Upload
vasudha
View
213
Download
0
Embed Size (px)
Citation preview
PAPER www.rsc.org/loc | Lab on a Chip
Publ
ishe
d on
08
Sept
embe
r 20
09. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 2
2/06
/201
4 14
:26:
35.
View Article Online / Journal Homepage / Table of Contents for this issue
Enhancing DNA hybridization kinetics through constriction-baseddielectrophoresis
Nathan Swami,*a Chia-Fu Chou,*bc Venkatraman Ramamurthyc and Vasudha Chaureya
Received 29th May 2009, Accepted 10th August 2009
First published as an Advance Article on the web 8th September 2009
DOI: 10.1039/b910598k
The enhancement of signal sensitivity through the scaling-down of sensors presents mass transport
limitations that can arrest the sensitivity gains obtained as a result of miniaturization. To alleviate these
limitations, we study the application of constriction-based dielectrophoresis methods to enhance
transport through pre-concentration of target DNA in the vicinity of the diffusion layer of the sensor,
on which capture probe DNA molecules were immobilized. We demonstrate that constriction-based
DEP pre-concentration was not impeded by scaling-down of the sensor, as long as the sensor electrode
was composed of nanostructured edges and was coupled to an equally scaled down insulating
constriction within a microfluidic channel to enhance the focusing effects of the constrictions and
edges. Furthermore, as a result of the high focusing fields, pre-concentration of single-stranded target
DNA occurred in the vicinity of the sensor pad within the relatively high ionic strength buffers required for
DNA hybridization, with minimal degradation of capture probe molecules. Finally, constriction-based
DEP resulted in an almost immediate pre-concentration of target DNA in the vicinity of the sensor
electrode diffusion layer, resulting in a ten-fold enhancement of the DNA hybridization kinetics at target
concentration values down to the sensitivity limit of 10 pM for the sensor platform.
I. Introduction
The role of mass transport limitations in the sensitivity of
biosensor assays, especially for miniaturized sensor arrays, has
been recognized in prior work.1–3 For many detection platforms,
such as electrochemical, surface enhanced Raman spectroscopy
(SERS),4 electrical, resonance frequency shifts, and surface
plasmon resonance5 based methods, miniaturization improves
signal transduction and sensitivity to varying degrees. However,
at successively lower DNA target concentration values, fewer
target DNA molecules are available for diffusion towards the
miniaturized sensor surface for hybridization with surface-bound
capture probe DNA molecules. This slows down binding
kinetics, thereby considerably delaying signal onset and satura-
tion (i.e. complete hybridization of all capture probes on the
sensor surface). Hence, while signals barely above the noise can
be measured for various biomolecular assays at picomolar or
lower target DNA concentration values,6 these assays operate
under mass transport limitations of the analyte rather than in
a signal transduction limit. This suggests that, in order to further
improve sensitivity within practical assay time scales, methods to
direct the transport of target biomolecules towards the sensor
surface are necessary. For protein sensors, improved transport
becomes even more crucial for proteins with �100 kDa molec-
ular weight, which diffuse at a rate which is an order of magni-
tude slower. Furthermore, miniaturization imposes penalties on
aDepartment of Electrical Engineering, University of Virginia,Charllottesville, VA, 22904. E-mail: [email protected] of Physics; Genomics Research Center; Research Center forApplied Sciences; Academia Sinica, Taipei, 11529, Taiwan. E-mail:[email protected] Institute, Arizona State University, AZ, 85284, USA
3212 | Lab Chip, 2009, 9, 3212–3220
passive transport of biomolecules to the sensor surface for
hemispherical and disk electrodes and presents no significant
advantages for transport to nanowire electrodes.2 Hence, there is
a need for active transport methods where miniaturization aids,
rather than aggravates, transport, so that the sensor can
capitalize on improved signal transduction, as well as improved
transport as a result of miniaturization.
The improvement of mass transport through higher flow rates7
or stirring,8 were reported to enhance hybridization kinetics at
nanomolar or higher target concentration values.9 However, no
more than a 3–4-fold enhancement can be obtained for the
typical flow rates supported in a microfluidic channel,2 since
the enhacement of binding kinetics requires the stirring of the
diffusion layer around the sensor electrode, which is difficult to
achieve for surface-based miniaturized sensor assays. Electroki-
netic methods such as electrophoresis have been demonstrated to
direct transport;10 however, while significant pre-concentration
may be achieved, the requirement of a ‘‘permeation layer’’ and
a non-conducting buffer limits its application towards various
assays. Enhanced transport through magnetic methods,11 as well
as through asymmetric electric-field assisted methods have also
been demonstrated, although these have not been performed
within microfluidic systems at sub nano-molar target concen-
tration values.12 In all cases, methodologies for directed trans-
port of analyte biomolecules to the sensor surface to enable
simultaneous DNA pre-concentration, hybridization and real-
time sensing of the enhancement to DNA hybridization kinetics
are a pressing need.
Dielectrophoresis (DEP) has been widely used for the
manipulation, separation and characterization of cells, DNA,
viruses, and colloid particles (at both micro- and nano-scales) in
various microfluidic platforms,13 since it has numerous unique
This journal is ª The Royal Society of Chemistry 2009
Publ
ishe
d on
08
Sept
embe
r 20
09. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 2
2/06
/201
4 14
:26:
35.
View Article Online
advantages. First, DEP utilizes the intrinsic dielectric properties
of the material, and can therefore be applied to a wide category
of materials. Hence, applications of DEP, notably in bioparticle
sorting or manipulation in microdevices, do not require labelling
of the bioparticles under study, although labelling could often
assist in improved characterization of the DEP response of
bioparticles or the efficiency of bioparticle separation. Second, it
is relatively straightforward to implement DEP in microdevices
where high-fields and field gradients may be generated by thin-
film deposited metal electrodes separated by a few microns. The
conventional implementation of metal electrodes-based DEP
(MDEP) for force generation has been extended recently by
using dielectric constrictions or insulating obstacles to generate
localized DEP force fields, called electrodeless or insulator-based
dielectrophoresis (EDEP or iDEP, repectively),14,15 and by the
addition of electrically floating metal electrodes within these
constrictions, to further enhance the electric field gradient at
electrode edges.16 In this paper, these so called ‘‘constriction-
based’’ DEP methods are applied to trap bio-molecules, not at
the drive electrodes, where the voltage may damage them, but at
specially designed constricted insulator and metal edge points,
where electric field gradients and force fields are highest, even
though these points are electrically floating.
From a modelling perspective, the constriction-DEP configu-
ration simplifies the interpretation of the DEP phenomena due to
the symmetry of the system (3D symmetry of the force field
rather than 2D planar symmetry in MDEP). Consequently the
physical picture of the force field is less obscure, since in
comparison to MDEP, electrode polarization and electrochem-
istry occurring at metal electrodes do not complicate the picture.
Our prior work has shown the great promise of constriction-
based DEP technology applied to various functionalities of
a micro-total analysis system for molecular diagnostics, such as
cell trapping and separation, cell lysing, and DNA pre-concen-
tration for detection in a microarray platform.14,17,18 However, in
prior work, these operations were carried out within different
chambers on the chip. Simultaneous DNA pre-concentration
and sensing is a major challenge due to the highly focused field
and current conditions of constriction-based DEP, which can
effect sensor performance due to DNA capture probe degrada-
tion on the sensor electrode at the constriction, Joule heating
associated with the focused current, and the effects of convective
fluid flow induced by the temperature gradients from this local
heating effect. On the other hand, the higher force field due to
constriction-based DEP could enable the trapping of low
dielectric strength biomolecules (such as ss-DNA), even under
the relatively high salt conditions required for DNA hybridiza-
tion, in spite of the lower biomolecular polarizability under
such conditions. Hence, to study the relative enhancement of
DNA-hybridization kinetics due to constriction-based DEP
pre-concentration of single-stranded target DNA molecules in
relatively high salt buffers, the simultaneous pre-concentration,
binding and sensing of these molecules is examined herein.
In this paper, we examine the application of constriction-based
DEP methods towards the directed pre-concentration of target
DNA molecules within a micro-constricted fluidic channel onto
a microelectrode sensor with nanostructured edges, on which
DNA capture probes were immobilized. Electric field simulations
were used to guide the development of device designs by
This journal is ª The Royal Society of Chemistry 2009
understanding the trade-offs in scaling of the sensor electrode
with respect to the constriction size, so that miniaturization was
optimized for the directed transport of biomolecules, as well as
for improved signal transduction. These simulations were used to
develop conditions and designs for the trapping of target DNA
molecules due to constriction-based DEP, by imaging the pre-
concentration of fluorescently-labelled DNA target molecules in
solution. Finally, to measure the enhancement of DNA hybrid-
ization under constriction-based DEP, we used a two-potential
electrochemical probe19 to simultaneously probe, in real-time,
the signal from capture probe DNA molecules immobilized on
the sensor surface and target DNA molecules hybridized to the
capture probes. This platform allowed us to optimize electric-
field conditions, so that pre-concentration of target molecules
was enhanced on one hand, while capture probe degradation due
to the highly focussed fields was minimized on the other hand.
This enabled an accurate measurement of the effect of DNA pre-
concentration on hybridization kinetics. To enable the develop-
ment of a versatile biomolecule trapping methodology that
may be applied to various detection platforms, we aim to develop
constriction-based DEP methods that circumvent the use of
a ‘‘permeation layer’’ to protect against capture-probe degrada-
tion (as required within electrophoresis based assays), and
that can simultaneously carry out DNA preconcentration,
hybridization and sensing in the same high-conductivity elec-
trochemical buffer, without any wash steps. As a result, it is
anticipated that the results from the modelling and experiments
described herein will aid in the development of design rules for
effective implementation of constriction-based DEP methodo-
logies within various biosensor platforms.
II. Materials and methods
2.1. Modelling methodology
The classical DEP theory states that the dielectrophoretic force
arises from the interaction of the induced dipole of a polarizable
object and an external non-uniform electric field (dc or ac).20 The
DEP force is given by:
FDEP ¼ aðuÞE vE
vy(1)
a(u) is the polarizability of the object at the angular frequency u,
and y is the direction of the applied external field, E. For
a spherical object of radius a, the dielectrophoretic force may be
solved analytically in the form:14
FDEP ¼ 2pa33m Re
3*
p � 3*m
3*p þ 23*
m
!VE2 (2)
3*p and 3*
m are the complex permittivity of the dielectric object and
the medium, and E may be replaced by Erms, namely the root-
mean-square of the external field, in an alternating field. In
general, the DEP force results in the transportation of the
polarizable object towards (positive DEP) or away from (nega-
tive DEP) the high field gradient region, depending on the
contrast of the dielectric response between the material and the
immersed medium to the applied field. In an aqueous solution,
positive or negative DEP occurs when the dielectric response of
Lab Chip, 2009, 9, 3212–3220 | 3213
Publ
ishe
d on
08
Sept
embe
r 20
09. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 2
2/06
/201
4 14
:26:
35.
View Article Online
the object is higher (3*p � 3*
m > 0) or lower (3*p � 3*
m < 0) than the
solvent (in this case, water), respectively.
To elucidate the device-level computation of the dielectric
response, we focussed on the interplay of multiphysical param-
eters of a given constriction and sensor design with a set of
experimental inputs. The Laplace equation was solved simulta-
neously with the current continuity equation, which, in the
frequency domain, is V$[s + ju3]Vf. If s [ u3, this reduces to
DC case: V$sVf. Accordingly, the potential, field, current and
field gradient distributions were computed to simulate measur-
able physical parameters that were compared to our experi-
mental observations. Based on this, we aimed to develop a set of
design rules for constriction-DEP applications in microfluidic
systems.
We implemented electric, thermal, and fluidic modules from a
multiphysics computational fluidics dynamics software package
CFD-ACE+ (ESI R&D U.S., Huntsville, AL) using the device
layout in Fig. 1. The layout consists of a single insulator
constriction in a unit cell of 500 � 500 � 5 (xyz) mm. The device
structure of the constriction is formed by insulating plastic
constriction capped by PDMS (poly-dimethyl-siloxane)-coated
cover slide, on a substrate with metal electrodes insulated by
Si3N4 on a substrate of SiO2/Si. The gap of the constriction
and size of the sensor electrode were gradually varied to study
the resulting trends of field focusing factors due to the variation
of these sizes. The electric potential was applied across the
constriction at both ends of the unit cell in the y direction. The
buffer condition was set at 10 mM NaCl in an aqueous solution
with 1 S cm�1 conductivity.
2.2. Experimental methodology
The device structure for the DEP and detection experiments is
shown schematically in Fig. 1. Pairs of platinum electrode arrays
were patterned by photolithographic lift-off methods (10 mm
long across the fluidic channel width and to 100 nm thickness),
and separated by 500 mm to serve as the ‘‘driving electrodes’’ in
Fig. 1, onto which the DEP voltage was applied. Gold electrodes
were vapor deposited at the center of the Pt electrodes to serve as
Fig. 1 Schematic of constriction-based DEP device, showing Pt driving
electrodes and a Au sensor electrode with exposed edges patterned on an
SiO2/Si substrate, with an insulator (PDMS or oxide) defined microfluidic
channel that is constricted at the sensor electrode. DNA capture probes
were immobilized on the sensor electrode, and the target DNA pre-
concentration was characterized by fluorescence probes and DNA
hybridization was measured by electrochemical methods in separate runs.
3214 | Lab Chip, 2009, 9, 3212–3220
the sensor electrodes for DNA immobilization, and these were
varied in size from 1 to 50 mm as an experimental and model
parameter. A high quality silicon nitride layer was deposited by
PECVD methods to insulate the interconnecting lines of Pt and
Au from the electrolyte, and dry etching was used to only expose
the active areas for Pt and Au electrodes. Special care was taken
in designing etch masks to expose the Au electrode edges of
�50 nm thickness to enable bending of the electric field lines. The
syntheses of DNA capture probes, signaling probes, target
mimics, and ferrocene attachment to DNA for signaling at
different formal potentials have been described in detail in prior
publications.21–23 The deposition procedure for immobilization
of DNA capture probe monolayers involves the incubation of
30 nL drops of the deposition solution consisting of DNA
capture probe and other components of the self-assembled
monolayer in a high-salt buffer for 30 min, as described previ-
ously.19 To enable deposition of different DNA capture probes
on neighboring sensor electrodes to serve as the ‘‘positive’’ and
‘‘negative’’ controls, plasma treatments involving oxygen,
hydrogen, and silane were optimized to enable a contact angle of
35–40� on the insulating silicon nitride layer, adjacent to the
clean and hydrophilic gold sensor electrode surface. After
capture probe DNA spotting, the substrates were placed in
a high-humidity chamber for 30 min, and no significant evapo-
ration or accumulation was observed. The substrates were then
rinsed in de-ionized water, dried under a stream of nitrogen and
stored in nitrogen-filled foil-lined polythene bags until use.
PDMS spin-casting on a silicon master was used to fabricate 100,
200 and 500 mm microfluidic channels, with constrictions of
variable sizes. The channel depth was estimated to be 5 � 1 mm.
After a short oxygen plasma treatment cycle, the PDMS layer
was bonded to the device substrate, so that the constriction was
partially on top of the etched silicon nitride opening on the Au
sensor electrode. The variability of the PDMS constriction sizes
were estimated to be within �1 mm, however, upon bonding to
the substrate, an additional �1 mm variation was observed in
some cases. The mis-alignment of the constriction from the
center of the sensor electrode was estimated at �1 mm.
Following DNA immobilization on Au sensor pads and device
assembly with the microfluidic channel with constrictions, AC
fields of 10–200 Vpp cm�1 at 100–1000 Hz frequency were
applied to the Pt electrodes. Initial experiments to observe pre-
concentration at the constriction points using fluorescence
measurements were performed on dye-intercalated (YOYO-1)
T5 double stranded DNA (38 kbp) at 0.1 mg mL�1 concentration
in 0.5 � TBE buffer with varying degrees of salt concentration
(1 to 50 mM). After optimizing pre-concentration conditions,
we used asymmetric PCR techniques (1 : 10 ratio of
forward : reverse primers) to obtain fluorescein-labeled single-
stranded model DNA targets (�200 bases) with a complemen-
tary sequence of 20 bases matching the capture probe DNA
sequence. The experiments on simultaneous pre-concentration,
hybridization and electrochemical sensing, were carried out in
a buffer containing target DNA of the appropriate concentration
(varied from 100 nM to 10 pM), with 50 mM sodium chloride,
and added components of 1 mM mercaptohexanol (to keep the
DNA monolayer intact), 10% v/v fetal calf serum, and 10 mM
trizma buffer to adjust pH to 6.5. Based on prior work, we
optimized the capture probe concentration on the Au sensor
This journal is ª The Royal Society of Chemistry 2009
Publ
ishe
d on
08
Sept
embe
r 20
09. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 2
2/06
/201
4 14
:26:
35.
View Article Online
surface at �2 � 1013 molecules cm�2, where hybridization
efficiency was highest, for a large range of target concentration
values.19
III. Results
Field focusing due to dielectric constrictions
To optimize the force fields during DEP pre-concentration, the
electric potential and field distribution around a single dielectric
constriction (i-DEP configuration), without gold sensor elec-
trode edges, was modeled in Fig. 2(a). This shows that the electric
field (the derivative of the potential) was highly focused at the
constriction. To ensure a high field-focusing factor we set the
channel width constriction from 500 to 1 mm, within a few
microns of channel length.
Field focusing due to dielectric and metal-edge constrictions
Fig. 2(b) shows the field distribution around a single constriction
trap, in the presence of an additional metal electrode edge
(Au sensor electrode) to enhance bending of the fields, and the
normalized DEP force: V(E2)/2. In absence of the metal edge, the
focusing field was peaked only at the tips of the constriction trap
and decreased towards the centre of the trap due to attenuation
of the field. As a result of the metal-edge in addition to the
constriction trap, the focusing field was peaked not only at the
constriction tips, but also towards the edges of the sensor elec-
trode; thereby enhancing the extent of focusing.
Constriction size versus sensor size
It is apparent from Fig. 2(a) that a narrower insulator constric-
tion is preferred since this would enhance the force fields as per
eqn (2). However, an understanding of the relationships and
spatial extents arising from the focusing effects of the insulator
constriction versus the sensor electrode edge is necessary to
develop designs that optimally exploit both methodologies for
local enhancement of the force fields. For this purpose, we
coupled Au sensor electrodes of four different sizes (1, 5, 10 and
20 mm in diameter) to a 1 mm insulator constriction to simulate
focusing fields of the constricted DEP device at the plane of the
Fig. 2 Electric field distribution around a single i-DEP trap: (a) without
Au sensor edge; and (b) with Au sensor edge. In (a), the field focusing
occurs due to the dielectric constriction of the trap, while in (b) the sensor
edge and the tip of the constriction trap are responsible for the field
focusing.
This journal is ª The Royal Society of Chemistry 2009
sensor electrode edge and mid-way through the microfluidic
channel. As shown in Fig. 3(a) for profiles at the plane of the
sensor electrode edge, the focusing fields (electric field and field
gradient) for the larger sized Au sensor electrode edges were
greatly attenuated as was apparent from the double peaks cor-
responding to the respective electrode edges for each electrode
size. As the electrode size was decreased from 20 to 1 mm, the
DEP force increased about 100 times. Additionally, as the Au
sensor electrode size was decreased, the two focusing field peaks
due to the edges become closer, resulting in a force field that was
more focused towards the center of the constriction. Hence, for
the 1 mm sensor electrode, force fields due to the insulator
constrictions and metal edges merged, but the broad peak with
shoulders correspond to the continued presence of the metal-
edges on the field. Fig. 3(b) shows the electric field profile above
the plane of the electrodes, at a height corresponding to the
Fig. 3 (a): Electric field profile by finite volume analysis at the plane
of the sensor electrode, for a 1 mm dielectric constriction coupled to four
different Au sensor electrode sizes (Au pad on the channel floor was
centered at 250 mm between a pair of Pt driving electrodes, 500 mm apart
in a channel of 500 mm width and 5 mm height). For the 1 mm sensor
electrode, shoulders corresponding to the continued presence of the
metal-edges on the field were apparent. (b): Electric field profile by finite
volume analysis at the mid-way plane of the microfluidic channel, for
a 1 mm dielectric constriction coupled to four different Au sensor elec-
trode sizes (Au pad on the channel floor was centered at 250 mm between
a pair of Pt driving electrodes, 500 mm apart in a channel of 500 mm width
and 5 mm height). The inset shows an expanded view of the E-field at the
1 mm Au sensor electrode, to show shoulders (indicated by arrows)
corresponding to the sensor edge points.
Lab Chip, 2009, 9, 3212–3220 | 3215
Fig. 4 (a) Bright field image of a 1 mm constriction centered on a 50 mm
Au sensor electrode that is dry-etched to expose Au edges through a 1 mm
hole in the Si3N4 dielectric. (b) Fluorescence image of (a), showing the
initiation of pre-concentration of fluorescently tagged DNA within a few
seconds of E-field; (c) fluorescence image of (a), showing pre-concen-
tration of fluorescently-tagged DNA under field (100 V cm�1, 500 Hz,
2 min); (d)–(f) are similar to (c), with Au electrode sizes of 5 mm (d), 20 mm
(e), 40 mm (f). Note that in all these cases the preconcentration of double-
stranded DNA was carried out in the absence of salt in the buffers.
Publ
ishe
d on
08
Sept
embe
r 20
09. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 2
2/06
/201
4 14
:26:
35.
View Article Online
midpoint of the channel. Since the 1 mm constriction extended
through the entire thickness of the microfluidic channel, its
focusing field was observed at the mid-way plane of the micro-
fluidic channel, whereas the focusing field of the sensor electrode
edges was localized to its surface plane only. Hence, for a 1 mm
constriction, while the focusing fields from the 5 mm and 1 mm
sensor electrode edges were somewhat comparable at the plane of
the sensor (Fig. 3(a)), these focusing fields were far higher for the
1 mm sensor electrode edge when measured mid-way through the
microfluidic channel (Fig. 3(b)). It may be noted, however, that
the peak width for the focusing effect from the 1 mm constriction
coupled to the 1 mm sensor electrode was broadened at the floor
of the channel (Fig. 3(a)) due to the greater role of sensor edges
and was sharper midway through the channel (Fig. 3(b)) due to
the dominant role of the constriction, even though small shoul-
ders were also apparent in the latter case (inset of Fig. 3(b)). This
suggests that, for the purposes of field focusing at the floor of
the channel, such as would be necessary for a surface bound
DNA assay at the sensor electrode, the optimal combination of
focusing fields from the insulator constriction and metal edges
can be obtained by coupling a 1 mm constriction to a 5 mm
electrode edge. On the other hand, to extend the focusing field
through all planes of the microfluidic channel, the 1 mm
constriction should be coupled to a 1 mm electrode edge. In other
words, for an extremely narrow insulator constriction (1 mm), an
equally small Au sensor electrode size (1 mm) resulted in the
greatest field enhancement, but this occurred in a non-uniform
manner at the center of the sensor electrode. Alternatively, if
highly uniform field focusing at the plane of the sensor electrode
is sought, then the sensor electrode size should be marginally
larger than the constriction. Sensor electrode sizes greater than
5 mm did not result in optimal utilization of the respective field
enhancements. In summary, for optimal field focusing, the sensor
electrode size must be equal to or marginally greater than the size
of the constriction it was coupled to. Hence, miniaturization
of both sensor and constriction sizes aids biomolecule trapping
from microfluidic channels, thereby alleviating transport
limitations.
Experimental verification of pre-concentration
To verify results on the simulated focusing fields, we used
fluorescently labeled double-stranded DNA to follow DNA pre-
concentration for varying electrode sizes (1–40 mm) coupled to
a fixed constriction size (1 mm). DEP conditions of 100 V cm�1,
500 Hz, were applied within the microfluidic channel for 2 min.
From Fig. 4, it is apparent that the strongest degree of pre-
concentration was obtained for the 1 mm constriction coupled to
a 1 mm Au sensor electrode, followed by a significant, yet lesser
level of pre-concentration on a 5 mm Au sensor electrode. Upon
coupling of the same 1 mm constriction to larger sensor elec-
trodes, the focusing effect at the constriction becomes succes-
sively more faint, and much of the pre-concentration was
confined to the electrode edges. Note that the fluorescence signal
from electrode edges under the insulator was not due to DNA
pre-concentration, but due to refraction of light at the metal/
insulator bump. These results reiterate that for optimal combi-
nation of the focusing fields from the insulator constriction and
sensor electrode edge, the sizes of the respective structures must
3216 | Lab Chip, 2009, 9, 3212–3220
be almost equally small, otherwise the enhacement due to the
constriction is not fully realized. In these and other experiments
to characterize the degree of pre-concentration, the fluorescence
intensity of DEP-concentrated target molecules (after back-
ground subtraction for intensity from buffer, substrate, and
electronic noise) was compared that from non pre-concentrated
target molecules at equilibrium.
DNA pre-concentration in high-ionic strength buffers
While the pre-concentration data in Fig. 4 were acquired in the
absence of salt in the buffer, biomolecular sensors are usually
based on DNA hybridization assays, and these require a rela-
tively high salt concentration in the buffer (�50 mM NaCl).
Hence, in the next set of experiments we aimed to optimize the
appropriate AC field and frequency conditions for DNA pre-
concentration under varying salt concentration of buffer. The
pre-concentration of double-stranded DNA as a function of
constriction size and for different salt concentration values in the
buffer was studied in Fig. 5(a) and (b). While a good degree of
DNA pre-concentration in relatively short times (�1 min) was
obtained in buffers free of salt (Fig. 4) with 100 V cm�1 fields of
500 Hz frequency, higher field strengths were required for pre-
concentration from 2 mM and 50 mM salt containing buffers. In
the latter case, a field of �200 V cm�1 was required for
a comparable degree of pre-concentration within a minute.
However, signals from the immobilized DNA capture probes
(obtained by the two-potential electrochemical characterization
of capture probe) at the constriction were degraded due to the
highly focused fields at 200 V cm�1. Hence, we identified field
conditions of 140–150 V cm�1 at 500 Hz, where a reasonable level
of pre-concentration could be obtained within �5 min, with
minimal capture probe degradation. Upon application of these
field conditions for up to 60 min, the pre-concentration degree
This journal is ª The Royal Society of Chemistry 2009
Fig. 6 Pre-concentration of 1 nM single-stranded target DNA from high
ionic strength buffers (50 mM NaCl) using fields of 150 V cm�1 at 500 Hz,
and enhanced using a 1 mm constriction coupled to a 5 mm sensor
electrode edge, measured after 30 min for indicated channel widths.
Fig. 5 Fluorescence signals of preconcentration of double-stranded
DNA at various PDMS constricted microfluidic channels for 2 and
50 mM NaCl salt concentration values in buffer. The 500 and 200 mm
channels were constricted to 10, 5 and 1 mm, respectively, and images
were acquired using AC fields of 140 V cm�1 at 500 Hz, for 5 min.
Publ
ishe
d on
08
Sept
embe
r 20
09. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 2
2/06
/201
4 14
:26:
35.
View Article Online
was substantial. However, for higher pre-concentration times
�15–20% capture probe degradation was observed, with even
higher degradation at the higher ionic strength buffers, possibly
due to electrolysis. The results in Fig. 5 show significant pre-
concentration within 5 min from buffers with 2 mM NaCl for all
the three constriction sizes. The fluorescence signals were peaked
at the tips of the constriction for the 10 mm constriction, and the
peaks begin to merge for the smaller 1 and 5 mm constrictions.
Furthermore, no clear differences between the two channel sizes
(500 and 200 mm) were apparent. With 50 mM NaCl in the buffer
which is closer to the condition for DNA hybridization, the
utility of the narrower constriction from the larger channel for
net enhancement of DNA trapping per unit area of sensor was
apparent. In this case, the 1 mm constrictions from the 500 mm
channel showed the highest fluorescence intensities over all the
other cases. It should be noted that our observations of lower
fluorescence during DEP pre-concentration from high salt
buffers could arise from either a lower DNA polarizability in
conducting buffers or due to salt concentration dependent
lowering of the fluorescence intensity.
Fig. 7 Enhancement of hybridization kinetics due to constriction DEP
pre-concentration (DEP) applied for 30 and 60 min to various concen-
trations of target DNA, as compared to hybridization under passive
diffusion (No DEP). Inset has the same axes plotted over a smaller range
to show the enhancement down at 10 pM target.
Pre-concentration of single-stranded target DNA under high
ionic strength conditions
To further simulate the conditions of DNA hybridization assays, we
studied the DEP pre-concentration of single-stranded target DNA
from within high ionic strength buffers. Asymmetric PCR was used
to synthesize fluorescein-labeled single-stranded model DNA
targets of 200 bases. Electric field strengths of�150–200 V cm�1 at
500 Hz were required for the onset of pre-concentration. Using
fields of 150 V cm�1 at 500 Hz that were tested previously for
minimal capture probe degradation, a substantial degree of
pre-concentration was obtained after 30 min of the E-field. It is
apparent from Fig. 6, that the pre-concentration from 500 mm
channels constricted to a 1mm trap coupled to a 5 mm sensor
electrode edge was greater than that obtained using similar
conditions from 100 or 200 mm channels. Based on these fluo-
rescence results, to screen for device designs and field conditions
under which a considerable degree of DNA pre-concentration
could be obtained from high-salt buffers without significant
capture probe degradation, subsequent constriction-based DEP
pre-concentration experiments were carried out in 50 mM NaCl
buffers, by applying fields of 150 V cm�1 at 500 Hz, that were
This journal is ª The Royal Society of Chemistry 2009
focused from a 500 mm channel to a 1 mm constriction that was
coupled to a 5 mm sensor electrode (the 5 mm sensor electrode was
used instead of a 1 mm electrode to allow for easier alignment of
the sensor electrode to be within the center of constriction and
thereby obtain an array of working sensor electrodes).
Enhancement of DNA hybridization kinetics
The two potential electrochemical platform was applied to probe
the enhancement of DNA hybridization kinetics due to DEP pre-
concentration, since signals from immobilized DNA capture
probe and hybridized DNA target molecules could be quanti-
tatively detected simultaneously, in real-time.19 This allowed us
to screen out the pre-concentration conditions (electric field,
frequency, time, type of buffer, etc.) that may cause degradation
of the capture probe. Furthermore, the signals arise primarily
from hybridized target DNA molecules rather than from
pre-concentrated unbound or non-specifically bound target
molecules. Signals from this sensor platform displayed a three-
orders-of-magnitude dynamic range. Since capture probe
concentration on the Au sensor surface was �1013 molecules
cm�2, where hybridization efficiency was highest, the signal at the
saturation point was obtained from �1013 bound target mole-
cules cm�2, and signal at the detection limit was obtained from
�1010 bound target molecules cm�2. Hence, for a 5 mm sensor
pad, a signal was distinguishable from as few as �100 bound
target molecules on the surface. Fig. 7 shows the signals from
Lab Chip, 2009, 9, 3212–3220 | 3217
Publ
ishe
d on
08
Sept
embe
r 20
09. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 2
2/06
/201
4 14
:26:
35.
View Article Online
DNA hybridization under passive diffusion conditions (labelled
‘‘No DEP’’) for various target concentration values from 100 nM
down to 10 pM, in comparison to that obtained for DNA
hybridization under active DEP pre-concentration for 30 and
60 min. Considering the case of hybridization under passive
diffusion, while the signal from 100 nM target DNA saturates
within 30 min and that from 10 nM target DNA saturates within
an hour, target DNA concentration values of 1 nM and lower
run into diffusion problems that considerably delay signal satu-
ration. The signal from 1 nM target saturates after 4 h; that from
100 pM target reaches close to saturation after 10–12 h; and for
the case of 10 pM, the signals do not saturate in the time frame of
the experiment (the monolayer degrades after 16 h in the buffer
solution). Degradation of the signals from their saturation value
was observed upon repeat scanning due to capture probe
degradation. The target DNA concentration range for the DEP
pre-concentration experiments was chosen to be in the sub-
nanomolar range (10 pM to 1 nM), since the concentration
within this range was low enough to slow diffusion but high
enough to possibly reach signal saturation with DEP pre-
concentration. Hence, the time necessary to obtain signal satu-
ration could be used to judge the effect of DEP pre-concentration
on DNA hybridization kinetics. The constriction-based DEP
pre-concetration was applied with microfluidic channels con-
stricted from 500 mm to 1 mm on 20 mm and 5 mm sensor pads.
The DEP pre-concentration was applied for 30 and 60 min only,
since prolonged DEP and/or repeated electrochemical scans
caused some degradation of capture probe signals which posed
problems in signal comparison and quantification. Considering
the case of 1 nM target, it is apparent that, as a result of DEP
preconcentration, the hybridization kinetics were improved to
raise signal values to an equivalent level as the respective values
for 10 nM target DNA without DEP pre-concentration (only
passive diffusion). Similarly, signals from 100 pM target under
DEP pre-concentration were raised to an equivalent level as the
respective values from 1 nM target DNA without DEP pre-
concentration (only passive diffusion). Furthermore, from
a comparison of DEP preconcentration of 1 nM target on 20 mm
versus 5 mm sensor pads, it is apparent that the signals were
higher on the latter as would be expected from the pre-concen-
tration results within Fig. 3 and 4. These further re-affirm the
observation that a smaller sensor size was optimal for DEP pre-
concentration as long as the insulator constriction size was
considerably smaller. A final observation of interest from Fig. 7
was that while signals from DEP pre-concentration on 100 pM
target after 30 and 60 min were equivalent to the expected signal
value at 1 nM without DEP preconcentration, and similarly
(from inset of Fig. 7), the signal from 10 pM target pre-
concentrated under DEP for 30 min seems comparable to that
obtained from 100 pM without DEP pre-concentration, the same
signals for 10 pM target after 60 min of DEP pre-concentration
were significantly lower than the expected signal value at 100 pM
without DEP preconcentration. We attribute this difference to
the depletion of target DNA molecules within the microfluidic
channel at the longer pre-concentration times for the lower
target concentration values (10 pM or lower). A flow cell DEP
preconcentrator would alleviate this problem. In summary, the
results indicate that the role of DEP preconcentration is to
almost immediately increase the target DNA concentration by at
3218 | Lab Chip, 2009, 9, 3212–3220
least ten-fold in the vicinity of the sensor pad’s diffusion layer,
and the kinetics of the ten-fold pre-concentrated target were
followed.
IV. Discussion
Versatility of constriction-based DEP pre-concentration method
From the E-field simulation results, it is apparent that target
DNA pre-concentration through constriction-based DEP was
not impeded by scaling down the sensor. In fact, scaling down
sensor sizes for the purposes of improved signal transduction was
not detrimental to constriction-DEP based analyte transport as
long as the sensor electrode was composed of nanostructured
edges and was coupled to an equally scaled down insulator
constriction in a microfluidic channel to enhance the focusing
effects of the constrictions and edges. As a result of these higher
force fields due to constriction-based DEP, the trapping of low
dielectric strength biomolecules (such as ss-DNA), even under
the relatively high salt conditions required for DNA hybridiza-
tion was demonstrated in spite of the lower biomolecular
polarizability under such conditions. Furthermore, since target
DNA trapping does not take place at the driving electrodes,
where the voltage may damage them, but at specially designed
constricted insulator and metal edge points where electric field
gradients and force fields were highest, the degradation of
capture-probe DNA immobilized on sensor electrodes at the
constriction can be minimized since these electrodes are electri-
cally floating during DEP. This eliminates the need for a perme-
ation layer above the sensor electrode to arrest capture probe
degradation. As a result of these factors, constriction based DEP
can be applied towards analyte pre-concentration within a wide
variety of sensor platforms. Local heating and convective flow
due to the ensuing temperature gradients during constriction-
based DEP do not arrest target DNA pre-concentration. These
characteristics attest to the versatility of the constriction-based
DEP pre-concentration method.
Applying DEP pre-concentration to enhance DNA hybridization
kinetics
While target DNA pre-concentration under electric fields has
been demonstrated previously within various studies, its appli-
cation in conjunction with DNA hybridization and sensing has
proved to be a significant challenge. First, the optimal ionic
strengths of electrolytes for DNA pre-concentration do not
match those required for DNA hybridization and sensing.
Second, the force fields causing target DNA pre-concentration
can cause capture probe DNA degradation on the sensor. Third,
the three-dimensional configuration of the target DNA mole-
cules pre-concentrated under electric field may not be suitable for
hybridization with immobilized DNA capture probe molecules.
In this study, we get around the first limitation by enhancing the
force fields for DNA pre-concentration through the use of
constrictions and edges, so that a significant degree of pre-
concentration may be obtained within the high ionic strength
buffers required for DNA hybridization and sensing. Further-
more, the constriction-based DEP format was optimized to
screen out the pre-concentration conditions that cause capture
probe degradation to address the second limitation. Finally,
This journal is ª The Royal Society of Chemistry 2009
Publ
ishe
d on
08
Sept
embe
r 20
09. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 2
2/06
/201
4 14
:26:
35.
View Article Online
through the use of AC fields for pre-concentration at constriction
points rather than at the driving electrodes, we anticipate that the
three-dimensional configuration of pre-concentrated target
DNA molecules were not too constrained to participate in DNA
hybridization. However, it must be noted that to optimize DEP
pre-concentration to address each of these three limitations, the
degree of pre-concentration was somewhat reduced from its
highest level (�1000-fold) to enable the simultaneous testing of
target DNA pre-concentration, hybridization and sensing.
Degree of enhancement to DNA hybridization kinetics
In order to quantify the degree of enhancement to DNA
hybridization kinetics through DEP pre-concentration, we used
the two-potential electrochemical sensing platform. At the
outset, we realize that while nanowire and cantilever sensor
platforms may display higher signal sensitivity, this particular
electrochemical sensor platform was chosen due to its nearly
three-orders-of-magnitude dynamic signal range, a large target
concentration range where signal saturation could be obtained
(100 nM down to 100 pM), and the ability to probe the degra-
dation of capture probe DNA on the sensor surface as a function
of the constriction-based DEP field in real-time. These features
enabled a quantitative real-time assessment of the enhancement
to hybridization kinetics due to DEP pre-concentration.
Furthermore, while constriction based DEP for enhanced
hybridization kinetics may be observed for target concentration
down to the sub-picomolar range, we focus this kinetics study
in the 10 pM to 1 nM range, to enable a direct quantitative
comparison to the case of passive diffusion, since the concen-
tration within this range was low enough to slow diffusion but
high enough to eventually reach signal saturation with DEP-pre-
concentration. The enhancement to DNA hybridization kinetics
observed due to DEP pre-concentration in comparison to passive
diffusion occurred due to an almost immediate ten-fold target
DNA pre-concentration in the vicinity of the sensor pad’s
diffusion layer. It should be noted that while much of the prior
work on enhancement of DNA hybridization kinetics using
a similar sensor platform was reported at 10 nM,8,9 we report
a ten-fold enhancement to hybridization kinetics for target
concentration values down to the sensitivity limit of 10 pM for
this sensor platform. In principle, this enhancement should
continue down to even lower target concentration values, but the
sensitivity limitations of our platform prevented us from
reporting it here. Furthermore, since the sensor covers only the
floor of this constricted volume, we report only a ten-fold
enhancement in Fig. 7, whereas Fig. 6 shows a far greater pre-
concentration factor since it integrates signals from the entire
constricted volume. Hence, sensor methodologies based on the
entire constricted volume would potentially display greater
sensitivity enhancements due to constriction-DEP based
enhancement of hybridization kinetics. Even considering just the
floor of the constricted volume, it is likely that only a fraction of
the pre-concentrated target molecules are bound to the sensor
surface due to losses arising from capture probe degradation and
three-dimensional conformation of arriving target molecules
under the AC field, resulting in only a limited measurement of the
enhancement of DNA hybridization kinetics. Nevertheless,
based on the enhanced hybridization kinetics observed, in spite
This journal is ª The Royal Society of Chemistry 2009
of limitations of the sensor platform, we envision that constric-
tion-based DEP methods can enhance sensitivities within
a variety of sensor platforms through alleviating transport limi-
tations that arise due to sensor miniaturization.
Conclusions
In this study, we have optimized the application of constriction-
based DEP methodologies to alleviate transport limitations
arising from sensor miniaturization and to enhance DNA
hybridization kinetics. The chief conclusions are:
(1) Constriction-based DEP alleviates transport limitations
without being impeded by scaling down of the sensor, as long as
the sensor electrode was composed of nanostructured edges and
was coupled to an equally scaled down insulator constriction
within a microfluidic channel to enhance the focusing effects of
the constrictions and edges.
(2) Due to the high focusing fields obtained as a result of
constriction-based DEP, the pre-concentration of low polariz-
ability biomolecules such as single-stranded target DNA was
demonstrated within the relatively high ionic strength buffers
required for DNA hybridization, with minimal capture probe
degradation.
(3) Through optimization of the constriction-based DEP device
for simultaneous DNA pre-concentration, hybridization and
sensing, a ten-fold enhancement of the DNA hybridization kinetics
was observed for target concentration values down to the sensi-
tivity limit of 10 pM for this sensor platform, due to an almost
immediate ten-fold enhancement of target DNA pre-concentration
in the vicinity of the sensor electrode’s diffusion layer.
Acknowledgements
N. Swami acknowledges support for this work from NSF CBET
Award 0403963 and NSF EFRI 0736002. C. F. Chou acknowl-
edges support from AZ Prop 301, Academia Sinica and NSC 96-
2112-M-001-024-MY3, Taiwan, ROC.
References
1 J. Bishop, A. Chagovetz and S. Blair, Nanotechnology, 2006, 17,2442–2448.
2 P. R. Nair and M. A. Alam, Appl. Phys. Lett., 2006, 88, 233120,233120: 1–3.
3 P. E. Sheehan and L. J. Whitman, Nano Lett., 2005, 5, 803–807.4 A. G. Brolo, E. Arctander, R. Gordon, B. Leathem and
K. L. Kavanagh, Nano Lett., 2004, 4, 2015–2018.5 T. Rindzevicius, Y. Alaverdyan, A. Dahlin, F. Hook,
D. S. Sutherland and M. Kall, Nano Lett., 2005, 5, 2335–2339.6 Y. W. C. Cao, R. C. Jin and C. A. Mirkin, Science, 2002, 297, 1536–
1540.7 C. Phillips, M. Jakusch, H. Steiner, B. Mizaikoff and A. G. Fedorov,
Anal. Chem., 2003, 75, 1106–1115.8 R. H. Liu, R. Lenigk, R. L. Druyor-Sanchez, J. N. Yang and
P. Grodzinski, Anal. Chem., 2003, 75, 1911–1917.9 R. Lenigk, R. H. Liu, M. Athavale, Z. J. Chen, D. Ganser,
J. N. Yang, C. Rauch, Y. J. Liu, B. Chan, H. N. Yu, M. Ray,R. Marrero and P. Grodzinski, Anal. Biochem., 2002, 311, 40–49.
10 M. J. Heller, A. H. Forster and E. Tu, Electrophoresis, 2000, 21,157–164.
11 R. L. Edelstein, C. R. Tamanaha, P. E. Sheehan, M. M. Miller,D. R. Baselt, L. J. Whitman and R. J. Colton, Biosens. Bioelectron.,2000, 14, 805–813.
12 F. Fixe and et al, Nanotechnology, 2005, 16, 2061–2071.13 R. Pethig and G. H. Markx, Trends Biotechnol., 1997, 15, 426–432.
Lab Chip, 2009, 9, 3212–3220 | 3219
Publ
ishe
d on
08
Sept
embe
r 20
09. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 2
2/06
/201
4 14
:26:
35.
View Article Online
14 C. F. Chou, J. O. Tegenfeldt, O. Bakajin, S. S. Chan, E. C. Cox,N. Darnton, T. Duke and R. H. Austin, Biophys. J., 2002, 83,2170–2179.
15 E. B. Cummings and A. K. Singh, Anal. Chem., 2003, 75, 4724–4731.16 M. Washizu, Electrostatics, 2004, 178, 89–94.17 C. F. Chou and F. Zenhausern, IEEE Engineering in Medicine and
Biology, 2003, 22, 62–67.18 J. Gu, R. Gupta, C.-F. Chou, Q. Wei and F. Zenhausern, Lab Chip,
2007, 7, 1198–1201.
3220 | Lab Chip, 2009, 9, 3212–3220
19 N. Swami, C. F. Chou and R. Terbrueggen, Langmuir, 2005, 21,1937–1941.
20 H. A. Pohl, Dielectrophoresis: The Behavior of Neutral Matter inNonuniform Electric Fields. Cambridge, UK: Cambridge UniversityPress, 1978.
21 C. Bamdad, Biophys. J., 1998, 75, 1997–2003.22 C. J. Yu and et al, J. Am. Chem. Soc., 2001, 123, 11155–11161.23 R. M. Umek and et al, Journal of Molecular Diagnostics, 2001, 3,
74–84.
This journal is ª The Royal Society of Chemistry 2009