FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT™ SYSTEM

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT SYSTEM SYSTEM SYSTEMON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACT SYSTEMON THE RADIXACT SYSTEMON THE RADIXACT SYSTEMON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACT SYSTEMON THE RADIXACT SYSTEMON THE RADIXACT SYSTEMON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACT SYSTEM SYSTEM SYSTEM SYSTEM SYSTEM SYSTEM SYSTEM SYSTEM SYSTEMON THE RADIXACT SYSTEMON THE RADIXACT SYSTEMON THE RADIXACT SYSTEM

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACTON THE RADIXACT

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT

Eric Schnarr, Etienne Lessard, Matt Beneke, Dylan Casey, Edward Chao, Jonathan Chappelow, Doug Henderson, Petr Jordan, Dan Lucas, Andriy Myronenko, Calvin Maurer

Accuray Incorporated White Paper08 April 2017

Page 2: FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE … · 2019. 4. 24. · FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT™ SYSTEM Abstract In this study, we adapted the

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT™ SYSTEM

Abstract

In this study, we adapted the motion tracking technology from the CyberKnife® Treatment Delivery System to a system capable of delivering helical tomotherapy treatments and performed motion correction using the jaws and binary multileaf collimator (MLC). To enable motion modeling and tracking, a kV flat-panel imaging system was added to an existing gantry, mounted perpendicular to the MV treatment beamline, and an optical camera was mounted above the foot of the couch. Flat-panel images were acquired every few seconds to determine internal target positions and the target positions were correlated with breathing amplitudes detected continuously by the camera view of external markers. Jaw positions and MLC leaf patterns were updated continuously to follow the target motion. Film measurements were taken of deliveries, with and without motion management enabled, and compared against the planned (no-motion) dose. With motion correction enabled, dose delivered to a respiratory motion target closely matched the planned dose. For a target with non-periodic motion (e.g., prostate-like motion), dose delivered with motion correction enabled was substantially improved compared to delivery when no motion correction was used.

Introduction

Targets move during radiotherapy treatment for many reasons, including bodily processes, such as respiration and digestion. It is clinically beneficial to account for this motion, to ensure that the target receives the prescribed dose, as well as to minimize toxicity associated with irradiation of surrounding healthy tissue. Accounting for motion is particularly important when higher doses of radiation are delivered in fewer sessions than conventional radiotherapy, such as for stereotactic body radiation therapy (SBRT).

Various pre-treatment and in-treatment approaches have been used to manage target motion. Pre-treatment alternatives include expanding planning target volume (PTV) margins, immobilizing the patient or target, or both. Immobilization techniques include the use of deep inspiration breath-hold,1 abdominal compression,2 bladder and bowel preparation3 and other similar methods of physically preventing motion. But these methods are generally inconvenient and negatively impact patient comfort. Margin expansion for respiratory motion is usually done through the definition of an internal target volume (ITV) that covers the target throughout its entire range of motion. Whether using an ITV for respiratory motion, or simply expanding PTV margins for other motion-related setup uncertainty, the resulting target volumes can be large and include additional healthy tissue. Furthermore, it has been shown that target motion can change significantly, from day to day and even during treatment, which means that an ITV defined during treatment planning may not accurately cover the target range of motion during treatment delivery.4

In-treatment techniques focus on adjusting the treatment delivery to the patient, rather than forcing patient anatomy to stay in the planned path of the treatment beam. This requires technology to track the target position and hardware capable of changing the planned treatment delivery in real time to correct for the motion. For conventional radiotherapy systems, this usually means stopping treatment delivery when a motion larger than a specified threshold is detected, repositioning the couch, acquiring additional images to verify the patient’s position and then resuming treatment delivery.

In the case of targets that move with respiration, some conventional radiotherapy systems gate the treatment delivery based on the patient’s respiratory cycle, i.e., these systems turn the radiation on only when the target is in the path of the treatment beam. Respiratory gating is time consuming, because the radiation is turned off for most of the respiratory cycle; the duty cycle is typically less than 30%. Wider gating windows – the portion of the respiratory cycle during which radiation is turned on – means faster treatment, but also reduces the precision of delivery and results in more normal tissue being irradiated.5

Page 3: FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE … · 2019. 4. 24. · FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT™ SYSTEM Abstract In this study, we adapted the

3

The most advanced technique goes further and re-targets the treatment beam in real-time to follow the moving target. The treatment beam moves as the target gradually drifts (common for intracranial and spine targets), unpredictably shifts (common for prostate and gynecological targets) or moves with respiration (common for abdominal and thoracic targets). This technique maintains the delivery accuracy throughout treatment delivery and enables planning with minimal PTV margin expansion.6, 7, 8 For targets that move with respiration, real-time re-targeting enables PTV margins similar to those used with very short gating windows, but without increasing delivery times (a 100% duty cycle).9, 10, 11, 12, 13

The CyberKnife System is the only radiotherapy device that can maintain sub-millimeter accuracy throughout treatment delivery with fully integrated image guidance, automatic motion tracking, and real-time robotic beam re-targeting to correct for motion. To track motion during delivery it uses an orthogonal imaging system with diagnostic kV sources fi xed to the ceiling and fl at-panel detectors located in the fl oor, along with an optical camera positioned above the couch.14 The system also includes a suite of proprietary tracking algorithms specifi cally designed for particular clinical applications. Translational and rotational motion of soft tissue targets, like the prostate, can be tracked using implanted fi ducial markers. Xsight® Lung Tracking can track lung targets without implanted fi ducial markers, eliminating the risk of pneumothorax due to marker implantation.15, 16, 17 The sub-millimeter accuracy of these and other tracking algorithms on the CyberKnife System has been demonstrated and reported in the literature.18, 19

In addition, targets that move with respiration, such as lung and liver, are tracked with the Synchrony® Respiratory Motion Tracking System. This system dynamically aligns the treatment beam to the moving target and synchronizes the treatment delivery to the patient’s normal breathing, without compression and/or breath holding. A robotic arm moves the treatment beam based on a correlation between the target position, determined from periodic kV image acquisitions, and the breathing amplitude continuously monitored with external optical markers. This respiratory correlation model is built before each treatment and periodically updated during treatment delivery with additional kV image acquisitions to adjust to the patient’s gradually changing respiratory pattern. Synchrony Respiratory Motion Tracking is the most advanced respiratory motion management system, and is unique to the CyberKnife System. Its sub-millimeter accuracy has been demonstrated and reported.20, 21, 22

Accuray Incorporated has unique expertise in image-guidance technologies, anatomical site-specifi c tracking algorithms and robotic beam delivery correction. It is our intention to transfer the currently approved and proven effective technology and expertise from the CyberKnife System to the Radixact™ Treatment Delivery System. This white paper describes experimental work that has been done to evaluate the feasibility of this technology transfer.

Experimental Methods and Results

Helical tomotherapy treatment delivery provides precise treatments with conformal dose distributions, but the system’s continuous motion makes this modality incompatible with some forms of motion management. For example, the continuously rotating gantry is unlikely to be at the correct angle to synchronize dose delivery with a narrow window of breathing phases, as required for gated delivery.

However, real-time re-targeting, similar to that performed by the CyberKnife System, is compatible with helical delivery modalities. As the couch progressively moves into the gantry and the linear accelerator continuously rotates around the patient, the jaw positions could be modifi ed to compensate for superior-inferior target motion (Figure 1) and the binary MLC leaf patterns could be modifi ed to compensate for anterior-posterior and mediolateral motions (Figure 2).

Figure 1a. Jaws open

symmetrically on target

Figure 1b. Target shifted in

superior direction; jaws move to compensate

Page 4: FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE … · 2019. 4. 24. · FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT™ SYSTEM Abstract In this study, we adapted the

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT™ SYSTEM

Figure 2. As the couch moves the patient into the gantry, the jaws can move in the IEC-y direction to compensate for superior-inferior target motion.

Figure 3. Experimental setup showing additional kV imaging components mounted on a gantry. The drawing on the left shows the conceptual setup, illustrating the perpendicular MV and kV beam lines and optical camera. On the right is our

actual experimental system, highlighting the kV imaging components mounted orthogonal to the MV beam-line.

Before the system can correct for motion, it must fi rst be able to detect the position of the target. To track internal target motion like the CyberKnife System, we mounted kV fl at-panel imaging hardware onto a gantry capable of helical tomotherapy deliveries, as shown in Figure 3. A kV tube and detector were mounted at 90 degrees offset from the MV treatment beam-line. In addition, an optical camera was installed above the foot of the couch, looking into the bore.

An in-house robotic phantom was used to simulate realistic target motion. The phantom consists of three linear actuators supporting an acrylic cylinder, in which can be placed a fi lm and ion chamber to acquire dosimetric measurements. Gold fi ducial markers – visible in kV images – were implanted inside the acrylic cylinder and light-emitting diode (LED) markers visible to the optical camera were attached to the moving stage of the robotic platform.

10 20 30 40 50 60 10 20 30 40 50 60

(2a) Planned MLC leaf pattern for a 5 cm target at isocenter, with no motion

(2b) Delivered MLC leaf pattern for a 5cm target that starts at isocenter, and then experiences a 2.5 cm shift in the transverse plane

5 cm targetat isocenter

Target moved2.5 cm away from isocenter in the transverse plane

LINACJaws

and MLC

MVDetector

kVDetector

kVDetector

kVGenerator

MLC Leaf Number

Time (gantry rotates)

Time (gantry rotates)

MLC Leaf Number

kVTube

MVDetector

LINAC

Jawsand MLC

kVTube

kVDetector

kVGenerator

(not visible; mounted on the back side of the gantry – the mounting plate is shown here)

Page 5: FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE … · 2019. 4. 24. · FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT™ SYSTEM Abstract In this study, we adapted the

5

The experimental system acquires periodic kV images to track the gold fi ducials, then correlates the fi ducial positions with breathing amplitudes continuously monitored by the optical camera following the external LED markers. These sequential monoscopic kV images are acquired as the gantry rotates to build an initial correlation model. The model is updated by acquiring additional images throughout the subsequent treatment delivery. The impact of MV scatter on kV fl at-panel images was negligible, so kV images can be acquired without interrupting the treatment delivery. Jaw positions and MLC leaf patterns are updated continuously in real-time, re-shaping (effectively re-pointing) the treatment beam to follow the target motion. Jaw and leaf adjustments are made every 10 ms with only a few tens of milliseconds end-to-end latency.

Respiratory Motion Experiment

To test the system’s effi cacy in correcting for respiratory motion, we delivered a helical tomotherapy treatment plan to a small moving target. A small target was chosen since dose delivery uncertainties resulting from motion are more noticeable in the dose profi les measured for smaller targets. The treatment plan was prepared to cover at least 95% of a 1 cm target volume with 2.5 Gy, which resulted in a dose distribution just over 2 cm wide at the half-max isodose line. The treatment duration was 126 seconds.

To simulate respiratory motion, the phantom was programmed to move linearly in all three axes using the motion trace in Figure 4, which has a maximum amplitude of approximately 15 mm. This motion trace corresponds to a typical respiratory breathing pattern recorded during treatment delivery by the Synchrony® Respiratory Tracking system.

Figure 4: Respiratory motion trace used to drive all three axes of the robotic phantom in this experiment.

Figure 5: Respiratory motion experiment fi lm measurement results. Superior-inferior dose profi les measured through the center of the target, with no motion (blue), motion but no correction (orange) and motion with real-time correction (green). The blue and green lines closely match, demonstrating the system’s ability to correct for respiratory motion.

Superior-inferior dose profi les were measured through the center of the target using EBT3 radiochromic fi lm and compared against the same plan in three different scenarios: with no phantom motion, with phantom motion without correction and with phantom motion with correction. Results are shown in Figure 5.

We observe that when no motion correction was applied (orange line), the dose profi le is blurred and shifted, and the peak drops by 19% compared the reference with no phantom motion (blue line). When motion correction is enabled (green line), the dose profi le closely matches the reference, demonstrating the experimental system’s ability to correct for respiratory motion.

0

5

10

15

0 20 40 60 80 100 120

mm

Seconds

0.0 Gy

0.5 Gy

1.0 Gy

1.5 Gy

2.0 Gy

2.5 Gy

-4 cm -3 cm -2 cm -1 cm 0 cm 1 cm 2 cm 3 cm 4 cm

No MotionMotion Without CorrectionMotion With Correction

Superior-Inferior Dose Profiles

Superior-Inferior Dose Profiles

No MotionMotion Without CorrectionMotion With Correction

-4 cm -3 cm -2 cm -1 cm 0 cm 1 cm 2 cm 3 cm 4 cm 5 cm0.0 Gy1.0 Gy2.0 Gy3.0 Gy4.0 Gy5.0 Gy

6.0 Gy

L/RS/IU/D

3303002702402101801501209060300

0

-5

5

10

15

mm

Seconds

0

5

10

15

0 20 40 60 80 100 120

mm

Seconds

0.0 Gy

0.5 Gy

1.0 Gy

1.5 Gy

2.0 Gy

2.5 Gy

-4 cm -3 cm -2 cm -1 cm 0 cm 1 cm 2 cm 3 cm 4 cm

No MotionMotion Without CorrectionMotion With Correction

Superior-Inferior Dose Profiles

Superior-Inferior Dose Profiles

No MotionMotion Without CorrectionMotion With Correction

-4 cm -3 cm -2 cm -1 cm 0 cm 1 cm 2 cm 3 cm 4 cm 5 cm0.0 Gy1.0 Gy2.0 Gy3.0 Gy4.0 Gy5.0 Gy

6.0 Gy

L/RS/IU/D

3303002702402101801501209060300

0

-5

5

10

15

mm

Seconds

Page 6: FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE … · 2019. 4. 24. · FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT™ SYSTEM Abstract In this study, we adapted the

FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT™ SYSTEM

Prostate Motion Experiment

We also evaluated the system’s ability to correct for non-respiratory motion, such as the gradual drift and unpredictable shifts that can occur in prostate treatments. The robotic phantom was programmed to reproduce a real prostate motion trace previously acquired with the Calypso® system (Figure 6). The range of motion was +8 mm to -5 mm with intermittent excursions up to 14 mm.

Figure 6. Prostate motion trace acquired from a real patient using Calypso, left-right motion (blue), superior-inferior motion (orange) and up-down motion (green).

In the case of non-periodic motion, there is no need to use the external camera. The system used only the sequential monoscopic kV images, acquired as the gantry rotated, to track the positions of the gold fi ducials. The system acquired two kV images per rotation, 90 degrees apart, with a gantry rotation period of 12 seconds. Jaw positions and MLC leaf patterns were automatically adjusted to account for detected target motion after each image acquisition.

A typical hypo-fractionated prostate helical treatment was delivered. The treatment plan was prepared to cover at least 95% of a 5.5 cm diameter volume with 5.0 Gy, which resulted in a dose distribution approximately 6.5 cm wide at half-max. The treatment plan used TomoEDGE™ and the 2.5 cm fi eld. The treatment duration was 340 seconds. The superior-inferior dose profi les were measured through the center of the target using EDR2 fi lm, and compared against the same plan in three different scenarios: with no phantom motion, with phantom motion without correction and with phantom motion with correction.

Figure 7. Prostate motion experiment fi lm measurement results. Superior-inferior dose profi les measured through the center of the target, with no motion (blue), motion but no correction (orange) and motion with correction (green).

Film measurements are shown in Figure 7. We observe that when no motion correction is applied (orange line), the dose profi le signifi cantly changes compared to the reference with no phantom motion (blue line). The dose profi le central region is hotter by up to 16%, and both superior and inferior ends are shifted toward the middle by 0.5 cm and 0.3 cm, respectively. When motion correction is enabled (green line), the dose profi le more closely matches the reference, demonstrating the system’s ability to correct for a typical prostate motion with gradual drift and intermittent excursions.

The limiting factor in this case is the frequency of kV image acquisition. Unlike respiratory motion tracking, where the optical camera provides a continuous motion signal, new target position information is only available from the

0

5

10

15

0 20 40 60 80 100 120

mm

Seconds

0.0 Gy

0.5 Gy

1.0 Gy

1.5 Gy

2.0 Gy

2.5 Gy

-4 cm -3 cm -2 cm -1 cm 0 cm 1 cm 2 cm 3 cm 4 cm

No MotionMotion Without CorrectionMotion With Correction

Superior-Inferior Dose Profiles

Superior-Inferior Dose Profiles

No MotionMotion Without CorrectionMotion With Correction

-4 cm -3 cm -2 cm -1 cm 0 cm 1 cm 2 cm 3 cm 4 cm 5 cm0.0 Gy1.0 Gy2.0 Gy3.0 Gy4.0 Gy5.0 Gy

6.0 Gy

L/RS/IU/D

3303002702402101801501209060300

0

-5

5

10

15

mm

Seconds

0

5

10

15

0 20 40 60 80 100 120

mm

Seconds

0.0 Gy

0.5 Gy

1.0 Gy

1.5 Gy

2.0 Gy

2.5 Gy

-4 cm -3 cm -2 cm -1 cm 0 cm 1 cm 2 cm 3 cm 4 cm

No MotionMotion Without CorrectionMotion With Correction

Superior-Inferior Dose Profiles

Superior-Inferior Dose Profiles

No MotionMotion Without CorrectionMotion With Correction

-4 cm -3 cm -2 cm -1 cm 0 cm 1 cm 2 cm 3 cm 4 cm 5 cm0.0 Gy1.0 Gy2.0 Gy3.0 Gy4.0 Gy5.0 Gy

6.0 Gy

L/RS/IU/D

3303002702402101801501209060300

0

-5

5

10

15

mm

Seconds

Page 7: FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE … · 2019. 4. 24. · FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT™ SYSTEM Abstract In this study, we adapted the

7

periodic kV images. Motion that occurs between acquisitions (alternately at 3 second and 9 second intervals) is not corrected for, and results in some residual delivery inaccuracy. More frequent imaging would further improve the motion corrected delivery accuracy.

Discussion and Conclusion

This experimental work describes a possible adaptation of tracking and motion compensation technology from the CyberKnife System to the Radixact System, including experimental results demonstrating the feasibility of this approach. This could benefi t many clinical indications, taking advantage of Accuray’s proprietary tracking algorithms to position the target pre-treatment, and automatically adapt the treatment during delivery to account for gradual drifts, unpredictable shifts, and respiratory motion. As the adaptation is real-time, the beam-on time for a motion-corrected treatment does not take additional time compared to a non-motion-corrected treatment. This would mean faster motion corrected treatment delivery compared to gated treatment delivery References

1. Latty D, Stuart K E, Wang W, Ahern V, Review of deep inspiration breath-hold techniques for the treatment of breast cancer. J Med Radiat Sci. 2015; 62(1): 74–81, 2015.

2. Negoro Y, Nagata Y, Aoki T, Mizowaki T, Araki N, Takayama K, Kokubo M, Yano S, Koga S, Sasai K, Shibamoto Y, Hiraoka M, The effectiveness of an immobilization device in conformal radiotherapy for lung tumor: reduction of respiratory tumor movement and evaluation of the daily setup accuracy. Int J Radiat Oncol Biol Phys. 2001; 50:889–98.

3. Yahya S, Zarkar A, Southgate E, Nightingale P, Webster G, Which bowel preparation is best? Comparison of a high-fi bre diet leafl et, daily microenema and no preparation in prostate cancer patients treated with radical radiotherapy to assess the effect on planned target volume shifts due to rectal distension. The British Journal of Radiology 2013; 86(1031):20130457.

4. Minn A Y, Schellenberg D, Maxim P, Suh Y, McKenna S, Cox B, Dieterich S, Xing L, Graves E, Goodman K A, Chang D, Koong A C, Pancreatic tumor motion on a single planning 4D-CT does not correlate with intrafraction tumor motion during treatment. Am J Clin Oncol 2009; 32(4): 364-8.

5. Jiang S B, Technical aspects of image-guided respiration-gated radiation therapy. Med Dosim. 2006; 31: 141–51.6. Xie Y, Djajaputra D, King C R, Hossain S, Ma L, Xing L, Intrafractional motion of the prostate during hypofractionated radiotherapy. Int J Radiat Oncol Biol Phys

2008; 72: 236-46.7. van de Water S, Valli L, Aluwini S, Lanconelli N, Heijmen B, Hoogeman M, Intrafraction Prostate Translations and Rotations During Hypofractionated Robotic

Radiation Surgery: Dosimetric Impact of Correction Strategies and Margins. Int J Radiat Oncol Biol Phys 2012; 88(5):1154-60.8. Hoogeman M S, Nuyttens J J, Levendag P C, Heijmen B J, Time dependence of intrafraction patient motion assessed by repeat stereoscopic imaging. Int J

Radiat Oncol Biol Phys 2008; 70(2):609-18.9. Yoon K, Kwak J, Cho B, Park J H, Yoon S M, Lee S W, Kim J H, Gated Volumetric-Modulated Arc Therapy vs. Tumor-Tracking CyberKnife Radiotherapy as

Stereotactic Body Radiotherapy for Hepatocellular Carcinoma: A Dosimetric Comparison Study Focused on the Impact of Respiratory Motion Managements. PloS one 2016; 11(11).

10. Rault E, Lacornerie T, Dang H P, Crop F, Lartigau E, Reynaert N, Pasquier D, Accelerated partial breast irradiation using robotic radiotherapy: a dosimetric comparison with tomotherapy and three-dimensional conformal radiotherapy. Radiation oncology 2016; 11(29).

11. Hoogeman M S, Prevost J B, Nuyttens J, Poll J, Levendag P, Heijmen B, Clinical accuracy of the respiratory tumor tracking system of the CyberKnife: assessment by analysis of log fi les. Int J Radiat Oncol Biol Phys 2009; 74(1): 297-303.

12. Korreman S S, Juhler-Nottrup T, Boyer A L, Respiratory gated beam delivery cannot facilitate margin reduction, unless combined with respiratory correlated image guidance. Radiother Oncol 2008; 86:61–68.

13. Seppenwoodle Y, Berbeco R I, Nishioka S, Shirato H, Heijmen B, Accuracy of tumor motion compensation algorithm from a robotic respiratory tracking system: A simulation study. Med Phys 2007; 34(7): 2774-84.

14. Kilby W, Dooley J, Kuduvalli G, Sayeh S, Maurer C R, The CyberKnife Robotic Radiosurgery system in 2010. TCRT 2010; 9(5):433-452.15. Jung I H, Song S Y, Jung J, Cho B, Kwak J, Je H U, Choi W, Jung N H, Kim S S, Choi E K, Clinical outcome of fi ducial-less CyberKnife radiosurgery for stage I non-

small cell lung cancer. Radiat Oncol J 2015; 33(2):89-97.16. Bahig H, Campeau M P, Vu T, Doucet R, Béliveau Nadeau D, Fortin B, Roberge D, Lambert L, Carrier J F, Filion E, Predictive Parameters of CyberKnife Fiducial-less

(XSight Lung) Applicability for Treatment of Early Non-Small Cell Lung Cancer: A Single-Center Experience. Int J Radiat Oncol Biol Phys 2013; 87(3):583-9.17. Bibault J-E, Prevost B, Dansin E, Mirabel X, Lacornerie T, Lartigau E, Image-Guided Robotic Stereotactic Radiation Therapy with Fiducial-Free Tumor Tracking for

Lung Cancer. Radiation Oncology 2012; 7:102.18. Drexler C, Furweger C, Quality assurance of a robotic, image guided radiosurgery system. WC 2009, IFMBE Proceedings 25/I, 492-5.19. Antypas C, Pantelis E, Performance evaluation of a CyberKnife G4 image-guided robotic stereotactic radiosurgery system. Phys Med Biol 2008; 53:4697-4718.20. Hoogeman M S, Prevost J B, Nuyttens J, Poll J, Levendag P, Heijmen B, Clinical accuracy of the respiratory tumor tracking system of the CyberKnife: assessment

by analysis of log fi les. Int J Radiat Oncol Biol Phys 2009; 74(1): 297-303.21. Nioutsikou E, Seppenwoodle Y, Symonds-Tayler J R, Heijmen B, Evans P, Dosimetric investigation of lung tumor motion compensation with a robotic respiratory

tracking system: an experimental study. Med Phys 2008; 35(4): 1232-40.22. Seppenwoodle Y, Berbeco R I, Nishioka S, Shirato H, Heijmen B, Accuracy of tumor motion compensation algorithm from a robotic respiratory tracking system:

A simulation study. Med Phys 2007, 34(7): 2774-84.

Page 8: FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE … · 2019. 4. 24. · FEASIBILITY OF REAL-TIME MOTION MANAGEMENT ON THE RADIXACT™ SYSTEM Abstract In this study, we adapted the

Important Safety Information: Most side effects of radiotherapy, including radiotherapy delivered with Accuray systems, are mild and temporary, often involving fatigue, nausea, and skin irritation. Side effects can be severe, however, leading to pain, alterations in normal body functions (for example, urinary or salivary function), deterioration of quality of life, permanent injury, and even death. Side effects can occur during or shortly after radiation treatment or in the months and years following radiation. The nature and severity of side effects depend on many factors, including the size and location of the treated tumor, the treatment technique (for example, the radiation dose), and the patient’s general medical condition, to name a few. For more details about the side effects of your radiation therapy, and to see if treatment with an Accuray product is right for you, ask your doctor.

© 2017 Accuray Incorporated. All Rights Reserved. Accuray, the stylized Accuray logo, TomoTherapy, H Series, Tomo, TomoH, TomoHD, TomoHDA, TomoEDGE, TomoHelical, TomoDirect, Hi Art, PlanTouch, PreciseART, PreciseRTX, Radixact, Accuray Precision and iDMS are trademarks or registered trademarks of Accuray Incorporated, in the United States and other countries and may not be used or distributed without written authorization from Accuray Incorporated. Use of Accuray Incorporated’s trademarks requires written authorization from Accuray Incorporated. Other trademarks used and identified herein are the property of their respective owners. MKT-RX-0417-0063

UNITED STATES

Accuray Corporate Headquarters 1310 Chesapeake TerraceSunnyvale, CA 94089 USATel: +1.408.716.4600Toll Free: 1.888.522.3740Fax: +1.408.716.4601Email: [email protected]

Accuray Incorporated1240 Deming WayMadison, WI 53717 USATel: +1.608.824.2800Fax: +1.608.824.2996

ASIA

Accuray Japan K.K.Shin Otemachi Building 7F2-2-1 Otemachi, Chiyoda-kuTokyo 100-0004JapanTel: +81.3.6265.1526Fax: +81.3.3272.6166

Accuray Asia Ltd.Suites 1702 - 1704, Tower 6The Gateway, Harbour City9 Canton Road, T.S.T. Hong KongTel: +852.2247.8688Fax: +852.2175.5799

Accuray Accelerator Technology #39 Huatai Road, Longtan Industrial Zone,Section 2 East, 3rd Ring Road, Chengdu, Sichuan 610051People's Republic of China

EUROPE

Accuray International SarlRoute de la Longeraie 9 (3rd floor)CH – 1110 MorgesSwitzerlandTel: +41.21.545.9500Fax: +41.21.545.9501