to Tube with
3-5 degree angle
System limiting resolution must be 8 lp/mm +/- 2 lp/mm (6 to 10 lp/mm)
2. Evaluation of System Resolution
X-ray Tube Manufacturer: Varian (1/02) Model #: B-115/M-113R
Visible black points: No NoAnode material Mo MoNominal kVp setting 27 25Nominal mA setting 30 100Density control setting 0 0mAs 57 124S Value (Digital) 58 63Magnification factor 1.80 CONTACTLimiting resolution bars parallel to A-C axis 10 7in line-pairs per mm bars perpendicular to A-C axis 10 7
Action Limit: For Fuji FCRm, if the limiting resolution with the bars parallelto the anode-cathode axis is < 8 +/2 line-pairs/mm orperpendicular to the anode-cathode axis,then a more detailed investigation of the reasonshould be made and corrective action should betaken.
Fuji CRm
FCRm Density Control Function
4 cm Acrylic, Accreditation Phantom Technique
Repeat at -2, -1, 0, 1, 2, etc. density settings
Record mAs
mAs Changes should be 5 to 15% per step
Density Control Function:
Imaging mode: Film/Screen Focal spot: 0.3 mm MOmA: kVp:Phantom thickness: 4 cm BR-12 Cassette ID: QC
Relative to NormalDensity Selector % mAs Density
Setting Change Change5 11.0% Pass4 11.5% Pass3 7.0% Pass2 11.2% Pass1 12.0% Pass0 0% Pass-1 14.4% Pass-2 12.8% Pass-3 9.8% Pass-4 12.2% Pass-5 12.6% Pass
Action Limit: Each step should result in a 5 to 15% change in mAs.If not, seek service.
75.4
125.0109.098.386.3
201.0187.0166.0146.0
mAs255.0227.0
FCRm Reproducibility and Image Mode Tracking• 4 cm Acrylic with Clinical Technique• Position Ion Chamber in Beam• Record mAs and Exposure• Repeat 3 Times• Repeat in each Mode (small, large, mag and no grid)
Action Limits:Coefficient of Variation for Exposure must not exceed 0.05No Significant Difference in Exposure between small and
large bucky when using similar grid
4. Automatic Exposure Control (AEC) System PerformanceAEC position: #3 Density control: 0Small cassette ID: QC Large cassette ID: 24 x 30
Performance Capability:
CNR PER OBJECT THICKNESS Image Mode: Auto TimeFocal spot: LargeDensity Control 0
Phantom thickness Target/Filter kVpmAs
(Auto) CNR
CNR (relative to
4cm PMMA)Acceptable
Level2 cm Mo/Mo 25 26.5 16.3 1.11 >100%4 cm Mo/Mo 25 131 14.7 1.00 100%6 cm Mo/Rh 30 218 12.5 0.85 >75%8 cm
PASS
Image Mode TrackingmA: 100Phantom thickness: 4 cm BR-12Anode/Filter: Mo/Mo
Image Mode AEC MODE Target Filter kVp mAs Commentssmall grid Auto Filter Mo Mo 26 146.0 HTC Gridlarge grid Auto Filter Mo Mo 25 113.0 HTC Grid
Magnification/no grid Auto kV Mo Mo 27 57.7 No Grid
Fuji CRm System
Dynamic Range Test
FCRm Dynamic Range
• To confirm dynamic range of reader and plate
• Use 2 and 4 cm acrylic• Technique for 6 cm• View image to determine if 3 regions are
visible and discernable
Fuji CRm System - CNR Test
PASS
>75%91.6%13.765280.0282.030.0RhMo6 cm
100%100.0%15.030130.0135.025.0MoMo4 cm
>110%114.4%17.19728.027.225.0MoMo2 cm
Acceptable level
CNR (relative to 4cm
PMMA)CNRmAs
(Manual)mAs
(Auto)kVpFilterTargetPhantom Thickness
Auto TimeAEC Mode:
0Density Control:
3Focal Spot:
Auto TimeImage Mode:
CNR PER OBJECT THICKNESS
FCRm Reader Scanner Performance
• Establish that IP Reader and optics do not exhibit scan or print jitter
• Non-Grid Exposure• Position rulers in “T” formation• Expose and ProcessAction Limit:Image must have smooth borders free from jagged
edges or defects
3. CR Reader Scanner Performance Test
Date:
Parallel to Chest Wall
Workstation MonitorCorrective Action Needed?Laser Film Printer*Corrective Action Needed?Corrective Action Description?Corrective Action Taken?Date of Corrective ActionResults of New Test
Perpendicular to Chest Wall
Workstation MonitorCorrective Action Needed?Laser Film Printer*Corrective Action Needed?Corrective Action Description?Corrective Action Taken?Date of Corrective ActionResults of New Test
*If hard copy is used for final interpretation.
November 21, 2007
Image Reader ID: Fuji ClearView CSm # 5722213(Required for Annual Surveys; not required for Mammography Equipment Evaluations of new units. However,
Pass
NoNA
NA
NANANA
NA
Acceptance Limits
Acceptance Limits
NANA
NoPass
PassNo
PassNo
FCRm Primary Erasure
• To assess the erase performance of reader and plates
Additive: Shoot phantom, process, wait 1 minute, reprocess, change S value to 10X original, inspect for visibility of phantom image
Multiplicative: Shoot & process phantom, shoot 4 cm acrylic on same cassette, process and inspect for visibility of phantom image
LSG BH: FTarget Filter kVp mAs
Mo Mo 25 139 Printer:
Acceptable?CORRECTIVE ACTIVE NEEDED?CORRECTIVE ACTIVE DESCRIPTCORRECTIVE ACTIVE TAKEN?CORRECTIVE DATE:RESULTS OF NEW TEST:
Target Filter kVp mAsMo Mo 25 139Mo Mo 25 132
Acceptable?CORRECTIVE ACTIVE NEEDED?CORRECTIVE ACTIVE DESCRIPTCORRECTIVE ACTIVE TAKEN?CORRECTIVE DATE:RESULTS OF NEW TEST:
The mAs noted on the generator read-out must not change by more than ±15% from the previous test.
Percent Change in mAs:Acceptable?
A uniform 4 cm sheet imageACR MAPP
NA
Multiplicative Lag testExposure conditions
NA
PassNoNANA
ACR MAPP
52
520
Artifact Evaluation
S value of the ACR Phantom Image:S value applied to non-exposure image (10 Times Phantom Image S Value)
Cassette / IP Number A30389595C
Additive Lag testExposure conditions
Artifact EvaluationPassNoNA
Pass
NANANA
5.2%
FCRm Inter-Plate Consistency
• To confirm x-ray absorption & SNR consistency• Exposure plates with 4 cm acrylic using clinical
technique• Record mAs (must be within 10%)• Calculate SNR
(must be within 15%)
Yes69.662Yes133A777, #6
Yes69.521Yes132A111, #5
Yes70.166Yes132A222, #4
Yes70.152Yes132A123, #1
Yes70.952Yes132A234, #2
Yes70.946Yes132A456, #3
AcceptableSNRAcceptablemAsCassette ID
Group: Small Cassettes
25MoMoAuto TimeExposure conditions
kVpFilterTargetAEC-mode
Lorad M-IIIExposure Unite ID:
SNR limit is plus or minus 15 % from mean.
mAs limit is plus or minus 10 % from mean.
Inter PlateConsistency
FCRm CNR per Object Thickness• CNR using clinical techique for 2 cm• Repeat for 4 and 6 cm
Action Limits:CNR of 2 cm relative to 4 cm must be > 100%CNR of 6 cm relative to 4 cm must be > 75%
Image Mode: Focal Spot:Density Control:AEC Mode:
Phantom Thickness Target Filter kVp mAs (Auto)
mAs (Manual) CNR
CNR (relative to
4cm PMMA)
Acceptable level
2 cm Mo Mo 25.0 26.2 26.0 13.270 100.8% >100%4 cm Mo Mo 25.0 134.0 135.0 13.165 100.0% 100%6 cm Mo Rh 30.0 227.0 225.0 10.890 82.7% >75%
PASS
CNR PER OBJECT THICKNESS Auto Time
L0
Auto Time
FCRm Contrast to Noise Ratio Test (CNR)• To examine consistency of CNR ratio
measured over time• Use 4 cm Acrylic and 0.2 mm Aluminum• Manual Technique (Mo/Mo, 26 kVp, 125 mAs)• Calculate CNR using softwareAction Limit:+/- 20% of baseline
4. Automatic Exposure Control (AEC) System PerformanceAEC position: #3 Density control: 0Small cassette ID: QC Large cassette ID: 24 x 30
Performance Capability:
CNR PER OBJECT THICKNESS Image Mode: Auto TimeFocal spot: LargeDensity Control 0
Phantom thickness Target/Filter kVpmAs
(Auto) CNR
CNR (relative to
4cm PMMA)Acceptable
Level2 cm Mo/Mo 25 26.5 16.3 1.11 >100%4 cm Mo/Mo 25 131 14.7 1.00 100%6 cm Mo/Rh 30 218 12.5 0.85 >75%8 cm
PASS
Image Mode TrackingmA: 100Phantom thickness: 4 cm BR-12Anode/Filter: Mo/Mo
Image Mode AEC MODE Target Filter kVp mAs Commentssmall grid Auto Filter Mo Mo 26 146.0 HTC Gridlarge grid Auto Filter Mo Mo 25 113.0 HTC Grid
Magnification/no grid Auto kV Mo Mo 27 57.7 No Grid
FCRm Viewing Conditions
• Hardcopy Interpretation: Same conditions as to be used for screen-film mammography images. Film masking devices and hot lights must be available to the radiologist.
• Softcopy Interpretations: Follow the monitor manufacturer’s QC procedures. If no monitor manufacturer’s QC procedures available, follow procedures outlined in Fuji QC Manual.
Fuji CRm - Monitor TestsSite: Med. Img Ctr of So Cal Date of Survey: 4/19/2007
Room #: Reading Room Date of Installation: Apr-07
DOME©E5
Target Value = 500 cd/sq-m
499.2
Serial Numbers
Luminance Value Setting (cd/sq-m)
Left Monitor Right Monitor
703PNKN00028
500.88
703PNKN00011
Is the CXtra icon present in thetaskbar and shown as a green
check mark?
Manual Conformance report forboth displays attached?
Is DICOM calibration Graph forboth displays attached?
NoYes
Yes
Yes
No
No
Laser Printer -must be able to print images on mammography laser film
FCRm Printer Annual QC
• Follow the printer manufacturer’s QC program for test procedures, frequency, performance and corrective action.
• If no printer manufacturer’s QC program, follow the QC program outlined in Fuji QC Manual.
• Tests must be performed initially prior to interpreting hard copy mammograms and after major repairs and then annually.
OD Requirements for OD Requirements for HiHi--Resolution Laser ImagersResolution Laser Imagers
Dmax > 3.5 OD
Mid-density > 1.5 OD
10. Breast Entrance Exposure, AEC Reproducibility, Average GlanduDose, and Radiation Output Rate
Imaging mode: SID (cm):Screen type: Source-detector distance (cm):Film type: Source-bucky distance (cm):Cassette size (cm): Dosimeter used:Field Restriction: Energy correction factor:
Breast thickness (cm)PhantomNominal kVp settingAnode materialFilterAEC modeAEC density control settingMeasured HVL (mm Al)
Breast Entrance Exposure and AEC Reproducibility:R mAs R mAs R mAs R mAs
Exposure #1 1.581 147.0 1.480 121.0 1.231 89.1 0.705 75.2Exposure #2 1.498 139.0Exposure #3 1.521 142.0Exposure #4 1.501 140.0
Mean values 1.525 142.0 1.480 121.0 1.231 89.1 0.705 75.2Standard deviations (SD) 0.039 3.56
0.025 0.025
Action limit: If coefficient of variation for either R or mAs exceeds 0.05, seek service.
Average Glandular Dose:Inv Sq corrected skin exp 1.525 1.480 1.231 0.705
153 159 164 217
233 235 202 153
Action limit: If average glandular dose exceeds 300 mrad (3 mGy) for 4.2 cm effective breastthickness, seek service or technique adjustment. Corrective action must be takenbefore further examinations are performed if the test results fail MQSA regulations.
Radiation Output Rate:
3 sec, kVp Anode Filter SID (cm)
Exp (mR) mAs Time
(sec)Rate
(mR/s)Kerma
(mGy/s)4.5 cm above 28 Mo Mo 65 4740 300 3 1580 13.8breast support 28 Mo Mo
Air Kerma (mGy/sec) = Exp Rate (mR/s) x 0.00873 mGy/mR
Action limit: If output rate is less than 800 mR/sec (7.0 mGy air kerma/sec), seek service.
Coefficient of variation (CV)
Dose conversion factor from Tables 1-3 (mrad/R)Computed average glandular dose (mrad)
MoAuto Filter
00.3
1.00
DigitalFuji CR
CR18 x 24
656064
Keithley TriadCollimators
RhAuto Time
00.43
4.2RMI 156-15914
28Mo
MoAuto Time
00.32
4.2RMI 156-15914
27Mo
MoAuto Time
00.31
25Mo
4.2RMI 156-15914
26Mo
RMI 156-159144.2
Phantom Image Quality Check
Possible Items ( 6 )
( 5 )
( 5 )
Proper Scoring Techniques of the ACR Mammography Accreditation Phantom
Fibers• Full Point Score
- Full length of fiber visible (+/- 1 mm) -use first fiber as a reference for length- Correct location- Correct orientation
Fibers• Half Point Score
- Not all, but > 50% visible- Correct location- Correct orientation- Stop if orientation or borders not visible
Fibers - Artifacts• In Wrong Location of Phantom Insert
– Fiber-like– At least as apparent as last fiber scored– Diameter and length > last scored fiber
Fibers - Artifacts -Deductions
• Deduct 1.0 - last fiber is scored as whole and artifact is > 75% of full fiber length
• Deduct 0.5 - last fiber is scored as 1/2 and artifact is > 50% of full fiber length
• Only deduct from last fiber scoredAction Limits: Minimum of 4 fibers for all units
except Hologic and Siemens FFDM units. Minimum of 5 fibers for Hologic and Siemens FFDM units
Speck GroupsFull Point Score
- > 4 specks visible- Correct locations
Half Point Score- 2 or 3 specks visible- Correct locations
Speck Groups - Artifacts• Noise or speck-like• As apparent as last specks scored (i.e. - at
least as bright and > diameter)• Do not subtract bright artifacts caused by
dust or emulsion pickoff
Speck Groups - Deductions• Deduct 1-for-1 from scored specks• Only deduct from the specks(s) in the last
group scored (with either a 1.0 or 0.5 score)
Speck GroupsAction Limit:
Minimum score of 3 speck groups for all units except Hologic and Siemens FFDM unitsMinimum score of 4 speck groups for Hologic and Siemens FFDM units
Masses Score• Full Point
- Density difference visible- Generally circular
(> 75% circumference)- Correct location
Masses Score• Half Point
- Density difference visible- Not generally circular (< 75%
circumference)- Correct location
Masses Score - ArtifactsIn Wrong location of Phantom Insert
- Mass - like- As apparent as last mass scored
DeductionOnly deduct from last mass scored
Masses ScoresAction Limit:
Minimum score of 3 masses for all units except Hologic and Siemens FFDM unitsMinimum score of 4 masses for Hologic and Siemens FFDM units
Acceptable Phantom ImageWW = 200 (suggestion)
Description: AcceptableACR Phantom image
Minimum Acceptable Phantom Image Scorefor the Fuji FCRm systemis:4 fibers, 3 speck groups,3 masses
Image Quality Values
• The Phantom Image Quality on the Fuji CRm system will reflect very closely what was achieved previously on the film/screen system in use - expect no significant change in quality.
Site Name Report DateAddress Survey DateMedical Physicist's Name SignatureX-Ray Unit Manufacturer ModelDate of Unit Installation Room IDFFDM Image Receptor Mfr FFDM ModelFFDM QC Manual Version: (check one; must use version applicable to system tested; contact mfr if questions)
Accessory Equipment:Review Workstation*
Laser Film Printer*
Survey Type: Mammo Eqpt Evaluation of new unit (include MQSA Rqmts for Mammo Eqpt checklist) Annual Survey
PASS/FAIL1. S Value Confirmation ( < 120 + 20% [96 < corrected S value < 144])2. System Resolution (8 lp/mm + 2 lp/mm in both directions)
3. CR Reader Scanner Performance4. Imaging Plate (IP) Fog5. Mammography Unit Assembly Evaluation6. Collimation Assessment Test date if different from above:
Chest wall edge of X-ray field extends to edge of IRDeviation between X-ray field and light field <2% of SIDX-ray field does not extend beyond any side of the IR by more than 2% of SIDPaddle chest wall edge not beyond IR by more than 1% of SID or appear on the image
7. Automatic Exposure Control (AEC) System Performance AssessmentAEC density control function meets Fuji performance criteriaReproducibility (CV) for either exposure or mAs is <0.05Image mode tracking meets Fuji performance criteriaCNR per object thickness meets Fuji performance criteria
8. System Artifact Evaluation9. Phantom Image Quality Evaluation Fibers Specks Masses (at least 4 fibers, 3 speck groups & 3 masses)
Phantom IQ (printed images)Phantom IQ (softcopy)Other tests meet Fuji performance criteria (mAs, OD & DD for hardcopy, S value for soft copy)
10. Dynamic Range11. Primary Erasure12. Inter-Plate Consistency (variation of mAs within + 10%; SNR within + 15%)
13. kVp Accuracy and Reproducibility Test date if different from above:Measured average kVp within +5% of indicated kVpkVp coefficient of variation <0.02
14. Dose (average glandular dose for average breast is < 3 mGy [300 mrad]) mrad15. Beam Quality Assessment & HVL Measurement Test date if different from above:
HVL >kVp/100 mm Al16. Radiation Output Test date if different from above:
Radiation output rate is >800 mR/sec (7.0 mGy/sec) at 28 kVp with Mo/Mo17. Viewing and Viewing Conditions Test date if different from above:
Mammographic viewbox luminance >3000 cd/m^2 (nit)Room illuminance (viewbox surface & seen by observer) <50 lux
18. Review Workstation (RWS) Tests** (for all RWS, even if located offsite)Overall Results ("Pass" means all tests pass; indicate "Fail" if any test fails)
Fuji FCRm
MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARYFull-Field Digital – Fuji
Model QC Manual Version
Medical Physicist's QC Tests
Manufacturer Location
*FDA recommends that only monitors and printers specifically cleared for FFDM use by FDA’s Office of Device Evaluation (ODE) be used. See FDA's Policy Guidance Help System www.fda.gov/CDRH/MAMMOGRAPHY/robohelp/START.HTM
*** YOUR MEDICAL PHYSICIST MUST SUMMARIZE HIS/HER RESULTS ON THIS FORM ***
**FDA requires that all RWS comply with a QC program that is substantially the same as that recommended by the image receptor manufacturer . If the RWS is FDA-approved, the RWS's QC manual is considered to be "substantially the same" and you may follow it. (Check with the RWS manufacturer for their FDA clearance status and QC manual.) If the RWS is not FDA-approved for FFDM, you must follow the QC manual provided by the image receptor manufacturer. (Check with the image receptor manufacturer for their required tests.)
1st edition, 07.2006 897N0602, 2006
On-site Off-site
On-site Off-site
OTHER (write in):
Click in boxes to use drop-down lists
Evaluation of Site's Technologist QC Program
PASS/FAILMQSA Regs
1. CNR Weekly Check Weekly2. Phantom Image Weekly3. Visual Checklist Monthly4. Repeat Analysis Quarterly5. Compression Semi-annually6. Printer QC* See FDA guidance7. Review Workstation QC-Overall* See FDA guidance
Medical Physicist's Recommendations for Quality Improvement
MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY
(Required for Annual Surveys; not required for Mammography Equipment Evaluations of new units. However, medical physicists must review the site's technologist QC program within 45 days and complete this section so that the facility may submit this form along with the entire
Mammography Equipment Evaluation report with their phantom and clinical images to the ACR.)
(Fuji, continued)
Fuji Digital Mammography QCTechnologist’s Tests
TESTCNR-Contrast/NoisePhantom ImagePrinter/MonitorVisual ChecklistRepeat AnalysisCompressionIP Fog
FREQUENCYWeeklyWeekly
Per ManufacturerMonthly
QuarterlySemi-annualSemi-annual
Fuji Digital Mammography QCTechnologist’s Tests
Printer TestsNo film in cassetteImaging Plate should be erased after 8
hours of non-useImaging Plate should be processed or
erased after each exposurePrinters are monitored according to the
manufacturer’s approved QC program
Fuji Digital Mammography QCTechnologist’s Tests
Phantom Images TestConfirms image quality using ACR PhantomTools required - QC Cassette & ACR PhantomMethod - AEC Exposure, process IP using the
Physics, ACR MAPP MenuVisually inspect the image. Same scoring method
as MQSACorrection Period - Before any further
examinations are performed
Fuji Digital Mammography QCTechnologist’s Tests
CNR (Contrast/Noise Ratio) TestEvaluates Noise and Contrast using a fixed x-
ray exposureTools needed - 4 cm acrylic block & 0.2 mm AlMethod - Fixed manual techniqueCNR Measurement must be within +/- 20%Correction Period - Before any further
examinations are performed
Fuji Digital Mammography QCTechnologist’s Tests
Tests in Accordance with ACR/MQSAVisual Checklist MonthlyRepeat Analysis QuarterlyCompression Test Semi-Annual
Same Action Limits and Corrective Action as MQSA
IP Fog replaces Darkroom Fog test - Semi Annual
Review Workstation QC
(See FDA guidance)
Compression(25-45 lb)
(semi-annually)
Repeat Analysis(≤2% change)
(quarterly)
Visual Checklist (monthly)
DECNOVOCTSEPAUGJULJUNMAYAPRMARFEBJANMonth
Year
Visual Checklist - Technologist QC
FULL-FIELD DIGITAL MAMMOGRAPHY QUALITY CONTROL CHECKLIST –FUJI FCRmDaily and Weekly Tests
Review Workstation QC
(See FDA guidance)
Printer QC (See FDA guidance)
Phantom Image (weekly)
CNR Weekly Check (weekly)
Initials
Date
Month
Year
Fuji Digital Mammography QCTechnologist’s Tests
Imaging Plate FogConfirm that the storage conditions of the exposure
room protects the IP cassettes from scatterTools required - IP CassetteMethod - Tape coin to front of cassette and place in
storage location during test exposuresVisually inspect image for visibility of coinCorrection Period - Before any further examinations
are performed
Fuji Digital Mammography QCTechnologist’s Tests
Inter-Plate Consistency - As NeededThe Inter-plate Consistency Test replaces the
Screen Speed TestMust be performed when new IPs and/or
cassettes are introducedCan be performed by the QC technologist under
Qualified Medical Physicist oversight
Fuji Digital Mammography QCTechnologist’s Tests
Inter-Plate Consistency - As NeededConfirm that Imaging Plates used in the facility
are similar in sensitivity and image qualityTools Required - new and existing IPs and
Cassettes, 4 cm acrylic blockMethod - Expose cassettes using AEC (time)
ModeAcceptable results: mAs +/- 10$, SNR +/- 15%Correction Period - Within 30 days of the test
Thank You !!!
Melissa C. Martin, M.S., FACRTherapy Physics Inc.
879 West 190 St., Ste 419Gardena, CA 90248
Office Phone: 310-217-4114Cell Phone: 310-612-8127
e-mail: [email protected]