Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
Experimental Methods for High Pressure VLE Measurement
by
Rajpuri Sai Kiran Singh", Varun Bhalla*, S.NKaul**, VA.Shinde***
*UG Students; ** HOD., Dept. of Chemical Engineering; *** Assistant Professor, Dept.
of Chemical Engineering; Bharati Vidyapeeth University, College of Engineering, Pune.
Abstract:
Accurate measurement of high pressure vapour liquid equilibria is a most
demanding task. The number and variety of approaches adopted atleast to the exacting,
nature of the task of finding the most reliable, accurate and cost effective approach. But
there exists no firm opinion as to which type of equipment is superior. Each type has its
own advantages and drawbacks. The purpose of this study is to summarize in some detail
the equipments and procedures used to measure high pressure VLE. We discuss some of
the dynamic and static methods developed for high pressure VLE measurement.
Introduction:
Deiters and Schneider distinguished between synthetic method and analytical
method for high pressure VLE measurement. In the former method, a mixture of known
composition prepared and its behavior is observed as a function of temperature and
pressure, the problem of analyzing the equilibrium compositions is thereby avoided. In
analytical method overall compositions are not at issue. Temperature and pressure is
adjusted to bring about phase separation and withdrawn samples are then analyzed.
A simpler classification is based on whether either the liquid or vapour, or both,
are circulated through equilibrium chamber. If circulation takes place it is classified as
dynamic or flow method, otherwise it is a static method. [5]
A schematic diagram of the apparatus used in the analytical method is as shown in
the figure lIt consists of the following:
1. An equilibrium cell in which the vapour and liquid phases of the mixture are III
equilibrium.
2. An environment that controls the temperature of the equilibrium cell, e.g. air bath, oil
or water bath etc.
3. A procedure for agitating and mixing the cell contents, e.g. stirrer.
4. A method for sampling the liquid and vapour phases.
5. A means for accurate analysis of the withdrawn samples.
6. Pressure and temperature measuring devices.
PRESSURE: AND TEMPERATUREMEASURING DEVICE
[--~----~--~------~I
VAPOUR SAMPLINGSYSTEM :-:1
SAMPLING SYSTEM
LlOUIQ SAMPLlNG-1SYSTEM
CONTROLLED ENVIRONMENT
Figure 1: Features of typical analytical method.
Static Method for VLE Measurement:
The features of a typical static apparatus are shown in figure 2. An equilibrium
cell is filled with each component of the mixture to be studied. The mass of each
component introduced is determined by weighing. The cell is kept at constant
temperature in an air thermostat and the mixture is stirred with magnet. At equilibrium,
pressure is read and sample of both the phases are withdrawn and analyzed.
The cleaned and degassed cell is weighed. Then the liquid component IS
introduced, cooled and degassed by stirring under vacuum. After degassing, the cell is
weighed to determine the mass of the liquid component. Then the cell is connected to the
filling equipment to introduce the gaseous component. A known amount of gas is
introduced. The exact mass of the gaseous component is determined by weighing the cell
accurately. The filled cell is connected to the apparatus and fixed on a rigid support to
allow easy handling of the sampling valves without constraints on the tuning and welded
parts of the cell. The air thermostat and magnetic stirring are activated to reach the
equilibrium conditions. When the temperature and pressure are constant, the samples are
withdrawn through the sampling valves to perform phase analysis .The pressure inside
the cell is higher than that in chromatographic circuit, and therefore the sample flows
through the stem of the sampling valve. After the sampling valve is closed, the sample is
taken out by the carrier gas upto thermal conductivity detector. [1,5]
CG
Figure 2: Static apparatus for VLE measurement.
Nomenclature:AT-Air Thermostat;CG-Carrier Gas;CO-Coupling;GC-GasCylinder;GLC-Gas Liquid Chromatographic Analysis Setup;PD-Pressure ElectronicDisplay;PT -Pressure Transducer;S-Solenoids;TD- Temperature Electronic Display.Th.&Th2-Thermocouples;VA-Feeding Valve;VS-Sampling Valves.
Dynamic VLE Measurement Using Flow Type Apparatus:
The apparatus is composed of a feed system, preheater, equilibration system and
sampling system. All parts exposed to high temperature are made of SS316. The details of
the apparatus are shown in figure 3.A gaseous component, such as CO2, was cooled and
liquefied in a water heat exchanger. The liquefied component was then pressurized
through a plunger pump and supplied to a buffer tank immersed in a water bath which
allowed the component to be heated to its supercritical temperature. A liquid component
was provided through a non pulsating pump and joined in a tee zone with the gaseous
component supplied from the buffer tank. This mixture was heated when passing through
a preheater tubing fitted with an electric line heater .In another tubing immersed in an air
bath, the temperature of the mixture was adjusted to a desired equilibrium temperature.
Then it flowed into a static mixer and was well mixed in order to ascertain the attainment
of equilibrium state in a short time before the separation of vapour and liquid phases in an
equilibrium cell. The liquid phase flowed out constantly from the bottom of the cell
through a metering valve, while the vapour phase exited from the top of the cell. A small
amount of liquid and vapour phases are taken in the separate samplers and are analyzed
by gas chromatography. [3,4]
Back PressureRegulator -r
v~~~t:P=~+=~ Overflow Line
I I
~
a k Pressure':' Requloror
MV2
Vent®
GasCylinder
5PlungerPump
,-4- Metering ValveiI-t><l- Stop Valve
Graduated@Cylinder Wet Test
ft1eter(1)
Figure 3: Schematic diagram of flow-type experimental apparatus.
VLE Measurement Using Thin Film Evaporator:
The experiment was conducted using aqueous solution of formaldehyde and methanol.
The apparatus used for investigating VLE is a thin film evaporator in which a rotating
coil spreads the liquid feed on the inner surface of a tube surrounded by a heating jacket.
The liquid partially evaporates. The coexisting phases are separated and collected in
vials. Experiments can be carried out at small evaporation ratios and high residence times
so that chemical equilibrium in the liquid phase leaving the apparatus is achieved. The
temperature of the coexisting phases is measured by platinum resistance thermometry and
pressure using differential pressure transducers. Either gas chromatography or wet
chemistry was used for analyzing the composition of the coexisting phases. [2]
Conclusion:
Several procedures and many different types of equipment used for the measurement of
high pressure VLE have been surveyed. The choice of equipment will be influenced by
the systems to be investigated, e.g. expected relative volatilities or possible multiple
liquid phase formation. Some type of equipment may produce data more rapidly but may
be associated with certain drawbacks. Thus proper selection of method and equipment for
the VLE measurement is the most difficult task. This survey of equipment and procedure
should prove useful to those entering the field or to those seeking to improve their present
equipment and procedures.
References:
• Shou-Ming Hwang, Ming-Jer Lee, Ho-mu Lin, New group interaction parametersof the UNIFAC model: Aromatic Methoxyl Binaries, Ind.Eng.Chem, Res. 2001,40, 1740-1747.[1]
• Michael Albert,Baudilio Coto Garcia,Christian Kuhnert,Roger Peschla,GerdMaurer,Vapour-Liquid equilibrium of aqueous solutions of Formaldehyde andMethanol,AICHE Journal,46,August 2000.[2]
• Hiroshi Inomata, Kazuhiro Tuchiya, Kunio Arai, Shozaburo Saito, Measurementof Vapour-Liquid Equilibrium at elevated temperature and pressures using a flowtype apparatus, journal of chemical engineering of Japan,19,1986.[3]
• Armando del Rio,Baudilio Coto,Concepcion Pando,Jaun A.R. Renuncio,Ind. Eng.Chern. REs.2001,40,689-695.[4]
• J.D.Raal, A.L. Muhlbauer, The measurement of high pressure Vapour-LiquidEquilibria: Dynamic Methods, Static Methods, Developments in Chemicalengineering and mineral processing, 2, 1994.[5]