Dicom Print x700

  • Upload
    skarpen

  • View
    257

  • Download
    1

Embed Size (px)

DESCRIPTION

Configuracion de Impresora equipo Siemens

Citation preview

  • Siemens, 2014 US25-102.843.01.02.02 ACUSON X70004.14 Ultrasound

    DICOM Configuration 31

    Page 31 of 82

    1. Click New under the Printer list to configure a new DICOM BW printer.

    Fig. 24: DICOM - BW Printer Shown

  • 32 DICOM Configuration

    ACUSON X700 US25-102.843.01.02.02 Siemens, 201404.14 Ultrasound

    Page 32 of 82

    2. Enter the data as described in step 3 through step 26 below. Use the values recorded in (US25-103.843.01 / Appendix A: Preparation for DICOM networking).

    Fig. 25: DICOM BW Printer Setup Window3. For Film Orientation, select either Portrait or Landscape.4. For Display Format, select the number of images per film and their display arrange-

    ment.

    NOTE The available choices include 1x1, 1x2, 2x2, 2x3, 3x2, 3x3,3x5, 4x5, 4x6, and 5x6 (choices are listed by row / column).

    5. For Copies, enter a default number for the number of copies to be produced by each print request.

    6. For Film Size, select the appropriate film size from the following listing, as well as according to the specific printer the customer is installing.

  • Siemens, 2014 US25-102.843.01.02.02 ACUSON X70004.14 Ultrasound

    DICOM Configuration 33

    Page 33 of 82

    - 8 x 10 in- 8.5 x 11 in- 10 x 12 in- 10 x 14 in- 11 x 14 in- 11 x 17 in- 14 x 14 in- 14 x 17 in- 24 x 24 cm- 24 x 30 cm- A3- A4

    7. Select More for more specified configuration. The system displays the DICOM BW Printer Setup window.

    Fig. 26: DICOM - BW Printer More Button Shown8. For Medium Type, select paper, clear film, blue film, transparency, or current as

    required.

  • 34 DICOM Configuration

    ACUSON X700 US25-102.843.01.02.02 Siemens, 201404.14 Ultrasound

    Page 34 of 82

    NOTE If values are not specified for Min. Density and Max. Den-sity, use the following default values: Min. Density 20 andMax Density 299

    9. For Min. Density, enter a printer-specific value for the density of white between 1 and 349. If no value is specified, use 20 as the starting default value.

    10. For Max Density, enter a printer-specific value for the density of black between 1 and 349. If no value is specified, use 299 as the starting default value.

    11. For Border Density, select Black or White to define the color of the border that will be printed around and between all images.

    12. For Trim, select or uncheck the box to specify whether a trip box should be printed around each image.

    13. For Print Priority, select the appropriate value for the priority assigned to the print job: high, medium, or low.

    14. For Film Destination, select magazine, processor, or current (default value = current).15. For Magnification, select the type of interpolation that the printer uses to magnify or

    decrease the size of an image to fit it in the allocated area on the film; the choices include replicate, bilinear, cubic and bob (default value = cubic).

    16. For Smoothing Type, enter alphanumeric characters to define the type of interpolation to be used for cubic magnification.

    NOTE This setting is optional. In most cases, no value is required.

    17. For Empty Image, select Black or White to define the color of the border printed around an image location that does not contain an image.

    18. For Polarity, select Normal to print a black image as black; select Reverse to print a black image as white and a white image as black (as a photographic negative).

    19. For Configuration: If required, enter a printer-specific string that must be sent to the printer as entered.

    NOTE This setting is optional; in most cases, no value is requiredhere.

    20. Click Save to save the new or edited data and to return to the Active Setup window.21. Click Ping to perform an internet protocol Ping of the selected printer.

  • Siemens, 2014 US25-102.843.01.02.02 ACUSON X70004.14 Ultrasound

    DICOM Configuration 35

    Page 35 of 82

    NOTE A successful Ping indicates that the selected device is con-nected to the network, energized, and correctly configured tocommunicate using TCP/IP. The Ping and Echo functionsbecome operational only after the DIMAQ host has been reg-istered from the Host Setup window. Refer to Ping Test Fail-ure for more information.

    Fig. 27: Presets - DICOM - BW Printer Ping Shown

  • 36 DICOM Configuration

    ACUSON X700 US25-102.843.01.02.02 Siemens, 201404.14 Ultrasound

    Page 36 of 82

    Fig. 28: TCP/IP Successfully Contacted System Message22. Click Echo to perform a DICOM network Echo test similar to the Ping test

  • Siemens, 2014 US25-102.843.01.02.02 ACUSON X70004.14 Ultrasound

    DICOM Configuration 37

    Page 37 of 82

    NOTE After the destination device is configured, the Echo functionattempts to verify that the selected storage server is capableof DICOM communications. Refer to Echo Test Failure inthe Troubleshooting section for more information.

    Fig. 29: Presets - DICOM BW Printer - Echo Shown

  • 38 DICOM Configuration

    ACUSON X700 US25-102.843.01.02.02 Siemens, 201404.14 Ultrasound

    Page 38 of 82

    Fig. 30: DICOM Successfully Contacted System Message23. On the Active Setup window, select either Print When Page is Full or Print at End of

    Exam for each printer to designate when printing is to be performed by that printer. 24. Click OK to accept the configuration and exit from the DICOM configuration windows

    NOTE The system will reboot at this time. Please wait for the sys-tem to restart (approximately three (3) minutes).

    25. After the message box System is rebooting, please wait.... disappears, power off the system, wait 15 seconds, then power on the system.

  • Siemens, 2014 US25-102.843.01.02.02 ACUSON X70004.14 Ultrasound

    DICOM Configuration 39

    Page 39 of 82

    DICOM Color Printer Setup Window 0

    1. Click New under the Printer list to configure a new DICOM Color printer.

    Fig. 31: DICOM - Color Printer Shown

  • 40 DICOM Configuration

    ACUSON X700 US25-102.843.01.02.02 Siemens, 201404.14 Ultrasound

    Page 40 of 82

    2. Enter the data as described in step 3 through step 26 below. Use the values recorded in (US25-103.843.01 / Appendix A: Preparation for DICOM networking).

    Fig. 32: DICOM - Color Printer Setup Window3. For Film Orientation, select either Portrait or Landscape.4. For Display Format, select the number of images per film and their display arrange-

    ment.

    NOTE The available choices include 1x1, 1x2, 2x2, 2x3, 3x2, 3x3,3x5, 4x5, 4x6, and 5x6 (choices are listed by row / column).

    5. For copies, enter a default number for the number of copies to be produced by each print request.

    6. For Film Size, select the appropriate film size from the following listing, as well as according to the specific printer the customer is installing.

  • Siemens, 2014 US25-102.843.01.02.02 ACUSON X70004.14 Ultrasound

    DICOM Configuration 41

    Page 41 of 82

    - 8 x 10 in- 8.5 x 11 in- 10 x 12 in- 10 x 14 in- 11 x 14 in- 11 x 17 in- 14 x 14 in- 14 x 17 in- 24 x 24 cm- 24 x 30 cm- A3- A4

    7. Select More for more specified configuration. The system displays the DICOM Color Printer Setup window.

    Fig. 33: DICOM - Color Printer More Button Shown8. For Medium Type, select paper, clear film, blue film, transparency, or current as

    required.

  • 42 DICOM Configuration

    ACUSON X700 US25-102.843.01.02.02 Siemens, 201404.14 Ultrasound

    Page 42 of 82

    NOTE If values are not specified for Min. Density and Max. Den-sity, use the following default values: Min. Density 20 andMax Density 299

    9. For Min. Density, enter a printer-specific value for the density of white between 1 and 349. If no value is specified, use 20 as the starting default value.

    10. For Max Density, enter a printer-specific value for the density of black between 1 and 349. If no value is specified, use 299 as the starting default value.

    11. For Border Density, select Black or White to define the color of the border that will be printed around and between all images.

    12. For Trim, select or uncheck the box to specify whether a trip box should be printed around each image.

    13. For Print Priority, select the appropriate value for the priority assigned to the print job: high, medium, or low.

    14. For Film Destination, select magazine, processor, or current (default value = current).15. For Magnification, select the type of interpolation that the printer uses to magnify or

    decrease the size of an image to fit it in the allocated area on the film; the choices include replicate, bilinear, cubic and bob (default value = cubic).

    16. For Smoothing Type, enter alphanumeric characters to define the type of interpolation to be used for cubic magnification.

    NOTE This setting is optional. In most cases, no value is required.

    17. For Empty Image, select Black or White to define the color of the border printed around an image location that does not contain an image.

    18. For Polarity, select Normal to print a black image as black; select Reverse to print a black image as white and a white image as black (as a photographic negative).

    19. For Configuration: If required, enter a printer-specific string that must be sent to the printer as entered.

    NOTE This setting is optional; in most cases, no value is requiredhere.

    20. Click Close to save the new or edited data and to return to the Active Setup window.21. Click Ping to perform an internet protocol Ping of the selected printer.

  • Siemens, 2014 US25-102.843.01.02.02 ACUSON X70004.14 Ultrasound

    DICOM Configuration 43

    Page 43 of 82

    NOTE A successful Ping indicates that the selected device is con-nected to the network, energized, and correctly configured tocommunicate using TCP/IP. The Ping and Echo functionsbecome operational only after the DIMAQ host has been reg-istered from the Host Setup window. Refer to Ping Test Fail-ure for more information.

    Fig. 34: DICOM - Color Printer Ping Button Shown