Nearfield IR Spectroscopy:

Enhancing the molecular sensitivity using compressed sensing and other techniques

B. Kästner, P. Hermann, A. Hornemann, G. Ulrich, A. Hoehl,

C. Elster, G. Wübbeler, F. Schmähling

Funded by

Page 2: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

PTB:

• National Metrology Institute (NMI)

• Federal Ministry of Economics and Technology (BMWi)

• 170 Mio. € budget, plus external funding

• Approx. 1300 permanent staff and 550 non-permanent staff including 110 PhD students

• 600 scientific papers per year

Metrology:

• Science and application of correct

measurement

• Traceability of results to the SI through

national standards

• Determination of results with verification of uncertainty

Hermann von Helmholtz

Physikalisch-Technische Bundesanstalt

Page 3: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

Berlin-Adlershof

Adlershofer PTB Labs

3

L. Novotny, S. J. Stranick; Annu. Rev. Phys. Chem. 2006

PTB laboratory at BESSY II

Metrology Light Source (MLS)

Page 4: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

4

IR Spectrometry Group - Activities

Understand interactions of drug with cell membrane

Where? Nanometer scale resolution - nearfield !

→Metrology: quantify/localize drugs in membranes & bacterial cells

Page 5: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

5

IR Spectrometry Group - Activities

Measurement requirements, example:

Antibiotic: Triclosan Escherichia coli

Required sensitivity: 20 molecules in 50 nm voxel

State of the art : Dj = 0.25° , tint 500 s (see below, NeaSNOM 2018)

→ 500 molecules (derived from PTB measurements on triclosan)

Page 6: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

6

Enhancing the sensitivity

o Nearfield IR Spectroscopy using s-SNOM

o Compressed Sensing

o Radiation Sources and Detection

o Tailored AFM Probes

o Conclusion

Page 7: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

7

Nearfield IR Spectroscopy: s-SNOM

Combination of

IR Spectroscopy and Atomic Force Microscopy

B. Kästner et al. ACS Omega 2018

Page 8: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

8

s-SNOM Principle

IR Radiation incidenton metal cone (antenna)

Focus << l

A. Huber et al. Nano Lett. 2008

Detection of radiation scatteredfrom antenna as dressed dipole

→ scattering-type SNOM

Page 9: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

9

Background suppression

o Strong field between tip and

substrate (of high refractive index)

o Field enhancement depending exponentially on the distance

o In tapping mode: periodic emission with higher harmonics

o Background suppression: higher harmonic demodulation

Detection voltage signal nU

Page 10: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

10

Nearfield Spectroscopy

FTn nU U=

nU

P. Hermann et al., Optics Express 2014

Page 11: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

11

Scattered Signal: Interpretation

→Absorption in grazing incidence, see Mastel et al. Appl. Phys. Lett. 2015

Wavenumber

Abs nU

Arg nU

Arg nU

Interpretation:

Page 12: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

12

Reflectivity R forweak oscillator (organics):

For strong oscillator (e.g. SiC):

Scattered Signal: Interpretation

Abs nU Interpretation:

R

Abs nU

R

Abs nU

Measuring Absorption and Reflectivity in one measurement

Page 13: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

13

Enhancing the sensitivity

o Nearfield IR Spectroscopy using s-SNOM

o Compressed sensing

o Radiation sources and detection

o Tailored tips

o Conclusion In collaboration withM. Raschke

Page 14: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

14

What is Compressed Sensing

Without Compressed Sensing -> compresses only the data

y Compression Storage/Transmission

Decompression y

Measurement

Example: digital imaging / JPEG

• DCT of Image y

• Only largest k components are stored

• Decompression before viewing

Page 15: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

15

m Measurements and matrix A (m x n, n>>m)

What is Compressed Sensing

Requirement for Reconstruction:

y must be sparse in a suitable basis, i.e.:

𝑦 𝑡1 ,… , 𝑦(𝑡𝑚)

𝜙1 𝑡 ,𝜙2 𝑡 ,…

𝑦 𝑡 =

𝑖=1

𝐾

൯𝑧𝑖 𝜙𝜋 𝑖 (𝑡

Compressed Sensing -> Measurement itself is compressed

y Compressed Measurement Storage/Transmission

Reconstruction y

min𝑧

||z||0 𝑦 = 𝐴𝑧under

with 𝐴𝑖𝑗=𝜙𝜋𝑗𝑡𝑖 , 𝑖 = 1,… ,𝑚 , 𝑗 = 1,… , 𝑛

Principle:

Page 16: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

16

Compressed Sensing -Example

Application to a 2D map of SNOM spectra:

44 x 41 = 1804 spectra

How many interferometer samples:

Spectral range

Resolution

→ Sweep interferogram over

x = 1600 mm at 1024 samples

Datapoints: 1804 x 1024 1.8 x 106

with typical integration time of 10 ms→more than 5 hours

16 cm −D =

13200 cm −

Page 17: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

17

Compressed Sensing - Example

Sample:

6H-SiC with Graphene

Indent to generate

2D strain:

Strain induced

spectral shift

B. Kaestner, F. Schmähling, A. Hornemann, G. Ulrich, A. Hoehl, M. Kruskopf, K. Pierz, M. B. Raschke, G. Wübbeler, C. Elster, Optics Express 2018

1 in cm −

3Abs U

Page 18: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

18

Compressed Sensing - Example

Complete sampling ofn = 1024 points

Simulate a compressed measurement:

2U

Page 19: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

19

Compressed Sensing - Example

Random sub-sampling ofm = 114 points at xi

n = 1024

→ Only use 1/9th of measurement samples

2U

Page 20: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

20

Compressed Sensing - Example

Reconstruction of

complete interferogram:

( )2 1 2( ),..., ( )T

my U x U x=

A

- the vector of weightsz

( )ij j iA x=- matrix Solve optimization problem:

0min z z

y Az=subject toFourier basis function(in this case)

0z number of nonzero elements of z

2U

Page 21: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

21

Result of reconstruction

Compressed Sensing - Example

Measured

Sub-sampled

Reconstructed

Fourier Transform

2U

2Abs U 2Arg U

Page 22: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

22

➢Out of 1024 basis functions choose the 50 most frequent ones

➢ Fix this sub-set of 50 functions for the whole 2D map

Compressed Sensing - ExampleIncorporation of further prior knowledge:

1. Similar spectral components across 2D map (1804 pixel)

Page 23: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

23

➢The determination of the z-coefficients by fitting to

fixed basis set, i.e. solve

➢Augment by regularization functional which favours

similar values for 8 neighboring pixels

➢Regularization parameter determined by L-curve

principle

Compressed Sensing - ExampleIncorporation of further prior knowledge:

2. Coefficients of basis functions vary smoothly

y Az=

Page 24: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

24

Compressed Sensing - Example

Hypothesis: sensitivity enhancement by measuring 1/9th

samples at 9 fold increased integration time

Page 25: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

25

Enhancing the sensitivity

o Nearfield IR Spectroscopy using s-SNOM

o Compressed sensing

o Radiation sources and detection

o Tailored tips

o Conclusion

Page 26: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

26

Synchrotron radiationAdvantage of synchrotron radiation: ultra-broadband spectrum

Page 27: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

27

Metrology Light Source, Berlin-Adlershof

For comparison: Bechtel et al, PNAS 2014

Synchrotron radiation

Pulsed at 500 MHz, 30 ps

Page 28: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

28

Detection

Requirements: scattered nearfield from tip VERY WEAK!

Judge by NEP:

Signal power for SNR = 1 for ½ sec integration time

Page 29: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

29

Detection

Ultimate limit: shot noise of optical radiation t

2

2p c

h Pf l

D=

D

For monochromnatic light of l = 10 mmAverage power P = 200 mW x ¼ (due to beamsplitter)

1210 W / Hz−

Page 30: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

30

Best experience with

Mercury Cadmium Telluride photoconductive detectors

Manufacturer specifies 12NEP =10 W/ Hz−

Detection

But

Typicalmeasurement range

Novel detector conceptor

Increase signalscattered from tip

~ radiation intensity

in VnU m

P. Hermann et al.Optics Express 2017

Page 31: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

31

Enhancing the sensitivity

o Nearfield IR Spectroscopy using s-SNOM

o Compressed sensing

o Radiation sources and detection

o Tailored tips

o Conclusion

Page 32: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

32

Tailored TipsAntenna model: see Mastel et al. Nano Lett. 2017:

Choose: 4 mm long tip forresonance around l = 10 mm

A A AZ R iX= +

gap AX X− =

Resonance for:

gap gap gap 2

ihZ R iX

D= + = −

Page 33: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

33

Tailored Tips

600 800 1000 1200 1400 1600 1800 2000 2200

0.1

1

10

100 NanoAndMore

Neaspec

FIB

S

4 /

mV

Frequency / cm-1

Result: Measured spectrum on Au surface

But:alignment ciritical; need robust determination of field enhancement

Tailored tip

Gain: almost x 10

Luca Boarino, Giulia April

Page 34: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

34

Field enhancement - determination

E. Pfitzner, X. Hu, H. W. Schumacher, A. Hoehl, D. Venkateshvaran, M. Cubukcu, J.-W. Liao, S. Auffret, J. Heberle, J. Wunderlich, and B. KästnerarXiv1808.10767, 2018.

In collaboration with:

J. Wunderlich, Hitachi Cambridge LaboratoryE. Pfitzner, J. Heberle-group, FU Berlin

Page 35: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

35

Field enhancement - determination

By scanning a magnetic

nanowire:

E. Pfitzner et al., arXiv1808.10767, 2018.

Anomalous Nernst Effect

leads to generation of VT

How to acess field enhancement:

• Tip heats chip surface at its position (x0,y0), map VT(x0,y0)

• Relate VT(x0,y0) – map to temperature profile T(x,y) by comparing

to on-chip heaters and sensors

• Obtain power dissipation underneath tip, and corresponding field

Page 36: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

36

Field enhancement - determination

VT

M

( ) ( )0 0 0 0, , , ,TV x y T x y x y→

How to obtain:

E. Pfitzner et al., arXiv1808.10767, 2018.

Page 37: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

37

Field enhancement - determination

Comsol simulations by Xiukun Xu, PTB

Obtain using external heaters:

Heaters andT-sensors

Magnetic wire

( ) ( )0 0 0 0, , , ,TV x y T x y x y→

ANE ANE 0

ANE 0.054 V/KT

S

TV N M l

y

N

m

m

=

=

Page 38: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

38

Field enhancement - determination

Comsol simulationsby Xiukun Xu, PTB

300 K

325 K

-0.5 0 0.5Tip position y

0 (µm)

Simulation

-0.5 0 0.5Tip position y

0 (µm)

-0.8

-0.4

0

0.4

VT (

µV

)

Power loss at tip apexPloss = ?

Power loss at tip apexPloss = 4 GW/m2

-0.6

0

0.6

VT (

µV

)

Pin = 0.01 GW/m2

Field enhancement: 25

E. Pfitzner et al., arXiv1808.10767, 2018.

Page 39: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

39

ConclusionsStandard s-SNOM ignores any prior knowledge.

Compressed sensing: exploit prior knowledge at measurement stage

• Spectral sparsity

• Spatial smoothness

• Future: Spectra of individual components

→Rapid hyperspectral imaging or sensitivity enhancement

Further strategies for enhancing molecular sensitivity:

• Novel detector concepts

• Tailored Nearfield-AFM tips

Page 40: Enhancing the molecular sensitivity using compressed ...empir.npl.co.uk/metvbadbugs/wp-content/uploads/... · Nearfield IR Spectroscopy: Enhancing the molecular sensitivity using

Peter Hermann, Arne HoehlAndrea Hornemann

Georg Ulrich, Gerhard UlmClemens Elster, Gerd Wübbeler

Marko SchmählingXiukun Xu, Hans SchumacherKlaus Pierz, Mattias Kruskopf

Emanuel PfitznerPiotr PatokaEckart Rühl

Jörg FeikesMarkus Ries

Tobias GoetschGodehard Wüstefeld