Drägèr - Veterinary Technics · Intended Use The Dräger minute Volumeter 3000 shows the patient's respiratory minute volume and tidal volume independent ot the gas composition,

Embed Size (px)

Citation preview

3850'D

,;",.:>

<:c

Drägèr

"..<..~.:+~~

~~~~~t!J()~~~~~ .

_~.;;:,A~ t"v$'~.'OPERATING MANUAL

Minute Volumeter 3000

From Dräger:Minute Volumeter 3000OPERATING INSTRUCTIONS

Important NoticeFor correct and effective use of the ap­pliance, and to avoid hazards, we wouldpoint out the following:

Any use of the appliance requiresprecise knowledge and observationof these operating instructions.

2 The appliance is intended only for thepurposes specified in the OperatingManual or for purposes confirmed inwriting by Drägerwerk AG.

3 The appliance should be inspected byexperts at regular time intervals. Anofficial report of the inspectionsshould be drawn up.

4 Only original Dräger spare partsr' ,uld be used for maintenance and'_,Jairs. Repairs and maintenance,and the replacement of spare partsshould only be carried out by experts.

5 We recommend having inspectionsand repair work carried out by the

2

Technical Customer Service of yourDräger Branch or Agent.Regular inspection is best ensured byentering into an Inspection ServiceContract with the Technical Custo­mer Service ofyour Dräger Branch orAgent.

6 Responsibility for the reliable func­tion of the appliance passes to theowner or operator in all cases wherethe appliance has been inexpertlymaintained or repaired by personsnot employed by the Dräger Organi­sation or where it has been used in amanner which does not conform tothe normal conditions of use.

We would also point out that the natio­nar recommendations, regulations andlaws governing the use of technicalequipment should be observed.

DRÄGERWERK AG LÜBECK

ContentsPage

Important Notice 2

Intended Use 3

Approvals 3

Technical Data 3

Initial Preparation 4

Operational Use 4

Cleaning and Disinfection 6

Maintenance and Care 7

Trouble Shooting 7

Order List.. 8

Intended UseThe Dräger minute Volumeter 3000shows the patient's respiratory minutevolume and tidal volume independentot the gas composition, and it therebyappreciably simplities the determina­tion ot respiratory values needed tortherecord.

Application ot Volumeter 3000• In assisted or controlled ventilation;• During anaesthesia;

• In the determination ot vita I capacity;• To check spontaneous respiration

(e. g. during postoperative recovery).

No heating is required tor the Volume­ter 3000.

It is resistant to corrosion in respect tothe usual inhalation anaesthetics (e. g.Halothane, Entlurane, or Methoxy­tlurane).

Technical Data

ApprovalsThe Volumeter 3000 is exclusivelymechanical in operation and it is there­tore explosion proot.

Fig.1 Minute Volumeter 300036270

Measurement range1 turn ot large hand1 turn ot small hand

Necessary initial ti ow

Max. permissible peak tlow

Flow resistance

Heat stability

Connections

Air inlet (above)Air outlet (below)

Weight (without heating)

Dimensions

Diameter/Length

1000 cm315 I

3.5 I/min

120 I/min

1 mbar at 80 I/min

up to 120°C

Male thread M 33x 1Union nut M 33x 1

0.9 kg

80 0/90 mm

3

Initial PreparationThe Volumeter is connected to the ex­piratory side of a ventilator or to a circlesystem of an anaesthetic machine. Theexpiratory air flows from top to bottom.

The Volumeter can be directly screwedonto the Dräger ventilators- andanaesthetic machines (Fig. 2).

Operational UseMeasurement of Minute Volume

When the left button is depressed (Fig.3), a timer begins to run, which after oneminute wil! stop the hand ofthe volume­ter. In this manner, the minute volume isdetermined. When the left button ispushed, a black spot wil! also sim ul­taneously appear in the left aperture,and after one minute, a black disc wil! fil!the rectangular space, located belowthe hand of the indicator.

Measurement of Tidal Volume

When the right button is depressed (Fig.4) a black spot wil! appear in the rightaperture and the tidal volume wil! bemeasured.

To stop the indicator hand after acertainnumber of respiratory cycles, the rightbutton is depressed half way. To restartthe Volumeter, the same button iscompletely depressed.

38271 38272

Fig.2 Minute Volumeter 3000 in circle Fig. 3 Release of indicator hand for minutesystem of anaesthetic machine volume

4

38213

Fig. 4 Release of indicator hand for tidalvolume

Measurement of Vital Capacity

(Special equipment 2M 18476 for theminute Volumeter 3000 in a kit with ac­cessories).

The accessory kit (Fig. 5) enables mea­surement of vital capacity using a facemask (Fig. 6) or a moufh piece (Fig. 7).

During postoperative recovery, the mi­nute volume of a patient with an indwel­ling endotracheal tube can be deter­mined using the same special equip­ment without further expenditure.

38625

Fig. 6 Measuring the vita I capacity using aface piece

38626

Fig. 7 Measuring the vital capacity using amouth piece

38621

Fig.5 Dräger Minute Volumeter 3000 with special equipment kit

1 Minute Volumeter 30002 Face mask size 1 (large)3 Face mask size 2 (smalI)4 Head st rap5 Valve6 Mouth piece7 Nose clamp8 Corrugated hose with

connecting nozzleconsisting of:GasketThreaded connectorCorrugated hoseUnion collarConnecting nozzle

9 Hose for endotracheal tubes10 Bag with 5 cardboard

mouth pieces11 Cone shaped adapter

2 M 18250M 8983M 9126M 8135

2 M 21032 M 2353

R 5868

2 M 2346

M 92572 M 2328

M 12558M 9226

2 M 2091M 10071

2 M 23558400851

5

Cleaning and DisinfectionCleaning

The Volumeter is to be cleaned aftereach day's use. Run hot water from thetap through the Volumeter measure­ment chamber, the water flowing framtop to bottom, so that none of the watercan enter the bleed vents of the contralmechanism (Fig. 8). Do not immerse theVolumeter in water.

Shake out the Volumeter manually (Fig.9) and then dry it using the DrägerVolumeter Dryer 2M 14850 (Fig. 10) or

the Dräger Dryer 2M 8220. The connect­ing nozzle 2M 13921 is to be used to at­tach the Volumeterto the 2M 8220Dryer.

Disinfection in the Dräger Aseptor

Clean and dry the Volumeter as de­scribed before. Then attach it to theholding screw in the Aseptor and it willbe disinfected with formaldehyde.

Autoclave Sterilization

Clean the Volumeter as described be-

fore. It does not need to be dried.Sterilize in an autoclave set to a glovepragramme at 120°C.

..Q

38628

Fig. 8 Flushing out Volumeter

6

38629

Fig. 9 Shaking out remaining water38274

Fig.10 Dräger Volumeter-Dryer 2 M 14850

Maintenance and CareThe bearings of the Volumeter's rotatingelements become accessible through 4lateral bores; these bores are marked byprinted arrows, This makes it possible tolubricate the bearings with a speciallysuited lubricant »Oxigënoex S 4« (ap­proved for fittings carrying oxygen).A lubricating set 2M 18180 is suppliedwith each Volumeter.

Lubrication must be carried out afterabout 30 steam sterilisations or at least

every 2 months (Fig. 11).Note!

Only use pipette supplied with the kit toavoid damage to the bearings.

Each bearing must be lubricated as fol­lows:

• Remove protective cover frompipette and open lubricant bottle.

• Dip tip of pipette into lubricant andsuck one drop of »Oxigenoex S 4«into pipette by pulling the pin up­wards.

• Insert pipette into one of the fourbores as far as it will go (bores indi­cated by arrows on the side of theVolumeter). Push the pin forward tothe stop whereby lubricant is pressedinto the hole.

Fig. 11 Lubrication of the bearings38275

Trouble ShootingIf the timer mechanism of the Volume­ter 3000 does not start followingsterilization using superheated steam,th is is due to the small shafts sticking intheir bearings. This can be corrected bygiving the Volumeter a sharp turnaround its longitudinal axis.

When a new piece of tempered plateglass hast to be put in, remove the re­taining ring over the glass using apointed instrument (such as forceps) bylevering it out to the front. When theVolumeter is tipped. the glass will fallout.

When putting in a new glass:

• Clean the dial face;

• Clean the new glass before putting itin;

• Snap in the retaining ring.

7

Order List

DescriptionDesignationOrder-No.

Minute Volumeter 3000

Volumeter 30002 M 18250

Minute Volumeter 3000

Volumeter 3000 in

in carrying case for measurementcarrying case2 M 18476

of vital capacity; complete with 2 face masks, valve, mouth pieceand connecting hosesSpare and wearing partsTempered plate glass

GlassE9285

Retaining ring

Retaining ringE 9284

Gasket

GasketsM 22154

(on upper connecting piece)(pack of 10)

Bag with cardboard mouth pieces;

Bag mouthpiece2M2355

bag contains five pieces (belongs to Volumeter in carrying case)Lubrication-Set for

Lubrication-Set2 M 18180

Volumeter 3000

Subject to alterations!

8