Upload
r-haas
View
213
Download
0
Embed Size (px)
Citation preview
Journal of Oral Rehabilitation 1999 26; 19–24
Experimental study of the damping behaviour of IMZimplantsR . H A A S , T . B E R N H A R T, O . D O R T B U D A K & G . M A I L AT H Department of Oral Surgery,School of Dentistry of the University of Vienna, Austria
SUMMARY Measurements of the damping behaviour
of dental implants with the Periotest® device are
considered to be an objective means to assess the
mobility of implants. The effects of the position of
an implant in the maxilla or mandible, the period
of time passing between the measurements and
implant placement and the height at which the
Periotest® measurements are performed on the
damping behaviour of implants have been discussed
controversially. This experimental study examined
the influence of the use of different measuring
devices, the measuring height and the embedding
depth on the damping behaviour of IMZ implants.
The implants were embedded in resin at different
depths and damping measurements were carried out
at different measuring heights. It was found that the
values rose with an increasing measuring height and
a decreasing embedding depth. Analysis of variance
was used to assess the influence of the embedding
depth and revealed that the embedding depth had
a significant impact on the measuring values at each
Introduction
Ankylotic anchorage of the implant in bone isconsidered an essential criterion for the success of animplant (Albrektsson et al., 1981; Brånemark, Zarb &Albrektsson, 1985). For a long time, the mobility of animplant was assessed through subjective palpitation.Later, attempts were made to use the dampingcharacteristics for an objective evaluation of the bondbetween the implant and bone.
© 1999 Blackwell Science Ltd 19
measuring height, above 6 mm. Moreover, it was
found that the higher the measuring height, the
higher the measured values and the greater the
differences between the values obtained at the
individual depths. The different measuring devices
had no influence on the measuring results (P J 0·79).
The results of this study suggest that a longitudinal
follow-up of the peri-implant residual bone height
around individual implants is possible. Single
measuring values by themselves do not allow any
conclusions about the prognosis of an implant. The
assessment of the peri-implant bone height through
Periotest® measurements is conceivable only when
a table of damping values taking into account the
physical length of the implant, the embedding depth
and the measuring height for the examined implant
system is available. In cylindrical implants, the head
of the available prefabricated measuring post can be
recommended as a constant measuring point for
further studies, especially when the results are to be
compared with those obtained by other study
groups.
The Periotest® method (Schulte et al., 1983; D’Hoedtet al., 1985) was designed for the objective evaluationof the periodontal health status of natural teeth. ThePeriotest® device consists of a handpiece connected toa unit, which analyses the braking time of a rod appliedto a tooth or an implant surface. The rod, which islocated inside the handpiece and held in low frictionbearings, is accelerated until it reaches a constant speedof 0·2 m/s, which is maintained to compensate forfriction and gravitation until contact is made with the
20 R . H A A S et al.
Table 1. Range of PTV of implant systems
Implant type Author Location Range of PTV Reference
Brånemark Teerlinck et al., 1991 mand 1 max 25/15 21
mand 24/12
Olive & Aparicio 1990 max 20·8/12 13
mand 22/16
Haas et al., 1995b mand 25/15 8
max 28/19
Tricio et al., 1995 mand 26/14 22
max 26/15·5
IMZ Schramm-Scherer 1987 mand 1 max 25/18 15
Haas et al., 1995a mand 27/17 7
max 25/17
ITI Schramm-Scherer 1987 max 1 mand 25/1515
Mericske-Stern et al., 1995 max 1 mand 28/21 12
TPS D’Hoedt & Schulte 1989 max 1 mand 26/18 10
Tubingen Schramm-Scherer 1987 max 1 mand 25/15 15
surface being tested. The deceleration of the rod when
tapping the surface is recorded by a miniature
accelerometer contained in the tapping head. The
information is analysed by the measuring unit and
tabulated as contact time (CT) and given as the
Periotest® value (PTV) (Schulte & Lukas, 1992). Several
reports on measurements carried out in different
implant systems are available (Table 1).
Various authors (Olive & Aparicio, 1990; Cramer,
1992; Schulte & Lukas, 1993; Haas et al., 1995a)
observed PTVs of 1 8 and higher in clinically mobile
implants that had to be removed. Therefore, implants
with a PTV of more than 1 8 should be subject to
further investigation.
Aside from fibrous encapsulation, other factors have
been reported to affect the damping behaviour of
implants:
(i) After an initial rise following second-stage
surgery, the PTV values of an implant show a
tendency to drop, corresponding to increasing
ankylosis of the implant (Schramm-Scherer,
1987; van Steenberghe et al., 1995).
(ii) Maxillary implants show higher values than
mandibular ones because of the different bone
qualities in these regions (Olive & Aparicio,
1990; Buser, Weber & Lang, 1990; Salonen
et al., 1993; Haas et al., 1995a,b; Tricio et al.,
1995).
© 1999 Blackwell Science Ltd, Journal of Oral Rehabilitation 26; 19–24
(iii) An implant may show varying PTVs,
depending on the point in which the
measurements are carried out (Tricio et al.,
1995; Derhami et al., 1995).
The measuring heights used for Periotest®
measurements vary considerably between the
individual study groups and for different implant
systems. For example, measurements of Brånemark
implants were performed at the upper abutment margin
(Teerlinck et al., 1991; Tricio et al., 1995; van
Steenberghe et al., 1995), at the closest point to the
junction (Olive & Aparicio, 1990) or at a height of
3 mm in a gold cylinder. In IMZ implants, measurements
were carried out either at the crown equator or as
close to the gingival margin as possible (Haas et al.,
1995a) or at the measuring post (Schramm-Scherer,
1987). A measuring point located in the middle of the
implant head was used for TPS screws (Schramm-
Scherer, 1987). Some authors do not give the exact
measuring height (Mericske-Stern et al., 1995). Because
different measuring points have been used for different
implant systems and for measurements carried out in
one and the same implant system, a comparison of the
results is not possible. Therefore, no definite statements
can be made about factors influencing the PTVs of
implants.
This experimental pilot study therefore aimed to
examine the impact of the measuring height and the
D A M P I N G B E H AV I O U R O F I M Z I M P L A N T S 21
embedding depth on the Periotest® values of IMZ
implants.
Materials and methods
Ten plasma-flame-sprayed IMZ implants* with a
diameter of 4 mm and a length of 15 mm were used
for the present study. The implants were embedded in
autopolymerizing methyl methacrylate† , each one at a
different depth, i.e. at 1, 2, 3, 4, 5, 6, 8, 10, 12,
and 14 mm, beginning from the tip of the implant.
Therefore, only 1 mm of the first implant was apically
anchored in the resin, whereas the last implant was
fully submerged in the embedding material, except for
a 1 mm portion. The selected embedding material has
an E-modulus of 3353 6 235 N/mm2 and is thus within
the range of 2000 N/mm2 and 4000 N/mm2 of
cancellous bone (Staudter, 1992). The mixing ratio used
to produce the resin block was 2 units powder and 1
unit liquid. The resin was applied in two layers to
guarantee homogenous curing. The curing time was
5–10 min. Prefabricated measuring posts for the IMZ
implant system with a total length of 11 mm above the
upper implant margin were then placed and fixed with
20 Ncm. The resin block was fixed in such a way that
the fixtures were perpendicular to the floor.
Three different Periotest® devices‡ were used during
all experiments and were calibrated individually before
the actual measurement was started. All registrations
were performed by the same investigator. A series of
seven measurements per measuring height and
measuring device was carried out in each implant. The
handpiece of the Periotest® device was held in such a
way that the horizontally swinging rod hit the implant
surface from a distance of 2 mm. Because of the
dimensions of the handpiece, the PTV measurements
began at 3 mm above the resin and then continued at
1 mm increments until the end of the post was reached.
Statistical data analysis
Each measurement was repeated seven times per device
to achieve a greater measuring accuracy and only the
mean value of each series of measurements was used for
further analysis. A repeated measures ANOVA (Crowder &
Hand, 1990) was carried out with the SAS procedure
*Friatec, Friedrichsfeld, Germany.†Technovit 4071, Kluzer, Germany.‡Periotest®; Siemens, Benzheim, Germany.
© 1999 Blackwell Science Ltd, Journal of Oral Rehabilitation 26; 19–24
GLM to analyse the impact of the embedding depth as
well as of the measuring height and the three measuring
devices, which were defined as repeated factors. A
Huynh/Feldt correction ( Huynh & Feldt 1970) was used
because several measurements of the same implant at
increasing heights and with three different devices result
in specific dependencies. Since a completely crossed
design was needed in order to evaluate interactions in
the ANOVA, only measurements carried out at measuring
heights between 3 mm and 12 mm were used.
To assess the impact of the embedding depth at
different measuring heights, additional subgroup
analyses were carried out using repeated measures
ANOVA (Crowder & Hand, 1990). P-values , 0·05 were
considered statistically significant.
Results
The mean values of the PTVs of the three devices
and corresponding standard deviations for the different
embedding depths and measuring heights are listed in
Table 2. All values ranged between –8 and 125·7 PTV.
The lowest values were found near the embedding
margin and in implants embedded at the greatest depth.
The use of the three different devices had no
significant influence on the measuring values (P 5 0·79).
The analysis clearly revealed that the embedding
depth (P 5 0·0005) and the measuring height (P 5
0·0001) had a significant influence on the PTVs. A
linear and quadratic trend was observed with respect
to the measuring height. The subgroup analyses for
each measuring height that were carried out because
of significant interactions (P 5 0·0001), showed that
the embedding depth had a significant influence on
the values in measuring heights above 6 mm. No
statistically significant linear trend was observed below
this measuring height. The deeper the implant was
embedded, the lower the measuring value.
The higher the measuring height, the greater the
differences between the measuring results obtained at
the different embedding depths.
Although the statistical statements are based on the
data obtained for measuring heights between 3 and
12 mm only, Table 2 seems to indicate that these
statements apply also to greater measuring heights.
Discussion
Periotest® measurements have been recommended in
the literature as a valid tool to judge osseointegration
22 R . H A A S et al.
Table 2. Mean periotest values of three devices with regard to different embedding depths and measuring heights for 15 mm IMZ
plasma-flame sprayed implants. The values in parentheses denote the standard deviation
Measuring Embedding depths (mm)
height
(mm) 1 2 3 4 5 6 8 10 12 14
3 26·3 (0·5) 26·3 (0·5) 27·0 (0·0) 28·0 (0·0) 28·0 (0·0) 28·0 (0·0) 25·8 (1·0) 27·3 (0·5) 27·0 (0·0) 27·3 (0·5)
4 26·1 (0·7) 26·6 (0·5) 27·0 (0·1) 28·0 (0·0) 28·0 (0·0) 24·6 (2·8) 25·0 (0·0) 26·0 (0·0) 27·0 (0·0) 27·0 (0·1)
5 26·8 (0·2) 0·3 (6·3) 27·5 (0·5) 27·9 (0·1) 28·0 (0·0) 24·5 (2·7) 25·0 (0·0) 26·0 (0·0) 27·0 (0·0) 27·0 (0·0)
6 0·8 (6·7) 0·4 (6·4) 27·8 (0·2) 22·3 (4·9) 25·6 (4·0) 23·0 (0·0) 23·4 (4·8) 26·0 (00) 26·3 (1·1) 26·0 (0·0)
7 4·3 (0·7) 4·6 (0·5) 24·7 (4·9) 2·6 (1·5) 0·4 (2·6) 23·0 (0·0) 23·0 (0·0) 25·3 (0·5) 24·6 (0·5) 25·6 (0·5)
8 3·4 (0·5) 4·1 (0·7) 2·9 (2·0) 1·8 (1·6) 0·6 (1·5) 21·0 (1·7) 21·6 (1·1) 24·0 (0·0) 24·3 (1·1) 25·0 (0·0)
9 4·2 (1·4) 4·8 (0·7) 3·1 (0·7) 0·0 (0·1) 20·8 (0·2) 21·6 (1·1) 20·5 (0·8) 23·3 (0·5) 23·3 (1·1) 24·3 (1·1)
10 5·0 (0·0) 4·7 (0·2) 1·3 (1·5) 2·6 (2·2) 0·3 (2·3) 0·3 (0·2) 0·3 (1·5) 23·0 (0·0) 23·0 (1·0) 3·0 (0·0)
11 5·0 (1·1) 5·3 (1·5) 4·0 (0·4) 3·0 (2·7) 3·4 (0·5) 1·3 (1·4) 1·8 (1·0) 21·6 (0·5) 21·0 (1·7) 23·0 (0·0)
12 5·6 (1·8) 7·4 (3·6) 4·0 (2·6) 5·9 (1·7) 3·1 (0·2) 2·3 (0·5) 2·5 (0·5) 21·0 (0·0) 20·3 (0·5) 23·0 (0·0)
13 7·8 (1·3) 8·1 (3·0) 5·2 (1·1) 7·5 (2·9) 4·4 (2·1) 3·9 (0·8) 3·3 (1·8) 0·3 (1·1) 0·0 (1·0)
14 11·2 (2·3) 10·7 (3·5) 5·0 (1·0) 9·2 (3·4) 5·3 (0·5) 4·4 (0·5) 3·6 (1·1) 1·0 (0·0) 0·3 (0·5)
15 11·0 (0·5) 11·1 (3·5) 6·9 (2·0) 10·3 (3·3) 5·6 (0·5) 4·9 (0·1) 5·0 (0·1) 1·5 (0·5)
16 12·6 (1·2) 13·5 (3·3) 8·3 (2·7) 11·8 (3·2) 6·2 (0·4) 6·0 (0·8) 5·3 (0·5) 2·2 (0·4)
17 13·9 (2·2) 14·6 (2·5) 8·4 (2·4) 12·4 (2·9) 8·8 (2·4) 6·0 (1·0) 5·6 (1·5)
18 14·8 (2·8) 17·0 (3·7) 11·4 (2·3) 14·9 (3·4) 11·5 (1·5) 7·4 (0·7) 6·0 (0·0)
19 16·4 (3·3) 16·3 (2·4) 12·9 (2·0) 16·4 (3·8) 12·7 (1·5) 8·3 (0·5)
20 18·8 (5·1) 19·9 (2·1) 13·4 (2·8) 17·4 (2·6) 13·4 (0·5)
21 20·5 (4·2) 20·1 (2·3) 17·0 (2·0) 19·4 (1·2)
22 22·4 (3·3) 20·7 (0·3) 18·9 (1·4)
23 23·6 (3·2) 23·4 (0·5)
24 25·7 (1·1)
(Cramer, 1992; Haas et al., 1995a; van Steenberghe
et al., 1995). Values between –7·0 and 17 are considered
normal values. However, because the measuring heights
used by the different authors often vary considerably,
the results obtained in different implant centres cannot
be compared. The aim of this experimental study was
to examine factors that might influence the PTV.
The use of three different measuring devices had no
statistically proved influence on the measuring values.
This is in accordance with the results of other authors
obtained on screw-shaped implants (Derhami et al.,
1995).
Variable measuring heights and embedding depths
were examined to simulate the effects of peri-implant
bone resorption on the damping behaviour of IMZ
implants.
When the findings of this study are transferred to
in vivo conditions, it should be taken into account that
variable bone qualities were not simulated. Moreover,
the examinations were based on the assumption that
the implants are fully ‘osseointegrated’ in the part
covered by the resin.
© 1999 Blackwell Science Ltd, Journal of Oral Rehabilitation 26; 19–24
The results of the study show that the embedding
depth and the measuring height have a significant
influence on the PTV. The measuring results were the
higher, the higher the point at which the measurements
were carried out and the less deep the implant was
anchored. The differences between PTVs obtained at
the same level but in implants submerged into the resin
at differing depths were more pronounced at higher
measuring heights than near the embedding margins.
For example, the mean values of the measurements
carried out at 3 mm were between –8·0 and –5·8,
whereas those obtained at a measuring height of
11 mm were between –3·0 and 5·3 (Table 2). A linear
trend of the individual embedding depth was confirmed
statistically in measuring heights above 6 mm.
The present results confirm the findings of Derhami
who observed that the measuring height had an
influence on the PTVs of screw-shaped implants placed
in different regions of the facial bones of a human
cadaver (Derhami et al., 1995). This fact had not been
stated for IMZ implants so far.
Clinical experience has shown that implants with a
D A M P I N G B E H AV I O U R O F I M Z I M P L A N T S 23
PTV of more than 17 up to 18 can be considered afailure. Table 2 shows that this value corresponds to a6-mm-long implant anchored within bone when themeasurements are carried out 10 mm above the edgeof the implant (measuring height 19). On the otherhand, this critical value is achieved if a 2-mm portionof the implant is still anchored within bone and themeasurements are carried out at the upper edge ofthe implant (measuring height 13, embedding depth2 mm).
The different conclusions about the influence of theimplant portion still anchored within bone drawn bydifferent authors might be due to the fact that theyused different measuring points:
Cramer (1992), for example, found that peri-implantbone resorption had an influence on the PTV whencarrying out measurements at the crown of single-toothrestorations on Tubingen implants. In contrast, Haaset al. (1995a) did not observe any such influence whenthey carried out Periotest® measurements at the sitewhere the implant passes through the mucousmembrane in 392 IMZ implants.
Therefore, a standardized measuring point forPeriotest® measurements of implant mobility is requiredto allow a comparison of the results of different studygroups.
A prefabricated measuring post is available for theIMZ system. The centre of this measuring post is located10 mm above the edge of the implant. In the presentexperimental design, the lowest possible measuringresult for 15-mm-long IMZ implants would be a PTVof –3 (measuring height 11 mm). A PTV of 18, whichhas shown to be the threshold value for normallyfunctioning implants in clinical routine, was obtainedat an embedding depth of 6 mm in the present study.Increasing bone resorption would also affect the apicalperforation of the implant and have a negative effecton possible attempts to restore peri-implantitis. Thismeasuring point therefore seems to be very suitable forthe implant length examined.
In summary, the following conclusions can be drawnfrom the results of this study:
(i) A standardized measuring point, which
constitutes an essential criterion in PTV
measurements, should be established as
reference value for the respective implant
system examined.(ii) Measurements carried out at changing
measuring heights are not suitable to follow-
up an implant.
© 1999 Blackwell Science Ltd, Journal of Oral Rehabilitation 26; 19–24
(iii) Single PTVs do not allow any conclusions
about the state of peri-implant bone resorption
because of a lack of experimental data.
(iv) Standardized measuring sites would allow a
longitudinal follow-up of single implants. The
head of the measuring post is suitable as
constant measuring point for the assessment
of 15-mm-long IMZ implants.
(v) Further experimental PTV measurements will
be needed to obtain analogous tables for all
available implant lengths and diameters of the
different implant systems used throughout the
world.
It is conceivable that single measurements carried
out at a standardized measuring height allow relevant
conclusions about the state of peri-implant bone
resorption if the above criteria are fulfilled.
Acknowledgments
The authors wish to express their gratitude to Dr M.
Mittlbock and Mag. A. Kaider, Department of Medical
Computed Sciences of the University of Vienna, for
their valuable assistance in the statistical data analysis.
References
ALBREKTSSON, T., BRÅNEMARK, P.-I., HANSSON, H.-A. & LINDSTROM, J.
(1981) Osseonintegrated titanium implants. Requirements for
ensuring a long-lasting, direct bone-to-implant anchorage in
man. Acta Odontologica Scandinavia, 52, 155.
BRÅNEMARK, P.-I., ZARB, G. & ALBREKTSSON, T. (1985) Tissue Integrated
Prostheses: Osseointegration in Clinical Dentistry, 1st edn, pp. 11–
76. Quintessence Publishing, Chicago.
BUSER, D., WEBER, H.-P. & LANG, N.P. (1990) Tissue integration of
non-submerged implants. 1-year results of a prospective study
with 100 ITI hollow-cylinder and hollow-screw implants. Clinical
Oral Implants Research, 1, 33.
CRAMER, A. (1992) Periotestwerte der Tubinger Implantate aus
Aluminiumoxidkeramik. Dissertation, 1st edn, University of
Tubingen, Tubingen.
CROWDER, M.J. & HAND, D.J. (1990) Analysis of Repeated Measures,
1st edn. Chapman & Hall, London.
DERHAMI, K., WOLFAARDT, J.-F., FAULKNER, G. & GRACE, M. (1995)
Assessment of the periotest device in baseline mobility
measurements of craniofacial implants. International Journal of
Oral and Maxillofacial Implants, 2, 221.
D’HOEDT, B. & SCHULTE, W. (1989) A comparative study of results
with various endosseous implant systems. International Journal
of Oral and Maxillofacial Implants, 4, 95.
HAAS, R., SABA, M., MENSDORFF-POUILLY, N. & MAILATH, G. (1995a)
Examination of the damping behavior of IMZ implants.
International Journal of Oral and Maxillofacial Implants, 4, 410.
24 R . H A A S et al.
HAAS, R., SABA, M., MENSDORFF-POUILLY, N., SCHIEBEL, H. & MAILATH,
G. (1995b) Einflußfaktoren auf das Dampfungsverhalten von
IMZ- und Brånemark-Implantaten. Zeitschrift fur Zahnarztliche
Implantologie, 11, 15.
D’HOEDT, B., LUKAS, D., MUHLBRADT, L., SCHOLZ, F., SCHULTE, W.,
QUANTE, F. & TOPKAYA, A. (1985) Das Periotestverfahren –
Entwicklung und klinische Prufung. Deutsche Zahnarztliche
Zeitschrift, 40, 113.
HUYNH, H. & FELDT, L.S. (1970) Conditions under which mean
square ratios in repeated measurements designs have exact F-
distributions. Journal of the American Statistic Association, 65, 1582.
MERICSKE-STERN, R., MILANI, D., MERICSKE, E. & OLAH, A. (1995)
Periotest measurements and osseointegration of mandibular ITI
implants supporting overdentures. Clinical Oral Implants Research,
6, 73.
OLIVE, J. & APARICIO, C. (1990) The periotest method as a measure
of osseointegrated oral implant stability. International Journal of
Oral and Maxillofacial Implants, 5, 390.
SALONEN, M., OIKARINEN, K., VIRTANEN, K. & PERNU, H. (1993)
Failures in the osseointegration of endosseous implants.
International Journal of Oral and Maxillofacial Implants, 8, 92.
SCHRAMM-SCHERER, B. (1987) Untersuchungen zum Damp-
fungsverhalten von Metall und Keramikimplantaten. Zeitschrift
fur Zahnarztliche Implantologie, 3, 22.
SCHULTE, W., D’HOEDT, B., LUKAS, D., MUHLBRADT, L., SCHOLZ, F.,
BRETSCHI, J., FREY, D., GUDAT, H., KONIG, M., MARKL, M., QUANTE,
© 1999 Blackwell Science Ltd, Journal of Oral Rehabilitation 26; 19–24
F., SCHIEF, D. & TOPKAYA, A. (1983) Periotest – ein neues
Meβverfahren und Gerat zur Messung der Funktion des
Parodontiums. Zahnarztliche Mitteilungen, 73, 1229.
SCHULTE, W. & LUKAS, D. (1992) The Periotest method. International
Dental Journal, 42, 433.
SCHULTE, W. & LUKAS, D. (1993) Periotest to monitor
osseointegration and to check the occlusion in oral implantology.
Journal of Oral Implantology, 19, 23.
STAUDTER, M. (1992) Verformungsmessungen und Verformungs-
berechnungen an Zahnimplantaten. Diplomarbeit. Mannheim.
VAN STEENBERGHE, D., TRICIO, J., NAERT, I. & NYS, M. (1995) Damping
characteristics of bone-to–implant interfaces. Clinical Oral
Implants Research, 6, 31.
TEERLINCK, J., QUIRYNEN, M., DARIUS, P. & VAN STEENBERGHE, D.
(1991) Periotest: an objective clinical diagnosis of bone
apposition toward implants. International Journal of Oral and
Maxillofacial Implants, 6, 55.
TRICIO, J., LAOHAPAND, P., VAN STEENBERGHE, D., QUIRYNEN, M. &
NAERT, I. (1995) Mechanical state assessment of the implant-
bone continuum: A better understanding of the Periotest
method. International Journal of Oral and Maxillofacial Implants,
10, 43.
Correspondence: Dr Robert Haas, Department of Oral Surgery,
School of Dentistry of the University of Vienna, Wahringerstraße
25a, A-1090 Vienna, Austria.