23
Now also available in preparative configuration : Faster More sensitive CIEF analyses

Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

  • Upload
    others

  • View
    8

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Now also available in

preparative configuration

: Faster More sensitive

CIEF analyses

Page 2: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Topics

Conventional CIEF vs iCIEF

CEInfinite – novelties:

1. Improved sensitivity

2. Higher sample throughput

3. Preparative iCIEF

Page 3: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Applications

What?

Charge-based characterisation of proteins, pI range 2.5-11

VLP, mAb, ADC, fusion proteins, peptides, …

Lot release & stability analysis

Structural & degradation pathway characterisation

Kinetic & dynamic studies

For whom?

QA/QC, analytical & process development

www.isogen-

Page 4: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Slide courtesy of Agilent Technologies

Chemical Mobilization

Conventional CIEF

Page 5: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

iCIEF Whole-Column

Detection

Separation capillary

Sample transfer capillary

+HV -

CMOS

Membrane or frit

Light Source (Laser LED)

High Resolution Lens

Separation capillary: 5 cm

Page 6: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

iCIEF run of

hemoglobin

Focusing current

0.5 min

1 min

1.5 min 2 min 2.5 min

0.5 min

1 min

1.5 min

2 min

2.5 min

3 min

3.5 min

4 min

5 min

6 min

Sample: Hemoglobin A, F, S, C

8% Pharmalyte 3-10

600 V/cm

Page 7: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

www.isogen-

Page 8: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Quality Control

www.isogen-

All consumables delivered

with certificate

Cartridges

Carrier ampholytes

Electrolytes

Standards

Page 9: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Whole-Column

Detection - iCIEF

Quantitative protein analysis

Peak patterns: reproducible

No hydrostatic flow

Wide pI range (pH 2.5-11)

High resolution - ∆pI > 0.02

Short analysis: > 5 runs/hr

Method development: hr - days

Quantitative protein analysis

Low resolution - mobilisation

Long: +/- 1 run/hr

Method development: months

Lower reproducibility

Requires additives

Advantages of iCIEF Conventional CIEF

Page 10: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Practical iCIEF

www.isogen-

Method development

Page 11: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

1. Improved sensitivity

The cartridge diameter

www.isogen-

Page 12: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

High-sensitivity

cartridge

200 μm

100 μm

HR3-10, Hemoglobin standardDB coated cartridge

Page 13: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

High-sensitivity

cartridge

200 μm

100 μm

HR3-10 : HR6-7.5 (1:3)Fusion proteinFC coated cartridge

Page 14: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

2. Increased throughput

The cartridge coating

www.isogen-

Page 15: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Higher sample

throughput

Limitations of traditional cartridges:

FC-coating need methylcellulose viscosity ↑

Solution: other coatings

PA or DB: no methylcellulose needed! viscosity ↓

Improved pipetting reproducibility

Increased injection speed

Reduction in separation time CIEF

From 5 runs per hour to 10 runs per hour

www.isogen-

Page 16: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

3. Preparative iCIEF

The CEInfinite

www.isogen-

Page 17: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

www.isogen-

Prep-iCIEF

Recuperate fractions with the software-controlled pump and

then inject in your MS or other device for more

Page 18: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Prep-iCIEF

-+

Focused proteinsTransfer capillary

Separation capillarySyringe pump

Changing diameter with factor 4 increases length peak

factor 16 in outlet capillary better time control

Page 19: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Prep-iCIEF

A F S C C

S

F

A

High voltage during mobilization. Preserve sharp peaks

Page 20: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Prep-iCIEF

pI 7.9pI 6.61

A

F

S

CC

S

F

A

Page 21: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Prep-iCIEF - MS

Page 22: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

Reproducibility

www.isogen-

Page 23: Faster More sensitive CIEF analyses courtesy of Agilent Technologies Chemical Mobilization Conventional CIEF. iCIEF Whole-Column Detection Separation capillary Sample transfer capillary

For further information

on iCIEF in Europe

Questions?

Isogen Life Science

Tel: +31 30 688 07 71

E-mail: [email protected]