41
User’s Manual DPharp BRAIN Communication Type (EJXA, EJAE) IM 01C25T03-01E IM 01C25T03-01E 6th Edition

DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

  • Upload
    hakien

  • View
    222

  • Download
    0

Embed Size (px)

Citation preview

Page 1: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

User’sManual DPharp

BRAIN Communication Type(EJXA, EJAE)

IM 01C25T03-01E

IM 01C25T03-01E6th Edition

Page 2: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

i

IM 01C25T03-01E

DPharpBRAIN Communication Type

IM 01C25T03-01E 6th Edition

6th Edition: June 2014 (KP)All Rights Reserved, Copyright © 2004, Yokogawa Electric Corporation

Contents1. Introduction ............................................................................................... 1-1

Regarding This Manual ................................................................................................1-11.1 Safe Use of This Product .................................................................................1-11.2 Warranty .............................................................................................................1-21.3 ATEX Documentation .......................................................................................1-3

2. Connection ................................................................................................ 2-12.1 Connecting the BT200 ......................................................................................2-12.2 Communication Line Requirements ...............................................................2-12.3 Power Supply Voltage and Load Resistance .................................................2-22.4 Integral Indicator Display When Powering On ..............................................2-2

3. Operation ................................................................................................... 3-13.1 BT200 Operating Procedures ..........................................................................3-1

3.1.1 Key Layout and Screen Display .........................................................3-1

3.1.2 Operating Key Functions ...................................................................3-1

(1) Alphanumeric Keys and Shift Keys ........................................... 3-1

(2) Function Keys ............................................................................3-2

3.1.3 Calling Up Menu Addresses Using the Operating Keys .................... 3-3

3.1.4 Printout (for BT200 printer option) .....................................................3-3

(1) Printout of All Parameters ..........................................................3-3

(2) Printout by Menu Item ...............................................................3-3

3.2 Setting Parameters Using the BT200 .............................................................3-43.2.1 Parameter Usage and Selection ........................................................3-4

3.2.2 Menu Tree ..........................................................................................3-5

3.2.3 Setting Parameters ............................................................................3-6

(1) Tag No. Setup ...........................................................................3-6

(2) Calibration Range Setup ...........................................................3-6

(3) Damping Time Constant Setup ................................................3-7

(4) Output Mode and Integral Indicator Display Mode Setup ........ 3-8

(5) Output Signal Low Cut Mode Setup ......................................... 3-8

(6) Integral Indicator Scale Setup ...................................................3-9

(7) Unit Setup for Displayed Temperature .................................... 3-11

(8) Operation Mode Setup ........................................................... 3-11

(9) Impulse Line Connection Orientation Setup .......................... 3-11

Page 3: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

ii

IM 01C25T03-01E

(10) CPU Failure Burnout Direction and Hardware Write Protect . 3-11

(11) Software Write Protect ............................................................3-12

(12) Output Status Setup when a Hardware Error Occurs ............3-13

(13) Bi-directional Flow Measurement Setup .................................3-13

(14) Range Change while Applying Actual Inputs .........................3-13

(15) Sensor Trim .............................................................................3-14

(16) Test Output Setup ...................................................................3-17

(17) Signal Characterizer ................................................................3-17

(18) Process Alarm ........................................................................3-18

(19) Status Output (option code AL) ...............................................3-18

(20) Capillary Fill Fluid Density Compensation .............................3-19

(21) Adjustment Information and User Memo Fields .....................3-20

3.3 Displaying Data Using the BT200 ..................................................................3-203.3.1 Displaying Measured Data...............................................................3-20

3.3.2 Display Transmitter Model and Specifications .................................3-21

4. Self-diagnostics ........................................................................................ 4-14.1 Checking for Problems ....................................................................................4-1

4.1.1 Identifying Problems with BT200 .......................................................4-1

4.1.2 Checking with Integral Indicator .........................................................4-2

4.2 Alarms and Countermeasures ........................................................................4-2

5. Parameter Summary ................................................................................ 5-1Appendix 1. Safety Instrumented Systems Installation ............................A1-1

A1.1 Scope and Purpose ....................................................................................... A1-1A1.2 Using the Transmitter for an SIS Application ............................................. A1-1

A1.2.1 Safety Accuracy ...............................................................................A1-1

A1.2.2 Diagnostic Response Time ..............................................................A1-1

A1.2.3 Setup ................................................................................................A1-1

A1.2.4 Required Parameter Settings ..........................................................A1-1

A1.2.5 Proof Testing ....................................................................................A1-1

A1.2.6 Repair and Replacement .................................................................A1-2

A1.2.7 Startup Time .....................................................................................A1-2

A1.2.8 Firmware Update .............................................................................A1-2

A1.2.9 Reliability Data .................................................................................A1-2

A1.2.10 Lifetime Limits ..................................................................................A1-2

A1.2.11 Environmental Limits .......................................................................A1-2

A1.2.12 Application Limits .............................................................................A1-2

A1.3 Definitions and Abbreviations ...................................................................... A1-3A1.3.1 Definitions ........................................................................................A1-3

A1.3.2 Abbreviations ...................................................................................A1-3

Revision Information

Page 4: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<1. Introduction> 1-1

IM 01C25T03-01E

1. IntroductionThank you for purchasing the DPharp EJX series pressure transmitter/EJA series pressure transmitter (“transmitter”).

The transmitters are precisely calibrated at the factory before shipment. To ensure both safety and efficiency, please read this manual carefully before operating the instrument.

This manual describes the BRAIN protocol communication functions of the transmitter and explains how to set the parameters for the transmitters using the BT200 handheld terminal. For information on the installation, wiring, and maintenance of the transmitters, please refer to the user’s manual of each model.

WARNING

When using the transmitter in a Safety Instrumented Systems (SIS) application, refer to Appendix 1 in this manual. The instructions and procedures in the appendix must be strictly followed in order to maintain the designed safety integrity of the transmitter.

Regarding This Manual• This manual should be provided on to the end

user.

• The contents of this manual are subject to change without prior notice.

• All rights reserved. No part of this manual may be reproduced in any form without Yokogawa’s written permission.

• Yokogawa makes no warranty of any kind with regard to this manual, including, but not limited to, implied warranty of merchantability and fitness for a particular purpose.

• If any question arises or errors are found, or if any information is missing from this manual, please inform the nearest Yokogawa sales office.

• The specifications covered by this manual are limited to those for the standard type under the specified model number break-down and do not cover custom-made instruments.

• Please note that changes in the specifications, construction, or component parts of the instrument may not immediately be reflected in this manual at the time of change, provided that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint.

• The following safety symbols are used in this manual:

WARNING

Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

CAUTIONIndicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices.

IMPORTANTIndicates that operating the hardware or software in this manner may damage it or lead to system failure.

NOTEDraws attention to information essential for understanding the operation and features.

1.1 Safe Use of This Product For the safety of the operator and to protect the instrument and the system, please be sure to follow this manual’s safety instructions when handling this instrument. If these instructions are not heeded, the protection provided by this instrument may be impaired. In this case, Yokogawa cannot guarantee that the instrument can be safely operated. Please pay special attention to the following points:

Page 5: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<1. Introduction> 1-2

IM 01C25T03-01E

(a) Installation

• This instrument may only be installed by an engineer or technician who has an expert knowledge of this device. Operators are not allowed to carry out installation unless they meet this condition.

• With high process temperatures, care must be taken not to burn yourself by touching the instrument or its casing.

• Never loosen the process connector nuts when the instrument is installed in a process. This can lead to a sudden, explosive release of process fluids.

• When draining condensate from the pressure detector section, take appropriate precautions to prevent the inhalation of harmful vapors and the contact of toxic process fluids with the skin or eyes.

• When removing the instrument from a hazardous process, avoid contact with the process fluid and the interior of the meter.

• All installation shall comply with local installation requirements and the local electrical code.

(b) Wiring

• The instrument must be installed by an engineer or technician who has an expert knowledge of this instrument. Operators are not permitted to carry out wiring unless they meet this condition.

• Before connecting the power cables, please confirm that there is no current flowing through the cables and that the power supply to the instrument is switched off.

(c) Operation

• Wait 10 min. after the power is turned off, before opening the covers.

(d) Maintenance

• Please carry out only the maintenance procedures described in this manual. If you require further assistance, please contact the nearest Yokogawa office.

• Care should be taken to prevent the build up of dust or other materials on the display glass and the name plate. To clean these surfaces, use a soft, dry cloth.

(e) Modification

• Yokogawa will not be liable for malfunctions or damage resulting from any modification made to this instrument by the customer.

1.2 Warranty• The warranty shall cover the period noted on

the quotation presented to the purchaser at the time of purchase. Problems occurring during the warranty period shall basically be repaired free of charge.

• If any problems are experienced with this instrument, the customer should contact the Yokogawa representative from which this instrument was purchased or the nearest Yokogawa office.

• If a problem arises with this instrument, please inform us of the nature of the problem and the circumstances under which it developed, including the model specification and serial number. Any diagrams, data and other information you can include in your communication will also be helpful.

• The party responsible for the cost of fixing the problem shall be determined by Yokogawa following an investigation conducted by Yokogawa.

• The purchaser shall bear the responsibility for repair costs, even during the warranty period, if the malfunction is due to:

- Improper and/or inadequate maintenance by the purchaser.

- Malfunction or damage due to a failure to handle, use, or store the instrument in accordance with the design specifications.

- Use of the product in question in a location not conforming to the standards specified by Yokogawa, or due to improper maintenance of the installation location.

- Failure or damage due to modification or repair by any party except Yokogawa or an approved representative of Yokogawa.

- Malfunction or damage from improper relocation of the product in question after delivery.

- Reason of force majeure such as fires, earthquakes, storms/floods, thunder/lightening, or other natural disasters, or disturbances, riots, warfare, or radioactive contamination.

Page 6: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<1. Introduction> 1-3

IM 01C25T03-01E

1.3 ATEX DocumentationThis section is only applicable to the countries in the European Union.

GB

DK

I

E

NL

SF

P

F

D

S

LT

LV

PL

EST

SLO

H

BG

RO

M

CZ

SK

GR

Page 7: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<2. Connection> 2-1

IM 01C25T03-01E

2. ConnectionThe BRAIN communication signal is superimposed onto the 4 to 20 mA DC analog signal. Since the modulated wave is a communication signal, superimposing it on the normal signal will, from basic principles, cause no error in the DC component of the analog signal. Thus, monitoring can be performed via the BT200 while the transmitter is on-line.

2.1 Connecting the BT200

IMPORTANTAnalog output may change temporally in connecting with BRAIN terminal due to an initial current flowed to it. To prevent communication signal affecting the upper system, it is recommended to install a low-pass filter (approximately 0.1s)

Connection to the transmitter with the BT200 can be made by either connecting to the BT200 connection hooks in the transmitter terminal box or by connecting to a relaying terminal or a terminal board.

Relaying terminals

Distributor

Control room

Terminal board

F0201.ai

Figure 2.1 Connecting the BT200

2.2 Communication Line Requirements

[Protocol specification] Yokogawa original protocol

[Modulation] Burst modulation 0: 2400Hz 1: Signal without carrier

[Baud rate] 1200bps

[Communication signal]

host to device: +/- 0.5V (load resistance 250Ω)

device to host: +/- 2mA

DPharp

BT200

Cable resistance Rc

Cable resistance Rc

Load resistance R

ccPower supply

Loop resistance = R + 2Rc = 250 to 600 Ω Loop capacitance = 0.22 µF max.

F0202.ai

Figure 2.2 Communication Line Requirements

Page 8: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<2. Connection> 2-2

IM 01C25T03-01E

2.3 Power Supply Voltage and Load Resistance

When configuring the loop, make sure that the external load resistance is within the range in the figure below.(Note) With an intrinsically safe transmitter, external load

resistance includes safety barrier resistance.

600

250

0 10.5 16.6 25.2 42

External load

resistanceR (Ω)

Power supply voltage E (V DC)F0203.ai

Communication applicable range

R= E–10.50.0244

Figure 2.3 Relationship between Power Supply Voltage and External Load Resistance

2.4 Integral Indicator Display When Powering On

For models with the integral indicator code “D”, the display shows all segments in the LCD and then changes to the displays shown below sequentially.

All segments display Model name (3 s)

Communication Protocol (3 s) – – – – (3 s) F0200.ai

NOTEFor output signal code “D”, this function is available for software revision 2.02 or later.Software revision can be checked by the parameter M15: SOFT REV.Refer to section 3 “Operation” how to call up the parameter.

NOTELCD display can be set to all segments display only by the parameter I41: POWER ON INF.

ON Show All segments display, Model name and Communication Protocol when powering on.

OFF Show All segments display when powering on.

Refer to section 3 “Operation” how to call up the parameter.

Page 9: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-1

IM 01C25T03-01E

3. OperationThe transmitter is equipped with BRAIN communications capabilities, so that range changes, Tag No. setup, monitoring of self-diagnostic results, and zero point adjustment can be handled remotely via the BT200 BRAIN TERMINAL, the FieldMate Versatile Device Management Wizard or the CENTUM CS console. This section describes procedures for setting parameters using the BT200. For further information on the BT200, see the BT200 User’s Manual (IM 01C00A11-01E).

IMPORTANTCommunication signal is superimposed on analog output signal. It is recommended to set a low-pass filter (approximately 0.1s) to the receiver in order to reduce the output effect from communication signal. Before online-communication, confirm that communication signal does not give effect on the upper system.

3.1 BT200 Operating Procedures

3.1.1 Key Layout and Screen DisplayFigure 3.1 shows the arrangement of the operating keys on the BT200 keypad, and figure 3.2 shows the BT200 screen.

F0301.ai

LCD(21 characters × 8 lines)

Movement keys

Power ON/OFF key

Function keys

ENTER key

Alphanumeric keys

Shift keys

Figure 3.1 BT200 Key Layout

PARAM A10:OUTPUT 0.0 % A11:PRES 0.0000 kPa A15:OUTPUT mA 4.000 mA

Menu choices

MessagesMENU SCREEN

Screen title

MENU A:DISPLAY B:SENSOR TYPE C:BASIC SETUP D:AUX SET1 E:AUX SET2 G:ALARM SET

BATTERY

HOME SET ADJ ESC

Functioncommands

PARAMETER SCREEN

Parameters

DATA DIAG PRNT ESC

F0302.ai

Figure 3.2 BT200 Screen

3.1.2 Operating Key Functions

(1) Alphanumeric Keys and Shift KeysUse the alphanumeric keys in conjunction with the shift keys to enter numbers, symbols, and alphabetic characters.

Shift keys

Alphanumeric keys

F0303.ai

a. Entering Numbers, Symbols, and Spaces

Simply press the alphanumeric keys.

Entry Key-in sequence

–4

0.3

1 –9F0304.ai

Page 10: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-2

IM 01C25T03-01E

b. Entering Alphabetic Characters

Press either the left or right shift key and then an alphanumeric key to enter the desired alphabetic character. The shift key must be pressed each time an alphabetic character is entered.

Entry Key-in sequence

W

IC

J. BF0305.ai

Letter on left side of the alphanumeric key

Letter on right side of the alphanumeric key

Use the function key [F2] CAPS to select uppercase and lowercase (for alphabetic characters only). The case toggles between uppercase and lowercase each time [F2] CAPS is pressed.

Entry

Boy

Key-in sequence

F0306.ai

CODE CAPS CLR ESC

Entering uppercase

CODE caps CLR ESC

Entering lowercase

( B ) ( y )( o )

to lower case

Use the function key [F1] CODE to enter symbols. The following symbols will appear in sequence, one at a time, at the cursor each time [F1] CODE is pressed:

/ . – , + * ) ( ’ & % $ # ” !

To enter characters next to these symbols, press [ > ] to move the cursor.

( / )

Entry

l/m

Key-in Sequence

F0307.ai

( m )( I )

symbol command

(2) Function KeysThe function command carried out by each function key is displayed directly above the key.

MENU A:DISPLAY B:SENSOR TYPE C:BASIC SETUP D:AUX SET1 E:AUX SET2 G:ALARM SETHOME SET ADJ ESC Function commands

Function keysF0308.ai

Function Command List

Command FunctionADJ Displays the ADJ menuCAPS/caps Selects uppercase or lowercaseCODE Selects symbolsCLR Erases input data or deletes all dataDATA Updates parameter dataDEL Deletes one characterDIAG Calls the self-check panelESC Returns to the most recent displayHOME Displays the menu panelNO Quits setup and returns to the previous

displayOK Proceeds to the next panelPARM Enters the parameter number setup

modeSET Displays the SET menuSLOT Returns to the slot selection panelUTIL Calls the utility panel*COPY Prints out parameters on display*FEED Paper feed*LIST Lists all parameters in the menu*PON/POFF Automatic printout mode on or off*PRNT Changes to the print mode*GO Starts printing*STOP Cancels printing

* Available on BT200-P00 (with printer).

Page 11: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-3

IM 01C25T03-01E

3.1.3 Calling Up Menu Addresses Using the Operating Keys

DATA DIAG PRNT ESC

PARAM A60:SELF CHECK GOOD

CODE CAPS CLR ESC

--WELCOME--BRAIN TERMINAL ID: BT200

check connection push ENTER key

UTIL FEED esc

PARAM 01:MODEL EJX110A-DM 02:TAG NO. YOKOGAWA 03:SELF CHECK GOOD

OK

STARTUPSCREEN

UTILITY SCREEN

FUNCTION SCREEN

MENUSCREEN

SETUPSCREEN

INITIAL DATA SCREEN

PARA-METER SCREEN

(HOME MENU SCREEN) (SET MENU SCREEN) (ADJ MENU SCREEN)

See the BT200 user’s manual for details concerning uploadingand downloading parameters and printouts (BT200-P00).

(UTIL)

(SET)

(ESC)

HOME SET ADJ

(ADJ)

The utility screen contains the following items.1. BT200 ID settings2. Security code settings3. Switching language of messages (Japanese or English)4. LCD contrast setting5. Adjusting printout tone (BT200-P00 only)

FUNC 1.MENU 2.UPLOAD TO HHT 3.DOWNLOAD TO INST 4.PRINT ALL DATA

HOME SET ADJ ESC

MENU A:DISPLAY B:SENSOR TYPE C:BASIC SETUP D:AUX SET1 E:AUX SET2 G:ALARM SET

HOME SET ADJ ESC

MENU C:BASIC SETUP D:AUX SET 1 E:AUX SET 2 G:ALARM SET H:AUTO SET I:DISP SET

HOME SET ADJ ESC

MENU J:ADJUST K:TEST M:DEVICE INFO P:RECORD T:CHARACTERIZR

DATA DIAG PRNT ESC

DATA DIAG PRNT ESC DATA DIAG PRNT ESC

PARAM C60:SELF CHECK GOOD

PARAM C22:PRES URV 100.00 kPa C23:PRES POINT DATA DIAG PRNT ESC

DATA DIAG PRNT ESC

PARAM J60:SELF CHECK GOOD

PARAM J12:P SPAN ADJ 100.000 % J15:P ZERO DEV

DATA DIAG PRNT ESC

PARAM J09:ADJ UNIT % J10:ADJ PRES 0.02563 % J11:P ZERO ADJ 0.04792 %

SET C10:TAG NO. YOKOGAWA YOKOGAWA

F0309.ai

DATA DIAG PRNT ESC

PARAM A10:OUTPUT 0.0 % A11:PRES 0.0000 kPa A15:OUTPUT mA 4.000 mA

DATA DIAG PRNT ESC

PARAM C10:TAG NO. YOKOGAWA C20:PRES UNIT kPa C21:PRES LRV 0.00000 kPa

PARAM A16:ENGR.OUTPUT 0.0000 mmHg A17:ENGR.EXP

UTILITY 1.ID 2.SECURITY CODE 3.LANGUAGE SELECT 4.LCD CONTRAST 5.PRINTER ADJUST

3.1.4 Printout (for BT200 printer option)

(1) Printout of All ParametersSelect 4. PRINT ALL DATA from the function screen to output a list of all parameters. It takes about 10 minutes to complete the printout.

(2) Printout by Menu ItemTo printout the parameters for a specific screen, push the function key corresponding to screen’s PRNT.

Page 12: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-4

IM 01C25T03-01E

3.2 Setting Parameters Using the BT200

3.2.1 Parameter Usage and SelectionBefore setting a parameter, please see the following table for a summary of how and when each parameter is used.

Table 3.1 Parameter Usage and Selection

Setup item DescriptionTag No. setup P.3-6 Sets the Tag No. (using 16 alphanumeric characters).Calibration range setup P.3-6

Sets the calibration range for 4 to 20 mA DC. Sets the following items: range unit, input value at 4 mA DC (LRV), input value at 20 mA DC (URV), and decimal point position.Note: LRV and URV can be specified with range value specifications up to 5 digits (excluding any

decimal point) within the range of –32000 to 32000.Damping time constant setup P.3-7

Adjusts the output response speed for 4 to 20 mA DC at amplifier.Can be set from 0.50 to 100.00 s. (from 0.00 to 100.00 s with quick response mode on)

Output and integral indicator display mode setup P.3-8

Sets modes for output signal and integral indicator to Linear mode (proportional to input differential pressure) or to Square root mode (proportional to flow).

Output signal low cut mode setup P.3-8

Used mainly to stabilize output near 0% if the output signal is square root mode.Two modes are available: forcing output to 0% for input below a specific value, or changing to proportional output for input below a specific value.

Integral indicator display function P.3-9

Available from the following 5 types of integral indicator scale ranges and units: input pressure, % of range, user set scale, input static pressure, % of static pressure range, and

alternating among any four of the above.Configure the following when using the user set scale; user set scale setting, unit (BT200 only), display value at 4 mA DC (LRV), and display value at

20 mA DC (URV).Note: LRV and URV can be specified with range value specifications up to 5 digits (excluding any

decimal point) within the range of –32000 to 32000.Static pressure setup P.3-11

Sets the parameters concerned with static pressure such as unit, calibration range, upper and lower range values, decimal point position, damping time constant.

Unit setup for displayed temperature P.3-11

Sets the unit for temperatures displayed on the BT200.

Operation mode (normal/reverse signal) setup P.3-11

Reverses the direction for 4 to 20 mA DC output relative to input.Reverse mode is used for applications in which safety requires that output be driven toward 20 mA if input is lost.

Impulse line connection orientation (higher pressure on right/left side) setup P.3-11

Used where installation conditions make it imperative to connect high pressure side impulse line to low pressure side of transmitter. Reversal of orientation should be dealt with by reversing impulse line wherever possible. Use this function only where there is no alternative.

CPU Failure burnout direction and hardware write protect P.3-11

Displays the status of 4 to 20 mA DC output when a CPU fails. The direction is selectable by the hardware switch on the amplifier. It also physically prevents parameter access

Software write protect P.3-12 Configured data can be protected by setting a password.Output status setup when a hardware error occurs P.3-13

Sets the status of the 4 to 20 mA DC output when an abnormal status is detected with the capsule or the amplifier as the result of self-diagnosis. Either the last held, high limit, or low limit values status, can be selected.

Bi-directional flow measurement P.3-13

Used to measure bi-directional flows. Output at zero flow is 12 mA DC, with output range equally divided between forward and reverse flow. Can be used with square root mode.

Range change while applying actual inputs P.3-13

Range for 4 to 20 mA DC signal is set with actual input applied. Sets 20 mA DC output precisely with respect to user’s reference instrument output. Note that the transmitter is calibrated with high accuracy before shipment, so span should be set using the normal range setup.

Sensor trim P.3-14 Adjusts zero point and span of the sensor.Test output (fixed current output) setup P.3-17

Used for loop checks.Output can be set freely from –2.50% to 110.00% in 0.01% steps.

Signal characterizer P.3-17 Used to compensate the output for the non-linear application.Process alarm P.3-18 Used for alarm generation on the integral indicator.Status output P.3-18 Outputs an on/off digital signal based on the settings of process alarm.Capillary fill fluid density compensation P.3-19

Compensates the zero shift by the ambient temperature effect on the capillary tubes.

User memo fields P.3-20

Allows user to enter up to 3 items, each containing any combination of up to 16 alphanumeric characters.

IMPORTANTAfter setting and sending data with the BT200, wait 30 seconds before turning off the transmitter. If it is turned off too soon, the settings will not be stored in the transmitter.

Page 13: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-5

IM 01C25T03-01E

3.2.2 Menu Tree

A: DISPLAYHOME

A10: OUTPUTA11: PRESA15: OUTPUT mAA16: ENGR. OUTPUTA17: ENGR. EXPA20: SP %*1

A21: SP*1

A30: CAPSULE TEMPA60: SELF CHECK

B10: MODELB11: STYLE NO.B20: PRES LRLB21: PRES URLB22: P MIN SPANB30: SP LRL*1

B31: SP URL*1

B32: SP MIN SPAN*1

B60: SELF CHECK

C: BASIC SETUP

C10: TAG NO.C20: PRES UNITC21: PRES LRVC22: PRES URVC23: PRES POINTC30: AMP DAMPINGC40: OUTPUT MODEC60: SELF CHECK

D10: LOW CUTD11: LOW CUT MODED15: H/L SWAP*1

D16: H2O UNIT SELD20: OUT LIMIT(L)D21: OUT LIMIT(H)D22: REV OUTPUTD25: BURNOUTD26: ERROR OUTD30: SP UNIT*1

D31: SP A/G SLCT*1

D32: ATM. PRESS*1

D33: SP LRV*1

D34: SP URV*1

D35: SP POINT*1

D36: SP DAMPING*1

D37: SP SELECT*1

D40: TEMP UNITD50: QUICK RESPD55: WRT PROTECTD56: WRT ENABLED57: NEW PASSWORDD58: SOFTWR SEALD60: SELF CHECK

E10: T. ZERO CMPE11: TEMP ZEROE30: BI DIRE MODEE50: DO SELECT*2

E51: DO SIG.TYPE*2

E52: D OUTPUT*2

E60: SELF CHECK

G10: P AL MODEG11: P HI. AL VALG12: P LO. AL VALG20: SP AL MODE*1

G21: SP HI. AL VAL*1

G22: SP LO. AL VAL*1

G30: T AL MODEG31: T HI. AL VALG32: T LO.AL VALG50: AUTO RECOVERG60: SELF CHECK

H10: AUTO P LRVH11: AUTO P URVH20: AUTO SP LRV*1

H21: AUTO SP URV*1

H60: SELF CHECK

I10: DISP OUT1I11: DISP OUT2I12: DISP OUT3I13: DISP OUT4I20: P DISP MODEI21: PRES % RESOI30: ENGR. UNITI31: EASY EU SETI32: ENGR. EXPI33: ENGR. LRVI34: ENGR. URVI35: ENGR. POINTI40: BAR INDICATRI41: POWER ON INF*3

I60: SELF CHECK

J09: ADJ UNITJ10: ADJ PRESJ11: P ZERO ADJJ12: P SPAN ADJJ15: P ZERO DEVJ16: P SPAN DEVJ20: ADJ SP*1

J21: SP ZERO ADJ*1

J22: SP SPAN ADJ*1

J25: SP ZERO DEV*1

J26: SP SPAN DEV*1

J40: OUTPUT 4mAJ41: OUTPUT 20mAJ45: AMP TEMPJ50: ADJ WHOJ51: ADJ DATEJ52: ADJ LOCJ53: ADJ DESCJ55: EXT ZERO ADJJ56: CLEAR ADJJ60: SELF CHECK

K10: OUTPUT X %K40: DO TEST*2

K45: TEST TIME*3

K50: TEST KEY1K51: TEST KEY2K52: TEST KEY3K53: TEST KEY4K60: SELF CHECK

M10: SERIAL NO.M11: MFTR. DATEM12: EXTRA NO.M15: SOFT REVM16: BRAIN REVM17: MEMO1M18: MEMO2M19: MEMO3M20: ISOL MATLM21: FILL FLUIDM22: GASKET MATLM23: PRO CON MATLM24: D-VENT MATLM25: PRO CON TYPEM26: RS ISOL MATLM27: PRO CON SIZEM28: NUM RSM29: RS FILL FLIDM30: RS TYPEM50: MS CODE 1M51: MS CODE 2M52: MS CODE 3M53: MS CODE 4M54: MS CODE 5M55: MS CODE 6M60: SELF CHECK

P10: ERROR REC 1P12: ERROR REC 2P14: ERROR REC 3P16: ERROR REC 4P50: REC CLEARP60: SELF CHECK

T10: S. C. ENABLET11: NUM OF POINTT20: X START (FIX)T21: Y START (FIX)T22: X1T23: Y1T24: X2T25: Y2T26: X3T27: Y3T28: X4T29: Y4T30: X5T31: Y5T32: X6T33: Y6T34: X7T35: Y7T36: X8T37: Y8T38: X9T39: Y9T40: X END (FIX)T41: Y END (FIX)T60: SELF CHECK

B: SENSOR TYPE

D: AUX SET 1 E: AUX SET2 G: ALARM SET H: AUTO SET I: DISP SET

J: ADJUST K: TEST M: DEVICE INFO P: RECORD T: CHARACTERIZR

SET

ADJ

F0310.ai*1: Available for differential pressure transmitter.*2: Available for EJX series only.*3: Available for software revision 2.02 or later. Software revision can be checked by the parameter M15: SOFT REV.

Page 14: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-6

IM 01C25T03-01E

3.2.3 Setting ParametersSet or change the parameters as necessary. After completing these, do not fail to use the “DIAG” key to confirm that “GOOD” is displayed for the self-diagnostic result at _60: SELF CHECK.

(1) Tag No. Setup (C10: TAG NO)

Use the procedure below to change the Tag No. Up to 16 alphanumeric characters can be entered.

Press the key to turn on the BT200.

Connect the transmitter and BT200 using a communication cable and press the key.

PARAM 01:MODEL EJX 02:TAG NO. YOKOGAWA 03:SELF CHECK GOOD

OK

MENU A:DISPLAY B:SENSOR TYPE C:BASIC SETUP D:AUX SET1 E:AUX SET2 G:ALARM SETHOME SET ADJ ESC

MENU C:BASIC SETUP D:AUX SET 1 E:AUX SET 2 G:ALARM SET H:AUTO SET I:DISP SETHOME SET ADJ ESC

MENU C10:TAG NO. YOKOGAWA C20:PRES UNIT kPa C21:PRES LRV 0.00000 kPaDATA DIAG PRNT ESC

<When power is off>

Select C: BASIC SETUP and press the key.

Select C10: TAG NO. and pressthe key.

Displays the model name of connected transmitter, TAG NO. and diagnostics information. Press the (OK) key after confirmation.

Press the (SET) key to display the SET menu panel.

--WELCOME--BRAIN TERMINAL ID: BT200

check connection push ENTER key

UTIL FEED

F0311_1.ai

SET C10:TAG NO. YOKOGAWA YOKOGAWA

CODE CAPS CLR ESC

When you have made an entry mistake, return the cursor using the key, then reenter.

Set the new TAG NO. (FIC-1a).FOKOGAWA

FIKOGAWA

FICOGAWA

FIC-GAWA

FIC-1AWA

FIC-1aWA

FIC-1a

Set TAG NO. and press the key.

SET C10:TAG NO. YOKOGAWA FIC-1a _

CODE caps CLR ESC

• Example: Set a Tag No. to FIC-1a

SET C10:TAG NO. FIC-1a

FEED NO OK

DATA DIAG PRNT ESC

This is the panel for confirming set data. The set data items flash.When all items have been confirmed, press theagain. (To go back to the setting panel, press the (NO) key.

SET C10:TAG NO. YOKOGAWA FIC-1a

print off F2:printer on

PARAM C10:TAG NO. FIC-1a C20:PRES UNIT kPa C21:PRES LRV 0.00000 kPa

FEED POFF NO

The DPharp TAG NO. was overwritten.Press the (OK) key to return to the parameter panel.Press the (NO) key to return to the setting panel.

F0311_2.ai

(2) Calibration Range Setup

a. Setting Calibration Range Unit (C20: PRES UNIT)

The unit parameter is set at the factory before shipment if specified at the time of order. Follow the procedure below to change the unit parameter.

• Example: Change the unit from mmH2O to kPa.

ESC

SET C20:PRES UNIT kPa

FEED NO OK

SET C20:PRES UNIT mmH2O < mmWG > < mmHg > < Torr > < kPa >

Use the or key to select kPa.Press the key twice to enter the setting.

Press the (OK) key.

F0312.ai

Note that the Yokogawa default setting for the standard temperature is 4°C (39.2°F). For the units of mmH2O, mmAq, mmWG, inH2O, and ftH2O, the pressure varies according to the standard temperature definition. When a standard temperature of 20°C (68°F) is required, select @20degC (68.0F) at the parameter D16:H2O UNIT SEL.

Available pressure units are shown below.

mmH2OmmAqmmWGmmHgTorrkPa

MPambarbargf/cm2kgf/cm2inH2O

inHgftH2OpsiatmPahPa

Page 15: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-7

IM 01C25T03-01E

b. Setting Calibration Range Lower Range Value and Upper Range Value (C21: PRES LRV, C22: PRES URV)

These range values are set as specified in the order before the instrument is shipped. Follow the procedure below to change the range.

• The measurement span is determined by the upper and lower range limit values. In this instrument, changing the lower range value also automatically changes the upper range value, keeping the span constant.

• Example 1: With present settings of 0 to 30 kPa, set the lower range value to 0.5 kPa.

Span = Upper range value – Lower range value

FEED NO OK

SET C21:PRES LRV 0.5 kPa

DATA DIAG PRNT ESC

PARAM C20:PRES UNIT kPa C21:PRES LRV 0.5 kPa C22:PRES URV 30.5 kPa

DEL CLR ESC

SET C21:PRES LRV 0 kPa + 0.5

The upper range value is changed while the span remains constant.

Set 0.5.Press the key twice to enter the setting.

Press the (OK) key.

F0313.ai

• Entering the range values as LRV>URV reverses the direction of the output signal of 4-20 mA to 20-4 mA corresponding to the calibration range of 0 to 100%.

• Calibration range can be specified with range value specifications up to 5 digits (excluding any decimal point) for lower or upper range limits within the range of –32000 to 32000.

• Note, however, that changing the upper range value does not cause the lower range value to change. Thus, changing the upper range value also changes the span.

• Example 2: With present settings of 0 to 30 kPa, set the upper range value to 10 kPa.

FEED NO OK

SET C22:PRES URV 10 kPa

DATA DIAG PRNT ESC

PARAM C20:PRES UNIT kPa C21:PRES LRV 0 kPa C22:PRES URV 10 kPa

DEL CLR ESC

SET C22:PRES URV 30 kPa + 10

The lower range value is not changed, so the span changes.

Set 10.Press the key twice to enter the setting.

F0314.ai

Press the (OK) key.

(3) Damping Time Constant Setup (C30: AMP DAMPING)

When the instrument is shipped, the damping time constant is set at 2.00 seconds unless otherwise specified in the order. Follow the procedure below to change the damping time constant.

Note that setting the quick response parameter (D50: QUICK RESP) ON enables you to set the damping time constant between 0.00 to 0.49 second.

• Example: Change from 2.00 to 4.00 seconds.

ESCCLR

SET C30:AMP DAMPING 4.00 sec

FEED NO OK

SET C30:AMP DAMPING 2.00 sec + 004.00

Enter 4.Press the key twice to enter the setting.

F0315.ai

Press the (OK) key.

Note 1: The damping time constant set here is the time constant for the amplifier assembly. The damping time constant for the entire transmitter is the sum of the values for the amplifier assembly and for the capsule assembly.

Note 2: When the damping time constant is set to less than 0.5 second, communication may occasionally be unavailble during the operation, especially while output changes dynamically.

Page 16: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-8

IM 01C25T03-01E

(4) Output Mode and Integral Indicator Display Mode Setup (C40: OUTPUT MODE, I20: P DISP MODE)

The mode setting for the output signal and the integral indicator can be performed independently.

This mode is set as specified in the order when the instrument is shipped. Follow the procedure below to change the mode.

If the instrument is equipped with an integral indicator and the display mode is SQUARE ROOT, “ ” is displayed on the integral indicator.

• Output mode for 4-20 mA output

• Example: Set output mode from Linear to Square root.

ESC

SET C40:OUTPUT MODE SQUARE ROOT

FEED NO OK

SET C40:OUTPUT MODE LINEAR < LINEAR > < SQUARE ROOT >

Use the or key

to select SQUARE ROOT.

Press the key twice to

enter the setting.

Press the (OK) key.

F0316.ai

• Integral indicator display mode

• Example: Set display mode from Linear to Square root.

ESC

SET I20:P DISP MODE SQUARE ROOT

FEED NO OK

SET I20:P DISP MODE LINEAR < LINEAR > < SQUARE ROOT >

Use the or key

to select SQUARE ROOT.

Press the key twice to

enter the setting.

Press the (OK) key.

F0317.ai

(5) Output Signal Low Cut Mode Setup (D10: LOW CUT, D11: LOW CUT MODE)

Low cut mode can be used to stabilize the output signal near the zero point. The low cut point can be set in a range from 0 to 20%, the direct ratio corresponding to the output signal of 4 to 20 mA. (Hysteresis: ±10% of the cut point)

Either LINEAR or ZERO can be selected as the low cut mode. Unless otherwise specified, the cut mode is set to LINEAR at the factory.

Note that when the output modes of the output signal and the integral indicator are selected as SQUARE ROOT and LINEAR accordingly, the low cut function is not available for the integral indicator display.

• Example: Change the low cut setting range from 10% to 20%, and the low cut mode from LINEAR to ZERO in the SQUARE ROOT output mode.

FEED NO OK

SET D11:LOW CUT MODE ZERO

Use the or key

to select ZERO.

Press the key twice to

enter the setting.

Press the (OK) key.

F0318.ai

FEED NO OK

SET D10:LOW CUT 20.00 %

ESC

SET D11:LOW CUT MODE LINEAR < LINEAR > < ZERO >

DATA DIAG PRNT ESC

PARAM D10:LOW CUT 20.0 % D11:LOW CUT MODE ZERO D15:H/L SWAP NORMAL

CLR ESC

SET D10:LOW CUT 10.00 % + 20.00

Set 20.

Press the key twice to

enter the setting.

Press the (OK) key.

Next, the [D11: LOW CUT MODE]

setting panel is displayed.

(%)50

20

0 50 (%)

Output

Low cut mode “LINEAR”

Input

(%)50

20

0 50 (%)

Low cut mode “ZERO”

LOW CUT at 20%

Input

Page 17: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-9

IM 01C25T03-01E

The low cut point has hysteresis so that the output around the point is behaved as below figure.

<Example> Output mode: Linear Low cut mode: Zero Low cut: 20.00%

F0319.ai

Setting range: 0 to 20%

2% 2%4mA

Output Low cut point

Input

Hysteresis fixed at 10%of the cut point

7.2mA(20%)

(6) Integral Indicator Scale SetupThe following five displays are available for integral indicators: input pressure*1, % of range, user set scale, input static pressure, and % of static pressure range*1. A cycle of up to four displays can be shown by assigning variables to the parameters I10 to I13: DISP OUT1 to DISP OUT4.

F0320.ai

Available displays

Input pressure(PRES)

% of range(PRES %)

User set scale(ENGR. PRES)

Input static pressure(SP)*1

% of static pressure range(SP %)*1

Indicates values of input pressurewith the indication limits -32000 to 32000.

Indicates input pressure in -2.5 to 110% range depending on the measuring range (C21, C22).

Indicates values depending on the engineering range (I33, I34) with the unit (I30).

Indicates input static pressure with the indication limits -32000 to 32000.Reference pressure is factory-set in absolute.

Indicates input static pressure in -10 to 110% range depending on the measuring range (D33, D34).

A11:PRES

456 kPa

Descriptionand related parameters

A10:OUTPUT

45.6 %

A16:ENGR.OUTPUT

20.5 m3/min

A17:ENGR.EXP

×100

A21:SP

4.000 MPa

A20:SP %

52.6 %

*1: Available for differential pressure transmitter.

See (a.) through (d.) for each setting procedure.

Page 18: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-10

IM 01C25T03-01E

a. Display Selection (I10: DISP OUT1)

Select the variable for the parameter I10: DISP OUT1 to display on the integral indicator.

• Example: Change the integral indicator scale from % of range to input pressure display.

ESC

SET I10:DISP OUT1 PRES

FEED NO OK

Use the or key

to select PRES.

Press the key twice to

enter the setting.

Press the (OK) key.

F0321.ai

SET I10:DISP OUT1 PRES % < PRES > < PRES % > < ENGR.PRES > < SP >

b. Cyclic Display (I11: DISP OUT2, I12: DISP OUT3, and I13: DISP OUT4)

In addition to the display set at I10: DISP OUT1, displays can be set at I11: DISP OUT2, I12: DISP OUT3, and I13: DISP OUT4 for cyclic display in the order of the parameter number.

c. User Setting of Engineering Unit and Scale (I30: ENGR.UNIT, I31: EASY EU SET, I33: ENGR.LRV, and I34: ENGR.URV)

These parameters allow the entry of the engineering units and scale to be displayed. The engineering unit can be selected from the parameter I31: EASY EU SET as listed below. Alternately, up to eight alphanumerics, spaces, and a slash “/” can be input on keypad at I30: ENGR.UNIT; only first six are displayed on the integral indicator.

Select the unit from the list of I31: EASY EU SET.

kPaMPambarbarpsipsiammH2OmmHgmmHgAmmAqmmWGTorrinH2OinHginHgA

ftH2Ogf/cm2kgf/cm2kg/cm2Gkg/cm2Aatmkg/ht/hm3/hm3/minl/hl/minkl/hkl/minNl/h

Nl/minNm3/hNm3/minACFHACFMSCFHSCFMGPHGPMmmminftkg/m3g/cm3

Follow the procedure below to change the settings.

• Example: Set an engineering unit M.

CODE CAPS CLR ESC

SET I30:ENGR.UNIT M

FEED NO OK

SET I30:ENGR.UNIT M_

Set M.

Press the key twice to

enter the setting.

Press the (OK) key.

F0322.ai

Note that following symbols are not available.

. – , + * ) ( ’ & % $ # ” !

The transmitter integral indicator shows “-- -- -- -- -- --” when these are entered.

• Example: Set lower range value (LRV) to –50 and upper range value (URV) to 50.

FEED NO OK

SET I34:ENGR.URV 50 M

Press the (OK) key.

F0323.ai

DEL CLR ESC

DATA DIAG PRNT ESC

PARAM I32:ENGR.EXP --- I33:ENGR.LRV - 50 M I34:ENGR.URV 50 M

DEL CLR ESC

SET I33:ENGR.LRV 0.00 M - 50

Set –50.

Press the key twice to

enter the setting.

Setting LRV

Setting URVSet 50.

Press the key twice to

enter the setting.

SET I34:ENGR.URV 100.00 M + 50

FEED NO OK

SET I33:ENGR.LRV - 50 M

Press the (OK) key.

Page 19: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-11

IM 01C25T03-01E

d. Setting Static Pressure Unit and Scale (D30: SP UNIT, D33: SP LRV, and D34: SP URV)

Static pressure can be displayed in measured input static pressure or in %, independent from the 4-20 mA output signal of measured pressure or differential pressure. These parameters allow the entry of the static pressure unit and scale to be displayed.

Note that the parameter D37: SP SELECT can be used to select either the high or low pressure side of the capsule to monitor the static pressure.

(7) Unit Setup for Displayed Temperature(D40: TEMP UNIT)

When the instrument is shipped, the temperature units are set to degC. Follow the procedure below to change this setting. Note that changing the unit here changes the unit for A30: CAPSULE TEMP (capsule temperature) and J45: AMP TEMP (amplifier temperature).

• Example: Change the unit for the temperature display degC to degF.

ESC

SET D40:TEMP UNIT degC < degC > < degF > < K >

Use the or key

to select degF.

Press the key twice to

enter the setting.

F0324.ai

Then press the (OK) key.

(8) Operation Mode Setup (D22: REV OUTPUT)

This parameter allows the direction of the 4-20 mA output to be reversed with respect to input. Follow the procedure below to make this change.

• Example: Change 4 to 20 mA output to 20 to 4 mA output.

ESC

SET D22:REV OUTPUT NORMAL < NORMAL > < REVERSE>

Use the or key

to select REVERSE.

Press the key twice to

enter the setting.

F0325.ai

Then press the (OK) key.

(9) Impulse Line Connection Orientation Setup (D15: H/L SWAP)

This function reverses the impulse line orientation. Follow the procedure below to make this change.

• Example: Assign the high pressure impulse line connection to the L side of the transmitter.

ESC

SET D15:H/L SWAP NORMAL < NORMAL > < REVERSE>

Use the or key

to select REVERSE.

Press the key twice to

enter the setting.

F0326.ai

Then press the (OK) key.

(10) CPU Failure Burnout Direction and Hardware Write Protect (D25: BURNOUT)

There are two slide switches on the CPU assembly board. One sets the burnout direction at CPU failure, and the other sets a write protection function which disables parameter changes through the use of a handheld terminal or some other communication method.

HIGH LOW

CPU assembly

Slide switch

Burnout direction switch

Write protection switch

Write ProtectionSwitch Position

Burnout DirectionSwitch Position

BO H L

WR E D

H L

E D

H L

E D

H L

E D

H L

E D

Hardware write protection switch (WR)

Burnout direction switch (BO)

Burnout Direction

Write Protection YES(Write disabled)

NO(Write enabled)

F0327.ai

The parameter D25: BURNOUT displays the status of 4-20 mA DC output if a CPU failure occurs. In case of a failure, communication is disabled.

Page 20: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-12

IM 01C25T03-01E

Standard specifications

The burnout direction switch is set to HIGH. If a failure occurs, the transmitter outputs a 110% or higher signal.

Option code /C1

The burnout direction switch is set to LOW. If a failure occurs, a –5% or lower output is generated.

• Example: Standard specificationsSlide switch position: HD25: BURNOUT

HIGH

F0328.ai

• Example: Option code /C1Slide switch position: LD25: BURNOUT

LOW

(11) Software Write Protect (D55: WRT PROTECT, D56: WRT ENABLE, D57: NEW PASSWORD)

Transmitter configured data can be saved by the write protect function. Write protect status (D55: WRT PROTECT) is set from NO to YES when eight alphanumerics are entered in the parameter D57: NEW PASSWORD. Accordingly, the transmitter does not accept any parameter changes. When the eight alphanumeric password is entered in the parameter D56: WRT ENABLE, the transmitter accepts parameter changes during a 10 minute period.

To cancel the transmitter for the software write protection completely, use D56: WRT ENABLE to first release the write protect function and then enter eight spaces in the D57: NEW PASSWORD field.

The software write protection does not affect the function of external zero adjustment screw.

To disable the external zero adjustment screw, set the parameter J15: EXT ZERO ADJ to INHIBIT before activating the software write protection.

a. Setting Password (D57: NEW PASSWORD)

Enter 1234ABCD.

Then status of parameter D55: WRT PROTECT becomes YES.

F0329.ai

• Example: Set the password to 1234ABCD.

SET D57:NEW PASSWORD 1234ABCD

CODE CAPS CLR ESC

SET D57:NEW PASSWORD 1234ABCD

FEED NO OK

Press the key twice to enter the setting.

Press the (OK) key.

b. Entering Password to Enable Parameter Change (D56: WRT ENABLE)

Enter the password.

Parameter changes are availablefor 10 minutes.

F0330.ai

• Example: Enter the password of 1234ABCD.

SET D56:WRT ENABLE 1234ABCD

CODE CAPS CLR ESC

SET D56:WRT ENABLE PASS

FEED NO OK

Press the key twice to enter the setting.

Press the (OK) key.

c. Releasing Password (D57: NEW PASSWORD)

To release the password, enter eight spaces at D57: NEW PASSWORD during the period that the parameter change is possible.

d. Software Seal (D58: SOFTWR SEAL)

When you lose the password that has been registered, it is possible to release the write protect function by using general password. Contact Yokogawa about the general password. When the password is used, the status shown in the parameter D58: SOFTWR SEAL is changed from KEEP to BREAK. The status returns to KEEP by entering a newly set password at D56: WRT ENABLE.

Page 21: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-13

IM 01C25T03-01E

(12) Output Status Setup when a Hardware Error Occurs (D26: ERROR OUT)

This parameter allows the setting of the output status when a hardware error occurs. The following selections are available.

(a) BURNOUT DIR; Outputs the corresponding values of 110% or –5% of output signals according to the setting by burnout direction switch (BO) on the CPU board.

(b) HOLD; Outputs the last value held before the error occurred.

Note: A hardware error means CAP.ERR of AL.01 or AMP.ERR of AL.02 which are shown in table 4.1 Alarm Message Summary.

• Example: Set the output status to HOLD when a hardware error occurs.

ESC

SET D26:ERROR OUT BURNOUT DIR < BURNOUT DIR > < HOLD >

Use the or key

to select HOLD.

Press the key twice to

enter the setting.

F0331.ai

Then press the (OK) key.

(13) Bi-directional Flow Measurement Setup(E30: BI DIRE MODE)

(a) This parameter enables selection of 50% output at an input of 0 kPa.

Procedure is shown in the figure below.(b) Combining this with C40: OUTPUT MODE

provides a square root output computed independently for 0% to 50% output and for 50% to 100% output.

• Example: If measurement range is 0 to 10 kPa (LRV=0 kPa, URV=10 kPa)

The measurement range changes to –10 to 0 to 10 kPa (output 0% to 50% to 100).Note that C21: PRES LRV and C22: PRES URV are not changed.

ESC

SET E30:BI DIRE MODE OFF < OFF > < ON >

Use the or key

to select ON.

Press the key twice to

enter the setting.

F0332.ai

Then press the (OK) key.

20 mA (100% display)

4 mA (–100% display)

Output mode “LINEAR”

LRV HRV

F0333.ai

20 mA (100% display)Low Cut

4 mA (–100% display)

Output mode “SQUARE ROOT”

LRV HRV

(14) Range Change while Applying Actual Inputs (H10: AUTO P LRV, H11: AUTO P URV)

This feature allows the lower and upper range values to be set up automatically with the actual input applied. If the lower and upper range values are set, C21: PRES LRV and C22: PRES URV are changed at the same time.

Follow the procedure in the figure below. The measurement span is determined by the upper and lower range values. Changing the lower range value results in the upper range value changing automatically, keeping the span constant.

• Example 1: When changing the lower range value to 0.5 kPa for the present

setting of 0 to 30 kPa, take the following action with input pressure of 0.5 kPa applied.

FEED NO OK

SET H10:AUTO P LRV 0.5000 kPa

DATA DIAG PRNT ESC

PARAM H10:AUTO P LRV 0.5000 kPa H11:AUTO P URV 30.500 kPa H20:AUTO SP LRV 0.0 MPa

ESC

SET H10:AUTO P LRV 0 kPa + 0

F0334.ai

The upper range value is changed

keeping the span constant.

Parameters C21 and C22 are

changed at the same time.

Press the key twice.

The lower range value is changed

to 0.5 kPa.

Press the (OK) key.

Page 22: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-14

IM 01C25T03-01E

Note that changing the upper range value does not cause the lower range value to change but does change the span.

• Example 2: When the upper range value is to be changed to 10 kPa with the

present setting of 0 to 30 kPa, take the following action with an input pressure of 10 kPa applied.

FEED NO OK

SET H11:AUTO URV 10.000 kPa

DATA DIAG PRNT ESC

PARAM H10:AUTO P LRV 0 kPa H11:AUTO P URV 10.000 kPa H20:AUTO SP LRV 0.0 MPa

ESC

SET H11:AUTO P URV 30 kPa + 30

The lower range value is not changed, so the span changes.Parameter C22 is changed at the same time.

Press the key twice.

The upper range value is

changed to 10 kPa.

Press the (OK) key.

F0335.ai

(15) Sensor TrimEach transmitter is factory characterized. Factory characterization is the process of comparing a known pressure input with the output of each transmitter sensor module over the entire pressure and temperature operating range. During the characterization process, this comparison information is stored in the transmitter EEPROM. In operation, the transmitter uses this factory-stored curve to produce a process variable output (PV), in engineering units, dependent on the pressure input.

The sensor trim procedure allows you to adjust for local conditions, changing how the transmitter calculates process variables. There are two ways to trim the sensor: a zero trim and a full sensor trim. A zero trim is a one-point adjustment typically used to compensate for mounting position effects or zero shifts caused by static pressure. A full sensor trim is a two-point process, in which two accurate end-point pressures are applied (equal to or greater than the range values), and all output is linearized between them.

a. Zero Trim (J11: P ZERO ADJ, J15: P ZERO DEV, J55: EXT ZERO ADJ)

The transmitter supports several adjustment methods. Select the method best suited for the conditions of your application.

Adjustment Method Description

Using the BT200 Set the present input to 0%. Adjust for 0% output at input level

of 0%.Adjust output to the reference value obtained using other means. If the input level cannot easily

be made 0% (because of tank level, etc.), adjust output to the reference value obtained using other means, such as a sight glass.

Using the external zero-adjustment screw

Adjust zero point using the zero-adjustment screw on the transmitter.This permits zero adjustment without using the BT200. Accurately adjust the output current to 4 mA DC or other target output value using an ammeter that accurately reads output currents.

When using BT200, the output signal can be adjusted either in % or pressure unit. The unit can be selected by the parameter J09: ADJ UNIT. Output signal can be changed by displaying parameter A10: OUTPUT for % or J10: OUTPUT for pressure unit.

This section discribes the zero adjustment procedure by using the pressure unit.

a-1. Zeroing

Setting the parameter J11: P ZERO ADJ carries out the zero adjustment and automatically sets the applied “0” input values to the transmitter’s output value of zero, keeping the span constant. Use this setting when the LRV is known to be 0 kPa.

Page 23: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-15

IM 01C25T03-01E

FEED NO OK

SET J11:P ZERO ADJ 0.00000 kPa

CLRDEL ESC

SET J11:P ZERO ADJ 0.00000 kPa + 0

A pressure of 0 kPa is applied.

Press the key twice

after the pressure has become stable.

Zero adjustment is completed.

Press the (OK) key.

F0336.ai

Transmitter measures pressure of 0.03585 kPa.

Transmitter measures pressure of 0.00000 kPa.

A11:PRES 0.03585 kPa

A11:PRES 0.00000 kPa

a-2. Level Adjustment

The zero adjustment by the parameter J11: P ZERO ADJ calibrates the transmitter output corresponding to the actual tank level. To perform this adjustment, first use a glass gauge or the like to determine the actual tank level, then enter the correct data as shown below.

F0337.ai

25.00 kPa

0.00 kPa

Actual level13.50 kPa

DPharp span: 0 to 25.00 kPaActual level: 13.50 kPaTransmitter output: 13.83 kPa

DPharp

CLRDEL ESC

SET J11:P ZERO ADJ 0.00000 kPa + 045.0 + 13.5

Enter the present actual level, 13.5 kPa.

Press the key twice.

The measured pressure is

changed to 13.5 kPa.

F0338.ai

Transmitter measures presentpressure of 13.83 kPa.

A11:PRES 13.83 kPa

A11:PRES 13.5000 kPa

a-3. Using External Zero-adjustment Screw

This method permits zero adjustment without the BT200. Use a slotted screwdriver to turn the zero-adjustment screw. See the hardware manual for details.

Note that the parameter J55: EXT ZERO ADJ must be ENABLE to perform this adjustment.

Follow the procedure below to enable or inhibit zero point adjustment from the zero-adjustment screw on the transmitter. This is set to ENABLE when the instrument is shipped.

• Example: Inhibiting zero adjustment by the external zero-adjustment screw.

ESC

SET J55:EXT ZERO ADJ ENABLE < ENABLE > < INHIBIT>

Use the or key

to select INHIBIT.

Press the key twice to

enter the setting.

F0339.ai

Then press the (OK) key.

b. Full Sensor Trim (J11: P ZERO ADJ, J12: P SPAN ADJ, J15: P ZERO DEV, J16: P SPAN DEV)

Full sensor trim is carried out with a series of the procedure of J11: P ZERO ADJ and J12: P SPAN ADJ. Also, you can manually perform the trimming procedure by using J15: P ZERO DEV and J16: P SPAN DEV.

The full sensor trim is a two-point adjustment, and the lower point adjustment should always be performed before the upper point adjustment in order to maintain the pitch between the zero and 100% points within the calibration range.

In the manual method, the reference pressure should also be applied to the transmitter at both lower and upper point of trim ends. Without the reference pressure, J15: P ZERO DEV and J16: P SPAN DEV may not represent the correct value of adjustment point for each.

Page 24: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-16

IM 01C25T03-01E

b-1. Auto Sensor Trim

• Example: For the range of 10 to 30 kPa.

FEED NO OK

SET J11:P ZERO ADJ 10.0000 kPa

ESCCLRDEL

ESCCLRDEL

SET J11:P ZERO ADJ 9.94000 kPa + 10

Set 10.After obtaining a stable pressure of 10 kPa, press key twice.

F0340.ai

Press the (OK) key.

ESC

SET J10:ADJ PRES 9.94000 kPa

Transmitter indicates 9.94 kPa as its output when applying a standard pressure of 10 kPa.

ESC

SET J10:ADJ PRES 30.1500 kPa

Transmitter indicates 30.15 kPa as its output when applying a standard pressure of 30 kPa.

ESC

SET J10:ADJ PRES 10.0000 kPa

Check the output becomes 10 kPa.

ESC

SET J10:ADJ PRES 30.0000 kPa

Check the output becomes 30 kPa.

FEED NO OK

SET J12:P SPAN ADJ 30.0000 kPa

Press the (OK) key.

Set 30.After obtaining a stable pressure of 30 kPa, press key twice.

SET J12:P SPAN ADJ 30.1500 kPa + 30

Setting a lower point

Setting an upper point

b-2. Manual Sensor Trim

• Example: For the range of 10 to 30 kPa. J15: P ZERO DEV = –0.04 kPa J16: P SPAN DEV = –0.03 kPa

Suppose that a standard pressure of 10 kPa is applied and the value of the parameter J10: ADJ PRES is 9.94 kPa.Correct for this output error of 0.06 kPa by adding 0.06 to J15: P ZERO DEV.

–0.04 + 0.06 = +0.02

FEED NO OK

SET J15:P ZERO DEV 0.02000 kPa

Suppose that a standard pressure of 30 kPa is applied and the value of the parameter J10: ADJ PRES is 30.15 kPa.Firstly, obtain the slope error for the span as follows;

Then correct for this slope error of –0.1 by adding –0.1 to J16: P SPAN DEV.

–0.03 + (–0.1) = –0.13

ESCCLRDEL

SET J15:P ZERO DEV -0.04000 kPa + 0.02

Set 0.02.

Press key twice.

Press the (OK) key.

F0341.ai

FEED NO OK

SET J16:P SPAN DEV -0.13000 kPa

ESCCLRDEL

SET J16:P SPAN DEV -0.03000 kPa - 0.13

Set –0.13.

Press key twice.

Press the (OK) key.

Applied Pres Value – Measured Pres ValueApplied Pres Value

Slope Error × (URV – LRV)=

30.00–30.1530.00

× (30.00–10.00)= = –0.1

Page 25: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-17

IM 01C25T03-01E

c. Sensor Trim for Static Pressure (J21: SP ZERO ADJ, J22: SP SPAN ADJ, J25: SP ZERO DEV, J26: SP SPAN DEV)

For the transmitters (Except for EJX120A/EJA120E), zeroing and full sensor trim of the static pressure is performed in the same way as with the primary process variable (PV). Note that the static pressure sensor trim should be done only after trimming the PV.

d. Reset Trim Adjustment to Factory Setting (J56: CLEAR ADJ)

Use PRES or SP of J56: CLEAR ADJ parameter to reset the trim adjustment to the initial calibrated values that were set. When PRES is selected to clear the adjustment, the amount of the adjustment by the external zero-adjustment screw is returned to the initial setting as well.

• Example: Reset the trim adjustment of pressure to factory set characterization curve.

ESC

SET J56:CLEAR ADJ ---

FEED NO OK

Use the or key

to select PRES.

Press the key twice to

enter the setting.

Press the (OK) key.

F0342.ai

SET J56:CLEAR ADJ --- < --- > < PRES > < SP > < 4-20mA >

(16) Test Output Setup (K10: OUTPUT X %)

This feature can be used to output a fixed current for loop checks. The available range for test output depends on the setting at parameters D20: OUT LIMIT (L) and D21: OUT LIMIT (H), whose limit is from 3.6 mA (–2.5%) to 21.6 mA (110%).

• Example: Output 12 mA (50%) fixed current.

CLR ESC

SET K10:OUTPUT X % 50.00% ACTIVE

FEED NO OK

SET K10:OUTPUT X % 0.00 % + 050.00

Set 50.00%.

Press the key twice to

output a fixed current at 50%.

ACTIVE is displayed while this is

being executed.

Press the (OK) key to cancel

the fixed current output.

F0343.ai

NOTE• Fixed current output and DO Test continue

for a given holding time, then is released automatically. Even if the BT200 power supply is turned off or the communication cable is disconnected, the test output will continue for that time.

The holding time can be selected from 10 min*, 30 min, 60 min, 3 hour, 6 hour or 12 hour by the parameter K45: TEST TIME.

*: Default value.• Press the (OK) key to release test output

immediately.

(17) Signal CharacterizerThis function is used to compensate the output for non-linear applications. The characterized values are applied to the 4-20 mA output. For the measured pressure, a maximum of nine coordinates can be specified between 0-100%. Perform the coordinate settings while the T10: S. C. ENABLE parameter is INHIBIT.

To apply the settings to the output, set the T10: S. C. ENABLE parameter to ENABLE.

Note that the transmitter rejects the activation of the function by AL. 60 with the following transmitter’s status:

• When the specified coordinates of x and y are not incremental as the input increases.

• When the output mode of the output signal is set as SQUARE ROOT; at the same time, the low cut mode is set to LINEAR.

F0344.ai

Y

X100%0%

100%

INPUT OUTPUTInput pressure in % Characterized value

Page 26: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-18

IM 01C25T03-01E

• Example: Set the number of coodinates on the line graph to 5.

ESCCLR

SET T11:NUM OF POINT 5

FEED NO OK

SET T11:NUM OF POINT 9 5

F0345.ai

Set 5.

Press the key twice to

enter the setting.

Press the (OK) key.

• Example: Set the first coordinates (X1, Y1) as (12, 14) in %.

ESCCLR

SET T23:Y1 10.00% + 014.00

CLR ESC

SET T22:X1 10.00% + 012.00

F0346.ai

PARAM T21:Y 0.00% T22:X1 12.00% T23:Y1 14.00%

PRNTDATA DIAG ESC

Set 12 for X1.

Press the key twice to

enter the setting.Press the (OK) key.

Set 14 for Y1.

Press the key twice to

enter the setting.Press the (OK) key.

• Example: Set the signal characterizer ENABLE.

ESC

SET T10:S.C.ENABLE ENABLE

FEED NO OK

SET T10:S.C.ENABLE INHIBIT < INHIBIT> < ENABLE >

F0347.ai

Use the or key

to select ENABLE.

Press the key twice to

activate the function.

Press the (OK) key.

(18) Process Alarm (G10: P AL MODE, G11: P HI.AL VAL, G12: P LO.AL.VAL)

The function is used to display the alarm codes when the input pressure exceeds the specified value within the calibration range. The same is available for the input static pressure and the capsule temperature on the pressure sensor. Refer to table 4.1 Alarm Message Summary for the specific alarm code to be generated.

• Example: Set alarm mode from OFF to HI.AL DETECT.

ESC

SET G10:P AL MODE HI.AL DETECT

FEED NO ESC

SET G10:P AL MODE INHIBIT < INHIBIT > < HI.AL DETECT > < LO.AL DETECT > < HI/LO.AL DETECT >

Use the or key

to select HI. AL DETECT.

Press the key twice to

enter the setting.

Press the (OK) key.

Alarm code is generated when the

output goes beyond the value set

at G11: P HI. AL VAL.F0348.ai

• Example: Set the higher alert value of 75 for alarm generation.

CODE CAPS CLR ESC

SET G11:P HI.AL VAL 75.0000 kPa

FEED NO OK

SET G11:P HI.AL VAL 100.000 kPa + 75

Set 75.

Press the key twice to

enter the setting.

Press the (OK) key.

Then the alarm code “AL.35 P. HI”

is generated when the input

pressure goes beyond that value.

F0349.ai

(19) Status Output (option code AL)This feature is used for a transistor output (open collector) of an on/off signal according to the status of high and low alarm limits, which are user-configurable values as shown in (18) Process Alarm. The status output can be assigned as any combination of the high or low limits of the input pressure, input static pressure, or capsule temperature.

CAUTIONExecute DO testing by the parameter "K40: DO test" whenever turning on the transmitter or detecting the short interruption in order to check that the alarm contact output is correctly configured.

NOTENo status output signal has been defined for a CPU failure or hardware error. Use a 4-20 mA signal to indicate a transmitter’s failure.

Page 27: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-19

IM 01C25T03-01E

• Example: Set the status output to output an off signal when the input pressure exceeds 75 kPa with its alert mode of HI. AL DETECT.

ESC

SET E50:DO SELECT PRES

FEED NO OK

SET E50:DO SELECT INHIBIT < INHIBIT > < PRES > < SP > < TEMP >

Use the or key

to select PRES.

Press the key twice to

enter the setting.

Press the (OK) key.

F0350.ai

ESC

SET E51:DO SIG.TYPE OFF WHEN ALARM

FEED NO OK

SET E51:DO SIG.TYPE ON WHEN ALARM < ON WHEN ALARM > < OFF WHEN ALARM >

Use the or key

to select OFF WHEN ALARM.

Press the key twice to

enter the setting.

Press the (OK) key.

Status output for higher alert value

Status output for lower alert value

F0351.ai

Output(%)

Status output

Status output

Time (t)

Time (t)

Settingvalue

Settingvalue

5%* of hysteresis(5°C for heat)

Output(%)

5%* of hysteresis(5°C for heat)

On Off

Example: Status output operation of ON WHEN AL. DETECT

On

On

On

Off

*5% of the setting span of differential pressure / pressure.

(20) Capillary Fill Fluid Density Compensation (E10: T.ZERO CMP, E11: TEMP ZERO)

For transmitters with diaphragm seals, this function is used to compensate the zero shift caused by the ambient temperature effect on the capillary tubes.

The following equation indicates the relationship between the calculated output value and the compensating constant K (%/°C) with the measured ambient temperature at the capsule module.

Compensated output = output + K × Tamb

(1) Temperature Compensation Mode Setup (E10: T. ZERO CMP) When using this function, set T. ZERO CMP to ON to enable or OFF to disable. To set to ON, follow the procedure below.

ESC

SET E10:T.ZERO CMP OFF < OFF > < ON >

OKFEED NO

SET E10:T.ZERO CMP ON

• Example: Set the temperature compensation mode to ON.

Use the or key

to select ON.

Press the key twice to

enter the setting.

Press the (OK) key.

F0352.ai

(2) Zero Shift Compensation Setup (E11: TEMP ZERO) Obtain the K compensating value from the equation(1) below.

K = − ×100 ……… (a)h×Bspan

where,B: Constant value of fill fluid (See Table A.)span: |URV–LRV|h: Distance from high pressure side to low

pressure side (m)

EJX118A/EJA118E: Distance from high side of diaphragm seal to low side of diaphragm seal.

EJX438A/EJA438E: Distance from diaphragm seal (high side) to position of transmitter (low side).

Page 28: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-20

IM 01C25T03-01E

Transmitter

Lh

0H

(+)

EJX118A/EJA118E

F0353.ai

h

hL

L

H

Transmitter

Transmitter

0

(+)

(–)

EJX438A/EJA438E

Note: When the transmitter is positioned lower than the diaphragm seal part, the value of “h” must have a negative sign (–).

Enter “–0.149.”

Press the key twice

to enter the setting.

F0354.ai

ESC

SET E11:TEMP ZERO 0.000 %/degC - 00.149

CLR

OK

SET E11:TEMP ZERO -0.149 %/degC

NOFEED

K=−(+3)×0.00745÷15×100=−0.149

Press the (OK) key.

• Example: Enter K value obtained from the equation (a). A value having up to 3 decimal places may be specified.When h=+3 m, Fill fluid code A, span=15 kPa,

Note 1: The function is performed using a built-in temperature sensor in the transmitter body. The temperature deviation between the transmitter body and capillaries should be minimized to achieve optimal performance of the function.

Note 2: When the span changes, reenter the newly obtained value of K to E11: TEMP ZERO.

Table A. Constant value [B] of fill fluid

Fill fluid code

A, C, 1, 2, 4 B D E

Con

stan

t val

ue [B

]

mmH2O 0.76 0.87 1.45 0.75kgf/cm2 0.000076 0.000087 0.000145 0.000075kPa 0.00745 0.00853 0.01422 0.00736mBar 0.07453 0.08532 0.14220 0.07355atm 0.000074 0.000084 0.000140 0.000073inH2O 0.02992 0.03425 0.05709 0.02953psi 0.00108 0.00124 0.00206 0.00167mmHg 0.05592 0.06401 0.10669 0.05518

Note 3: Select the unit of constant value of [B] from the actual unit used for the transmitter in operation.

(21) Adjustment Information and User Memo Fields (J50: ADJ WHO, J51: ADJ DATE, J52: ADJ LOC, J53: ADJ DESC, M17 to M19: MEMO1 to MEMO3)

This feature provides four fields for instrument adjustment information at maintenance: inspection date, inspector, location, and description. Also three user memo fields are provided, each holding up to 16 alphanumeric characters.

• Example: Save an inspection date of October 21, 2003.

DATA DIAG PRNT ESC

SET J51:ADJ DATE MM-DD-YYYY 10-21-2003

ESCCODE CAPS CLR

PARAM J50:ADJ WHO

J51:ADJ DATE

J52:ADJ LOC

Set “10-21-2003” in the order of

month, day, and year.

Press the key twice to

enter the setting.

F0355.ai

3.3 Displaying Data Using the BT200

3.3.1 Displaying Measured DataThe BT200 can be used to display measured data.

The measured data is updated automatically every seven seconds. In addition, the display can be updated to the present data value at any time by pressing the (DATA) key. For parameters associated with the display of measured data, see chapter 5 Parameter Summary.

• Example: Display output.

DATA DIAG PRNT ESC

HOME SET ADJ ESC

F0356.ai

Data is updated automatically at 7-second intervals.

Display “A10: OUTPUT.”

communi

PARAM A10:OUTPUT 0.0 % A11:PRES 0.0000 kPa A15:OUTPUT mA 4.000 mA

PARAM A10:OUTPUT 0.0 % A11:PRES 0.0000 kPa A15:OUTPUT mA 4.000 mA

MENU A:DISPLAY B:SENSOR TYPE C:BASIC SETUP D:AUX SET1 E:AUX SET2 G:ALARM SET

Page 29: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<3. Operation> 3-21

IM 01C25T03-01E

3.3.2 Display Transmitter Model and Specifications

The BT200 can be used to display the model and specifications of the transmitter.

• Example: View transmitter model name.

HOME SET ADJ ESC

DATA DIAG PRNT ESC

PARAM B10:MODEL EJX110 M B11:STYLE NO. 1.00 B20:PRES LRL - 98.07 kPa

Press .

F0357.ai

For the associated parameters, see Chapter 5, Parameter Summary.

MENU A:DISPLAY B:SENSOR TYPE C:BASIC SETUP D:AUX SET1 E:AUX SET2 G:ALARM SET

Page 30: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<4. Self-diagnostics> 4-1

IM 01C25T03-01E

4. Self-diagnostics4.1 Checking for Problems

4.1.1 Identifying Problems with BT200The following four areas can be checked.

(a) Whether connections are good.(b) Whether BT200 was properly operated.(c) Whether settings were properly entered.(d) History of the errors.

See examples below.

• Example 1: Connection errors

ESC

communication error

Press the key.

When the panel shown on the

left appears, press the

key.

Since communications will be

unsuccessful if there is a problem

in the connection to the BT200,

the display at the left will appear.

Recheck the connection.

Press the (OK) key.

--WELCOME--BRAIN TERMINAL ID: BT200

check connection push ENTER key

UTIL FEED

F0401.ai

• Example 2: Setting entry errors

F0402.ai

The initial data panel shows the

result of current transmitter

diagnostics.

DATA DIAG PRNT ESC

PARAM C20:PRES UNIT kPa C21:PRES LRV 600 kPa C22:PRES URV 600 kPa

FEED PRNT ESC

DIAG C60:SELF CHECK ERROR < ERROR > < ILLEGAL LRV >

Press the (DIAG) key in the

parameter panel to go to the

diagnostics panel

(C60: SELF CHECK).

An error message is displayed

when an error occurs in the

diagnostics panel.

OK

PARAM 01:MODEL EJX110 M 02:TAG NO. YOKOGAWA 03:SELF CHECK ERROR

• Example 3: Checking the history of the errors

Connect the BT200 to the transmitter, and call item “P.”

The history of up to four errors can be stored. When the 5th error has occurred, it is stored in “P10.” The error stored in “P16” will be deleted, and then, the error in “P14”will be copied to “P16.” In this sequence, the history of the most previously occurred error will be removed from memory. “GOOD” will be displayed if there was no previous error.

F0403.ai

HOME SET ADJ ESC

MENU J:ADJUST K:TEST L:CALIBRATION M:MEMO P:RECORD T:CHARACTERIZR

DATA DIAG PRNT ESC

PARAM P10:ERROR REC 1 ERROR P12:ERROR REC 2 ERROR P14:ERROR REC 3 GOOD

For the details of the messages, see Table 4.1 Alarm Message Summary.

ESC

SET P10:ERROR REC 1 ERROR < ERROR > <50:P ILLEG LRV >

<(a) SETUP PANEL>

P10: “ERROR REC 1” displays the last error.P12: “ERROR REC 2” displays the error one time before

the last error occurred.P14: “ERROR REC 3” displays the error two times before

the last error occurred.P16: “ERROR REC 4” displays the error three times

before the last error occurred.

Select P10: ERROR REC1 and press the key to display the error message.

Note 1: Press the key twice in the setup panel (a) to clear all error message (P10 to P16) information.

Note 2: When the error occurs, the self-diagnostic detects errors and records them in two ways depending on

the types of errors. The amplifier/capsule failures are recorded immediately after the occurrence, while the minor errors such as warnings of inappropriate parameter settings are periodically recorded at an interval of minimum five minutes to twenty four hours.

Note that the interval extends as the number of access counts to EEPROM increases.

Page 31: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<4. Self-diagnostics> 4-2

IM 01C25T03-01E

4.1.2 Checking with Integral Indicator

NOTEIf an error is detected by running self-diagnostics, an error number is displayed on the integral indicator. If there is more than one error, the error number changes at three-second intervals.See table 4.1 regarding the alarm codes.

F0404.ai

Figure 4.1 Integral Indicator

4.2 Alarms and CountermeasuresTable 4.1 Alarm Message Summary

Indicator BT200 display Cause Output operation during error Countermeasure

None GOODAL. 01CAP. ERR

01: P-SENSOR ERR Sensor problem. Outputs the signal (Hi or Low) set with parameter D26.[Status output: undefined]

Replace capsule when error keep appearing error even after restart.

01: CT-SENSOR ERR Capsule temperature sensor problem.

Replace capsule.

01: C-EEPROM ERR Capsule EEPROM problem.AL. 02AMP. ERR

02: AT-SENSOR ERR Amplifier temperature sensor problem.

Replace amplifier.

02: A-EEPROM ERR Amplifier EEPROM problem.02: CPU BOARD ERR Amplifier problem.

AL. 10PRESS

10: P OVER SPEC Input is outside measurement range limit of capsule.

Outputs upper range limit (URL) or lower range limit (LRL).

Check input or replace capsule when necessary.

AL. 11ST. PRSS

11: SP OVER SPEC Static pressure exceeds limit. Continues to operate and output.

AL. 12CAP. TMP

12: CT OVER SPEC Capsule temperature is outside range (–50 to 130°C).

Use heat insulation or make lagging to keep temperature within range.

AL. 13AMP. TMP

13: AT OVER SPEC Amplifier temperature is outside range (–50 to 95°C).

AL. 30RANGE

30: P OVER RANGE Output is outside upper or lower range limit value.

Outputs upper range value (URV) or lower range value (LRV).

Check input and range setting, and change them as needed.

AL. 31SP. RNG

31: SP OVER RANGE Static pressure exceeds specified range.

Continues to operate and output.

AL. 35P. HI

35: P HIGH ALARM Input pressure exceeds specified threshold.

Check input.

AL. 36P. LO

36: P LOW ALARM

AL. 37SP. HI

37: SP HIGH ALARM Input static pressure exceeds specified threshold.

AL. 38SP. LO

38: SP LOW ALARM

AL. 39TMP. HI

39: CT HIGH ALARM Detected temperature exceeds specified threshold.

Check capsule temperature.

AL. 40TMP. LO

40: CT LOW ALARM

Page 32: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<4. Self-diagnostics> 4-3

IM 01C25T03-01E

Indicator BT200 display Cause Output operation during error Countermeasure

AL. 50P. LRV

50: P ILLEG LRV Specified value is outside of setting range.

Holds output immediately before error occurred.

Check settings and change them as needed.

AL. 51P. URV

51: P ILLEG URV

AL. 52P. SPN

52: P ILLEG SPAN

AL. 53P. ADJ

53: P SPAN ADJ Continues to operate and output.

Adjust settings and change them as needed.53: P ZERO ADJ

AL. 54SP. RNG

54: SP ILLEG LRV Continues to operate and output holding static pressure in %.

Check settings and change them as needed.54: SP ILLEG URV

54: SP ILLEG SPANAL. 55SP. ADJ

55: SP SPAN ADJ Continues to operate and output.

Adjust settings and change them as needed.55: SP ZERO ADJ

AL. 60SC. CFG

60: SC CONFIG ERR Specified values or settings do not meet the conditions.

Continues to operate and output without signal characterizing.

Check settings and change them as needed.

AL. 79OV. DISP

— Displayed value exceeds limit. Continues to operate and output.

Page 33: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<5. Parameter Summary> 5-1

IM 01C25T03-01E

5. Parameter SummaryInstruments to which applicable:

F: Differential pressure transmittersP: Absolute and gauge pressure transmittersL: Flange mounted differential pressure transmitters

No. Parameter name Item*1

R/W Content Default valueApplicable

model

Upl

oad

data

F P L01 MODEL Model R EJX (for EJX series)

EJA (for EJA series) —

02 TAG No. Tag number R As specified —03 SELF CHECK Self-diagnostics R GOOD —A DISPLAY Measured data displayA10 OUTPUT Output (in %) R -2.5 to 110% —A11 PRES Measured pressure after zero

adjustmentR Unit specified in C20 —

A15 OUTPUT mA Output current R 3.600 to 21.600 mA —A16 ENGR. OUTPUT User scaled value R Unit specified in I30 —A17 ENGR. EXP Exponents R Unit specified in I32 —A20 SP % Static pressure (in %) R -10 to 110% — —A21 SP Static pressure after zero

adjustmentR Unit specified in D30 — —

A30 CAPSULE TEMP Capsule temperature R Unit specified in D40 —A60 SELF CHECK Self-diagnostics R Refer to Table 4.1 Alarm

Message Summary —

B SENSOR TYPE Sensor typeB10 MODEL Model and capsule type R Model and capsule type —B11 STYLE NO. Style number R Style number of product —B20 PRES LRL Lower range limit R Unit specified in C20 —B21 PRES URL Upper range limit R Unit specified in C20 —B22 P MIN SPAN Minimum span R Unit specified in C20 —B30 SP LRL Lower range limit for static

pressureR Unit specified in D30 — —

B31 SP URL Upper range limit for static pressure

R Unit specified in D30 — —

B32 SP MIN SPAN Minimum span for static pressure

R Unit specified in D30 — —

B60 SELF CHECK Self-diagnostics R See A60 —C BASIC SETUP Setting dataC10 TAG NO. Tag number W 16 alphanumeric characters As specified

C20 PRES UNIT Measurement range unit W mmH2O, mmAq, mmWG, mmHg, Torr, kPa, MPa, mbar, bar, gf/cm2, kgf/cm2, inH2O, inHg, ftH2O, psi, atm, Pa, hPa

kPa

C21 PRES LRV Lower range value W -32000 to 32000 within measurement range

As specified

C22 PRES URV Upper range value W -32000 to 32000 within measurement range

As specified

C23 PRES POINT Decimal place W 0 to 4 2

C30 AMP DAMPING Damping time constant at amplifier

W 0.50(0.00) to 100.00 seconds, see D50

2.00 seconds or as specified

C40 OUTPUT MODE Output mode W LINEAR or SUQARE ROOT LINEAR or as specified

C60 SELF CHECK Self-diagnostics R See A60 —*1: R/W: R = Read only, W = Read & Write*2: The default value shows MWP (Maximum working pressure) of the capsule.

Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.

Page 34: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<5. Parameter Summary> 5-2

IM 01C25T03-01E

No. Parameter name Item*1

R/W Content Default valueApplicable

model

Upl

oad

data

F P LD AUX SET 1 Auxiliary setting data 1D10 LOW CUT Low cut W 0.00 to 20.00% 10.00%

D11 LOW CUT MODE Low cut mode W LINEAR or ZERO LINEAR

D15 H/L SWAP Impulse piping accessing direction

W NORMAL or REVERSE NORMAL —

D16 H2O UNIT SEL H2O unit select W @4degC or @20degC (68.0F)

@4degC

D20 OUT LIMIT (L) Low side output limiter W -2.50 to 110.00% -2.50%

D21 OUT LIMIT (H) High side output limiter W -2.50 to 110.00% 110%

D22 REV OUTPUT Output reversal W NORMAL or REVERSE NORMAL

D25 BURNOUT CPU error R HIGH or LOW —D26 ERROR OUT Hardware error W BURNOUT DIR or HOLD BURNOUT DIR

D30 SP UNIT Static pressure unit W See C20 MPa —

D31 SP A/G SLCT Gauge/Abs select for static pressure

W GAUGE or ABSOLUTE ABSOLUTE —

D32 ATM. PRESS Coefficient for given gauge pressure

W Unit specified in D30 0.10133 MPa —

D33 SP LRV Lower limit of static pressure W -32000 to 32000 within measurement range

0.0 MPa —

D34 SP URV Upper limit of static pressure*2 W -32000 to 32000 within measurement range

D35 SP POINT Decimal place of static pressure

W 0 to 4 1 —

D36 SP DAMPING Damping time constant of SP W 0.00 to 100.00 seconds 2.00 seconds —

D37 SP SELECT H/L select for static pressure W HIGH or LOW HIGH —

D40 TEMP UNIT Temperature setting unit W degC, degF, or K degC

D50 QUICK RESP Quick response W OFF or ON (enable 0.00 to 0.50 seconds at C30)

OFF

D55 WRT PROTECT Write protect indicator R NO or YES NO —D56 WRT ENABLE Write protect release W 8 alphanumeric characters None —D57 NEW PASSWORD User set password for write

protectW 8 alphanumeric characters None —

D58 SOFTWR SEAL Software seal R BREAK or KEEP KEEP —D60 SELF CHECK Self-diagnostics R See A60 —E AUX SET 2 Auxiliary setting data 2E10 T. ZERO CMP Temperature compensation

modeW OFF or ON OFF —

E11 TEMP ZERO Zero shift compensation W -99.999 to 99.999%/degC 0.000%/degC —E30 BI DIRE MODE Bidirectional mode W OFF or ON OFF —E50 DO SELECT Contact output select W INHIBIT, PRES, SP, TEMP,

PRES/SP, PRES/TEMP, SP/TEMP, or PRES/SP/TEMP

INHIBIT —

E51 DO SIG. TYPE Signal type select W OFF WHEN ALARM or ON WHEN ALARM

ON WHEN ALARM

E52 D OUTPUT Contact output R OFF or ON OFF —E60 SELF CHECK Self-diagnostics R See A60 —*1: R/W: R = Read only, W = Read & Write*2: The default value shows MWP (Maximum working pressure) of the capsule.

Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.

Page 35: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<5. Parameter Summary> 5-3

IM 01C25T03-01E

No. Parameter name Item*1

R/W Content Default valueApplicable

model

Upl

oad

data

F P LG ALARM SET Alarm settingG10 P AL MODE Alert mode W INHIBIT, HI. AL DETECT, LO.

AL DETECT, or HI/LO. AL DETECT

INHIBIT —

G11 P HI. AL VAL High side alert value W -32000 to 32000, unit specified in C20

100.000 kPa —

G12 P LO. AL VAL Low side alert value W -32000 to 32000, unit specified in C20

-100.000 kPa —

G20 SP AL MODE Static pressure alert mode W INHIBIT, HI. AL DETECT, LO. AL DETECT, or HI/LO. AL DETECT

INHIBIT — —

G21 SP HI. AL VAL High side alert value of SP*2 W -32000 to 32000, unit specified in D30

— —

G22 SP LO. AL VAL Low side alert value of SP W -32000 to 32000, unit specified in D30

0.00000 MPa — —

G30 T AL MODE Temperature alert mode W INHIBIT, HI. AL DETECT, LO. AL DETECT, or HI/LO. AL DETECT

INHIBIT —

G31 T HI. AL VAL High side alert value of temperature

W -50 to 130 120 degC —

G32 T LO. AL VAL Low side alert value of temperature

W -50 to 130 -40 degC —

G50 AUTO RECOVER Auto-recover from sensor error W OFF or ON ON —G60 SELF CHECK Self-diagnostics R See A60 —H AUTO SET Automatic setupH10 AUTO P LRV Lower range value auto setup W -32000 to 32000, unit

specified in C20As specified —

H11 AUTO P URV Upper range value auto setup W -32000 to 32000, unit specified in C20

As specified —

H20 AUTO SP LRV SP lower range value auto setup

W -32000 to 32000, unit specified in D30

0.00000 MPa — —

H21 AUTO SP URV SP upper range value auto setup*2

W -32000 to 32000, unit specified in D30

— —

H60 SELF CHECK Self-diagnostics R See A60 —I DISP SET Display settingI10 DISP OUT1 LCD output 1 W PRES, PRES %, ENGR.

PRES, SP, or SP %PRES %

I11 DISP OUT2 LCD output 2 W PRES, PRES %, ENGR. PRES, SP, SP %, or ---

---

I12 DISP OUT3 LCD output 3 W See I11 ---

I13 DISP OUT4 LCD output 4 W See I11 ---

I20 P DISP MODE % display mode W LINEAR or SQUARE ROOT LINEAR

I21 PRES % RESO % display resolution W NORMAL or HIGH RESOLUTION

NORMAL

I30 ENGR. UNIT User set engineering unit W 8 alphanumeric characters

I31 EASY EU SET Engineering unit select W —I32 ENGR. EXP Exponents W ---, ×10, ×100, ×1000 ---

I33 ENGR. LRV User set lower range limit W -32000 to 32000, unit specified in I30

0.00

I34 ENGR. URV User set upper range limit W -32000 to 32000, unit specified in I30

100.00

I35 ENGR. POINT Decimal place of user set W 0 to 4 1

I40 BAR INDICATR Bar indicator W OFF or ON ON

I41 POWER ON INF Display when powering on W OFF or ON ON —I60 SELF CHECK Self-diagnostics R See A60 —*1: R/W: R = Read only, W = Read & Write*2: The default value shows MWP (Maximum working pressure) of the capsule.

Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.

Page 36: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<5. Parameter Summary> 5-4

IM 01C25T03-01E

No. Parameter name Item*1

R/W Content Default valueApplicable

model

Upl

oad

data

F P LJ ADJUST Adjusting dataJ09 ADJ UNIT Pressure adjusting unit select W % or PRES UNIT PRES UNIT —J10 ADJ PRES Adjustment reference pressure R Unit specified in J09 —J11 P ZERO ADJ Automatic zero adjustment W -32000 to 32000, unit

specified in J090.00000 kPa —

J12 P SPAN ADJ Automatic span adjustment W -32000 to 32000, unit specified in J09

100.000 kPa —

J15 P ZERO DEV Manual zero adjustment W -32000 to 32000, unit specified in J09

100.000 kPa —

J16 P SPAN DEV Manual span adjustment W -32000 to 32000, unit specified in J09

100.000 kPa —

J20 ADJ SP Adjustment reference pressure of SP

R Unit specified in J09 — —

J21 SP ZERO ADJ Automatic SP zero adjustment W -32000 to 32000, unit specified in J09

0.00000 MPa — —

J22 SP SPAN ADJ Automatic SP span adjustment*2

W -32000 to 32000, unit specified in J09

— —

J25 SP ZERO DEV Manual SP zero adjustment W -32000 to 32000, unit specified in J09

0.00000 MPa — —

J26 SP SPAN DEV Manual SP span adjustment W -32000 to 32000, unit specified in J09

0.00000 MPa — —

J40 OUTPUT 4mA 4 mA adjustment W -10.000 to 10.000% 0.000% —J41 OUTPUT 20mA 20 mA adjustment W -10.000 to 10.000% 0.000% —J45 AMP TEMP Amplifier temperature R Unit specified D40 0 degC —J50 ADJ WHO Adjustment information W 8 alphanumeric characters —J51 ADJ DATE Adjustment information W 16 alphanumeric characters —J52 ADJ LOC Adjustment information W 8 alphanumeric characters —J53 ADJ DESC Adjustment information W 16 alphanumeric characters —J55 EXT ZERO ADJ External zeroing permission W INHIBIT or ENABLE ENABLE —J56 CLEAR ADJ Clear adjustment W ---, PRES, SP, 4-20mA, or

ALL--- —

J60 SELF CHECK Self-diagnostics R See A60 —K TEST Test parametersK10 OUTPUT X % Test output % setting W Within a range between D20

and D210.00% —

K40 DO TEST Test contact output W OFF or ON OFF —K45 TEST TIME “OUTPUT X %” and “DO

TEST” duration time selectionW 10 min, 30 min, 60 min,

3 hour, 6 hour, 12 hour10 min —

K50 TEST KEY1 Special maintenance parameter

W —

K51 TEST KEY2 Special maintenance parameter

W —

K52 TEST KEY3 Special maintenance parameter

W —

K53 TEST KEY4 Special maintenance parameter

W —

K60 SELF CHECK Self-diagnostics R See A60 —M DEVICE INFO Device informationM10 SERIAL NO. Serial number R —M11 MFTR. DATE Manufactured date R —M12 EXTRA NO. Customization number R —M15 SOFT REV Software revision R —M16 BRAIN REV BRAIN protocol revision R —M17 MEMO1 Memo W 16 alphanumeric characters —M18 MEMO2 Memo W 16 alphanumeric characters —M19 MEMO3 Memo W 16 alphanumeric characters —M20 ISOL MATL Capsule material W —M21 FILL FLUID Fill fluid W —*1: R/W: R = Read only, W = Read & Write*2: The default value shows MWP (Maximum working pressure) of the capsule.

Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.

Page 37: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<5. Parameter Summary> 5-5

IM 01C25T03-01E

No. Parameter name Item*1

R/W Content Default valueApplicable

model

Upl

oad

data

F P LM22 GASKET MATL Gasket material W —M23 PRO CON MATL Flange material W —M24 D-VENT MATL Vent plug material W —M25 PRO CON TYPE Process connection type W —M26 RS ISOL MATL Remote seal material W —M27 PRO CON SIZE Flange size W —M28 NUM RS Number of remote seal W —M29 RS FILL FLUID Fill fluid of remote seal W —M30 RS TYPE Remote seal type W —M50 MS CODE 1 Model and suffix code 1 W —M51 MS CODE 2 Model and suffix code 2 W —M52 MS CODE 3 Model and suffix code 3 W —M53 MS CODE 4 Model and suffix code 4 W —M54 MS CODE 5 Model and suffix code 5 W —M55 MS CODE 6 Model and suffix code 6 W —M60 SELF CHECK Self-diagnostics R See A60 —P Record History of errorsP10 ERROR REC 1 Last error W See A60 GOOD —P12 ERROR REC 2 Second recent error W See A60 GOOD —P14 ERROR REC 3 Third recent error W See A60 GOOD —P16 ERROR REC 4 Forth recent error W See A60 GOOD —P60 SELF CHECK Self-diagnostics R See A60 —T CHARACTERIZR Signal characterizer settingT10 S. C. ENABLE Signal characterizer permission W INHIBIT or ENABLE INHIBIT —T11 NUM OF POINT Number of coordinates W 0 to 9 0 —T20 X START (FIX) Start point of X R 0.00% —T21 Y START (FIX) Start point of Y R 0.00% —T22 X1 Coordinate 1 of X W 0.00 to 100.00% 10.00 —T23 Y1 Coordinate 1 of Y W 0.00 to 100.00% 10.00 —T24 X2 Coordinate 2 of X W 0.00 to 100.00% 20.00 —T25 Y2 Coordinate 2 of Y W 0.00 to 100.00% 20.00 —T26 X3 Coordinate 3 of X W 0.00 to 100.00% 30.00 —T27 Y3 Coordinate 3 of Y W 0.00 to 100.00% 30.00 —T28 X4 Coordinate 4 of X W 0.00 to 100.00% 40.00 —T29 Y4 Coordinate 4 of Y W 0.00 to 100.00% 40.00 —T30 X5 Coordinate 5 of X W 0.00 to 100.00% 50.00 —T31 Y5 Coordinate 5 of Y W 0.00 to 100.00% 50.00 —T32 X6 Coordinate 6 of X W 0.00 to 100.00% 60.00 —T33 Y6 Coordinate 6 of Y W 0.00 to 100.00% 60.00 —T34 X7 Coordinate 7 of X W 0.00 to 100.00% 70.00 —T35 Y7 Coordinate 7 of Y W 0.00 to 100.00% 70.00 —T36 X8 Coordinate 8 of X W 0.00 to 100.00% 80.00 —T37 Y8 Coordinate 8 of Y W 0.00 to 100.00% 80.00 —T38 X9 Coordinate 9 of X W 0.00 to 100.00% 90.00 —T39 Y9 Coordinate 9 of Y W 0.00 to 100.00% 90.00 —T40 X END (FIX) End point of X R 100.00% —T41 Y END (FIX) End point of Y R 100.00% —T60 SELF CHECK Self-diagnostics R See A60 —*1: R/W: R = Read only, W = Read & Write*2: The default value shows MWP (Maximum working pressure) of the capsule.

Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.

Page 38: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<Appendix 1. Safety Instrumented Systems Installation> A1-1

IM 01C25T03-01E

Appendix 1. Safety Instrumented Systems Installation

WARNING

The contents of this appendix are cited from exida.com safety manual on the transmitters specifically observed for the safety transmitter purpose. When using the transmitter for Safety Instrumented Systems (SIS) application, the instructions and procedures in this section must be strictly followed in order to preserve the transmitter for that safety level.

A1.1 Scope and PurposeThis section provides an overview of the user responsibilities for installation and operation of the transmitter in order to maintain the designed safety level for Safety Instrumented Systems (SIS) applications. Items that will be addressed are proof testing, repair and replacement of the transmitter, reliability data, lifetime, environmental and application limits, and parameter settings.

A1.2 Using the Transmitter for an SIS Application

A1.2.1 Safety AccuracyThe transmitter has a specified safety accuracy of 2%. This means that the internal component failures are listed in the device failure rate if they will cause an error of 2% or greater.

A1.2.2 Diagnostic Response TimeThe transmitter will report an internal failure within 5 seconds of the fault occurrence.

A1.2.3 SetupDuring installation the transmitter must be setup with engineering units parameters. This is typically done with a handheld terminal. These parameters must be verified during the installation to insure that the correct parameters are in the transmitter. Engineering range parameters can be verified by

reading these parameters from the optional local display or by checking actual calibration of the transmitter.

The calibration of the transmitter must be performed after parameters are set.

A1.2.4 Required Parameter SettingsThe following parameters need to be set in order to maintain the designed safety integrity.

Table A1.1 Required Parameter Settings

Item DescriptionBurnout direction switch

To specify if the output should go 21.6 mA or higher or 3.6 mA or lower upon detection of an internal failure.

Write protection switch

The write function should be disabled.

A1.2.5 Proof TestingThe objective of proof testing is to detect failures within the transmitter that are not detected by the diagnostics of the transmitter. Of main concern are undetected failures that prevent the safety instrumented function from performing its intended function. See table A1.2 for proof testing method.

The frequency of the proof tests (or the proof test interval) is to be determined in the reliability calculations for the safety instrumented functions for which the transmitter is applied. The actual proof tests must be performed more frequently or as frequently as specified in the calculation in order to maintain required safety integrity of the safety instrumented function.

The following tests need to be specifically executed when a proof test is performed. The results of the proof test need to be documented and this documentation should be part of a plant safety management system. Failures that are detected should be reported to Yokogawa.

The personnel performing the proof test of the transmitter should be trained in SIS operations including bypass procedures, transmitter maintenance, and company management of change procedures.

Page 39: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<Appendix 1. Safety Instrumented Systems Installation> A1-2

IM 01C25T03-01E

Table A1.2 Proof Testing

Testing method Tools required Expected outcome RemarksFunctional test:

1. Follow all Management of Change procedures to bypass logic solvers if necessary.

2. Execute HART/BRAIN command to send value to high alarm (21.5 mA) and verify that current has reached this level.

3. Execute HART/BRAIN command to send value to low alarm (3.6 mA) and verify that current has reached this level.

4. Restore logic solvers operation and verify.

• Handheld terminal Proof Test Coverage =52% The output needs to be monitored to assure that the transmitter communicates the correct signal.

Perform three point calibration along with the functional test listed above.

• Handheld terminal • Calibrated pressure

source

Proof Test Coverage =99%

A1.2.6 Repair and ReplacementIf repair is to be performed with the process online the transmitter will need to be bypassed during the repair. The user should setup appropriate bypass procedures.

In the unlikely event that the transmitter has a failure, the failures that are detected should be reported to Yokogawa.

When replacing the transmitter, the procedure in the installation manual should be followed.

The personnel performing the repair or replacement of the transmitter should have a sufficient skill level.

A1.2.7 Startup TimeThe transmitter generates a valid signal within 1 second of power-on startup.

A1.2.8 Firmware UpdateIn case firmware updates are required, they will be performed at factory. The replacement responsibilities are then in place. The user will not be required to perform any firmware updates.

A1.2.9 Reliability DataA detailed Failure Mode, Effects, and Diagnostics Analysis (FMEDA) report is available from Yokogawa with all failure rates and failure modes.

The transmitter is certified up to SIL2 for use in a simplex (1oo1) configuration, depending on the PFDavg calculation of the entire Safety Instrumented Function.

The development process of the transmitter is certified up to SIL3, allowing redundant use of the transmitter up to this Safety Integrity Level, depending the PFDavg calculation of the entire Safety Instrumented Function.

When using the transmitter in a redundant configuration, the use of a common cause factor (β-factor) of 2% is suggested. (However, if the redundant transmitters share an impulse line or if clogging of the separate impulse lines is likely, a common cause factor of 10% is suggested.)

Note that the failure rates of the impulse lines need to be accounted for in the PFDavg calculation.

A1.2.10 Lifetime LimitsThe expected lifetime of the transmitter is 50 years. The reliability data listed the FMEDA report is only valid for this period. The failure rates of the transmitter may increase sometime after this period. Reliability calculations based on the data listed in the FMEDA report for the transmitter lifetimes beyond 50 years may yield results that are too optimistic, i.e. the calculated Safety Integrity Level will not be achieved.

A1.2.11 Environmental LimitsThe environmental limits of the transmitter are specified in the user’s manual IM 01C25.

A1.2.12 Application LimitsThe application limits of the transmitters are specified in the user’s manual IM 01C25. If the transmitter is used outside of the application limits, the reliability data listed in A1.2.9 becomes invalid.

Page 40: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

<Appendix 1. Safety Instrumented Systems Installation> A1-3

IM 01C25T03-01E

A1.3 Definitions and Abbreviations

A1.3.1 Definitions

Safety

Freedom from unacceptable risk of harm

Functional Safety

The ability of a system to carry out the actions necessary to achieve or to maintain a defined safe state for the equipment/machinery/plant/apparatus under control of the system

Basic Safety

The equipment must be designed and manufactured such that it protects against risk of damage to persons by electrical shock and other hazards and against resulting fire and explosion. The protection must be effective under all conditions of the nominal operation and under single fault condition

Verification

The demonstration for each phase of the life-cycle that the (output) deliverables of the phase meet the objectives and requirements specified by the inputs to the phase. The verification is usually executed by analysis and/or testing

Validation

The demonstration that the safety-related system(s) or the combination of safety-related system(s) and external risk reduction facilities meet, in all respects, the Safety Requirements Specification. The validation is usually executed by testing

Safety Assessment

The investigation to arrive at a judgment -based on evidence- of the safety achieved by safety-related systems

Further definitions of terms used for safety techniques and measures and the description of safety related systems are given in IEC 61508-4.

A1.3.2 Abbreviations

FMEDA

Failure Mode, Effects and Diagnostic Analysis

SIF

Safety Instrumented Function

SIL

Safety Integrity Level

SIS

Safety Instrumented System

SLC

Safety Lifecycle

Page 41: DPharp BRAIN Communication Type (EJX A, EJA E) · that postponement of revisions will not cause difficulty to the user from a functional or performance standpoint. ... Communication

i

IM 01C25T03-01E

Revision Information Title : DPharp BRAIN Communication Type Manual No. : IM 01C25T03-01E

Edition Date Page Revised Item1st Apr. 2004 — New publication.2nd Oct. 2004 3-19 3.2.3(20) • Add capillary fill fluid density compensation setting procedure.3rd Aug. 2009 3-8 3.2.3(5) • Add example for hysteresis.

3-14 3.2.3(15) • Correct misprint.3-18 3.2.3(19) • Add CAUTION. Add note for hysteresis.

4th Jun. 2012 2-2 2.4 • Add integral indicator display when powering on3-5 3.2.2 • Add parameters in the menu tree (I41, K45)3-6 3.2.3(1) • Correct errors3-12 3.2.3(11)d • Change description for SOFTWARE SEAL3-15 3.2.3(15)a-2 • Correct the figure3-17 3.2.3(16) • Correct the NOTE3-20 3.2.3(20) • Add EJA model name5-1 5 • Add EJA to default value of parameter No. "01"5-3 5 • Add parameter I415-4 5 • Add parameter K45

5th Jun. 2013 3-20 3.2.3(20) • Add constant value of fill fluid for high vacuum use diaphragm sealed differential pressure/pressure transmitters.

6th June 2014 2-1 2.1 • Change terminal drawing.2-1, 3-1 Add Note for BRAIN communication.