Form 2066-2.20/030807 Page 1 PRIMUS Clinical – Large Size Sterilizers These sterilizers are designed for all Clinical applications including surgery, central sterile supply. They are pit-mounted, allowing the optional floor carts to roll directly into the chamber. PRIMUS sterilizers are designed, developed and manufactured in the USA according to Quality Management System which are in compliance with ISO 9001:2000, ISO 13485:2003, CMDCAS and FDA’s Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21CFR Part 820). Large size sterilizers are one segment of 10 standard clinical sizes manufactured by PRIMUS. SPECIFICATION DATA SHEET The pressure vessel is the heart of any sterilizer. PRIMUS offers simple, straight forward, design. All PRIMUS pressure vessels are constructed of solid stainless steel and are fabricated by PRIMUS in our quality controlled, ASME facility. The interior chamber surface is polished to a mirror finish of <10μ in. Ra. The brilliant Pri-Mirror ® reflective chamber finish sets the highest standard for cleanliness and offers easy spill clean-up and resistance to staining and degradation. PRIMUS rectangular chambers eliminate the wasted space and high utility cost common to old-style cylindrical or elliptical chambers and jackets. This provides savings in operating cost and service. All large sizes of PRIMUS sterilizers have pneumatically powered horizontal operating doors, designed to be efficient, reliable and inherently safe. To close the door, depress and hold the door button, if button is released door opens. An added safety feature, the doors stop automatically if an obstacle is encountered. All PRIMUS steam sterilizers include Gravity, Vacuum and Liquid cycle settings. A Test/Vac cycle is provided for utilizing Bowie-Dick™ tests. For surgical packs, select one of the vacuum cycles. For the sterilization of liquids in open or vented containers, select a liquids cycle. The time-tested and reliable PRIMUS PSS500 Microcomputer Control is an industrial microprocessor providing accuracy and automation for all customer requirements. This simple, versatile, control has become the first choice for both sterilizer users and service technicians. STERILIZER DESIGN STERILIZER CYCLES Pit-mounted sterilizers with matching floor loading carts are guided into the chamber by floor tracks which match cart casters. Pass- through models are available PRIMUS Sterilizer Company, LLC • 117 South 25 th Street • Omaha, NE 68131 Phone (402) 344-4200 Fax (402) 344-4242 info@primus-sterilizer .com www .primus-sterilizer .com Two 26” x 63” Pit Mounted Units

F2066 Large Clinical Spec Sheet 2 20

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: F2066 Large Clinical Spec Sheet 2 20

Form 2066-2.20/030807 Page 1

PRIMUS Clinical – Large Size SterilizersThese sterilizers are designed for all Clinical applications including surgery, central sterile supply. They arepit-mounted, allowing the optional floor carts to roll directly into the chamber. PRIMUS sterilizers aredesigned, developed and manufactured in the USA according to Quality Management System which are incompliance with ISO 9001:2000, ISO 13485:2003, CMDCAS and FDA’s Good Manufacturing Practice forMedical Devices: General (GMP) regulation (21CFR Part 820). Large size sterilizers are one segment of 10standard clinical sizes manufactured by PRIMUS.

SPECIFICATIONDA

TA

SH

EET

The pressure vessel is the heart of any sterilizer.PRIMUS offers simple, straight forward, design. AllPRIMUS pressure vessels are constructed of solidstainless steel and are fabricated by PRIMUS in ourquality controlled, ASME facility. The interior chambersurface is polished to a mirror finish of <10µ in. Ra.The brilliant Pri-Mirror® reflective chamber finish setsthe highest standard for cleanliness and offers easyspill clean-up and resistance to staining anddegradation.

PRIMUS rectangular chambers eliminate the wastedspace and high utility cost common to old-stylecylindrical or elliptical chambers and jackets. Thisprovides savings in operating cost and service.

All large sizes of PRIMUS sterilizers havepneumatically powered horizontal operating doors,designed to be efficient, reliable and inherently safe.To close the door, depress and hold the doorbutton, if button is released door opens. An addedsafety feature, the doors stop automatically if anobstacle is encountered.

All PRIMUS steam sterilizers include Gravity,Vacuum and Liquid cycle settings. A Test/Vaccycle is provided for utilizing Bowie-Dick™ tests.

For surgical packs, select one of the vacuumcycles. For the sterilization of liquids in open orvented containers, select a liquids cycle.

The time-tested and reliable PRIMUS PSS500Microcomputer Control is an industrialmicroprocessor providing accuracy andautomation for all customer requirements. Thissimple, versatile, control has become the firstchoice for both sterilizer users and servicetechnicians.

STERILIZER DESIGN

STERILIZER CYCLES

Pit-mounted sterilizers with matching floorloading carts are guided into the chamber byfloor tracks which match cart casters. Pass-through models are available

PRIMUS Sterilizer Company, LLC • 117 South 25th Street • Omaha, NE 68131Phone (402) 344-4200 • Fax (402) 344-4242 • [email protected] • www.primus-sterilizer.com

Two 26” x 63” Pit Mounted Units

Page 2: F2066 Large Clinical Spec Sheet 2 20

Form 2066-2.20/030807 Page 2

GENERAL (Options italicized) All models shall include Vacuum, Gravity andLiquids cycles. A Low Temperature Flowing Steamoption provides flowing steam at low temperaturesfor Research laboratories for specific applications.

VESSEL MATERIAL AND CONSTRUCTIONThe sterilizer will have a double-wall designproviding for a fully jacketed pressure vessel. Thechamber cross section will be rectangular with widthand height dimensions to include:

J - 26” x 63” x 48”K - 26” x 63” x 76”

The vessel material will be non-laminated solidstainless steel plate/sheet. The chamber, head-ringand door plate material will be 316L stainless steelwith an interior Pri-Mirror® finish of <10µ in. Ra.

DESIGN and MANUFACTURING APPROVALSThe vessel will be designed and constructed toASME code Section VIII Division I and pressurerated for 45 PSIG and full vacuum. ASME CodeStamp and U-1 form will bear the name of PRIMUSas the specified US manufacturer.

The sterilizer will be manufactured according toQuality Management Systems which are incompliance with ISO 9001:2000, ISO 13485:2003,CMDCAS, FDA and will be UL listed and inconformance to CSA requirements.

DOOR CONSTRUCTIONA pneumatically operated horizontally sliding doorwill be operable from the touchpad push button. Toclose door, depress/hold door button . If door buttonis released while closing, door reverses directionand opens. In the open position, the insulated doorwill be secured behind a panel preventing theoperator from contact with hot surfaces. An interlockwill prevent cycle start unless the door is fully closedand secured.

GENERAL CONSTRUCTIONThe vessel shall be insulated with 1" semi-rigid hightemperature fiberglass board/blanket overlaid withformed aluminum paneling and mounted in astructural steel frame. The frame shall be enamelcoated and fitted with adjustable legs on self-centering floor pads. The fascia and side panels willbe 16-gauge, type 304 stainless-steel, removable foreasy service access.

WATER CONSERVATIONTo conserve water and assure effluent exhausted todrain is 140oF or below, PRIMUS provides quenchwater on demand for exhaust effluent.

PURCHASING SPECIFICATION DATA SHEETEQUIPMENT WARRANTYSterilizer pressure vessels manufactured byPRIMUS are warranted against defects inworkmanship and materials under normal use andoperation for fifteen years where the sterilizer iscontinually maintained under PRIMUS servicecontract.

SERVICE and EQUIPMENT ACCESSStandard service access, when facing the unit will befrom left side and top. Wiring shall be laid side-by-side and mechanically secured flat against metalinsulations cover. All wiring will be clearly labeledand readily visible for visual tracing. Pipingcomponents will be threaded rigid brass and flaredcopper fittings, positioned with sufficient space forremoval and replacement without disassembly of theentire piping assembly. Wiring and pipingcomponents will be nonproprietary, industrial gradeavailable through Authorized Service Agencies, localsupply house, or direct from PRIMUS.

STEAM SOURCESteam will be from an in-house steam source in thequantity and quality specified. An optional electricboiler may be specified. Contact PRIMUS for specificboiler information

CONTROLSA PSS500 microcomputer will offer a selection of 8(eight) separate programs for the sterilization ofwrapped goods, hard-goods, liquids and test cycles.Cycle parameters may be set by the user and areadaptable to a wide range of products. Alarms areannunciated, displayed and printed. Cycle processparameters will be displayed and recordedthroughout the cycle. A side-mounted control paneland a thermal printer dot-matrix technology, and 32characters per line printing is provided. A secondcopy of the sterilization cycle report will be availableat the end of the run.

Standard Control Panel /Printer Arrangement

SPEC

IFIC

ATIO

N D

ATA

SH

EET

Page 3: F2066 Large Clinical Spec Sheet 2 20

Form 2066-2.20/030807 Page 3

DIMENSIONS, CUBAGE & WEIGHTS

ledoMeziSreziliretS

sretemillim/sehcni

J"84x"36x"62

2.9121x2.0061x4.066

K"67x"36x"62

4.0391x2.0061x4.066

emarFfo,snoisnemiDllarevO 06x09x144251x6822x4.1401

88x09x142.5322x6822x4.1401

rooDelgniS,thgieWSTHGIEWROFSUMIRPTCATNOC

rooDelbuoD,thgieW

thgieWdetarC YLNOTNEMPIHSDETACIDED

snoisnemiDdetarC 16x29x244.9451x8.6332x8.6601

98x29x246.0622x8.6332x8.6601

ebuCdetarC m4/tfuc731 3 m6/tfuc002 3

)straCroolF(tnempiuqEgnidaoL gk96/bl251 gk901/bl042

relioB gk653/bl587 gk953/bl097

HVAC DATA Heat loss, at ambient of 700 F.

LEDOM RH/S'UTBK:ROODELGNIS

,llawenohguorhTaicsafta

JK

55

:ROODELGNIS,llawenohguorhT

aeraecivres

JK

8.1121

:ROODELGNIS,gnidnatseerF

latottenibac

JK

6161

LEDOM RH/S'UTBK

:ROODELBUOD,llawenohguorhT

aicsafta

JK

55

:ROODELBUOD,llawenohguorhT

aeraecivres

JK

6161

:ROODELBUOD,sllawowthguorhT

aicsafhcaeta

JK

55

:ROODELBUODsllawowthguorhT

aeraecivres

JK

0101

SPEC

IFIC

ATIO

N D

ATA

SH

EET

Page 4: F2066 Large Clinical Spec Sheet 2 20

Form 2066-2.20/030807 Page 4

VOLUME/DIMENSIONS CHART

1. Fascia extends 1” beyond wall opening on each side overlapping the wall and sealing the opening.2. Allow minimum 2” clearance at rear of recessed unit only. Cabinet side models have rear clearance built in.3. Contact PRIMUS for alternative wall opening.4. Standard left side service/equipment access show. Optional right side access is available.

PRIMUS Clinical - Large Sterilizers

Refer to General Arrangement drawings for details and final connection point to utility services. (S-Steam, W-Water, D-Drain, E-Electrical, A-Air)

SPEC

IFIC

ATIO

N D

ATA

SH

EET

LEDOM J K

eziSrebmahC)lxhxw(

sretemillim/sehcni

84x36x629121x0061x066

67x36x620391x0061x066

yticapaCrebmahC tf5.54 3 m92.1/ 3 tf0.27 3 m40.2/ 3

htdiWllarevO 3 53.2201/52.04 53.2201/52.04

thgieHllarevO 1 6.8672/00.901 6.8672/00.901

)DS(htgneLllarevO 2 6.5261/00.46 8.6332/00.29

htdiWgninepOllaW 6.3312/00.48 6.3312/00.48

thgieHgninepOllaW 6.7832/00.49 6.7832/00.49

htgneLgninepOllaW 3 51.1851/52.26 53.2922/52.09

htdiWaicsaF 4.4812/00.68 4.4812/00.68

thgieHaicsaF 0.3142/00.59 0.3142/00.59

A DS

EW

W

AE

S

D

SINGLE DOOR

WE

A DS

WE

A DS

TYPICAL ACCEPTED RANGES

Service access space

Typical floor drain and

ancillary equipment area.

Page 5: F2066 Large Clinical Spec Sheet 2 20

Form 2066-2.20/030807 Page 5

Refer to General Arrangement drawings for details and final connection point to utility services.(S-Steam, W-Water, D-Drain, E-Electrical, A-Air)

VOLUME/DIMENSIONS CHART

PRIMUS Clinical - Large Sterilizers

1. Fascia extends 1” beyond wall opening on each side overlapping the wall and sealing the opening.2. Contact PRIMUS for alternative wall opening.3. Standard left side service/equipment access show. Optional right side access is available.

SPEC

IFIC

ATIO

N D

ATA

SH

EETLEDOM J K

eziSrebmahC)lxhxw(

sretemillim/sehcni

84x36x629121x0061x066

67x36x620391x0061x066

yticapaCrebmahC tf5.54 3 m92.1/ 3 tf0.27 3 m40.2/ 3

htdiWllarevO 2 53.2201/52.04 53.2201/52.04

thgieHllarevO 1 6.8672/00.901 6.8672/00.901

)DD(htgneLllarevO 81.1871/31.07 5.2942/31.89

htdiWgninepOllaW 2 6.3312/00.48 6.3312/00.48

thgieHgninepOllaW 2.2632/00.49 6.7832/00.49

htgneLgninepOllaW 19.5851/44.26 11.7922/44.09

htdiWaicsaF 4.4812/00.68 4.4812/00.68

thgieHaicsaF 0.3142/00.59 0.3142/00.59

E A

S

W

D

W

SDA

E

DOUBLE DOOR

EW

A DS

EW

A DS

Typical floor drain and

ancillary equipment area.

Service access space

TYPICAL ACCEPTED RANGES

Page 6: F2066 Large Clinical Spec Sheet 2 20

Form 2066-2.20/030807 Page 6

PRIMUS Clinical - Large Sterilizers

LEDOM J K

eziSrebmahC)lxhxw(

sretemillim/sehcni

84x36x629121x0061x066

67x36x620391x0061x066

yticapaCrebmahC tf5.54 3 m92.1/ 3 tf0.27 3 m40.2/ 3

htpeDtiP 8.403/00.21 8.403/00.21

)DS(htgneLtiP 4.9651/00.16 6.0622/00.98

)DD(htgneLtiP 3.1151/05.95 5.2222/05.78

htdiWtiP 0.2302/00.08 0.2302/00.08

Drawing is general example of PRIMUS’ pit installation. Refer to PRIMUS’ General Arrangement (GA)drawings for exact dimensions and pit installation details. Prior to installing your PRIMUS sterilizer verify thepit dimensions are correct with the PRIMUS GA Drawings, ensure the pit is square, clean and sealed.Contact PRIMUS for further pit installation information.

PIT DIMENSIONS CHART

SPEC

IFIC

ATIO

N D

ATA

SH

EET PIT INSTALLATION DETAILS

Page 7: F2066 Large Clinical Spec Sheet 2 20

Form 2066-2.20/030807 Page 7

STERILIZER UTILITY SERVICESStub-in utility services within 6”-0” of final connection to sterilizer. Optimum performance requires the specified utilities.

ARCHITECTURAL NOTES:S

PEC

IFIC

ATIO

N D

ATA

SH

EET

S maetylppuSmaetSgnidliuB

. epiP eziS wolebeeS:

. :ytilauQ etasnednoCeerf 001ot%79 %

s ropavdetarutaotdeppartylbatius(

dnamaetsyrderusneevomerotderetlif

)setalucitrap. erusserP : GISP08-05

cimanyD

RETAWylppuSretaWdloC

. epiP eziS woleBeeS:

. erutarepmeT : < 07 o F

. erusserP : GISP07-05cimanyD

NIARDmetsySniarDgnidliuB

"2muminiM

. noitacoL : nietacoLaeraecivres

:etoN.1 ) egrahcsidtsuahxE

otdeloocsi041< oF

"8x"21x"21).2siknisroolf

ybdednemmocerSUMIRP

riAriAtnemurtsnI

. noitcennoC : woleBeeS

. ytilauQ eerfliodnayrD:

. erusserP : ISP08-06cimanyD

LACIRTCELErewoPgnidliuB

ylppuS. stloV 011:. esahP elgniS:. spmA : 01

:etoNrofstiucriclanoitiddAlanoitpodnayrallicna

,.e.itnempiuqetsoob,pmupmuucav

.cte,reliob,pmup

ledoM TPN RH/SBL)RH/GK(

TPN )XAM(MPG)retiL(

TPN)eziSepiPegrahcsiD(

TPN

J "1 )1.561(0.463 "4/3 )35(41 "1 "4/1K "1 )27.712(0.084 "1 )59(52 "4/11 "4/1

ELECTRIC BOILERSLEDOM J K

eziSrebmahCsretemillim/sehcni)lxhxw(

84x36x629121x0061x4.066

67x36x620391x0061x4.066

eziSreiloB .rH/Wk 0.631 0.081

tuptuOmaetSrelioB rH/.sbl 0.014 0.245

largetnI ledoM A/N A/N

enolAdnatS ledoM J-BE K-BE

hP3,802.C.A.V serepmA 873 A/N

hP3,042.C.A.V serepmA 823 434

hP3,084.C.A.V serepmA 461 712

.zH06,011.C.A.V 1 serepmA 01 01

W

D

AE

S

1. Allow sufficient space for traps, shut-off’s, filters and other utility supply components.2. Utility (service disconnects) shall be provided and installed “By Others”.3. Building or structure modifications to accommodate the sterilizer, as well as, sterilizer shoring, rigging,

cribbing and/or crane requirements into the facility performed “By Others”.4. Provide maximum mechanical and service access space (refer to page 4 and 5), additional space required

when boiler specified. See General Arrangement drawing, for placement of ancillary equipment and serviceaccess.

5. Some options affect utility services and overall dimensions.6. Water Quality - refer to page 67. The Manufacturer’s Equipment Warranty does not cover failure due to improper utility provisions.8. Drawings not to scale.9. Wall thickness must be provided on single and double door models recessed through 1 or 2 walls, with

cabinet sides.

1. Controls Current

Page 8: F2066 Large Clinical Spec Sheet 2 20

Form 2066-2.20/030807 Page 8

SPEC

IFIC

ATIO

N D

ATA

SH

EET

SELECT CONFIGURATION

Use the following pages to custom design your sterilizer by checking the boxes of the configurationand options required for your project. Transfer all items checked to the Specification Worksheet(Page 12) and fax to PRIMUS Sales Department or to your local Sales Representative for a quotation.If you do not find an option or size listed to meet your requirements, please contact our ProjectManager for additional information and assistance.

Model Number NotationSD = Single DoorDD = Double Door

PSS5-__-M__ __ __ J S = House Steam K E = Electric Boiler

Two models/sizes are offered:J - 26" x 63" x 48"K - 26" x 63" x 76"

A detailed formatted specification, suitable forinclusion in formal contract documents, isavailable on request.

Note: Use the above format to determine PRIMUS Model numberand insert below and on Page 12, Specification Worksheet.

PRIMUS Clinical Steam Pressure Sterilizer, Model Number PSS5___________.

SELECT CONFIGURATION

CabinetCD - Panels Both SidesCCL - Left Side PanelCCR - Right Side Panel

Steam SourceHouse SupplyElectric Boiler1, 2, 3, 5

EB1 - 208/3 phaseEB2 - 240/3 phaseEB3 - 480/3 phaseOther

EB04- Auto. Blowdown

DAR - Single Door (Right)DAL - Single Door (Left)DBR - Double Door (Right)DBL - Double Door (Left)

Door RecessedCA - One WallCB - Two Walls

Service and Equipment AccessLeft Side (Standard)Right Side

1. Electric boilers are available in Carbon Steel or Stainless Steel.2. Contact PRIMUS for overall dimensions and utility connections.3. All models are stand-alone4 Low Water Cutoff is standard and the “automatic reset” feature is disabled with this option. The boiler will need

to be manually reset.5. Water Quality - for best results, the feed water supply should be evaluated prior to initial start-up by a

reputable water conditioning company. If the mineral content exceeds recommended limits, various externaltreatment processes (water softener, water conditioning, etc.) may be used to correct the problem. ContactPRIMUS for further recommendations.

Page 9: F2066 Large Clinical Spec Sheet 2 20

Form 2066-2.20/030807 Page 9

SPEC

IFIC

ATIO

N D

ATA

SH

EET

CLINICAL LABORATORY CONFIGURATIONS

Low Temperature Flowing Steam P14For media preparation, formula and similar uses.Pasteurization is accomplished with flowing steam.In addition to standard sterilizer temperatureranges, this option allows cycles to operate in therange of 1680F (760C) to 2200F (1040C). Notintended for reprocessing reusable medicaldevices

PSS500 Control C6PSS500 microcomputer with eight (8) separatefactory preset cycles; seven (7) programmable and 1preset Test cycle are standard. A vertical touchpadcontains a Liquid Crystal Display (LCD) mounted on aside-mounted control panel with a thermal printer.Cycle progression is displayed by Light EmittingDiodes (LEDs) on the operator panel and printed, 32characters per line printing. A second copy of theentire sterilization cycle printable at the end of therun.

Cycles may be programmed by the end user but anydeviation from the factory preset values must bevalidated by the end-user.

CONTROLS AND RELATED OPTIONS(Contact PRIMUS for details and performance specifications)

1. Gravity cycles must be loaded in a manner to ensure no airis entrapped in the load. E.g., invert pans, beakers andsimilar items.

SELCYCTESERPYROTCAF

epyTdaoL elcyCrebmuN

elcyCepyT

eziliretSpmeT

eziliretSemiT)nim(

depparWelbuoDhcaesbl61stnemurtsnI

yart

1 muucaV 372 oF431( o )C

51

suoropnoNdepparwnUtnemurtsnIelgniS

2 muucaV 372 oF431( o )C

4

depparWelbuoDsbl61syarTtnemurstnI

yarthcae

3 muucaV 372 oF431( o )C

4

sbl21skcaPelitxeT 4 muucaV 372 oF431( o )C

4

suoropnoNdepparwnUtnemurtsnIelgniS

5 ytivarG 372 oF431( o )C

4

mumixamskcaPcirbaFsbl21,02x21x21:ezis

6 ytivarG 352 oF8.221( -o

)C

03

etacilisoroBdetneVlm005,sreniatnocssalg

rellamsro

7 sdiuqiL 352 oF8.221( -o

)C

03

tseTkciD-eiwoB 8 tseT)CAV(

372 oF431( o )C

2/1-3

1 - Liquid Crystal Display (LCD)2 - Light Emmitting Diode (LED)3 - Cycle Buttons4 - Heart Buttons

1

2

4

3

Thermal Printer R7Thermal dot-matrix printer with take-up reel and 32character per line printing is standard. Secondsterilization cycle report available at the end of therun.

Remote Mount Control Panel C10Mount control panel in separate housing adjacent toor up to 35 feet away from the sterilizer.

Serial Data Output C11Supervisory use to capture electronic record of eachsterilization cycle. Cycle hard copy and display datais transmitted via RS232 in real time to remotecomputer, up to 100 feet away, for display andrecording. Hard copy records may be printed fromthe computer.

Contact PRIMUS for additional configurationoptions.

Page 10: F2066 Large Clinical Spec Sheet 2 20

Form 2066-2.20/030807 Page 10

SPEC

IFIC

ATIO

N D

ATA

SH

EET

UTILITY - RELATED OPTIONS(CON’T)

Uninterruptible Power Supply E2In the event of electrical power loss and no emergencyelectrical power to sterilizer, this option provides electri-cal power to operate control system for up to 30 min-utes, to complete the cycle. This option is a dedicatedpower source for the electronic controls only and will notsupport any power requirements for heavy power loadcomponents (boiler, pumps, compressors, etc.).

LOADING EQUIPMENT

Loading cart includes one bottom and one intermedi-ate shelf with four adjustable levels. Additionalshelves are available. Cart frames and shelves are316L stainless steel welded, ground and polished.Shelf surfaces are stainless steel wire mesh. Trans-fer carriages include swivel casters with swivel locksand 5” wheels with wheel brakes.

L4 Loading Cart (Floor Mounted Only) L5 Additional Cart Shelf L6 Transfer Carriage (Floor Mounted Only)

L7 Floor CartL8 Additional Floor Cart Shelves

NOTE: Loading cart and transfer carriages arerecommended for chamber lengths 48 inches orlonger. Pit Mounted units with chamber length equalor greater than 60” use one set of 2 floor carts

Contact PRIMUS for special loading equipmentrequirements.

FLOOR LOADER CART DIMENSIONS

GENERAL OPTIONS

Validation Port, 1.5 inch V4.1Provides chamber penetration to accommodatevarious monitoring/control probes. Two 1.5” valida-tion ports are standard. Option V5 provides for 2”validation ports.

Seismic Restraints V8Required in areas prone to seismic hazards.Secures sterilizer to building but allows for leveling.Designed to current California Code.

UTILITY - RELATED OPTIONS

Boost Pump for Low House Water Pressure P3Delivers required dynamic water pressure forefficient operation. Requires specified GPM flow.Contact PRIMUS for details and performancespecifications.

Transformers E1 Reduces line voltage to required 110 VAC foroperation of sterilizer controls.

E1A 220 vac to 110 vacE1B 480 vac to 110 vacE1C 480/240 to 240/120, 1 phase,

0.5 KVA

EZISREBMAHC ROOLFYTQTRAC

A 1B 2B C D

"84x"36x"62-J 7L-1 "42 "28.35 "91.06 "64 "61

"67x"36x"62-K 7L-2 "42 "28.35 "91.06 "5.63 "61

Page 11: F2066 Large Clinical Spec Sheet 2 20

Form 2066-2.20/030807 Page 11

This document and data contained within is intended for the exclusive use of PRIMUS customers, includingarchitects or designers. It shall not be reproduced in whole or in part without permission of PRIMUS SterilizerCompany. Any authorized reproduction must bear the entire page to include Headers and Footers.

SPEC

IFIC

ATIO

N D

ATA

SH

EET

Three 26” x 63” Pit-mounted Units

QUALITY AND INTEGRITY FORM THE FOUNDATION OFPRIMUS STERILIZER COMPANY

OUR VISION FOR CONTINUAL IMPROVEMENT IS REALIZEDTHROUGH THE ESTABLISHMENT OF QUALITY OBJECTIVES

ACHIEVED THROUGH TEAMWORK WITH A GOAL OFRETURNING SIGNIFICANT VALUE TO OUR CUSTOMERS.

“PRIMUS Sterilizer... ...First In Quality Sterilization Products”

PRIMUS Sterilizer Company, LLC • 117 South 25th Street • Omaha, NE 68131Phone (402) 344-4200 • Fax (402) 344-4242 • [email protected] • www.primus-sterilizer.com

Page 12: F2066 Large Clinical Spec Sheet 2 20

Form 2066-2.20/030807 Page 12

Complete the following by selecting or transferring those checked on preceding pages, required tomeet your needs. Tear off and forward completed worksheet to PRIMUS Sales Department,fax 402-344-4251, or your local PRIMUS Representative. If you do not find an option or size listed tomeet your requirement, please contact our Sales Department for additional information andassistance with a custom design to meet your specifications.

PRIMUS Specification Data Worksheet

REBMUNLEDOMSUMIRP YTITNAUQ

EMANTCEJORP

SSERDDATCEJORP

EMANYNAPMOC ENOHP

EMANTCATNOC XAF

LIAMETCATNOC

NOITCESNOITACIFICEPS .ONMETI

.ONGNIWARDNONWOHS .ONMOOR

.ONETOUQSUMIRP

GENERAL CONFIGURATIONSDA Single DoorDB Double DoorCD Cabinet, 2 sidesCCL Cabinet, 1 Side LeftCCR Cabinet, 1 Side RightCA Recessed, 1 WallCB Recessed, 2 WallNPN Left Side Service (Std)C30 Right Side Service Access & ControlC31 Control Panel opposite side of ServicePM Pit Mounted (Std)FM Floor Mounted (Special Order with ramp

or carriage/cart)STEAM SOURCE

EB1 Boiler, 208/3phEB2 Boiler, 240/3phEB3 Boiler, 480/3phEBO Automatic boiler blow-downLWC Low Water Cutoff, Manual Reset (Std)

CLINICAL LABORATORY OPTIONS

P14 Low Temperature Flowing SteamCONTROLS & RELATED OPTIONS

C6 PSS500 Control (Std)C10 Remote Mounted Control PanelC11 Serial Data Output (RS 232)R7 Thermal Printer, 32 Column (Std)R1 Impact Printer, Flush Mount 24 ColumnR4 Thermal Printer, Take-up ReelR4.1 Thermal Printer, Flush Mount 24 Column

Manufacturer reserves the right to modify materials or specification without notice.

Std = Standard ComponentN/A = Not Available

FOR ADDITIONAL ASSISTANCETo discuss your sterilization design needs or to requesta Sterilizer Design Guide, please contact PRIMUS.

YOUR LOCAL PRIMUS REPRESENTATIVE:

GENERAL OPTIONSV4.1 Two Validation Ports, 1.5 “ diameter (Std)V5 Validation Port, 2.0” diameterV8 Seismic RestraintP10 Compressed Air to GasketP25 Drain Line Strainer/ValveEAC Air Compressor

UTILITY - RELATED OPTIONSP3 Boost Pump for Low House WaterE1A Transformer from 220V to 110VE1B Transformer from 480V to 110VE1C Transformer from 480/240 to 240/120V,

1 ph, 0.5KVAE2 UPS maintains control functionality for

30 minutes during power outageE3 GFI Receptacle

LOADING EQUIPMENTL4 Loading Cart (Floor Mount Only)L5 Additional Cart ShelvesL6 Transfer Carriage (Floor Mount Only)L7 Floor CartL8 Additional Shelves

VESSEL OPTIONC4 Jacket Pressure Display, Analog

SPEC

IFIC

ATIO

N D

ATA

SH

EET

PRIMUS Sterilizer Company, LLC • 117 South 25th Street • Omaha, NE 68131Phone (402) 344-4200 • Fax (402) 344-4242 • [email protected] • www.primus-sterilizer.com