Upload
frederick
View
226
Download
4
Embed Size (px)
Citation preview
Prq. N-PsychopharmacoL & BioL Psychtat 1994. Vol. 18. pp. 497-502
Copyrtght 0 1994 Elsevter Science Ltd
Prtnted In Great Brttaln. All riahts resewed
027&5646(94)E0026-D 0278 - 5&k/94 $26.00
EEG SEIZURE DURATION MONITORING OF ECT
MICHAEL LAMBERT and FREDERICK PETTY
Department of Veterans Affairs Medical Center and Department of Psychiatry, University of Texas
Southwestern Medical Center at Dallas, Dallas, TX, U.S.A.
(Final form, June 1993)
Lambert, Michael and Frederick Petty: EEG Seizure Duration Monitoring of ECT. Prog. Neuro-Psychopharmacol. & Biol. Psychiat. 1994, 18(3) 497502.
The use of electroencephalographical monitoring in convulsive therapy has gained widespread acceptance in the United States where devices with this capability are readily available. In other countries, clinical monitoring alone is more common. Discrepancies between electroencephalographic and clinical observation (such as the “cuff method” described below) have been known to exist. Duration guidelines have evolved based on concerns about decreased effectiveness of very short seizures and neurotoxicity of overly long seizures. While these guidelines need further research to verify the assumptions they are based on, they have been recommended for use in clinical decision making. Currently available guidelines do not indicate how to manage disparate results when two measurement methods are used simultaneously. The authors examine the controversy and report the frequency of discrepancies across several guidelines.
Kev Words: electroconvulsive therapy, electroencephalogram
Abbreviations: electroconvulsive therapy (ECT), electroencephalogram (EEG)
Introduction
Electroconvulsive therapy (ECT) is an effective treatment of severe depression and mania.
One important recent technical development has been the addition of
electroencephalographic (EEG) monitoring to ECT devices made in the United States. Prior
to the addition of this capability most clinicians monitored seizure duration by observing the
497
498 M. Lambert and F. Petty
motor movements in a limb that had been isolated from the paralyzing agent by inflating a
blood pressure cuff above the systolic blood pressure (Addersly and Hamilton, 1953). In
current practice, many American clinicians utilize both monitoring methods concurrently.
Clinicians in Great Britain more often rely on clinical assessment alone due to the limited
availability of ECT devices with EEG monitoring capability (Scott et al., 1989).
Generally accepted guidelines for seizure duration are evolving. The most recent
American Psychiatric Association (APA) report recommends that seizures be at least 20 to 30
seconds in duration, but no longer than 180 seconds (American Psychiatric Association,
1990). Since that report, Abrams has re~mmend a more conservative upper limit of 120
seconds (Abrams, 1990). Based on concerns that a very short seizure may be ineffective, it
has been recommended that the patient be restimulated if the duration is less than about 25
seconds (Fink, 1989). If the seizure is prolonged, it is recommended that it be
pharmacologically terminated because of concerns about possible neurologic sequelae
(Abrams, 1990). While the exact duration re~mmendations are largely empirical and many of
the assumptions on which they are based need further research, they do serve to emphasize
the importance of seizure duration monitoring. With different interventions recommended
based on the duration estimate, discrepancies between methods of estimating seizure
duration could have an impact on clinical decision making and possibly outcome and side
effects.
The earliest report comparing EEG and cuff duration found a remarkable 90% agreement
between the two techniques (Fink and Johnson, 1982). Generally, subsequent studies have
shown the cuff measurement averages about 70% of the EEG seizure duration (Liston et al.,
1988) but large variations between the two methods make simple e~rapolation of any single
cuff estimate impossible. In Ries’ (1985) study the period of EEG activity after the cessation
of motor movements varied from 4.4 seconds to 116.2 seconds. Couture et al. (1988) also
found a wide range of variation, stating “variable differences in the two duration estimates
caution against the use of formulae that predict from one technique to the other.” For longer
seizures the discrepancy may be more marked. The cuff method has been shown less
sensitive in detecting prolonged seizures (Greenberg, 1985). McCreadie et al. (1989) only
found serious discrepancies between the EEG duration and clinical observation in patients
treated with unilateral ECT.
In the present study, the authors examined the frequency of discrepancies across currently
recognized thresholds for simultaneously measured EEG and cuff techniques. The data was
examined to see if electrode placement had an effect on the frequency of discrepancies.
Monitoring ECT 499
Four hundred consecutive ECT seizures that had been ~minister~ with a MECTA SR-1
(MECTA Corporation, Portland, Oregon, USA, 1986) under the supervision of experienced
clinicians were retrospectively reviewed. The treatments had been administered to 30
patients (28 males, 2 females) with an average age of 48.5 years. All of the patients were
medication-free during the treatments. The diagnoses included major depressive disorder
(15), bipolar disorder, depressed (9), s~izoaff~ive disorder, depressed (5), and chronic
undifferentiated schizophrenia (I). Treatments were modified in a standard manner with
succinylcholine, methohexital, glycopyrrolate, and 100% oxygen (Fink, 1979). Two hundred
and twenty-six of the treatments were unilateral (dElia position, nondominant hemisphere)
and 174 were bilateral.
Duration estimates for simultaneously measured motor cuff method and single lead EEG
were available. A cross-threshold discrepancy was defined as disagreement between the two
methods across a duration guideline. This data was reviewed for cross-threshold
discrepancies at a lower threshold of 25 seconds, and at upper thresholds of 120 seconds
(Abrams, 1990) and 180 seconds (American Psychiatric Association, 1990).
Results
At the lower threshold of 25 seconds, 21.8% of the treatments had cross-threshold
discrepancies between the two methods. Seizure length measured by EEG averaged 33
seconds longer than the cuff method for the treatments in which this occurred. In every
instance the EEG duration was equal to or exceeded the cuff duration.
Using an upper threshold of 120 seconds, the methods were discrepant in 4% of the
treatments. The cuff method did not measure any seizure as over 120 seconds. For
treatments discordant across this threshold, the EEG duration averaged 101 seconds longer
than the motor method.
Using the APA Task Force upper limit of 180 seconds, only three seizures (0.75%) were
discrepant, by an average of 163 seconds (Table 1).
500 M. Lambert and F. Petty
The EEG estimate was equal to or longer than the motor estimate in all 400 treatments.
The incidence of discrepant results was similar between unilateral and bilateral electrode
placements, (X2 = 1.9, n.s.).
Table 1
Discrepanices Between Two ECT Measurement Techniques
Discrepancy Threshold Unilateral Bilateral Combined (N=226) (N=174) (N=400)
25 Seconds 55 (24.3%) 32 (18.4%) 87 (21.8%) 120 Seconds 8 (3.5%) 8 (4.6%) 16 (4.0%) 180 Seconds 0 3 (1.7%) 3 (.75%)
Total Treatments = 400
For all discrepanices the EEG duration exceeded the cuff duration.
Discussion
Given that a significant percentage of “adequate” seizures by EEG are simultaneously read
as “inadequate” by the cuff method, these results would argue against relying on the cuff
method as the sole determination method. If restimulation of seizures shorter than 25
seconds had been performed on this sample using the cuff method alone, one fifth of the
patients could have received restimulation while the electrical seizure was continuing. When
data from both sources are available, most clinicians utilize the longer of the two
determinations when deciding whether to restimulate or terminate a seizure. For this sample,
that would have been the EEG for all 400 treatments. The ill effects, if any, of unnecessary
restimulation or undetected prolonged seizures cannot be determined except by a prospective
study design.
While this study does not answer the question of which data source is most valid, the
common clinical dilemma is the patient with a short or nonexistent motor seizure. In this
sample the EEG alone could have served as a guide whether or not to restimulate. In the
smaller percentage of patients with prolonged ictal activity, the EEG again appears to be the
more sensitive technique. While the measurement of motoric seizure activity clearly has
limitations in sensitivity and in detecting the true extent of prolonged seizures, in our
experience the single lead EEG has been an excellent tool for measuring seizure response.
In this study, there were no patients who manifested a motoric seizure without a
Monltorhg ECT 501
corresponding EEG activity. The reverse was not true. For these reasons, the routine use of
EEG monitoring is recommended.
The result of this report support those who question the cuff methods role in seizure
duration monitoring (Chesen, 1983; Greenberg, 1985, Couture et al., 1988). The risk of leak
across the cuff of succinylcholine, metabolic exhaustion of the limb muscle in longer seizures,
and non-motoric electrical seizures decrease its utility for accurately measuring seizure
length.
Conclusion
The cuff method may retain a useful and unique role in confirming the bilateral
generalization of unilaterally induced ECT seizures. However, routine EEG monitoring should
become the standard.
Acknowledaments
Supported by the Department of Veterans Affairs and a Department of Veterans Affairs
Research Development Award to Frederick Petty, Ph.D., M.D. Also supported by an NIMH
Research Grant (MH37899) NIAAA (AA07234) and by MHCRC Grant (MH41115) to the
Department of Psychiatry, University of Texas Southwestern Medical Center at Dallas. The
authors thank Dinah Turner-Knight for excellent secretarial support in preparing the
manuscript.
References
ABRAMS R [I9901 Termination of ECT-induced prolonged seizures. Convulsive Therapy &54-58.
ADDERSLY DJ and HAMILTON M [I9531 Use of succinylcholine in E.C.T. with particular reference to its effect on blood pressure. Br Med J 1: 195-I 97.
AMERICAN PSYCHIATRIC ASSOCIATION (19901 The Practice of ECT: Recommendations for Treatment, Training, and Privileging. Washington, DC, American Psychiatric Press, pp. 11.8.2-I 1.8.4.
502 M. Lambert and F. Petty
CHESEN ES [I9831 EEG monitoring of ECT preferred to cuff method. Am J Psychiat 140 (letter): 1648-l 649.
COUTURE LJ, LUCAS LF, LIPPMANN SB, SHALTOUT T, PALOHEIMO, MPJ and EDWARDS, HL. [1988] Monitoring seizure duration during electroconvulsive therapy. Convulsive Therapy &206-214.
FINK M and JOHNSON L [1982] Monitoring the duration of electroconvulsive therapy seizures. Arch Gen Psychiatry %:1189-l 191.
FINK M [I 9891 An adequate treatment? Convulsive Therapy 3:31 l-31 3.
FINK M [I9791 Convulsive Therapy: Theory and Practice. New York, Raven Press pp. 105 227.
GREENBERG LB [I 9851 Detection of prolonged seizures during electroconvulsive therapy: A comparison of electroencephalogram and cuff monitoring. Convulsive Therapy _1:32-37.
LISTON EH, GUZE BH, BAXTER LR, RICHEIMER SH and GOLD ME [1988] Motor versus EEG seizure duration in ECT. Biol Psychiatry 394-96.
MCCREADIE RG, PHILLIPS K, ROBINSON ADT, GILHOOLY G and CROMBIE W [1989] Is electroencephalographic monitoring of electroconvulsive therapy clinically useful? Br J Psychiatry -:229-231.
RIES RK 119851 Poor interrater reliability of MECTA EEG seizure duration measurement during ECT. Biol Psychaitry 2094-98.
SCOTT AIA, SHERING A and DYKES S [1989] Would monitoring by electroencephalogram improve the practice of electroconvulsive therapy? Br J Psychiatry =:853-857.
Inquiries and reprint requests should be addressed to:
Michael Lambert, M.D. Psychiatry Service (116A) Veterans Affairs Medical Center 4500 So. Lancaster Road Dallas, TX 75216 (214) 376-5451, Ext. 5474