BARNSTEAD|THERMOLYNE CORPORATION

EASYpure™ LFOPERATION MANUAL

AND PARTS LISTSeries 1052

Model D7383 Model D7381 Model D7382-33100 Volts 120 Volts 230 Volts

LT1052X1 • 3/27/98

Page 2: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

2

Safety Information .............................................................................................................................................. 3 Alert Signals ..................................................................................................................................................... 3 Warnings .......................................................................................................................................................... 3Introduction ........................................................................................................................................................ 5Specifications ..................................................................................................................................................... 6 Dimensions and Clearance Requirements. ...................................................................................................... 6 Feedwater Requirements ................................................................................................................................. 6 Product Water .................................................................................................................................................. 6 Electrical Requirements ................................................................................................................................... 6 Declaration of Conformity ................................................................................................................................. 7Unpacking and Installation ................................................................................................................................. 8 Tubing Adapter Installation ............................................................................................................................... 9 Wall Mounting ................................................................................................................................................ 10Initial Operation ................................................................................................................................................ 13 Cartridge Installation and Rinse Up ............................................................................................................... 13Normal Operation ............................................................................................................................................. 15 Water Draw off ............................................................................................................................................... 15 Run and Standby Modes................................................................................................................................ 15Installing Float or Pressure Switch ................................................................................................................... 17Maintenance and Servicing .............................................................................................................................. 18 Cartridge Replacement .................................................................................................................................. 18 Cartridge Removal ....................................................................................................................................... 18 0.2 Micron Filter Replacement ....................................................................................................................... 19 System Sanitization ........................................................................................................................................ 20 Fuse Replacement ......................................................................................................................................... 22 Printed Circuit Board Replacement ................................................................................................................ 23 Cleaning the Resistivity Cell ........................................................................................................................... 26 Shutdown ....................................................................................................................................................... 28Parts and Supplies ........................................................................................................................................... 29Troubleshooting ............................................................................................................................................... 31Figures ............................................................................................................................................................. 33Ordering Procedures ........................................................................................................................................ 34Warranty ........................................................................................................................................................... 39

Table of Contents

Page 3: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

3

Your Barnstead EASYpure LF has beendesigned with function, reliability, and safety inmind. It is your responsibility to install it inconformance with local electrical codes. Forsafe operation, please pay attention to the alertboxes throughout the manual.Important Information

This manual contains important operating andsafety information. You must carefullyread and understand the contents of thismanual prior to the use of this equipment.Water purification technology employs one ormore of the following: chemicals, electricaldevices, mercury vapor lamps, steam andheated vessels. Care should be taken wheninstalling, operating or servicing Barnsteadproducts. The specific safety notes pertinent tothis Barnstead product are listed on thefollowing page.

WarningsTo avoid electrical shock, always:

1. Use a properly grounded electricaloutlet of correct voltage and currenthandling capacity.

2. Do not locate the EASYpure LFdirectly over equipment that requireselectrical service. Routinemaintenance of this unit may involvewater spillage and subsequentelectrical shock hazard if improperlylocated.

3. Replace fuses with those of the sametype and rating.

Safety Information

WarningWarnings alert you to a possibilityof personal injury.

CautionCautions alert you to a possibilityof damage to the equipment.

NoteNotes alert you to pertinent factsand conditions.

Hot SurfaceHot surfaces alert you to apossibility of personal injury if youcome in contact with a surfaceduring use or for a period of timeafter use.

Alert Signals

Page 4: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

4

4. Do not disassemble water lines orremove cartridges where spilled watercould contact equipment that requireselectrical service. Disassembly of waterlines and removal of cartridges will resultin water spillage. Electrical shock hazardcould result.

5. Disconnect from the power supply priorto maintenance and servicing.

To avoid personal injury:

1. Do not use in the presence of flammableor combustible materials; fire orexplosion may result. This devicecontains components which may ignitesuch materials.

2. This device is to be used with waterfeeds only. Sanitizing/cleaning agentsmust be used in compliance withinstructions in this manual. Failure tocomply with the above could result inexplosion and personal injury.

3. Avoid splashing disinfecting solutions onclothing or skin.

4. Ensure all piping connections are tight toavoid chemical leakage.

5. Ensure adequate ventilation.

6. Carefully follow manufacturer’s safetyinstructions on labels of chemicalcontainers and material safety datasheets.

7. Depressurize system prior to openingcartridge access door.

8. Refer servicing to qualified personnel.

SAFETY INFORMATION

Page 5: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

5

The Barnstead EASYpure LF is a line-fed water purification system designed to provide TypeI reagent-grade water with extremely low organic content. It uses a three-stage deionizationprocess combined with a 0.2 micron filter to polish pretreated water (distilled, deionized, orreverse osmosis) to produce water with a resistivity of up to 18.3 megohms-cm and witha total organic carbon content of less than 15 ppb. Water resistivity is continuously sensedby a resistivity cell and displayed on a digital display.

Introduction

Figure 2 EASYpure LF Back

Power Entry Module w/ Fuse Drawer

Pump Interlock Door Latch

Door

Feedwater InletConnection

Figure 1 EASYpure LF Front

Draw-Off Valve Lever

ResistivityMonitor

0.2 Micron Filter w/ Bell

Page 6: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

6

Dimensions and Clearance Requirements.EASYpure LF dimensions - 12 7/8" W X 19" D X 17 1/8" H (32.7 cm X 48.3 cm X 43.5 cm).

Clearances:Sides - 4" (10.2 cm) minimum to allow air flow.Above - 12" (30.5 cm) minimum for removal of the top cover.

Cartridge replacement requires that you be able to access the back of the unit and open the cartridge accessdoor (total depth, unit + open door, = 34") (86.4 cm).

Feedwater RequirementsThe EASYpure LF requires water pretreated by either distillation, deionization or reverse osmosis meeting thefollowing criteria.TOC - Less than 1.0 ppm.Turbidity - 1.0 N.T.U. maximumTemperature - 40°F - 120°F (4.4°C - 48.8°C)(Pressure: Gravity - PSIG)Resistivity (Minimum):

Distilled - 300,000 Ohms-cmDeionized - 1.0 Megohms-cmReverse osmosis - 50,000 Ohms-cm, (100,000 Ohms-cm)

Product WaterQuality

Resistivity: ASTM Type I, less than 18.0 Megohm-cm at 25°CTOC: Less than 10.0 ppb, (15.0 ppb)

Flow Rate: 1.5 LPM at 30 PSIG inlet, feedwater pressure at 60Hz and with a new final filter.

Electrical RequirementsThe EASYpure LF is equipped with a power cord to be plugged into an electrical outlet of the appropriatevoltage.

Model D7383 - 100 VAC +5%, -10%, 47-63 Hz.Model D7381 - 120 VAC +5%, -10%, 47-63 Hz.Model D7382-33 - 230 VAC + 10%, - 10%, 47-63 Hz.

WarningUse a properly grounded electrical outlet of correct voltage and current handling capacity.

Environmental ConditionsOperating: 4°C - 49°C; 20% to 80% relative humidity, non-condensing. Installation Category II (overvoltage) inaccordance with IEC 664. Pollution Degree 2 in accordance with IEC 664. Altitude limit 5,000 meters.Storage:-25°C to 65°C; 20% to 80% relative humidity.

Specifications

Page 7: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

7

SPECIFICATIONS

Declaration of ConformityBarnstead/Thermolyne hereby declares under its soleresponsibility that this product conforms with thetechnical requirements of the following standards(applies to -33 model(s) only):

EMC: EN 50081-1 Generic Emission StandardEN 50082-1 Generic Immunity Standard

Safety: EN 61010-1 Safety requirements forelectrical equipment formeasurement, control andlaboratory use; Part I:General Requirements

per the provisions of the Electromagnetic CompatibilityDirective 89/336/EEC, as amended by 92/31/EEC, andper the provisions of the Low Voltage Directive 73/23/EEC, as amended by 93/68/EEC.

The authorized representative located within theEuropean Community is:

European ManagerBarnstead ThermolyneSaarbrückener Str. 248D-38116 BraunsweigGermany

Copies of the Declaration of Conformity are availableupon request.

Page 8: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

8

1. Remove the unit from its shipping con-tainer. Ensure that the tubing, adapterand power cord are removed from thepackaging materials before discarding.

Note: Cartridges and the 0.2 micron final filter arenot provided with the EASYpure LF and must beordered separately.

Description Catalog #

EASYpure High Purity/Low TOC Cartridge D50229

Pretreatment Cartridge DI Feed D50230

Pretreatment Cartridge R/O - Distilled Feed D50231

Ultrapure Mixed Bed Cartridge 50233

Filter, 0.2 Micron FL703X2

2. Place the EASYpure LF on a bench topconvenient to your work area, noting theClearance Requirements.

3. Tape the 1/4" OD x 1/4" NPT tubingadapter using 1 1/2 to 2 turns of Teflon®

tape, and install it on your incomingwater line or your storage reservoir.

4. Install one end of the tubing to thisadapter, but do not install the other endon the EASYpure LF until cartridgeshave been installed. You will connect thewater supply to the quick disconnectbody in the Initial Operation section.Refer to the Tubing AdapterInstallation section of this manual.

Unpacking and Installation

®Teflon is a registered trademark of Dupont.

NoteIf you purchased a wall mountbracket for the EASYpure LF,refer to the Wall Mounting sectionof this manual.

WarningDo not locate the EASYpure LFdirectly over equipment thatrequires electrical service.Routine maintenance of this unitmay involve water spillage andsubsequent electrical shockhazard if improperly located. Donot use in the presence offlammable or combustiblematerials; fire or explosion mayresult. This device containscomponents which may ignitesuch materials. This device is tobe used with water feeds only.Sanitizing/cleaning agents mustbe used in compliance withinstructions in this manual. Failureto comply with the above couldresult in explosion and personalinjury.

NoteThe outlet of your storagereservoir must be level with theinlet of the EASYpure LF.

Page 9: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

9

Tubing Adapter Installation1. Completely disassemble the fitting.

Refer to Figure 3 to familiarize yourselfwith the names of the componentparts.

2. Make sure the tubing is cut offreasonably square and that no plasticburrs or ridges are present.

3. Place the grab ring and backup ring inthe hex nut in the order and orientationshown in Figure 3. Thread the nut ontothe adapter. DO NOT use the o-ring atthis time.

4. Push the tubing through the nut until itbottoms out in the adapter.

5. Remove the adapter nut and tubing.Place the o-ring over the tubing. Becareful not to push the backup ring orgrab ring further back on the tubingwhen installing the o-ring.

6. Install the hex nut on the adapter andhand tighten.

NoteTo disrupt the flow of water to theEASYpure LF, press the stainlesssteel thumb pad on the quickdisconnect fitting. The insert withthe valve is easily removed. Toresupply the EASYpure LF withwater, push the insert into thecoupling body and listen for theclick that indicates full closure.

UNPACKING AND INSTALLATION

Figure 3Typical Polypropylene Tubing Adapter Installation

Adapter

O-Ring

Backup Ring

Stainless SteelGrab Ring

NutTubing

Page 10: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

10

Wall MountingInstall the wall bracket on the wall in a locationthat is accessible to water and convenient to use.A minimum of 4 fasteners must be used.

1. Remove the pump interlock plug fromthe rear of the unit.

2. Open cartridge access door in the rearof the unit.

3. Remove the cover of the EASYpure LFby removing the screws that secure it tothe unit and lifting it straight up.

4. Remove tubing attached to quickdisconnect body.

5. Remove the inlet quick disconnect bodyfrom the rear of the EASYpure LFcabinet.

6. On both sides of the wall bracket thereare plugs. Remove the plug from theside that is most accessible to yourfeedwater source. Install the inlet quickdisconnect body fitting into the hole byplacing the inlet quick disconnect intothe hole and secure in place with thelocknut inside. Replace the plug in the

UNPACKING AND INSTALLATION

CautionWall composition, condition andconstruction, as well as fastenertype, must be considered whenmounting this unit. The mountingsurface and fasteners selectedmust be capable of supporting aminimum of 150 lbs. Inadequatesupport and/or fasteners mayresult in damage to mountingsurface and/or equipment. If youare unsure of mounting surfacecomposition, condition andconstruction or correct fasteners,consult your buildingmaintenance group or contractor.

WarningDo not locate the EASYpure LFdirectly over equipment thatrequires electrical service.Routine maintenance of this unitmay involve water spillage andsubsequent electrical shockhazard if improperly located.

CautionDo not tighten tube fitting hex nutwith a wrench. Tight connectionscan be easily made by hand.

Page 11: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

11

hole generated in the EASYpure LF(where the disconnect body wasinstalled).

7. Remove the four feet from theEASYpure LF and retain the screws.

8. Place the EASYpure LF on the wallbracket swivel base so the screwholes where the feet were attachedline up with the holes in the wallbracket. There are guides on the wallbracket that will mate with theEASYpure LF.

9. Install the four screws removed in step7 through the bottom of the wallbracket and into the EASYpure LF.

10. Snake the tubing removed in step 4through the hole in the EASYpure LFfloor and the wall bracket and attach itto the quick disconnect body installedon the side of the bracket.

11. Installing feed tubing from the watersource to the quick disconnect bodyon the outside of the bracket cannotbe done until cartridges have beeninstalled. You will connect the watersupply to the quick disconnect body inthe Initial Operation section of thismanual.

12. Replace the EASYpure LF cover.

13. Reinstall the pump interlock plug.

UNPACKING AND INSTALLATION

NoteEnsure there are no kinks in thetubing that could restrict water flow.

NoteOutlet of feedwater storage tankmust be above or at the same levelas the inlet valve on the EASYpureLF.

Figure 4Orientation of Swivel Base

Page 12: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

12

Figure 5Cartridge Installation

UNPACKING AND INSTALLATION

1. Wet o-rings on cartridge.2. Insert upper cartridge end cap into upper socket in keyhole.3. Slide upper cartridge end cap down in keyway.4. Firmly insert lower cartridge end cap into lower socket.

O-rings

FlangesO-rings

Lower End Cap

Upper End Cap

High Purity/Low TOC (D50229) or Ultrapure Mixed Bed (D50233)

Ultrapure Mixed Bed (D50233)

Pretreatment Cartridge (D50230 or D50231)

(Door removed for clarity.)

Page 13: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

13

Cartridge Installation andRinse Up

1. Open cartridge access door in the rearof the unit by pushing the door latchdown.

2. Remove a new Pretreatment cartridge(Catalog No. D50230 or D50231) fromits plastic bag.

3. Wet the o-rings on both end caps.

4. Press the upper end cap into theupper right position until it bottoms out.

5. Lower the cartridge and insert thelower end cap into the lower socketuntil it is firmly seated.

6. Repeat steps 2 - 5 with the EASYpureULTRApure and High Purity/Low TOCcartridges, placing them in the centerand left-hand positions. Do not install0.2 micron filter and bell assembly atthis time.

7. Close cartridge access door.

8. Open draw-off valve.

9. Connect water supply to quick discon-nect body.

10. Plug power cord into the unit’s powerentry module and plug into a liveoutlet. Turn on power entry moduleswitch.

Initial Operation

NoteIf the EASYpure LF is wall mounted,rotate the EASYpure LF until thecartridge access door faces forwardand the EASYpure LF locks intoplace.

NoteThe EASYpure LF’s cartridges mustbe installed in the proper order.

NoteThe upper end cap is the one withthe rightangle turn and the twoflanges. The lower end cap extendsstraight out from the cartridge.

NoteThe two flanges on the end capshould be able to slide down oneach side of the keyway wall.

WarningUse a properly grounded electricaloutlet of correct voltage and currenthandling capacity.

Page 14: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

14

11. Press “START”.

12. Rinse 5-10 liters of water through thecartridges to drain.

13. Remove a new 0.2 micron filter and bellassembly from its bag and insert it intothe Luer fitting. Gently turn it clockwiseuntil it is fully seated in the Luer fitting.

14. Remove the protective cap from the filterbell.

15. Flush 5-10 liters of water through thefilter to waste.

INITIAL OPERATION

CautionDo not allow EASYpure LF tooperate unless water is availableto unit.

NoteFor more demanding applicationswhere low TOC water is required,a rinse of 15-20 liters through thecartridges and filter may benecessary.

Page 15: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

15

1. Press the “START” button on the frontof the EASYpure LF.

The EASYpure LF’s pump will begin to run andthe Purity meter will display the resistivity of thewater in megohm-cm.

2. Allow the water’s resistivity to rise tothe desired purity before drawing offwater.

Water Draw off1. Remove the protective cap from the

filter bell.

2. Depress the draw-off lever.

3. When draw off is complete, lift thedraw-off lever and replace theprotective cap on the filter bell.

Run and Standby ModesSince not all qualities of permissible feedwaterwill reach maximum resistivity after one passthrough the unit’s cartridges (especially as thecartridges near exhaustion), the EASYpure LFhas two operational modes.

Normal Operation

NoteOn initial start-up, the purity metermay display “ERR.” This is causedby air in the cell and should bereplaced by a resistivity readingalmost immediately. If “ERR” doesnot go out after the pump has run fora minute or if it appears any timewhile the EASYpure LF is inoperation, refer to theTroubleshooting section of thismanual.

NoteFor low organic applications, drawoff 50 to 100 ml of water fromsystem and discard prior to drawingwater for use.

Page 16: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

16

In the run mode, the pump continuouslyrecirculates water through the cartridges. This isthe mode that the unit enters upon start-up. Ifwater will be drawn from the EASYpure LF on acontinuous basis throughout the day, it isrecommended that the EASYpure LF be left in therun mode. In the run mode, the purity meterdisplay indicates the resistivity of the wateravailable for draw off.

In standby mode, the pump runs for ten minutesout of every hour (i.e. ten minutes on, fiftyminutes off). If water will be drawn from the unitonly infrequently, it is recommended that the unitbe put into standby mode. To put the unit intostandby mode, press the “STANDBY” button onthe front of the unit. The purity meter’s digitaldisplay will display “Sby” to indicate that the unitis in standby mode.

NORMAL OPERATION

Page 17: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

17

Accessories D0603, D0606 (float switches) andD2706 (pressure switch) are designed to pro-tect the EASYpure LF pump by alerting theEASYpure LF of an inadequate feedwatercondition so that the pump can be shut down.Use the following instructions for installation.

1. Disconnect electrical power.

2. If using D0603 or D0606 float switch,follow the installation instructionsincluded with the float switch forinstallation to tank.

3. If using D2706 low pressure switch,install the PVC tee (supplied withD2706) in incoming water line. Screwthe switch into the top of the tee, thenconnect the inlet tubing to theEASYpure LF with the remainingopening.

4. Route cable from float or low pressureswitch to the rear of the EASYpure LF.

5. Remove jumper plug and save forfuture use.

6. Plug cable into jumper plug outlet.

7. Reconnect electrical cord.Figure 6Float and Pressure Switch Installation

Installing Float or PressureSwitch

Page 18: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

18

Cartridge ReplacementThe frequency with which you will need to replacecartridges is dependent on your feedwater’scharacteristics, your purity requirements and yourusage. Replace the cartridges when the productwater purity drops below acceptable levels ofresistivity or when organic levels become toohigh.

Cartridge Removal1. Turn unit off.

2. Disconnect the unit from the powersupply.

3. Disconnect incoming water line at quickdisconnect. Depressurize system byopening outlet valve.

4. Open the cartridge access door in therear of the unit by sliding the latch downand pulling the door toward you. Thedoor will swing down.

5. Grasp one of the cartridges at the bot-tom and pull it straight up to disconnectthe lower end cap from the lower socket.

6. Move cartridge upward until uppersocket is in keyhole of keyway.

7. Pull cartridge straight out from unit todisconnect upper end cap from uppersocket.

8. Repeat steps for the other cartridges.

Maintenance and Servicing

WarningDisconnect from the power supplyprior to maintenance andservicing. Refer servicing toqualified personnel. Do notdisassemble water lines orremove cartridges where spilledwater could contact equipmentthat requires electrical service.Disassembly of water lines andremoval of cartridges will result inwater spillage. Electrical shockhazard could result.

NoteThe cartridges will still containwater when removed. Therefore,you will want to have a sink,bucket or other waterproofcontainer available to place themin after removal.

WarningDepressurize system prior toopening cartridge access door.

NoteIf the EASYpure LF is wallmounted, remove the power cordand rotate the EASYpure LF untilthe cartridge access door facesforward and the EASYpure LFlocks into place.

Page 19: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

19

9. Discard the used cartridges. (See notebelow.)

10. Install new cartridges and rinseaccording to the instructions forCartridge Installation in the InitialOperation section.

0.2 Micron Filter ReplacementReplace the 0.2 micron final filter whenever anyof the following conditions occur: every 30 days,theproduct water flow rate is reduced or bacteriabreak through. The 0.2 micron filter is shippedassembled with a bell. To replace the 0.2micron filter assembly:

1. Remove the old 0.2 micron filterassembly by turning it counterclock-wise until it is free from the Luer fitting.

2. Remove the new 0.2 micron filterassembly from its bag and insert it intothe Luer fitting. Gently turn it clockwiseuntil it is fully seated in the Luer fitting.

3. Rinse at least 10 liters of waterthrough the filter to drain prior to usingthe product water.

NoteA small amount of water will drainfrom the cartridge when it is discon-nected from the lower socket. Plugthe cartridge’s lower opening withyour finger to minimize water spill-age while you finish removing thecartridge.

NoteUsed cartridges may be recycled.See P.U.R.E. information packedwith new cartridges.

MAINTENANCE AND SERVICING

CautionDo not overtighten the 0.2 micronfilter assembly onto the Luer fittingor use excessive force in seating it.The filter and/or Luer fitting can bedamaged by overtightening orexcessive force.

NoteIf a newly installed 0.2 micron filterclogs rapidly after installation, theEASYpure LF may need to besanitized to remove bacterialcontaminants. See SystemSanitization.

Page 20: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

20

General CleaningWipe exterior surfaces with lightly dampenedcloth containing mild soap solution.

System SanitizationThe frequency with which you will need to cleanyour unit and replace your cartridges isdependent on your feedwater’s characteristics,your purity requirements and your usage.Sanitize your EASYpure LF and replace thecartridges when the product water purity dropsbelow acceptable levels of resistivity, whenorganic levels become too high, or if a new 0.2micron filter clogs rapidly after installation eventhough the cartridges were thoroughly rinsedbefore the 0.2 micron filter was installed. Tosanitize the EASYpure LF, the purificationcartridges must be replaced with a sanitizationcartridge and the two empty cartridges. Thesimple-to-use sanitization cartridge (CatalogNumber D50245) and two empty cartridges(D7034) are available fromBarnstead|Thermolyne or your localrepresentative.

1. Turn the unit off.

2. Disconnect the unit from the powersupply. Disconnect the unit from thewater supply.

3. Depressurize system by opening thedraw-off valve.

4. Open the cartridge access door in therear of the unit by sliding the latch downand pulling the door toward you. Thedoor will swing down.

Maintenance and Servicing

WarningDisconnect from the power supplyprior to maintenance andservicing. Refer servicing toqualified personnel. Avoidsplashing disinfecting solutions onclothing or skin. Ensure all pipingconnections are tight to avoidchemical leakage. Ensureadequate ventilation. Carefullyfollow manufacturer’s safetyinstructions on labels of chemicalcontainers and material safetydata sheets.

NoteThe cartridges will still containwater when removed. Therefore,you will want to have a sink,bucket or other waterproofcontainer available to place themin after removal.

WarningDepressurize system prior toopening cartridge access door.

Page 21: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

21

5. Remove exhausted cartridge in theright-hand position by pulling it first upand then out.

6. Remove a D50245 sanitization car-tridge from its packaging. Press theupper end cap of the D50245 sanitiza-tion cartridge into the upper rightposition until it bottoms out.

7. Lower the cartridge and insert thelower end cap into the lower socketuntil it is firmly seated.

8. Repeat steps 2 - 6 with the two emptycartridges supplied with yourEASYpure LF, placing them in thecenter and left-hand positions.

9. Close the cartridge access door.Remove the 0.2 micron filter and bellassembly.

10. Sanitize, install new cartridges andrinse according to the instructions forCartridge Installation and Rinse Up inthe Initial Operation section.

MAINTENANCE AND SERVICING

NoteThe two flanges on the end capshould be able to slide down oneach side of the keyway wall.

NoteIf the EASYpure LF is wall mounted,remove the power cord and rotatethe EASYpure LF until the cartridgeaccess door faces forward and theEASYpure LF locks into place.

WarningDepressurize system prior toopening cartridge access door.

WarningReplace fuses with those of thesame type and rating. (See PartsListing)

Page 22: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

22

Fuse Replacement1. Turn the system off.

2. Disconnect incoming water line at quickdisconnect. Depressurize system byopening outlet valve.

3. Disconnect the EASYpure LF from thepower supply.

4. Remove the power cord from the powerentry module, and remove the pumpinterlock plug from the rear of the unit.

5. Remove the screws securing theEASYpure LF cover.

6. Remove the cover by lifting it straight up.

7. Pull out the fuse drawer located in thepower entry module.

8. Remove old fuses and replace withfuses of the same type and rating. (SeeParts Listing.)

9. Replace fuse drawer.

MAINTENANCE AND SERVICING

CautionDo not operate the pump dry. Dryrunning will damage the pump.

WarningDepressurize system prior toopening cartridge access door.

Page 23: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

23

10. Replace cover, power cord, and pumpinterlock plug.

11. Reconnect unit to power supply, andconnect incoming water line at quickdisconnect.

Printed Circuit BoardReplacement

1. Turn system off.

2. Disconnect incoming water line atquick disconnect. Depressurize sys-tem by opening outlet valve.

3. Disconnect the EASYpure LF from thepower supply.

4. Remove the power cord from thepower entry module, and remove thepump interlock plug from the rear ofthe unit.

5. Remove the screws securing theEASYpure LF cover.

MAINTENANCE AND SERVICING

WarningDepressurize system prior toopening cartridge access door.

Page 24: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

24

6. Remove the cover by lifting it straight up.

7. Remove the filter and the Luer fitting.

8. Remove the screws securing the frontcover. Remove the front cover.

9. Disconnect the membrane switch leadfrom the printed circuit board.

10. Disconnect the resistivity cell lead fromthe printed circuit board.

11. Disconnect the printed circuit boardrelay plug by squeezing the retainingclip while pulling the plug straight outfrom the frame.

12. Remove the nuts holding the printedcircuit board in place.

13. Remove the printed circuit board.Disconnect power supply (3 wires) fromprinted circuit board.

MAINTENANCE AND SERVICING

WarningReplace fuses with those of thesame type and rating.

Page 25: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

25

14. Reconnect power supply to newprinted circuit board and install thenew printed circuit board. Refer toelectrical diagrams.

15. Reconnect the resistivity cell lead andprinted circuit board relay plug.

16. Reconnect the membrane switch lead.

17. Reinstall the front cover.

18. Re-tape the Luer fitting with newTeflon® tape and reinstall fitting.

19. Reinstall the EASYpure LF’s cover.

20. Reconnect the power cord and thepump interlock plug to the unit, andreconnect the unit to the power supply.Connect the incoming water line atquick disconnect.

MAINTENANCE AND SERVICING

Figure 7Printed Circuit Board

Page 26: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

26

Cleaning the Resistivity Cell1. Turn unit off.

2. Disconnect incoming water line at quickdisconnect. Depressurize system byopening outlet valve.

3. Disconnect the EASYpure LF from thepower supply.

4. Remove the power cord from the powerentry module, and remove the pumpinterlock plug from the rear of the unit.

5. Remove the screws securing theEASYpure LF cover.

6. Remove the cover by lifting it straight up.Remove the filter and Luer fitting. Care-fully remove the front cover. Disconnectmembrane switch lead from the printedcircuit board.

7. Remove the screw holding the cell-cableretaining clip.

8. Disconnect the cell lead from the printedcircuit board and gently pull the cableout of the EASYpure LF frame.

MAINTENANCE AND SERVICING

WarningDepressurize system prior toopening cartridge access door.

Page 27: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

27

9. Unscrew and remove the cell.

10. Carefully remove the O-ring beforecleaning the cell.

11. Wash the cell in a mild detergentsolution or a 10% Hydrochloric orSulfuric acid solution (follow acidmanufacturers warnings andrecommended handling procedure).This may be done in an ultrasoniccleaner or with a soft brush.

12. Thoroughly rinse the cell in deionizedor distilled water following thedetergent or acid cleaning.

13. After cleaning, reinstall and check theo-ring on cell; replace if necessary.

14. Reinstall the cell into the cell well andhand tighten. Reroute the cable upthrough the housing and reconnect.

15. Reinstall the screw holding thecell-cable retaining clip. Reinstallmembrane switch lead. Replace thefront cover. Re-tape the Luer fittingwith new Teflon® tape and reinstallfitting. Replace the top cover.

MAINTENANCE AND SERVICING

CautionThe cell electrodes are etched toimprove wetting characteristics. Donot mechanically abrade or damagethis surface (i.e. do not clean with awire brush, sandpaper, etc.)

WarningCarefully follow manufacturer’ssafety instructions on labels ofchemical containers and materialsafety data sheets.

CautionDo not immerse the entire cellassembly in cleaning solution, onlythe electrode portion.

Page 28: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

28

16. Reconnect the power cord and thepump interlock plug to the unit, andreconnect the unit to the power supply.Connect incoming water line at quickdisconnect.

ShutdownIf the EASYpure LF is to be shut down for anextended period of time, the unit should becompletely drained and the cartridges removed toprevent the growth of bacteria.

If the system has remained inactive and full ofwater for more than 96 hours, the unit should bedrained, sanitized and new cartridges installedprior to use.

MAINTENANCE AND SERVICING

Page 29: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

29

ConsumablesConsumable parts are those required to support the day-to-day operation of this equipment.Barnstead|Thermolyne establishes two types of consumables; those items that mustperiodically be replaced to maintain performance (filters, resin cartridges, etc.) and otheritems of limited life (indicator lights, fuses, etc.) that you can expect to replace on a more orless random basis. Where practical, Barnstead|Thermolyne recommends the frequency ofreplacement, or provides information on life expectancy from which you may calculate areplacement interval compatible with your usage pattern.

The replacement of consumable parts is discussed in the MAINTENANCE AND SERVICINGsection to assist you in accomplishing your own service.Consumables may be ordered separately and in some cases, as an expendables kit. Checkwith your Barnstead|Thermolyne representative for additional information on theexpendables kit.

Description Catalog No. Recommended Qty.Pretreatment Cartridge DI Feed D50230 1Pretreatment Cartridge R/O - Distilled Feed D50231 1EASYpure High Purity/Low TOC Cartridge D50229 1Ultrapure Mixed Bed Cartridge D50233 10.2 micron Final Filter and Bell Assembly FL703X2 2Fuse, 100/120 Volt Models - 2.0 Amp, 250V FZX37 2 eachFuse, 230 Volt Model - Type T, 1.0 Amp, 250V 5120-0016 2 eachEmpty Cartridges for sanitization (set of 3) D70341 1Sanitization Cartridge D50245 1O-rings, cartridges 06162 8 8

General MaintenanceGeneral maintenance parts are defined as laboratory level repair parts which do not requiregreat expertise or special tools for installation. Barnstead|Thermolyne recommends that youstock the general maintenance parts as an aid to ensuring the continued operation of thisequipment.

Description Catalog No. Recommended QuantityCheck valve 02214 1Luer Fitting PM703X3 1Pressure Regulating Valve 02280 1Quick Disconnect, body CUX12 1Quick Disconnect, insert CUX10 1

Parts and Supplies

Page 30: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

30

Safety StockFor critical applications where performance with minimum downtime is required,Barnstead|Thermolyne recommends that you maintain a local stock of those parts listed inthe GENERAL MAINTENANCE PARTS and SAFETY STOCK sections.

Description Catalog No. Recommended QuantityDisplay PCB PC703X1 1Main PCB100-120 Volt PC703X2 1240 Volt PC703X3 1Pump and Motor Assembly.100-120 Volt PU738X1A 1240 Volt PU738X2AResistivity cell E703X1A 1Draw-off valve Assembly. PM703X9A 1

PARTS AND SUPPLIES

Page 31: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

31

Troubleshooting

FLerupYSAE.evitcaniyletelpmoc,gnitarepotonpmup(

,tiltonlenaplortnoc.cte

otrewoplacirtceleoN.FLerupYSAE

.nwolbsesuF

sidrocrewopFLerupYSAEehttahterusnEdnaecruosrewopevilaotdetcennoc

.teltuolacirtceleotnideggulpyletelpmoc

.nosihctiwseludomyrtnerewoperusekaM

esuFehtnidetacidnisasesufehtecalpeR.noitcestnemecalpeR

ontub,snurpmuPlatigidon(yalpsid

.)yalpsid

dnaBCPniaMtonBCPyalpsid

.detcennoc

dnakcehC.rewopmorftinutcennocsiD.sdraobtcennocer

.nurtonseodpmuP.tilyalpsiD

niretawwoLwolroknatretawdeef.erusserpretawtelni

secivedrorepmuJpmupnidellatsniton

.kcolretni

rotuonrowpmuP.evitcefed

enilretawtelnikcehC.retawdeefhsinelpeR.sevlavdesolcrosegakcolb,snoitcirtsnocrof

.llatsnI

.pmupecalpeR

"rrE"sdaeryalpsiDgnikcehcnehw

.ytivitsiser

daelllecytivitsiseR.detcennocsid

.metsysniriA

.ytridllecytivitsiseR

.daelllecytivitsiserkcehC

retawffogniwardybmetsysmorfriaegruPnoitarepOehtnisnoitcurtsniehtotgnidrocca

.noitces

.llatsnierdnanaelC

lliwretawdetalucriceRderisedotpuesnirton

.levelytirup

.egdirtracdetsuahxE

fotuosegdirtraC.redro

ehtnidetacidnisasegdirtracehtecalpeR.noitcestnemecalpeRegdirtraC

saredroreporpnisegdirtracehtllatsnI.noitcesnoitallatsnIegdirtraCehtnidetacidni

onrodecudeRehtmorfwolftcudorp

.ylbmessaretlif2.0

retliflanifnorcim2.0.deggolc

ylreporpmisegdirtraC.desnir

nidetacidnisaylbmessaretlif2.0ehtecalpeR.noitcestnemecalpeRretliFnorciM2.0eht

.retlifwenllatsni;segdirtracesniR

Page 32: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

32

TROUBLESHOOTING

sgolcretliflanifnorcim2.0.tnemecalperretfayldipar

detanimatnocFLerupYSAE.airetcabhtiw

ehtotgnidroccaFLerupYSAEezitinaS.noitazitinaSmetsySnisnoitcurtsni

saylbmessaretlif2.0ehtecalpeRretliFnorciM2.0ehtnidetacidni

.noitcestnemecalpeR

.efilegdirtractrohS

eradesugniebsegdirtraC.etadnoitaripxednoyeb

retawdeefniegnahC.scitsiretcarahc

nigebsegdirtraC.etadnoitaripxeehtkcehCowtderotsgniebretfayticapacesolot

.erutcafunamfoetadehtmorfsraeyderipxenuhtiwsegdirtracehtecalpeR

.seno

retawdeefehtsierupORdaetsnraBafIsienarbmemehttahtkcehc,ecruos

.ylreporpgninoitcnuf

,ecruosretawdeefehtsillitSdaetsnraBafIseoderutarepmetetallitsidehttahterusne

ehtotdeddanehwF°021deecxeton.riovreserretawdeefFLerupYSAE

retawlartnecamorfsiretawdeeffIdnaytilauqretawyfirev,metsysnoitacifirup

.metsysehtfogninoitcnufreporp

edisniegakaelretaW.FLerupYSAE

gnibutgnitcennocegdirtraCotnidetaesyllufton

.stekcos

evitcefedrognissiM.sgnir-Oegdirtrac

.stekcosotniylmrifgnibutgnitcennocsserP

.sgnir-OegdirtracecalperrollatsnI

Page 33: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

33

Ordering Procedures

Please refer to the Specification Plate for the complete model number, serial number, andseries number when requesting service, replacement parts or in any correspondenceconcerning this unit.

All parts listed herein may be ordered from the Barnstead Thermolyne dealer from whomyou purchased this unit or can be obtained promptly from the factory. When service orreplacement parts are needed we ask that you check first with your dealer. If the dealercannot handle your request, then contact our Customer Service Department at 319-556-2241 or 800-553-0039.

Prior to returning any materials to Barnstead Thermolyne Corp. , please contact ourCustomer Service Department for a “Return Goods Authorization” number (RGA). MaterialReturned without an RGA number will be returned.

Page 34: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

34

Figures

Figure 8Flow Diagram

Page 35: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

35

FIGURES

Figure 9100 Volt Wiring Schematic

Page 36: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

36

FIGURES

Figure 10120 Volt Wiring Schematic

Page 37: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

37

FIGURES

Figure 11230 Volt Wiring Schematic

Page 38: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

38

Page 39: EASYpure™ LF - Cole-Parmer · PDF fileThe Barnstead EASYpure LF is a line-fed water purification system designed to provide Type ... inlet of the EASYpure LF. 9 ... the valve is

39

Barnstead|Thermolyne Corporation warrants that if a product manufactured byBarnstead|Thermolyne and sold by it within the continental United States or Canadaproves to be defective in material or construction, it will provide you, without charge, for aperiod of ninety (90) days, the labor, and a period of one (1) year, the parts, necessary toremedy any such defect. Outside the continental United States and Canada, the warrantyprovides, for one (1) year, the parts necessary to remedy any such defect. The warrantyperiod shall commence either six (6) months following the date the product is sold byBarnstead|Thermolyne or on the date it is purchased by the original retail consumer,whichever date occurs first.

All warranty inspections and repairs must be performed by and parts obtained froman authorized Barnstead|Thermolyne dealer or Barnstead|Thermolyne (at its owndiscretion). Heating elements, however, because of their susceptibility to overheating andcontamination, must be returned to our factory, and if, upon inspection, it is concluded thatfailure is not due to excessive high temperature or contamination, warranty replacement willbe provided by Barnstead|Thermolyne . The name of the authorizedBarnstead|Thermolyne dealer nearest you may be obtained by calling 1-800-446-6060 orwriting to:

Barnstead|ThermolyneP.O. Box 797

2555 Kerper BoulevardDubuque, IA 52004-0797

USAFAX: (319) 589-0516

E-MAIL ADDRESS: [email protected]|Thermolyne’s sole obligation with respect to its product shall be to repair

or (at its own discretion) replace the product. Under no circumstances shall it be liable forincidental or consequential damage.