Display Quality Assurance: Considerations When Establishing

a Display QA Program

Mike Silosky, M.S.

8/3/2017

Page 2: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Objectives and Outline

Why, Who, What, When, Where?

• Discuss the resources that may be needed for a display QA program and how these resources may be utilized

How do flat panel displays fail?

• Review some common modes of failure for LCDs and discuss testing strategies

What about existing guidelines and requirements?

• Compare and contrast the TG-270 recommendations with requirements provided by accreditation organizations (ACR and TJC)

2

Page 3: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Why perform display QA?

• Display performance has a direct impact on the information that physicians have available when making a diagnosis and may effect decision making

• Ensure consistent display of medical images from acquisition to interpretation (and beyond)

• Accrediting bodies have implemented requirements regarding quality assurance testing of displays used in medical imaging (TJC, ACR)

3

Page 4: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Who performs display QA?

4

Physicists/Assistants

Imaging Technologists

PACS Personnel

Biomedical Engineering

Page 5: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

What to test and when

5

Page 6: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

What and When

6

Page 7: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

What and When

7

Page 8: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

What and When

8

Page 9: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

What and When

9

Page 10: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Where?

10

• Reading Rooms

• Clinical Areas

• Personal Offices

• Storage Rooms/ isolated areas

• Houses???

Page 11: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Tools: Photometers/Colorimeters

11https://www.aapm.org/pubs/reports/OR_03.pdf

Page 12: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Tools: Test Patterns

12

Page 13: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Tools: Vendor/3rd Party Software

• What Software are you going to use?

• What measurements can be made?

• How reliable is this software?

• Who is going to review the results?

• How is QA going to be performed on vendor provided tools?

13

Page 14: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Common Modes of Failure

14

Page 15: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Example: Imaginary Hospital

• 200 Diagnostic displays

• 50-60 Modality Workstations

• 10 Clinical Specialist Workstations

• Innumerable Electronic Health Record Workstations

15

Goals:

1. Maximize the diagnostic value of acquired images

2. Ensure consistent presentation of medical images from acquisition to interpretation.

Page 16: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

GSDF Compliance

16

• Primary displays are typically compliant

• Some modality displays are compliant

• Most other devices require calibration to become GSDF compliant

Page 17: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Example 1

17

0.100

1.000

0 100 200 300 400 500 600 700

dL

/L

JND

DICOM

IR16

Page 18: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Example 2

18

0.100

1.000

0 100 200 300 400 500 600 700

dL

/L

JND

DICOM

IR14

Page 19: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Example 3

19

Page 20: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

20

Are you relying on a built-in photometer?

Page 21: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

As displays age, built-in photometers drift

21

-10

0

10

20

30

40

50

60

70

80

0 10000 20000 30000 40000 50000

Dif

fere

nce

in M

eas

ure

d

Wh

ite

Lev

el L

um

inan

ce (

%)

BLH

Page 22: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Contrast Perception

22

Page 23: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Ambient Luminance and Luminance Ratio

Maintaining the same Luminance Ratio and Luminance response function is essential for the consistent presentation of medical images

23

𝑅𝐿 =𝐿′𝑚𝑎𝑥

𝐿′𝑚𝑖𝑛

=𝐿𝑚𝑎𝑥 + 𝐿𝑎𝑚𝑏

𝐿𝑚𝑖𝑛 + 𝐿𝑎𝑚𝑏

RL = 350

Page 24: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

24

Perceived contrast is highly dependent on the average luminance of the scene

Page 25: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Uniformity

• Quantitative-Acceptance Only

25

• Qualitative-Routinely

Page 26: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Spatial Resolution

26Graphics driver set to the native resolution of the display

Page 27: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Spatial Resolution

27Graphics driver set to non-native resolution of the display

Page 28: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

SAMS Question #1

In order to obtain similar image contrast between different displays, what two performance characteristics should be the same?

a: Minimum Luminance, Maximum Luminance

b: Maximum Luminance, Maximum Luminance Deviation

c: Minimum Luminance, Display Function

d: Luminance Ratio, Display Function

e: Display Function, Ambient Luminance

28

Page 29: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

SAMS Question #1

Answer: (d). Luminance Ratio, Display Function

Reference:

American Association of Physicists in Medicine, “Assessment of display performance for medical imaging systems,” AAPM On-line Report No. 3 (AAPM, College Park, MD, 2005), p. 78

29

Page 30: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

SAMS Question #2

Luminance measurements reported by internal luminance meters used for GSDF calibration of primary displays have been shown to deviate as displays exceed their backlight warranty by up to what percent?

a: 10%

b: 20%

c: 40%

d: 60%

e: > 70%

30

Page 31: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

SAMS Question #2

Answer: (e). > 70%

Reference:

M. Silosky and R.M. Marsh, “Constancy of built-in luminance meter measurements in diagnostic displays,” Medical Physics 40 (12), 121902 (2013).

31

Page 32: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

SAMS Question #3

According to the ACR-AAPM-SIIM Technical Standard for Electronic Practice of Medical Imaging, it is recommended that the luminance ratio for displays used for primary interpretation should exceed what value?

a. 150

b. 250

c. 350

d. 450

e. 550

32

Page 33: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

SAMS Question #3

Answer: (c). 350

Reference:

ACR-AAPM-SIIM Technical Standard for Electronic Practice of Medical Imaging, 2014, p. 9

33

Page 34: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

What about existing guidance/requirements?

34

Page 35: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

35

Page 36: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Luminance Response Methodology

TG-270 and ACR-AAPM-SIIM Technical Standard for the Practice of Medical Imaging

1. Determine Lamb

2. Measure Lmin,-Ensure Lmin > 4 x Lamb

3. Set L’min determines L’max- RL = 350

4. Determine intermediate gray levels

36

The minimum value of L’min is 1.0 cd/m2

Page 37: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

37

ACR 2016 Digital Mammography QC Manual

Lmin (Modality and Diagnostic)

• Within 30% of manufacturers recommendations or ≤ 1.5 cd/m2

• Not based on Lamb

Page 38: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

38

ACR 2016 Digital Mammography QC Manual

Lmax (Modality and Diagnostic)• Within 10% of manufacturers recommendations or• 150 cd/m2 (Modality)• 420 cd/m2 (Diagnostic)

To maintain a luminance ratio of 350, L’min should be 0.43 cd/m2 (Modality) and 1.2 cd/m2 (Diagnostic)

Its difficult to meet the Lmin > 4 x Lamb requirement

Page 39: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

ACR – CT (Modality Only)

• Visual Evaluation

- Resolution (via line pairs)

- Contrast (Both patches visible)

- Geometric Distortion

- Black/White Transitions

- Loss of Bit Depth (Need a gray ramp)

39

Page 40: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

ACR – CT Scanner Monitors

• Quantitative Evaluation

- Lmin ( < 1.2 cd/m2)

- Lmax ( > 90 cd/m2)

- Establish Luminance Response Curve

- Luminance Uniformity (MLD < 15%)

(Test Pattern not specified)

40

Page 41: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

ACR – Magnetic Resonance Imaging

41

• Visual Evaluation

- Resolution (via line pairs)

- Contrast (Both patches visible)

- Geometric Distortion

- Black/White Transitions

- Loss of Bit Depth

Page 42: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

ACR – Magnetic Resonance Imaging

42

• Quantitative Evaluation

- Lmin ( < 1.2 cd/m2)

- Lmax ( > 90 cd/m2)

- Establish Luminance Response Curve

- Luminance Uniformity (MLD < 30%)

(at Maximum Luminance)

Page 43: Display Quality Assurance: Considerations When ...amos3.aapm.org/abstracts/pdf/127-38284-418554-126965...Display Quality Assurance: Considerations When Establishing a Display QA Program

Summary

• Implementation of display QA requirements and recommendations will vary from facility to facility

• Qualified Medical Physicists or other personnel with appropriate training should oversee display QA programs

• Individual sites must determine how to best navigate contradictory guidance

43