Development of Isoelectric Focusing

Techniques for Protein Analyses

By

Yanwei Zhan

A thesis

presented to the University of Waterloo

in fulfillment of the

thesis requirement for the degree of

Master of Science

in

Chemistry

Waterloo, Ontario, Canada, 2008 © Yanwei Zhan 2008

Page 2: Development of Isoelectric Focusing Techniques for Protein

Declaration

I hereby declare that I am the sole author of this thesis. This is a true copy of the thesis,

including any required final revisions, as accepted by my examiners.

I understand that my thesis may be made electronically available to the public.

ii

Page 3: Development of Isoelectric Focusing Techniques for Protein

Abstract

Isoelectric focusing (IEF) is a powerful approach in separations of zwitterionic

substances such as proteins, peptides and amino acids. It is important in proteomic research.

Generally, in IEF, carrier ampholytes (CAs) are necessary to establish a stable pH gradient.

However, CAs also bring attendant problems such as a decrease in detection sensitivity and

suppression of ionization of analytes in mass spectrometry (MS) detection. It is desirable to build

a pH gradient without using CAs. A simple slab based design was developed to establish a pH

gradient using the electrolysis of water and the strength of free flow electrophoresis (FFE). The

simple and robust CA free FFE-IEF design was applied in protein fractionation.

In capillary format, capillary isoelectric focusing (CIEF), coupled to MS is a promising

hyphenated technique for biomolecular analysis based on the combination of the high separation

power of CE and the high specificity of MS. Coupling of the instruments is usually achieved

with a coaxial sheath liquid interface, which decreases the detection sensitivity because of the

dilution of sample by the sheath liquid. In this project, nano-electrospray, a sheathless interface,

was used for coupling. Additionally, another major challenge is the presence of CAs which

suppresses the ionization of analytes and contaminates the MS. In order to complete this project,

a microcross union was chosen to couple CIEF with MS. A makeup solution was introduced to

dilute the concentration of CAs after IEF to assist the ionization for MS detection. The makeup

solution could replace the sheath liquid and could be maintained at a low flow rate so that

iii

Page 4: Development of Isoelectric Focusing Techniques for Protein

nanoelectrospray could be performed.

Monoliths can be described as integrated continuous porous separation media for micro

scale separation columns. CAs were immobilized at different positions in the column according

to their pIs, generating a monolithic immobilized pH gradient (M-IPG). In this project, carrier

ampholytes was immobilized in poly (GMA-co-EDMA) based monolithic capillary and poly

(GMA-co-acrylamide) based monolithic capillary to form a pH gradient. Two proteins were

separated by IEF, which was implemented in poly (GMA-co-acrylamide) based monolithic

capillary without CAs. The interface to MS was performed following the use of a microcross

union as described previously. No typical noise of CAs was observed in the MS spectrum.

iv

Page 5: Development of Isoelectric Focusing Techniques for Protein

Acknowledgment

I would like to express my gratitude to my supervisor, Professor Pawliszyn for offering

me the opportunity to do research in his group and giving me such a stimulating project, also for

his encouragement and guidance.

I would like to express my gratitude to my committee members, Professor Chong and

Professor Palmer. I thank them for their interest in my project, encouragement and valuable

advice. I especially thank them for their time in my thesis revision.

I would like to convey heartfelt thanks to my parents, brother, sister-in-law for their

inspiration and support. Specifically, I would like to thank my husband, Jixian, for his

unconditional support, encouragement and patience. I also would like to thank my daughter,

Eliya. She drew two figures for my thesis.

In addition, I also appreciate the support and suggestions from my colleagues in our

group: Dr. Ou, Weng, Tib, Marcel, Erasmus, Peipei, Shine, Jibao, Dajana, Francois, Melissa,

Leslie, Lisa, Gerald, Joanne, Ying, Ehsan, Jet, Dawei, Joy, Xiang, Vadoud, Shokouh, Sanja,

Fumi, Ouyang, Rina, Cherie, Saharnaz, In Yong, Simon and Heather. I appreciate the help from

Tomasz Glawdel.

v

Page 6: Development of Isoelectric Focusing Techniques for Protein

Dedication

I dedicate this thesis to my parents, my husband Jixian and my loved daughter Eliya.

vi

Page 7: Development of Isoelectric Focusing Techniques for Protein

TABLE OF CONTENTS

List of Figures…………………………………………………………………………………….x

List of Abbreviations…………………………………………………………………………...xiii

Chapter 1 Introduction…………………………………………………………..........................1

1.1 Free flow electrophoresis (FFE)……………………………………………………………....1

1.2 Capillary isoelectric focusing (CIEF)………………………………………………………....3

1.2.1 UV, Fluorescence and whole-column imaging detection………………………..……...4

1.2.2 Mass spectrometry……………………………………………………………………….6

1.2.3 Interfaces between CE and MS………………………………………………………….7

1.2.4 Strategies for coupling CIEF with MS…………………………………………………11

1.2.4.1 CA-free system…………………………………………………………………..11

1.2.4.2 Monolithic column with immobilized pH gradient……………………………...13

1.3 Objective of project…………………………………………………………………………..15

References……………………………………………………………………………………18

Chapter 2 Preparation and application of ampholyte-free FFE-IEF for protein

fractionation

2.1 Introduction………………………………………………………………………………….21

2.2 Experimental Sections…………………………………………………………………….....22

2.2.1 Chemicals and materials……………………………………………………………….22

2.2.2 Fabrication of the separation chamber…………………………………………………22

2.2.3 Sample preparation and the IEF procedure…………………………………………….23

2.2.4 CIEF analysis of fractions……………………………………………………………...24

2.3 Result and Discussion………………………………………………………………………..25

2.3.1 Preparation of chamber for free-flow electrophoresis…………………………………25

2.3.2 Formation of the pH gradient………………………………………………………….26

vii

Page 8: Development of Isoelectric Focusing Techniques for Protein

2.3.3 Application for protein separation……………………………………………………..28

2. 4 Conclusions………………………………………………………………………….……....35

References…………………………………………………………………………………...37

Chapter 3 Coupling CIEF with mass spectrometry

3.1 Introduction………………………………………………………………………………….38

3.2 Experimental Sections……………………………………………………………………….38

3.2.1 Instrument…………………………………………………………………………….38

3.2.2 Chemicals and materials……………………………………………………………...38

3.2.3 Sample preparation…………………………………………………………………...39

3.2.4 Method………………………………………………………………………………..39

3.2.4.1 CIEF-WCID method……………………………………………………………39

3.2.4.2 MS optimization………………………………………………………………...39

3.2.4.3 Coupling CIEF with MS………………………………………………………..40

3.3 Results and Discussion……………………………………………………………………..41

3.3.1 Off-line nanoelectrospray with MS…………………………………………………...41

3.3.2 Modification the off-line device for on-line coupling………………………………..44

3.3.3 Optimization of electrospray device………………………………………………….46

3.3.4 Optimization of CIEF condition for its coupling with MS……………………………50

3.3.5 Optimization of MS condition for coupling CIEF……………………………….……52

3.3.6 Coupling CIEF with MS………………………………………………………………56

3. 4 Conclusions…………………………………………………………………………………61

References…………………………………………………………………………………..62

Chapter 4 Preparation of capillary monolithic column with immobilized pH gradient for

coupling CIEF with mass spectrometry

4.1 Introduction………………………………………………………………………………...63

4.2 Experimental Sections……………………………………………………………………...63

viii

Page 9: Development of Isoelectric Focusing Techniques for Protein

4.2.1 Chemicals and materials……………………………………………………………...64

4.2.2 Preparation of monolithic capillary column………………………………………….64

4.2.3 Immobilization of CAs onto the monolithic columns………………………………..65

4.2.4 Coupling M-IPG based CIEF with MS…………………………………………………..66

4.3 Result and Discussion……………………………………………………….……………..67

4.3.1 Preparation and characterization of M-IPG…………………………………………..67

4.3.2 Immobilization of CAs……………………………………………………………….68

4.3.3 Coupling (CIEF) M-IPG with MS……………………….…………………………...68

4. 4 Conclusions…………………………………………………………………………………73

References…………………………………………………………………………………..74

ix

Page 10: Development of Isoelectric Focusing Techniques for Protein

LIST OF FIGURES Figure 1.1 Schematic illustration of FFE.5 Figure 1.2 Illustrations of a capillary isoelectric focusing cartridge. (a) side view, (b)

top view.14

Figure 1.3 Illustration of the basic concept of the whole-column-imaging detection

for CIEF.14

Figure 1.4 Schematic illustration of CE/MS interfaces to an ESI source. (a) Coaxial

sheath-flow interface (b) liquid-junction interface (c) sheathless interface.

Figure 1.5 Schematic illustration of an off-line nano-electrospray device.27

Figure 1.6 Illustration of hollow fiber-based microdialysis device for carrier

ampholytes removal in a CIEF-MS system.30

Figure 1.7 Schematic illustration of FFE for removing CAs.32 Figure 1.8 Schematic illustration of CIEF-RPLC-MS system.34 Figure 1.9 Fundamental reactions of M-IPG.36 Figure 2.1 Schematic designs of the apparatus for free-flow electrophoresis. Figure 2.2 Measured pH values of the sample fractions in each outlet. Figure 2.3 Photographic images show the isoelectric focusing process of hemoglobin

and β-lactoglobulin at different periods.

Figure 2.4 Photographic images show the isoelectric focusing process of myoglobin

and β-lactoglobulin at different periods.

Figure 2.5 Photographic images show the isoelectric focusing process of hemoglobin

and cytochrome c at different periods.

Figure 2.6 CIEF electropherograms of hemoglobin and β-lactoglobulin separation.

Figure 2.7 CIEF electropherograms of myoglobin and β-lactoglobulin separation.

x

Page 11: Development of Isoelectric Focusing Techniques for Protein

Figure 2.8 CIEF electropherograms of hemoglobin and cytochrome c.

Figure 3.1 Schematic connection of CIEF with MS via pressure from syringe.

Figure 3.2 Picture of Protana interface with API 3000.

Figure 3.3 Positive nano-electrospray ionization mass spectrum of testosterone. Figure 3.4 Positive nano-electrospray ionization mass spectrum of cytochrome c. Figure 3.5 Positive nano-electrospray ionization mass spectrum of myoglobin. Figure 3.6 Picture of modified upchurch microcross union.

Figure 3.7 Picture and detailed structure of inside of mocrocross union.

Figure 3.8 Picture of adapter for mounting the spray union.

Figure 3.9 Mass spectrum of 10 μg/mL myoglobin in 99% water with 1% acetic acid.

Figure 3.10 Mass spectrum of 10 μg/mL myoglobin in 99% methanol with 1% acetic

acid.

Figure 3.11 Mass spectrum of 10 μg/mL myoglobin in 50% methanol, 49% water, and

1% acetic acid.

Figure 3.12 Mass spectrum of 1 μg/mL myoglobin in 50% methanol, 49% water, and

1%acetic acid.

Figure 3.13 CIEF electropherograms of myoglobin in different concentration of CAs.

Figure 3.14 CIEF electropherograms of myoglobin in 0.01% of CAs. Figure 3.15 CIEF electropherograms of myoglobin in 0.5% of CAs. Figure 3.16 CIEF electropherograms of myoglobin in 1% of CAs.

Figure 3.17 Mass spectrum of 20 μg/mL myoglobin in 50% methanol, 49% water, and

1% acetic acid with 1% CAs.

Figure 3.18 Mass spectrum of 20 μg/mL myoglobin in 50% methanol, 49% water, and

1% acetic acid with 0.5% CAs.

Figure 3.19 Mass spectrum of 20 μg/mL myoglobin in 50% methanol, 49% water, and

xi

Page 12: Development of Isoelectric Focusing Techniques for Protein

1% acetic acid with 0.1% CAs.

Figure 3.20 Mass spectrum of 20 μg/mL myoglobin in 50% methanol, 49% water, and

1% acetic acid with 0.01% CAs.

Figure 3.21 Mass spectrum of 20 μg/mL myoglobin in water with 0.1% CAs. Figure 3.22 0.5% CAs, 40 μg/mL myoglobin in water mixed with 50% methanol, 49%

water, and 1% acetic acid solution.

Figure 3.23 MS total ion current of samples after the separation in CIEF.

Figure 3.24 Mass spectrum of 10 μg/mL β-lactoglobulin in 50% methanol 49% water

1% acetic acid.

Figure 3.25 Average mass spectrum of time A.

Figure 3.26 Average mass spectrum of time B.

Figure 3.27 Average mass spectrum of time C.

Figure 4.1 Schematic connection of coupling M-IPG with MS. Figure 4.2 SEM imaging of poly (GMA-co-EDMA) based monolithic capillary

(×1000, ×3000).

Figure 4.3 SEM imaging of poly (GMA-co-acrylamide) based monolithic capillary

(×1200, ×2500).

Figure 4.4 Current VS Time during the immobilization procedure.

Figure 4.5 MS total ion current of samples after the separation in CIEF (M-IPG).

Figure 4.6 Average mass spectrum of time A.

Figure 4.7 Average mass spectrum of time B.

Figure 4.8 Average mass spectrum of time C.

xii

Page 13: Development of Isoelectric Focusing Techniques for Protein

LIST OF ABBREVIATIONS

AIBN Azobisisobutyronitrile

CAs Carrier ampholytes

CE Capillary electrophoresis

CEC Capillary electrochromatography

CGC Capillary gel electrophoresis

CIEF Capillary isoelectric focusing

CZE Capillary zone electrophoresis

DMSO Dimethyl sulfoxide

EDMA Ethylenedimethacrylate

EOF Electroosmotic flow

ESI Electrospray ionization

FFE Free flow elctrophoresis

GMA Glycidyl methacrylate

IEF Isoelectric focusing

IPG Immobilized pH gradient

ITP Isotachophoresis

LC/MS Liquid chromatography coupled with mass spectrometry

LIF laser induced fluorescence

MALDI Matrix-assisted laser desorption/ionization

M-IPG Monolithic immobilization pH gradient

xiii

Page 14: Development of Isoelectric Focusing Techniques for Protein

xiv

MS Mass spectrometry / Mass spectrometer

pI Isoelectric point

PVP polyvinypyrrolidone

RPLC Reversed-phase liquid chromatography

2-D gel 2-dimensional gel electrophoresis

UV Ultraviolet

WCID Whole-column imaging detection

γ-MAPS 3-methacryloxypropyltrimethoxysilane

Page 15: Development of Isoelectric Focusing Techniques for Protein

1

Development of Isoelectric Focusing Techniques for Protein

Analyses

Chapter 1 - Introduction

Isoelectric focusing (IEF) is an electrophoretic method in which proteins are separated

based on their isoelectric points (pIs). Usually, a pH gradient is established by carrier

ampholytes (CAs). When a protein is placed in a medium with a linear pH gradient and

subjected to an electric field, it will initially move toward the electrode with the opposite

charge. During the migration through the pH gradient, the protein will either pick up or lose

protons. When the protein eventually arrives at the point in the pH gradient equivalent to its

pI, it will stop migrating because it becomes uncharged. Because the pI is a unique

characteristic of amphoteric substances, IEF is a powerful tool in zwitterionic separation with

high resolution. Slab gel IEF has been widely used in separation and characterization of

proteins and peptides in biotech laboratories.1 In 1985, Hjerten and Zhu first reported an IEF

method performed in a capillary column format.2 In 1989, Chmelik reported an application of

pH gradient in free-flow electrophoresis (FFE) format.3 Two kinds of IEF are applied in this

project. They are FFE-IEF and CIEF.

1.1 Free flow electrophoresis (FFE)

FFE separates charged particles, which are injected continuously into a thin carrier

buffer film flowing between two parallel plates. The electric field, applied perpendicular to

the flow direction, leads to deflection of charged sample components according to their

Page 16: Development of Isoelectric Focusing Techniques for Protein

2

mobility or pI.4 The sample and electrolyte used for the separation enter the separation

chamber at one end and the fractionated sample and the electrolyte are collected at the other

end as shown in Figure 1.1.5 Depending on the electrolyte system applied, FFE can be

performed in three basic modes: (1) zone electrophoretic (ZE), (2) isotachophoresis (ITP), and

(3) IEF. In ZE mode, a narrow sample beam is injected into a background electrolyte of a

constant composition, pH and conductivity, and individual components of the sample are

separated according to the charge-to-size ratios. In ITP mode, the sample is introduced

between leading and terminating electrolytes having the highest and lowest mobility of their

ions in respect to the mobility of analytes. From the principle of the method, it therefore

follows that the sample components can be both separated and concentrated during the

separation process. In IEF, the electrolyte used for the separation is created from a set of

carrier ampholytes, which leads to the formation of a linear pH gradient. The sample can be

introduced either in the mixture with the electrolyte or in a narrow zone. The method is aimed

at the separation of amphoteric solutes, which migrate in the electric field until they reach the

position where the pH of the electrolyte is equal to its pI and become immobile and focused.

FFE offers some important advantages. Firstly, FFE allows a continuous collection of

fractions combined with continuous sample feeding. For example, in preparative applications,

the separation is performed continuously so that potentially hundreds of milligrams of

samples can be separated. Secondly, FFE can be easily applied in different formats. There are

several commercial apparatus for preparative purposes. Furthermore, small-scale separation

is feasible in microfabricated free flow electrophoresis (mFFE). Thirdly, analytes can be

easily collected at the end of the separation area, which can decrease the cost of further

Page 17: Development of Isoelectric Focusing Techniques for Protein

treatment following the separation.

Figure 1.1: Schematic illustration of FFE5

FFE has been applied to a great variety of charged species, from low molecular ions

up to proteins, membranes, and cells.6, 7 The application of FFE in proteomics has been

reviewed by Chmelik.8

1.2 Capillary isoelectric focusing (CIEF)

When IEF is performed in capillary format, a separation column is filled with the

mixture of a protein sample and carrier ampholytes (CAs). The two ends of the column are

dipped into catholyte and anolyte. Under the separation voltage, a pH gradient is established

along the capillary column and the proteins migrate to the point where their pI values are

equivalent to the pH values and the migration stops. The proteins are focused into narrow

zones at their pIs.

3

Page 18: Development of Isoelectric Focusing Techniques for Protein

4

CIEF is expected to have both the high resolution of slab gel IEF and the advantages

of a column-based separation technology that include automation and quantification. CIEF is

widely applied in protein analysis. It can separate the analytes, and at same time, concentrate

them. It resolves the problem of low sample loading in capillary electrophoresis. However, a

major limitation is the presence of CAs, which is necessary in CIEF. High CA concentrations

decrease the sensitivity of MS and UV detection.9

1.2.1 UV, fluorescence, and whole-column imaging detection

Usually, a conventional CIEF instrument is only equipped with a single-point, on-

column detector, which is located close to one end of the column. UV and fluorescence are

common detection modes.10 Therefore, a mobilization process is necessary following the

focusing process. There are three ways to perform the mobilization: hydrodynamic

mobilization, electrophoretic (salt) mobilization, and EOF-driven mobilization.11,12

Mobilization of the focused analytes requires additional time for the analysis, deforms the

established pH gradient, and tends to reduce resolution and reproducibility of separation.

Imaging detection systems and scanning detection systems can free CIEF from these

problems. In particular, real-time whole-column imaging detectors appear to be ideal for

detection of the stationary zones focused within the capillary.

WCID-CIEF was invented by Wu and Pawliszyn in 1994,13 and it has been

commercialized by Convergent Bioscience Ltd.14 The separation column, shown in Figure

1.2, is internally coated, and its outer polyimide coating is removed to let light through. Two

pieces of hollow-fibre membrane are glued to the ends of the separation column. Two

Page 19: Development of Isoelectric Focusing Techniques for Protein

connection capillaries for sample introduction are glued to either end of the hollow-fibre

membrane. The membrane here has two functions. First, the membranes in the electrolyte

reservoirs allow small ions such as H+ and OH- to pass freely, and thus, the CIEF to occur

normally. Second, the membrane can prevent the proteins from going out to the electrolyte

reservoirs while proteins and CAs mixtures are injected to the column. The whole-column

imaging detection system uses ultraviolet (UV) absorbance imaging detector operated at 280

nm. The IEF process in the entire separation column is monitored by the whole-column

detection system, as shown in Figure 1.3. Since the whole-column detector monitors in an

online fashion, all sample zones within the column are recorded by the detector

simultaneously without disturbing the separation. Sample precipitation and aggregation are

the two most common problems in CIEF. Different additives may be selected to improve the

performance in CIEF. These features facilitate fast method development.15

Figure 1.2 Illustrations of a capillary isoelectric focusing cartridge. (a) side view, (b) top view14.

5

Page 20: Development of Isoelectric Focusing Techniques for Protein

Figure 1.3: Illustration of the basic concept of the whole-column imaging detection for CIEF14.

1.2.2 Mass spectrometry

Detection in CE is usually carried out on-column using UV absorbance or

laser-induced fluorescence (LIF).16 However, UV detection is not very sensitive, while LIF

might require solute derivatization with a fluorescent tag. Also, UV and LIF detection do not

provide the information necessary to determine directly the structure of the detected analytes.

A mass spectrometer (MS) is a device that measures the mass-to-charge ratio of ions.

Mass spectrometry is not only a sensitive detection technique; it can be applied for the

detection of wide range of analytes without derivatization and gives information to determine

the structure of the analytes of interest as well. The use of a mass spectrometer offers

unambiguous information of molecular weight and provides structural information helping

with the identification of unknowns. CIEF-MS, which combines the high efficiency and

resolution power of CIEF, with the high selectivity and sensitivity of MS, is an attractive

analytical technique.17

6

Page 21: Development of Isoelectric Focusing Techniques for Protein

7

1.2.3 Interfaces between CE and MS

The predominant ionization method for CE-MS is electrospray ionization (ESI).17

Matrix-assisted laser desorption ionization (MALDI) has also been used extensively.18

Another ionization method called sonic spray ionization (SSI) has also applied in coupling

CE with MS.19

Creating the actual interface between the two techniques is not simple. As the

capillary is held at a high voltage, there are many issues with providing a stable current for

both the CE separation and the MS ionization. Another factor is that CE flow is usually not

high enough to maintain stable ionization within the mass spectrometer source. The interface

is critical.

Three kinds of interfaces have been developed since the first CE-MS interface was

built in 1987 by Smith and his group. Coaxial liquid sheath-flow, sheathless, and

liquid-junction interfaces have been constructed for coupling CE to ESI-MS, as shown in

Figure 1.4.

In the coaxial sheath-flow interface configuration, the outlet of the CE capillary is

simply inserted into the ESI emitters (commonly referred to as the ESI needle). The sprayer is

present as a triple tube; the CE capillary in the centre surrounded by two stainless steel tubes.

The inner one delivers a flow of extra solvent to make up the flow to a suitable level. The

outer tube carries nebulizer gas to assist in droplet formation in the electrospray process.

Electrospray is the process by which the solute ions are vaporized and ionized so that they

can be separated by the MS. This kind of configuration has several advantages, including

simple fabrication, reliability, and ease of implementation. The coaxial liquid sheath interface

uses a makeup flow to provide stable electrospray and complete the CE electrical circuit.

Page 22: Development of Isoelectric Focusing Techniques for Protein

Although the interface suffers from dilution of the analytes by the sheath liquid, it is the most

useful interface for stable and long lasting performance of electrospray. These advantages

make coaxial sheath-flow the most common approach in connecting CE with ESI-MS.

Almost all commercial CE-MS instruments have coaxial sheath-flow interfaces.20,21

Figure 1.4: Schematic illustration of CE/MS interfaces to an ESI source. (a) Coaxial

sheath-flow interface (b) liquid-junction interface (c) sheathless interface.55

.

The liquid junction interface is formed by a tee junction. The tee forms a junction

between the CE terminus and a makeup flow line. The tee transfers the flow to the ion source

of the ESI-MS. A liquid junction is considered to have approximately 5% lower resolution

than a coaxial flow interface. This may be due to the dead volume in the liquid junction

between the CZE terminus and the sprayer.

The main disadvantage of sheath-flow interface is the low sensitivity of detection due

8

Page 23: Development of Isoelectric Focusing Techniques for Protein

9

to the dilution of the analyte by the sheath liquid. To eliminate this weakness of the

sheath-flow interface, a variety of sheathless interfaces have been introduced.22 Mainly two

different approaches are used to fabricate sheathless interface. The first type consists of a

nano-spray needle, a separation capillary and a connection unit. The second approach is using

a separation capillary with a tapered end that is coated with an electrically conducting layer.

The main advantage of a sheathless interface is its high sensitivity because there is no sheath

liquid to dilute the CE effluent. However, sheathless interfaces also have their weaknesses.

For example, the metal coating tip has a short lifetime due to electrochemical degradation.

Moreover, dead volume and air bubbles make sheathless coupling probably not yet

sufficiently robust for extended routine analysis.21-23

The designation “nano-electrospray” was first introduced by Wilm and Mann because

of the low nanolitre per minute flow and droplet size in the nanometre range produced by a

new electrospray technique.24 It is different from a conventional electrospray source by the

small spray tip, the low flow rate, the small size of droplets generated and the absence of

solvent pumps and inlet valves. In conventional ESI, the spray tip is around 100 µm in Ø,

sample flow is around 1-20 µL/min.25 In nano-electrospray, the spray tip is around 1-2 µm in

Ø and the sample flow is around 20-40 nL/min. The diameter of droplets produced by

nano-electrospray is less than 200 nm, which is about 100-1000 times smaller than those

generated by a conventional ESI source.

The advantage of such small droplets is that they have a high surface-to-volume ratio

resulting in high efficiency of ionization. Since no drying gas is needed and the tip can be

brought close to the MS inlet, the sensitivity is better. Since the typical flow rate in CE is

Page 24: Development of Isoelectric Focusing Techniques for Protein

about 20-200 nL/min,26 the flow rate from a nano-electrospray device better matches the CE

flow rate.

The Protana nano-electrospray (Odense, Denmark) is an offline device, as shown in

Figure 1.5.27 The sample is loaded into a metal-coated glass capillary with a tip internal

diameter that ranges from 1 to 4 µm. A voltage is applied to the metal coating on the tip and

an air-filled syringe provides a constant backpressure to initiate and maintain electrospray.

The flow rate is dictated by the electrospray process itself. The nano-electrospray also works

as an ion source, which disperses the sample solution purely by electrostatic means, without

nebulizing gas and without any solvent pump. A low flow rate can be achieved by

electrostatic force through the orifice of the capillary.

Figure 1.5: Schematic illustration of an offline nano-electrospray device.27

The MS instrument, API 3000, available in our lab is equipped with an offline

nano-electrospray interface. Our objective is to modify this offline interface so that online 10

Page 25: Development of Isoelectric Focusing Techniques for Protein

11

CIEF-MS can be performed.

1.2.4 Strategies for coupling CIEF with MS

An MS is the best choice as the detector for CIEF because it can provide the

information of molecular weight and structure. In other words, CIEF has the ability to

separate proteins in a small volume with high efficiency and MS has the compound

identification capability based on accuracy of mass determination.

One traditional bioanalytical and biochemical approach to protein characterization is

two-dimensional (2-D) gel electrophoresis. Protein samples are separated first by pI and then

by size in a two-dimensional gel. 2-D gel electrophoresis can resolve several thousand

proteins. However, it is time-consuming, labor-intensive, and provides only semi-quantitative

information. Following the 2-D idea, coupling CIEF with MS was first introduced by Tang

and colleagues in 1995.28 They used a commercial sheath flow CE-MS instrument for

coupling CIEF to MS. A catholyte reservoir was placed in the ion source house while

focusing and, after focusing, ionization was completed using a sheath flow liquid and

nebulizer gas. In 1996, the same group reported a small modification in which

gravity-assisted mobilization was used to compensate for moving-boundary effects caused by

the sheath liquid ions in the capillary.29

1.2.4.1 CAs-free system

The approaches described above have an obvious limitation. The presence of carrier

ampholytes (CAs) compromises the performance of MS detection because CAs can

contaminate the ionization source, suppress the analyte signal intensity, and degrade the mass

resolution. To solve this problem, the second stage began with the work completed by

Lamoree and colleagues in 1997.30,31 They devised an online microdialysis system between

Page 26: Development of Isoelectric Focusing Techniques for Protein

the separation capillary and a transfer capillary connected to an ESI-MS as shown in Figure

1.6. They still used a sheath liquid flow to assist the ESI.

Figure 1.6: Illustration of hollow fibre-based microdialysis device for CA removal in a

CIEF-MS system.30

In 2000, Chartogne et al. developed a chip-based free-flow electrophoresis device to

remove the carrier ampholytes.32 The device consisted of a 0.1 mm × 10 mm × 25 mm

chamber with one inlet and three outlets. An electric field is applied orthogonally to the

direction of buffer flow.

Figure 1.7 Schematic illustration of FFE for removing CAs.32

In recent years, a chromatographic separation has been inserted after CIEF as not only

a means of removing ampholytes but also means of obtaining additional resolution before MS 12

Page 27: Development of Isoelectric Focusing Techniques for Protein

analysis. In 2003, Jinzhi Chen et al. developed an online combination of CIEF with capillary

reversed-phase liquid chromatography (CRPLC).33 They applied this device to the analysis of

a protein and peptide mixture. In 2004, Zhou at el. also developed an online CIEF-RPLC MS

approach as shown in Figure1.8.34 The advancement demonstrated with this device is the use

of a microselection valve, which can control the delivery of every fraction of the CIEF

capillary. Other applications reported by this method include characterization of human saliva

proteins and examination of membrane proteins.35

Figure 1.8: Schematic illustration of a CIEF-RPLC-MS system.34

1.2.4.2 Monolithic column with immobilized pH gradient

Macroporous materials consist of an interconnected array of polymer microglobules

separated by pores. Monoliths can be described as an integrated continuous porous separation

media in micro-scale separation columns. The mobile phases are forced through the

13

Page 28: Development of Isoelectric Focusing Techniques for Protein

14

separation media, which results in mass transfer between the stationary and mobile phases.37

Monoliths have many advantages such as easy preparation, versatile surface modification,

and high capacity. Because of these advantages, they have been widely used in scientific and

industrial fields.38 In this project, two simple methods are introduced to accomplish an IPG in

the form of a monolith.

EDMA based M-IPG36

Carrier ampholyte is a complex mixture containing thousands of different oligoamino

and oligocarboxylic acids with molecular weights from about 300 to more than 1000. When

CAs and glutaraldehyde are mixed together under proper conditions, CAs can react with

aldehyde groups to produce Schiff bases.

A Schiff base (or azomethine), named after Hugo Schiff, is a functional group that

contains a carbon-nitrogen double bond with the nitrogen atom connected to an aryl or alkyl

group—but not hydrogen. Schiff bases are of the general formula R1R2C=N-R3, where R3 is

an aryl or alkyl group that makes the Schiff base a stable imine.

This reaction is fundamental in M-IPG (Figure 1.9). Certainly the double bond C=N

should be reduced to stabilize the attachment. A focusing procedure fulfills the key role to

achieve the immobilized pH gradient. It locates CA molecules in different positions

according to their pI. This kind of immobilized pH gradient can be applied in IEF separation.

Poly acrylamide-based M-IPG39, 40

The major problem encountered in protein separation is the adsorption to the inner

wall or stationary phase of the column. Polyacrylamide is a neutral and hydrophilic polymer

that has been widely used to solve this problem. Therefore, such a matrix was used to prepare

M-IPG columns for protein analysis.

Page 29: Development of Isoelectric Focusing Techniques for Protein

Figure 1.9: Fundamental reactions of M-IPG.36

Preliminary experiments with poly GMA-EDMA, a hydrophobic monomer as the

supporting material, did not produce good results. Thus, to improve the hydrophilicity of the

polymer, GMA, acrylamide and N, N’-methylenebisacrylamide were chosen as monomers.

Acrylamide and N, N’-methylenebisacrylamide could be dissolved in water but hardly in

aliphatic alcohols, while the reverse was true for GMA. Therefore, DMSO was chosen as a

porogen so that all monomers could be dissolved completely. In addition,

1, 4-butanediol proved helpful in improving the homogeneity, and a long-chain aliphatic

alcohol, dodecanol, could improve the permeability of the monolith.

1.3 Objective of project

Sample preparation is often necessary to separate and concentrate various compounds

prior to analysis of complex samples. In this regard, IEF is one of the best sample preparation

methods. With this approach, however, CAs have to be introduced into the samples, which

may result in matrix interferences. However, both gel IEF and solution IEF need CAs and

some other chemicals to operate. For example, proteins are always dissolved in a solution with 15

Page 30: Development of Isoelectric Focusing Techniques for Protein

16

high concentrations of ampholytes and often urea or detergents. These compounds have to be

removed in the end because they interfere with protein analysis and they are incompatible with

mass spectrometry.41 Therefore, a stable pH gradient without using CAs is considered

desirable.42-48 Methods such as temperature gradient, concentration gradient of neutral

substance, and electrochemical pH gradient are a few techniques explored in the quest for

developing alternative approaches.

Creating pH gradients by electrolysis is not a new technique.46,49 The hydrogen ions

and hydroxyl ions were generated by the decomposition of electrolyte at the electrodes.

Under the electric field, H+ and OH- ions could migrate to cathode and anode, respectively,

and thus create a pH gradient. The theory describing such a system was published by

Hagedorn and Kuhr.50 Huang and Pawliszyn used the electrolysis of water to create a pH

gradient in a capillary for IEF of proteins.51 The pH gradient was created in a capillary

without CAs. However, it was difficult to extract the separated analytes from the capillary. To

solve this issue, a combination of isotachophoresis (ITP) with carrier ampholyte-free IEF was

performed in a commercial isotachophoretic apparatus.48,52 The combination of electrolysis

and FFE was also performed in a microfluidic device in Yager’s group.53,54 In their

microfluidic channel, pH gradients were electrochemically formed. The pH gradients were

visible observed and quantified by using acid-base indicators. These stable pH gradients have

been applied for continuous concentration and separation. In this project, a simple

ampholyte-free IEF free-flow electrophoresis design was developed for the separation of

proteins.

Generally, CE-MS coupling is usually achieved with a coaxial sheath liquid interface,

Page 31: Development of Isoelectric Focusing Techniques for Protein

17

which decreases the detection sensitivity because of the dilution of sample by the sheath

liquid. The advantages of nanospray have been introduced previously. In order to accomplish

the combination of CIEF and MS in our lab, modification of an offline apparatus for online

coupling has to be done first. Another challenge is how to decrease the concentration of CAs

before MS detection. The approaches involve (1) optimization of the condition of CIEF for

MS coupling and (2) introduction of a makeup solution to dilute the CAs and assist ESI. To

complete this project, research has been divided into five stages. These five stages include: (1)

offline set-up of Protana nano-electrospray (offline) with MS; (2) modification of offline

device for online coupling; (3) optimization of performance of electrospray device; (4)

optimization of the CIEF; and finally (5) coupling CIEF with MS.

The other approach is using M-IPG to remove the CAs. CAs were immobilized in

poly-based (GMA-co-EDMA) monolithic capillary and poly-based (GMA-co acrylamide)

monolithic capillary to form a pH gradient. Mixed proteins samples would be separated after

the preparation of monolithic column and the immobilization of CAs. The separation

efficiency of IEF in M-IPG will be proved by MS detection.

Page 32: Development of Isoelectric Focusing Techniques for Protein

18

References

(1) Righetti,P. G.; Bossi, A. Anal. Chim. Acta 1998, 372, 1-19.

(2) Hjerten, S.;Zhu, M. J. Chromatogr. 1985, 346, 265-270.

(3) Chmelik, J.; Deml, M.; Janca, J. Anal. Chem. 1989, 61, 912-914.

(4) Krivankova, L.; Bocek, P. Electrophoresis 1998, 19, 1064-1074.

(5) Weber, G.; Bocek, P. Electrophoresis 1996, 17, 1906-1910.

(6) Reschiglian, P.; Zattoni, A.; Cinque, L.; Roda, B.; Dal Piaz, F.; Roda, A.; Moon, M. H.;

Min, B. R. Anal. Chem. 2004, 76, 2103-2111.

(7) Sanzgiri, R. D.; McKinnon, T. A.; Cooper, B. T. Analyst 2006, 131, 1034-1043.

(8) Chmelik, J. Proteomics 2007, 7, 2719-2728.

(9) Tang, S.; Nesta, D. P.; Maneri, L. R.; Anumula, K. R. J. Pharm. Biomed. Anal. 1999,

19, 569-583.

(10) Lim, H. B.; Lee, J. J.; Lee, K.-J. Electrophoresis 1995, 16, 674-678.

(11) Busnel, J. M.; Kilar, F.; Kasicka, V.; Descroix, S.; Hennion, M. C.; Peltre, G. J.

Chromatogr. A 2005, 1087, 183-188.

(12) Mazzeo, J. R.; Krull, I. S. Anal. Chem.1991, 63, 2852-2857.

(13) Wu, J. Q.; Pawliszyn, J. American Laboratory 1994, 26, 48-52.

(14) Wu, X. Z.; Wu, J. Q.; Pawliszyn, J. Lc Gc North America 2001, 19, 526-545.

(15) Wu, J.; Huang, T.; Pawliszyn, J. Handb. Capillary Microchip Electrophor. Assoc.

Microtech. (3rd Ed.) 2008, 563-579.

(16) Huang, T. M.; Pawliszyn, J. Analyst 2000, 125, 1231-1233.

(17) Smith, R. D.; Udseth, H. R.; Barinaga, C. J.; Edmonds, C. G. J. Chromatogr. 1991,

559, 197-208.

(18) Lechner, M.; Seifner, A.; Rizzi, A. M. Electrophoresis 2008, 29, 1974-1984.

(19) Hirabayashi, Y.; Hirabayashi, A.; Koizumi, H. Rapid Commun. Mass Spectrom. 1999,

13, 712-715.

(20) Ferranti, P.; Pizzano, R.; Garro, G.; Caira, S.; Chianese, L.; Addeo, F. J. Dairy

Research 2001, 68, 35-51.

Page 33: Development of Isoelectric Focusing Techniques for Protein

19

(21) Moini, M. Anal. Bioanal. Chem. 2002, 373, 466-480.

(22) Moini, M. Anal. Chem. 2001, 73, 3497-3501.

(23) Xiao, Z.; Conrads, T. P.; Lucas, D. A.; Janini, G. M.; Schaefer, C. F.; Buetow, K. H.;

Issaq, H. J.; Veenstra, T. D. Electrophoresis 2004, 25, 128-133.

(24) Wilm, M.; Mann, M. Anal. Chem. 1996, 68, 1-8.

(25) Shi, M. H.; Peng, Y. Y.; Yu, S. N.; Liu, B. H.; Kong, J. L. Electrophoresis 2007, 28,

1587-1594.

(26) Jussila, M.; Sinervo, K.; Porras, S. P.; Riekkola, M. L. Electrophoresis 2000, 21,

3311-3317.

(27) Troxler, H.; Kuster, T.; Rhyner, J. A.; Gehrig, P.; Heizmann, C. W. Anal. Biochem.

1999, 268, 64-71.

(28) Tang, Q.; Harrata, A. K.; Lee, C. S. Anal. Chem. 1995, 67, 3515-3519.

(29) Tang, Q.; Harrata, A. K.; Lee, C. S. Anal. Chem. 1996, 68, 2482-2487.

(30) Lamoree, M. H.; Tjaden, U. R.; vanderGreef, J. J. Chromatogr. A 1997, 777, 31-39.

(31) Lamoree, M. H.; van der Hoeven, R. A. M.; Tjaden, U. R.; van der Greef, J. J. Mass

Spectrom. 1998, 33, 453-460.

(32) Chartogne, A.; Tjaden, U. R.; Van der Greef, J. Rapid Commun. Mass Spectrom. 2000,

14, 1269-1274.

(33) Chen, J. Z.; Balgley, B. M.; DeVoe, D. L.; Lee, C. S. Anal. Chem. 2003, 75,

3145-3152.

(34) Zhou, F.; Johnston, M. V. Anal. Chem. 2004, 76, 2734-2740.

(35) Wang, W., Guo, T., Rudnick, P.A., Song, T. J., Zhuang, Li, Z., Zheng, W., DeVoe, D.

L., Lee, C. S., Balgley, B. M. Anal. Chem. 2007, 79, 1002-1009.

(36) Yang, C.; Zhu, G. J.; Zhang, L. H.; Zhang, W. B.; Zhang, Y. K. Electrophoresis 2004,

25, 1729-1734.

(37) Wu, R.; Hu, L.; Wang, F.; Ye, M.; Zou, H. J. Chromatogr. A 2008, 1184, 369-392.

(38) Wu, R.; Zou, H.; Fu, H.; Jin, W.; Ye, M. Electrophoresis 2002, 23, 1239-1245.

(39) Zhu, G. J.; Yuan, H. M.; Zhaol, P.; Zhang, L. H.; Liang, Z.; Zhang, W. B.; Zhang, Y. K.

Electrophoresis 2006, 27, 3578-3583.

Page 34: Development of Isoelectric Focusing Techniques for Protein

20

(40) Zhu, G. J.; Zhang, W. B.; Zhang, L. H.; Liang, Z.; Zhang, Y. K. Sci. China Ser.

B-Chem. 2007, 50, 526-529.

(41) Tang, Q.; Harrata, A. K.; Lee, C. S. Anal. Chem. 1995, 67, 3515-3519.

(42) Huang, T.; Pawliszyn, J. Electrophoresis 2002, 23, 3504-3510.

(43) Lochmuller, C. H.; Ronsick, C. S. Anal. Chim. Acta 1991, 249, 297-302.

(44) Englund, A. K.; Lundahl, P.; Elenbring, K.; Ericson, C.; Hjerten, S. J. Chromatogr. A

1995, 711, 217-222.

(45) Fang, X.-H.; Adams, M.; Pawliszyn, J. Analyst 1999, 124, 335-341.

(46) Pospichal, J.; Chmelik, J.; Deml, M. J. Microcolumn Sep. 1995, 7, 213-219.

(47) Rilbe, H. Electrophoresis 1981, 2, 261-267.

(48) Budilova, J.; Pazourek, J.; Krasensky, P.; Pospichal, J. J. Sep. Sci. 2006, 29,

1613-1621.

(49) Pospichal, J.; Deml, M.; Bocek, P. J.Chromatogr. A 1993, 638, 179-186.

(50) Hagedorn, R.; Fuhr, G. Electrophoresis 1990, 11, 281-289.

(51) Huang, T.; Wu, X.-Z.; Pawliszyn, J. Anal. Chem. 2000, 72, 4758-4761.

(52) Prochazkova, B.; Glovinova, E.; Pospichal, J. Electrophoresis 2007, 28, 2168-2173.

(53) Macounova, K.; Cabrera, C. R.; Holl, M. R.; Yager, P. Anal. Chem. 2000, 72,

3745-3751.

(54) Cabrera, C. R.; Yager, P. Electrophoresis 2001, 22, 355-362.

(55) Brocke, A.V.; Nicholson, G..; Bayer, E. Electrophoresis 2001, 22, 1251-1266.

Page 35: Development of Isoelectric Focusing Techniques for Protein

Chapter 2 Preparation and application of ampholyte-free FFE-IEF

for protein fractionation

This chapter is in press: Yanwei Zhan, Tibebe Lemma, Marcel Florin Musteata, and Janusz

Pawliszyn, J. Chromatogr. A (2008), doi:10.1016/ j.chroma.2008.07.093

Development of a Simple Ampholyte-Free Isoelectric Focusing Slab Electrophoresis for

Protein Fractionation

Yanwei Zhan, Tibebe Lemma, Marcel Florin Musteata, and Janusz Pawliszyn.

The contributions of Tibebe Lemma involved experimental suggestions and helping with

photograph. The contributions of Marcel Florin Musteata involved experimental suggestions and

manuscript revision.

2.1 Introduction Among the most useful chromatographic and electrophoretic techniques, isoelectric

focusing (IEF) is a powerful separation tool. However, the limitations of using CAs are obvious.

The omitting of CAs will facilitate coupling IEF with other techniques.

As mentioned in the introduction, FFE offers some advantages. One of them is easy

collection after the separation. In this chapter, a simple and robust design is developed for the

separation of proteins using FFE-IEF techniques without CAs. Dialysis membranes with 2%

agarose gel are used to establish a stable pH gradient and four proteins are selected as model

compounds. The colour of the selected proteins helps to visually observe the separation process.

Following FFE, the collected samples were analyzed by capillary isoelectric focusing with whole

column image detection in order to determine the fluid composition at each outlet; this approach

21

Page 36: Development of Isoelectric Focusing Techniques for Protein

gives an accurate depiction of the separation that takes place within the developed device.

2.2 Experimental Sections 2.2.1 Chemicals and materials

β-Lactoglobulin (bovine), hemoglobin (human), myoglobin (horse heart), cytochrome c

(horse heart), carrier ampholytes (Pharmalytes 3.0-10.0), tris (hydroxymethyl)-aminomethane

(Tris), dialysis membrane (tubing, COMW 12,000), PVP (MW 360,000) were purchased from

Sigma (St. Louis, MO, USA). Agarose 15 was purchased from BDH Electran (BDH Laboratory

Supplies). pI Markers (4.65, 8.40) were purchased from Convergent Bioscience (Toronto,

Canada), silanized glass beads (20/30 MESH) were purchased from Chromatographic Specialties

Inc., glass wool was purchased from Fisher Scientific. All solutions were prepared using

deionized water (NanoPurity). Unless otherwise stated, all chemicals were of analytical grade.

2.2.2 Fabrication of the separation chamber

The schematic design of the apparatus is shown in Figure2.1. Two microscope slides (75

mm × 50 mm) were glued with two electrolyte reservoirs (plastic pipettes with stoppers) to form

a chamber. The dimensions of this separation chamber are 75 × 50 × 3 mm. Dialysis membranes

were mounted at the two ends of glass slides, which are in the electrolyte reservoirs. Two percent

agarose gel was sealed as a thin film on the dialysis membrane. Eleven outlets (small plastic

pipette tips) were glued at the bottom part of the chamber. Soft plastic tubes with water clamps

were connected to the outlets (tips).The water clamp worked as a valve which could control the

flow rate of the outlets. Glass wool was put on the top of outlets. The chamber was filled with

silanized glass beads as support medium. Glass wool can prevent the glass beads from falling

down through the outlets and blocking tubes. The whole device was placed between two holders.

The chamber was vertical when electrolysis and IEF was performed.

22

Page 37: Development of Isoelectric Focusing Techniques for Protein

Figure 2.1: Schematic design of the apparatus for free-flow electrophoresis.

2.2.3 Sample preparation and the IEF procedure

Four proteins β-lactoglobulin (pI 5.1-5.3), hemoglobin (pI 7.0-7.2), myoglobin (pI 6.9-

7.3) and cytochrome c (pI 9.6) were dissolved in 25 mmol/L Tris-HCl buffer (pH 7.25). All

protein solutions were filtered using Acrodisc Syringe Filter (0.2 µm super membrane, Life

Science). Sample I was a mixture of hemoglobin and β-lactoglobulin (10 mg/mL each), Sample

II was a mixture of myoglobin and β-lactoglobulin (10 mg/mL each), and Sample III was a

mixture of hemoglobin and cytochrome c (10 mg/mL each). Each model sample was separated

using the proposed technique. The samples were directly introduced at the top of the separation

chamber.

Before each run, the chamber and glass beads were rinsed with 25 mmol/ L Tris-HCl (pH

7.25) buffer. Unless otherwise stated, Tris-HCl buffer was always used at this concentration and

pH. Two electrolyte reservoirs were filled with the same buffer. Two platinum electrodes were

inserted into the electrolyte reservoirs as anode and cathode. A constant voltage of 160 V was

23

Page 38: Development of Isoelectric Focusing Techniques for Protein

applied with an EC 105 power supply (EC Apparatus Inc.). The current was also monitored by

the same instrument. The temperature of the buffer in the chamber was monitored by a suitable

thermometer (54 П Thermometer, FLUKE). At the beginning of the experiment, both current and

temperature were higher and gradually decreased as time progressed. Twenty milliampere (mA)

was the maximum current in our FFE system and 65°C was the maximum temperature in the

separation chamber. Usually, the current dropped to1-2 mA, and the temperature decreased to

30°C after 30 minutes, then 0.2 mL mixed proteins was injected in the chamber at the top. The

clamps of the outlets were opened after the sample injection. Gravity made the solution flow

perpendicularly to the electric field. The flow rate of solution from outlets was controlled by

clamps. Tris–HCl buffer was added continuously to the chamber from the top as make up

solution in order to keep the chamber filled with liquid. Fractions from each outlet were collected

in small tubes. Subsequently, each collected fraction was mixed with CAs (3-10) and PVP

solution, and all the fractions were tested by iCE280 instrument to identify the sample

components that eluted at each outlet. The pH value of each collected fraction in small tubes was

tested by a MP 200 pH meter (Melter-Toledo, Sonnerbergstrasse, Switzerland).

2.2.4 CIEF analysis of fractions

CIEF was carried out using a whole column detection system (iCE280 analyzer,

Convergent Bioscience, Toronto, Canada). The detection system consisted of a whole column

optical absorption imaging detector operated at 280 nm. The light source of the whole-column

UV absorption detector was a deuterium (D2) lamp. The separation cartridge was a silica

capillary (50-mm long with 100 µm i.d. and 200 µm o.d. from Convergent Bioscience) and the

external polyimide coating was removed. The inner wall of the capillary was coated with

fluorocarbon (Restek, Bellefonte, PA, USA) to suppress the EOF and to prevent protein

adsorption onto the inner wall of the capillary. The two ends of the column were connected to

24

Page 39: Development of Isoelectric Focusing Techniques for Protein

two pieces of dialysis hollow fibre membrane to separate the protein sample from the external

electrolytes. The two sections of the hollow membrane were inserted into the electrolyte

reservoirs. The further details about the commercial iCE280 instrument have been well described

elsewhere.1,2 Protein samples mixed with CAs could then be continuously injected into the

separation column without changing the electrolytes in the two electrolyte tanks. 100 mmol· L-1

phosphoric acid and sodium hydroxide were used as anolyte and catholyte, respectively.

2.3 Result and Discussion 2.3.1 Preparation of chamber for free-flow electrophoresis

In FFE system, the electric field is applied perpendicularly to the direction of flow. High

resolution separation depends on stable voltage, stable laminar flow and effective chamber

cooling system. Furthermore, the generation of O2 and H2 by electrolysis requires the segregation

of electrolytes and separation chamber. In order to accomplish this task, the chamber was

designed as in Figure 2.1.

Usually, stable laminar flow is achieved with a precision pump. In this design, the pump

was replaced by gravity and flow controlling valves. When the chamber was in a vertical

position, the solution flowed downstream to the outlets because of gravity. The flow rate was

controlled by the outlet valve, which simplified the design. The electrolyte segregation was

achieved by the dialysis membrane and agarose gel. A dialysis membrane and agarose gel was

used to prevent the diffusion of the protein mixtures into the electrolyte reservoirs and to prevent

air bubbles from heading into the separation chamber. Another reason for using agarose gel was

to decrease the conductivity of the chamber. Low conductivity could slow down the electrolysis

reaction, which is very necessary in this system. Otherwise, it takes a long time to cool down the

chamber, as no cooling system was attached to this chamber. At the beginning, the joule heat

25

Page 40: Development of Isoelectric Focusing Techniques for Protein

resulted in high temperature in the electrolyte reservoirs and separation chamber. As time went

on, the electrolysis slowed down and the whole system cooled. In order to avoid the effect of

joule heat on separation, samples were injected after the system cooled down.

The chamber was filled with silanized glass beads, which worked as support medium and

slowed down the diffusion of proteins. In order to prevent the glass beads from falling down

through the outlets, glass wool was put on the top of outlets. The flow rate in this system was

controlled by the water clamp working as a valve. When the electric field was applied, the outlets

were closed. The electrolysis and the migration of cations and anions can last about 30 minutes

without the perpendicular laminar flow. The pH gradient was created during this period. The

samples can be injected into the chamber after the formation of the pH gradient. When the

sample injection was finished, the outlets were opened. The samples were driven by two forces.

One was the gravity, which leads the samples going down vertically. Another one was

electrophoretic mobility under the electric field, which caused the analytes to move to cathode or

anode and focusing due to their different charges. Driven by these two forces, analytes can focus

on a certain position where the pH gradient is pre-established. A suitable flow rate was optimized

by using hemoglobin as an analyte. Hemoglobin could be observed focused in a narrow line

before it went out of the chamber through the outlets.

2.3.2 Formation of the pH gradient

In the presence of an electric field, redox reactions occur at the anode and cathode.3,4 2H2O - 4e- O2 + 4H+ (Anode) (1) 4H2O +4e- 4OH- +2H2 (Cathode) (2) The production rate at the electrodes is high at beginning. Currents are high and a lot of gas

bubbles come off the electrodes. Joule heat leads to high temperatures in the electrolyte

reservoirs and separation chamber. As the electrolysis goes on, more and more H+ and OH- ions

26

Page 41: Development of Isoelectric Focusing Techniques for Protein

collect in the electrolyte reservoirs, respectively. The anode becomes acidic and the cathode

becomes basic. A continuous pH gradient in the separation chamber is established by allowing

the H+ and OH- to migrate into the separation chamber from opposite ends, via diffusion and

electromigration. The pH gradient produced by electrolysis during these experiments was

measured by testing the pH value of samples from each outlet one by one. As indicated in Figure

2.2, the pH values of outlet 1 to 11 were 2.3, 3.1, 3.7, 5.4, 6.6, 7.6, 8.1, 8.4, 8.5, 8.8 and 8.9,

respectively. The slope is steep in the acidic area and is even in the basic area. The generation of

the nonlinear pH gradient is due to several reasons. One possible reason is that the

electrophoretic mobility of protons is different from that of hydroxyl ions. Another important

influence on pH comes from the Tris-HCl buffer. This buffer has strong buffer capacity in pH

range 7 to 9 because the pKa of Tris is 8.05. It can be estimated that different buffer solutions

will generate different pH gradient slope which can be utilized to separate different proteins. The

generated pH gradient was applied to separate mixed proteins, based on their pI points.

0

2

4

6

8

10

1.1 2.1 3.1 4.1 5.1 6.1 7.1

Distance from anode towards cathode (cm)

pH

1 3 5 7 9 1

Outlet order from anode to cathode

1

Figure 2.2: Measured pH values of the sample fractions in each outlet.

27

Page 42: Development of Isoelectric Focusing Techniques for Protein

2.3.3 Application for protein separation

Due to the nature of the samples (brown colour or red colour), the focusing dynamics

could be observed visually as the focusing proceeded. The whole separation process was

recorded by taking photos at different intervals. Four photos that were marked (A), (B), (C) and

(D) in Figures 2.3, 2.4 and 2.5 showed the different phenomena that occurred in electrophoresis

as time went on. As shown in Figure 2.3, when mixed samples of β-lactoglobulin and

hemoglobin were introduced into the separation chamber, the diffusion of proteins after injection

could be clearly observed in Figure 2.3 A. After about 3 minutes, hemoglobin began to focus, as

shown in Figure 2.3 B (a dark dot could be seen). Because not all the hemoglobin focused at this

time, the coloured mass around the dark dot was still brown hemoglobin. As the electrophoresis

went on, it was very clear that the hemoglobin focused into a line without any coloured mass

shown in the Figure 2.3 C. Figure 2.3 D shows the elution of focused hemoglobin into outlet 6.

The focusing process of β-lactoglobulin could not be seen because this protein is colourless.

Similar to Figure 2.3, the process of myoglobin focusing is clearly shown in Figure 2.4.

The process of focusing and separation of Sample Ш (hemoglobin and cytochrome c) is

presented in Figure 2.5. Hemoglobin has brown colour and cytochrome c has red colour. Figure

2.5 A recorded the diffusion of proteins after the sample injection. After about 3 minutes,

hemoglobin began to gradually focus into a dark line while cytochrome c did not focus. Figure

2.5 B shows the dark line in a red coloured mass. Because the pH gradient formed in this

chamber was from 2.3 to 8.9, the pI of hemoglobin was covered and its focusing could be

observed. However, the pI of cytochrome c (9.5) was not in the range of the pH gradient so that

the focusing of cytochrome c could not be seen. Nevertheless, cytochrome c was positively

charged when it was located in the middle of chamber as expected, as the process of

28

Page 43: Development of Isoelectric Focusing Techniques for Protein

electrophoresis continued, hemoglobin remained in its focusing place while cytochrome c moved

gradually to the cathode. Figure 2.5C recorded the trend clearly. Finally, hemoglobin and

cytochrome c could be separated completely and eluted from different outlets as shown in Figure

2.5D. Although the generated pH gradient was out of range for cytochrome c to focus (as

explained in the discussion part), the photos still showed complete separation of the two proteins.

The separation can be confirmed further by testing all the collected fractions from different

outlets using CIEF. Figure 2.6A exhibits the whole electropherogram of hemoglobin and β-

lactoglobulin with two pI markers by CIEF-WCID. It can be seen from Figure 2.6B that β-

lactoglobulin comes out of outlet 4. It is also clear that a lot of hemoglobin and some β-

lactoglobulin come out of outlet 6 as shown in Figure 2.6D. Hemoglobin and β-lactoglobulin are

still mixed in outlet 5 as shown in Figure 2.6C. The other samples collected from the rest of

outlets such as 1 to 3 and 6 to 11 have not been shown in Figure 2.6 because no protein peak

could be observed. The same situation can be observed in Figure 2.7. Myoglobin and β-

lactoglobulin can be separated in this chamber. In Figure 2.8, hemoglobin comes out of outlet 6

and cytochrome c comes out of outlet 8. All these results did match the pH gradient. The pH

value of outlet 4 is around 5.4 which is close to the pI of β-lactoglobulin. The pH value of outlet

6 is around 7.6, which is close to pI of hemoglobin. Cytochrome c went out of outlet 8; the pH

value was 8.1, which did not match the pI of cytochrome c. This was due to the fast flow rate of

samples.

A stable pH gradient was established only after electrolysis. If the sample was injected

while the electric field was applied, no focus phenomena could be observed, and the coloured

samples gradually went out to the outlets. The pH gradient created in chamber could last for

some time. When the proteins from the first injected sample went out from the outlets, a new

29

Page 44: Development of Isoelectric Focusing Techniques for Protein

protein sample was injected again into the chamber, and focusing of proteins could still be

observed.

The pH resolution of the chamber is limited by its width. This resolution of the presently

described chamber was not high enough to separate hemoglobin from myoglobin. Once the

mixture of hemoglobin and myoglobin was injected into the chamber, only one focusing line was

observed.

A C

Hb and Lac

Hb focused

DB

Hb began focusing

Figure 2.3: Photographic images show the isoelectric focusing process of hemoglobin and β-

lactoglobulin at different periods. Photo A shows the diffusion of protein samples after the

injection. Photo B shows the beginning of focusing. Photo C shows the clear focusing of

hemoglobin as time goes on and Photo D shows the elution of focused hemoglobin into outlet 6.

30

Page 45: Development of Isoelectric Focusing Techniques for Protein

Figure 2.4: Photographic images show the isoelectric focusing process of myoglobin and β-

lactoglobulin at different periods. Photo A shows the diffusion of protein samples after the

injection. Photo B shows the beginning of focusing. Photo C shows the clear focusing of

myoglobin as time goes on and Photo D shows the elution of focused myoglobin into outlet 6.

31

Page 46: Development of Isoelectric Focusing Techniques for Protein

Figure 2.5: Photographic images show the isoelectric focusing process of hemoglobin and

cytochrome c at different periods. Photo A shows the diffusion of protein samples after the

injection. In Photo B, hemoglobin began its focus (dark line) while cytochrome c did not focus.

In Photo C, as the electrophoresis time increased, hemoglobin still stayed at its focusing place

while cytochrome c moved gradually to the cathode. Finally, hemoglobin and cytochrome c

could be separated completely and eluted from different outlets as shown in Photo D.

32

Page 47: Development of Isoelectric Focusing Techniques for Protein

Figure 2.6: CIEF electropherograms of hemoglobin and β-lactoglobulin separation: (A)

hemoglobin and β-lactoglobin and two pI markers (control) (B) collected sample from outlet 4

(C) collected samples from outlet 5 and (D) collected sample from outlet 6. In profile A, Hb and

Lac (0.2mg/mL each) were mixed with ampholyte buffer containing 2% carrier ampholyte pH 3-

10, 0.5% pvp and 2µL of pI markers (pH 4.65 and 8.40). In profile B, sample fraction from

outlet 4 was mixed with ampholyte buffer containing 2% carrier ampholyte pH 3-10, 0.5% pvp

and 2µL of pI markers (pH 4.65 and 8.40). Profile C and D are for outlets 5 and 6, respectively.

The profile A is used as a reference for profile B, C and D. The voltage used for focusing was

500V for the first 2 minutes and 3000V for the next 15 minutes. Peaks in the electropherograms

are labelled.

33

Page 48: Development of Isoelectric Focusing Techniques for Protein

Figure 2.7: CIEF electropherograms of myoglobin and β-lactoglobulin separation: (A)

myoglobin and β-lactoglobulin and two pI markers (control) (B) collected sample from outlet 4

(C) collected sample from outlet 5 and (D) collected sample from outlet 6. In profile A, Myo and

Lac (0.2mg/mL each) were mixed with ampholyte buffer containing 2% carrier ampholyte pH 3-

10, 0.5% pvp and 2µL of pI markers (pH 4.65 and 8.40. Other conditions are the same as in

Figure2.6.

34

Page 49: Development of Isoelectric Focusing Techniques for Protein

Figure 2.8: CIEF electropherograms of hemoglobin and cytochrome c: (A) hemoglobin and

cytochrome c and one pI marker (control) (B) collected sample from outlet 6 (C) collected

samples from outlet 8. In profile A, Hb and Cyt (0.2mg/mL each) were mixed with ampholyte

buffer containing 2% carrier ampholyte pH 3-10, 0.5% pvp and 2µL of pI marker (pH 4.65). In

profile B, sample fraction from outlet 6 was mixed with ampholyte buffer containing 2% carrier

ampholyte pH 3-10, 0.5% pvp and 2µL of pI marker (pH 4.65). Profile C is for outlet 8. The

profile A is used as a reference for profile B and C. Other conditions are the same as in Figure

2.6.

2.4 Conclusions

A simple and practical apparatus that can generate a pH gradient capable of separating

proteins according to their pI has been developed. Even though the pH gradients have no buffer

capacity and are not linear, our design still shows potential for easy and rapid sample

fractionation or preliminary fractionation. In addition, samples can be withdrawn from the

35

Page 50: Development of Isoelectric Focusing Techniques for Protein

appropriate outlets for further investigations (characterization) after the separation. This can save

the cost of CAs and resolve the problem of having to remove the ampholytes before MS analysis.

Therefore, it is ideal for coupling with MS.

Several factors can influence the pH gradient generated by the device, such as

composition of buffer, time of electrolysis and the length of chamber.5,6 After further research on

these factors is finished, it is anticipated that this approach will be very practical for industrial

applications. The medium can be reused and the design can be easily automated. It is hoped that

this method will be useful in designing microchip devices for micro-preparative separation of

proteins as well as preparative-scale separation.

36

Page 51: Development of Isoelectric Focusing Techniques for Protein

37

References

(1) Mao, Q.; Pawliszyn, J. J. Biochem. Biophy. Methods 1999, 39, 93-110.

(2) Wu, J.; Pawliszyn, J. Electrophoresis 1995, 16, 670-673.

(3) Zumdahl, S. S. Chemical Principles; Lexington, Mass.: D.C. Heath and Co., c1992, 1992.

(4) Corstjens, H.; Billiet, H. A.; Frank, J.; Luyben, K. C. Electrophoresis 1996, 17, 137-143.

(5) Macounova, K.; Cabrera, C. R.; Holl, M. R.; Yager, P. Anal. Chem.2000, 72, 3745-3751.

(6) Macounova, K.; Cabrera, C. R.; Yager, P. Anal. Chem. 2001, 73, 1627-1633.

Page 52: Development of Isoelectric Focusing Techniques for Protein

Chapter 3 Coupling CIEF with mass spectrometry

3.1 Introduction

The objective of this project is to integrate CIEF with MS using a nano-electrospray

interface for protein analysis. Modification of an offline interface for the online coupling of

CIEF-MS is the main task of this project. Because a high concentration of CAs and most

additives used in CIEF are not compatible with MS,1,2 decreasing the concentration of CAs is

another task. The approaches involve (1) optimization of the condition of CIEF for MS coupling

and (2) introduction of a makeup solution to dilute the CAs and assist ESI.

The advantage of a liquid-junction interface is the independent control of the separation

and spray capillaries.3 Therefore, a microcross union, which is similar to liquid-junction interface,

was selected to perform the coupling.

3.2 Experiments

3.2.1 Instrument: Protana nano-electrospray (Odense, Denmark), API 3000 (Applied

Biosystems, MDS Sciex), iCE 280 (Convergent Bioscience), Syringe pump (Kd Scientific

780100 USA).

3.2.2 Chemicals and materials: Pharmalytes, cytochrome c (horse heart,), myoglobin (horse

heart) were obtained from Sigma Aldrich (Canada), testosterone in acetonitrile was obtained

from Cerilliant (USA), ammonia; methanol, acetic acid and ethanol were obtained from Fisher

Scientific (Canada).

Nano ES spray Capillary and modified upchurch microcross union was obtained from Proxeon

38

Page 53: Development of Isoelectric Focusing Techniques for Protein

Biosystems (USA), Genuine Eppendorf® GeloaderTM Tips (1-10μL) was obtained from

Eppendorf (Germany). Spray capillary for modified union (Pico TipTM) was obtained from New

Objective (USA). Capillary with 100 μm i.d. × 360 μm o.d. for monolithic column and

connection was obtained from Polymicro Technologies (USA).

All solutions were prepared using deionized water (NanoPurity). Unless otherwise stated, all

chemicals were of analytical grade.

3.2.3 Sample preparation: 1mg/mL testosterone standard solution was diluted to 1μg/mL

testosterone in acetonitrile:water (50:50) solution,

Myoglobin and cytochrome c were dissolved in deionized water (1 mg/mL). Concentrated

samples were diluted to desired concentration using methanol:water:acetic acid (50:49:1)

solution or water.

All solutions were filtered using an Acrodisc syringe filter (0.2 µm super membrane, Life

Science).

3.2.4 Methods

3.2.4.1 CIEF-WCID method: Please see Chapter 2 Experimental Section 2.2.4

3.2.4.2 MS optimization

The MS used was an API 3000 (Sciex Canada) triple quadrupole equipped with a Protana

nano-electrospray ionization source. A home-modified electrospray interface was used to

perform the direct infusion experiment and also to couple CIEF with MS. To obtain a stable

spray, some parameters had to be carefully adjusted and optimized. Samples were pushed

through the capillary and spray tip to MS by syringe pump. The quadrupole was scanned from

m/z 500 to 1500 at a scan rate of 3 second/scan.

Some important parameters such as composition of background solution, spray voltage,

39

Page 54: Development of Isoelectric Focusing Techniques for Protein

spray tip position, and concentration of analytes were studied.

3.2.4.3 Coupling CIEF with MS

When CIEF was completed with an iCE instrument, the cartridge was taken out. One

connecting capillary next to the anode was connected to a syringe pump using a capillary adapter

and a sleeve, another connecting a capillary next to the cathode was connected to one end of the

microcross union using a capillary adapter and sleeve. The makeup solution (methanol, water,

acetic acid) was introduced from the other end by capillary and syringe pump. The mixture was

pushed to the spray tip for ESI. Figure 3.1 shows the schematic connection. The MS instrument

can record spectra of the whole process. From the total ion current spectrum, any time point, and

any time range can be chosen to check the mass spectrum of analyte at that time point or check

the average mass spectrum within that time range.

Figure 3.1 Schematic connection of CIEF with MS via pressure from syringe.

40

Page 55: Development of Isoelectric Focusing Techniques for Protein

3.3 Results and Discussion

3.3.1 Offline nano-electrospray with MS

Figure 3.2: Picture of Protana interface with API 3000.

The Protana nano-electrospray interface system (Figure 3.2) consists of three major

components: the ion source, microscope and monitors, and spray tip holder. The holder has an

arm for mounting the spray tip. The arm is controlled by an X-Y-Z manipulator. This manipulator

and monitors are used for fine adjustment of the position of spray tip in front of the orifice of a

MS

Loading samples into spray tips without introducing any air bubbles is not easy and

requires practice.

41

Page 56: Development of Isoelectric Focusing Techniques for Protein

Testosterone, cytochrome c, and myoglobin were tested with the Protana nano-

electrospray device. In positive mode, the strongest signals for testosterone were obtained at m/z

289.1 and m/z 289.4 (shown in Figure 3.3) in the conventional ESI and nano-electrospray,

respectively. It clearly shows that the nano-electrospray source has been successfully set up with

the MS.

+Q1: 23 MCA scans from Sample 1 (TuneSampleID) of MT20070313180752.wif f (Ion Spray) Max. 6.8e7 cps.

150 160 170 180 190 200 210 220 230 240 250 260 270 280 290 300m/z, amu

5.0e6

1.0e7

1.5e7

2.0e7

2.5e7

3.0e7

3.5e7

4.0e7

4.5e7

5.0e7

5.5e7

6.0e7

6.5e7

6.8e7 289.4

215.3

239.3222.3

170.9156.9 225.1185.3227.1187.3 279.4241.1 245.2 266.0263.2181.8 271.2199.3 281.2 291.3213.3203.2 228.5217.3160.0 230.5163.3 168.9 197.2191.3 277.3257.4152.4 237.3 293.3247.2

Figure 3.3: Positive nano-electrospray ionization mass spectrum of testosterone.

Electrospray produces multiple charged ions for proteins. There are several equations to

calculate the number of charges and the molecular weight.4 For adjacent peaks, assume N2 = N1+

1, where N is number of charges (raw value is rounded to the nearest integer). The detected M

(m/z) is given by

M1 = (MW+N1) /N1 (2)

where MW is the actual mass. The measured M is the sum of the mass plus the mass of the

42

Page 57: Development of Isoelectric Focusing Techniques for Protein

protons forming the positive ion. Then

N2 = (M1-1)/ (M2–M1) (3)

And

MW = N2 (M2-1) (4)

+Q1: 9 MCA scans from Sample 1 (TuneSampleID) of MT20070313161924.wiff (Ion Spray) Max. 2.6e7 cps.

500 550 600 650 700 750 800 850 900 950 1000 1050 1100 1150 1200 1250 1300 1350 1400 1450 1500m/z, amu

1.0e6

2.0e6

3.0e6

4.0e6

5.0e6

6.0e6

7.0e6

8.0e6

9.0e6

1.0e7

1.1e7

1.2e7

1.3e7

1.4e7

1.5e7

1.6e7

1.7e7

1.8e7

1.9e7

2.0e7

2.1e7

2.2e7

2.3e7

2.4e7

2.5e72.6e7 728.1

773.4

687.8

825.0

651.4883.7

831.6

619.1891.0 951.8693.1

569.4525.5 657.0786.0589.7 959.2838.0740.5613.4 1030.9503.3

701.5 792.8745.8555.3 967.1677.0 900.8635.3 846.7 1124.8529.1 1047.3 1142.9 1246.6983.7 1256.6 1374.3 1396.01086.6 1303.8 1455.1

Figure 3.4: Positive nano-electrospray ionization mass spectrum of cytochrome c.

From these equations, in Figure 3.4, it can be calculated that m/z of 1030.9 is the result of

12 positive charges on cytochrome c. MW of cytochrome c is 12,359 Dalton. The charge range is

from 12-21. From Figure 3.5, it can be calculated that m/z of 1305.1 is the result of 13 positive

charges on myoglobin. MW of myoglobin is about 16,953 Dalton. These results acquired with

the Protana nano-electrospray are in good agreement of related reports in the literature. This

indicates that off line Protana nano-electrospray is working properly.

43

Page 58: Development of Isoelectric Focusing Techniques for Protein

+Q 1 : 1 8 M CA s ca n s fr o m S a m p le 1 (T u n e S a mp le ID) o f MT 2 0 0 7 0 31 3 1 7 5 8 1 2 .wif f (Io n S p ra y) M a x. 9 .1 e 7 c p s.

4 0 0 4 5 0 5 0 0 5 5 0 60 0 6 5 0 70 0 7 5 0 8 00 8 5 0 9 0 0 9 5 0 1 0 00 1 0 5 0 1 1 0 0 1 1 5 0 1 2 0 0 1 2 5 0 1 3 0 0 1 3 5 0 1 4 0 0 14 5 0 1 5 0 0m /z, a m u

5.0 e 6

1.0 e 7

1.5 e 7

2.0 e 7

2.5 e 7

3.0 e 7

3.5 e 7

4.0 e 7

4.5 e 7

5.0 e 7

5.5 e 7

6.0 e 7

6.5 e 7

7.0 e 7

7.5 e 7

8.0 e 7

8.5 e 7

9.0 e 76 1 6 .3

8 9 3 .38 4 8 .8

8 0 8 .2 9 42 .9

7 7 1 .7

9 9 8 .3 10 6 0 .57 3 8.2

11 3 1 .31 2 1 1 .9

89 8 .57 0 7 .4 8 5 3 .49 4 8.28 1 2 .96 7 9 .0

1 3 0 5 .177 5 .9 1 0 6 6 .61 1 37 .96 5 3 .1

6 9 8 .45 6 9 .55 25 .64 81 .34 3 7 .4 62 9 .1 1 3 1 2.51 4 1 3.75 4 7 .6 1 0 10 .17 4 3 .8 1 1 4 4 .49 0 4 .7 1 2 2 1.47 2 0.4 1 2 3 1 .69 5 4 .78 8 3 .05 9 1 .6 8 4 1 .1 1 0 38 .86 7 4 .34 2 7.1 4 6 0 .6 1 3 5 8 .3 1 4 3 2.31 2 7 9 .9

Figure 3.5: Positive nano-electrospray ionization mass spectrum of myoglobin

3.3.2 Modification of offline device for online coupling

In chromatography ESI-MS coupling, a sufficient flow of the eluent is assured and

electrospray high voltage can be selected independently of the separation. However, in CE-ESI-

MS, coupling requires finding the proper balance between the liquid flow, separation voltage,

and the electrospray conditions. A variety of ESI interfaces have been previously developed for

the CE-MS including coaxial sheath flow, sheathless, and liquid junction arrangements. The

coaxial sheath flow interface is the only design available on commercial instruments. It is robust

yet its sensitivity is usually lower due to the dilution of the solutes by the makeup liquid.

In our lab, only this offline design is available. To be compatible with the MS instrument,

the electro power provided by the MS for ESI has to be utilized. WCID-CIEF has its own electric

power system. After focusing, analytes are mobilized toward to the MS by pressure application

via a syringe pump. It is similar to LC-ESI-MS. Based on the above description; a liquid-

junction interface was chosen to replace the offline interface as shown in Figures 3.6 and 3.7.

44

Page 59: Development of Isoelectric Focusing Techniques for Protein

This kind of union has four ends. One end has an integrated platinum electrode, which

the electricity from MS for the ESI can be connected to. One end can be used for inserting the

spray tip with the sleeve and the opposite end can be used for connecting the CIEF capillary. The

fourth end is flexible. This flexibility is the advantage of this design, especially for coupling

CIEF with MS.

Figure 3.6: Picture of modified upchurch microcross union.

Figure 3.7: Picture and detailed structure of inside of microcross union.

After the union is chosen, the adapter for mounting the union on the arm of the Protana is

45

Page 60: Development of Isoelectric Focusing Techniques for Protein

designed and fabricated in the machine shop of University of Waterloo as shown in Figure 3.8

3.3.3 Optimization of electrospray device

To obtain stable electrospray, some parameters have to be carefully adjusted and

optimized. An infusion method was used. Samples were pushed through the capillary and spray

tip to the MS by a syringe pump. Some important parameters such as composition of background

solution, spray voltage, spray tip position, and concentration of analytes have been studied.

Figure 3.8: Picture of adapter for mounting the spray union.

Generally, the composition of background solution should depend on the analyte of

interest. There are still some general rules. Our results confirm the general rule: moderately polar

organic solvent is often required to achieve a stable spray. For example, the ESI of myoglobin

under the condition of 99% water with 1% acetic acid is fairly poor. In the mass spectrum

(Figure 3.9), no fingerprint of myoglobin can be obtained. This is likely because the surface

tension of water is large so that it is difficult to get fine spray under regular voltage. The ESI of

myoglobin in 99% methanol with 1% acetic acid shows some improvement because the surface

46

Page 61: Development of Isoelectric Focusing Techniques for Protein

tension is decreased due to the organic solvent. Some of the fingerprint of myoglobin such as

m/z 998 and 1060 can be recognized from Figure 3.10. The slight improvement is due to the fact

that the efficiency of ionization of acetic acid in pure methanol is poor so that the protonation of

myoglobin is not sufficient. In contrast, ESI of myoglobin under 50% methanol 49% water 1%

acetic acid is very stable. The fingerprint of myoglobin is very clear in Figure 3.11.

The spray voltage is also important. Usually, there is a critical voltage below which no

stable ESI can be obtained. When the spray voltage is higher than the critical, stable ESI can be

obtained. However, higher voltage will give higher noise.

The position of spray tip is very important. From experience, if the tip is too far from the

MS orifice, too few ions can enter the MS for detection. If the tip is too close to the orifice, the

curtain gas will directly block the ions from getting into MS. There is a best position where best

total ion current (TIC) and mass spectra of analytes are obtained.

After the optimization of position of spray tip and spray voltage, different concentrations

of myoglobin, such as 100 μg/mL, 10 μg/mL, and 1 μg/mL in 50% methanol, 49% water, and

1% acetic acid, were tested. The mass spectrum of myoglobin at 100 μg/mL is similar to the

mass spectrum of myoglobin at 10 μg/mL (Figure 3.11). The only difference is the peak intensity.

When the concentration of myoglobin decreases to 1 μg/mL, the mass spectrum still showed the

fingerprint of myoglobin, but the multiple charge characterization is not obvious as shown in

Figure 3.12.

Stable ESI only can be acquired from a solution which has both organic solvent and water.

This means that the protein sample after IEF has to be in solution with methanol:water:acetic

acid. Therefore, it is necessary to introduce makeup solution after CIEF.

47

Page 62: Development of Isoelectric Focusing Techniques for Protein

+Q1: 4.007 min from S ample 1 (TuneSampleID) of MT20071017121007.wiff (Ion S pray) Max. 3.6e5 cps.

Figure 3.9: Mass spectrum of 10 μg/mL myoglobin in 99% water with 1% acetic acid.

Figure 3.10: Mass spectrum of 10 μg/mL myoglobin in 99% methanol with 1% acetic acid.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu

548.8

597.5662.8

771.3

551.1

805.6668.2

631.1 645.1

5.0e4

1.0e5

1.5e5

2.0e5

2.5e5

3.0e5

3.5e5

514.4586.4 909.1773.9

531.6 650.7963.2737.4 851.2 1485.4

751.6 937.6 990.3713.2611.4 931.3564.3 707.3 1034.3

757.9 875.8622.8 947.1 1086.3866.4 1133.3523.9 1008.3 1195.6 1346.31060.0 1270.5896.3886.9 1105.1 1403.81017.4832.7 970.4 1177.61281.2

+Q1: 0 .751 m in from S ample 1 (TuneSam pleID ) o f M T20071017151528.w iff ( Ion S pray) M ax. 9 .7e5 c ps.

500 6 00 7 00 800 9 00 1000 11 00 12 00 13 00 1400 150 0m /z, am u

1.0e5

2.0e5

3.0e5

4.0e5

5.0e5

6.0e5

7.0e5

8.0e5

9.0e5

9.7e5 822.0

794.1

616.1513.2

535.6

765.9709.9 848.6508.7 905.9

737.6631.4

541.3687.7

581.3 711.0557.3 601.8 893.9

583.1 647.9567.1 998.0843.8807.91060.9

877.9548.9 645.6 865.3619.6 943.1659.8738.4 1012.8837.8787.3 916.8695.8 956.6

969.9673.4 1040.8725.5 760.4958.6 1050.01069.11019.5755.3 1332.21240.8

1211.21130.61096.11203.0 1477.01360.51114.3 1270.81156.3 1396.5

1285.01353.7 1483.1

48

Page 63: Development of Isoelectric Focusing Techniques for Protein

+Q1: 4.958 min from S ample 1 (TuneSampleID) of MT20071015213038.w iff ( Ion S pray) Max. 2.1e6 cps.

Figure 3.11: Mass spectrum of 10 μg/mL myoglobin in 50% methanol, 49% water, and 1% acetic

acid. +Q1: 4.958 min from S ample 1 (TuneSampleID) of MT20071015204632.wiff (Ion S pray) Max. 2.6e6 cps.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu

2.0e5

4.0e5

6.0e5

8.0e5

1.0e6

1.2e6

1.4e6

1.6e6

1.8e6

2.0e6

2.2e6

2.4e6

2.6e6 616.4

648.5

514.8 942.8848.8737.8

808.2698.2 893.2679.4 1060.6775.6707.6 998.1555.5 1211.31131.0597.1 823.5753.3

564.2 953.2909.3868.6657.1 1137.4605.5 688.8 1304.51218.71078.8845.1 879.6 967.5800.4 1005.4923.0 1148.71089.7 1311.71225.3

934.3 1240.0 1413.61049.9 1339.41102.6 1328.3 1421.91292.4

Figure 3.12: Mass spectrum of 1 μg/mL myoglobin in 50% methanol, 49% water, and 1%acetic

acid.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu

2.0e5

4.0e5

6.0e5

8.0e5

1.0e6

1.2e6

1.4e6

1.6e6

1.8e6

2.0e62.1e6 892.9

942.8848.9 1060.3898.6

998.0807.8 1211.71066.7948.1771.5

1131.1853.7

738.11304.7

812.71218.6

616.3

1312.4775.71009.8

707.4

741.9 953.9 1072.3

1421.7

860.0

1320.11145.7653.0 908.4684.0 1228.2514.9

1018.7533.1 714.8632.8 822.0754.7545.2 837.3575.2 968.2698.2 873.2 1086.4586.7 727.7 1430.31240.91027.8 1157.8760.4 916.9881.2 1326.9662.11039.0925.7 1165.41098.8983.7

1447.71347.11174.5 1264.71195.41365.81290.5 1456.4

49

Page 64: Development of Isoelectric Focusing Techniques for Protein

3.3.4 Optimization of CIEF conditions for its coupling with MS

Coupling CIEF with MS can combine the high separation resolution from CIEF and the structure

information from MS. However, the major difficulty in coupling is that high concentration of

CAs and most of the additives used in CIEF are not compatible with MS. Usually, the buffer

capacity of CAs is determined by their concentration. However, CAs suppress the ionization of

analytes in ESI. There should be a trade-off between the concentration required for CIEF and the

concentration tolerable for MS. A workable compromise can be found.

0 1 2 3 4 5

- 4 0 0 0

- 2 0 0 0

0

2 0 0 0

4 0 0 0

6 0 0 0

8 0 0 0

1 0 0 0 0

1 2 0 0 0

1 4 0 0 0

1 6 0 0 0

1 8 0 0 0

2 0 0 0 0

2 2 0 0 0

UV

abs

orba

nce

D i s t a n c e f o r m a n o d e ( c m )

0.1%CAs

0.25%CAs

0.5%CAs

1%CAs

2% CAs

0 1 2 3 4 5

- 4 0 0 0

- 2 0 0 0

0

2 0 0 0

4 0 0 0

6 0 0 0

8 0 0 0

1 0 0 0 0

1 2 0 0 0

1 4 0 0 0

1 6 0 0 0

1 8 0 0 0

2 0 0 0 0

2 2 0 0 0

UV

abs

orba

nce

D i s t a n c e f o r m a n o d e ( c m )

0.1%CAs

0.25%CAs

0.5%CAs

1%CAs

2% CAs

0.1%CAs

0.25%CAs

0.5%CAs

1%CAs

2% CAs

Figure 3.13: CIEF electropherograms of myoglobin in different concentration of CAs

Some proteins are easily precipitated and aggregated at their pI value. Many chemicals

can be used as additives to reduce the chance of protein precipitation and aggregation during the

CIEF. In CIEF methods, polymer solutions are usually added into the samples, such as methyl

50

Page 65: Development of Isoelectric Focusing Techniques for Protein

cellulose (MC) and polyvinylpyrrolidone (PVP). The purpose of the polymer is to modify the

surface of the separation column and to enhance the resolution of the separation by decreasing

sample diffusion. In this research, they were omitted because their high viscosity can

contaminate the MS. After taking all this into consideration, myoglobin was used as a standard

analyte to test the effects of CAs with different concentrations without any additives and

polymers. A total five of different concentrations of CAs (2%, 1% 0.5%, 0.25%, and 0.1 %) were

evaluated with the same concentration of myoglobin added to each cartridge. The CIEF

electropherograms are shown in Figure 3.13. From Figure 3.13, it can be found that when the

concentration of carrier ampholytes decrease to 0.1%, the separation (three peaks of myoglobin)

is still acceptable. There is no significant change of resolution and peak shape.

From Figure 3.13, some interesting phenomenon can be found. When the concentration

of CAs gradually decreased, the peaks of myoglobin moved away from the anode. One reason

that might cause the shift is because the inner coating is not strong enough to completely

suppress the EOF. Another reason is that no polymer was added into the mixture of analyte and

CAs. It shows the carrier ampholytes function not only to form the pH gradient and control the

capacity of buffer. It also might have the capability of suppress the EOF. To test this hypothesis,

solutions with three different concentration of CAs (0.01%, 0.5%, and 1%) and the same

concentration of myoglobin were applied to CIEF using the same cartridge. A total of 30 minutes

of focusing electropherograms were recorded. Four different times, 5, 10, 20, and 30 minutes,

were chosen to demonstrate the whole process. Figures 3.14, 3.15, and 3.16 show the

electropherograms of myoglobin with 0.01%, 0.5%, and 1% concentration of CAs.

From Figure 3.14, the electropherogram of myoglobin at 30 minutes has almost

disappeared from the range of detection. In Figure 3.15, the electropherogram of myoglobin at

51

Page 66: Development of Isoelectric Focusing Techniques for Protein

30 minutes was still in the detection range. It shows that higher concentration of CAs has high

capability to suppress EOF.

10 minutes

0 1 2 3 4 5-2000

0

2000

4000

6000

8000

10000

12000

14000

UV

abs

orba

nce

at 2

80nm

Distance from anode (cm)

5 minutes

20 minutes

30 minutes

10 minutes

0 1 2 3 4 5-2000

0

2000

4000

6000

8000

10000

12000

14000

UV

abs

orba

nce

at 2

80nm

Distance from anode (cm)

5 minutes

20 minutes

30 minutes

0 1 2 3 4 5-2000

0

2000

4000

6000

8000

10000

12000

14000

UV

abs

orba

nce

at 2

80nm

Distance from anode (cm)

5 minutes

20 minutes

30 minutes

5 minutes

20 minutes

30 minutes

Figure 3.14: CIEF electropherograms of myoglobin in 0.01% of CAs.

3.3.5 Optimization of MS conditions for coupling CIEF

To prepare sample solutions with different concentration of CAs, different amounts of

CAs were added to a series of solutions with the same concentration of myoglobin (20 μg/mL).

An infusion method was used. The effects of CAs on ESI were tested by trying different CAs

concentration such as 1%, 0.5%, 0.4%, 0.2%, 0.1%, and 0.01%.

From Figure 3.17, it is apparent that the ionization of myoglobin is totally suppressed in

1% of CAs. It showed that when the concentration of CAs decreased to 0.5%, the ionization of

myoglobin can be clearly observed as shown is Figure 3.18. The ratio of signal-to-noise is not

high enough. The ratio of signal-to-noise is improved when the concentration of CAs was

decreased to 0.1% as shown in Figure 3.19.

52

Page 67: Development of Isoelectric Focusing Techniques for Protein

10 minutes

0 1 2 3 4 5

-2000

0

2000

4000

6000

8000

10000

12000

14000

16000

18000

UV

abs

orba

nce

at 2

80nm

D istance from anode (cm)

5 minutes

20 minutes

30 minutes

10 minutes

0 1 2 3 4 5

-2000

0

2000

4000

6000

8000

10000

12000

14000

16000

18000

UV

abs

orba

nce

at 2

80nm

D istance from anode (cm)

5 minutes

20 minutes

30 minutes

0 1 2 3 4 5

-2000

0

2000

4000

6000

8000

10000

12000

14000

16000

18000

UV

abs

orba

nce

at 2

80nm

D istance from anode (cm)

5 minutes

20 minutes

30 minutes

Figure 3.15: CIEF electropherograms of myoglobin in 0.5% of CAs.

0 1 2 3 4 5

-4000

-2000

0

2000

4000

6000

8000

10000

12000

14000

16000

UV

abs

orba

nce

at 2

80nm

D istance from anode (cm )

5 minutes

10 minutes

20 minutes

30 minutes

0 1 2 3 4 5

-4000

-2000

0

2000

4000

6000

8000

10000

12000

14000

16000

UV

abs

orba

nce

at 2

80nm

D istance from anode (cm )

5 minutes

10 minutes

20 minutes

30 minutes

5 minutes

10 minutes

20 minutes

30 minutes

Figure 3.16: CIEF electropherograms of myoglobin in 1% of CAs.

53

Page 68: Development of Isoelectric Focusing Techniques for Protein

+Q1: 3.806 m in from S ample 1 (TuneSam pleID ) of M T20071015214534.w iff ( Ion S pray) M ax. 3.0e6 c ps.

Figure 3.17: Mass spectrum of 20 μg/mL myoglobin in 50% methanol, 49% water, and 1% acetic

acid with 1% CAs

+Q1 : 2 .003 to 2 .855 m in from Sam p le 16 (0 .5 C A m yo in m wa 3) o f 20080610.wiff (Io ... M ax. 2 .6e6 cps.

500 6 00 7 00 800 900 1000 1100 1200 13 00 14 00 1500m /z, am u

0.0

2.0e5

4.0e5

6.0e5

8.0e5

1.0e6

1.2e6

1.4e6

1.6e6

1.8e6

2.0e6

2.2e6

2.4e6

2.6e6 5 06 .4

5 20.4

5 14 .7

5 6 2.5

6 0 9.55 68 .6 6 37 .5

607 .7 66 7.6547 .5695 .4

6 63 .5629 .8 7 07 .7727 .7 76 8 .7

786 .677 4.4

8 40.6790 .5 998.1942.6859 .5 106 0.411 31 .29 17.6 12 11 .8999 .5 1032.9 13 05.01 247 .51 18 3.1 1413.7

Figure 3.18: Mass spectrum of 20 μg/mL myoglobin in 50% methanol, 49% water, and 1% acetic

acid with 0.5% CAs.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m /z, amu

2.0e5

4.0e5

6.0e5

8.0e5

1.0e6

1.2e6

1.4e6

1.6e6

1.8e6

2.0e6

2.2e6

2.4e6

2.6e6

2.8e63.0e6 506.2

522.3 536.5

552.6582.6

563.5

609.7651.4 665.5590.5

667.2 695.4 713.7

634.7

692.5754.5 784.6

741.6 766.8 844.8810.6855.8871.6827.6 997.8885.5 899.2 958.3 1057.9 1131.1

983.8924.4 937.6 1211.21135.01030.6 1046.4 1269.8 1459.31249.8 1304.71182.2 1354.01119.0

1404.91382.71106.0

54

Page 69: Development of Isoelectric Focusing Techniques for Protein

+Q1: 2.204 to 3.406 min from Sample 7 (0.1C A m yo after myo2) of 20080610.wiff (Io.. . Max. 3.3e6 cps.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu

0.0

5.0e5

1.0e6

1.5e6

2.0e6

2.5e6

3.0e6

506.4

536.3

564.5 609.5560.5 576.6

639.5598.4

695.4661.7 727.7

751.6 1060.4998.2733.6 1131.2810.4790.6 1211.8

1305.1831.2836.7 899.5976.8 1413.7920.5 1132.7983.5

1163.4 1240.71114.9 1310.1 1417.4

Figure 3.19: Mass spectrum of 20 μg/mL myoglobin in 50% methanol, 49% water, and 1% acetic

acid with 0.1% CAs.

+Q1: 1.152 to 3.256 min from Sample 6 (0.01%CAs Myo 2) of 20080721.wiff (Ion Spr... Max. 2.7e6 cps.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu

0.0

2.0e5

4.0e5

6.0e5

8.0e5

1.0e6

1.2e6

1.4e6

1.6e6

1.8e6

2.0e6

2.2e6

2.4e6

2.6e6616.4

506.4998.1

1060.4942.7893.1 1131.0 1211.8522.4

553.6 848.7526.2 1305.0

552.6 609.6 808.3567.6583.4 771.5

572.5 665.5 1413.7634.7 1000.7602.7 656.4 738.4 1061.7945.2

1134.1850.7721.4 1214.9702.6840.5 1308.4854.5761.5 900.7 1005.8 1417.31231.5916.6 1068.8 1255.9 1352.31173.51052.8 1465.1

Figure 3.20: Mass spectrum of 20 μg/mL myoglobin in 50% methanol, 49% water, and 1% acetic

acid with 0.01% CAs.

55

Page 70: Development of Isoelectric Focusing Techniques for Protein

It is a very interesting observation that when the concentration of CAs was decreased to

0.01%, the effect of suppression was still observed from the mass spectrum and Figure 3.20

shows the effect of suppression is still obvious. This result argued against the merit of coupling

using a dialysis system, therefore, the introduction of methanol, water, and acetic acid solution to

dilute to concentration of CAs after focusing was evaluated. As mentioned previously, the

methanol, water, and acetic acid solution can assist with ionization. The following experiments

proved introducing makeup solution could end with better results.

A CIEF solution, 20 μg/mL myoglobin in water with 0.1% CAs, was prepared. This

solution was infused through the spray tip. The acquired mass spectrum of myoglobin, shown as

Figure 3.21, is quite different. The fingerprint of myoglobin could not be observed. It proved

again that the ESI effect in water solution without methanol and acid was poor.

Another CIEF solution, 40 μg/mL myoglobin in water with 0.5% CAs, was infused from

one end, the makeup solution (methanol, water, and acetic acid) was introduced from the other

end. Both of them were mixed together in the cross section before they were being eluted

through the spray tip for ESI. From Figure 3.22, the acquired mass spectrum of myoglobin

clearly showed the fingerprint of myoglobin. It shows that ESI is good. From this result, it can be

proved that the microcross union design is a proper design for coupling CIEF with MS.

3.3.6 Coupling CIEF with MS

After all the optimization was done, the WCID-CIEF separation was performed using

0.5% CAs without any additives and polymers. β-Lactoglobulin and myoglobin were selected as

model analytes. After the focusing in iCE280 instrument, the cartridge was taken out. One

connecting capillary next to the anode was connected to syringe pump using capillary adapter

56

Page 71: Development of Isoelectric Focusing Techniques for Protein

+Q1: 4.257 to 4.357 min from Sample 11 (0.1CA myo (water)4) of 20080610.wiff (Ion... Max. 8.9e5 cps.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu

0.0

1.0e5

2.0e5

3.0e5

4.0e5

5.0e5

6.0e5

7.0e5

8.0e5

526.1552.4

535.9566.6

550.9

798.4562.2612.6

598.5617.9 782.7

539.6 854.7 1115.3872.7641.2 786.5631.8 1305.1690.0 1112.6520.5 936.7 1210.2

611.2 852.7 1011.2635.4 718.6 1489.81240.91089.8 1262.9934.6571.2 1216.4885.5590.0 984.5681.4 1442.5707.5 1248.5 1392.1868.0 1057.8961.7

Figure 3.21: Mass spectrum of 20 μg/mL myoglobin in water with 0.1% CAs.

+Q1: 12.621 to 16.378 min from Sample 24 (mix mwa and myo+ca) of 20080610.wiff ... Max. 2.2e6 cps.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu

0.0

2.0e5

4.0e5

6.0e5

8.0e5

1.0e6

1.2e6

1.4e6

1.6e6

1.8e6

2.0e6

2.2e6506.4

536.5

514.5

519.6 582.5609.6

625.6627.6558.6 648.5

681.6633.7 697.5

712.6 1060.6721.5 998.2848.7 1131.2731.5

765.6 1211.8828.5 854.5 1304.9957.5 1063.0874.7 1134.0 1413.81041.6 1214.8 1234.6

Figure 3.22: 0.5% CAs, 40 μg/mL myoglobin in water mixed with 50% methanol, 49% water,

and 1% acetic acid solution.

57

Page 72: Development of Isoelectric Focusing Techniques for Protein

and sleeves, capillary next to the cathode was connected to one end of the microcross union

using capillary adapter and sleeves too.The MWA (methanol, water, and acetic acid) was

introduced by capillary and syringe pump. The mixture was pushed to the spray tip for ESI. MS

can record the spectra of the whole process. From the total ion current spectrum, any time point

and any time range can be chosen to check the mass spectrum of analyte at that time point or

check the average mass spectrum within that time range. Figure 3.23 is the TIC spectrum of β-

lactoglobulin and myoglobin which were focused in WCID-CIEF before they were sent to MS

for detection. Theoretically, myoglobin will be closer to the cathode than β-lactoglobulin because

pI of myoglobin is 6.8-7.2 while the pI of β-lactoglobulin is 5.1 to 5.3. Accordingly, the

myoglobin will be eluted before the β-lactoglobulin. From Figure 3.23, three different three time

intervals which were marked as A, B, and C were chosen to check the mass spectrum for

identification. Before the identification, the fingerprint of β-lactoglobulin should be tested and

shown.

TIC o f +Q 1: fro m S a m ple 2 4 (sam ple fina l) o f 2 0 08 0 6 23 .w iff (Io n Sp ra y) M a x. 4 .2 e 9 cp s.

1 2 3 4 5 6 7 8 9Tim e, min

0.0

5.0e8

1.0e9

1.5e9

2.0e9

2.5e9

3.0e9

3.5e9

4.0e94.2e9

A B C

Figure 3.23: MS total ion current of samples after the separation in CIEF.

58

Page 73: Development of Isoelectric Focusing Techniques for Protein

+Q1 : 5 .559 to 6 .711 m in from Sam p le 4 (Lactog lobu lin 2 ) o f 20080707.wif f (Ion S p ray.. . M ax. 1 .4e6 cps.

Figure 3.24: Mass spectrum of 10 μg/mL β-lactoglobulin in 50% methanol, 49% water, and 1%

acetic acid. +Q1: 0.401 to 0.852 min from Sample 24 (sample final) of 20080623.wiff (Ion Spray) Max. 2.2e6 cps.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu

0.0

2.0e5

4.0e5

6.0e5

8.0e5

1.0e6

1.2e6

1.4e6

1.6e6

1.8e6

2.0e6

616.3506.1

522.0

553.5

595.5552.5

625.3651.5560.5 998.2 1060.4942.7653.4

1131.1893.2669.0680.6 1211.9683.6

808.3 1305.01413.6701.6 728.4 753.5

1061.7828.6790.7 858.4 948.51415.51312.51137.6 1214.81009.6918.5 940.6 1080.4 1319.81150.5 1429.9

Figure 3.25: Average mass spectrum of time A.

500 6 00 7 00 8 00 9 00 10 00 11 00 12 00 13 00 1400 150 0m /z, am u

0.0

1.0e5

2.0e5

3.0e5

4.0e5

5.0e5

6.0e5

7.0e5

8.0e5

9.0e5

1.0e6

1.1e6

1.2e6

1.3e6

1.4e61 22 5.3

1 148 .7

1 21 9 .513 12 .8

1 14 3 .5

1 41 3 .7

13 06 .710 81 .2

1 0 76 .2 14 0 7.01 2 31 .7

1 31 2.012 18 .71 4 21 .111 54 .85 7 3.3

12 46 .610 21 .16 30 .9 6 8 1.3 1 16 8 .8513 .3 14 16 .484 0 .4 1 336 .51 10 0 .2 1 25 3.15 94.7 9 08 .07 79 .1 9 67 .5 99 8.77 03 .2 14 5 5.31 13 1.6

59

Page 74: Development of Isoelectric Focusing Techniques for Protein

+Q1: 2.304 to 2.955 min from Sample 24 (sample final) of 20080623.wiff (Ion Spray) Max. 1.7e6 cps.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu

0.0

2.0e5

4.0e5

6.0e5

8.0e5

1.0e6

1.2e6

1.4e6

1.6e6506.3

616.4536.5

505.01225.2

552.5581.7

541.6 1312.6611.3 651.4 1148.8

584.1606.8 1413.8695.2 848.7 893.2

713.4 738.3 808.3 998.2 1131.0754.0

1211.9751.31305.0814.4 898.3 1415.3820.3 1154.81082.6999.5886.7 1246.5967.5 1438.81327.81184.51108.8

Figure 3.26: Average mass spectrum of time B.

+Q1: 4.708 to 5.309 min from Sample 24 (sample final) of 20080623.wiff (Ion Spray) Max. 6.3e5 cps.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu

0.0

5.0e4

1.0e5

1.5e5

2.0e5

2.5e5

3.0e5

3.5e5

4.0e5

4.5e5

5.0e5

5.5e5

6.0e5

616.4508.4

536.5

564.2

514.9 1225.3

567.11148.5 1312.7595.1572.1

611.1602.6659.4 689.3

1081.0502.4 706.8 737.9 1413.6808.0 1231.7893.2780.5758.8 1319.7998.2 1150.2905.4 1421.3878.4 1130.9

1233.1921.8 976.6 1252.51107.4 1339.4 1459.2

Figure 3.27: Average mass spectrum of time C.

Figure 3.23 indicates the MS TIC of samples after the IEF separation of β-lactoglobulin

60

Page 75: Development of Isoelectric Focusing Techniques for Protein

and myoglobin, and it was stable. It could be seen from Figure 3.25 that the MS fingerprints of

myoglobin emerged slowly, which indicates myoglobin was focused in stage A; similarly, the β-

lactoglobulin was focused in stage C (Figure 3.27). However, both the fingerprints of β-

lactoglobulin and myoglobin emerged in stage B at the same time as shown in Figure 3.26,

which indicates some β-lactoglobulin and myoglobin are still mixed in stage B. It might be due

to the dead volume in the connection capillary that destroyed the separation performance,

resulting in the focused zone mix a little. It is anticipated that the separation efficiency could be

improved if another mobilization method could be applied.

3.4 Conclusions

The microcross design is suitable for coupling CE with MS, especially in coupling CIEF

with MS. A low CA concentration demonstrated efficient focusing, and at the same time, does

not cause significant signal interference of MS detection. CIEF has been coupled with MS using

nanospray.

61

Page 76: Development of Isoelectric Focusing Techniques for Protein

62

References (1) Tang, Q.; Harrata, A. K.; Lee, C. S. Anal. Chem.1995, 67, 3515-3519.

(2) Tang, Q.; Harrata, A. K.; Lee, C. S. Anal. Chem. 1996, 68, 2482-2487

(3) Jussila, M.; Sinervo, K.; Porras, S. P.; Riekkola, M. L. Electrophoresis 2000, 21, 3311-

3317.

(4) R.Weinberger Practical Capillary Electrophoresis; Academic Press Inc.,(San Diego),

1993.

Page 77: Development of Isoelectric Focusing Techniques for Protein

Chapter 4 Preparation of capillary monolithic column with

immobilized pH gradient for coupling CIEF with mass spectrometry

4.1 Introduction

Capillary isoelectric focusing is a mode of capillary electrophoresis that exhibits

powerful separation ability for ampholytic molecules such as proteins. The separation principle is

based on the differential isoelectric points in a chemically formed pH gradient under the

influence of an applied electric field. Commercial carrier ampholytes (CAs) are commonly used

for establishing a pH gradient. After finishing the IEF process, the focused analytes zones are

mobilized to pass through the detection point for UV absorbance detection. Due to the high

absorbance of CAs at low wavelength, the detection sensitivity is very low.

Much effort has been made to perform CIEF without CAs including autofocusing and

temperature- and electrolysis-generated pH gradient.1-4 Furthermore, omitting CAs will facilitate

coupling of IEF with mass spectrometry (MS), which is a wonderful detection technique for

proteomics research. Hochstrasser et al. prepared a CA-immobilized pH gradient (CA-IPG) in a

glass tube.5 Zhang’s group developed an immobilized pH gradient in a monolithic capillary

column.6 CAs were immobilized at different positions in the column according to their pIs,

generating a monolithic immobilized pH gradient (M-IPG). The IEF was implemented in such a

column without soluble CAs. In our case, the M-IPG columns were prepared and coupled with

MS after IEF of protein separation.

63

Page 78: Development of Isoelectric Focusing Techniques for Protein

4.2 Experimental Sections

4.2.1 Chemicals and materials

Glycidyl methacrylate (GMA), ethylenedimethacrylate (EDMA), glutaraldehyde (50% in

water), dodecanol, cyclohexanol, 1,4-butanediol, n-propanol, azobisisobutyronitrile (AIBN),

sodium borohydride (NaBH4), acrylamide, N,N’-methylenebisacrylamide, dimethyl sulfoxide

(DMSO), 3-methacryloxypropyltrimethoxysilane (γ-MAPS), and 1, 4-butanediol were obtained

from Sigma (Canada). Ammonia, methanol, acetic acid, and ethanol were obtained from Fisher

Scientific (Canada). Modified upchurch microcross union was obtained from Proxeon

Biosystems (USA), spray capillary for modified union (Pico TipTM) was obtained from New

Objective (USA). Capillary with 100 μm i.d. × 360 μm o.d. for monolithic column and

connection was obtained from Polymicro Technologies (USA).Water bath 150D is purchased

from VWR.

4.2.2 Preparation of monolithic capillary column

Fused-silica capillary was successively washed with NaOH (1.0 mol/L), water, HCl (1.0

mol/L), water, and methanol for 30 min, then dried in a gas chromatographic oven with nitrogen

flush at 70 °C for 3 hours. The cleaned capillary was filled with a solution of γ-MAPS (30% v/v

in methanol) and kept at room temperature with both the ends sealed by silicon septa. About 24

hours later, unreacted γ-MAPS was washed off using 0.5 mL of methanol. The capillary was

rinsed by nitrogen at room temperature for 30 min to remove the residual methanol.

Poly-base (GMA-co-EDMA) monolithic capillary column was synthesized by filling the

capillary with the mixture of 30 μL GMA, 10 μL EDMA, 110 μL cyclohexanol, 10 μL dodecanol,

and 0.3 mg AIBN. The capillary that was about 10 cm long was filled with the solution and then

was put in a water bath at 60 °C. The polymerization was stopped after 18 hours, and unreacted

64

Page 79: Development of Isoelectric Focusing Techniques for Protein

monomers were removed by flushing the capillary with methanol using an HPLC pump.

Similarly, the polyacrylamide-based monolithic capillary column was also synthesized.

EDMA was replaced by acrylamide and N,N’-methylenebisarylamide. 1.4 Butanediol and

DMSO was used as porogenic solvent. GMA (18.9 mg), acrylamide (16.7 mg), and N,N’-

methylenebisacrylamide (40.7 mg) were dissolved in the mixture of 333.3 mg of DMSO, 148.2

mg of 1, 4-butanediol, and 185.1 mg of dodecanol. After the addition of AIBN (% wt with

respect to monomers), the mixture was vortexed for a few minutes, and then degassed with

ultrasonic for 2 minutes. Subsequently, the mixture was injected into the pretreated capillary.

With both ends sealed, the capillary was put in an oven at 75 °C for 24 hours. The monolith in

the capillary was then washed with ethanol for 2 hours, followed by water for another 1 hour

using an HPLC pump.

4.2.3 Immobilization of CAs onto the monolithic columns

For preparation of poly-based (GMA-co-EDMA)-based M-IPG, ammonia (1 mol/L) was

pressed through to react with the epoxy groups for 4 hours. A solution of 5% glutaraldehyde was

pumped in and the column was kept at room temperature for 12 hours. Both ends of the capillary

without polymer were cut away. The capillary was filled with 10% pharmalyte. The two ends of

capillary were immersed into cathode and anode electrolytes. 100 mmol· L-1 phosphoric acid and

sodium hydroxide were used as anolyte and catholyte, respectively. 1500 V was applied on to the

column for the first 3 minutes, and then followed by 3000 V 27 minutes. The capillary was kept

at room temperature for another 12 hours after the IEF. A solution of NaBH4 (20 mmol/L in 100

mmol/L tris-HCl, pH 8.1) was pumped through the column for 9 hours. Finally the capillary was

washed with methanol for 10 minutes and then with H2SO4 (10 mmol/L) for 2 hours followed by

65

Page 80: Development of Isoelectric Focusing Techniques for Protein

flushing with water for 10 minutes. After these treatments, the capillary could be used in CIEF.

For preparation of polyacrylamide-based M-IPG, the capillary column was directly filled

with 10% pharmalyte. A 1500 voltage was applied on to the column for the first 3 minutes, and a

3000 voltage was applied and kept for the rest 27 minutes. The current of the whole process was

recorded by the instrument as shown in Figure 4.3.

The capillary was kept at room temperature for another 12 hours after the immobilization.

After immobilization, the capillary was washed with water for 10 minutes.

4.2.4 Coupling M-IPG-based CIEF with MS

β-Lactoglobulin and myoglobin were selected as model analytes. β-Lactoglobulin and

myoglobin with 40 μg/mL concentration were mixed in water solution. The mixture was injected

into monolithic column by the syringe and the injection lasted for 3 minutes to fill the monolithic

column. After sample introduction, the column was inserted into the electrolyte reservoirs, which

are similar to the iCE 280 separation cartridge. The end with higher pH immobilized was dipped

into the cathode while the end with low pH immobilized was dipped into the anode. 100 mmol·

L-1 phosphoric acid and sodium hydroxide were used as anolyte and catholyte, respectively. The

voltage used for focusing was 500 V for the first 2 minutes and 3000 V for the next 8 minutes.

The voltage was provided by iCE 280 instrument.

After the focusing in the iCE280 instrument, the electric potential was turned off and the

cartridge was taken out. One connecting capillary next to the anode was connected to a syringe

pump using capillary adapter and sleeves at a flow rate of 167 nL/min, while the capillary next to

the cathode was connected to one end of the microcross union using capillary adapter and

sleeves. The makeup solution, consisting of 50% methanol, 49% water, and 1% acetic acid

66

Page 81: Development of Isoelectric Focusing Techniques for Protein

(v/v/v), was introduced from the other end by capillary at a flow rate of 167 nL/min using a

syringe pump. The mixture was pushed to the spray tip for ESI. MS can record the spectra of the

whole process as shown in figure 4.5

Figure 4.1 Schematic connection of coupling M-IPG with MS.

4.3 Results and discussion

4.3.1 Preparation and characterization of M-IPG

Monolith has been widely used in various scientific and industrial fields such as

supporting of catalysts, enzyme immobilization, molecular imprinting, and separation supports.7

In this project, two kinds of monolithic columns were prepared to accomplish IPG in the form of

monoliths. Figure 4.2 and 4.3 showed the SEM images of the prepared poly-based (GMA-co-

EDMA) and polyacrylamide-based monolithic capillary columns. As shown in Figure 4.2A and

Figure 4.3A, the organic polymer monolith was well immobilized onto the inner wall of capillary,

which is helpful when run several times. The macro pores with diameters of several micrometers

67

Page 82: Development of Isoelectric Focusing Techniques for Protein

can be clearly observed in Figure 4.2 B and Figure 4.3 B. Additionally; some micropores with

diameters of several hundreds of nonometers are seen. The large surface area can be used for

further derivatization. Thus, the monolithic columns were used to immobilize the CAs.

BA

Figure 4.2: SEM imaging of poly (GMA-co-EDMA) based monolithic capillary (×1000, ×3000)

A B

Figure 4.3: SEM imaging of poly-based (GMA-co-acrylamide) monolithic capillary (×1200, ×2500)

4.3.2 Immobilization of CAs

As mentioned in the method of immobilization of CAs onto the monolithic columns, a

1500 voltage was applied on to the column for the first 3 minutes, a 3000 voltage was applied

and kept on for the rest of the 27 minutes. Figure 4.4 recorded the current profile during the

68

Page 83: Development of Isoelectric Focusing Techniques for Protein

whole immobilization process. The dropping down of current indicated the CAs gradually

moved to the points, which equal to their pIs.

0 5 1 0 1 5 2 0 2 5 3 0

4

6

8

1 0

1 2

1 4

1 6

1 8

2 0

Cur

rent

/uA

T i m e / m i n

Figure 4.4: Current vs Time during the immobilization procedure. The capillary was filled with

10% pharmalyte. The two ends of capillary were immersed into anode and cathode electrolytes.

100 mmol· L-1 phosphoric acid and sodium hydroxide were used as anolyte and catholyte,

respectively. A 1500 V was applied on to the column for the first 3 minutes, followed by 3000 V

for 27 minutes. The power supply is provided by iCE280.

4.3.3 Coupling M-IPG with MS

β-Lactoglobulin and myoglobin were selected as model analytes. The monolithic

capillary was filled with a mixture of β-lactoglobulin and myoglobin water solution. The

concentration of β-lactoglobulin and myoglobin was 40 μg/mL. After filling, the end of low pH

was dipped into the anode reservoirs; the other end was dipped into the cathode reservoirs. A

3000 V was applied for 10 minutes. After focusing, the monolithic capillary was taken out. The

69

Page 84: Development of Isoelectric Focusing Techniques for Protein

anode end of monolithic was connected to syringe pump using capillary adapter and sleeves, the

cathode end was connected to one end of the microcross union using capillary adapter and

sleeves. The makeup solution (methanol, water, and acetic acid) was introduced from the other

end by capillary and syringe pump. The mixture was pushed to the spray tip for ESI. MS can

record the spectra of the whole process. From the total ion current spectrum, any time point and

any time range can be chosen to check the mass spectrum of analyte at that time point or check

the average mass spectrum within that time range.

TIC of +Q1: from Sample 11 (focus 51) of 20080722.wiff (Ion Spray) Max. 5.9e9 cps.

4.0 4.2 4.4 4.6 4.8 5.0 5.2 5.4 5.6 5.8 6.0 6.2 6.4 6.6 6.8 7.0 7.2 7.4 7.6Time, min

0.0

5.0e8

1.0e9

1.5e9

2.0e9

2.5e9

3.0e9

3.5e9

4.0e9

4.5e9

5.0e9

5.5e9

5.9e9

A B C

Figure 4.5: MS TIC of samples after the separation in CIEF (M-IPG).

70

Page 85: Development of Isoelectric Focusing Techniques for Protein

+Q1: 4.608 to 5.309 min from Sample 11 (focus 51) of 20080722.wiff (Ion Spray) Max. 2.2e6 cps.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu

0.0

2.0e5

4.0e5

6.0e5

8.0e5

1.0e6

1.2e6

1.4e6

1.6e6

1.8e6

2.0e6

2.2e6998.1

942.71060.6

1131.2893.2 1212.0

573.5 1304.9848.7

1063.0616.4 808.3 945.1

771.6595.5

1413.7738.2 1005.6 1219.0898.6853.0 1137.8538.5 1308.3694.0 772.6953.9

1011.7903.7679.2 1226.0806.9 956.3 1417.11144.4611.2 884.3525.1 823.3 1078.8727.1 1320.11232.9711.2 961.6 1151.0 1425.3581.0562.8 982.8656.3 939.6670.9 1235.31033.2 1335.1927.2 1163.8 1438.1

1296.8 1368.5 1408.6

Figure 4.6: Average mass spectrum of time range A.

+Q1: 6.010 to 6.812 min from Sample 11 (focus 51) of 20080722.wiff (Ion Spray) Max. 6.2e6 cps.

500 600 700 800 900 1000 1100 1200 1300 1400 1500m/z, amu

0.0

5.0e5

1.0e6

1.5e6

2.0e6

2.5e6

3.0e6

3.5e6

4.0e6

4.5e6

5.0e6

5.5e6

6.0e6616.2

1231.5614.4893.1 942.7848.7714.4 1060.4

1131.1900.6 1211.8771.6 1306.5 1413.8956.6738.2 841.61143.51051.6573.5 638.3 1218.7948.4 1329.5743.6 860.5 1416.6838.4 997.0730.6 1164.0546.9 1284.01205.4646.4 1486.3

Figure 4.7: Average mass spectrum of time range B.

71

Page 86: Development of Isoelectric Focusing Techniques for Protein

+Q1: 7.212 to 7.563 min from Sample 11 (focus 51) of 20080722.wiff (Ion Spray) Max. 3.5e6 cps.

1100 1150 1200 1250 1300 1350 1400m/z, amu

0.0

5.0e4

1.0e5

1.5e5

2.0e5

2.5e5

3.0e5

3.5e5

4.0e5

4.5e5

5.0e5

5.5e5

6.0e5

6.5e5 1225.3

1219.4 1312.81231.4

1306.6

1143.7

1131.11413.51142.9

1211.7 1407.01076.41304.9

1246.91229.4 1313.91154.71145.8 1308.21234.61082.7 1182.8 1330.01227.61320.01152.5 1262.71094.6 1189.9 1415.11328.71255.01120.6 1370.91209.9 1288.31177.1 1410.71358.31272.21207.8 1393.51282.6

1405.9

Figure 4.8: Average mass spectrum of time range C.

Figure 4.5 indicated the MS TIC of samples after the IEF separation of β-lactoglobulin

and myoglobin. The mass spectrum should be stable, however, in this case, it was not stable. It

might be because of the adapters that connected the syringe pump, the separation capillary, and

the spray tip. The adapter may not prevent air bubbles from being trapped in the capillary, which

may lead to the instability of total ion current. It could be seen from Figure 4.6 that the MS

fingerprints of myoglobin were the sole emergents, which indicated that myoglobin was focused

in stage A; similarly, the β-lactoglobulin was focused in stage C. These results prove that β-

lactoglobulin and myoglobin could be successfully focused under the electrical field in the

72

Page 87: Development of Isoelectric Focusing Techniques for Protein

prepared monolithic IPG column. However, both the fingerprints of β-lactoglobulin and

myoglobin emerged in stage B at the same time as shown in Figure 4.7, which indicated that

some β-lactoglobulin and myoglobin are still mixed in stage B. It might be due to air bubbles

trapped in the adapter or the dead volume in the connection capillary, which destroyed the

separation performance, resulting in the focused zone mixed a little.

It is anticipated that the separation efficiency could be improved if connectors could be

improved.

4.4 Conclusions

Two kinds of monolithic columns were prepared and two methods were used to

immobilize a pH gradient into two kinds of monolithic capillary column by attaching CAs

(Phamalytes) to porous polymer monolith. A poly-based (GMA-co-EDMA) monolithic column

did not produce good results. Isoelectric focusing of two proteins was only carried out in

polyacrylamide-based M-IPG monolithic capillary column without additional CAs. MS was

coupled to the M-IPG as the detector of CIEF. No typical noise of CAs was observed in the MS

spectrum.

73

Page 88: Development of Isoelectric Focusing Techniques for Protein

74

References

(1) Huang, T. M.; Pawliszyn, J. Electrophoresis 2002, 23, 3504-3510.

(2) Fang, X. H.; Adams, M.; Pawliszyn, J. Analyst 1999, 124, 335-341.

(3) Pospichal, J.; Chmelik, J.; Deml, M. J. Microcolumn Sep.1995, 7, 213-219.

(4) Budilova, J.; Pazourek, J.; Krasensky, P.; Pospichal, J. J. Sep. Sci. 2006, 29, 1613-1621.

(5) Hochstrasser, D., Augsburger, V., Funk, M., Appel, R., Pellegrini, C., Muller, A.F.

Electrophoresis 1986, 7, 505-511.

(6) Yang, C.; Zhu, G. J.; Zhang, L. H.; Zhang, W. B.; Zhang, Y. K. Electrophoresis 2004, 25,

1729-1734.

(7) Wu, R.; Hu, L.; Wang, F.; Ye, M.; Zou, H. J. Chromatogr. A 2008, 1184, 369-392.