Thermo Scientific DRI Oxycodone Assay Drugs of Abuse Testing Oxycodone, a semi-synthetic opioid, is prescribed for pain management for patients with moderate to severe pain. The analgesic properties of oxycodone are similar to codeine and morphine, but more potent than morphine. Oxycodone has a higher dependency potential. Drug abusers crush the pills into powder and snort for a faster effect which may result in a potentially fatal outcome. The DRI ® Oxycodone assay is a liquid, ready-to-use assay with specific antibodies that detect oxycodone and oxymorphone without significant cross reactivity to other opiate compounds. Par t of Thermo Fisher Scientific

DRI Oxycodone Assay Drugs of Abuse Testing - Thermo Scientific

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Thermo ScientificDRI Oxycodone AssayDrugs of Abuse Testing

Oxycodone, a semi-synthetic opioid, is prescribed for pain management for patients with moderate to severe pain. The analgesic properties of oxycodone

are similar to codeine and morphine, but more potent than morphine. Oxycodone has a higher dependency potential. Drug abusers crush the pills into powder and

snort for a faster effect which may result in a potentially fatal outcome.The DRI® Oxycodone assay is a liquid, ready-to-use assay with specific

antibodies that detect oxycodone and oxymorphone without significant cross reactivity to other opiate compounds.

Part of Thermo Fisher Scientific

© 2008 Thermo Fisher Scientific Inc. All rights reserved. All the trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.

2008 08 01

Thermo ScientificDRI Oxycodone AssayDrugs of Abuse Testing

DiagnosticsU.S.A.46360 Fremont Blvd. Fremont, CA 94538 USATel. 800-232-3342Fax: 510-979-5002CanadaTel. [email protected]. + 49 851 88689 [email protected] Tel. + 61 2 9649 [email protected] Tel. + 34 93589 [email protected] KingdomTel. + 44 1727 [email protected]

• Outstanding sensitivity and specificity

• Exhibits 100% cross-reactivity to oxymorphone with very low cross reactivity to other opiate compounds

• Excellent correlation with GC/MS• Optimized for 2 cutoff levels,

300 ng/mL and 100 ng/mL• Fully automated results in

minutes• Increased value to drug test

panel for clients• Liquid, ready-to-use reagents in

convenient packaging for use on a variety of chemisty analyzers

Specific Immunoassay for Detecting Oxycodone Qualitative

100 ng/mL Cutoff 75 ng/mL 100 ng/mL 125 ng/mL

Mean (mA/min) = 348 370 389

SD = 2.1 1.9 2.0

% CV = 0.6 0.5 0.5

300 ng/mL Cutoff 225 ng/mL 300 ng/mL 375 ng/mL

Mean (mA/min) = 429 458 479

SD = 2.2 2.4 2.4

% CV = 0.5 0.5 0.5

Semi-Quantitative

75 ng/mL 100 ng/mL 125 ng/mL

Mean (ng/mL) = 73 98 123

SD = 2.4 2.9 3.0

% CV = 3.3 2.9 2.4

225 ng/mL 300 ng/mL 375 ng/mL

Mean (ng/mL) = 227 303 375

SD = 5.0 9.0 10.2

% CV = 2.2 3.0 2.7

Within Run Precision

Qualitative

100 ng/mL Cutoff 75 ng/mL 100 ng/mL 125 ng/mL

Mean (mA/min) = 348 370 389

SD = 2.9 3.2 3.1

% CV = 0.8 0.9 0.8

300 ng/mL Cutoff 225 ng/mL 300 ng/mL 375 ng/mL

Mean (mA/min) = 429 458 479

SD = 3.7 4.1 3.8

% CV = 0.9 0.9 0.8

Total Precision

Semi-Quantitative

75 ng/mL 100 ng/mL 125 ng/mL

Mean (ng/mL) = 73 98 123

SD = 2.9 3.6 4.9

% CV = 4.0 3.7 4.0

225 ng/mL 300 ng/mL 375 ng/mL

Mean (ng/mL) = 227 303 375

SD = 8.2 11.5 14.7

% CV = 3.6 3.8 3.9

Compound Concentrations (µg/mL)

6-Acetylmorphine 50

Codeine 500

Dihydrocodeine 100

Heroin 300

Hydrocodone 75

Hydromorphone 30

Levorphanol 200

Morphine 350

Morphine-3-glucuronide 900

Norcodeine 1000

Normorphine 1000

SensitivityThe sensitivity of the assay using the negative calibrator is 4.9 ng/mL:

Cross ReactivityThe following structurally related compounds tested negative with the DRI Oxycodone assay using a cutoff of 100 ng/mL:

Catalog Number Product Description Size (mL)

100248 Oxycodone Reagent Kit 70

100249 Oxycodone Reagent Kit 500

100250 Oxycodone 100 Calibrator 10

100251 Oxycodone 300 Calibrator 10

100252 Oxycodone 500 Calibrator 10

100253 Oxycodone 1000 Calibrator 10

100254 Oxycodone 100 Control 2 x 10

100255 Oxycodone 300 Control 2 x 10

Order 1-800-232-3342

Interferent Level (mg/dL) Interferent Level (mg/dL)

Acetone 1000 Hemoglobin 300

Ascorbic acid 1500 Human serum albumin 500

Creatinine 500 Oxalic acid 100

Ethanol 1000 Riboflavin 7.5

Galactose 10 Sodium Chloride 1000

Glucose 3000 Urea 2000

InterferenceNo significant interference was observed from the following substances:

Method Comparison

Qualitative vs. GC/MSIn sample correlation studies using 144 clinical samples, the overall concordance at both cutoff levels, 100ng/mL and 300ng/mL, is 97.2%.Semi-Quantitative vs. GC/MSDRI Oxycodone correctly detected 59 samples as positive and 84 as negative using 100 ng/mL as cutoff calibrator.

GC/MS vs DRI OxycodoneDemings y = 1.00 - 1, r = 0.982

Least Square Regression:Slope: 0.98Intercept: 2.00r: 0.982N: 139

GC/MS (ng/mL)

DRI (

ng/m

L)

1000

800

600

400

200

00 200 400 600 800 1000