63
g Presented by Adam Kesner, Ph.D., DABR Deviceless respiratory motion correction in PET imaging – exploring the potential of novel data driven strategies Assistant Professor, Division of Radiological Sciences, Department of Radiology, University of Colorado School of Medicine AAPM Spring clinical meeting, 2015

Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

g

Presented by Adam Kesner, Ph.D., DABR

Deviceless respiratory motion correction in PET imaging – exploring the potential of novel

data driven strategies

Assistant Professor, Division of Radiological Sciences, Department of Radiology, University of Colorado School of Medicine

AAPM Spring clinical meeting, 2015

Page 2: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Introduction • Respiratory and cardiac motion

are inherent problems in medical imaging

• Limits scan qaulity

• Resolution

• Quantification

• Lesion detectability

• AC artifacts

Simulation

AAPM draft slides 2-2-2015

Page 3: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

0

2

4

6

8

10

12

14

16

1975 1980 1985 1990 1995 2000 2005 2010 2015

FWHM

(mm

)

Evolution of PET Spatial Resolution

Respiratory motion (thorax)

AAPM draft slides 2-2-2015

Page 4: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Introduction • Considered the resolution limiting factor in

thorax imaging • Gating is a potential solution

Image s borrowed from Philips Healthcare website AAPM draft slides 2-2-2015

Page 5: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Introduction • State of respiratory gating technology in nuclear

medicine: o 10+ years of research o Major vendors sell integrated systems o Many clinics own necessary equipment

• Respiratory gating rarely used in routine imaging • (my) question: why is respiratory motion correction

failing its transition into the clinic? • (my) answer: cost / benefit • (my) solution: stick around for the talk!

AAPM draft slides 2-2-2015

Page 6: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Introduction • Cost / benefit of gating

o Most respiratory gating is implemented using hardware based respiratory tracking equipment

o Negatives of such equipment include • Patient discomfort • Prone to setup error • Slower throughput • Increased costs (hardware, training) • Increased radiation dose

o Overall gating represents a change towards complexity when considered for use in routine scanning

AAPM draft slides 2-2-2015

Page 7: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Introduction • Cost / benefit of gating

o There is a fundamental tradeoff when gating Improved resolution comes with the loss of image statistics, o Benefit uncertain

Simulation AAPM draft slides 2-2-2015

Page 8: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Introduction • Gating comes with a cost and has uncertain benefit

If we can bring the cost of gating to nil, and the benefit to guaranteed – that could change the equation.

• We propose to do this with 2 independent /

integrateable steps o Both based on utilizing information currently unused

AAPM draft slides 2-2-2015

Page 9: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Presentation overview Software driven motion control

1. Software driven gating

analogus to hardware

2. “Gating+” method for signal optimization

3. Recovery of continuous motion method for decoupling data from gates it was created with

4. Summary/implications

Automated workflow

AAPM draft slides 2-2-2015

Page 10: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Software driven gating Section I

AAPM draft slides 2-2-2015

Page 11: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Respiratory gating in PET • Hardware driven gating is the field standard • In recent years several software-based

methods have been presented to extract respiratory signal to be used for gating

• Software driven methods appeal o Ease of use o Operator independent o None of the errors in the application of hardware o If integrated properly, their implementation

would be a software add-in, and require no change to current clinical protocols

AAPM draft slides 2-2-2015

Page 12: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Introduction

• The idea behind software based algorithms: There is a lot of information in list mode data

not being utilized

- signal from respiratory motion

• The challenge: How to sort out signal from the noise?

AAPM draft slides 2-2-2015

Page 13: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

IVF method (2009) • Image Voxel Fluctuation method for

extracting respiratory signal from data o Use the fluctuating signals per voxel over time

• ~2*107 voxels in scan o Signal extracted from each voxel evaluated o Global respiratory signal is created as a

combination of many individual voxel contributions

• Different than traditional image based methods of following structural movement o fully automated

AAPM draft slides 2-2-2015

Page 14: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Am

plitu

de

Time

Trigger

voxel

Page 15: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Am

plitu

de

Time

Trigger

Mean signal in delineated area of sinogram

Page 16: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Acquisition of respiratory signal (SRF method)

• Patient receives scan

AAPM draft slides 2-2-2015

Page 17: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Acquisition of respiratory signal (SRF method)

• Data is transferred to computer

PET listmode output

AAPM draft slides 2-2-2015

Page 18: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Acquisition of respiratory signal (SRF method)

• Raw list-mode data is binned into short time sinogram frames (representing consecutive 500 ms acquisitions).

AAPM draft slides 2-2-2015

Page 19: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Acquisition of respiratory signal (SRF method)

• These sinograms are collapsed in the ρ and θ dimensions

~0.04% memory

AAPM draft slides 2-2-2015

Page 20: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Acquisition of respiratory signal (SRF method)

• A one dimensional blurring kernel is applied along the z axis of the resized sinogram data to improve statistics.

AAPM draft slides 2-2-2015

Page 21: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Acquisition of respiratory signal (SRF method)

• The time activity curves for each of the small sinograms’ voxels is analyzed and scored by the ratio of its signal in respiratory frequencies (periodicity 2-9 seconds) relative to its signal in non-respiratory frequencies.

18

1 4

13 22

15

17 31

19

AAPM draft slides 2-2-2015

Page 22: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Acquisition of respiratory signal (SRF method)

• Using the order of highest to lowest score (from step 4) the time activity curves of each of the voxels in the small sinograms are strategically combined using the below steps

AAPM draft slides 2-2-2015

Page 23: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Acquisition of respiratory signal (SRF method)

• Final trace output to be used for sorting data

Time

Am

plitu

de

Time

Am

plitu

de Patient Respiratory Trace

AAPM draft slides 2-2-2015

Page 24: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Acquisition of respiratory signal (SRF method)

• Summary

AAPM draft slides 2-2-2015

Page 25: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Results – SRF method • Comparison of hardware based and software

based signals

AAPM draft slides 2-2-2015

Page 26: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Results

AAPM draft slides 2-2-2015

Page 27: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Results – summary in population

• 22 FDG Human scans

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22Scan

Pear

son

Cor

rela

tion

Coe

ffic

ient

IVF

SRF

GSG

GSGe

AAPM draft slides 2-2-2015

Page 28: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Discussion • Advantages of fully automated data based

algorithms for acquiring respiratory signal from PET data: o Machine independent o Operator independent o Can be written into software package

• Can be used with existing scans or scanners o Fast, we project they can be run in “real time”

• “cost” -> virtually nil o Require no extra efforts or deviation from routine clinical

procedures o If not used alone, can be used as a second check for

vendor hardware systems o If not used in the clinic, can be used in research for large

population studies

AAPM draft slides 2-2-2015

Page 29: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Gated signal optimization: Gating+

Section II

AAPM draft slides 2-2-2015

Page 30: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Introduction • Separating available statistics into phase-bins results

in decreased image quality – less statistics per bin

Simulation

AAPM draft slides 2-2-2015

Page 31: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Introduction

AAPM draft slides 2-2-2015

• A center with gating equipment has a choice

Gated images

Ungated image

Non-linear image morphing correction

Raw scan data

Improved resolution, inferior contrast *Benefits condition specific

Dependable (time tested)

Improved resolution and contrast, uncertain accuracy *Benefits condition specific

It appears utilization of extra

motion information comes with

uncertain risk

Page 32: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Methods • Non-linear image morphing has been proposed for

recombining gated data • We present an alternative strategy for utilizing the

additional information provided by motion characterization -“gating+” o Basic precept: Movement of signal in space is expressed in

intensity fluctuations in individual voxels over the gated frames

o Our methods are based on isolating the fluctuations in voxels, and modulating them according to their reliability

AAPM draft slides 2-2-2015

Presenter
Presentation Notes
Mention : movement of signal in space OVER GATES IMAGES
Page 33: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

• A gated scan can provide two sets of information per voxel

Methods

Correctly gated Randomly gated Triggers

Voxel Fluctuations = motion + noise

Voxel Fluctuations = noise

Simulations AAPM draft slides 2-2-2015

Page 34: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Methods Information in voxel signals can characterize voxel

Sample Voxelx,y,z

FFT ->

FFT -> Uniform amplitudes of all frequencies

Frequency amplitudes will deviate from mean

AAPM draft slides 2-2-2015

Page 35: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Methods Filter non-supported frequencies

Sample Voxelx,y,z

FFT-1 ->

AAPM draft slides 2-2-2015

Page 36: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Methods

AAPM draft slides 2-2-2015

• By looking at the effective “signal” to “noise” ratio at every voxel, we can selectively accept fluctuation information in voxels that benefit from gating, and filter fluctuation information in voxels that do not, thus optimizing information at every voxel.

Voxel at liver lung boundary benefits from gating -> preserve fluctuations

Voxel signal in background tissue is degraded from gating –> dampen fluctuations

Page 37: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Methods • Gating+ protocol:

1. Look at activity over gates in every voxelx,y,z for volume 2. Characterize real fluctuations (correctly gated scan) and noise

(randomly gated scan) through frequency amplitude analysis 3. Accept only those frequencies which are supported by statistics

• Method verification o Simulations o Population gated small animal PET

• Rats, 16 gates, n=85, acquired with various tracers using Siemens Inveon μPET

• Acquired using data driven gating methods (no hardware)

AAPM draft slides 2-2-2015

Presenter
Presentation Notes
Concept described by this equation
Page 38: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Results Relative count statistics

1 4 13 55 2000

Ungated

Gated

Gating+

Motion Map

• Simulation

Relative Counts 1 4 13 55 2000

ungated gated gating+ ungated gated gating+ ungated gated gating+ ungated gated gating+ ungated gated gating+

Max 76% 174% 76% 58% 112% 87% 58% 94% 90% 55% 79% 79% 55% 76% 75%

Volume (70% max) 29% 4% 29% 154% 4% 29% 171% 46% 71% 188% 96% 92% 183% 100% 96%

SUV (mean/bckgrnd) 63% 187% 63% 49% 112% 77% 49% 73% 72% 47% 66% 67% 47% 65% 65%

FWHM 130% 27% 130% 177% 50% 85% 169% 102% 104% 216% 105% 95% 189% 100% 101%

Lesion/background ratio = 3 Upper diaphragm/background = 1.5 Lower diaphragm/background = 3.0

AAPM draft slides 2-2-2015

Page 39: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Results Example PET slice (18F-DMDPA)

AAPM draft slides 2-2-2015

Page 40: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Results

AAPM draft slides 2-2-2015

Page 41: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

• Our gating+ gate combination algorithm offers an alternative approach to optimizing information acquired in a gated scans

• All correction comes down to (simple) 1dimensional equation • Characterizable/reproducible • Fast

o ~20 sec processing for gating+ (µPET volume * 16 gates) • Accuracy:

o All corrected voxels in simulation have a higher probability of being closer to truth than uncorrected voxels

o Corrected image is derived from a selective use of raw information • No offset vectors created from assumptions

• 100% Fully automated

Discussion A

B

1D signal optimization

algorithm

4D non-linear image morphing algorithms

AAPM draft slides 2-2-2015

Page 42: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Discussion • Algorithm works with effective signal

o Irrespective of reconstruction algorithm, smoothing, etc o Irrespective of quality of signal

• Areas not benefiting from gating, or entire scans not benefiting from gating, will be returned to their ungated embodiment

• Algorithm utilizes available information and optimizes its transformation into Cartesian space. o Does not preclude the use of non-linear morphing algorithms

• Potential applications: o Support use of routine gating thorax imaging o Respiratory, Cardiac imaging o Human, small animal o PET, SPECT, CT (low dose 4D CT), MRI…

AAPM draft slides 2-2-2015

Page 43: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Motion-gate information decoupling

Section III

AAPM draft slides 2-2-2015

Page 44: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Introduction • When information is optimized in frequency space,

during the gating+ processing, there is an opportunity to shift the phase of the signal by rotating the frequency components in real and imaginary space. This allows a user to extract a voxel value at any and all phases of the cycle. o Values adhere to the optimized frequency information

• By repeating process for all voxels, can reconstruct phase shifted images o ~0.02 seconds processing per slice

• With this process, we can reconstruct continuous motion image (CMI) sequences

AAPM draft slides 2-2-2015

Page 45: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Voxel time-activity curve phase shifting example

AAPM draft slides 2-2-2015

Page 46: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Phase shifted curve validation

• We generated 106 simulations of true gate-activity o Signals < Nyquist frequency

• Gated (step function) values were derived from the true curves, CMI values were derived from gated curves

• In 100% of the simulations the CMI curves correlated better with truth than the respective gated curves.

Example randomly generated voxel activity vs gate curve simulation.

AAPM draft slides 2-2-2015

Page 47: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Combined data driven workflow results

Data driven gating +

Gating+ +

Phase shifted CMI frame images

AAPM draft slides 2-2-2015

Page 48: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Workflow results • All images created using standard, non-gated,

small animal PET rat acquisitions • Animations created with 90 frames/cycle, displayed

with 30 frames/second

AAPM draft slides 2-2-2015

Page 49: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Movie 1

AAPM draft slides 2-2-2015

Page 50: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Movie 2

AAPM draft slides 2-2-2015

Page 51: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Movie 3

AAPM draft slides 2-2-2015

Page 52: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Results quantified Note – all numbers are average values of all appropriate scans Measurements

Gated Gating

+

Gating+ with

phase shift

Ungated

Number of frames (360ᵒ/n)

16 16 90 1

Kidney VOI n=102

Average kidney uptake [relative values]

Average VOI value, averaged over all gates

1.01 1.01 1.01 1.00

% SD Average uptake across gates

2.69% 1.47% 1.47%

Max displacement [mm]

Average over all scans (SD over all scans)

1.50 (0.56)

1.10 (0.75)

1.10 (0.75)

Liver profile n=34 Liver boundary displacement [mm]

3.25 3.04 3.05

Shoulder VOI n=84 % SD in VOI [relative values]

Average over all gates

1.53 1.05 1.05 1.00

Global COM n=84 Max COM displacement [mm]

Average over all scans (SD over all scans)

0.71 (0.46)

0.43 (0.41)

0.43 (0.41)

AAPM draft slides 2-2-2015

Page 53: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Can it all work in Humans?

AAPM draft slides 2-2-2015

Page 54: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Human work

AAPM draft slides 2-2-2015

Page 55: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Human work

AAPM draft slides 2-2-2015

Page 56: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Human work

AAPM draft slides 2-2-2015

Page 57: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Presentation summary • There is information in PET data that is not being utilized • We present data driven gating, signal optimization, and

information decoupling strategies o Can be used separately or combined in an automated workflow. o Can be implemented in clinical setting with minimal impact o Can be used with minimal risk of degradation of care

• Our strategies can reframe the boundaries of motion control o # of gates vs noise paradigm o Characterization of motion control strategies o Risks of using motion correction o Visualization of motion

• Further validation needed – we provided proof of principles and small population measurements

• Still room for improvement o Not seen limits in accuracy or speed o As technology advances (sensitivity and resolution), so will

potential of such algorithms • Still areas of application to explore

AAPM draft slides 2-2-2015

Page 58: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

AAPM draft slides 2-2-2015

Page 59: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Additional material

AAPM draft slides 2-2-2015

Page 60: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Methods Sinogram Slice

Sinogram Volume Projection

Histogram of Counts in 500ms Sinogram

1

10

100

1000

10000

100000

1000000

10000000

0 1 2 3

Counts

Sinogram SliceSinogram Volume

Projection

Histogram of Counts in 500ms Sinogram

1

10

100

1000

10000

100000

1000000

10000000

0 1 2 3

Counts

Axial Image SliceImage Volume

Projection

Histogram of Counts in 500ms Image

1

10

100

1000

10000

100000

1000000

0 7.7304E-06 1.5461E-05 2.3191E-05

Counts

Axial Image SliceImage Volume

Projection

Histogram of Counts in 500ms Image

1

10

100

1000

10000

100000

1000000

0 7.7304E-06 1.5461E-05 2.3191E-05

CountsAAPM draft slides 2-2-2015

Page 61: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

IVF method (2009) We can then “grow” a motion amplitude curve through sequential combination of information in voxels

Software trace

Hardware trace

AAPM draft slides 2-2-2015

Page 62: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

Human Scan (Standard Clinical FDG scan)

Ungated Gated

Volume Projection

Slice

Slice (zoomed)

IVF method (2009)

AAPM draft slides 2-2-2015

Page 63: Deviceless respiratory motion correction in PET imaging ...amos3.aapm.org/abstracts/pdf/97-25916-353470... · AAPM draft slides 2-2-2015 . Introduction • Cost / benefit of gating

IVF method (2009) • IVF method employs a fully automated software

algorithm o Image based o Require no external equipment or hardware o Come at the “cost” of processing

• Mostly due to image reconstruction

AAPM draft slides 2-2-2015