L234 test report compliance to EN 12469
Document
Number: 20 460 234 Document
Version:
1.0 Document
Responsible
TGA
Page 1 of 12
K:\UDVIKLING INF\Projekter\Projekt 55 L200 opdatering\Analyse\L234 test report compliance to EN 12469.docx
L234 test report compliance to EN 12469
Document
Number: 20 460 234 Document
Version:
1.0 Document
Responsible
TGA
Page 2 of 12
K:\UDVIKLING INF\Projekter\Projekt 55 L200 opdatering\Analyse\L234 test report compliance to EN 12469.docx
Content
1 Scope ............................................................................................................................................ 3
2 Definitions and Acronyms ............................................................................................................ 3
3 How to obtain compliance to the Standard ................................................................................. 3
3.1 Type testing ........................................................................................................................... 3
3.1.1 Retention at front aperture ........................................................................................... 3
3.1.2 Leaktightness of carcass ................................................................................................. 3
3.1.3 Filters .............................................................................................................................. 4
3.1.4 Product protection ......................................................................................................... 4
3.1.5 Cross contamination ...................................................................................................... 4
3.2 Installation testing ................................................................................................................. 4
3.2.1 Volumetric airflow rate measurements ......................................................................... 4
3.2.2 Check airflow patterns ................................................................................................... 4
3.3 Guidance on manufacture ..................................................................................................... 4
3.3.1 Front aperture height .................................................................................................... 4
3.3.2 Lightning ......................................................................................................................... 4
3.3.3 Sound ............................................................................................................................. 5
3.3.4 Vibration ......................................................................................................................... 5
3.3.5 Stability........................................................................................................................... 5
3.3.6 Materials, general .......................................................................................................... 5
3.3.7 Glass and front window ................................................................................................. 5
3.3.8 Electrical safety .............................................................................................................. 5
3.3.9 Gas supply safety ........................................................................................................... 6
3.3.10 Ergonomics ..................................................................................................................... 6
3.3.11 Temparature .................................................................................................................. 6
4 Tests done .................................................................................................................................... 6
5 Person .......................................................................................................................................... 6
6 Instrument / Series number / calibrated ..................................................................................... 6
7 Conclusion .................................................................................................................................... 7
8 Annex A ........................................................................................................................................ 8
9 Annex B ........................................................................................................................................ 9
9.1 Normal conditions ................................................................................................................. 9
9.2 If MSC is turned off after use ................................................................................................ 9
10 Annex C ...................................................................................................................................... 10
11 Annex D ...................................................................................................................................... 12
L234 test report compliance to EN 12469
Document
Number: 20 460 234 Document
Version:
1.0 Document
Responsible
TGA
Page 3 of 12
K:\UDVIKLING INF\Projekter\Projekt 55 L200 opdatering\Analyse\L234 test report compliance to EN 12469.docx
1 Scope
The scope of this document is to describe the compliance of the K-Systems L234 workstation to
the standard EN 12469:2000.
In general three types of performance testing can be distinguished: Type testing, Installation
testing and routine maintenance testing. (See table 5 in EN 12469)
Routine maintenance testing is a sub part of the installation testing as described in the standard.
We have therefore chosen to verify that L234 comply type testing and relevant installation testing,
as stated in table 5 in the standard.
2 Definitions and Acronyms
15 060 04 Definitions
Testing is passed (P).
Not applicable (n/a).
Testing is failed (F)
3 How to obtain compliance to the Standard
The following paragraphs refers back to the test specifications listed in table 5 of the standard.
See table 5 in annex D.
3.1 Type testing
3.1.1 Retention at front aperture
KI method chosen and done testing. (See annex A) P
3.1.2 Leaktightness of carcass
Soap solution test method chosen and done testing. (See annex B) P
L234 test report compliance to EN 12469
Document
Number: 20 460 234 Document
Version:
1.0 Document
Responsible
TGA
Page 4 of 12
K:\UDVIKLING INF\Projekter\Projekt 55 L200 opdatering\Analyse\L234 test report compliance to EN 12469.docx
3.1.3 Filters
Aerosol challenge method chosen and done testing. (See annex C) P
3.1.4 Product protection
Microbiological method or alternative KI method. n/a
3.1.5 Cross contamination
Microbiological method or alternative KI method. n/a
3.2 Installation testing
3.2.1 Volumetric airflow rate measurements
Measurements must be made in the horizontal plane 50 mm to 100 mm above the top
edge of the front aperture. Requirements in the standard states that the mean downflow
velocity to achieve product protection must be within 0,25 m/s to 0,50 m/s. No individual
measurements should differ from the mean by more than 20%.
We measure mean 0,35 m/s. with a difference at min 9% and max 13% (See annex C). P
3.2.2 Check airflow patterns
Same as 3.2.1. (See annex C) P
3.3 Guidance on manufacture
3.3.1 Front aperture height
The height of the front aperture should be between 160 mm and 250 mm.
We have 163,5 mm. P
3.3.2 Lightning
Lightning at the work surface should be at least 750 lux.
L234 test report compliance to EN 12469
Document
Number: 20 460 234 Document
Version:
1.0 Document
Responsible
TGA
Page 5 of 12
K:\UDVIKLING INF\Projekter\Projekt 55 L200 opdatering\Analyse\L234 test report compliance to EN 12469.docx
We measure around 900 lux at the working positions. P
3.3.3 Sound
The sound level should not exceed 65 dB A-weighted.
We have measured 57 dB A-weighted. P
3.3.4 Vibration
We have less than 0,005 mm RMS. At work surface between 20 Hz and 20000 Hz. P
3.3.5 Stability
Equipment is tilted in each direction to an angle of 10° from its normal position. Equipment has a force applied at its top, or at a height of 2 m if the equipment has a height more than 2 m. The force is 250 N, or 20 % of the weight of the equipment, whichever is less, and is applied in all directions except upwards. Conforming EN 61010-1 for stability. P
3.3.6 Materials, general
The material where micrio-organisms have contact whould be corrosion resistant,
non-flammable and non-adsorbing.
The tabletop where micro-organisms should be at is made of stainless steel, which
is corrosion resistant, non-flammable and non-adsorbing. P
3.3.7 Glass and front window
Glass side windows complying to EN 61010-1. P
Transparent front window, can be taken off while system is not running, but must
be mounted while in use. When working with IVF only a small opening is needed.
3.3.8 Electrical safety
The MSC has been designed to, and are being tested at UL in Denmark, to conform
EN 61010-1. In test
L234 test report compliance to EN 12469
Document
Number: 20 460 234 Document
Version:
1.0 Document
Responsible
TGA
Page 6 of 12
K:\UDVIKLING INF\Projekter\Projekt 55 L200 opdatering\Analyse\L234 test report compliance to EN 12469.docx
3.3.9 Gas supply safety
No flammable gas installed n/a
3.3.10 Ergonomics
The design should ensure that maintenance can safety be carried out on the
cabinet after installation. P
- Easy to remove back panel for filter replacement
- Easy to open tray for cleaning
- Powder coated cabinet for easy cleaning
- Stainless steel tabletop accept all normally use detergents and alcohol for cleaning
3.3.11 Temparature
With no heat in tabletop temperature will not rise by more than 8 ˚C. P
4 Tests done
- Downflow measurements
- Inflow measurements
- DOP (HEPA filter leakage)
- KI-Discus
5 Person
Mechanical Engineer Thomas Grau-Andersen
6 Instrument / Series number / calibrated
- Testo 425 / 02052833 / 01-08-2012
- TDA-2G / 15747 / 03-09-2012
- Ki-Discus MK2 / K03/05-0188 / 22-08-2012
L234 test report compliance to EN 12469
Document
Number: 20 460 234 Document
Version:
1.0 Document
Responsible
TGA
Page 7 of 12
K:\UDVIKLING INF\Projekter\Projekt 55 L200 opdatering\Analyse\L234 test report compliance to EN 12469.docx
7 Conclusion
Relevant tests from EN 12469:2000 is listed and show how L234 workstation comply to the class II
standard.
L234 workstation have successful ergonomics and the design ensure that maintenance can safety
be carried out on the cabinet after installation.
Performance results in brief:
All airflow measurements such as downflow and inflow are complying to EN 12469:2000.
DOP test (HEPA filter test) are complying to EN 12469:2000.
The sound level are complying to EN 12469:2000.
The vibration level are complying to EN 12469:2000.
General design are complying to EN 12469:2000.
L234 test report compliance to EN 12469
Document
Number: 20 460 234 Document
Version:
1.0 Document
Responsible
TGA
Page 8 of 12
K:\UDVIKLING INF\Projekter\Projekt 55 L200 opdatering\Analyse\L234 test report compliance to EN 12469.docx
8 Annex A
The KI Discus test was performed:
By: Thomas Grau-Andersen Date: 27-08-2012 On L234 S/N: Proto 2
Instrumentation / calibration: Ki-Discus MK2 / 22-08-2012
The photo above shows the 4 filters suction in front of the cabinet, at the KI-Discus setup.
Conclusion: We have less than 62 brown dots at each filter.
L234 test report compliance to EN 12469
Document
Number: 20 460 234 Document
Version:
1.0 Document
Responsible
TGA
Page 9 of 12
K:\UDVIKLING INF\Projekter\Projekt 55 L200 opdatering\Analyse\L234 test report compliance to EN 12469.docx
9 Annex B
9.1 Normal conditions
Under normal running conditions non-filtered air under
positive pressure is encapsulated by surroundings
negative pressure. Therefore positive pressure leaking to
the surroundings under running conditions is only
theoretical possible through sample port for DOP testing.
This sample port is double secured with both a valve and
a silicone cap.
9.2 If MSC is turned off after use
Soap solution test with an internal pressure at approximately 250 Pa and observe the soap
applied.
L234 test report compliance to EN 12469
Document
Number: 20 460 234 Document
Version:
1.0 Document
Responsible
TGA
Page 10 of 12
K:\UDVIKLING INF\Projekter\Projekt 55 L200 opdatering\Analyse\L234 test report compliance to EN 12469.docx
The Soap solution test was performed:
By: Thomas Grau-Andersen Date: 20-08-2012 On L234 S/N: Proto 2
Instrumentation: Soap solution
The previous photos shows how soap solution is added to the cabinet.
Conclusion: If the MSC is turned off nothing indicating leaks.
10 Annex C
Downflow is measured by use of an anemometer. With the anemometer inside the MSC, make air
velocity measurements in the horizontal plane 50 mm to 100 mm above the top edge of the front
aperture. Make measurements over a period of at least 1 min in each position for at least eight
positions. The exact height and placement can be seen in the tables(s) below.
Inflow is measured by use of an anemometer. With the anemometer outside the MSC, make air
velocity measurements pointing towards front opening. The exact height and placement can be
seen in the tables(s) below.
DOP test (HEPA filter test) is measured by use of an aerosol photometer. With this device test
filters and edges along workspace filter and exhaust filter. If an aerosol photometer is used, the
local penetration of a HEPA filter should be not more than 0,01 %. Test results can be seen in the
tables(s) below.
L234 test report compliance to EN 12469
Document
Number: 20 460 234 Document
Version:
1.0 Document
Responsible
TGA
Page 11 of 12
K:\UDVIKLING INF\Projekter\Projekt 55 L200 opdatering\Analyse\L234 test report compliance to EN 12469.docx
L234 test report compliance to EN 12469
Document
Number: 20 460 234 Document
Version:
1.0 Document
Responsible
TGA
Page 12 of 12
K:\UDVIKLING INF\Projekter\Projekt 55 L200 opdatering\Analyse\L234 test report compliance to EN 12469.docx
11 Annex D