1
Development of High-Precision Gas Analyzers for Measurements of N 2 O, CO, CH 4 , CO 2 , and H 2 O D. Baer 1 , R. Provencal 1 , A. Crotwell 2 , E. Dlugokencky 2 , P. Novelli 2 , T. Owano 1 , M. Gupta 1 and D. Baer 1 1 Los Gatos Research, Mountain View, CA 94041; 650-965-7772, E-mail: [email protected] 2 NOAA Earth System Research Laboratory, Boulder, CO 80305 We report the continued development of two continuous-flow gas analyzers, based on cavity enhanced laser absorption spectroscopy, which measure important atmospheric constituents in real time, with high precision and accuracy, and over a wide range of mixing ratios. One instrument (N 2 O+CO Analyzer) uses a tunable continuous-wave quantum cascade laser operating in the wavelength region near 4.56 microns to record measurements of carbon monoxide, nitrous oxide and water vapor in ambient air. The typical precision for continuous measurements at typical ambient levels of either CO or N 2 O is about 0.1 ppbv (1-sigma, 10-second measurement time). The instrument also reports water vapor mixing ratio measurements simultaneously and employs those measurements to determine the effects due to dilution and line broadening to allow accurate determination of mixing ratios (CO, N 2 O) on a dry basis. The other instrument uses two tunable continuous-wave near-infrared telecommunications-grade diode lasers operating in the wavelength regions near 1.6 and 1.65 microns to record measurements of carbon dioxide and methane in ambient air. This instrument employs an internal thermal control system to allow extremely low drift despite changes in ambient temperatures over the range from 0 to 40 degrees C. The instrument also reports water vapor mixing ratio measurements simultaneously and employs those measurements to accurately quantify the effects due to dilution and line broadening and provide accurate determination of the measured mixing ratios (CO 2 , CH 4 ) on a dry basis. Both of these instruments require relatively low-power (150-250 watts) and no liquid nitrogen thus allowing for easy measurements in the field. Figure 1. Thermally-stabilized gas analyzer provides simultaneous measurements of methane, carbon dioxide and water vapor.

Development of High-Precision Gas Analyzers for

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Development of High-Precision Gas Analyzers for Measurements of N2O, CO, CH4, CO2, and H2O

D. Baer1, R. Provencal1, A. Crotwell2, E. Dlugokencky2, P. Novelli2, T. Owano1, M. Gupta1 and D. Baer1

1Los Gatos Research, Mountain View, CA 94041; 650-965-7772, E-mail: [email protected] Earth System Research Laboratory, Boulder, CO 80305

We report the continued development of two continuous-flow gas analyzers, based on cavity enhanced laserabsorption spectroscopy, which measure important atmospheric constituents in real time, with high precisionand accuracy, and over a wide range of mixing ratios. One instrument (N2O+CO Analyzer) uses a tunablecontinuous-wave quantum cascade laser operating in the wavelength region near 4.56 microns to recordmeasurements of carbon monoxide, nitrous oxide and water vapor in ambient air. The typical precision forcontinuous measurements at typical ambient levels of either CO or N2O is about 0.1 ppbv (1-sigma, 10-secondmeasurement time). The instrument also reports water vapor mixing ratio measurements simultaneously andemploys those measurements to determine the effects due to dilution and line broadening to allow accuratedetermination of mixing ratios (CO, N2O) on a dry basis.

The other instrument uses two tunable continuous-wave near-infrared telecommunications-grade diode lasersoperating in the wavelength regions near 1.6 and 1.65 microns to record measurements of carbon dioxide andmethane in ambient air. This instrument employs an internal thermal control system to allow extremely lowdrift despite changes in ambient temperatures over the range from 0 to 40 degrees C. The instrument alsoreports water vapor mixing ratio measurements simultaneously and employs those measurements to accuratelyquantify the effects due to dilution and line broadening and provide accurate determination of the measuredmixing ratios (CO2, CH4) on a dry basis. Both of these instruments require relatively low-power (150-250 watts)and no liquid nitrogen thus allowing for easy measurements in the field.

Figure 1. Thermally-stabilized gasanalyzer provides simultaneousmeasurements of methane, carbondioxide and water vapor.