Professor : Byoung-Kwan Cho Development of Laboratory based portable spectrometer Lalit Kandpal |Kwon kyung-do |Min Hyun jung |Dewi Kusumaningrum |Collins
1ContentsSpectrometer developemnt1Sample spectra collection2Data analysis 43Wavelength calibrationResults5System components
Accessories for spectrometer development
System hardware
General principle of spectrometer
crossed czerny-turner spectrograph arrangement
Laboratory developed spectrometer
Orange juice Sample Preparation10 samples each concentrationTotal :100 samples
3%5%7%10%13%15%17%20%25%22%
For the development of PLS model large amount of samples require, therefore in this study we used 100 samples. Where, Calibration 70 & validation 30 samples.
SPLICCO
CSharpData acquisition softwareHardware calibrationData acquisition
DarkReferenceSample123Spectra collection procedureWavelength calibration with LED light
Waveband (Hyper)Pixel (spectrometer)459.4448531.1626550.1660736.41064893.91449
HSI spectra of LED light LED spectra from spectrometer Correlation between waveband (hyper) & pixel (spectrometer)
Mean spectra
Raw spectra
600 nm: due to the yellow color of juice samples Absorbance decreases as the concentration increases Raw spectra of orange juice
Spectral data (X) Juice concentration (Y)PLS finds a set of orthogonal components that :maximize the level of explanation of both X and Yprovide a predictive equation for Y in terms of the XsThis is done by:fitting a set of components to X (as in PCA)similarly fitting a set of components to Yreconciling the two sets of components so as to maximize explanation of X and Y
X = TP + EY = UQ + FWhere, T= scores/latent vector of X, P=loadings for XU= Scores of Y, Q=loadings of Y, E & F =error of X &Y Data matrix for PLS analysis Data processing (PLS)CalibrationPredication Results: Predicted concentration
Optimal PC no. computed by cross-validation (leave one out)ResultsBeta coefficient of PLS model
colorcolorConclusion & further modificationProper focusing and handling was difficult task during spectrometer development therefore in further modification accurate calibration of spectra and light is highly essential.In future work LABVIEW program will be developed for calibration and data acquisition.Thank You !Chungnam National University, Daejeon, S. KoreaNon-destructive Bio Sensing Laboratory