C511-E016A
Printed in Japan 6295-09801-30A-NS
2 3
Navigating Radiography Preparation
EZy-Rad Pro, an X-ray radiography system that combines
compact dimensions to allow installation in tight spaces
and makes imaging possible with simple operations, has arrived.
Its innovative "one-touch guide" informs the operator
of the proper exposure distance and cassette size in addition to
the X-ray conditions, and thereby facilitates stress-free preparation.
It offers superior image quality
and ease of use that meets the demands of medical facilities.
This is Shimadzu's new standard
for diagnostic X-ray imaging systems.
The ultra compact console is an example of Shimadzu’s leading-edge technology. It is designed to save space while enhancing operator-friendly interface. A standard feature is the productivity-boosting 432 anatomical programs. Entering and changing exposure parameters in these programs is straight-forward and easy. Additionally, automatic exposure control (AEC) and phototimer sensitivity selection are available as optional features.
432 Easy-to-SetAnatomical Programs
User-Friendly Console
Easy Operation
GENERAL RADIOGRAPHIC SYSTEM
The Arrival of a New "One-Touch Guide"
Functions That Incorporate Wisdom and Tenderness
Ensuring stress-free imaging by navigating radiography preparation
Optimum setting of the exposure conditions like the exposure distance, cassette size, and grid requirements in accordance with the region to be examined is extremely difficult and requires experience. With EZy-Rad Pro, the wisdom of experienced operators is incorporated in the console. The pre-programmed exposure conditions are displayed on the console for each region, enabling stress-free preparation even for unfamiliar regions.
Once APR is selected …
X-ray conditions
Exposure distance
Cassette size
Grid requirements
One-touch guide
Confirmed with display
NEW
The Communication option enables automatic linking with the radiography program selected for CR (optional).
The table surface is covered with a soft synthetic leather mattress to alleviate the hard, cold feeling that is experienced with standard tables. Naturally, it has antibacterial specifications.The heavy-duty wooden table top is selectable as well.
NEWPatient-ComfortSoft Table Top
No Hardness, No Cold Feeling
CHEST
X-rayconditions
SID 2mGrid35cm X 35cmcassette
ABDOMEN
X-rayconditions
SID 1mGrid43cm X 35cmcassette
ELBOW
X-rayconditions
SID 1mNo Grid24cm X 30cmcassette
(cassette size display is selectable, cm / inch)
4 5
Compact SizeWith the generator tucked under the examination table and no requirement for wall or ceiling supports, the EZy-Rad Pro requires a floor area of only 2.7m × 1.8m as the minimum space.
1.8m2.7m
Example of minimum installation space
* Lieder cassette stand is not included. Rails do not have to be fixed to walls. Units can be delivered through 70cm wide entrance.
Smoothly perform radiography of arm and leg joints, thanks to an X-ray tube arm design that allows easy forward and backward positioning. X-ray focus points can be set near the patient, when hand or elbow joint radiography is performed from the front of the table, allowing the patient to enjoy a relaxed, natural posture. Easy positioning allows leg and foot joint radiography to be performed without requiring the patient to turn left or right.
Wide transverse coverage for orthopedic radiography(M-Pack and MH-Pack only)
Clear images provide high-quality diagnosis
A design based on a 60kHz high-frequency, inverter, X-ray high-voltage generator and a 140kHU high-capacity X-ray tube unit allows high-current short-time radiography. This produces sharp images, regardless of patient movement.
MH-Pack, AH-Pack : 100mAs(200mAs optional)
High output High current determinesimage quality
High mAs - for all types of patients
A-Pack, AH-Pack : 16kW 200mA
(M-Pack and MH-Pack only)
NEW
New Functions Unique to EZy-Rad Pro That Reflect Consideration ofDose Management
Assisting X-ray dose management Reference skin dose indicator
The exposure distance is measured with an ultrasonic distance meter
attached to front of the collimator.
NEWLinking radiography program withCommunication Option
After examination, the exposure conditions can be sent to the CR side.
(Note) Some CR models may be not compatible with EZy-Rad Pro.Please contact Shimadzu sales representative office for more details.
(A-Pack and AH-Pack)
24cm
105cm
59cm148cm
(M-Pack and MH-Pack only)
93cm
In response to the ever-increasing demand for lower doses, we have prepared a function that calculates and displays the predicted X-ray dose before exposure based on NDD method (optional).
Combining the reference skin dose indicator and Communication option makes it possible to notify the CR side of the actual exposure conditions and dose calculated after exposure.
The dose is calculated before exposure from
the set conditions and exposure distance
and is displayed in real time.
Accommodating a variety of locations, the EZy-Rad Pro AH-Pack and MH-Pack can be operated using only 1.5kVA power source. Neither special power contracts nor expensive transformer is needed.
Inverter provides a small capacity power source (1.5kVA) (AH-Pack & MH-Pack)
Note: The cassette tray cannot move in the transverse axis direction.
The linkage of X-ray tube and cassette unit ensure the cassette tray is always centered on the X-ray beam center. This means no time-consuming focus-point positioning, and provides accurate, worry-free radiography. This linkage can also be easily released, allowing oblique radiography.
Easy positioning via linked mechanism
OPTION
6 7
1868 (73.5)
1050 (41.3)
2000 (78.7)
90°
Units: mm (inch)Scale: 1/50
90°
720
(28.
3)
1025
(40
.4)
240
550
(21.
7)
820
(32.
3)93
0 (3
6.6)
680 (26.8)
430
(16.
9)52
0 (
20.5
)
55 (2.2)
170 (6.7)
220(8.7)
SpecificationsDimensions (M-Pack and MH-Pack)
Radiographic systemHigh voltage generator Control cabinet including the X-ray high voltage generator Control consoleX-ray tube unitCollimatorHigh voltage cablePower source - Control cabinet Cable 10mControl cabinet - Control console Cable 12m
1
1111111
Standard equipment Radiographic systemHigh Voltage generator Control cabinet including the x-ray high voltage generator Control consoleX-ray tube unit CollimatorHigh voltage cablePower source - Control cabinet Cable 10mControl cabinet - Control console Cable 12m
1
1111111
Specifications
Standard equipment
Dimensions (A-Pack and AH-Pack)
200mAs unit (AH-Pack)Bucky unitUD-PC, Phototimer controlDirect phototimer pickupHand switchCommunication option ( 2) Reference skin dose indicator
Optional items (A-Pack / AH-Pack)
200mAs unit (MH-Pack)Bucky unitUD-PC, Phototimer controlDirect phototimer pickupHand switchCommunication option ( 2) Reference skin dose indicator
Optional items (M-Pack / MH-Pack)
Units: mm (inch)Scale: 1/50
680 (26.8)
55 (2.2)
170(6.7)
220(8.7)
430
(16.
9)52
0 (2
0.5)
Control consoleCabinet
Control consoleCabinet
Ezy-RAD Pro A-Pack 240kg
Control console 1.1kg
Weight
(Including the control cabinet and high voltage generator)
Ezy-RAD Pro AH-Pack 265kg
Control console 1.1kg
Ezy-RAD Pro M-Pack 270kg
Control console 1.1kg
Weight
(Including the control cabinet and high voltage generator)
Ezy-RAD Pro MH-Pack 295kg
Control console 1.1kg
1050 (41.3)
90°
2000 (78.7)
90°
850
(33.
5)90
0 (3
5.4)
550
(21.
7)
1868 (73.5)
1020
(40
.4)
720
(28.
3)
X-ray high voltage generatorControl methodRating
Tube voltage setting rangeTube current setting rangemAs setting range
Timer setting rangeAnatomical program exposureDisplaySelf-diagnostic functionPower sourceSystem / FrequencyNominal supply voltageTransformer capacity X-ray tube unitEffective focus sizeMax. anode heat capacityX-ray tube support unitFocus movement
Radiographic table Size of table topTable top materialTable heightX-ray gridSize of loadable cassette
High frequency Inverter (maximum 60 kHz)16kW125kV 125mA [250mA]100kV 160mA [320mA] 80kV 200mA [400mA][ ]: Current converted into two-pulse generator40 - 125kV63 - 200mA0.5 - 320mAs
0.0025 - 5secMax. 432 user-programmable parametersLED display Error message indication on the LED display
Single Phase AC / 50Hz or 60Hz 200 / 220 / 230 / 240V 20kVA
1.2mm100kJ (140kHU)
Vertical travel (focus - floor distance) 850 - 1750mm (33.4” - 68.9”) Operation : Manual, Electromagnetic lockLongitudinal travel (Along the guide rail) 1050mm (41.3”) Operation : Manual, Electromagnetic lockX-ray tube rotations on the horizontal axis ±90° Click stop by 90° 1868×720mm (73.5”×28.3”)wood or soft synthetic leather (selectable)550mm (21.7”)R=10 : 1, N=40lines/cm, fo=100cm, spacer=Al14”×17” to 61/2”×81/2”cm or inch size cassette based on IEC standard 60406
0.5 - 100mAs (0.5 - 200mAs)*125kV 32mAs (63mAs)*100kV 40mAs (80mAs)*80kV 50mAs (100mAs)*40kV 100mAs (200mAs)*
0.0025 - 1.6sec (0.0025 - 3.2sec)*
100 / 110 / 120 / 200 / 220 / 230 / 240V 1.5kVA
A-Pack AH-Pack
*( ) : With 200mAs unit
X-ray high voltage generatorControl method Rating
Tube voltage setting rangeTube current setting rangemAs setting range
Timer setting rangeAnatomical program exposureDisplaySelf-diagnostic functionPower sourceSystem / FrequencyNominal supply voltageTransformer capacity X-ray tube unitEffective focus sizeMax. anode heat capacityX-ray tube support unitFocus movement
Radiographic table Size of table topTable top materialTable heightX-ray gridSize of loadable cassette
High frequency Inverter (maximum 60kHz)20kW 125kV 160mA [320mA] 100kV 200mA [400mA] 80kV 250mA [500mA] [ ]: Current converted into two-pulse generator
40 - 125kV80 - 250mA0.32 - 320mAs
0.0025 - 4secMax. 432 user-programmable parametersLED displayError message indication on the LED display
Single Phase AC / 50Hz or 60Hz 200 / 220 / 230 / 240V 25kVA
1.2mm100 kJ (140kHU)
Vertical travel (focus - floor distance) 820 - 1750mm (32.2” - 68.9”) Operation : Manual, Electromagnetic lockLongitudinal travel (Along the guide rail) 1050mm (41.3”) Operation : Manual, Electromagnetic lockTransverse travel (towards operator) 240mmX-ray tube rotation on the horizontal axis ±90° Click stop by 90° 1868×720mm (73.5”×28.3”)wood or soft synthetic leather (selectable)550mm (21.7”)R=10 : 1, N=40 lines/cm, fo=100cm, spacer=Al14”×17” to 61/2”×81/2”
0.32 - 100mAs (0.32 - 200mAs)*125kV 36mAs (71mAs)*100kV 50mAs (100mAs)*80kV 63mAs (125mAs)*45kV 100mAs (200mAs)*
0.0025 - 1.6 sec (0.0025 - 3.2sec)*
100 / 110 / 120 / 200 / 220 / 230 / 240V 1.5kVA
MH-PackM-Pack
*( ) : With 200mAs unit
cm or inch to size cassette based on IEC standard 60406
( 1) Lieder cassette stand is not included. ( 1) Lieder cassette stand is not included.
( 1)
( 1)
( 2) Some CR models may be not compatible with EZy-Rad Pro.Please contact Shimadzu sales representative office for more details.
( 2) Some CR models may be not compatible with EZy-Rad Pro.Please contact Shimadzu sales representative office for more details.