Dose Survey in Computed Tomography
DS /CM
Kampala
IAEA/RCA Kampala
Dose Surveys (Have you seen this slide?)
• Measure doses to patients.
• Compare measured doses to standards
• Decide if action is required
• Begin again
Kampala
Dose Survey – What to think about
• What should be Surveyed– Examinations– Equipment– Patients
• Data collection• Data analysis• Reporting• Investigating
Kampala
Surveys: Planning
• Identify scanners to be surveyed
• Identify procedures carried out in those rooms that you wish to survey
• Decide how to obtain Dose information
• Submit data for reference values
• OR compare against reference values
Kampala
Planning a dose Survey in CT
• Examinations• What examinations are you going to
survey?
Kampala
Criteria for inclusion
• Examinations must be performed reasonably frequently in your hospital.
• Data collection must be feasible
Kampala
Your Presentations
• Brain• Lung• Lumbar Spine• Abdomen
Kampala
Relative Contributions – 5 Scanners
Head Body Other N
NWIP 52% 40% 8% 1201
NWOP 24% 69% 7% 757
WGH 38% 58% 4% 1110
STJ 40% 54% 6% 866
DCN 85% 6% 9% 968
Kampala
Dundee Exams
Kampala
Edinburgh Exams
Kampala
An issue for PACS / RIS based surveys
• What are all these codes?
Kampala
There are a large number of CT exams
• Anatomical Groups / Protocol Types– Head– Abdomen– Chest– Chest & Pelvis– Angiography– Choice of correct
exam important
Kampala
Choice of exams - IAEA
• “In order to allow comparison of truly similar examinations conducted for similar purpose and requiring similar scan technique, you should specify detailed descriptions of CT procedures, including a clinical indication (such as CT abdomen in relation to liver metastases), rather than simply broad categories of examination (such as CT abdomen).”
• You must compare like with like, or as near as possible !! (This can be difficult / even impossible)
Kampala
Planning a dose Survey in CT
• Patient Selection
Kampala
Patient selection
• Standard sized patients
• Define weight range, or take everyone (cf Australia), but exclude very large and very small
• Make measurements / collect data for at least 10 patients – minimum of 20 preferable – more if no weight constraint.
• RIS enables thousands to be selected.Kampala
Dose Survey – What to think about
• What should be Surveyed– Examinations– Equipment– Patients
• Data collection• Data analysis• Reporting• Investigating
Kampala
Planning a dose Survey in CT
• How are you going to collect the dose data ?
• What Physical Quantities are you going to record?
Kampala
Kampala
Dose Length Product & CTDIvol
• Are the dose related quantity measured and displayed on all modern CT scanners.
• Stored on DICOM Header • Can be transcribed to RIS system• Can be looked up on PACS using DICOM
query retrieve interrogation.• Can be calculated, not going to show you
how – see TRS 457.
Data Collection – What quantity
• DLP• CTDIvol• What are the relative merits of each?
Kampala
Data Collection – What quantity• CTDIvol – In My Opinion CTDIvol is best for
optimising your protocol• DLP – In My Opinion, DLP will tell you about
variations introduced by different patients and is therefore best for dose audit.
• DLP is analagous to KAP in conventional radiology.
Kampala
Dose Summary Sheet - Commercial CT Scanner
Huda W , Mettler F A Radiology 2011;258:236-242
Kampala
Patient dose Survey in CT
• common scans• patient weight, sex• collect
– CTDIvol total DLP,
• (calculate exam CTDIvol
– Σ CTDIvol,i for overlapping ranges
– do not mix head and body CTDIvol)
Kampala
Planning a dose Survey in CT
• How are you going to collect the dose data ?
Kampala
Data Collection – What’s the protocol?
Kampala
How to get the data ?
Kampala
Data Collection Paper or Electronic
– Paper• Data tailored to needs• Prospective• ? Accuracy of data entry
– RIS• Retrospective• Patient numbers• ?? Patient weight, kV/mAs• ? Accuracy and consistency of data entry (units)
– PACS• Accurate and consistent data• Patient size?
Kampala
Very Simple Paper Entry Form
Kampala
Paper Entry – Get the design right!
Kampala
Form 2: Patient data acquisition form One entry of this form should be filled in for each patient that receives a ‘standard’ CT CAP exam. For each patient entry, the date, CHI and DLP should be recorded. Only patients receiving the ‘standard’ CAP exam should be included in the Survey. If the patient’s CT exam deviates from the standard protocol, e.g. receives additional scans, the patient should not be included in the Survey. Patient size: we ask that only patients of average size are included in the Survey. Very large and very small patients should be excluded from the Survey. We are not asking you to weigh the patients, simply exercise your professional judgement on what an ‘average’ sized patient is. The position of the patient’s arms should be consistent for all patients in the Survey. If the protocol states that the patient’s arms should be above their head for the exam, only patients who meet this criterion should be included. Data from 30 patients per site are required for the Survey.
Large scale data collection
• It will very likely that you will have transcription errors
• Incorrect data entry• Zeros and blanks• Multiple exposure asigned to single
exam• Lack of patient information
– Abnormal patients – eg very large
Kampala
Kampala
Clean up data
Kampala
Filter the data
• The results of the Survey will only be as good as the data that goes into it!!
• It can be quite time consuming to extract the good data!
Kampala
“Cleaned” – exclude ridiculous numbers and zeros
Kampala
Dose Survey Body Scans Ninewells Hospital, Dundee
Kampala
Patient Size.
• CTDI is a measure of scanner output.• CTDI does take patient size into account• CTDI is NOT patient dose• Important for paediatrics and possibly bariatrics.• You can correct to SSDE
– Only going to mention, not discuss.
Kampala
Patient Size
• CTDIvol tells you about scanner output.
• It does not address patient size, so no dose information.
• For smaller paediatrics, interpreting DLP as patient dose could lead to underestimate of patient dose levels by a factor of 2-3 if the 32cm phantom is used for reference.
Kampala
How is CTDIvol related to patient dose?
• CTDIvol is not patient dose
• The relationship between the two depends on many factors, including patient size and composition
• AAPM Report 204 introduces a parameter known as the Size Specific Dose Estimate (SSDE) to allow estimation of patient dose based on CTDIvol and patient size
• For the same CTDIvol, a smaller patient will tend to have a higher patient dose than a larger patient
How is CTDIvol related to patient dose?
Both patients scanned with the same CTDIvol Patient dose will be higher for the smaller patient
CTDIvol = 20 mGy CTDIvol = 20 mGy
120 kVp at 200 mAs
120 kVp at 200 mAs
32 cm Phantom
32 cm Phanto
m
How is CTDIvol related to patient dose?
Smaller patient scanned with a lower CTDIvol Patient doses will be approximately equal
CTDIvol = 10 mGy CTDIvol = 20 mGy
120 kVp at 100 mAs
120 kVp at 200 mAs
32 cm Phantom
32 cm Phantom
Equivalent cylinder diameter
• Patient diameter can be equated to a standard cylinder of the same length and weight, if other data not available
• Equivalent diameter = 2√[Weight/( Height)]
• OR 2√[APxLat] – easy in CT
h
Effective Diameter – 32cm reference
Kampala
Effective Diameter – 16 cm reference
Kampala
Free from ww.aapm.org and on your DVD
Kampala
32 cm Phantom
32 cm Phanto
m
How is CTDIvol related to patient dose?
Patients have equivalent SSDE
CTDIvol = 10 mGySSDE = 13.2 mGy
CTDIvol = 20 mGySSDE = 13.2 mGy
120 kVp at 100 mAs
27 cm9 cm
120 kVp at 200 mAs
Patient size specific protocols
• different protocols for adults and children– different image quality requirements
• differences in pathology (contrast, size)• poorer organ delineation• use of contrast agents
Kampala
Dose Survey – What to think about
• What should be Surveyed– Examinations– Equipment– Patients
• Data collection• Data analysis• Reporting ✔• Investigating
Kampala
When doses are too high or too low: check…
• scan range – DLP too high, exam
CTDIvol OK
• diagnostic task– e.g. brain CT
Kampala
When doses are too high or too low: check…
• scanner performance– tube current modulation
Kampala
When dose does not change with patient size: check…
• If mA modulation is enabled– patient size, z and xy modulation– appropriate to anatomy
• head• Pelvis
• If no mA modulation– protocol, protocol, protocol
• If mA modulation– Protocol, protocol, protocol
Kampala
When doses are too high or too low: reference protocols
• DO NOT transfer protocols between scanners– unless same model
• SET UP a new scan protocol– match
• kVp, slice thickness, beam collimation, beam shaping filter, pitch
– adjust mAs / noise indicator• to match CTDIvol and DLP
• may need phantom
– select reconstruction kernel to match resolution and image noise
Kampala
Summary
• Dose Survey required:– To set Local DRLs and test compliance– Optimise new equipment/ protocols
• Choice of examinations and equipment• Patient selection and sample size• Data collection: paper or RIS• Need for special attention for
paediatrics
Kampala
WHATEVER YOU DO
• Know your scanner…!
• Collect scanner displayed values of DLP and CTDIvol
• Adjust data collection forms and information for scanner type (so radiographers understand clearly which parameters to record)