Upload
others
View
6
Download
0
Embed Size (px)
Citation preview
1
FACULTY OF PHARMACEUTICAL SCIENCES
INSTITUTE FOR PHARMACEUTICAL TECHNOLOGY AND BIOPHARMACY
UNIVERSITY OF DUESSELDORF
Academic year 2008-2009
TOPEM, a new temperature modulated DSC technique: A critical review
Laurine PARMENTIER First Master of Pharmacy, Pharmaceutical Care
Promotor Dr. M. Thommes
Co-promotor Prof. Dr. C. Vervaet
Commissioners Prof. Dr. W. Baeyens
Dr. K. Remaut
2
“De auteur en de promotor geven de toelating deze masterproef voor consultatie beschikbaar te stellen en delen ervan te kopiëren voor persoonlijk gebruik. Elk ander gebruik valt onder de beperkingen van het auteursrecht, in het bijzonder met betrekking tot de verplichting uitdrukkelijk de bron te vermelden bij het aanhalen van de resultaten uit deze masterproef.”
28/05/2009
Promotor Auteur Prof. Dr. C. Vervaet Laurine Parmentier
3
Word of Thanks
I would like to thank Prof Dr. C. Vervaet and Prof. Dr. P. Kleinebudde
for generally leading my master thesis.
I would like to express my warmest thanks to Dr. M. Thommes for helping me throughout this
study, especially for his patience and great help conducting me into the right direction.
I would also like to thank the whole staff of the institute for the pleasant environment
and especially Ms. K. Matthée who stood by me from the first day.
I would like to thank all the friends,
especially F. Koppers for always being ready to help.
Further, I would like to thank my parents for their eternal support.
Finally, I would like to thank the Erasmus programm,
which made this whole experience possible.
4
LIST OF CONTENTS
WORD OF THANKS
LIST OF CONTENTS
LIST OF ABBREVIATIONS
1 INTRODUCTION ................................................................................................. 1
1.1 Thermal analysis ........................................................................................... 1
1.1.1 In general ................................................................................................ 1
1.2 Theory ........................................................................................................... 2
1.2.1 First law of thermodynamics ................................................................... 2
1.2.2 Second law of thermodynamics .............................................................. 3
1.3 Thermal analysis methods............................................................................. 4
1.3.1 Differential thermal analysis .................................................................... 4
1.3.2 Differential scanning calorimetry ............................................................. 4
1.3.2.1 General principles................................................................................ 4
1.3.2.2 Power compensation DSC................................................................... 5
1.3.2.3 Heat flux DSC...................................................................................... 6
1.3.3 Temperature modulated differential scanning calorimetry ...................... 7
1.3.3.1 General principles................................................................................ 7
1.3.3.2 Isothermal step method differential scanning calorimetry .................... 8
1.3.3.3 Alternating differential scanning calorimetry ........................................ 9
1.3.3.4 TOPEM.............................................................................................. 10
2 MATERIALS AND METHODS ........................................................................... 16
2.1 Materials...................................................................................................... 16
2.2 Methods....................................................................................................... 17
2.2.1 Differential scanning calorimetry ........................................................... 17
2.2.2 Balance................................................................................................. 17
5
2.2.3 α-mannitol ............................................................................................. 17
2.2.4 Water uptake of sugars......................................................................... 17
3 OBJECTIVES..................................................................................................... 18
4 RESULTS .......................................................................................................... 19
4.1 Preliminary tests .......................................................................................... 19
4.2 Reversing and non-reversing heat flow ....................................................... 20
4.2.1 In general .............................................................................................. 20
4.2.2 Griseofulvin ........................................................................................... 20
4.2.3 Sucrose-water mixture .......................................................................... 22
4.2.4 Moist substances .................................................................................. 23
4.3 Calculation window...................................................................................... 25
4.3.1 In general .............................................................................................. 25
4.3.2 Mannitol ................................................................................................ 26
4.3.3 Calculation window parameters ............................................................ 27
4.3.4 Effect of the pulse width on the calculation window .............................. 30
4.4 Heat capacity............................................................................................... 37
4.5 Frequency ................................................................................................... 38
4.6 Resolution ................................................................................................... 41
5 CONCLUSIONS................................................................................................. 42
6 LITERATURE LIST ............................................................................................ 43
6
LIST OF ABBREVIATIONS
ADSC Alternating Differential Scanning Calorimetry
Cp Heat capacity
Cp* Complex heat capacity
DMA Dynamic Mechanical Analysis
DSC Differential Scanning Calorimetry
DTA Differential Thermal Analysis
HR Heating rate
ISMDSC Isothermal Step Method Differential Scanning Calorimetry
m Sample mass
p Period
PET Polyethylene terephtalate
PH Pulse height
PVP Polyvinylpyrolidone
TGA Thermogravimetric Analysis
TMA Thermal Mechanical Analysis
TMDSC Temperature Modulated Differential Scanning Calorimetry
TOA Thermooptical Analysis
USP United States Pharmacopeia
1
1 INTRODUCTION
1.1 THERMAL ANALYSIS 1.1.1 IN GENERAL
In pharmaceutical research, the development of new medicines is the main
goal. Besides that, already existing products need to be evaluated. There are several
ways to determine the characteristics of drugs during and after the development.
Information is needed about several characteristics such as purity, interactions,
solubility, molecular weight, structure and melting point. All these physicochemical
properties can be determined with a variety of methods and devices. A few examples
are chromatographic methods, spectroscopy, titration and thermal analysis.
In thermal analysis, the behavior of a substance is analyzed when a certain
temperature program is applied. Thermal analysis is a group of techniques in which a
physical property is measured as a function of temperature. Therefore the substance
is subjected to a controlled temperature program. This can be a dynamic or an
isothermal program. The obtained curve is the measured characteristic in function of
the time or the applied temperature. (Schwarz & de Buhr, 1998)
Thermal analysis includes several techniques of which the most important are:
differential scanning calorimetry (DSC), differential thermal analysis (DTA)
thermogravimetric analysis (TGA), thermal mechanical analysis (TMA), dynamic
mechanical analysis (DMA) and thermooptical analysis (TOA). TOA is the visual
perception or measurement of the permeability of light or light reflection of a sample.
In DMA, a force is supplied to the sample whereby the sample deforms and
information about the stiffness (viscosity and elasticity) can be determined. In TMA,
the deformation (change in width or length) of the sample under a load is measured.
In TGA, the mass of the sample is measured, under a defined atmosphere. DTA
measures the temperature difference between a sample and a reference. DSC
contains several techniques which will be discussed later. All the techniques for
thermal analysis are measured with an underlying temperature program. (Schwarz &
de Buhr, 1998)
It is the latest new DSC method, TOPEM, which will be the subject of the
research in this study.
2
With thermal analysis methods, several characteristics can be measured:
melting point, crystallization behavior, glass transition, expansion, mass loss, thermal
stability, decomposition temperature, oxidation time, network formation, purity, visco-
elastic behavior and chemical reactions. (Schwarz & de Buhr, 1998)
1.2 THEORY 1.2.1 FIRST LAW OF THERMODYNAMICS
The first law of thermodynamics is an expression of the law of conservation of
energy. No energy can be destroyed or created, only changes in energy can take
place. The internal energy is the sum of heat and work done by the system. (Van der
Plaats, 1992)
U = Q + W (1)
Where: U: internal energy (J) Q: heat (J) W: work (J)
A calorimeter for example is a system in which mass and pressure are
normally constant. This means that the energy is only depending on temperature and
volume.
U = U (T, V) (2)
Where: U: internal energy (J) T: temperature (°C)
V: volume (m3)
A new state function is introduced, namely the enthalpy. State functions are
functions which always give the same result when going from point A to B,
independent from the way which is taken to get there. The change of enthalpy equals
the heat under constant pressure. Further is the work given by the pressure and
volume. The minus sign means that the work is done by the system. This means that
equation 1 can be rewritten using equations 3 and 4. (Van der Plaats, 1992)
(∆H)p = Q (3)
W = -pV (4)
(dH)p = dU + pdV (5)
3
Where: U: internal energy (J) H: enthalpy (J)
p: pressure (Pa) V: volume (m3)
Q: heat (J) W: work (J)
Under constant pressure, the heat given to the system is the same as the
increase in enthalpy. Therefore, the heat capacity of a sample under constant
pressure is the change of enthalpy in time. (Van der Plaats, 1992)
cp = p (6)
Where: cp: heat capacity (J/K) H: enthalpy (J)
T: temperature (K) p: pressure (Pa)
1.2.2 SECOND LAW OF THERMODYNAMICS
There are different ways to determine the second law of thermodynamics:
• It’s impossible to transfer heat from a cold to a warm place, without changing
the environment
• It’s impossible to transform heat completely without changing the environment
• A new state function is introduced, namely the entropy. The entropy is heat
divided by temperature, in a reversible process. (Van der Plaats, 1992)
S = (7)
Where: S: entropy (J/K) Q: Heat (J)
T: temperature (K)
Using equation 4 and 7, equation 1 can be rewritten as follows: (Van der
Plaats, 1992)
dU = TdS – pdV (8)
Where: U: internal energy (J) T: temperature (K)
S: entropy (J/K) p: pressure (Pa)
V: volume (m3)
4
Further is also the Gibbs free energy defined as G = H – TS. A system will
always try to find the lowest free energy state. According to the equation, this means
that the enthalpy has to be minimal and entropy has to be maximal to achieve this.
The Gibbs free energy is minimal when the system is at equilibrium under constant
pressure and temperature. However, pressure and temperature can be influenced in
thermal analysis. Therefore, the Gibbs free energy cannot reach the mininmum.
(Van der Plaats,1992)
G = H – TS (9)
Where: G: Gibbs energy (J) H: enthalpy (J)
T: temperature (K) S: entropy (J/K)
1.3 THERMAL ANALYSIS METHODS 1.3.1 DIFFERENTIAL THERMAL ANALYSIS
DTA is the oldest technique in thermal analysis. In DTA, a sample and
reference undergo the same temperature program. The sample and reference are
put in a pan. As a reference, mostly an empty pan is used. When a reaction occurs in
the sample, energy is needed (for an endothermal reaction) or released (for an
exothermal reaction) and therefore a temperature difference is measured. The
temperature difference is measured between the sample and the reference by two
thermal elements, one for each. When a temperature difference is measured, the
oven will compensate this, so that both sample and reference have the same
temperature. The temperature difference is used as a measured value. Any change
in temperature is detected with DTA. (Widman & Riesen, 1984)
1.3.2 DIFFERENTIAL SCANNING CALORIMETRY
1.3.2.1 General principles
As in DTA, in DSC a sample and reference, put into a pan, undergo a dynamic
or isothermal temperature program. Both sample and reference are placed on
identical measuring cells next to each other in an oven. The temperature difference is
compensated by the measurement cells and not by the oven. Because the sample
and reference have their own measuring cell, the detection of the temperature
difference goes faster and with more sensitivity as in DTA. The output value is the
5
heat flow. This heat flow is actually measured as a heat power, with the unit Watt. By
integration of the heat flow in time, the enthalpy of a sample is calculated, with the
unit Joule. (Widman & Riesen, 1984)
1.3.2.2 Power compensation DSC
The sample and the reference both have their own measuring cell and oven.
Both ovens separately follow a temperature program. When there is a thermal
symmetry between both systems, the same amount of power will be needed to heat
them. When a transition (endothermal or exothermal) occurs, this will lead to a
temperature difference. A proportioning controller will be activated and puts as much
heat in the sample or cools it down as needed for the transition. That’s why the
temperature difference is kept a constant. (Widman & Riesen, 1984)
The compensated power is proportional to the measured temperature
difference.
∆P = -K1∆T (10)
Where: P: power (W) T: temperature (K)
K: calibration factor, fixed value of the controller
When a good calibration is used, the temperature difference can be used to
measure the heat flow in the sample.
ΦS = -K2∆T (11)
Where: Φ: heat flow (W) T: temperature (K)
K: calibration factor
FIGURE 1.1. CROSS-SECTION OF THE OVENS USED IN POWER COMPENSATION DSC.
(HTTP://WWW.FCHPT.STUBA.SK)
6
1.3.2.3 Heat flux DSC
In heat flux DSC, the sample and reference are placed in one oven. The heat
flows from the oven, over a defined heat resistance to the sample and reference. The
driving force for the heat flow is the temperature difference over the heat resistance.
The heat flow to the sample is different to the one to the reference. (Widman &
Riesen, 1984)
When heat flows under steady state conditions, the temperature difference
between the reference and sample is proportional to the difference in heat flow from
the oven to the sample (OS) and to the reference (OR).
∆Φ = ΦOS – ΦOR = -K∆T = ΦS (12)
Where: Φ: heat flow (W) T: temperature (K)
O: oven R: reference S: sample
K: calibration factor that includes the asymmetry of the system
By measuring the temperature difference between a monster and reference, using
this equation, the heat flow ΦS can be measured.
A B
FIGURE 1.2. A) CROSS-SECTION OF THE OVEN USED IN HEAT FLUX DSC.
B) TOP-VIEW OF THE SENSOR USED IN HEAT FLUX DSC. S = SAMPLE,
R = REFERENCE (HTTP://WWW.MT.COM)
7
1.3.3 TEMPERATURE MODULATED DIFFERENTIAL SCANNING CALORIMETRY
1.3.3.1 General principles
In conventional DSC, the sample undergoes a linear temperature program.
The output value is the difference in heat flow between the sample and reference. In
TMDSC however, a temperature modulation is superimposed over the linear
temperature program. This allows us to separate the heat flow into two components,
the reversing and non-reversing heat flow. The separating of these can be useful
when overlaying effects occur or thermal events cannot be clearly identified with
classical DSC. (Cao, 2007)
The reversing heat flow is the sensible heat flow component. This is
associated with processes that can be reversed. Thermodynamically seen, these are
processes that occur close to a local metastable state. When a process is reversible,
the measured curve should be reproducible after cooling the sample down and
reheating it. For such processes the heat input results in a temperature change.
Some examples are: glass transitions, melting of polymers, temperature change
when no thermal transition occurs. (Schawe, 2005a)
The non-reversing heat flow is the latent heat flow component. It is associated
with irreversible processes. Thermodynamically seen, such processes start in a non-
equilibrium state and end in an equilibrium state. They are time dependent. The heat
input doesn’t change the temperature, because the heat is used for changes in the
sample. Such changes take all the energy of the heat and therefore no heat is left for
a temperature increase. Examples are: non-reversible chemical reactions, melting,
and crystallization. (Schawe, 2005a)
For measuring the change of heat capacity with time or temperature,
temperature changes of the sample are measured. Due to that temperature
difference, there’s a change in enthalpy, given by
∆H = m cp ∆T (13)
Where : H: enthalpy (J) m: sample mass (mg)
cp: heat capacity (J/K) T: temperature (K)
8
Next to the heat flow as a result from the heat capacity, there are other energy
changes due to chemical reaction or physical transitions. Therefore we should add
the heat related to chemical reactions, related to physical transitions and the drift of
the measuring system to equation 13. With respect to time, we can also split the heat
flow in the sensible heat flow and latent heat flow. (Riesen 1994)
Φ = m cp β Φb + Φr + Φt (14)
Φs Φl
Where: Φ: heat flow (W) Φr: heat flow of chemical reaction (W)
m: sample mass (mg) Φt: heat flow of physical transition (W)
cp: heat capacity (J/K) Φb: drift of the measurement (W)
β: heating rate (K/min) Φs:sensible or reversing heat flow (W)
Φl:latent or non-reversing heat flow(W)
In conventional DSC the temperature program is given by T = T0 + ß t. In
temperature modulated DSC, a temperature modulation is applied to a programmed
underlying heating rate and therefore a time-dependent function is added to the last
equation. This function is a saw tooth (ISM DSC), a sine wave (ADSC) or a
stochastic modulation (TOPEM).
T = T0 + ß t + f(t) (15)
Where: T: temperature ß: heating rate
T0: start temperature t: time
f(t): function that describes the temperature modulation
1.3.3.2 Isothermal step method differential scanning calorimetry
In ISM DSC, the temperature program consists of alternating heating and
isothermal segments. The isothermal segment should last until the heat flow remains
constant. If the temperature steps are too small, it is possible that the sample cannot
follow. When the steps are big enough, quasi-static conditions can be achieved. This
means that the sample can follow the set oven temperature and has time to
equilibrate at the isothermal parts. When the sample can follow well, there is a good
9
separation of reversing and non-reversing heat flow. If the sample doesn’t have the
time to equilibrate, this separation cannot be completed. (Schawe, 2005b)
FIGURE 1.3. TEMPERATURE PROFILE OF AN ISM DSC MEASUREMENT
(METTLER-TOLEDO 2009)
1.3.3.3 Alternating differential scanning calorimetry
The temperature program in ADSC is a periodic sinusoidal change. Several
frequencies can be chose, in a range from 0.001 to 0.1 Hz. The disadvantage
however is that for every new selected frequency, a new measurement has to be
done. When low pulse heights and low underlying heating rates are used, there is an
increase in resolution. (Schawe, 2005b; Jörimann et al. 1999)
T(t) = T0 + βt + ATsin(ωt) (16)
ω = 2π/p (17)
Where: T: temperature (K) t: time (s)
ω: angular frequency (rad/s) p: period (s)
T0: start temperature (K) β: heating rate (K/min)
AT: amplitude of temperature modulation (K)
FIGURE 1.4. SINUSOIDAL HEATING RATE OF AN ADSC MEASUREMENT
(METTLER-TOLEDO 2009)
10
From the raw data, the total heat flow is calculated, by averaging the signal.
This is the same approach as in conventional DSC measurements. The measured
heat flow is also a sine wave, but has a phase shift or phase lag compared to the
applied heating rate. (Jörimann, 1999)
Φ = AΦ cos(ωt + φ) (18)
Where: Φ: heat flow (W) AΦ: amplitude of the heat flow (K)
t: time (s) ω: angular frequency (rad/s)
φ: phase (rad/s)
For a certain frequency f, the heat capacity cp(f) can be calculated from
periodic changes in heat flow. Further on, the reversing heat flow and heat capacity
can be measured. The non-reversing heat flow is the difference between the total
heat flow and reversing heat flow. (Jörimann, 1999)
Φrev = β cp (19)
cp = AΦ (20)
Φrev = β AΦ (21)
Φnon = Φtot - Φrev (22)
Where: β: heating rate (K/min) cp: heat capacity (J/K)
p: period (s) AΦ: amplitude of heat flow (K)
AT: amplitude of temperature modulation (K)
Φtot: total heat flow (W) Φrev: reversing heat flow (W)
Φnon: non-reversing heat flow (W)
1.3.3.4 TOPEM
In general
In the TOPEM method, the temperature program is a stochastically changing
pulse. This means that the pulse widths change at random between a chosen
minimum and maximum. Also the pulse height has to be chosen. Figure 1.5. shows
an example of how the pulse can look like.
11
FIGURE 1.5. EXAMPLE OF POSSIBLE PULSES IN TOPEM. THE VERTICAL ARROW IS
THE PULSE HEIGHT. THE HORIZONTAL ARROWS ARE TWO POSSIBLE
PULSE WIDTHS. (METTLER-TOLEDO 2009)
The stochastic pulses contain many different frequencies. After the
measurement, the heat capacity is calculated. This curve is then used for the
frequency evaluation. For a certain frequency, out of the heat capacity curve a new
heat capacity curve is calculated, which is called the complex heat capacity. Several
frequencies can be chosen, ideal for analyzing frequency dependent effects. The
heat capacity is determined under quasi-static conditions. The calculated heat flow is
the sum of the sensible and latent heat flow. (Schawe, 2005b)
The chosen temperature program defines the input signal. The output signal is
the measured heat flow. A mathematical procedure calculates the correlation
between the input (heating rate) and output signal (heat flow). After this procedure, a
component is obtained related to the input signal and one that doesn’t. The
correlated one is the reversing heat flow, the non-correlated one is the non-reversing
heat flow. (Schawe et al., 2005e)
The temperature program is defined by the underlying heating rate, switching
time and pulse height. The heating rate cannot be too high so that enough data
points are obtained. The switching time of the pulse width is set between 15 and 30
seconds. For low frequency measurements, a high switching time is needed and vice
versa. Too low or too high switching times result in noisy results. The pulse height is
determined by the behavior of the sample. The sample has to be able to follow the
oven. A sample is able to follow a little pulse height better than a high pulse height
12
because it needs less time to equilibrate. On the other hand, more information is
gained with higher pulse heights because the sample is heated for a longer period.
This means that a compromise has to be found between higher and lower pulse
heights.
FIGURE 1.6. TYPICAL TEMPERATURE PROFILE (LEFT) AND CORRESPONDING HEAT
FLOW (RIGHT) IN A TOPEM MEASUREMENT (SCHAWE, 2005B)
Advantages and disadvantages of TOPEM
The new TMDSC method has many claimed benefits:
- In one measurement, there is a simultaneous determination of sample
properties as a function of time and temperature over a wide frequency range.
- Very accurate determination of the quasi-static specific heat capacity by
calculation of this heat capacity from the pulse response.
- High sensitivity and high resolution allows the measurement of low energy
transitions or effect close to another.
- Separation of sensible and latent heat flow. Heat capacities can be determined
even if the effects overlap.
- Simple interpretation of the curves. Effects depending on frequency can easily
be distinguished from those independent of frequency.
- The dynamics of the system can be analyzed over a broad range of
frequencies in one single measurement. (Schawe 2005b)
13
The disadvantages of TOPEM are:
- Melting of pure substances cannot be measured. During the melting, the
sample temperature doesn’t change and the sample can therefore not follow a
temperature modulation. (Schawe 2005b)
- For the evaluation, a calculation window has to be set. The recommended
width of that calculation window by Mettler-Toledo is one third of the transition
interval. This is however not strict enough. It should be less than about one
tenth of the transition interval. (Fraga et al. 2007)
- The limit frequency for TOPEM is 4 mHz, independent on the experimental or
calculation parameters. This means it is intrinsic to the TOPEM method.
(Fraga et al. 2007)
- The measured complex specific heat capacities at a selected frequency shift
to higher temperatures for higher sample mass. (Fraga et al. 2007)
- In comparison to ADSC, the measured temperature of a transition is
significantly larger. This is because both use a different temperature
modulation. (Fraga et al. 2007)
Theory
For the measured heat flow both the reversing and non-reversing heat flow
can be calculated. The total heat flow is the sum of the reversing and non-reversing
heat flow according to the equation
Φtot = Φrev + Φnon (23)
Φrev = m ß (24)
Where: Φtot: total heat flow (W) m: sample mass (mg)
Φrev: reversing heat flow (W) : quasi-static heat capacity
Φnon: non-reversing heat flow (W) ß: heating rate (K/min)
If the temperature modulation is sufficiently small, it is assumed that the
current state of the sample is almost unaffected and that it is in equilibrium.
Therefore, in a limited temperature range the sample can be described as a linear
system. (Schawe, 2005d)
14
u(t) g(t) y(t)
u(t) = T(t) linear system y(t) = Φmeasured (t)
Temperature as a function of
time
The system is defined
by the sample and
instrument
Heat flow as a function of
time.
FIGURE 1.7. GENERAL PRINCIPLE OF THE TOPEM TECHNIQUE
For a linear time dependent system with an input signal u(t) and output signal
y(t), the correlation is given as: (TOPEM – The new advanced multi-frequency
TMDSC technique)
y(t) = g(t) u(t) (25)
Where: y(t): output signal g(t): pulse response of the system
u(t): input signal t: time (s)
The pulse response is calculated with a z-transformation. Generally, a
transformation is the description of the signal with another set of basic units as was
the case before. The z-transformation is the discrete analog of the Laplace
transformation. It is used to convert differential equations which come up for time
discrete processes. The time-domain signal is converted by the z-transformation into
a complex frequency-domain representation. Equation 25 can be rewritten in the z-
plane as
y(z) = H(z)u(z) (26)
Equations of the time-domain can be described and solved more easily in the
z-plane. Several functions can be used to solve H(z), but often it can be exactly
described with a rational function.
(27)
Where B(z) and A(z) are polynomials of degree q or p in the variable z.
15
By using the last equation, y(z) can be rewritten as
y(z) = u(z) (28)
or A(z)y(z) = B(z)u(z) (29)
When the last equation is converted to the time domain, the unknown
parameters for A(z) and B(z) have to be defined. These parameters are determined
using the measured input and output quantities. If these parameters are known, the
values of the pulse response g(t) for a certain frequency can easily be calculated.
Once we get the pulse response, the quasi-static heat capacity is calculated from this
pulse response. (TOPEM – The new advanced multi-frequency TMDSC technique)
(30)
Where: m: sample mass (mg) : quasi-static heat capacity
g(t): pulse response t: time (s)
16
2 MATERIALS AND METHODS
2.1 MATERIALS
TABLE 3.1. LIST OF ALL USED CHEMICALS
Products Company City Country
Griseofulvin (USP
micronized)
Hawkins
Pharmaceutical
Group
Minneapolis USA
Mannitol (Pearlitol
60) Roquette Lestrem France
Indium Mettler-Toledo Gießen Germany
PET BASF Ludwigshafen Germany
Sucrose Pfeifer & Langen Cologne Germany
Glucose Roquette Lestrem Germany
Dextrose
(monohydrate ST) Roquette Lestrem France
Lactose (granulac
200) Meggleburg Wasserburg Germany
Sodium nitrate Grüssing Filsum Germany
PVP CL BASF Ludwigshafen Germany
PVP CL-M BASF Ludwigshafen Germany
PVP 17PF BASF Ludwigshafen Germany
PVP 25 BASF Ludwigshafen Germany
PVP 90F BASF Ludwigshafen Germany
PVP IR BASF Ludwigshafen Germany
PEG PVA
copolymer VA64 BASF Ludwigshafen Germany
17
2.2 METHODS 2.2.1 DIFFERENTIAL SCANNING CALORIMETRY
All experiments are done with the DSC1 from Mettler-Toledo (Gießen, Germany)
equipped with sample robot and a Haake Intra-Cooler EK/MT. Sample parameters
and data-analysis are realized with the STARe software. This software also
determines if the DSC is used as a conventional DSC, ADSC or as TOPEM. All
experiments were done using Aluminium pans without pin of 40μl (Mettler-Toledo,
Gießen, Germany). Al of the pans had pierced lids unless told else. All
measurements are done under N2 with a flow of 50ml/min. The used method will be
shown in the diagrams, when provided or else in the text. The used abbreviations are
HR for heating rate, PH for pulse height and p for period.
2.2.2 BALANCE
All samples are weighed on the balance MC210P from Sartorius AG
(Goethingen, Germany). First, sample pan holder, pan and lid are set to zero before
weighing the sample. After sealing the pan with the lid, the sample is weighed again
and this value is used.
2.2.3 α-MANNITOL
In order to make α-mannitol, ß-mannitol is heated until it is completely molten.
Then it’s poured into a mortar. When it’s completely crystallized, it is pulverized into
powder and stored at room temperature.
2.2.4 WATER UPTAKE OF SUGARS
For some measurements, moist substances were needed. Glucose, fructose,
dextrose and sucrose were heated until completely molten and poured into a petri
dish. The glassy substances were equilibrated at 23°C at 60% relative humidity for at
least 10 hours.
18
3 OBJECTIVES
In thermal analysis, differential scanning calorimetry (DSC) is a commonly
used technique. Later, temperature modulated DSC techniques made their entrance,
namely isothermal step DSC and alternating DSC. Recently, TOPEM, a new
temperature modulated DSC technique has been developed. However, not much
research has been done on this technique. In addition, most available information is
linked to the company which developed TOPEM. The goal of this study is to make a
critical review about this new technique. TOPEM will be compared to regular DSC
and ADSC in order to evaluate its possible advantages and drawbacks.
The main question is: does TOPEM has any added value for differential
scanning calorimetry? In the literature, some advantages and disadvantages can be
found (see also introduction p.13) and based on the described advantages and
disadvantages, the following points should be investigated:
- separation of reversing and non-reversing heat flow: measurements of several
examples
- calculation window: determination of a width of calculation window and the
influence of the pulse width on it
- heat capacity: comparison of several methods to calculate heat capacity and
evaluation of the shift in complex heat capacity
- frequency: comparison with ADSC, determination of its limits
- resolution
19
4 RESULTS
4.1 PRELIMINARY TESTS To start with, there was a search for a substance to do the investigations on
TOPEM with, because no standard substances are described in the literature. As a
first substance, Indium was measured because it is commonly used as a standard for
regular DSC. The melting peak with high intensity was found at the same
temperature as DSC. No conclusions could be drawn from the reversing and non-
reversing heat flow curves because of the big intensity of the melting over small
temperature range. In addition, pure substances such as Indium cannot follow a
temperature modulation. (Schawe 2005b)
Next step was to take substances described in the literature and do
measurements with the same parameters. Own measurements were compared with
measurements described in literature. The goal was to work on with a certain
substance when comparable results were found. A lactose-water mixture containing
4,5% lactose was measured as described by Schawe (2005a). The article shows a
peak at -30°C of which the origin is not clear. There was no transition found, even
after several measurements. However, the described peak in the literature is small. It
is possible that the DSC used by Schawe (2005a) is able to detect smaller effects
because it is more sensitive.
Afterwards, polyethelene terephtalate was measured as described by Schawe
(2005c). Pet was crystallized for 10 min at 170°C. Compared to the article, smaller
curves are obtained and there is noise even after repeating the measurement. The
reason for this is that polymers often have a batch-to-batch variability. Therefore, the
next measured substance was a crystalline substance, namely sodium nitrate as
described by Schawe (2005a). The measurement showed comparable curves to the
literature with a step in the curves at the same temperature (see figure 4.1.).
From these tests was concluded to do no measuring of melting of pure
substances. Pure substances can be used when effects overly or are close to each
other. Because polymers often have differences between batches, it is better to take
crystalline substances. Also mixtures will be used for investigation.
20
A B
FIGURE 4.1. A. TOTAL (UPPER), NON-REVERSING (MIDDLE) AND REVERSING (LOWER) HEAT
FLOW CURVES OF A TOPEM MEASUREMENT OF SODIUM NITRATE (SCHAWE,
2005C)
B. TOTAL (BLACK), REVERSING (GREEN) AND NON-REVERSING (BLUE) HEAT
FLOW CURVE OF A TOPEM MEASUREMENT OF SODIUM NITRATE
4.2 REVERSING AND NON-REVERSING HEAT FLOW 4.2.1 IN GENERAL
Separation of reversing and non-reversing heat flow is the main goal of using
temperature modulated DSC. To evaluate this, first griseofulvin was measured
because it is expected that TOPEM can give more information about the results
obtained from regular DSC. Also a 40:60 mixture of sucrose and water was
measured and compared to the literature. There were also several moist sugar
samples measured. Moist substances give in regular DSC only a peak between 30°C
and 120°C because of the evaporation of water. This peak overlies glass transitions
occurring in this range. TOPEM should be able to detect this glass transition.
4.2.2 GRISEOFULVIN
The conventional DSC curve of griseofulvin shows a small peak before the
melting peak which could be due to the sublimation of griseofulvin. The clue for that
is the description of a weight loss due to sublimation at about 200°C, measured with
thermogravimetry. (Analytical profiles of drug substances and excipient)
21
Since sublimation is a process related to latent heat flow, it should be seen in
the non-reversing heat flow curve of a TOPEM measurement. As shown in figure
4.2.B the total heat flow curve also shows this second peak but the non-reversing
curve doesn’t. The reason for this could be that the gas phase escaped through the
hole in the pierced lid of the pan and therefore no effect can be measured. Another
measurement was done with a closed pan. As can be seen in figure 4.2.C the curves
are comparable to the curves of the measurement with a pierced pan. This means
that no extra information was gained using TOPEM.
The peak at about 18°C is due to melting. Griseofulvin is a pure substance so
its melting cannot be evaluated using TOPEM. Griseofulvin was investigated for the
small peak under the first part of melting peak. It was not used for the investigation of
the melting peak.
A B C
FIGURE 4.2. TOTAL HEAT FLOW CURVE OF A DSC MEASUREMENT OF GRISEOFULVIN
(A) AND TOTAL (BLACK), REVERSING (GREEN) AND NON-REVERSING
(BLUE) HEAT FLOW CURVES OF A TOPEM MEASUREMENT OF
GRISEOFULVIN WITH PIERCED (B) AND SEALED (C) AL PAN
22
4.2.3 SUCROSE-WATER MIXTURE
A TOPEM measurement of a mixture of 40% sucrose and 60% water was
measured as described by Schawe (2005c). The total heat flow curves are
comparable to our measurement. However, the curve for the non-reversing heat flow
is not entirely published. In our measurement, a peak in the non-reversing heat flow
curve at about -5°C can clearly be seen. The little peak around -35°C is explained in
the article to be the melting of non-equilibrated crystals. The peak at -5°C is
explained by Schawe (2005c) to be an artifact.
It is interesting that the peak at -5°C, which is more than ten times larger than
the peak at -35°C is not reported in the article. The peak at -5° is also present at all
the measurements which were taken. Therefore we assume that it is not an artifact
but that the peak is due to crystallization.
FIGURE 4.3. TOTAL (BLACK), REVERSING (GREEN) AND NON-REVERSING (BLUE)
HEAT FLOW CURVES OF A TOPEM MEASUREMENT OF A 40:60
SUCROSE-WATER MIXTURE
23
A B
FIGURE 4.4. A. TOTAL HEAT FLOW (BLACK) AND MEASURED HEAT FLOW (RED) OF A
TOPEM MEASUREMENT OF A 40:60 SUCROSE-WATER MIXTURE
(SCHAWE, 2005C)
B. TOTAL (BLACK), REVERSING (GREEN) AND NON-REVERSING (BLUE)
HEAT FLOW CURVES OF A TOPEM MEASUREMENT OF A 40:60
SUCROSE-WATER MIXTURE
(SCHAWE, 2005C)
4.2.4 MOIST SUBSTANCES
Moist substances show in regular DSC a peak between 30°C and 120°C
because of the evaporation of water. When some underlying effects occur at the
same range, such as a glass transition, this cannot be seen in the heat flow curve.
TOPEM however can separate the latent and sensible heat flow. The water peak will
still be seen in the total heat, but as a glass transition is a reversible process, it
should be seen in the reversing heat flow. To examine the effects underlying the
evaporation of water, several samples of polyvinylpyrolidone (PVP) were evaluated.
All the selected samples showed a large water peak in measurements obtained by
regular DSC in the past. All the PVP samples were measured between 0°C and
140°C with a heating rate of 1°C/min, a pulse height of 0.5°C and had a sample
weight of about 3.00 mg.
There was no underlying effect found for none of the PVP samples. Only the
reversing heat flow curve of the TOPEM measurement of PVP VA64 does show a
24
glass transition at about the glass transition temperature of the DSC measurement.
At that point however, it is not overwhelmed by a peak of evaporation of water. This
means that for all the PVP samples, TOPEM cannot separate between reversing and
non-reversing heat flow.
The next measured substances were glucose, fructose, dextrose and sucrose.
All have a defined glass transition temperature and take up water fast and easily. All
samples were measured from 50°C below to 50°C above the glass transition
temperature, with a heating rate of 1°C/min, pulse height of 0.5°C and a sample
mass between 2.00 and 6.00 mg. In all curves, the water peak is visible, but no
underlying effects were detected. For all the substances there is an extra peak at
about the middle of the peak due to the desorption of water. It occurs at about 70°C
for dextrose and 90°C for glucose (see figure 4.6.). Desorption means that, in this
case, water molecules leave the surface to join the gas phase. It can be concluded
that also for the sugars, TOPEM cannot separate between reversing and non-
reversing heat flow.
FIGURE 4.5. TOTAL (BLACK), REVERSING (GREEN) AND NON-REVERSING (BLUE)
HEAT FLOW CURVES OF A TOPEM MEASUREMENT OF PVP VA64
25
FIGURE 4.6. TOTAL (BLACK), REVERSING (GREEN) AND NON-REVERSING (BLUE)
HEAT FLOW CURVES OF A TOPEM MEASUREMENT OF DEXTROSE
(UPPER) AND GLUCOSE (LOWER)
4.3 CALCULATION WINDOW 4.3.1 IN GENERAL
For the evaluation of the data, several parameters have to be defined. The
width of calculation window, sample and pan weight, parameters for fitting the points
in a curve and the width of smoothing window have to be determined.
The calculation window is a set of data points which are considered for the
calculation of the heat flow curves out of the raw data. The chosen width is a time
interval which is considered for calculation for each data point. The shift of calculation
window is how far the calculation window moves each time. The width of the
calculation window is recommended by Mettler-Toledo to be one third of the
transition whereas it is said to be about one tenth by Fraga (2007). This window width
and its effect on the eventual curves will be investigated. When a narrow calculation
window is used, the calculation goes faster, but less information is obtained. Using
bigger calculation windows more information is obtained, but the calculation requires
more time. Large window widths need low heating rates. A small shift in calculation
window gives a good analysis but takes a long time.
26
FIGURE 4.7. WIDTH AND SHIFT OF CALCULATION WINDOW (METTLER-TOLEDO 2008)
The sample and pan weight can be filled in to quantify measurements or to
define a heat capacity. These weights are without sample and are a blank. The
sample and instrument response parameters are the degrees of the polynomials
used to fit the data points in a curve. After the curves are calculated with the
calculation window, they still have to be smoothened. Therefore the width of
smoothing window has to be set. It has to be smaller than the calculation window
because it smoothens within this window.
4.3.2 MANNITOL
For the investigation of the calculation window, a 50:50 mixture of α-ß mannitol
was measured. Mannitol can crystallize in several forms of which two polymorphic
forms are widely described in the literature: α-mannitol and ß-mannitol. ß-mannitol is
the most stable one and the metastable α-mannitol transforms to the ß-form over
time. The orthorombic α and ß-forms have many similarities. Their physical
properties such as melting temperature, enthalpy of fusion, density and specific heat,
only differ little. They can be separated using X-ray diffraction and infrared
spectroscopy. It’s not possible to differentiate them with conventional DSC
measurements because their melting points are too close to each other: α-mannitol
melts at 166°C and ß-mannitol at 166.5°C. (Burger et al., 1999)
27
When a mixture of α and ß-mannitol is being melted, the metastable α-form
will melt first. Because of the presence of ß-mannitol, the molten α-mannitol will
recrystallize into ß-mannitol. This is a non-reversible effect because ß-mannitol will
never crystallize into the less stable α-form when the more stable ß-form is present.
This means that this effect should be seen in the non-reversing heat flow.
4.3.3 CALCULATION WINDOW PARAMETERS
During the evaluation of mannitol, it was found that the width of the calculation
window influenced the curves. The correct width of calculation window is the one
which gives a good fit between the total and the measured heat flow. That is always
the case when the total heat flow is calculated from the measured heat flow, as in
ADSC for example. But in TOPEM, it is the reversing and non-reversing heat flows
which are calculated from the measured heat flow and the total heat flow is the sum
of both. Afterwards, it has to be checked if this total heat flow and the measured heat
flow do fit. If not, no conclusions can be drawn. To make these curves fit, the
evaluation parameters have to be set.
With every new calculation window a new curve is obtained. The window that
gives the best fitted total heat flow curve should be taken for interpretation of the
data. To see if there is a good fit, the measured heat flow and the total heat flow were
subtracted. The obtained curve should not have large fluctuations. These curves
were compared for several calculation windows for several mixtures of α-ß mannitol.
To analyze the fluctuations in the subtraction curves, the amplitudes at each point are
summed up. If there are large fluctuations, the sum of the amplitudes is high. The
lowest sum of amplitudes accords to the best fit for total and measured heat flow and
the corresponding width of calculation window is the best to analyze the sample.
28
FIGURE 4.8. TOTAL HEAT FLOW (BLACK), MEASURED HEAT FLOW (RED) AND
SUBTRACTION OF BOTH (PURPLE) OF A TOPEM MEASUREMENT OF Α-ß
MANNITOL; CALCULATION WINDOW IS 120S (UPPER), 300S (MIDDLE)
AND 550S (LOWER)
The calculated window should correspond to the recommended calculation
window, namely one third of the transition interval. The transition interval for every
mixture is divided by three to obtain the recommended width of the calculation
window. As can be seen in table 4.1. the recommended and optimal calculation
windows are different. Only for the 80:20 mixture, one third of the transition interval is
close to the optimal width of the calculation window. For all the others, the window
should be bigger. In figure 4.9. can be seen that for all the mixtures, except the 60:40
mixture, several calculation window widths will yield acceptable results, because their
sums of amplitudes are in the same range.
From this data was concluded that it is important to determine the correct
calculation window before any conclusions are drawn. However, it is difficult to
determine the optimal calculation window width since similar results were obtained.
When the average of these was made, the calculation window width should be
around one fifth of the transition interval, but this should be evaluated for each
sample separately.
29
A B
C D
E
FIGURE 4.9. SUM OF THE AMPLITUDES OF THE SUBTRACTION CURVE BETWEEN THE
TOTAL HEAT FLOW AND MEASURED HEAT FLOW FOR SEVERAL α-ß
MANNITOL MIXTURES: 20:80 (A), 40:60 (B), 50:50 (C), 60:40 (D) AND
80:20 (E) α-ß MANNITOL; THE LOWEST SUM IS MARKED (RED).
30
TABLE 4.1. COMPARISON OF CALCULATED BEST AND RECOMMENDED CALCULATION
WINDOW (1/3 OF TRANSITION INTERVAL)
Mixture Transition
interval (s)
1/3 of
transition
interval (s)
Calculated best
calculation
window (s)
Part of
transition
interval
20:80 1200 400 250 1/5
40:60 1200 400 200 1/6
50:50 1200 400 250 1/5
60:40 1440 480 375 1/4
80:20 1440 480 475 1/3
4.3.4 EFFECT OF THE PULSE WIDTH ON THE CALCULATION WINDOW
As already pointed out, the width of the calculation window has a big influence
on the outcome of the final heat flow curves. The goal was to find a defined width of
calculation window. The problem is that TOPEM uses a stochastically changing pulse
width. In addition, TOPEM does not provide the raw data, there has already been a
mathematical modulation executed. In order to overcome the problem with the pulse
width and the data, a TOPEM simulation was carried out with regular DSC. The
modulation was done with a periodic step function of 60s.
The measured substance was a 50:50 mixture of α and ß-mannitol. The
following parameters were used: heating rate 0.05°/min, pulse height 0.02°C, period
60s and sample mass between 2.00 and 5.50 mg. We are especially interested in
this substance because the ADSC measurement shows no reversing effect while the
TOPEM measurement does show a peak in the reversing heat flow. This is shown in
figure 4.10. Figure 4.11. shows the raw data of the sample, an empty pan and the
reference pan. Not the heat flow but the temperature is taken for analysis because
the temperature is the only true measured value and the heat flow is correlated to it.
31
A B
FIGURE 4.10. MEASURED (RED), TOTAL (BLACK), REVERSING (GREEN) AND NON-
REVERSING (BLUE) HEAT FLOW CURVE OF A TOPEM MEASUREMENT
(A) AND AN ADSC MEASUREMENT (B) OF A 50:50 MIXTURE OF α-ß
MANNITOL
FIGURE 4.11. RAW DATA OF THE SAMPLE, EMPTY PAN AND REFERENCE PAN OF A
PERIODIC TMDSC MEASUREMENT. CURVES ARE A FUNCTION OF TIME
AND TEMPERATURE.
32
The three curves are still a function of time and of temperature. The aim is to
eliminate the factor temperature so that the slope of the curve becomes zero. In order
to do that, the average of the pan curve is calculated and subtracted of all the curves.
This average is actually the heating rate of the measurement. After this is done, the
peak of melting can be seen in the sample curve (pink curve in figure 4.12.). This
curve is considered for further analysis of the calculation window.
Since this is a periodic modulation, the period should be used as a calculation
window width. For comparison, two other widths are considered: 200s and 330s. As
transition interval the segment between 2600 and 3600 seconds is considered. The
width of 200 corresponds to about one fifth of the transition interval, which was found
to be the correct width. The width of 330s corresponds to about one third of the
transition interval, which is the recommended calculation window width. As can be
seen in figure 4.12. the width of the period gives the best fit. The width of 200s and
330s are not flat anymore.
FIGURE 4.12. RAW DATA OF THE SAMPLE OF A PERIODIC TMDSC MEASUREMENT.
AVERAGE CURVES OF THE SAMPLE ARE GIVEN FOR A PERIOD, 200S
AND 330S. CURVES ARE ONLY A FUNCTION OF TIME.
33
To take a closer look at how the sample responses to the set value and in
order to compare each segment to another, the melting peak had to be eliminated.
For further calculations, figure 4.12. is considered. The sample curve (pink) and its
averages were subtracted for all three average curves. This means that a subtraction
was done for the width of the period (pink minus blue curve), the width of 200s (pink
minus green curve) and for the width of 330s (pink minus brown curve). These
subtraction curves were compared to the reference. The resulting curves are shown
in figure 4.13. The same colors are used for the same widths of calculation window.
In figure 4.13. can again be seen that different curves are created for the different
calculation windows and that the period gives the best width.
To compare the three window widths better, every period was overlapped to
see any changes in the curves. This means that all the curves in figure 4.13 were cut
every 60s and all these pieced were laid over each other. The result is shown in
figure 4.14. Only the first half of period is shown, but again the curves are different for
the three window widths. For the width of 330s, the curves even go below zero. The
sample is able to follow the oven temperature quickly.
FIGURE 4.13. BEHAVIOR OF THE SAMPLE; COMPARISON OF DIFFERENT CALCULATION WINDOWS
34
FIGURE 4.14. BEHAVIOUR OF THE SAMPLE; COMPARISON OF DIFFERENT
CALCULATION WINDOWS; BLUE CURVES ARE BEFORE THE TRANSITION,
PINK ARE DURING AND YELLOW ARE AFTER
In order to understand the shape of the curves in figures 4.13. and 4.14., the
behavior of the sample should be investigated more. This is explained in figure 4.15.
There are two effects which have to be considered: the effect of melting (green) and
the effect of the diffusion of the heat through the sample (brown). Both curves result
in the sample curve (blue). In the mathematical evaluation of the data (see
introduction under 1.3.3.4), these curves are part of the equation 27. The equation
y(t) = g(t)u(t) was rewritten in the z-plane as y(z) = H(z)u(z) in which H(z) can be
described as a rational function.
y(t) = g(t) u(t)
Where: y(t): output signal g(t): pulse response of the system
u(t): input signal t: time (s)
Where B(z) and A(z) are polynomials of degree q in the variable z.
35
The last equation holds the functions which describe the melting part and the
heat diffusion part of the sample. B(z) is the melting part and A(z) is the heat diffusion
of the sample. Using these equations, in the time domain, y(t) becomes the following:
y(t) = locally
y(t) = u(t) in general
The output value or sample curve y(t), the input value or set value u(t) and the
heat diffusion A(t) are known variables, so the melting B(t) can be determined.
FIGURE 4.15. BEHAVIOR OF THE SAMPLE, THE EFFECT OF MELTING AND EFFECT OF
HEAT DIFFUSION
A last step in comparison of the different calculation windows was to evaluate
the melting behavior. For the three different sample curves (blue, green and brown
curve) as shown in figure 4.13. the average is calculated. Since there was no
reversing effect found in the ADSC measurement, it is also expected not to see a
reversing effect in this measurement. This means that the calculated average curve
36
should be flat. In figure 4.16. can be seen that this is almost the case for the period-
curve, but the bigger the calculation window, the more the curves deviate.
FIGURE 4.16. AVERAGE OF THE SAMPLE CURVE FOR DIFFERENT CALCULATION
WINDOWS
It was already clear that several widths of calculation window give different
heat flow curves. The curves shown in figure 4.13, 4.14. and 4.16. also prove this. In
addition, several ways proved that the period is the best width for the calculation
window. The problem with TOPEM is that, because of the stochastic pulses, there is
no fixed period. It is because of this that the calculation of the heat flow curves
becomes difficult. Therefore we would recommend using TOPEM in a periodic way to
overcome the problem with the calculation window. This means however that no
frequency evaluation can be done since a periodic modulation holds only one
frequency.
The difference with ADSC is that TOPEM uses a step function instead of a
sine wave as modulation. A sample cannot follow a sine wave well because there is
no time for equilibration. Because of the many fluctuations, a blank and calibration
curve need to be executed for every new method in ADSC. Using a step function as
in TOPEM, the sample has the time to equilibrate every step upwards and
downwards. That’s why there are no blank and calibration curves needed for TOPEM
as there is for ADSC. When TOPEM is used in a periodic way, more measurements
37
will be needed to do a frequency evaluation. There is however no need for blank and
calibration curves, so a TOPEM evaluation will still be less time consuming than an
ADSC evaluation.
4.4 HEAT CAPACITY In TOPEM, the heat capacity (Cp) is calculated from the pulse response. The
heat capacity is the enthalpy divided by temperature and sample mass (see also
introduction). The enthalpy is given by integration of the heat flow. For a heat
capacity evaluation in an ADSC measurement, a blank and calibration curve need to
be executed for every new method. As been said before, those curves aren’t needed
in TOPEM. The sample and pan weight can be filled in in the calculation window and
this is used as a blank. The fact that no extra curves are needed in TOPEM leads
also to a simpler interpretation of the curves.
A 50:50 mixture of α and ß-mannitol is used for the evaluation of Cp. For the
calculation of Cp out of the raw data, data from the measurements under 4.3.4 are
used. To calculate Cp out of the heat flow for TOPEM and ADSC, the data from the
measurements under 4.3.3 (see also figure 4.10.) was used. All three curves are
shown in figure 4.17. The peak middle is for all three at the same time. This means
that it doesn’t matter if the heat capacity is calculated from the pulse response or
from the heat flow.
FIGURE 4.17. COMPARISON OF HEAT CAPACITY CALCULATED FOR TOPEM FROM THE
RAW DATA AND THE HEAT FLOW AND FOR ADSC FROM THE HEAT FLOW
38
4.5 FREQUENCY The TOPEM technique promises a frequency evaluation in just one single
measurement. Frequency is the reciprocal of the period.
Where: f: frequency (s)
τ: period (Hz)
This frequency evaluation is actually the calculation of the complex heat
capacity out of the heat capacity for a set frequency. In TOPEM, the set value for the
stochastic pulses is 15 to 30 seconds and can go up to 500s. Theoretically, this
means that the frequency can take values from 33.3 mHz to 66.7 mHz. A 40:60
mixture of sucrose and water was used for the frequency evaluation. The upper limit
is 50 mHz. The 60 mHz curve cannot be seen because it’s covered by the 50 mHz
curve. The lower limit is 10 mHz. Also the 5 mHz curve cannot be seen because it’s
covered. The complex heat capacity curves can be considered between 10 and 50
mHz.
Fraga et al. (2007) mentioned a limit of 4 mHz for the frequency evaluation.
This accords to a period of 25 seconds. Because the period goes from 15 to 30
seconds, we assume that 4 mHz as a limit only counts for the substance used in that
article. Because the theoretical limit, the limit in our measurement and in the literature
differ, we assume that every substance has its own limit. The same article also
mentioned that the complex heat capacity shifts to higher temperatures for higher
sample masses for a selected frequency. This shift was also the case with the
sucrose-water mixture (see figure 4.18.). The complex heat capacity shifts to lower
temperatures because the measurement is a cooling, not a heating.
39
FIGURE 4.18. COMPARISON OF MIDPOINT TEMPERATURES OF COMPLEX HEAT
CAPACITY FOR DIFFERENT SAMPLE MASSES OF A 40:60 MIXTURE OF
SUCROSE-WATER IN A TOPEM MEASUREMENT; COMPLEX HEAT
CAPACITY IS CALCULATED FOR 50 MHZ
The TOPEM frequency evaluation was compared to ADSC measurements
performed at periods according to frequencies of 10 to 50 mHz, as in TOPEM. The
results are shown in figure 4.19. and according values for the glass transition are
shown in table 4.2. In both techniques, the midpoint temperature moves to lower
temperatures for higher frequencies. The difference in the values for ADSC are
bigger but considered neglectable since the mean values for both techniques only
differ little. This is probably because both techniques use different temperature
modulations.
We can conclude that a frequency evaluation done by TOPEM is comparable
to ADSC. Of course, there has to be considered that also the heat capacity curve is
calculated from the raw data using a calculation window. First, as for any TOPEM
evaluation, a frequency evaluation can only be done if the total heat flow curve does
give a good fit with the measured heat flow. Second of all, we would recommend
doing a regular DSC measurement and comparing as well the total heat flow curve
40
as the heat capacity before doing a frequency evaluation. Still a TOPEM
measurement will be less time consuming than an ADSC measurement.
A B
FIGURE 4.19. FREQUENCY EVALUATION OF A TOPEM MEASUREMENT (A) AND AN
ADSC MEASUREMENT (B) OF A 40:60 MIXTURE OF SUCROSE-WATER
TABLE 4.2. GLASS TRANSITION TEMPERATURES FOR THE TOPEM AND ADSC
MEASUREMENTS ACCORDING TO FIGURE 4.19.
TOPEM ADSC
Glass transition Onset (°C) Midpoint (°C) Onset (°C) Midpoint (°C)
10 mHz -34.94 -33.95 -35.36 -34.18
20 mHz -34.99 -33.55 -35.23 -34.20
30 mHz -34.60 -33.17 -33.59 -31.93
40 mHz -34.10 -32.92 -33.57 -31.93
50 mHz -34.07 -32.86 -33.27 -31.28
41
4.6 RESOLUTION It is claimed that TOPEM has a good resolution. As can be seen in figure 4.20.
this is a true statement. To go in one period from the highest to the lowest point,
more than one hundred data points are in between. In order to have a good
resolution between the pulses, at least fifty data points are needed. It is also claimed
that TOPEM has a good sensitivity, but because the raw data cannot be obtained,
this could not be investigated.
FIGURE 4.20. MEASURED HEAT FLOW CURVE OF A TOPEM MEASUREMENT OF A
50:50 MIXTURE OF α AND ß-MANNITOL. USED PARAMETERS: HR
0.05°C/MIN, PH 0.02°C. SAMPLE MASS = 4.98 MG.
42
5 CONCLUSIONS
In this study, the temperature modulated DSC technique TOPEM was
investigated. During the investigation of the separation of reversing and non-
reversing heat flow was noticed that often no extra information is gained using
TOPEM. For measurements where it is suspected that separating reversing and non-
reversing heat flow is useful, we would recommend to do a quick run with regular
DSC for comparison.
The biggest problem seems to be setting the evaluation parameters, more
specifically the width of calculation window. In order to calculate the heat flow curves,
this calculation window has to be set. Because there is no fixed period as in ADSC, it
is not easy to define this window. Different articles recommend different calculation
windows. We’ve proven that the shape of the curves is very dependent on this
calculation window. Different windows can create different curves and therefore also
false curves.
It was however found, that the frequency evaluation and heat capacity were
comparable to ADSC and that there is a good resolution. The frequency evaluation
still has a limit for every substance despite the theoretical limits and the measured
complex heat capacity has a shift in temperature for higher sample masses. Further,
it can be concluded that the pulse response can be used to define the heat capacity.
For as well the heat flow as the frequency evaluation there is still the problem of the
calculation window. The correct width of this calculation window should be checked in
advance before doing any further evaluations.
When everything is put together, we would recommend using TOPEM not in a
stochastic way but by using a fixed period. This should solve the problem with the
calculation window. This means however that there is no frequency evaluation
possible. However, there is no need to do blank and calibration curves as in ADSC,
so it will still be less time consuming as an ADSC measurement. We also
recommend doing a regular DSC measurement in advance for comparison.
43
6 LITERATURE LIST
Analytical profiles of drug substances and excipients. Brittain, H.G. (Ed.) (2002),
Center for Pharmaceutical Physics, Milford, NJ, USA
Burger, A.; Henck, J.; Hetz, S.; Rollinger, J. M.; Weissnicht, A. A.; Stöttner, H. (1999).
energy/temperature diagram and compression behavior of the polymorphs of D-
manntiol. Journal of Pharmaceutical Sciences, 89 (4), 457-468
Cao, J. (2007). Numerical simulation of DSC and TMDSC curves as well as reversing
and nonreversing curve separation. Journal of Applied Polymer Science, 106, 3063-
3069
Fraga, I.; Montserrat, S.; Hutschinson, J.M. (2007). TOPEM, a new temperature
modulated DSC technique. Application to the glass transition of polymers. Journal of
Thermal Analysis and Calorimetry, 87, 119-124
Höhne, G. W. H. (1997). Remark on the interpretation of the imaginary part C″ of the
complex heat capacity. Thermicochimica Acta, 304/305, 121-123
Http://www.mt.com
Http://www.fchpt.stuba.sk
Izzard, M.J.; Ablett, S.; Lillford, P.J.; Hill, V.L.; Groves, I.F. (1996). A modulated
differential scanning calorimetry study. Journal of Thermal Analysis, 47, 1407-1418
Jörimann, U.; Widmann, G.; Riesen, R. (1999). Temperature modulated DSC
(TMDSC) applications and limits of phase information, cp determination and effect
separation. Journal of Thermal Analysis and Calorimetry, 56, 639-647
Martindale, the extra pharmacopoeia. Reynolds, J. E. F. (Ed.) (1996), Royal
pharmaceutical society of Great Brittan, London, UK, pp. 407, pp. 901
44
Mettler-Toledo (2008). Manual TOPEM. Mettler-Toledo GmbH, Gießen, Germany
Mettler-Toledo (2009). Seminar Thermal Analysis. Mettler-Toledo GmbH, Gießen,
Germany
Riesen, R.; Widmann, G; Truttmann, R. (1996). Alternating thermal analysis
techniques. Thermochimica acta, 272, 27-39
Sauerbrunn, S.; Jing, N.; Riesen, R. (2002). An updated tool to extend the utility of
DSC. American Laboratory, 34 (16), 15-21
Schawe, J. E. K. (2005a). The separation of sensible and latent heat flow using
TOPEM®. Thermische Analyse Usercom, 22, 16-19
Schawe, J. E. K. (2005b). TOPEM® - The new multi-frequency temperature-
modulated technique. Thermische Analyse Usercom, 22, 16-19
Schawe, J. E. K. (2005c). Analysis of melting processes using TOPEM®. Thermische
Analyse Usercom, 25, 16-19
Schawe, J. E. K. (2006). A quantitative DSC analysis of the metastable phase
behavior of the sucrose–water system. Thermicochimica Acta, 451, 115-125
Schawe, J. E. K.; Hess, U. (2002). Influence of moisture on the glass transition of a
spray-dried compound using the ISOSTEP® method. Journal of Thermal Analysis
and Calorimetry, 68, 741-749
Schawe, J. E. K.; Hütter, T. (2005d). TOPEM® - The latest innovation in temperature
modulated DSC. Thermische Analyse Bulletin, 27 (86), 19-25
Schawe, J. E. K.; Hütter, T.; Heitz, C.; Alig, I.; Lellinger, D. (2005e). Stochastic
temperature modulation: a new technique in temperature-modulated DSC.
Thermochimica Acta, 446, 147-155
45
Schawe, J. E. K.; Schubnell, M. (2009). Webinar: TOPEM® - A New Temperature-
Modulated Technique. Mettler-Toledo GmbH, Gießen, Germany
Schubnell, M. (1998). Tutorial kit. Metller-Toledo GmbH, Schwerzenbach,
Switserland.
Schubnell, M.; Heitz, C.; Hütter, T.; Sauerbrunn, S.; schawe, J. E. K. (2006). A
Multifrequency Temperature-Modulated Technique for DSC. American Laboratory,
38(1), 18-21
Schubnell, M.; Schawe, J. E. K. (2001). Quantitative determination of the specific
heat and the glass transition of moist samples by temperature modulated differential
scanning calorimetry. International Journal of Pharmaceutics, 217, 173-181
Schwarz, E.; de Buhr J. (1998). Pharmazeutica. Mettler-Toledo GmbH,
Schwerzenbach, Switserland
Simon, S. L. (2001). Temperature-modulated differential scanning calorimetry: theory
and application. Thermicochimica Acta, 374, 55-71
TOPEM® – The new advanced multi-frequency TMDSC technique. Mettler-Toledo
GmbH, Schwerzenbach, Switserland
Van der Plaats, G (1992). Die praxis der thermische analyse. Mettler-Toledo,
Schwerzenbach, Switserland
Widmann, G.; Riesen, R. (1984). Thermoanalyse: anwendungen, begriffe, methoden.
Dr. Alfred Hüthig verlag GmbH, Heidelberg, Germany