AAPM 2009 Breast Imaging Course - Digital Mammography Update July 27, 2009 Kanal 1 Digital Mammography Update: Design Digital Mammography Update: Design and Characteristics of Current Systems and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Kalpana M. Kanal, Ph.D., DABR Assistant Professor Assistant Professor Department of Radiology Department of Radiology University of Washington University of Washington Seattle, Washington Seattle, Washington AAPM Annual Meeting 2009 AAPM Annual Meeting 2009 – Anaheim, CA Anaheim, CA Kanal Kanal 2 Educational Objectives Educational Objectives Understand the physics of digital detector technology Understand the physics of digital detector technology Recognize that vendors use varying detector technology Recognize that vendors use varying detector technology in FFDM systems in FFDM systems Appreciate the advantages and disadvantages of digital Appreciate the advantages and disadvantages of digital mammography systems mammography systems Understand briefly the image quality metrics such as Understand briefly the image quality metrics such as DQE and MTF DQE and MTF

Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

  • Upload
    others

  • View
    5

  • Download
    1

Embed Size (px)

Citation preview

Page 1: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 1

Digital Mammography Update: Design Digital Mammography Update: Design and Characteristics of Current Systems and Characteristics of Current Systems

Kalpana M. Kanal, Ph.D., DABRKalpana M. Kanal, Ph.D., DABRAssistant ProfessorAssistant Professor

Department of RadiologyDepartment of RadiologyUniversity of WashingtonUniversity of Washington

Seattle, WashingtonSeattle, Washington

AAPM Annual Meeting 2009 AAPM Annual Meeting 2009 –– Anaheim, CAAnaheim, CA

KanalKanal 22

Educational ObjectivesEducational Objectives

Understand the physics of digital detector technologyUnderstand the physics of digital detector technologyRecognize that vendors use varying detector technology Recognize that vendors use varying detector technology in FFDM systemsin FFDM systemsAppreciate the advantages and disadvantages of digital Appreciate the advantages and disadvantages of digital mammography systems mammography systems Understand briefly the image quality metrics such as Understand briefly the image quality metrics such as DQE and MTFDQE and MTF

Page 2: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 2

KanalKanal 33

ResourcesResourcesDoug PfeifferDoug Pfeiffer, MS, DABR, MS, DABRDigital Mammography: An Digital Mammography: An overview overview –– Dr. MaheshDr. Mahesh(Radiographics (Radiographics 2004;24:17472004;24:1747--1760)1760)Fundamentals of Digital Fundamentals of Digital

Mammography Primer Mammography Primer –– Dr. Dr. SmithSmith (Hologic Inc)(Hologic Inc)Digital Mammography Digital Mammography ––Pisano and YaffePisano and Yaffe (Radiology (Radiology 2005; 234:3532005; 234:353--262)262)Bloomquist and Yaffe Bloomquist and Yaffe –– Med Med Phys 33 (3), 2006Phys 33 (3), 2006

IntroductionIntroduction

Page 3: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 3

KanalKanal 55

FullFull--Field Digital Mammography (FFDM)Field Digital Mammography (FFDM)

c.f. www.gehealthcare.com/rad/xr/education/dig_xray_intro.htmlc.f. www.gehealthcare.com/rad/xr/education/dig_xray_intro.html

- Screen-Film

KanalKanal 66

FullFull--Field Digital Mammography (FFDM)Field Digital Mammography (FFDM)

Wide dynamic range Wide dynamic range (1000:1) compared with (1000:1) compared with SFM (40:1)SFM (40:1)Dynamic image Dynamic image manipulationmanipulationAbility to postAbility to post--processprocessSoftSoft--copy read copy read accompanied by accompanied by computercomputer--aidedaided--diagnosis (CAD)diagnosis (CAD)3D imaging3D imaging

Radiographics 2004:24,1749

Page 4: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 4

KanalKanal 77

FullFull--Field Digital Mammography (FFDM)Field Digital Mammography (FFDM)

AdvantageAdvantageEach component of the mammographic process can Each component of the mammographic process can be optimized with digital mammographybe optimized with digital mammography

Radiographics 2004:24,1750

KanalKanal 88

SFM vs. FFDMSFM vs. FFDM

SFM: Half mAs, Automatic exposure control, Double mAs

FFDM: Same technique factors as SFM, W/L adjusted

Radiographics 2004:24,1750

Page 5: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 5

KanalKanal 99

SFM vs. FFDMSFM vs. FFDM

Radiographics 2004:24,1751

KanalKanal 1010

FDA and Digital MammographyFDA and Digital Mammography

Butler Butler -- http://www.acr.org/accreditation/mammography/rsna07presentationhttp://www.acr.org/accreditation/mammography/rsna07presentation.aspx.aspx

Page 6: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 6

KanalKanal 1111

MQSA ScorecardMQSA Scorecard

Certification statistics, as of July 1, 2009 Certification statistics, as of July 1, 2009 Total certified facilities / Total accredited unitsTotal certified facilities / Total accredited units

8,714 / 12,8758,714 / 12,875Certified facilities with FFDM units / Accredited FFDM Certified facilities with FFDM units / Accredited FFDM unitsunits

4,659 / 6,9644,659 / 6,96453%53% certified facilities with FFDM unitscertified facilities with FFDM units54%54% accredited FFDM units (38% last year)accredited FFDM units (38% last year)

http://www.fda.gov/cdrh/mammography/scorecardhttp://www.fda.gov/cdrh/mammography/scorecard--statistics.htmlstatistics.html

Technologies for FFDMTechnologies for FFDM

Page 7: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 7

KanalKanal 1313

Technologies for FFDMTechnologies for FFDMIndirect CaptureIndirect Capture

a scintillator such as a scintillator such as cesium iodide (CsI) cesium iodide (CsI) absorbs xabsorbs x--rays and rays and generates a light generates a light scintillationscintillationdetected by an array of detected by an array of photodiodes or chargephotodiodes or charge--coupled devices (CCDs)coupled devices (CCDs)Resolution degradation Resolution degradation

http://www.hologic.com/wh/pdf/R-LM-016_Radiology_Management.pdf

KanalKanal 1414

Technologies for FFDMTechnologies for FFDMDirect CaptureDirect Capture

XX--ray photons are directly ray photons are directly captured by a captured by a photoconductor such as photoconductor such as amorphous selenium (aamorphous selenium (a--Se), which converts the Se), which converts the absorbed xabsorbed x--rays directly to rays directly to a digital signala digital signalSpatial resolution limited Spatial resolution limited to pixel sizeto pixel size

http://www.hologic.com/wh/pdf/R-LM-016_Radiology_Management.pdf

Page 8: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 8

KanalKanal 1515

Various Approaches in development of Various Approaches in development of FFDM systemsFFDM systems

IndirectIndirectA single flatA single flat--panel scintillator and an amorphous silicon panel scintillator and an amorphous silicon (a(a--Si) diode array Si) diode array –– GE GE Slot scanning with scintillators and CCD arrays Slot scanning with scintillators and CCD arrays –– Fischer Fischer Imaging, now HologicImaging, now HologicPhotostimulable phosphor plates Photostimulable phosphor plates -- FujiFuji

DirectDirectA flatA flat--panel amorphous selenium (apanel amorphous selenium (a--Se) array Se) array –– Hologic, Hologic, SiemensSiemensTiled scintillators with fiberoptic tapers and mosaic CCD Tiled scintillators with fiberoptic tapers and mosaic CCD arrays arrays –– Hologic StereotacticHologic Stereotactic

KanalKanal 1616

Technologies for FFDM Technologies for FFDM –– Indirect CaptureIndirect CaptureGEGE

GE GE SenographeSenographe 2000D Full Field 2000D Full Field Digital Mammography System on Digital Mammography System on 01/28/0001/28/00

GE GE SenographeSenographe DS Full FieldDS Full FieldDigital Mammography (FFDM) Digital Mammography (FFDM) System System -- 02/19/0402/19/04

GE Senographe Essential Full GE Senographe Essential Full Field Digital Mammography Field Digital Mammography (FFDM) System (FFDM) System -- 04/11/0604/11/06 (larger detector)

Fig. Ref: GE DS QC manualFig. Ref: GE DS QC manual

Page 9: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 9

KanalKanal 1717

Technologies for FFDM Technologies for FFDM –– Indirect CaptureIndirect CaptureGEGE

In this system, the digital In this system, the digital detector array is constructed detector array is constructed from an afrom an a--Si thinSi thin--film transistor film transistor (TFT) matrix deposited on a (TFT) matrix deposited on a glass substrateglass substrateThe CsI scintillator is deposited The CsI scintillator is deposited on the aon the a--Si detectorSi detectorEach lightEach light--sensitive diode sensitive diode element is connected by TFTs element is connected by TFTs to a control and a data line so to a control and a data line so that charge produced in the that charge produced in the diode is read out in response to diode is read out in response to light emission from the scintillatorlight emission from the scintillator

Radiographics 2004:24,1753

KanalKanal 1818

Technologies for FFDM Technologies for FFDM –– Indirect CaptureIndirect CaptureGEGE

Radiographics 2004:24,1753

141414141414Bit DepthBit Depth

5 lp/cm5 lp/cm5 lp/cm5 lp/cm5 lp/cm5 lp/cmLimiting Limiting Spatial Spatial

ResolutionResolution

2394 x 3062 2394 x 3062 pixels (14 MB)pixels (14 MB)

1914 x 2294 1914 x 2294 pixels (9 MB)pixels (9 MB)

1914 x 2294 1914 x 2294 pixels (9 MB)pixels (9 MB)

Image sizeImage size

100 100 µµmm100 100 µµmm100 100 µµmmPixel sizePixel size

24.0 x 30.724.0 x 30.719.2 x 23.019.2 x 23.019.2 x 23.019.2 x 23.0Detector Detector sizesize

EssentialEssentialDSDS2000D2000D

Page 10: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 10

KanalKanal 1919

Technologies for FFDM Technologies for FFDM –– Indirect CaptureIndirect CaptureGE GE –– Advantages and DisadvantagesAdvantages and Disadvantages

Close bonding between CsI and aClose bonding between CsI and a--Si ensures minimal Si ensures minimal light losslight lossStrong signal from the Si diode array yields higher Strong signal from the Si diode array yields higher detective quantum efficiencydetective quantum efficiencyDetector is linear over a wide range (10Detector is linear over a wide range (1055) of exposures) of exposures

Limiting factor is the large pixel size (100 Limiting factor is the large pixel size (100 μμm)m)Smaller pixel sizes improve spatial resolution but at the Smaller pixel sizes improve spatial resolution but at the cost of increased image noise and decreased SNR for cost of increased image noise and decreased SNR for the same breast dosethe same breast dosePossibility of ghosting in imagesPossibility of ghosting in images

Radiographics 2004:24,1753

KanalKanal 2020

PSP PSP

Base supportBase support

plate exposure:plate exposure:create latent imagecreate latent image

xx--ray exposureray exposure

Technologies for FFDM Technologies for FFDM –– Indirect CaptureIndirect CaptureComputed Radiography Computed Radiography –– Fuji (7/10/2006)Fuji (7/10/2006)

(Barium fluorobromide)(Barium fluorobromide)

Technology introduced in 1981Technology introduced in 1981

Pfeiffer

Page 11: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 11

KanalKanal 2121

plate erasure:plate erasure:remove residual signalremove residual signal

light erasurelight erasure

plate readout:plate readout:extract latent imageextract latent image

laser beam scanlaser beam scan

Technologies for FFDM Technologies for FFDM –– Indirect CaptureIndirect CaptureComputed Radiography Computed Radiography –– FujiFuji

Technology introduced in 1981Technology introduced in 1981

Pfeiffer

KanalKanal 2222

Technologies for FFDM Technologies for FFDM –– Indirect CaptureIndirect CaptureComputed Radiography Computed Radiography –– FujiFuji

Fuji FCRm, Dual-side reader

10 lp/mm10 lp/mmSpatial Spatial ResolutionResolution

14 bits14 bitsDynamic Dynamic RangeRange

3328 x 4096 3328 x 4096 pixels (24 MB)pixels (24 MB)Image sizeImage size

50 50 µµmmPixel sizePixel size

18 x 2418 x 2424 x 3024 x 30

Detector Detector sizesize

http://www.fujimed.com/

Page 12: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 12

KanalKanal 2323

Fuji Fuji -- Dual Side ReadoutDual Side Readout

laser beamlaser beam

mirrormirror

Protective Protective layerlayer

photodetectorphotodetector

photodetectorphotodetector

opticalopticalguideguide

TransparentTransparentsupport support

PhosphorPhosphorlayerlayer

emissionemission

opticalopticalguideguide

Image courtesy: Doug Pfeiffer

KanalKanal 2424

Technologies for FFDM Technologies for FFDM –– Indirect CaptureIndirect CaptureFuji Fuji –– Advantages and DisadvantagesAdvantages and Disadvantages

Advantage of the CR mammo system is that filmAdvantage of the CR mammo system is that film--screen screen cassettes can be replaced by CR cassettes without replacing cassettes can be replaced by CR cassettes without replacing the entire systemthe entire systemBoth small and large cassettes can be accommodated by the Both small and large cassettes can be accommodated by the readerreaderDual side reader, 50 Dual side reader, 50 μμm pixel sizem pixel size

Disadvantage Disadvantage -- Effective pixel size influenced by phosphor Effective pixel size influenced by phosphor thickness, light diffusion within phosphor, laser light scatter thickness, light diffusion within phosphor, laser light scatter & & diameter of laser beamdiameter of laser beamTechnologist time on processing of imagesTechnologist time on processing of imagesNoise associated with the low collection efficiency of emitted Noise associated with the low collection efficiency of emitted lightlight

Page 13: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 13

KanalKanal 2525

Technologies for FFDM Technologies for FFDM –– Indirect CaptureIndirect Capture

A narrow slotA narrow slot--detector and detector and a narrow fan beam of xa narrow fan beam of x--rays are scanned rays are scanned synchronously across the synchronously across the full field of view to cover full field of view to cover the entire breastthe entire breastSystem consists of System consists of phosphor (thalliumphosphor (thallium--activated CsI) with a activated CsI) with a fiberoptic coupling to a fiberoptic coupling to a CCDCCD

SenoScanSenoScan; Fischer Imaging ; Fischer Imaging –– 9/25/019/25/01

Radiographics 2004:24,1752

KanalKanal 2626

Technologies for FFDM Technologies for FFDM –– Indirect CaptureIndirect Capture

Seibert, AAPM Meeting Handout

SenoScanSenoScan; Fischer Imaging (now ; Fischer Imaging (now HologicHologic))

Page 14: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 14

KanalKanal 2727

Technologies for FFDM Technologies for FFDM –– Indirect CaptureIndirect Capture

Compact detector that is less expensive compared to Compact detector that is less expensive compared to othersothersExcellent scatter rejection due to small volume of breast Excellent scatter rejection due to small volume of breast exposed at any timeexposed at any timeNo grid needed therefore less doseNo grid needed therefore less doseLonger compression since scan times are longer Longer compression since scan times are longer (approx. 6 sec)(approx. 6 sec)Powerful tubes, elaborate signal readout and image Powerful tubes, elaborate signal readout and image reconstruction requiredreconstruction required

(SenoScan; Fischer Imaging (now Hologic))(SenoScan; Fischer Imaging (now Hologic))Advantages and DisadvantagesAdvantages and Disadvantages

KanalKanal 2828

Comparison Comparison –– Indirect CaptureIndirect Capture

3328 x 4096 3328 x 4096 pixels (24 MB)pixels (24 MB)

Fuji Fuji FCRFCRmm

PSP platePSP plate

10 lp/mm10 lp/mm

50 50 µµmm

18 x 2418 x 2424 x 3024 x 30

Page 15: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 15

KanalKanal 2929

Technologies for FFDM Technologies for FFDM –– Direct CaptureDirect Capture

aa--Se, a good photoconductor is deposited directly onto Se, a good photoconductor is deposited directly onto the athe a--Si TFT substrate enabling direct captureSi TFT substrate enabling direct captureThe aThe a--Se detector directly converts xSe detector directly converts x--rays to electronrays to electron--hole pairs hole pairs The aThe a--Si TFT converts the electronSi TFT converts the electron--hole pairs to hole pairs to electronic signalelectronic signal

aa--Se Array (Selenia, Se Array (Selenia, HologicHologic) ) –– 10/2/0110/2/01

http://www.hologic.com/wh/digisel.htm

KanalKanal 3030

Technologies for FFDM Technologies for FFDM –– Direct CaptureDirect CaptureFlatFlat--Panel aPanel a--Se Array (Selenia, Hologic)Se Array (Selenia, Hologic)

http://www.hologic.com/wh/digisel.htm

> 7 lp/cm> 7 lp/cmSpatial Spatial ResolutionResolution

14 bits14 bitsDynamic Dynamic RangeRange

3328 x 4096 3328 x 4096 pixels (24 MB)pixels (24 MB)Image sizeImage size

70 70 µµmmPixel sizePixel size24.0 x 29.024.0 x 29.0Detector sizeDetector size

Page 16: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 16

KanalKanal 3131

Technologies for FFDM Technologies for FFDM –– Direct CaptureDirect Capture

Advantage is that the detector response function Advantage is that the detector response function maintains its sharpness even with increasing thicknessmaintains its sharpness even with increasing thicknessHigh MTF and DQE can be achievedHigh MTF and DQE can be achieved

Potential weaknesses are the need for high biasing Potential weaknesses are the need for high biasing voltage, drifting of the dark signal and cost of detectorvoltage, drifting of the dark signal and cost of detectorInherent sharpness of detector may also increase the Inherent sharpness of detector may also increase the severity of aliasing artifacts associated with severity of aliasing artifacts associated with undersampling on any digital detectorundersampling on any digital detector

FlatFlat--Panel aPanel a--Se Array (Selenia, Hologic)Se Array (Selenia, Hologic)Advantages and DisadvantagesAdvantages and Disadvantages

KanalKanal 3232

Technologies for FFDM Technologies for FFDM –– Direct CaptureDirect Captureaa--Se Array (Mammomat Novation, Siemens Se Array (Mammomat Novation, Siemens –– 8/20/04)8/20/04)

http://www.medical.siemens.com

> 7 lp/mm> 7 lp/mmSpatial Spatial ResolutionResolution

70 70 µµmm

24.0 x 29.024.0 x 29.0

14 bits14 bitsDynamic Dynamic RangeRange

3328 x 4096 3328 x 4096 pixels (24 MB)pixels (24 MB)Image sizeImage size

Pixel sizePixel size

Detector sizeDetector size

Page 17: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 17

KanalKanal 3333

Comparison Comparison –– Direct CaptureDirect Capture

KanalKanal 3434

Technologies for FFDMTechnologies for FFDM

http://www.hologic.com/oem/pdf/DROverviewR-007_Nov2000.pdf

Fuji/Kodak/Fuji/Kodak/

Agfa/PhilipsAgfa/Philips

Fischer Fischer (Hologic)(Hologic)

GEGE HologicHologic

SiemensSiemens

Page 18: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 18

KanalKanal 3535

MTF, DQE, DOSEMTF, DQE, DOSE

KanalKanal 3636

MTFMTFMTF is a measure of signal transfer over a range of MTF is a measure of signal transfer over a range of frequencies and quantifies spatial resolutionfrequencies and quantifies spatial resolution

Yaffe - Radiology 2005:234,353

Type 1 Type 1 –– CsI (TFT)CsI (TFT)

Type 2 Type 2 –– CsI (CCD)CsI (CCD)

Type 3 Type 3 –– CRCR

Type 4 Type 4 –– aa--SeSe

http://www.hologic.com/wh/pdf/R-LM-016_Radiology_Management.pdf

Page 19: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 19

KanalKanal 3737

MTFMTFBloomquist et al Bloomquist et al -- DMIST trialDMIST trial

Bloomquist – Medical Physics 2006:33 (3), 719

KanalKanal 3838

DQEDQEDetective Quantum Efficiency (DQE) measures SNR Detective Quantum Efficiency (DQE) measures SNR transfer through the system as a function of spatial transfer through the system as a function of spatial frequency and is a good measure of dose efficiencyfrequency and is a good measure of dose efficiency

http://www.hologic.com/wh/pdf/R-LM-016_Radiology_Management.pdf

Page 20: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 20

KanalKanal 3939

FFDM FFDM –– Radiation DoseRadiation Dose

Bloomquist et al Bloomquist et al -- DMIST trialDMIST trial

Bloomquist – Medical Physics 2006:33 (3), 719

KanalKanal 4040

Storage and EconomicsStorage and Economics

Page 21: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 21

KanalKanal 4141

Storage of Digital ImagesStorage of Digital Images

Radiographics 2004:24,1755

KanalKanal 4242

Display of Digital ImagesDisplay of Digital Images

Radiographics 2004:24,1757

Page 22: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 22

KanalKanal 4343

Advantages and DisadvantagesAdvantages and Disadvantages

AdvantagesAdvantagesOptimize postOptimize post--processing of imagesprocessing of imagesPermit computerPermit computer--aided detection to improve the aided detection to improve the detection of lesionsdetection of lesionsStorage of images easierStorage of images easier

DisadvantagesDisadvantagesImage display and system costImage display and system costLimiting spatial resolution is inferior to film, 5Limiting spatial resolution is inferior to film, 5--13 13 lp/mm vs. 20 lp/mmlp/mm vs. 20 lp/mm

KanalKanal 4444

Economics of FFDMEconomics of FFDM

SFM systems cost well under $100,000SFM systems cost well under $100,000FFDM systems cost in the range of $300,000 FFDM systems cost in the range of $300,000 -- $450,000$450,000

Reimbursement rates approved by Medicare are Reimbursement rates approved by Medicare are decreasing every yeardecreasing every year

Plain FilmPlain Film ScreeningScreening 2008 2008 -- $83.03, 2009 $83.03, 2009 -- $81.51$81.51DigitalDigital ScreeningScreening 2008 2008 -- $133.69, 2009 $133.69, 2009 -- $129.84 $129.84

DiagnosticDiagnostic 2008 2008 -- $151.21, 2009 $151.21, 2009 -- $152.56$152.56CADCAD 2008 2008 -- $14.86, 2009 $14.86, 2009 -- $12.26$12.26

Page 23: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 23

KanalKanal 4545

BenefitsBenefits

KanalKanal 4646

Expected Benefits of FFDMExpected Benefits of FFDM

The costs of FFDM systems should be compared along The costs of FFDM systems should be compared along with the inherent benefits of the digital technology prior with the inherent benefits of the digital technology prior to the purchase:to the purchase:

Reduced recall ratesReduced recall ratesIncreased patient throughputIncreased patient throughputIncreased early detection of breast cancerIncreased early detection of breast cancerDecreased falseDecreased false--negative biopsy resultsnegative biopsy resultsDecreasing film and processing costsDecreasing film and processing costsIncreasing the caseload of each mammography roomIncreasing the caseload of each mammography room

Page 24: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 24

KanalKanal 4747

Clinical Trials and Phantom StudiesClinical Trials and Phantom StudiesLarge screening study screened 49,500 womenLarge screening study screened 49,500 womenDigital Mammographic Imaging Screening TrialDigital Mammographic Imaging Screening Trial (DMIST), (DMIST), funded by NCI and conducted by ACRIN funded by NCI and conducted by ACRIN (http://www.acrin.org/6652_protocol.html)(http://www.acrin.org/6652_protocol.html)

KanalKanal 4848

Future DevelopmentsFuture Developments

Page 25: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 25

KanalKanal 4949

Future DevelopmentsFuture Developments

Dual energy Dual energy CAD integrationCAD integrationTelemammographyTelemammographyTomosynthesisTomosynthesisContrast enhanced digital mammographyContrast enhanced digital mammography

KanalKanal 5050

Future DevelopmentsFuture DevelopmentsTomosynthesisTomosynthesis

Radiology 2005:234:353:362

Page 26: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 26

KanalKanal 5151

Future DevelopmentsFuture DevelopmentsContrast Enhanced Digital MammographyContrast Enhanced Digital Mammography

Radiology 2005:234:353:362

KanalKanal 5252

TAKE HOME POINTSTAKE HOME POINTS

Different technologies exist for digital systems Different technologies exist for digital systems –– indirect and indirect and directdirectCommercially available FFDM systems vary in technologyCommercially available FFDM systems vary in technologyMany advantages exist for FFDM in comparison to FSMMany advantages exist for FFDM in comparison to FSMMTF and DQE are typically used to characterize FFDM MTF and DQE are typically used to characterize FFDM systemssystemsDose is lower with FFDM compared to FSMDose is lower with FFDM compared to FSMAdvanced applications of FFDM include tomosynthesis, Advanced applications of FFDM include tomosynthesis, contrast enhanced mammography etccontrast enhanced mammography etc

Page 27: Digital Mammography Update: Design and Characteristics of … · 2009-07-27 · Digital Mammography Update July 27, 2009 Kanal 4 Kanal 7 Full-Field Digital Mammography (FFDM) Advantage

AAPM 2009 Breast Imaging Course -Digital Mammography Update

July 27, 2009

Kanal 27

KanalKanal 5353

THANK YOUTHANK YOU