18
DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar Hussain Biosensor Research Group 19 May 2014 1

DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

DQCM Beating the Standard Coagulometer in

the Domain of Sensitivity Range and

Information for Hemostasis of Human Plasma

Dr. Munawar Hussain

Biosensor Research Group

19 May 2014 1

Page 2: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 2

Instrumentation: qCell T

Photograph with permission from 3T Analytik

Page 3: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 3

Instrumentation: qCell T

• 10 MHz QCM (mass and visco-elastic sensitive transducer)

• Equally suitable for real time liquid and gas phase sensing

• Equipped with efficient thermostat with fast temperature control

• Powerful pumping system that eliminates any chance of bubble/

fluctuation in liquid phase analysis

• Fast and excellent baseline stability in few seconds

• The only DQCM instrument suitable for Hemostasis because of

excellent baseline stability

• User friendly software with latest features for data handling

Page 4: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

Literature

19 May 2014 4

* Y. Chunyan, Q. Ling, F. Weiling, Sensors. 2013, 13(6), 6946.

* R. S. Lakshmanan, V. Efremov, S. M. Cullen, A. J. Killard, Sens. Act. B. 2014, 192, 23.

• Some reports on fibrinogen concentration calculation

using QCM*

• No method for cumulative study of aPTT with fibrinogen

concentration in a single set of measurement.

• No study on comparison of aPTT with standard

coagulometer.

Page 5: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

Experimental

19 May 2014

5

VP-DVB

polymerization

under UV

λmax 350 nm

UV hardened

QCMs

aPTTqCell T

measurementsMeasurements

Spin coated QCMs coated

with thin film

UV λmax 350 nm

5000 rpm

Standard

Coagulometer

Page 6: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 6

DQCM Exemplary aPTT Curve

-750

-600

-450

-300

-150

0

150

300

450

600

0 20 40 60 80

Δ f

/H

Γ/

Hz

t/ s

aPTT Coagulation

aPTT CoagulationDamping

Negative Control withPlasma

Negative Control withPlasma Damping

Negative Control withoutPlasma

Negative Control withoutPlasma Damping

aPTT Starter

CaCl2 Addition

aPTT Point

Total Coagulation

Page 7: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 7

aPPT for Heparin Induced Plasma

-1200

-800

-400

0

400

800

1200

1600

0 100 200 300 400 500 600

Δ f

/H

Γ/

Hz

t/ s

Plasma

Plasma Damping

Plasma + 0.25 IU/mlHeparin

Plasma + 0.25 IU/mlHeparin Damping

Plasma + 0.5 IU/mlHeparin

Plasma + 0.5 IU/mlHeparin Damping

Plasma + 1 IU/ml Heparin

Plasma + 1 IU/ml HeparinDamping

aPTT Starter

Addition

Page 8: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 8

DQCM Vs Standard Coagulometer

25

50

75

100

125

150

25 50 75 100 125 150

tQC

M/

s

tCoag / s

Plasma

+ 0.25 IU ml-1 Heparin

+ 0.5 IU ml-1 Heparin

Page 9: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 9

aPTT for Heparin 1 IU/mL Plasma

0

50

100

150

200

250

300

1 2 3 4 5 6 7 8 9 10 11

tQC

M/

s

Samples Containing Heparin 1 IU/ml

tCoag / s Not Applicable

Page 10: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 10

Bland Altman Plot

-10.00

-5.00

0.00

5.00

10.00

25 50 75 100 125 150

tQC

M -

tCoa

g /

s

(tQCM + tCoag)/2 / s

Plasma

+ 0.25 IU ml-1 Heparin

+ 0.5 IU ml-1 Heparin

Mean-2SD

Mean

Mean +2SD

Page 11: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 11

Fibrinogen Gradient Curves for PPP

Page 12: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 12

Fibrinogen Calibration Curves

y = 20.755x2 - 30.875x - 59.939

R² = 0.9908

y = 6.4611x2 + 30.514x - 35.362

R² = 0.9983

y = 23.023x2 - 47.686x - 26.988

R² = 0.9767

y = 17.278x2 - 46.529x + 34.267

R² = 0.9757

y = -10.879x2 + 1.7767x - 158.05

R² = 0.9671

y = 1.1489x2 - 82.923x + 51.613

R² = 0.9892

y = -188.9ln(x) - 6.9266

R² = 0.9978

y = -13.209x2 + 41.728x - 146.64

R² = 0.9982

-600

-400

-200

0

200

400

600

0 1.5 3 4.5 6

Δ f

/ H

z

Δ Γ

/ H

z

Plasma Fibrinogen / (g/L)

Δ f /Δ Γ [From Start Point till aPTT Point]

Pl Poor

Pl 0.3 million

Pl 0.15 million

Pl 0.05 million

(A)

-750

-600

-450

-300

-150

0

150

300

450

600

0 20 40 60 80

Δ f

/H

Γ/

Hz

t/ s

Page 13: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 13

Fibrinogen Calibration Curves

y = 102.37x2 - 376.61x + 514.95

R² = 0.9772

y = 51.687x2 - 63.232x + 20.824

R² = 0.9995

y = 26.626x2 + 99.158x - 59.395

R² = 0.9982 y = 70.278x2 - 378x + 778.15

R² = 0.9929

y = -33.352x2 + 94.794x - 316.41

R² = 0.9871

y = -9.4979x2 - 53.289x + 55.646

R² = 0.9923

y = -5.2748x2 - 56.587x - 15.148

R² = 0.9753

y = -20.803x2 + 152.62x - 494.77

R² = 0.9972

-1000

-500

0

500

1000

1500

2000

0 1.5 3 4.5 6

Δ f

/ H

z

Δ

Γ/

Hz

Plasma Fibrinogen / (g/L)

Δ f /Δ Γ [From aPTT Point till Total Coagulation]

Pl Poor

Pl 0.3 million

Pl 0.15 million

Pl 0.05 million

(B)

-750

-600

-450

-300

-150

0

150

300

450

600

0 20 40 60 80

Δ f

/H

Γ/

Hz

t/ s

Page 14: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 14

Fibrinogen Calibration Curves

y = 123.12x2 - 407.49x + 455.01

R² = 0.9811

y = 58.149x2 - 32.718x - 14.538

R² = 0.9994

y = 49.649x2 + 51.471x - 86.383

R² = 0.9953

y = 87.556x2 - 424.53x + 812.42

R² = 0.9933

y = -44.231x2 + 96.57x - 474.46

R² = 0.9831

y = -8.349x2 - 136.21x + 107.26

R² = 0.9995

y = 6.8344x2 - 209.23x + 123.44

R² = 0.9802

y = -34.012x2 + 194.35x - 641.41

R² = 0.9994

-2000

-1500

-1000

-500

0

500

1000

1500

2000

2500

0 1.5 3 4.5 6

Δ f

/H

z

Δ Γ

/ H

z

Plasma Fibrinogen / (g/L)

Δ f /Δ Γ [From Start Point till Total Coagulation]

Pl Poor

Pl 0.3 million

Pl 0.15 million

Pl 0.05 million

(C)

-750

-600

-450

-300

-150

0

150

300

450

600

0 20 40 60 80

Δ f

/H

Γ/

Hz

t/ s

Page 15: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 15

Summary:

Excellent correlation at lower and mid range aPTT with

respect to standard mechanical coagulometer

Plasma-1 IU/mL heparin mechanical coagulometer N/A.

aPTT+Fibrinogen data from a single set of measurements.

Sensing layers are stable can be reused (10 times tested)

qCell T a suitable instrument for coagulation

DQCM has bright future of relevant clinical applications

Page 16: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

Thanks for your kind attention

19 May 2014 16

Page 17: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 17

Page 18: DQCM Beating the Standard Coagulometer in the …...DQCM Beating the Standard Coagulometer in the Domain of Sensitivity Range and Information for Hemostasis of Human Plasma Dr. Munawar

19 May 2014 18

Fibrinogen Gradient Curves

20

30

40

50

60

70

0 1.5 3 4.5 6

aP

TT

/ s

Plasma Fibrinogen / (g/L)

[Platelets Poor Plasma]

Coagulometer

QCM

WHO Fibrinogen Standard

in Healthy Human Plasma

(A)

20

30

40

50

60

70

0 1.5 3 4.5 6

aP

TT

/ s

Plasma Fibrinogen / (g/L)

[Platelets 0.3 million]

Coagulometer

QCM

WHO Fibrinogen Standard

in Healthy Human Plasma

(B)

20

30

40

50

0 1.5 3 4.5 6

aP

TT

/ s

Plasma Fibrinogen / (g/L)

[Platelets 0.15 million]

Coagulometer

QCM

WHO Fibrinogen Standard

in Healthy Human Plasma

(C)

20

30

40

50

60

0 1.5 3 4.5 6

aP

TT

/ s

Plasma Fibrinogen / (g/L)

[Platelets 0.05 million]

Coagulometer

QCM

WHO Fibrinogen Standard

in Healthy Human Plasma

(D)